1 Asked Questions 3M Sterilization Assurance Techline by Sandra Velte, BA, CSPDT FAQFrequently Objectives After completion of this self-study activity, the learner will be able to: 1. Develop a policy and procedure for monitoring implants based on recommended practices from the Association for the Advancement of Medical Instrumentation (AAMI) and the Association of perioperative Registered Nurses (AORN) and guidance from the Centers for Disease Control and Prevention (CDC). 2. Explain the purpose of running a positive biological indicator control. 3. Identify the information related to the steam sterilization process that should be documented and maintained in a record keeping system. 4. Comply with the Environmental Protection Agency s (EPA s) final rule on ethylene oxide (EO) emissions from hospital sterilizers. Many thanks to the team at 3M Health Care for working with Managing Infection Control to provide the following accredited course. IAHCSMM has awarded one (1) contact point for completion of this continuing education lesson toward IAHCSMM recertification. The CBSPD has preapproved this inservice for one (1) contact hour for a period of five (5) years from the date of publication, and to be used only once in a recertification period. This inservice is 3M Health Care Provider approved by the California Board of Registered Nurses, CEP 5770 for one (1) contact hour. This form is valid up to five (5) years from the date of publication. Instructions for submitting results are on page 86. Managing Infection Control and 3M Health Care will be working collaboratively to provide continuing education courses in monthly editions of Managing Infection Control. Test Questions Circle the correct answer. 1. In the event of a positive biological indicator (BI) result, all loads back to the last negative BI must be recalled. 2. Physical monitors are recommended for all steam sterilizers, including table-top sterilizers in clinics. 3. A positive control BI should be incubated with every load that is tested with a BI. 4. AAMI ST79 recommends the use of Class 6 chemical indicators (CIs) for routine release of implants. 5. The Food and Drug Administration (FDA) regulates sterilization monitoring products and the companies that manufacture and sell them, but not the healthcare facilities that use them. 6. It is recommended to place peel pouches flat on the sterilizer rack. 7. AORN recommends placement of two internal CIs in opposite corners of each level of multi-level rigid containers. 8. The reason for flash sterilization should be documented for each flash cycle. 68 MANAGING INFECTION CONTROL

2 9. Sterilization records are legal documents that should be stored for seven years. 10. The new EPA final rule on EO emissions from hospital sterilizers requires hospitals to adopt a management practice of running full loads. An acceptable alternative to this management practice is the use of an air pollution control device, such as an abator. Introduction As a member of the 3M Sterilization Technical Services Team, my job responsibilities include answering the 3M Sterilization Techline. In my career, I have responded to thousands of telephone and inquiries from healthcare professionals from across the country. These inquiries are not limited to questions about 3M Sterilization Assurance products, but also include requests for information related to sterilization standards and recommended practices. Therefore, it is important that I am familiar with the most current industry standards, such as the updated AAMI standards that were published at the end of 2008: 1. ANSI/AAMI ST79:2006/A1:2008 Comprehensive guide to steam sterilization and sterility assurance in health care facilities and 2. ANSI/AAMI ST41:2008 Ethylene oxide sterilization in health care facilities: Safety and effectiveness. Staying up-to-date with the most current industry standards, plus being well-informed about new product introductions to the marketplace, can be a challenge for sterile processing professionals and a wide variety of questions are asked by callers to the 3M Sterilization Techline. Although I am continually asked questions I have never heard before, there are several topics that seem to come up over and over again. Answers to the following 10 frequently asked questions are discussed in this self-study lesson. 1. What is the recommended frequency for routine testing with a biological indicator? 2. How should I monitor a table-top steam sterilizer? 3. How often do I need to incubate a positive control biological indicator? 4. Can I use a Class 5 Integrating Indicator or Class 6 Emulating Indicator to release implants? 5. Does the FDA regulate how sterilization monitoring products are used in healthcare facilities? 6. What is the recommended procedure for loading peel pouches in the sterilizer? 7. How many chemical indicators should be placed inside multi-level trays? 8. What information should be documented for each steam sterilization cycle? 9. How long should sterilization records be retained? 10. What do I need to do to comply with the new regulation from the EPA related to EO sterilization? What is the recommended frequency for routine testing with a biological indicator? For steam sterilization, routine sterilizer efficacy testing with a biological indicator process challenge device (BI PCD) should be done weekly, preferably each day the sterilizer is used. If a sterilizer is designed to be used for multiple types of cycles (e.g., 270 F gravity flash, 270 F vacuum-assisted, 250 F gravity, etc.), then each sterilization cycle type used at a facility should be tested. Additionally, all loads containing implants should also be monitored with a BI PCD. In section Using biological indicators, of AAMI ST79 it states: Biological indicators should be used within PCDs (see , , , ) for routine sterilizer efficacy monitoring at least weekly, but preferably every day the sterilizer is in use (see 10.7). Additionally, BIs within PCDs should be used to monitor every load containing implants (see ). If a sterilizer is designed to be used for multiple types of cycles (gravity-displacement at 132 C to 135 C [270 F to 275 F], gravity-displacement at 121 C [250 F], dynamic-air-removal at 132 C to 135 C [270 F to 275 F], flash at 132 C to 135 C [270 F to 275 F], flash with single wrapper or other packaging), then each sterilization cycle type used should be tested. 1 NOTE If a sterilizer will run the same type of cycle (e.g., dynamic-air-removal at 132 C to 135 C [270 F to 275 F] for different exposure times (e.g., 4 minutes and 10 minutes), then only the shortest cycle time needs to be tested. 1 Additionally, each type of tray configuration used for flash sterilization should be routinely tested with a BI PCD or BI challenge test tray. In section Composition of the PCD (BI challenge test tray) under Routine biological monitoring of flash sterilization cycles, of AAMI ST79 it states: A representative of the same type of tray to be routinely processed through the flash sterilizer should be selected to serve as the PCD (BI challenge test tray). Each type of tray configuration in routine use for flash sterilization should be tested separately MANAGING INFECTION CONTROL

3 The recommended frequency for routine testing with BI PCDs for all sterilization methods are summarized in the table below. Note that the recommendation to monitor implant loads with a BI PCD and quarantine until the BI result is known is universal for all methods of sterilization used to process implantable devices. Table 1. Frequency of Routine Testing with Biological Indicators Sterilization Method Steam Frequency Weekly, preferably daily, and with each load containing implantable devices which should be quarantined until the BI results are available. 1,2 of weekly, preferably daily, for BI monitoring of steam sterilizers and monitor every load with a BI PCD. Reasons why facilities adopt an every load monitoring program include the following, to: provide the same standard of care for all patients; improve patient safety; ensure all implants are monitored with a BI; ensure all cycle types used are tested with a BI; ensure all types of packaging used in flash sterilization are tested with a BI; reduce the risk and cost of a recall; reduce the risk and cost of healthcare-associated infections (HAIs); and reduce variability and the chance for errors. Ethylene oxide Every load, including each load containing implantable devices which should be quarantined until the BI results are available. 2,3 To achieve the highest level of sterility assurance, monitor every load with a BI PCD. Low temperature hydrogen peroxide gas plasma Ozone Liquid peracetic acid Dry-heat Daily, preferably each load, and with each load containing implantable devices which should be quarantined until the BI results are available. 2,4 Daily, preferably each load, and with each load containing implantable devices which should be quarantined until the BI results are available. 2,4 Daily 2 Weekly, preferably daily, and with each load containing implantable devices which should be quarantined until the BI results are available. 2 In the event of a positive BI result, all items processed in that sterilizer since the last negative BI result should be recalled and reprocessed. In section General procedure, of AAMI ST79 it states: The following actions should be taken if a BI tests positive: b) If it is determined that the sterilization failure was not the result of operator error (e.g., selection of the incorrect cycle), items processed in that sterilizer since the last negative BI result should be considered unsterile. They should be retrieved, if possible, and reprocessed (see 10.11). The sterilizer in question should be taken out of service. 1 To ensure the highest level of sterility assurance, many facilities go above and beyond the recommended frequency How should I monitor a table-top steam sterilizer? Many healthcare professionals in office-based settings believe that they are not subject to the same standards and guidelines as hospitals and surgery centers. This is not true. AAMI ST79 guidelines pertain to all healthcare facilities that perform steam sterilization including clinics and dental offices that use a table-top sterilizer. ST79 is a consolidation of five previous standards including ANSI/AAMI ST42, Steam sterilization and sterility assurance using table-top sterilizers in office-based, ambulatory-care medical, surgical, and dental facilities. Physical monitors such as recorders, displays, digital printouts and gauges that provide real-time measurements of time, temperature and pressure are a critical component of a steam sterilization monitoring program. After each cycle, a trained and knowledgeable operator should read and initial the physical monitor to verify that all cycle parameters were met. Some older model table-top sterilizers do not have physical monitors. According to section of AAMI ST79, Sterilizers that do not have recording devices should not be used. 1 Like other types of steam sterilizers, table-top sterilizers should be monitored routinely with a BI PCD weekly, preferably every day the sterilizer is used. The BI testing of a table-top sterilizer is conducted in a fully loaded chamber. The BI PCD should be representative of the load contents and placed in the cold point or area least favorable to steam sterilization. This area is typically the center of the load toward the front of the chamber but varies with sterilizer design; therefore, the manufacturer of the sterilizer should 72 MANAGING INFECTION CONTROL

4 be consulted about placement of the BI PCD. All different cycle types used should be tested. For example, if a table-top sterilizer is used to run cycles at both 250 F and 270 F, then both of these cycles should be routinely tested with a BI PCD. There are no commercially available BI PCDs designed for table-top sterilizers so the user must make his or her own BI PCD that represents the most challenging package configuration in the load. The BI PCD should also contain one or more CIs and items normally present during routine sterilization. For example, if routine loads contain single instruments in peel pouches then the appropriate BI PCD is a BI and a CI along with a representative instrument in a peel pouch. If wrapped sets are also routinely processed, then a dummy wrapped set that contains a BI and one or more CIs and representative instruments should be used as the BI PCD. An internal CI should be placed inside each package and an external CI should be placed on the outside of each item, unless the internal CI is visible. External CIs are used to distinguish items that have been processed through the sterilizer from items that have not been processed yet. Internal chemical indicators provide information about whether or not the steam penetrated inside the package to the location of the instruments. The results of internal chemical indicators should be interpreted by trained and knowledgeable healthcare professionals at the point of use before the items are used for patient care. As with all other types of steam sterilizers, qualification testing of table-top sterilizers should be conducted after sterilizer installation, relocation, major repairs and malfunctions, and after sterilization process failures. For qualification testing of a table-top sterilizer, a representative BI PCD should be run in three consecutive full loads and the load items should be quarantined until the BI results are negative. A table-top sterilizer should be routinely tested with a representative BI PCD run in a full load in all types of cycles used. How often do I need to incubate a positive control biological indicator? First let s discuss why it is important to run a positive control biological indicator (BI). It is recommended by AAMI and AORN as well as BI manufacturers. Incubating a positive control BI for a visual color-change result ensures that: the temperature of the incubator or Auto-reader is correct; the viability of the spores has not been altered due to improper storage; temperature, humidity, or proximity to chemicals; and the culture media is capable of supporting growth. A positive control is a BI that is not sterilized; therefore, the spores should be viable. When the control yields a positive result, you are assured that you are challenging your sterilizer with viable spores and that your system is able to detect them. This is important because you want to be sure that you are able to detect any spores that survive the sterilization process. Running a positive control BI is good science it is the quality assurance step that ensures the test BI results throughout the day are accurate. AAMI ST79 recommends that you incubate a positive control every day a test BI is run. If several test BIs from the same lot are run on the same day, only one control BI from that lot needs be incubated so incubating a control for every load that is tested with a BI is not necessary. The lot number of the control must match the lot number of the test BIs. It is important that the control and test BIs have experienced the same shipping and storage conditions. If a new lot of BIs is opened during the day, a new positive control should be incubated. Keep in mind that if you run two types of BIs and have several incubators or Auto-readers you need to run a positive control for each BI type in each incubator or Auto-reader used that day. Each day a sterilizer is tested with a BI, a positive control BI from the same lot number should be incubated. Can I use a Class 5 Integrating Indicator or Class 6 Emulating Indicator to release implants? Facility polices and procedures should be based on standards and recommended practices from organizations such as AORN and AAMI and guidance from the CDC. Let s review what these recommended practices say about monitoring implants. AORN recommends that implantable devices be monitored with a biological indicator and quarantined until the BI result is negative. each load containing an implantable device should be monitored with a BI and quarantined until the results of the BI testing are available 2 AAMI ST79 recommends that implant loads be monitored with a BI PCD that also contains a Class 5 integrating indicator, 74 MANAGING INFECTION CONTROL

5 and that the load should be quarantined until the BI results are known. In section Process monitoring devices, of AAMI ST79 it states: Every sterilization load containing implants should be monitored with a PCD containing a BI (a BI challenge test pack). A Class 5 integrating CI should be included in this PCD. 1 In section Release criteria for implants, of AAMI ST79 it states: The load should be quarantined until the results of the BI testing are available (CDC, 2003a). 1 In the 2008 CDC Guideline for Disinfection and Sterilization in Healthcare Facilities, it states: Use biologic indicators for every load containing implantable items and quarantine items, whenever possible, until the biologic indicator is negative. 5 The reason that only biological indicators can be used to release implants is because BIs provide the only direct measure of lethality of a sterilization cycle. Biological indicators are the monitoring tool that provides the greatest challenge to the sterilization process because they contain a large population of spores that are highly resistant to steam sterilization. If the cycle is able to kill all the spores in the biological indicator, then you have the highest assurance that the bioburden on the items in the load was also killed. In section General considerations, of AAMI ST79 it states: Rationale: The use of BIs provides evidence of efficacy by challenging the sterilizer with a large number of highly resistant bacterial spores. Biological monitoring provides the only direct measure of the lethality of a sterilization cycle. 1 If it is not possible to quarantine the implant until the biological indicator is negative, because of a documented medical emergency, then the implant can be prematurely released on the result of the Class 5 CI in the BI PCD provided that all the other monitoring results were acceptable. In Annex L of ST79 two forms are provided an implantable devices load record to document all implant loads and an exception form to document premature release of implants before the BI results are known. The exception form documents: the name of the patient, the name of the implant, the name of the surgeon, the reason the implant was released early, and what could have prevented the premature release. ST79 clearly states that premature release of implants is unacceptable and should be the exception, not the rule. In the event of an emergency requiring flash sterilization of an implant, AORN recommends the use of a rapid-action BI and quarantining of the implant until the BI result is known. In an emergency, when flash sterilization of an implant is unavoidable, a rapid-action BI and a Class 5 chemical integrating indicator (or enzyme-only indicator) should be run with the load. The implant should be quarantined on the back table and should not be released until the rapid-action BI provides a negative result. 2 (Note: The enzyme-only indicator product has been discontinued and is no longer available.) Class 6 emulating indicators are not recommended to be used as a substitute for biological indicators for routine sterilizer testing, qualification testing, or load release of implants. AAMI ST79 was updated at the end of 2008 and the updated document contains no information about Class 6 emulating indicators. Each healthcare facility develops policies and procedures based on laws and regulations, current scientific knowledge, and accepted practice guidelines, and these policies and procedures should be consistent throughout the healthcare facility. It is necessary for facilities to follow written policies and procedures so they can obtain Joint Commission accreditation. An easy-to-use sterilization process monitoring program should follow the most stringent recommended practices, standards and guidelines set forth by organizations such as AAMI, AORN, and the CDC. Implant loads should be monitored with a BI PCD that contains a Class 5 integrating indicator and quarantined until the BI result is known. Does the FDA regulate how sterilization monitoring products are used in healthcare facilities? Sterilization monitoring products, defined by the FDA as Class II medical devices, are cleared for sale. The FDA s role is to assure that products sold in the United States are safe and effective when the manufacturer s instructions are followed. The FDA regulates sterilization monitoring products and the companies who make and sell them, but not the healthcare professionals who use them. Therefore, the FDA clearance process is limited to assuring safety and efficacy, and does not offer guidance on how sterilization monitoring products should be used. The specific use and frequency-of-use of sterilization monitoring products, such as biological and chemical indicators, are determined by healthcare professionals, such as sterile processing and operating room personnel, at each healthcare facility. 76 MANAGING INFECTION CONTROL

6 Healthcare professionals establish policies and procedures that describe the use and frequency-of-use for the sterilization monitoring products used at their facility. Written policies and procedures should be based on industry standards, guidelines, and recommended practices, such as those published by AAMI, AORN and the CDC. Electing to develop policies and procedures that deviate from standards or recommended practices should be approached with caution and should include evaluation of relevant scientific data, consideration of potential legal issues, and involvement of all stakeholders throughout the facility (e.g., infection control, risk management, surgeons). Healthcare facilities establish policies and procedures based on industry standards, guidelines, and recommended practices from AAMI, AORN and the CDC. What is the recommended procedure for loading peel pouches in the sterilizer? Paper-plastic pouches should be placed on edge in the sterilizer and oriented so that the paper side of one pouch faces the plastic side of the pouch next to it (see Figure 1). This ensures adequate steam penetration and air removal from each pouch. Only small, lightweight items should be packaged in peel pouches because heavy metal instruments can cause problems with drying and sterility maintenance (i.e., heavy objects can cause the package to tear). In section Paper-plastic pouches, of AAMI ST79 it states: Paper-plastic pouches should stand on edge in relation to the cart or shelf, with the paper side of one pouch next to the plastic side of the next pouch; holding racks or baskets specifically designed for pouches can be used. 1 Additionally, AORN recommends that: Paper-plastic pouches should not be used within wrapped sets or containment devices because the pouches cannot be positioned to ensure adequate air removal, sterilant contact, and drying. 5 Due to the risk of inadequate air removal and steam penetration, it is not acceptable to use paper-plastic pouches inside wrapped trays or containers even if the pouch is not sealed. For the purpose of confining small items within wrapped or containerized sets, a packaging material that allows adequate air removal and steam penetration, such as a paper-paper pouch or wire mesh basket, should be used. It is always important to confirm that packaging materials are being used in accordance with the manufacturer s instructions and are FDA cleared for that usage. Double packaging with paper-plastic pouches is acceptable as long as the outside pouch is big enough to accommodate the inside pouch without folding and the packaging manufacture has FDA clearance for double packaging using their product. Folding of the inside pouch can cause sterilization process failures due to inadequate air removal and steam penetration. Both pouches should be sealed and oriented in the same direction (i.e., plastic facing plastic) (see Figure 2). Figure 1. Loading of paper-plastic pouches in sterilizer (placed on edge). Figure 2. Example of single- and double-packaging with paper-plastic pouches. Figure 9 (continued) Loading the sterilizer Reprinted from ANSI/AAMI ST79:2006 and A1:2008 with permission of Association for the Advancement of Medical Instrumentation, Inc. (C) 2008 AAMI All rights reserved. Further reproduction or distribution prohibited. Figure 8 Example of single-and double-packaging with paperplastic pouches Reprinted from ANSI/AAMI ST79:2006 and A1:2008 with permission of Association for the Advancement of Medical Instrumentation, Inc. (C) 2008 AAMI All rights reserved. Further reproduction or distribution prohibited. 78 MANAGING INFECTION CONTROL

7 In section Paper-plastic pouches, of AAMI ST79 it states: If the item is to be double-packaged, two sequentially sized pouches should be used (i.e., the sealed inner pouch should fit inside the other pouch without folding). The pouches should be positioned so that plastic faces plastic and paper faces paper. 1 Figure 3. Example of multi-level rigid container with two chemical indicators in opposite corners of each level. Peel pouches should be placed in a rack or basket to allow them to be standing on edge and oriented so that paper faces paper and plastic faces plastic. How many chemical indicators should be placed inside multi-level trays? AAMI ST79 recommends an internal chemical indicator in each package, tray or rigid container system placed in the area least accessible to steam penetration. In section Internal chemical indicators, of AAMI ST79 it states: The CI should be placed in that area of the package, tray, or containment device (rigid sterilization container system, instrument case, cassette, or organizing tray) considered to be least accessible to steam penetration; for a container system, the manufacturer s instructions for placement of the CI should be consulted. This location might or might not be the center of the package, tray, or containment device. 1 The packaging systems recommended practice of AORN offers the most detailed guidance on the placement of internal chemical indicators. The AORN guidelines for placement of internal chemical indicators are summarized below: The internal chemical indicator/integrator should be placed in the geometric center of the wrapped package, not on top. Two chemical indicators/integrators should be placed inside rigid containers, one in each of two opposite corners of the inside basket on each level (see Figure 3). A chemical indicator/integrator should be placed in the geometric center on each level of multi-level wrapped sets (see Figure 4). 6 Figure 4. Example of multi-level wrapped sets with chemical indicator in geometric center of each level. Two internal chemical indicators should be placed in opposite corners on each level of multi-level rigid containers. A chemical indicator should be placed in the geometric center of each level of multi-level wrapped sets. What information should be documented for each steam sterilization cycle? Each package must be traceable to the load in which it was sterilized. This is usually accomplished by labeling each package in 80 MANAGING INFECTION CONTROL

8 If the sterility of a load is later called into question, it is important that all of the packages from that load can be identified and pulled from use. each load with the sterilization date, the sterilizer number, and the load number. If the sterility of a load is later called into question, it is important that all of the packages from that load can be identified and pulled from use. In section Lot control number, of AAMI ST79 it states: Each item or package intended for use as a sterile product should be labeled with a lot control identifier. The lot control identifier should designate the sterilizer identification number or code, the date of sterilization, and the cycle number (cycle run of the sterilizer). 1 Lot identification enables personnel to retrieve items in the event of a recall and to trace problems (e.g., wet packs) to their source. 1 AAMI recommends that the following information should be recorded and maintained for each sterilization cycle: the lot number; the specific contents of the lot or load, including quantity, department, and a specific description of the items (e.g., towels, type/name of instrument sets); the exposure time and temperature, if not provided on the sterilizer recording chart; the name or initials of the operator; the results of biological testing, if applicable; the results of Bowie-Dick testing, if applicable; the response of the CI placed in the PCD (BI challenge test pack, BI challenge test tray, or CI challenge test pack), if applicable; and any reports of inconclusive or nonresponsive CIs found later in the load. 1 The physical monitors (e.g., digital printout) for each cycle should be dated and signed by the operator and saved in the sterilizer records. Additionally, a log of repairs and preventive maintenance should be maintained for each sterilizer. Sterilizer and load information may be maintained in a paper or electronic record keeping system. In section 9.7 Record-keeping, of AAMI ST79 it states: A maintenance record, in either paper or electronic format, should be kept for each sterilizer The maintenance record should include sufficient information to identify the equipment and to establish a continuous history of all scheduled and unscheduled service. At least the following information should be recorded: the date on which service was requested; the model and serial number of the sterilizer; the location of the equipment (hospital identification, if applicable); the name of the individual from the health care facility who requested and authorized the service; the reason for the service request; a description of the service performed (e.g., calibration, repair); the types and quantities of parts replaced; the name of the person who performed the service; the date the work was completed; the handwritten or electronic signature and title of the person who acknowledged completion of the work; and the results of any post-maintenance testing performed, if needed, before the sterilizer was returned to service. 1 Even though labels are typically not used on flash sterilized items, a lot number should be assigned to each flash sterilization load and a load record should be documented for each flash cycle. According to AORN, sterilization records for each flash load should include the following information: the item(s) processed; the patient receiving the item(s); the cycle parameters used (e.g., temperature, duration of cycle); the date and time the cycle is run; the operator information; and the reason for flash sterilization. 2 AAMI ST79 emphasizes that implantable devices should not be flash sterilized. However, if flash sterilization of implantable devices is unavoidable due to an emergency situation, full traceability to the patient should be maintained. 1 Label each package with the date of sterilization, the sterilizer number, and the cycle number. How long should sterilization records be retained? In section Sterilizer records, of AAMI ST79 it states: 82 MANAGING INFECTION CONTROL

9 The length of time that records must be retained varies throughout the country. Each health care facility is responsible for determining its recordretention policy based on state and local regulations, legal considerations (e.g., statutes of limitation for lawsuits), and its individual situation. Sterilization records should be retained according to the policy and procedure established by the individual health care facility. 1 Check with your risk management department to find out what the record-retention policy is at your facility. Each facility determines their own record-retention policy based on local laws and regulations. Legal considerations pertaining to any special circumstances are also taken into account. For example, a children s hospital may decide to keep their records longer than what is required by law. Each healthcare facility needs to determine its own record-retention policy based on local laws and regulations, legal considerations, and any special circumstances. What do I need to do to comply with the new regulation from the EPA related to EO sterilization? At the end of 2007, the U.S. EPA issued a final rule regulating emissions from hospital EO sterilizers. Under the new final rule, hospitals running EO cycles without an air pollution control device must sterilize full loads of items having common aeration time, except when running a partial load is medically necessary. The EPA recognizes that the practice of running full loads is being done at most facilities already. The use of an air pollution control device (e.g., 3M EO Abator) is an acceptable alternative to running full loads. If your hospital has an EO sterilizer without an air pollution control device, you must: Submit an Initial Notification of Compliance Status letter including: the name and address of your facility, the physical location of the EO sterilizer, a description of the EO sterilization done at your facility (such as typical number of sterilization cycles/year, etc.), and a statement that your facility will adhere to running full loads. Record the date and time of each EO sterilization cycle. Note any EO sterilization cycles that were not full loads, including an acknowledgement from a hospital CS staff member, administrator, or physician that it was medically necessary. If your hospital has an air pollution control device and is subject to state or local regulations, you must: Submit an Initial Notification of Compliance Status letter stating that you are operating the sterilizer in accordance with your state or local regulations and following the control device manufacturer s recommended procedures. If your hospital has an air pollution control device, but is not subject to state or local regulations, you must: Submit an Initial Notification of Compliance Status letter stating that you are venting the EO emissions from each sterilizer to an add-on air pollution control device and that this device is operating during all sterilization processes and in accordance with the manufacturer s recommended procedures. The compliance date for existing sterilizers was December 29, 2008 and upon start-up for new sterilizers. The Initial Notification Compliance Status letter is due 180 days after the compliance date. Use the following link to view the complete standard on the EPA s Web site /December/Day-28/a25233.pdf To be compliant with the EPA final rule on EO emissions, a healthcare facility that uses EO sterilization must use a pollution control device or document that each load was full or medically necessary. Summary Healthcare professionals have many questions related to sterilization ranging from how to conduct a recall after a positive BI result to what is the recommended procedure for monitoring implant loads. Consulting industry standards, recommended practices, and guidelines from AAMI, AORN and the CDC will help you make the best choices for your facility and patients. Use the information in this selfstudy lesson to update your facility s polices and procedures to ensure you are following accepted practice guidelines, laws, and regulations as well as improving patient care. 84 MANAGING INFECTION CONTROL

10 Sterile Process and Distribution CEU Information CEU Applicant Name Address City State Zip Code The CBSPD (Certification Board for Sterile Processing and Distribution) has pre-approved this inservice for one (1) contact hour for a period of five (5) years from the date of publication. Successful completion of the lesson and post test must be documented by facility management and those records maintained by the individuals until recertification is required. DO NOT SEND LESSON OR TEST TO CBSPD. For additional information regarding Certification contact: CBSPD, Inc. 148 Main St., Lebanon, NJ or call or visit the Web site at www. sterileprocessing.org. IAHCSMM has awarded one (1) Contact Point for completion of this continuing education lesson toward IAHCSMM recertification. Nursing CE Application Form This inservice is approved by the California Board of Registered Nurses, CEP 5770 for one (1) contact hour. This form is valid up to five years from the date of publication. 1. Make a photocopy of this form. 2. Print your name, address and daytime phone number and position/title. 3. Add the last 4 digits of your social security number or your nursing license number. 4. Date the application and sign. 5. Answer the true/false CE questions. KEEP A COPY FOR YOUR RECORDS. 6. Submit this form and the answer sheet to: Workhorse Publishing Managing Infection Control PO Box 25310, Scottsdale, AZ For questions or follow-up, contact craig@manageinfection.com. 8. Participants who score at least 70% will receive a certificate of completion within 30 days of Managing Infection Control s receipt of the application. Application Please print or type. Name Mailing Address City, State, Country, Zip Daytime phone ( ) Position/Title Social Security or Nursing License Number References 1. Association for the Advancement of Medical Instrumentation. Comprehensive guide to steam sterilization and sterility assurance in health care facilities. ANSI/AAMI ST79: 2006 and A1:2008 (consolidated text). 2. Association of perioperative Registered Nurses (AORN). Recommended Practices for Sterilization in Perioperative Practice Setting, Association for the Advancement of Medical Instrumentation. Ethylene oxide sterilization in health care facilities: Safety and effectiveness. ANSI/AAMI ST41: Association for the Advancement of Medical Instrumentation. Chemical sterilization and high-level disinfection in health care facilities. ANSI/AAMI ST58: Centers for Disease Control. Guideline for Disinfection and Sterilization in Healthcare Facilities, Association of perioperative Registered Nurses (AORN). Recommended Practices for Selection and Use of Packaging Systems for Sterilization, Sandra Velte, BA in Microbiology, is one of the voices at the 3M Sterilization Techline ( option 2). In this role, she answers technical questions about 3M products, troubleshoots sterilization process failures, and provides information related to sterilization best practices. Ms. Velte is a member of the Minnesota Healthcare Central Service Members Association (MHCSMA) and the International Association of Healthcare Central Service Materiel Management (IAHCSMM) and is a certified Central Sterile Processing and Distribution Technician. Date application submitted Signature Offer expires June 2014 On a scale of 1-5, 5 being Excellent and 1 being Poor, please rate this program for the following: 1) Overall content 2) Met written objectives 3) Usability of content ANSWERS 1. A 2. A 3. B 4. B 5. A 6. B 7. A 8. A 9. B 10. A Copyright 2009/Workhorse Publishing L.L.C./All Rights Reseved. <06/09> Reprint with permission from Workhorse Publishing L.L.C. 86 MANAGING INFECTION CONTROL

3M s Sterilization Tech Line FAQs Sandra Velte, BA, CSPDT

3M s Sterilization Tech Line FAQs Sandra Velte, BA, CSPDT 3M Sterile U Sterilization Assurance Continuing Education 3M s Sterilization Tech Line FAQs Sandra Velte, BA, CSPDT May 2013 Objectives After completion of this self-study activity, the learner will be

More information

Welcome! Thank you for attending today s meeting, it will begin shortly.

Welcome! Thank you for attending today s meeting, it will begin shortly. 3M Sterile U Web Meeting January 20, 2011 Welcome! Thank you for attending today s meeting, it will begin shortly. How do I get a CE Certificate? Next week, all of today s meeting participants will be

More information

Welcome! How do I get a CE Certificate?

Welcome! How do I get a CE Certificate? M Sterile U Network M Sterile U Web Meeting November 0, 0 Today s meeting times: 9:00 a.m., :00 a.m., and :00 p.m. Central Standard Time To hear audio, call 800-97-00 and enter access code 7 Phone lines

More information

What Every Infection Preventionist Should Know About the Critical Issues in the Steam Sterilization Process

What Every Infection Preventionist Should Know About the Critical Issues in the Steam Sterilization Process 3M Attest Sterile U Web Meeting August 20, 2009 What Every Infection Preventionist Should Know About the Critical Issues in the Steam Sterilization Process Updated Aug 2009 by Martha Young Welcome! Facilitator:

More information

2. Describe correct actions to take to address sterilization process failures. 3. List possible causes of sterilization process failures.

2. Describe correct actions to take to address sterilization process failures. 3. List possible causes of sterilization process failures. Objectives 1. Discuss the importance of sterilization process monitoring. 2. Describe correct actions to take to address sterilization process failures. 3. List possible causes of sterilization process

More information

ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment

ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment 3M Attest TM TM Products Sterile U Network T U T O R I A L S ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment Background: A second amendment (A2:2009) to the Association for the Advancement of

More information

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of Contents Glossary of equivalent standards... ix Committee representation... xi Acknowledgments...xiii Foreword... xiv Introduction: Need for the recommended practice...1 1 Scope...5 1.1 General...5 1.2

More information

Specialty Testing of Steam Sterilizers

Specialty Testing of Steam Sterilizers Specialty Testing of Steam Sterilizers Nancy Chobin, RN, AAS, ACSP, CSPM, CFER 2017 Sterile Processing University, LLC. **This in-service has been Approved by the CBSPD for 1.5 CEUs. Objectives 1. To describe

More information

Choose a category. You will be given the answer. You must give the correct question. Click to begin. Instructions for using this template.

Choose a category. You will be given the answer. You must give the correct question. Click to begin. Instructions for using this template. Instructions for using this template. Remember this is Jeopardy, so where I have written Answer this is the prompt the players will see, and where I have Question should be the player s response. To enter

More information

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية 1. POLICY STATEMENT: Sterilization process should be strictly monitored using, physical, chemical and biological indicator and the results of indicators should be supervised by infection control on daily

More information

Unloading, Storage, Distribution, Transportation and Aseptic Presentation of Sterile Items

Unloading, Storage, Distribution, Transportation and Aseptic Presentation of Sterile Items Unloading, Storage, Distribution, Transportation and Aseptic Presentation of Sterile Items Comprehensive guide to steam sterilization and sterility assurance in health care facilities (ANSI/AAMI ST79:2006)

More information

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of Contents Glossary of equivalent standards... ix Committee representation... xi Acknowledgments...xiii Background on Amendment 1...xiii Foreword... xiv Introduction: Need for the recommended practice...1

More information

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing What Dentistry Can Learn from AAMI ST79 Best Practices for device and instrument reprocessing Learning goals: Purpose and goals of AAMI organization. Overview of ST79. Review selected topics: IFUs, tools

More information

Quality Assurance with Rigid Containers Katrina Simpson, M.A., CST, CSPDT

Quality Assurance with Rigid Containers Katrina Simpson, M.A., CST, CSPDT Page 2 I HEARD IT THROUGH THE STEAMLINE Quality Assurance with Rigid Containers Katrina Simpson, M.A., CST, CSPDT Objectives: Discuss various quality assurance issues regarding rigid containers. Discuss

More information

In the Know about.appropriate Use of 3M Attest Rapid Readout Biological Indicators

In the Know about.appropriate Use of 3M Attest Rapid Readout Biological Indicators 3M Attest Sterile U Web Meeting Jan 22, 2009 In the Know about.appropriate Use of 3M Attest Rapid Readout Biological Indicators ww.3m.com/atteststerileuonline 7/15 Modified per Inma s input. SDF Prepared

More information

26/01/2016. A Day in the Life Steam Sterilization Reprocessing. Welcome! House Keeping. From the GoToWebinar page:

26/01/2016. A Day in the Life Steam Sterilization Reprocessing. Welcome! House Keeping. From the GoToWebinar page: SM 3M Health Care Academy A Day in the Life Steam Sterilization Reprocessing January 27, 2016 3M 2016. All Rights Reserved Welcome! Topic: A Day in the Life Steam Sterilization Reprocessing Facilitator:

More information

8/23/2012. Continuing Education Contact Hours. How much of your formal education focused on sterilization or sterile processing?

8/23/2012. Continuing Education Contact Hours. How much of your formal education focused on sterilization or sterile processing? Continuing Education Contact Hours The presenter is an employee of STERIS Corporation and receives no direct compensation other than normal salary for participation in this activity. STERIS Corporation

More information

Steam Sterilization Monitoring in the New CSA Standard (Z )

Steam Sterilization Monitoring in the New CSA Standard (Z ) Speaker Disclaimer Steam Sterilization Monitoring in the New CSA Standard (Z314.3-14) Rick Bauer, Ph.D. Senior Technical Services Advisor 3M Canada Employed in technical services for 3M Canada Education,

More information

Packaging for Sterilization

Packaging for Sterilization Packaging for Sterilization IT S NOT JUST A BOX, A BAG OR GIFT WRAPPING! by Rose Seavey, RN, BS, MBA, CNOR, ACSP Objectives After completion of this self-study activity, the learner will be able to: 1.

More information

Best Practice Monitoring for Sterile Processing Professionals

Best Practice Monitoring for Sterile Processing Professionals Best Practice Monitoring for Sterile Processing Professionals Sterile processing departments are faced with greater challenges than ever. Increasing demands for quick turnarounds. Pressure to streamline

More information

Sterilization Policy. Georgia Regents Medical Center Policy Library. Policy Owner: Epidemiology POLICY STATEMENT

Sterilization Policy. Georgia Regents Medical Center Policy Library. Policy Owner: Epidemiology POLICY STATEMENT POLICY STATEMENT The ability to sterilize instruments and equipment for use during operative or other invasive procedures is critical to promoting successful patient outcomes and preventing infections.

More information

Putting Confidence in Your Hands.

Putting Confidence in Your Hands. Putting Confidence in Your Hands. 3M Sterilization Products and Services Full Line Catalogue 1 Trusted 3M Sterilization Solutions Equipment Monitoring Exposure Monitoring Pack Monitoring Load Monitoring

More information

THE BASICS OF STERILIZATION

THE BASICS OF STERILIZATION THE BASICS OF STERILIZATION Objectives State the importance of sterilization to patient care Review three key essentials of the sterilization process Describe sterilization methods used, application, limitations,

More information

MANAGING INFECTION CONTROL

MANAGING INFECTION CONTROL To Kill or A biological indicator story Not to Kill by Sailaja Chandrapati, BS, MSc, MS, PhD, and Martha Young, BS, MS, CSPDT Objectives 1. Explain the historical evolution of the role of bacterial spores

More information

QUALITY ASSURANCE IN AN MDRD

QUALITY ASSURANCE IN AN MDRD QUALITY ASSURANCE IN AN MDRD MAINTENANCE AND PREVENTATIVE MAINTENANCE MDRD is a machine dependent department washers, pasteurizers, cart washer, sterilizer, ultrasonic, heat sealers, RO water systems All

More information

IT HAS BEEN A GOOD MORNING IN CENTRAL SERVICE (CS), AND THEN

IT HAS BEEN A GOOD MORNING IN CENTRAL SERVICE (CS), AND THEN CRCST Self-Study Lesson Plan Lesson No. CRCST 155 (Technical Continuing Education - TCE) Sponsored by: by Susan Klacik, BS, CRCST, ACE, CIS, FCS, and Linda Breadmont, CRCST, ACE Preventing Wet Packs LEARNING

More information

INSTRUMENT PROCESSING FROM PREP/PACKAGING TO STORAGE. January 2016

INSTRUMENT PROCESSING FROM PREP/PACKAGING TO STORAGE. January 2016 INSTRUMENT PROCESSING FROM PREP/PACKAGING TO STORAGE January 2016 Disclosures The opinions expressed in this presentation are those of the author and do not necessarily reflect the official position of

More information

IT HAS BEEN A GOOD MORNING IN CENTRAL SERVICE (CS), AND THEN

IT HAS BEEN A GOOD MORNING IN CENTRAL SERVICE (CS), AND THEN Lesson No. CRCST 155 (Technical Continuing Education - TCE) Sponsored by: by Susan Klacik, BS, CRCST, ACE, CIS, FCS, and Linda Breadmont, CRCST, ACE Preventing Wet Packs LEARNING OBJECTIVES 1. Learn how

More information

Low Temperature Processing. An Update on Guidelines, Standards and Requirements

Low Temperature Processing. An Update on Guidelines, Standards and Requirements Low Temperature Processing An Update on Guidelines, Standards and Requirements Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve successful completion

More information

Testing of routinely sterilized products should

Testing of routinely sterilized products should Product Quality Assurance Testing: One Hospital s Experience Testing of routinely sterilized products should be part of a healthcare facility s quality control program, and ANSI/AAMI ST79:2006, Comprehensive

More information

Steam Sterilization Validation for Healthcare Institutions

Steam Sterilization Validation for Healthcare Institutions Steam Sterilization Validation for Healthcare Institutions Jonathan A. Wilder, Ph.D. President, H & W Technology, LLC/Stericert Co. Managing Director, Quality Processing Resource Group, LLC 1 Introduction

More information

Medical Device Reprocessing Quality Assurance. 3M All Rights Reserved.

Medical Device Reprocessing Quality Assurance. 3M All Rights Reserved. Medical Device Reprocessing Quality Assurance Additional Content Qualification Testing Flash Sterilization Extended Cycles Recalls Wet Packs Finish 2 Qualification Testing and Required Monitoring Installation

More information

Immediate-Use Steam Sterilization (IUSS) Tuesday, October 4, :00 pm 1:30 pm EDT

Immediate-Use Steam Sterilization (IUSS) Tuesday, October 4, :00 pm 1:30 pm EDT Tuesday, October 4, 2011 12:00 pm 1:30 pm EDT Martha L. Young, BS, MS, CSPDT Martha L. Young, LLC marthalyoung1@aol.com Immediate-Use Steam Sterilization (IUSS) Martha Young Martha Young MS, BS, CSPDT

More information

Slide 1. Slide 2. Slide 3. Sponsored by Webber Training Inc. For more information contact Paul Webber

Slide 1. Slide 2. Slide 3. Sponsored by Webber Training Inc.  For more information contact Paul Webber Slide 1 QUALITY ASSURANCE IN STERILIZATION Susan Hadfield Director, Sterile Processing Department Health Sciences Centre Winnipeg, Manitoba Hosted by Paul Webber (paul@webbertraining.com) A teleclass sponsored

More information

WITH THE ADVENT OF THE EVENT-RELATED STERILITY PROCESS,

WITH THE ADVENT OF THE EVENT-RELATED STERILITY PROCESS, CRCST Self-Study Lesson Plan Lesson No. CRCST 157 (Technical Continuing Education - TCE) Sponsored by: by Deborah Bunn, BS, MS, CRCST, CIS, CHL, ACE, and Cindy Turney Smith, CRCST, CSPDT Expiration Dating

More information

Every Step. Every Detail.

Every Step. Every Detail. 3M Sterilization Products and Services Every Step. Every Detail. Every Day. 1 2 3 4 3M Sterilization: Comprehensive Monitoring Portfolio World Class Technical Support Expert Sales Support 3M Attest Sterile

More information

Loaner Instrumentation

Loaner Instrumentation Loaner Instrumentation Objectives Discuss issues and concerns relative to loaner instrument sets Discuss guidance from a professional organization paper and AAMI Examine the steps to take in managing the

More information

Your Sterile Processing Partner. Duraholder IPS

Your Sterile Processing Partner. Duraholder IPS Your Sterile Processing Partner The Bioseal Duraholder Instrument Protection System (IPS) was designed by a leading university hospital to prevent costly damage to their delicate neuro instruments. After

More information

Your Sterile Processing Partner. Duraholder IPS

Your Sterile Processing Partner. Duraholder IPS Your Sterile Processing Partner Table of Contents Item Page Duraholder Case Study 17 AAMI ST79 Reference 20 Steam Sterilization Efficacy Testing 21 PRODUCT INDEX Product Page Standard Pouches DURA050/100

More information

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar MONITORING THE SYSTEM 1E PROCESSOR Q: I want to confirm that there is no need for a Biological Indicator (BI) and that use

More information

Facility: Sterile Processing Assessment Date: CRITERIA ANSI/AAMI COMPLIANT Number Y N N/A Point of Use Gross contaminant is removed with

Facility: Sterile Processing Assessment Date: CRITERIA ANSI/AAMI COMPLIANT Number Y N N/A Point of Use Gross contaminant is removed with Point of Use 6.2-6.3 Gross contaminant is removed with water Instruments are sorted Sharps removed Instruments placed back in original container Un-used instruments placed in bottom of basket Towel placed

More information

Sterilization Overview from the CDC

Sterilization Overview from the CDC Sterilization Overview from the CDC Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008 New evidence-based topics and recommendations by Cynthia Hubbard RN, BS Objectives Identify

More information

Chemical Sterilants and High-Level Disinfectants: Managing Practical, Safe and Effective Use

Chemical Sterilants and High-Level Disinfectants: Managing Practical, Safe and Effective Use Chemical Sterilants and High-Level Disinfectants: Managing Practical, Safe and Effective Use by Donald P. Satchell, Ph.D. Objectives After completion of this self-study activity, the learner will be able

More information

Technical Data Monograph. VERIFY Assert Self-Contained Biological Indicators and VERIFY Assert STEAM Process Challenge Devices

Technical Data Monograph. VERIFY Assert Self-Contained Biological Indicators and VERIFY Assert STEAM Process Challenge Devices Technical Data Monograph VERIFY Assert Self-Contained Biological Indicators and VERIFY Assert STEAM Process Challenge Devices Table of Contents 1. Introduction....3 2. Product Use Application...4 Device

More information

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016 Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016 Ensuring our patients get clean, sterile instruments every time. Objectives Define the perioperative nurse and surgical tech

More information

McKesson Sterilization Products

McKesson Sterilization Products McKesson Sterilization Products Table of Contents Introduction Introduction... 3 Biological Indicators and Incubators Steam Integrator... 4 Sterilization Biological Indicator Vial...4 Self-Contained Biological

More information

The inaugural use of vaporized

The inaugural use of vaporized Lesson No. CRCST 164 (Technical Continuing Education - TCE) Sponsored by: Vaporized Hydrogen Peroxide Sterilization BY LARRY TALAPA, BS, MS, CQE - MICROBIOLOGIST AND TECHNICAL SERVICE SPECIALIST, 3M HEALTH

More information

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms Learning Objectives Compare and contrast the available methods of sterilization for the healthcare setting Describe the appropriate use and monitoring for sterilization methods Purpose of Sterilization

More information

Sterilization Packaging Systems, Preparation and Loading for Steam Sterilization

Sterilization Packaging Systems, Preparation and Loading for Steam Sterilization Sterilization Packaging Systems, Preparation and Loading for Steam Sterilization by Rose Seavey, RN, MBA, CNOR, ACSP Many thanks to the team at 3M Health Care for working with Managing Infection Contro

More information

Table-top Steam Sterilizers

Table-top Steam Sterilizers ANSI/ADA/AAMI ST55:2010 Approved by ADA Council on Scientific Affairs: November 2013 American National Standard/ American Dental Association/ Association for the Advancement of Medical Instrumentation

More information

Quality Control EQUIPMENT CONTROL. Monitor critical performance characteristics of sterilization, decontamination or cleaning equipment

Quality Control EQUIPMENT CONTROL. Monitor critical performance characteristics of sterilization, decontamination or cleaning equipment STERILITY ASSURANCE Monitor critical performance characteristics of sterilization, decontamination or cleaning equipment VERIFY Equipment Control Products EQUIPMENT CONTROL LOAD CONTROL PROCESS CONTROL

More information

A Sterile Processing User s Guide

A Sterile Processing User s Guide Product Manual A Sterile Processing User s Guide Safety, Assurance, Innovation, & Quality enabling perioperative professionals to achieve higher standards of performance and enhanced patient safety Phone:866-ONE-TRAY

More information

01.48 Receiving and Processing Vendor Owned Surgical Instrumentation in the Sterile Processing Department

01.48 Receiving and Processing Vendor Owned Surgical Instrumentation in the Sterile Processing Department 01.48 Receiving and Processing Vendor Owned Surgical Instrumentation in the Purpose To provide effective management of and ensure standardization of processing for all reusable surgical instruments that

More information

ASTM D10-F02 Workshop FDA Regulatory Perspective. Patrick Weixel FDA Center for Devices and Radiological Health

ASTM D10-F02 Workshop FDA Regulatory Perspective. Patrick Weixel FDA Center for Devices and Radiological Health ASTM D10-F02 Workshop FDA Regulatory Perspective Patrick Weixel FDA Center for Devices and Radiological Health 1 FDA Regulatory Perspective Focus of talk: Use of Standards Auditing Packaging Process Deficiencies

More information

The inaugural use of vaporized

The inaugural use of vaporized Lesson No. CRCST 164 (Technical Continuing Education - TCE) Sponsored by: Vaporized Hydrogen Peroxide Sterilization BY LARRY TALAPA, BS, MS, CQE - MICROBIOLOGIST AND TECHNICAL SERVICE SPECIALIST, 3M HEALTH

More information

Biological Indicators and Their Application in Clinical Sterilization Spores Don t Lie 1

Biological Indicators and Their Application in Clinical Sterilization Spores Don t Lie 1 Biological Indicators and Their Application in Clinical Sterilization 2010 Spores Don t Lie 1 SGM Biotech, Inc. 10 Evergreen Drive, Suite E Bozeman, MT 59715 Tel: 406-585-9535 Fax: 406-585-9219 email:

More information

Packaging & Sterilization

Packaging & Sterilization Packaging & Sterilization Packaging & Sterilization 1 Dr. Norman Clinical Consultant General Sterilization Packaging System in Hospitals Sterilization Packaging System Disposable Reusable Crepe paper Woven

More information

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities Sterilization Practices in Healthcare Facilities Virtual Tour of Central Processing William A. Rutala, Ph.D., M.P.H. Director, Hospital Epidemiology, Occupational Health and Safety Program, UNC Health

More information

Instructions for Validation of Autoclaves

Instructions for Validation of Autoclaves Instructions for Validation of Autoclaves A. Validation of autoclaves may accomplished by the following methods: 1. Autoclave tape applied to the outside of all packs to validate that the correct temperature

More information

Sterile Packaging & The End User

Sterile Packaging & The End User Sterile Packaging & The End User HealthPack Conference March 8-10, 2011 Cincinnati, Ohio Presented by: Ray Taurasi MBA, CRCST, CHL, FCS, ACE director clinical sales and services Healthmark Industries Inc.

More information

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 05/2018 PURPOSE: To test for detection of blood residue on

More information

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue on

More information

Lets Talk Indicators Challenging the process Sterilization. ICAN 5 th November Peter Newson

Lets Talk Indicators Challenging the process Sterilization. ICAN 5 th November Peter Newson Lets Talk Indicators Challenging the process Sterilization ICAN 5 th November Peter Newson Objectives In the context of the Decontamination process what is an indicator? What phases of the process do we

More information

Manufacturer s IFUs. What Are They and Why Are They So Important?

Manufacturer s IFUs. What Are They and Why Are They So Important? Manufacturer s IFUs What Are They and Why Are They So Important? Speaker Disclaimer Chuck Hughes VP, Infection Prevention Consulting Services SPSmedical Supply Corp. now part of Crosstex International

More information

Packaging Systems for Central Service Operations

Packaging Systems for Central Service Operations Lesson No. CRCST 132 (Technical Continuing Education - TCE) Sponsored by: by Carol Petro, RN, CRCST, CIS, CNOR OR Educator and Sterile Processing Educator Indiana University Health North Hospital Carmel,

More information

ARABIC CANADIAN MEDICAL CENTER CSSD POLICIES AND PROCEDURES

ARABIC CANADIAN MEDICAL CENTER CSSD POLICIES AND PROCEDURES CONTENTS Section Title 2 Disinfection, Packaging and Sterilization 3 Validation of Sterilization (Physical, Chemical and Biological) 4 Sterilizer Requalification 5 Sterile Package Shelf Life 6 Acquisition

More information

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY Lesson No. CRCST 162 (Technical Continuing Education - TCE) Sponsored by: by Susan Klacik, BS, CRCST, ACE, CIS, FCS IAHCSMM Clinical Educator Validation and Verification LEARNING OBJECTIVES 1. Describe

More information

Updates to AAMI s ST79 Steam Sterilization Standard Tuesday, January 15, :00 noon 1:30 pm Eastern

Updates to AAMI s ST79 Steam Sterilization Standard Tuesday, January 15, :00 noon 1:30 pm Eastern Updates to AAMI s ST79 Steam Sterilization Standard Tuesday, January 15, 2013 12:00 noon 1:30 pm Eastern Martha Young, MS, BS, CSPDT Martha L. Young, LLC marthalyoung1@aol.com Updates to AAMI s ST79 Steam

More information

One such problem is blood residual inside the channels of a robotic arm.

One such problem is blood residual inside the channels of a robotic arm. SUBJECT: Robotic Arm Testing for residue blood DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue inside and outside of a robotic arm and

More information

DUAL SPECIES BIOLOGICAL INDICATOR SPORE STRIPS For Monitoring Ethylene Oxide (EO), Steam and Chemiclave Processes

DUAL SPECIES BIOLOGICAL INDICATOR SPORE STRIPS For Monitoring Ethylene Oxide (EO), Steam and Chemiclave Processes DUAL SPECIES BIOLOGICAL INDICATOR SPORE STRIPS For Monitoring Ethylene Oxide (EO), Steam and Chemiclave Processes Reorder Code: STNS-65R Product Description Dual Species Biological Indicator Spore Strips

More information

Competency Guide: Care and Handling of Rigid Endoscopes

Competency Guide: Care and Handling of Rigid Endoscopes Competency Guide: Care and Handling of Rigid Endoscopes This document details the basic steps to correctly process and maintain a rigid endoscope. This document can be used to perform a competency review

More information

To Kill or Not to Kill

To Kill or Not to Kill To Kill or A biological indicator story Not to Kill by Sailaja Chandrapati, BS, MSc, MS, PhD, and Martha Young, BS, MS, CSPDT Objectives 1. Explain the historical evolution of the role of bacterial spores

More information

Obtaining Manufacturer s IFU: Challenges and Barriers

Obtaining Manufacturer s IFU: Challenges and Barriers 2011 AAMI/FDA Medical Device Reprocessing Summit 1 : Obtaining Manufacturer s IFU: Challenges and Barriers Rose Seavey RN, BS, MBA, CNOR, CRCST, CSPDT ANSI/AAMI ST79:2010 & A1:2010 7. Cleaning and other

More information

Options for load release

Options for load release Approaches to Release of Sterile Goods from Steam Sterilization: Parametric Release and Routine Monitoring Craig Wallace Senior Technical Specialist 3M Infection Prevention Division Today s Topics 1. Options

More information

Not to be reproduced - AGPAL & QIP

Not to be reproduced - AGPAL & QIP Keep it Clean! Practical methods for implementing correct sterilisation processes advanced level Presented by: Lin Lochead, B. App. Sc Validation A Documented Procedure for obtaining, recording and interpreting

More information

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation STERILIZATION VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial,

More information

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN DEPARTMENT OF THE ARMY TECHNICAL BULLETIN TBMED2 STERILIZING MEDICAL, SURGICAL, DENTAL, AND VETERINARY MATERIEL Headquarters, Department of the Army, Washington, DC 2 May 987 Approved for public release;

More information

Medline Gemini Sterilization Wrap. Test Studies

Medline Gemini Sterilization Wrap. Test Studies Medline Gemini Sterilization Wrap Test Studies Steam Efficacy : Steam PreVac Steam Efficacy: Steam Gravity Shelf Life: 180 Days Plasma Efficacy: STERRAD 100NX Plasma Efficacy: STERRAD NX Plasma Efficacy:

More information

Quality Control LOAD CONTROL. VERIFY Load Control Products. Use to monitor and release instrumentation with confidence RECORD KEEPING & LABELING

Quality Control LOAD CONTROL. VERIFY Load Control Products. Use to monitor and release instrumentation with confidence RECORD KEEPING & LABELING STERILITY ASSURANCE Use to monitor and release instrumentation with confidence VERIFY Load Control Products LOAD CONTROL PROCESS CONTROL PACK CONTROL EQUIPMENT CONTROL RECORD KEEPING & LABELING PRODUCT

More information

PREVENTING WOUND INFECTION BEGINS BEFORE THE PATIENT

PREVENTING WOUND INFECTION BEGINS BEFORE THE PATIENT Lesson No. CRCST 132 (Technical Continuing Education - TCE) Sponsored by: by Carol Petro, RN, CRCST, CIS, CNOR OR Educator and Sterile Processing Educator Indiana University Health North Hospital Carmel,

More information

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization)

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization) Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization) Date: Office Name: Office Number: Responsible Physician: Assessor: M.1. General Yes No N/A If No Please Comment

More information

Are My Rigid Containers. Maintaining Sterility?

Are My Rigid Containers. Maintaining Sterility? Are My Rigid Containers Maintaining Sterility? 1 Are My Rigid Containers Maintaining Sterility? Objectives 1. List the observations to be taken each time a rigid container is prepared for use 2. Describe

More information

STERILIZATION SOLUTIONS

STERILIZATION SOLUTIONS INFECTION PREVENTION IMS WRAP SPORECHECK STERILIZATION SOLUTIONS Products and Guidelines for Enhanced Office Safety BAGETTE POUCHES STERILIZATION MONITORING CDC recommends placing an indicator inside each

More information

REUSABLE MEDICAL DEVICE VALIDATION

REUSABLE MEDICAL DEVICE VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial, sub-lethal, or repetitive

More information

Getinge Steam Sterilizers From cycle to cycle through lifecycle. Reducing risk in Central Sterile.

Getinge Steam Sterilizers From cycle to cycle through lifecycle. Reducing risk in Central Sterile. Getinge Steam Sterilizers From cycle to cycle through lifecycle. Reducing risk in Central Sterile. Because uptime, over time, is the only time that counts. Central Sterile (CS) technicians and managers,

More information

PART 5. Sterilisation Consumables RFTHPV

PART 5. Sterilisation Consumables RFTHPV PART 5 STATEMENT OF REQUIREMENTS Sterilisation Consumables RFTHPV2015 010 Part 5 Statement of Requirements TABLE OF CONTENTS INTRODUCTION... 3 5.1 Purpose... 3 5.2 Scope... 3 5.3 Product Categories...

More information

Rev. 1 May 2018 Terragene offers a range of effective and easy-to-handle products to control the washing, hygiene, disinfection and sterilization processes in medical diagnostic laboratories. Available

More information

AAMI Quality Systems White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 1

AAMI Quality Systems White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 1 AAMI s White Paper Comparison of 21 CFR Part 820 to ISO 13485:2016 February 2017 AUTHORS Seb Clerkin, GMP Advisory Services Nicola Martin, Owner, Nicola Martin Consulting Jack Ward, Owner, Ward Sciences

More information

3/11/ M Sterile U Network 3M Sterile U Web Meeting March 11, Welcome! Topic: An Overview of ANSI/AAMI ST58: 2013

3/11/ M Sterile U Network 3M Sterile U Web Meeting March 11, Welcome! Topic: An Overview of ANSI/AAMI ST58: 2013 3M Infection Prevention Solutions 3M Sterile U Network 3M Sterile U Web Meeting March 11, 2014 Today s meeting times: 9:00 a.m., 11:00 a.m. and 1:00 p.m. CST To hear audio, call 800 937 0042 and enter

More information

The Significance of Validation to US CSSD Managers

The Significance of Validation to US CSSD Managers The Significance of Validation to US CSSD Managers Jonathan A. Wilder, Ph.D. President, H & W Technology, LLC/Stericert Co. Managing Director, Quality Processing Resource Group, LLC 1 Introduction This

More information

SterilAmp II Geobacillus stearothermophilus

SterilAmp II Geobacillus stearothermophilus SterilAmp II Geobacillus stearothermophilus TECHNICAL REPORT Complies to USP, ISO 11138, and all appropriate subsections Technical Data and Use of SterilAmp II Part #7709 Rev.5 25FEB14 Page 1 of 6 INTRODUCTION

More information

Timings: CSS Department is open all 365 days a year from 09:00 am to 04:00 pm.

Timings: CSS Department is open all 365 days a year from 09:00 am to 04:00 pm. Central Sterile Supply Department (CSSD) - Need of the Hour Introduction: The Central Sterile Supply Department (CSSD) is a service unit of the hospital responsible for providing guaranteed sterile equipments

More information

How is Sterilization Validation Performed by the Medical Device Manufacturer?

How is Sterilization Validation Performed by the Medical Device Manufacturer? How is Sterilization Validation Performed by the Medical Device Manufacturer? Rod Parker, Ph.D. Sr. Principal Scientist Stryker Instruments Division May 9, 2016 DISCLOSURE I am an employee of Stryker Corporation.

More information

MAINTAINING EFFECTIVE WORKSTATIONS IS CRUCIAL FOR

MAINTAINING EFFECTIVE WORKSTATIONS IS CRUCIAL FOR CRCST Self-Study Lesson Plan Lesson No. CRCST 163 (Technical Continuing Education - TCE) by Breanna Svenson, CRCST Sponsored by: Maintaining an Effective Workstation LEARNING OBJECTIVES 1. Review the importance

More information

Disinfection & Sterilization in the Ambulatory Surgery Center

Disinfection & Sterilization in the Ambulatory Surgery Center Disinfection & Sterilization in the Ambulatory Surgery Center 9500+ Patients Potential Exposed to Hepatitis B, Hepatitis C, or HIV Objectives Define important components in cleaning, disinfection, and

More information

gke Steri-Record Testsets and PCDs for validation, process and batch monitoring of sterilization processes

gke Steri-Record Testsets and PCDs for validation, process and batch monitoring of sterilization processes gke Steri-Record Testsets and PCDs for validation, process and batch monitoring of sterilization processes Application Process Monitoring Systems (PMS) are process challenge devices that are used to test

More information

Stericool H 2 O 2. plasma sterilizers Affordable low temperature solution

Stericool H 2 O 2. plasma sterilizers Affordable low temperature solution Stericool H 2 plasma sterilizers Affordable low temperature solution Getinge Stericool H 2 plasma sterilizers Meeting everyday demands of the CSSD An affordable low temperature solution Since many components

More information

Cool when it counts. 3M Steri-Vac Sterilizer/Aerator GS Series. For Health Care Facilities

Cool when it counts. 3M Steri-Vac Sterilizer/Aerator GS Series. For Health Care Facilities Cool when it counts 3M Steri-Vac Sterilizer/Aerator GS Series For Health Care Facilities Stay cool under pressure with state of the art temperature and humidity monitoring and a touch screen cycle display

More information

What is happening inside that magic box? Fundamentals of vaporized hydrogen peroxide sterilization in health care facilities

What is happening inside that magic box? Fundamentals of vaporized hydrogen peroxide sterilization in health care facilities SM What is happening inside that magic box? Fundamentals of vaporized hydrogen peroxide sterilization in health care facilities September 20, 2017 APIC, New York, New York Disclosure Judy Lambert BS, CSPDT

More information

Is Your Sterile Processing Department Ready for Survey?

Is Your Sterile Processing Department Ready for Survey? Is Your Sterile Processing Department Ready for Survey? Roberta Harbison Director of Sterilization and High Level Disinfection Services, AtlantiCare a Member of Geisinger Health System Linda Lucherine

More information