An Improved Method of Updating Retention Times for GCMS

Size: px
Start display at page:

Download "An Improved Method of Updating Retention Times for GCMS"

Transcription

1 An Improved Method of Updating Retention Times for GCMS Richard Whitney, Ph.D., GC/GCMS Senior Product Specialist, Ronald D. Snelling, Ph.D., GC/GCMS Senior Product Specialist, Clifford M. Taylor, GC/GCMS Product Manager, Shimadzu Scientific Instruments, Columbia, MD, USA

2 Abstract One of the common issues in gas chromatography-mass spectrometry (GCMS) is updating of retention time windows when column maintenance is performed. Updating 6 to 8 retention times manually is usually not time-consuming, but many methods have 50 or more target compounds. Manually updating that many retention times is very time consuming and, if not done properly, target compounds will not be determined accurately. A new technique has been developed to update retention times automatically in the data processing software. The technique is based on the use of retention indices, referenced to a hydrocarbon standard (series of n-alkanes). This is a multipoint correction method, so target analytes will always be bracketed by the retention index marker compounds to ensure accuracy of updated retention times for target compounds. Examples of this technique illustrate the accuracy of this technique using a subset of US EPA Method 8270D analytes. 2

3 Method Retention Time Updating 1.5 Method retention times (quantitation windows) (x1,000,000) When a capillary column is cut for maintenance, retention times shift. 1.5 (x1,000,000) ????? Manual updating of retention times can be time-consuming, especially when the number of target analytes is large (>50-100). 3

4 Method Retention Time Updating Two approaches have been employed to deal with updating of retention times: 4 Adjust chromatographic conditions to keep retention times constant after column maintenance. Use retention time markers (retention indices) to easily adjust retention times with changing conditions (column length), keeping other chromatographic conditions constant. Based on fundamental chromatographic principles Maintains optimum chromatographic performance Easy, inexpensive, accurate

5 What is Retention Index? The retention index is information unique to a compound for a given analytical column phase. It is the retention time of the compound expressed on a scale based on n-alkane retention times. Kovats proposed it in 1958.* It is based on a thermodynamic principle and it is correlated to chemical structure. *Kovats, E Helv. Chim Acta 41: (1958) min Decane (C10) min Undecane (C11) 9.733min Decane (C10) min Undecane (C11) min Limonene Retention Index min Limonene Retention Index Instrument A Retention index is constant. Instrument B

6 Carbon number x 100 Linear Retention Index LRI (T) = 100 n (RT T - RT Cn ) / (RT Cn +1 - RT Cn ) (Temperature Program Condition) Carbon number vs RT RT Cn, RT Cn+1 - retention times of C n and C n+1 alkanes uv(x100,000) Chromatogram C10 T C RT min 6 Note: under linear temperature programming conditions, n-alkanes show a linear relationship between retention time and carbon number

7 Principle of Retention Index (1) Method information C17 C18 C R.I Target compound 21.5 min (Method value) n-alkanes Determining retention indices (done only once for each method) A series of n-alkane standards are run under the standard method conditions used for target compounds. From the method retention times for the target compounds and the n-alkanes, retention indices are calculated. 7

8 Principle of Retention Index (2) Retention indices are used to accurately predict method retention times for new conditions (done only once for new conditions). n-alkane standards are run under the revised conditions. New method retention times are calculated for target compounds After column trimming C17 C18 C R.I min Calculated value Retention time n-alkanes

9 Principle of Retention Index (3) Original method information C17 C18 C19 After column trimming min (original method value) C17 C18 C19 R.I. n-alkanes R.I Retention time 21.1 min new method value

10 Advantages of using Retention Index Automatic Adjustment of Retention Times (AART) Retention times of up to 1000 compounds can be updated simultaneously with one analysis of an n- alkane Reduced input by human errors Easy operation Less work, regardless of the total number of compounds Multiple n-alkane standards Correction is performed at multiple points from low to high retention times, enabling accurate correction over a wide range of retention times. SIM time updating can be performed simultaneously AART can easily be applied to any existing method 10 for Retention Time Updating

11 AART Operation Step 1 Prepare GCMSsolution Method for Target Compounds (including compound table) 11

12 AART Operation Step 2 Analyze n-alkanes (AART) standard and prepare AART compound table (for n-alkanes) Same conditions used for analysis of target analytes C15 C SHIMADZU

13 AART Operation Step 3 Download retention indices to GCMSsolution method Run Wizard (modify) under create compound table Retention indices are calculated for the target analytes 13

14 AART Operation Step 4 Run n-alkanes standard under revised conditions and update retention times for target analytes Execute AART function to update target compound retention times Target compound retention times are updated SHIMADZU

15 AART Results (1) AART results: Target compound retention times are accurately predicted for the revised conditions (with 2-3min retention time change for late-eluters). All target compounds properly identified; all isomer assignments correct, indicating accurate updating of quantitation windows SHIMADZU

16 AART Results (2) AART results: After executing AART, GCMSsolution quantitation properly identified all 95 target analytes. All isomer assignments were correct in the processed data (ie, benzo (b&k)fluoranthenes). Retention times predicted very accurately across the entire 25min GC run. 16

17 Library Searching with Retention Index (1) Many compounds show similar spectra (isomers, homologs, etc.) Spectra of compounds with multiple isomers show library search results with multiple hits difficult to interpret the results (six hits for this example) 17

18 Library Searching with Retention Index (2) Use retention index as library search criterion (Retention indices are included in many libraries) Set Retention Index scale with data from n-alkane standard Set Retention Index allowance to limit library search 18

19 Library Searching with Retention Index (3) Using retention indices can qualify the library search results Library search result: Benzo(a)anthracene Library search result: Chrysene 19 Note: the library must include accurate, isomer-specific retention indices, corresponding to the column phase used for analyses

20 AART Summary 20 AART function is unique to Shimadzu GCMSsolution Based on fundamental science retention index Maintains optimum chromatographic conditions Easy, fast, inexpensive Can be applied to any GCMS method with up to 1000 analytes Extremely accurate - based on use of multi-point retention index standards Retention index is useful in library searching 2007 SHIMADZU

TurboMass GC/MS Software

TurboMass GC/MS Software TurboMass GC/MS Software GAS CHROMATOGRAPHY / MASS SPECTROMETRY P R O D U C T N O T E Today s demanding laboratory requires software that is easy to learn yet offers sophisticated instrument control, robust

More information

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications TECHNICAL BRIEF Gas Chromatography/ Mass Spectrometry The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications C 14 scans Introduction As gas chromatography

More information

Gas Chromatography SSI-GC-003

Gas Chromatography SSI-GC-003 Gas Chromatography SSI-GC-003 Trace Oxygenated Hydrocarbons in Liquid Hydrocarbon Streams by Multidimensional GC Introduction It has been widely accepted that oxygenates are related to corrosion and fouling

More information

New Diagnostic Method for Pancreatic Cancer by Profiling of Metabolites in Serum Using GC/MS

New Diagnostic Method for Pancreatic Cancer by Profiling of Metabolites in Serum Using GC/MS C146-E 6-E156 New Diagnostic Method for Pancreatic Cancer by Profiling of Metabolites in Serum Using GC/MS Dr. Masaru Yoshida, Division of Metabolomics Research, Division of Gastroenterology, The Integrated

More information

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730 application Note Gas Chromatography Author Terry Osenbach PerkinElmer, Inc. Shelton, CT 06484 USA Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730 Introduction Detailed hydrocarbon analysis

More information

Application Note. Abstract. Authors. Environmental

Application Note. Abstract. Authors. Environmental Maintaining Compound Retention Times with the Backflush Enabled Pressure Controlled Tee Configuration for Agilent 7890A GCs with Agilent 5975 Series MSD and Agilent 7000 Series Triple Quadrupole MS Systems

More information

QUANLYNX APPLICATION MANAGER

QUANLYNX APPLICATION MANAGER QUANLYNX APPLICATION MANAGER OVERVIEW INTRODUCTION QuanLynx, an Application Manager included with Waters MassLynx Software, is designed for quantitative analysis. Simply acquiring mass spectrometry data

More information

Expand Your Research with Metabolomics and Proteomics. Christine Miller Omics Market Manager ASMS 2017

Expand Your Research with Metabolomics and Proteomics. Christine Miller Omics Market Manager ASMS 2017 Expand Your Research with Metabolomics and Proteomics Christine Miller Omics Market Manager ASMS 2017 New Additions to Agilent Omics Workflows Acquisition Ion Mobility Q-TOF IM All Ions MS/MS Find Features

More information

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water NWTPH-Gx Volatile Petroleum Products Method for Soil and Water Summary The NWTPH-Gx method is intended to replace the Oregon s TPH-G and Washington s WTPH-G methods and to present a more comprehensive

More information

Target and Non-target Pesticide Residue Analysis Using GC/QQQ and GC/Q-TOF with a Discussion on Ways to Use Sandwich Injections

Target and Non-target Pesticide Residue Analysis Using GC/QQQ and GC/Q-TOF with a Discussion on Ways to Use Sandwich Injections Target and Non-target Pesticide Residue Analysis Using GC/QQQ and GC/Q-TOF with a Discussion on Ways to Use Sandwich Injections Philip L. Wylie Agilent Technologies Wilmington, DE 1 Topics Target vs. Non-target

More information

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis Sebastian Westrup, Nick Duczak, and Michal Godula Thermo Fisher

More information

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of

More information

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System Application Note Petrochemicals Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D50 on the Agilent 0 GC System Authors Jie Pan, Lukas Wieder,

More information

OpenLAB CDS Intelligent Reporting

OpenLAB CDS Intelligent Reporting OpenLAB CDS Intelligent Reporting Jon Welsh Application Specialist Agilent Technologies GC/GCMS User Meeting OpenLAB Intelligence Reporter - Overview Enable intelligent reporting and knowledge generation

More information

Improving Return on Innovation in Gas Chromatography

Improving Return on Innovation in Gas Chromatography Improving Return on Innovation in Gas Chromatography Incremental Economic of the Agilent Intuvo 9000 GC System White Paper Author Eric Denoyer, Ph.D. Director of Marketing, GC & Workflow Automation Agilent

More information

Gas Chromatography Assignment Chem 543/443

Gas Chromatography Assignment Chem 543/443 Gas Chromatography Assignment Chem 543/443 1. Introduction Capillary gas chromatography (GC) is one of the most popular analytical techniques used in today s research. Its popularity is mainly due to efficient

More information

FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE

FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE TYPES OF PESTICIDE ANALYSES: OrganoChlorine Pesticides OrganoPhosphorus Pesticides Nitrogen Containing Pesticides Carbamates Herbicides Standard

More information

ADVANCES IN MICRO GAS CHROMATOGRAPHY

ADVANCES IN MICRO GAS CHROMATOGRAPHY ADVANCES IN MICRO GAS CHROMATOGRAPHY Fast and Accurate Analysis of Extended Natural Gas Composition and Physical Properties Using a Temperature Programmable Micro GC Fusion PITTCON 2016 Debbie Alcorn INFICON,

More information

Simplifying Complex Multi-Residue Pesticide Methodology in GC-MS/MS

Simplifying Complex Multi-Residue Pesticide Methodology in GC-MS/MS Simplifying Complex Multi-Residue Pesticide Methodology in GC-MS/MS David Steiniger, Juan Carmona, Stacy Crain, Paul Silcock, and Jason Cole Thermo Fisher Scientific, Austin, TX, USA Technical Note 10319

More information

Thesis of Ph. D. dissertation. Prediction of moldy spoilage of fruits. Sági-Kiss Virág. Budapest

Thesis of Ph. D. dissertation. Prediction of moldy spoilage of fruits. Sági-Kiss Virág. Budapest Thesis of Ph. D. dissertation Prediction of moldy spoilage of fruits Sági-Kiss Virág Budapest 2012 PhD School Name: Field: Head: Supervisor: PhD School of Food Science Food Science Prof. Péter Fodor Corvinus

More information

ADVANCES IN MICRO GAS CHROMATOGRAPHY

ADVANCES IN MICRO GAS CHROMATOGRAPHY ADVANCES IN MICRO GAS CHROMATOGRAPHY Fast and Accurate Analysis of Extended Natural Gas Composition and Physical Properties Using a Temperature Programmable Micro GC Fusion Gulf Coast Conference 2015 Debbie

More information

A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery Using HPLC-Chip/ MS-Enabled ESI-TOF MS

A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery Using HPLC-Chip/ MS-Enabled ESI-TOF MS Application Note PROTEOMICS METABOLOMICS GENOMICS INFORMATICS GLYILEVALCYSGLUGLNALASERLEUASPARG CYSVALLYSPROLYSPHETYRTHRLEUHISLYS A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery

More information

What Does the Future Hold for Ambient Air Monitoring Regulations?

What Does the Future Hold for Ambient Air Monitoring Regulations? An Executive Summary What Does the Future Hold for Ambient Air Monitoring Regulations? A PAMS Site Instrumentation Evaluation for VOC Monitoring Nicola Watson Sales Support Manager - Americas Markes International

More information

C146-E096 TD-20. Shimadzu Thermal Desorption System

C146-E096 TD-20. Shimadzu Thermal Desorption System C146-E096 TD-20 Shimadzu Thermal Desorption System Thermal Desorption Thermal desorption systems trap target substances carried in mobile phase by adsorption in a sample tube packed with adsorbent. Then,

More information

Application Note: Overview. Introduction

Application Note: Overview. Introduction Application Note: 10014 Use of a Novel Large Volume Splitless Injection Technique and Sequential Full Scan/SIM for Simultaneous Screening and Confirmation of Toxicological Specimens Thermo Fisher Scientific

More information

Energy Saving Capillary Gas Chromatograph

Energy Saving Capillary Gas Chromatograph Energy Saving Capillary Gas Chromatograph C184-E026A Energy Saving Capillary Gas Chromatograph Eco-friendly Compact - All in One High Performance GCsolution Ver.2 Options. Specifications P. 4 5 P. 6 7

More information

Fusion Analytical Method Validation

Fusion Analytical Method Validation Fusion QbD Software Platform Fusion Analytical Method Validation The Only Software That Has It All! 100% aligned with FDA/ICH Quality by Design (QbD) guidances! Can be used for LC and Non-LC methods (e.g.

More information

SPECTROSCOPIC METHODS FOR VOCs DETERMINATION

SPECTROSCOPIC METHODS FOR VOCs DETERMINATION SPECTROSCOPIC METHODS FOR VOCs DETERMINATION Monica Culea 1, O.Cozar 1, Cora Craciun 1, R.Fechete 2, Andreea Iordache 1, E.Culea 2 1 Faculty of Physics, Babes-Bolyai University of Cluj-Napoca, 400084 Cluj-

More information

Fast Gas Chromatography using Heated Headspace Gas Auto Sampling Techniques: Polyethylene Pellets, Product

Fast Gas Chromatography using Heated Headspace Gas Auto Sampling Techniques: Polyethylene Pellets, Product Fast Gas Chromatography using Heated Headspace Gas Auto Sampling Techniques: Polyethylene Pellets, Product Purity Analysis Matt Holliday, Presenter Derrick Saul, Joe Perron - Falcon Analytical Abstract

More information

Pesticide Analysis in a Complex Matrix using a 2D GC/MS System

Pesticide Analysis in a Complex Matrix using a 2D GC/MS System Welcome to our E-Seminar: Pesticide Analysis in a Complex Matrix using a 2D GC/MS System Slide 1 Multidimensional (2-D) GC Very old (>25 yrs) but powerful separation technique Based on cutting peak(s)

More information

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use Liquid Chromatograph Liquid Chromatograph Mass Spectrometer CL Series for In Vitro Diagnostic Use Consistency. Performance. Reliability. Shimadzu s CL Series LC, LC-MS, and LC-MS/MS for in vitro diagnostic

More information

GC/MS (SIM) Determination of 14 Pesticides Including Cyhexatin and 2, 4, 5-T in Nuts (Received September 30, 1996)

GC/MS (SIM) Determination of 14 Pesticides Including Cyhexatin and 2, 4, 5-T in Nuts (Received September 30, 1996) June 1997 161 Note GC/MS (SIM) Determination of 14 Pesticides Including Cyhexatin and 2, 4, 5-T in Nuts (Received September 30, 1996) Masaru KAWASAKI, KatSuharu FUKUHARA, Rleko KATSUMURA, Norlko TAKASAKA

More information

Analysis of Polychlorinated Biphenyls and Chlorinated Pesticides by Gas Chromatography with Electron Capture Detection

Analysis of Polychlorinated Biphenyls and Chlorinated Pesticides by Gas Chromatography with Electron Capture Detection Analysis of Polychlorinated Biphenyls and Chlorinated Pesticides by Gas Chromatography with Electron Capture Detection Patricia Van Hoof and Jui-Lan Hsieh Great Lakes Environmental Research Laboratory

More information

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene Angela Smith Henry Kelly Beard George Gonzalez 1 Impurity Monitoring in Olefin Production Analysis of contaminants in ethylene and propylene

More information

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Quantification of genotoxic Impurity D in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 12 Series RRLC and 641B Triple Quadrupole LC/MS Application Note Manufacturing Process Development Author Siji Joseph Agilent

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Bruker s Scion Instrument s long tradition long of tradition innovation of innovation and product and reliability

More information

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and

More information

WHITEPAPER. Key Parameter Concepts. Part 1 - Effect of Column Temperature on Bioethanol High Performance Liquid Chromatography

WHITEPAPER. Key Parameter Concepts. Part 1 - Effect of Column Temperature on Bioethanol High Performance Liquid Chromatography WHITEPAPER Analytical and Measuring Instruments Key Parameter Concepts Part 1 - Effect of Column Temperature on Bioethanol High Performance Liquid Chromatography James Mott, Ph.D., Shimadzu Scientific

More information

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water Application Note Environmental Authors Min Cai and Yun Zou Agilent Technologies Co. Ltd, 412 Ying Lun Road Waigaoqiao

More information

Multiresidue Pesticide Analysis with the Agilent Intuvo 9000 GC and Agilent 7000 Series Mass Spectrometer

Multiresidue Pesticide Analysis with the Agilent Intuvo 9000 GC and Agilent 7000 Series Mass Spectrometer Multiresidue Pesticide Analysis with the Agilent Intuvo GC and Agilent 7 Series Mass Spectrometer Application Note Author Rebecca Veeneman, PhD and Joan Stevens, PhD Agilent Technologies, Inc. Abstract

More information

Manual Analytical Data Adjustment. Environmental Health Laboratories A division of MAS Technology Corporation

Manual Analytical Data Adjustment. Environmental Health Laboratories A division of MAS Technology Corporation Corporate Policy for Quality Assurance Manual Analytical Data Adjustment Environmental Health Laboratories A division of MAS Technology Corporation 110 S. Hill Street South Bend, IN 46617 Phone: (219)

More information

THERMAL SAMPLING TECHNIQUES FOR THE ANALYSIS OF POLYMERS BY GC-MS

THERMAL SAMPLING TECHNIQUES FOR THE ANALYSIS OF POLYMERS BY GC-MS JPACSM 100 THERMAL SAMPLING TECHNIQUES FOR THE ANALYSIS OF POLYMERS BY GC-MS CDS Analytical, Inc. Oxford, PA ABSTRACT Gas chromatography, especially when coupled with mass spectrometry, has become a ubiquitous

More information

Total Petroleum Hydrocarbons in Vapor: Seeing the Forest for the Trees

Total Petroleum Hydrocarbons in Vapor: Seeing the Forest for the Trees Total Petroleum Hydrocarbons in Vapor: Seeing the Forest for the Trees An Overview and Laboratory Perspective of Air-Phase TPH Suzie Nawikas & Kristin Beckley for H&P Mobile Geochemistry, Inc. Overview

More information

Impurity Control in the European Pharmacopoeia

Impurity Control in the European Pharmacopoeia Impurity Control in the European Pharmacopoeia Training Session Zagreb, Croatia, 24-25 May, 2018 Dr Ulrich Rose Head of Division European Pharmacopoeia Department, EDQM Agenda Which impurities are controlled?

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 5 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

IMPORTANCE OF CELL CULTURE MEDIA TO THE BIOPHARMACEUTICAL INDUSTRY

IMPORTANCE OF CELL CULTURE MEDIA TO THE BIOPHARMACEUTICAL INDUSTRY IMPORTANCE OF CELL CULTURE MEDIA TO THE BIOPHARMACEUTICAL INDUSTRY Research, development and production of biopharmaceuticals is growing rapidly, thanks to the increase of novel therapeutics and biosimilar

More information

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC UOP 63-13: Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC Fast Analysis in

More information

Fiehn GC/MS Metabolomics RTL Library IDENTIFY MORE METABOLITES

Fiehn GC/MS Metabolomics RTL Library IDENTIFY MORE METABOLITES Fiehn GC/MS Metabolomics RTL Library IDENTIFY MORE METABOLITES A Comprehensive GC/MS Solution for Metabolite Identification To further facilitate identification of known metabolites through untargeted metabolomics,

More information

Interlaboratory Validation Study for the Analysis of Nitrosamines in Drinking Water using GC-MS/MS

Interlaboratory Validation Study for the Analysis of Nitrosamines in Drinking Water using GC-MS/MS Interlaboratory Validation Study for the Analysis of Nitrosamines in Drinking Water using GC-MS/MS Diana Wong, PhD National Environmental Monitoring Conference Washington, D.C. August 7, 2017 1 Outline

More information

Fusion Analytical Method Validation

Fusion Analytical Method Validation Fusion QbD Software System Fusion Analytical Method Validation The Only Software That Has It All! 100% aligned with FDA/ICH Quality by Design (QbD) guidances! Can be used for LC and Non-LC methods (e.g.

More information

The Suite for Environmental GC Analysis

The Suite for Environmental GC Analysis The Suite for Environmental GC Analysis SGE Environmental GC Columns Performance Selectivity Delivery Promise www.sge.com SGE Environmental GC Columns The Suite for Environmental GC Analysis SGE GC Columns

More information

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor 229 10. Validated Normal Phase HPLC Method for the Determination of Fulvestrant in Pharmaceutical Dosage Forms 10.1 Introduction Fulvestrant is primarily used in the treatment of hormone receptor positive

More information

A Single Analyzer Solution for Refinery Flare Emissions Monitoring (40 CFR 60 Subpart Ja) Using Mass Spectrometry

A Single Analyzer Solution for Refinery Flare Emissions Monitoring (40 CFR 60 Subpart Ja) Using Mass Spectrometry A Single Analyzer Solution for Refinery Flare Emissions Monitoring (40 CFR 60 Subpart Ja) Using Mass Spectrometry Introduction New environmental regulations have recently prompted many oil refineries and

More information

Agilent Ultimate Plus Fused Silica Tubing

Agilent Ultimate Plus Fused Silica Tubing Agilent Ultimate Plus Fused Silica Tubing Technical Overview Introduction Deactivated fused silica tubing is widely used for guard columns, transfer lines, or long retention gaps in GC and GC/MS applications

More information

The Use of High Resolution Accurate Mass GC/Q-TOF and Chemometrics in the Identification of Environmental Pollutants in Wastewater Effluents

The Use of High Resolution Accurate Mass GC/Q-TOF and Chemometrics in the Identification of Environmental Pollutants in Wastewater Effluents The Use of High Resolution Accurate Mass GC/Q-TOF and Chemometrics in the Identification of Environmental Pollutants in Wastewater s Application Note Environmental Authors Anthony Gravell and Praveen Kutty

More information

Chemical Analysis. DHA Analyzer Family. Optimized Solutions for Detailed Hydrocarbon Analysis. Gas Chromatography. think forward

Chemical Analysis. DHA Analyzer Family. Optimized Solutions for Detailed Hydrocarbon Analysis. Gas Chromatography. think forward Chemical Analysis DHA Analyzer Family Optimized Solutions for Detailed Hydrocarbon Analysis think forward Gas Chromatography The DHA Analyzer is a complete high resolution gas chromatography solution for

More information

The Agilent StreamSelect LC/MS Solution: Increasing the Throughput of a Triple Quadrupole Mass Spectrometer

The Agilent StreamSelect LC/MS Solution: Increasing the Throughput of a Triple Quadrupole Mass Spectrometer The Agilent StreamSelect LC/MS Solution: Increasing the Throughput of a Triple Quadrupole Mass Spectrometer Technical Overview Introduction Liquid chromatography triple quadrupole mass spectrometry (LC/MS/MS)

More information

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS Puzzled About LCMS? Sensitivity in Mass Spec Analysis Sample Prep Adapting LC-UV Optimizing and Maintaining Your Mass Spec To LCMS Paul Altiero Alex Ucci Applications Phone Support Chemistry and Supplies

More information

Sulfur in Natural Gas /LPG GC configuration Quick Start March 2013

Sulfur in Natural Gas /LPG GC configuration Quick Start March 2013 This QuickStart document assumes that the user is familiar with the theory of operation of the Multiple- Gas#3 GC configuration. See SRI s website for this document. Step 1: FPD signal FID signal Inject

More information

UNIFI: The user environment Ken Eglinton Nordic User Training, September 2013

UNIFI: The user environment Ken Eglinton Nordic User Training, September 2013 UNIFI: The user environment Ken Eglinton Nordic User Training, September 2013 2013 Waters Corporation 1 What is UNIFI? Single Platform for Chromatography, Mass Spectrometry, Data Management and Laboratory

More information

Drinking Water Semivolatiles Analysis using the 6890N/5975B inert GC/MSD. Application. Author. Abstract. Instrument Operating Parameters.

Drinking Water Semivolatiles Analysis using the 6890N/5975B inert GC/MSD. Application. Author. Abstract. Instrument Operating Parameters. Drinking Water Semivolatiles Analysis using the 6890N/5975B inert GC/MSD Application Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808 USA Abstract The analysis

More information

Comparing gradient transfer of isocratic hold and delay volume addition using the Agilent 1290 Infinity LC with ISET

Comparing gradient transfer of isocratic hold and delay volume addition using the Agilent 1290 Infinity LC with ISET Comparing gradient transfer of isocratic hold and delay volume addition using the Agilent 1290 Infinity LC with ISET Technical Overview 180 Red 1290 Infinity LC with isocratic hold of 1 ute Blue 1 Series

More information

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2 Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP and DB-CLP Application Note Environmental Authors Doris Smith and Ken Lynam Agilent Technologies, Inc. Abstract Organohalide pesticides

More information

Tap Water Analysis by Atomic Absorption and Ion Chromatography Michelle Youn Chem. 31: Inorganic Quantitative Analysis Section 6

Tap Water Analysis by Atomic Absorption and Ion Chromatography Michelle Youn Chem. 31: Inorganic Quantitative Analysis Section 6 Tap Water Analysis by Atomic Absorption and Ion Chromatography Michelle Youn Chem. 31: Inorganic Quantitative Analysis Section 6 California State University, Sacramento April 30, 2018 Introduction The

More information

Instrumental Solutions for Metabolomics

Instrumental Solutions for Metabolomics Instrumental Solutions for Metabolomics Dr. Desislav Donchev US14702384-16012196010132-S01-A01.tif 300 400 500 600 700 800 3100 3200 3300 3400 3500 3600 3700 3800 Color Scale R = [52 1484] G = [26 514]

More information

AUTOMATIC PEROXYACETYL NITRATE (PAN) ANALYZER

AUTOMATIC PEROXYACETYL NITRATE (PAN) ANALYZER AUTOMATIC PEROXYACETYL NITRATE (PAN) ANALYZER The Instrument The PAN GC-was developed in 1992 with funding by the German EPA (UBA). Since that time the system has undergone a number of updates, partially

More information

Case Study on Application of Analytical Life Cycle Management and Risk Management

Case Study on Application of Analytical Life Cycle Management and Risk Management Case Study on Application of Analytical Life Cycle Management and Risk Management Jianmei, Ph.D. Genzyme, a Sanofi Company AAPS 2015 Dione Pompe disease Brazil www.genzyme.com Outline Introduction of Concepts

More information

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* UCT Products: ECUNIPAH (ENVIRO-CLEAN Universal PAH/DRO 2000 mg/83 ml cartridge) Method Summary Method 8015D may be used

More information

Agilent s NEW MassHunter Profinder

Agilent s NEW MassHunter Profinder Agilent s NEW MassHunter Profinder The Most Advanced Batch Feature Extraction Software for Metabolomics Theodore Sana, Ph.D. Metabolomics Marketing Manager 1 MassHunter Profinder A Batch Feature Extraction

More information

Agilent 7820A GC Agilent 5975E GC/MSD. Your best value for routine GC and GC/MS analysis. The Mea sure of Confidence

Agilent 7820A GC Agilent 5975E GC/MSD. Your best value for routine GC and GC/MS analysis. The Mea sure of Confidence Agilent 7820A GC Agilent 5975E GC/MSD Your best value for routine GC and GC/MS analysis The Mea sure of Confidence Introducing the Agilent 7820A GC and 5975E GC/MSD Reliability and value, day after day.

More information

Techniques for Reducing the Effects of Sample Solvent on UHPLC Analyses

Techniques for Reducing the Effects of Sample Solvent on UHPLC Analyses PO-CON1404E Techniques for Reducing the Effects of Sample Solvent Pittcon 2014 2050-11P Kenichiro Tanaka, William Hedgepeth Shimadzu Scientific Instruments, Inc., Columbia MD Introduction For HPLC, sample

More information

High temperature Simdist analysis using the new MXT-1 HT Simdist phase. Jaap de Zeeuw and Gary Stidsen Restek corporation, Bellefonte, PA, USA

High temperature Simdist analysis using the new MXT-1 HT Simdist phase. Jaap de Zeeuw and Gary Stidsen Restek corporation, Bellefonte, PA, USA High temperature Simdist analysis using the new MXT-1 HT Simdist phase Jaap de Zeeuw and Gary Stidsen Restek corporation, Bellefonte, PA, USA Simulated Distillation Determination of the boiling point

More information

How discovery activities can influence metabolic profiling in the regulatory space? C. DELATOUR EBF, 25 th September 2015

How discovery activities can influence metabolic profiling in the regulatory space? C. DELATOUR EBF, 25 th September 2015 How discovery activities can influence metabolic profiling in the regulatory space? C. DELATOUR EBF, 25 th September 2015 Outline 2 ן Background - Objectives - Shift in Workflow - Metabolite in safety

More information

Gas Filtration Catalogue

Gas Filtration Catalogue Gas Filtration Catalogue 2 SHIMADZU GAS FILTRATION CATALOGUE ABOUT SUPER CLEAN GAS FILTERS Cartridge systems make changing gas filters quick and easy. A base plate allows cartridges to be exchanged without

More information

Chromeleon 7.2 CDS The Easiest Way to Boost Your IC Productivity

Chromeleon 7.2 CDS The Easiest Way to Boost Your IC Productivity Chromeleon 7.2 CDS The Easiest Way to Boost Your IC Productivity Mississauga, ON, 10/13/15 Carl Fisher, Ph.D, Application Scientist The world leader in serving science 2 40 th Anniversary of Ion Chromatography

More information

Agilent 1290 Infinity LC with ISET under Waters Empower control Emulation of Agilent 1100 Series LC

Agilent 1290 Infinity LC with ISET under Waters Empower control Emulation of Agilent 1100 Series LC Agilent 1290 Infinity LC with ISET under Waters Empower control Emulation of Agilent 1100 Series LC Technical Overview Author A. G. Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract The Agilent

More information

SUGGESTIONS FOR REDUCTION OF ANALYTICAL COSTS BY ELIMINATION OF UNNECESSARY QUALITY CONTROL (QC) SAMPLES

SUGGESTIONS FOR REDUCTION OF ANALYTICAL COSTS BY ELIMINATION OF UNNECESSARY QUALITY CONTROL (QC) SAMPLES SUGGESTIONS FOR REDUCTION OF ANALYTICAL COSTS BY ELIMINATION OF UNNECESSARY QUALITY CONTROL (QC) SAMPLES Douglas M. Chatham 5413 Forest Ridge Dr., Loganville, GA 30052-3437 Environmental Specialist and

More information

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM)

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM) Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM) Mark Belmont, David Steiniger, Eric Phillips, Sergio Guazzotti, Pat O Brien,

More information

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *)

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *) VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *) Guideline Title Validation of Analytical Procedures: Methodology Legislative basis Directive 75/318/EEC as amended Date of first adoption December 1996

More information

NEW. 136 additives added to the additive library. Multi-functional Pyrolyzer. F-Search System. Multi-functional Pyrolysis System and GC/MS

NEW. 136 additives added to the additive library. Multi-functional Pyrolyzer. F-Search System. Multi-functional Pyrolysis System and GC/MS EW 136 additives added to the additive library ew search software F-Search (Ver. 3.5) supports analysis of polymers and additives with a complete list of libraries: pyrograms, THM pyrograms, thermograms,

More information

Agilent Automated Card Extraction Dried Blood Spot LC/MS System

Agilent Automated Card Extraction Dried Blood Spot LC/MS System Agilent Automated Card Extraction Dried Blood Spot LC/MS System Quick Start Guide What is the AACE DBS LC/MS System? 2 System Overview 3 Operational Overview 4 Method Development 5 Using the AACE DBS LC/MS

More information

NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778)

NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778) NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778) This Notice of Intended Decision to Enter into a Single Source Contract is posted in accordance with section 287.057(3), Florida

More information

SEMI-VOLATILE ORGANICS IN DRINKING WATER using EPA Method optimized for greater sensitivity

SEMI-VOLATILE ORGANICS IN DRINKING WATER using EPA Method optimized for greater sensitivity ENVIRONMENTAL application note SEMI-VOLATILE ORGANICS IN DRINKING WATER using EPA Method 525.2 optimized for greater sensitivity Elaine A. LeMoine Introduction The US EPA Drinking Water Manual contains

More information

APPENDIX III SAMPLE LAB REPORT. Experiment 1. High Performance Liquid Chromatography. Alfred E. Neuman

APPENDIX III SAMPLE LAB REPORT. Experiment 1. High Performance Liquid Chromatography. Alfred E. Neuman APPENDIX III SAMPLE LAB REPORT Experiment 1 High Performance Liquid Chromatography Alfred E. Neuman Chemistry 436 September 15, 1999 INTRODUCTION In this experiment, high performance liquid chromatography

More information

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract. The Use Of Automated Backflush on the 7890A/5975A GC-MS System Setup Example Using Biomarkers Application Hydrocarbon Processing Authors Courtney Milner and Russell Kinghorn BST International 41 Greenaway

More information

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines APPLICATION NOTE 64968 Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines Authors Ali Haghani, 1 Andy Eaton, 1 Neloni Wijeratne, Claudia Martins 1 Eurofins Eaton Analytical,

More information

Analytical Methods Development and Validation

Analytical Methods Development and Validation Understanding and Implementing Efficient Analytical Methods Development and Validation Jay Breaux, Kevin Jones, and Pierre Boulas Analytical methods development and validation play important roles in the

More information

Multiplexing Two Different Food Residue Methods using HILIC and Reversed Phase Chromatography in the Same LC-MS/MS Run

Multiplexing Two Different Food Residue Methods using HILIC and Reversed Phase Chromatography in the Same LC-MS/MS Run Multiplexing Two Different Food Residue Methods using HILIC and Reversed Phase Chromatography in the Same LC-MS/MS Run Adrian Taylor and André Schreiber AB SCIEX Concord, Ontario (Canada) Overview Multiplexing

More information

Polymer Degradation Mechanisms using Pyrolysis- GC/MS

Polymer Degradation Mechanisms using Pyrolysis- GC/MS FIELD APPLICATION REPORT GAS CHROMATOGRAPH Y Polymer Degradation Mechanisms using Pyrolysis- GC/MS Author: Thomas Wampler CDS Analytical, Inc. 465 Limestone Road Oxford, PA 19363-0277 Introduction The

More information

Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS with static and dynamic headspace

Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS with static and dynamic headspace PO-CON44E Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS ASMS 20 ThP 138 Sanket Chiplunkar, Ankush Bhone, Durvesh Sawant, Dheeraj Handique, Prashant

More information

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler PO-CON464E Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler ASMS 204 TP763 Ankush Bhone (), Dheeraj Handique (), Prashant Hase (), Sanket Chiplunkar (),

More information

MultiQuant Software 2.0 with the SignalFinder Algorithm

MultiQuant Software 2.0 with the SignalFinder Algorithm MultiQuant Software 2.0 with the SignalFinder Algorithm The Next Generation in Quantitative Data Processing Quantitative analysis using LC/MS/MS has benefited from a number of advancements that have led

More information

Ultra High Performance Liquid Chromatograph. Nexera XR C196-E082C

Ultra High Performance Liquid Chromatograph. Nexera XR C196-E082C Ultra High Performance Liquid Chromatograph Nexera XR C196-E82C The Most Accessible UHPLC Available Developed with expandability and compatibility in mind, the Nexera XR ultra high performance liquid chromatograph

More information

Ultra High Performance Liquid Chromatograph. Nexera XR C196-E082D

Ultra High Performance Liquid Chromatograph. Nexera XR C196-E082D Ultra High Performance Liquid Chromatograph Nexera XR C196-E82D The Most Accessible UHPLC Available Developed with expandability and compatibility in mind, the Nexera XR ultra high performance liquid chromatograph

More information

Development and evaluation of Nano-ESI coupled to a triple quadrupole mass spectrometer for quantitative proteomics research

Development and evaluation of Nano-ESI coupled to a triple quadrupole mass spectrometer for quantitative proteomics research PO-CON138E Development and evaluation of Nano-ESI coupled to a triple quadrupole mass spectrometer for quantitative proteomics research ASMS 213 ThP 115 Shannon L. Cook 1, Hideki Yamamoto 2, Tairo Ogura

More information

METHOD TRANSLATION OF HJ FOR INTUVO

METHOD TRANSLATION OF HJ FOR INTUVO METHOD TRANSLATION OF HJ679-2013 FOR INTUVO Translating a 530 µm id Column Method to 250 µm or 320 µm column Methods for the Agilent Intuvo 9000 GC Introduction HJ method 679-2013 describes the determination

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation

More information

Qualification of High-Performance Liquid Chromatography Systems

Qualification of High-Performance Liquid Chromatography Systems Qualification of High-Performance Liquid Chromatography Systems L. Huber Hewlett- Packard GmbH, Waldbronn December 1998 Manuscript of an article published in BioPharm, Vol 11 Number 11, November 1998,

More information

Pesticide & Environmental Pollutant (P&EP) GC/MS/MS Analyzer 2.0

Pesticide & Environmental Pollutant (P&EP) GC/MS/MS Analyzer 2.0 Pesticide & Environmental Pollutant (P&EP) GC/MS/MS Analyzer 2.0 Get on the Analyzer FAST TRACK Identify Analytes in Complex Matrices Obtain Analytical Results on DAY ONE! -- based on the most comprehensive

More information