Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy: Use of Grid-Controlled Pulsed Fluoroscopy

Size: px
Start display at page:

Download "Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy: Use of Grid-Controlled Pulsed Fluoroscopy"

Transcription

1 Giles W. L. Boland 1 Brian Murphy Ron Arellano Loren Niklason Peter R. Mueller Received January 24, 2000; accepted after revision May 3, Supported by a grant from Philips Medical Systems. 1 All authors: Division of Abdominal Imaging and Interventional Radiology, Massachusetts General Hospital, White 270, 55 Fruit St., Boston, MA Address correspondence to G. W. L. Boland. AJR 2000;175: X/00/ American Roentgen Ray Society Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy: Use of Grid-Controlled Pulsed Fluoroscopy OBJECTIVE. We evaluated the diagnostic accuracy of a grid-controlled fluoroscopy unit compared with a conventional continuous fluoroscopy unit for a variety of abdominal and pelvic fluoroscopic examinations. SUBJECTS AND METHODS. Seventy patients (29 men and 41 women; age range, years) were enrolled in one of seven abdominal and pelvic fluoroscopic examinations, including upper gastrointestinal series (n = 20), barium enema (n = 10), voiding cystourethrogram (n = 10), percutaneous abdominal catheter tube injection (n = 10), hysterosalpingogram (n = 10), and percutaneous needle insertion and catheter placement (nephrostomy, percutaneous biliary drainage) (n = 10). Each patient underwent at least 10 sec of continuous fluoroscopy that was randomly and blindly compared with 10-sec periods of pulsed fluoroscopy at 15, 7.5, and 3.75 frames per second. A radiologist outside the examination room, unaware of the frame rate per second, evaluated the procedure in real time on a television monitor. The radiologist assessed image quality and diagnostic acceptability using a scoring system. Statistical analysis was performed using the paired Student s t test. RESULTS. For all procedures at all frame rates, we found no statistically significant superiority of one frame rate over another. For most procedures, the slower frame rates were considered equivalent to continuous fluoroscopy when the images were assessed for image quality and diagnostic confidence. CONCLUSION. Our findings suggest that most abdominal and pelvic fluoroscopic procedures can be performed at substantially lower frame rates than those used for continuous fluoroscopy; adopting this procedure may lead to substantial dose savings for the patient and the fluoroscopy operator. F luoroscopic procedures contribute to patient and operator X-ray radiation in diagnostic radiology [1 8]; this is particularly true for long procedures (angiographic and general interventional procedures), but fluoroscopic doses for common procedures vary widely depending on the patient case mix and the skill of the performing radiologist. Additionally, radiologists may receive a high cumulative lifetime exposure to X-ray radiation from fluoroscopic procedures [1 8]. Although the radiologist can use simple dose-saving maneuvers to minimize fluoroscopic doses (X-ray field collimation, image intensifier placed close to the patient, spot films minimized), manufacturers have investigated methods to reduce the exposure dose emitted from the X-ray tube. Most radiographic machines use low-energy filters to remove unwanted low-energy X rays emitted from the anode because these X rays only increase patient dose and do not improve image quality. Several manufacturers have offered intermittent or pulsed fluoroscopy to reduce the dose. The premise is that a conventional continuous beam is not required for most fluoroscopic procedures, which can be adequately performed at slower frame rates. Although initial studies using pulsed fluoroscopy revealed the potential for reduced radiation dose when compared with continuous fluoroscopy, earlier models of pulsed fluoroscopy machines had significant limitations, and overall patient doses were similar to those of continuous fluoroscopy [5 8]. Because the pulses to the anode were controlled at the X-ray generator level, high resistance was induced in the long X-ray cables. This lead to an uneven current, with the production of a significant percentage of low-energy X-ray beams, which did not improve the AJR:175, November

2 Boland et al. radiographic image [9]. Therefore, many manufacturers abandoned this modification. More recent modification designs resulted in the pulses originating in the X-ray tube rather than in the X-ray generator [9]. With a grid-controlled technique, the resulting pulses are in the form of discrete and uniform current without wasteful low-energy radiation. This innovation permits multiple frame rates to be used from one to 30 frames per second (continuous fluoroscopy). Although lower frame rates have been shown in phantom and some cardiac models to reduce exposure doses by more than 50%, to our knowledge no study has compared the image quality, diagnostic fluoroscopic information, and accuracy of these lower frame rates with those of conventional continuous fluoroscopy [5 8]. We evaluated the diagnostic accuracy of a grid-controlled fluoroscopy unit compared with a conventional continuous fluoroscopy unit for a variety of abdominal and pelvic fluoroscopic examinations. Subjects and Methods Patient Selection Seventy prospective patients (29 men and 41 women; age range, years; mean age, 44 years; median, 52 years) were recruited for our study. Consecutive patents were asked to participate in the study if they were referred for the following examinations: upper gastrointestinal series (n = 20), barium enema (n = 10), voiding cystourethrography (n = 10), percutaneous abdominal catheter tube injection (n = 10), hysterosalpingography (n = 10), percutaneous nephrostomy (n = 5), and percutaneous transhepatic cholangiography and biliary catheter placement (n = 5). Of 20 patients referred for upper gastrointestinal series, 11 were referred for symptoms of peptic ulcer disease, six for postoperative assessment of esophagogastrectomy, and three for assessment of gastroesophageal reflux and esophagitis. Of 10 patients referred for barium enema, four were referred for assessment of colonic bleeding, three for large-bowel obstruction, and three for the assessment of a defunctionalized distal colon before colonic anastomosis. All 10 patients TABLE 1 Note. fps = frames per second. who underwent voiding cystourethrography were referred for the assessment of prior urinary tract infections. All 10 patients who underwent percutaneous abdominal catheter tube injections were assessed for evidence of abdominal fistula or for the size of abscess cavity. All 10 patients who underwent hysterosalpingography were referred for assessment of infertility. Of five patients who underwent nephrostomy, three were referred for ureteral obstruction and two for catheter and wire placement for percutaneous urethrolithotomy. Of five patients who underwent percutaneous biliary drainage, four were referred for biliary obstruction from malignant disease and one for balloon dilatation of a benign stricture. All patients were required to sign a consent form to enter the study, which was approved by the institutional review board at our hospital. Pregnant patients and those younger than 18 years were excluded from the study. Procedure Protocol All fluoroscopic procedures were performed using an EZ Diagnost scanner (Philips Medical Systems, Erlangen, The Netherlands). For all procedures, each patient underwent standard continuous fluoroscopy using three differing frame rates: 15, 7.5, and 3.75 frames per second. For each procedure, one of two radiologists was asked to assess the fluoroscopic image for diagnostic acceptability and image quality. To achieve this comparison, each patient underwent fluoroscopic exposures divided into three pairs or segments, with each segment consisting of a period of continuous exposure for 10 sec that was then compared with one of three different frame rates for a period of 10 sec. The order in which the frame rates were selected was random. For each fluoroscopic segment, the selection of the order of continuous fluoroscopy versus fluoroscopy at one of the three randomly chosen frame rates was also random. For example, a three-segment examination might consist of the following: first segment: continuous fluoroscopy (10 sec), then fluoroscopy at 7.5 frames per second (10 sec); second segment: fluoroscopy at 15 frames per second (10 sec), then continuous fluoroscopy (10 sec); and third segment: continuous fluoroscopy (10 sec), then fluoroscopy at 3.75 frames per second (10 sec). Three experienced gastrointestinal and genitourinary radiologists participated in the study. For Comparison of Image Quality for Differing Frame Rates in Specified Procedures Procedure No. of Examinations each procedure, two radiologists participated. One radiologist was required to perform the fluoroscopic examination and the other to review the images for analysis. The second radiologist reviewed the images on a separate monitor outside the examination room as the procedure was performed. Only the X-ray technologist knew the order of the fluoroscopic segment pairs. At the beginning of the examination, the technologist selected a sheet that listed a three-segment examination and included all three fluoroscopic pairs. One radiologist performed the examination with the X-ray technologist, who switched the frame rates according to the selected sheet. The X-ray technologist indicated to the radiologist analyzing the study (outside the room) that the frame rate was changing by notifying them when a new fluoroscopic frame rate was started. Each patient received a minimum of 60 sec of exposure (three segments with a pair of frame rates at 10 sec each). We believe that this procedure did not significantly increase the length of the examination because most fluoroscopic examinations last at least 60 sec. Even if a procedure could have been performed in a shorter time, the overall dose to the patient would likely remain the same because of the geometric design of the X-ray tube and image intensifier position. Image Analysis A scoring sheet for each of the three segments involved in the examination was devised and completed during the examination. The observing radiologists were requested to evaluate each pair of segments according to overall image quality and diagnostic acceptability. The scoring system consisted of three levels: level 1 was superior, level 2 was similar, and level 3 was inferior. Statistical analysis of the results was performed for continuous versus pulsed fluoroscopy at 3.75, 7.5, and 15 frames per second using the paired Student s t test. Results Our results are presented in Table 1. For all procedures at all frame rates, we found no statistically significant superiority of one frame rate over another. For all procedures, when the images were assessed for diagnostic acceptability, Best at 15 fps Best at 7.5 fps Best at 3.75 fps 15 fps 7.5 fps Upper gastrointestinal examination Barium enema Cystourethrography Hysterosalpingography Percutaneous catheter injection Percutaneous catheter insertion fps 1454 AJR:175, November 2000

3 Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy continuous fluoroscopy was considered equivalent to pulsed fluoroscopy at the slower frame rates. For most procedures, when the images were assessed for image quality, continuous fluoroscopy was considered nearly equivalent to pulsed fluoroscopy. However, we found continuous fluoroscopy to be superior to pulsed fluoroscopy at the 7.5 and particularly the 3.75 frame rates per second, but this finding was not statistically significant. This trend was more marked for interventional procedures that incorporated fine motor movements used in catheter and needle manipulation and for patient motion caused by turning in upper gastrointestinal series rather than for procedures that required less patient motion or static imaging (i.e., hysterosalpingography and barium enema). Similarly, however, this trend was not significant. Discussion fluoroscopy can lead to potentially large patient radiation doses and, to a lesser extent, radiation doses to the operator [1 8]. A variety of methods are used by diagnostic radiologists to minimize the exposure dose to the patient and operator, particularly for complex interventional procedures and pediatric examinations. Often these measures are simple; are based on the ALARA principle [2] (use a dose as low as reasonably achievable); and include minimizing exposure time, minimizing collimation, lowering tube current, increasing kilovoltage, and minimizing the number of spot radiographs obtained. However, in a continuing drive to reduce exposure dose, manufacturers have attempted either to modify equipment or to reduce output exposure without compromising image quality. These modifications traditionally involved improved detection devices (spot film and image intensifiers) that enable either more of the X-ray beam that exits the patient to be absorbed or greater signal output from imaging detectors. Additionally, equipment modifications have enabled conversion to pure digital imaging with the removal of the spot radiograph device, which reduces the distance of the patient to the image intensifier and, in turn, the dose. More recently, manufacturers have investigated methods of modifying the exposure from the X-ray tube. During conventional fluoroscopy, the X-ray beam is continuous and rapidly contributes to the total patient and operator dose. To minimize the dose, the operator must use fluoroscopy for the minimum time necessary to gain adequate diagnostic information. Unfortunately, because of a combination of operator experience and patient case mix, radiation doses for common procedures are extremely variable. Patient doses can vary markedly from one operator to another for the same procedure, despite adequate radiologic and equipment training. Therefore, manufacturers developed a method of reducing the dose by permitting the fluoroscopist to lower the exposure by reducing the number of exposures per second, so-called pulsed fluoroscopy. Instead of a continuous exposure, the X-ray beam is pulsed at a variable number of times per second, which can be controlled by the operator. Image perception would obviously alter because of the flicker or flicker effect on the images, depending on the number of frames used per second. The number of frames per second that would be acceptable for diagnostic fluoroscopy has never been scientifically established. Initially, pulsed fluoroscopy did not live up to its expectations, primarily because of limitations in producing a pulsed exposure of uniform current [9]. Conventional pulsed fluoroscopy involved rapid on off switching at the generator level. Because the generator has to be placed several yards away from the X-ray tube (both for safety reasons and the size of equipment), the cables between the X-ray tube and the generator can often be up to 20 ft (6 m) long. This length of cables induces a strong resistance that causes a strong capacitance in the cables. The effect produces an uneven current in the cables, with a slow rise in current after the exposure is initiated until the expected current is attained, then a slow Fig. 1. Drawing shows X-ray pulse output from conventional pulsed fluoroscopic system and grid-controlled fluoroscopic system. Note absence of ramping and trailing effect with grid-controlled fluoroscopy. fall back to zero current until fluoroscopy is turned off. The shape of the exposure curve is similar to a bell-shaped curve (Fig. 1), with the steepness of the curve dependent on the length of cabling. This rise and fall in current has been termed the ramp and trail effect of conventional fluoroscopy. Because much of the desired exposure level is below the chosen current, a large proportion of X rays are in the lower energy range. Because these lower energy X rays are absorbed by the patient, the result is an increased radiation dose without improvement in image quality. Therefore, conventional pulsed fluoroscopy offered little if any dose savings over continuous fluoroscopy, and many manufacturers discontinued offering it as an option on fluoroscopy machines. Recently, an innovation called grid-controlled fluoroscopy has been developed. This technique also uses pulsed fluoroscopy, but the pulses are controlled in the X-ray tube and not at the generator. Therefore, the ramp and trailing effect caused by cable capacitance is not observed (Fig. 1). Discrete uniform tube current is maintained by placing a negatively charged grid between the electron-emitting cathode and the X ray producing anode (Fig. 2). When the grid is switched on, the negative charge repels the electrons emitted from the cathode and prevents a current across the X-ray tube. Multiple pulses can be produced by rapidly switching the grid on and off. Because the on off switching mechanism is instantaneous, a uniform current can be AJR:175, November

4 Boland et al. Fig. 2. Drawing shows X-ray tube with additional negatively charged grid (white arrow) that can be switched on and off to generate discrete pulsed fluoroscopy. Note cathode (curved arrow) and anode (black arrow). produced across the X-ray tube (Fig. 2). This ensures a uniform X-ray exposure to the patient and minimizes unwanted radiation. Image quality is also improved because the X rays are produced in the desired diagnostic range. A further advantage of the grid-controlled fluoroscopic system is that the length of the exposure pulse can be varied depending on the clinical application. For standard abdominal applications, a 5- to 20-msec exposure pulse is typically used, whereas in pediatric applications, the pulse length can be reduced to 2 5 msec [10]. A secondary advantage of grid-controlled fluoroscopy is that the blooming effect of the image intensifier is reduced [10]. If during continuous fluoroscopy the image intensifier is moved (as it often is), then a small time elapses before the automatic exposure device corrects the new field of view to the correct exposure. As fluoroscopic panning is performed, a temporary overexposure occurs when moving from a dense area (mediastinum) to a less dense area (the lungs) (or from the pelvis to the abdomen), which causes image degradation with streaking and flaring of the image (image blooming). The opposite effect occurs when moving from the lungs to the mediastinum. This blooming effect is significantly reduced using grid-controlled fluoroscopy because each pulse automatically corrects to the desired exposure level. The grid-controlled fluoroscopic technique has several advantages. The most obvious is the ability to reduce the overall radiation dose by changing the selected frame rate. Higher frame rates lead to higher exposure. Previous studies have documented a substantially reduced dose using pulsed fluoroscopy in in vitro and in vivo studies [5, 9, 10]. Dose savings to the patient have been calculated to be 22% for frame rates of 15 frames per second, 38% for 10 frames per second, 49% for 7.5 frames per second, 80% for 3.75 frames per second, and as much as 86% for 2 frames per second [5 9, 10]. What is uncertain is whether these reduced frame rates can be used in clinical practice to produce diagnostically acceptable images. Our findings reveal that grid-controlled fluoroscopy can be used as an alternative to continuous fluoroscopy and can produce images of diagnostic quality. For this study to have relevant implications for abdominal fluoroscopic procedures, the procedures analyzed were chosen to reflect a range of examinations. Furthermore, to compare the effects of slower frame rates on different examinations, we included examinations that required static imaging (i.e., hysterosalpingography) and those in which the patient was required to move (i.e., upper gastrointestinal series). The grid-controlled fluoroscopic effect is most pronounced for studies involving little patient motion or image intensifier movement (i.e., hysterosalpingography and abdominal catheter injections) in which 3.75 frames per second is likely to be acceptable for most imaging. However, even for studies involving significant patient or image intensifier movement (upper gastrointestinal series and delicate catheter manipulations), the slower frame rates (3.75 and 7.5 frames per second) proved acceptable in most patients. In patients in whom a higher frame rate is desirable, the frame rate can easily be switched to 15 frames per second, which still offers a significant dose reduction of at least 22% when compared with continuous fluoroscopy. Our findings suggest that continuous fluoroscopy should rarely be used for abdominal and pelvic fluoroscopic procedures. Potential disadvantages of grid-controlled fluoroscopy might be expected because of the perceived flicker effect, particularly at lower frame rates. The flicker effect is observed when reducing the frame rate from continuous fluoroscopy. The image has an increasing tendency to flicker as the frame rate decreases. However, this effect did not appear to affect the ability of the radiologists to observe and diagnose normal and abnormal conditions. We noticed that some initial training and habituation to the lower frame rates may be required before they are considered acceptable. This initiation period is likely to vary from one observer to the next. A further potential disadvantage of this technique is that fine catheter and needle manipulations might not be seen to their best advantage using the slower frame rates. However, we found that in only three of 10 patients was a higher frame rate desirable; this was not the case for the other seven patients we examined. It is possible that a larger study might reveal that higher frame rates should be used for some procedures. This is the first clinical report to directly compare grid-controlled fluoroscopy with continuous fluoroscopy. Our results strongly suggest that grid-controlled fluoroscopy is acceptable for all abdominal fluoroscopic procedures, although a frame rate of 15 frames per second may be required for studies that require considerable motion or fine catheter or needle manipulations. Although a recognizable flicker effect occurs at lower frame rates (3.75 frames per seconds), our findings reveal that the lower frame rate does not cause a loss of diagnostic detail or prolonged examinations. This study did not evaluate potential quantitative dose reductions in each patient, although this has been confirmed by other phantom studies [5]. Further studies should be performed to quantitatively evaluate the actual dose savings to patients using grid-controlled fluoroscopy. Typical fluoroscopic doses for an upper gastrointestinal series range from 2 to 3 rad ( Gy) [1 5]. This study suggests that these doses can be reduced to less than 1 rad (0.01 Gy) for most examinations. This reduction is particularly important for pediatric patients and patients of 1456 AJR:175, November 2000

5 Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy reproductive age undergoing pelvic fluoroscopy. Many pelvic fluoroscopic examinations involve static imaging (hysterosalpingography and voiding cystourethrography), and our findings suggest that 3.75 frames per second may be used for most of these examinations. References 1. National Council on Radiation Protection and Measurement. Exposure of the US population from diagnostic medical radiation. Bethesda, MD: International Commission on Radiological Protection, Report International Commission on Radiological Protection. The ALARA principle. Bethesda, MD: International Commission on Radiological Protection, Report Berthelsen B, Cederblad A. Radiation dose to patients and personnel involved in embolization of intracerebral arteriovenous malformations. Acta Radiol 1991;32: Lindsey BD, Eichling JO, Ambos D, Cain ME. Radiation exposure to patients and medical personnel during radiofrequency catheter ablation for supraventricular tachycardia. Am J Cardiol 1992;70: Aufrichtig R, Thomas CW, Gilmore GC, Wilson DL. Perceptual comparison of pulse and continuous fluoroscopy. Med Phys 1994;21: Holmes DR Jr, Wondrow MA, Gray JE, Vetter RJ, Fellows JL, Julsrud PR. Effect of pulsed progressive fluoroscopy on reduction of radiation dose in the cardiac catheterization laboratory. J Am Coll Cardiol 1990;15: Grollman JH Jr, Klosterman Herman MW, Moler CL, Eber LM, MacAlpin RN. Dose reduction low pulsed rate fluoroscopy. Radiology 1972;105: Grollman JH Jr. Radiation reduction by means of low pulsed rate fluoroscopy during cardiac catheterization and coronary angiography. AJR 1974;121: Ammann E, Wiede G. Generators and tubes in interventional radiology. In: Categorical course in physics. Oak Brook, IL: Radiological Society of North America, 1995: Hahn H, Farber D, Allmendinger H, Brendler J. Grid-controlled fluoroscopy in pediatric radiology. Medicamundi 1997;41:12 17 AJR:175, November

Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children

Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children Mervyn D. Cohen, MB, ChB, MD Radiologists desire to keep radiation dose as low as possible. Pulsed fluoroscopy provides

More information

Interventional Fluoroscopy

Interventional Fluoroscopy NATIONAL CANCER INSTITUTE THE SOCIETY OF INTERVENTIONAL RADIOLOGY Interventional Fluoroscopy Reducing Radiation Risks for Patients and Staff CONTENTS Introduction 1 Increasing use and complexity of interventional

More information

2018/7/31. Primary Objective

2018/7/31. Primary Objective Andrew Kuhls-Gilcrist, PhD, DABR July 31, 2018 Primary Objective The purpose of this session is to allow vendors to present their solutions and describe how their product can empower the clinician and

More information

Reducing Fluoroscopic Doses in Routine Musculoskeletal Joint Related Procedures

Reducing Fluoroscopic Doses in Routine Musculoskeletal Joint Related Procedures Reducing Fluoroscopic Doses in Routine Musculoskeletal Joint Related Procedures Mital Patel MD, Pranav Chitkara MD, Vikram Kinni MD, Rati Patel MD Naveen Subhas MD, Ryan Fisher PhD Disclosure No relevant

More information

2015 Radiology Coding Survival Guide

2015 Radiology Coding Survival Guide 2015 Radiology Coding Survival Guide Chapter 7: Gastrointestinal Tract (74210-74363) If you report routine GI procedures, you'll find the codes you need here, such as 74240-74245 (Radiologic examination,

More information

Fluoroscopic Radiation Safety

Fluoroscopic Radiation Safety Fluoroscopic Radiation Safety Patient and Occupational Safety Michael J. Bohan, YNHH Radiation Safety Officer (203) 688-2950 or mike.bohan@yale.edu ACCREDITATION & DISCLOSURE The Yale School of Medicine

More information

Making the difference with Live Image Guidance Dose management in Allura systems

Making the difference with Live Image Guidance Dose management in Allura systems DoseWise Allura Making the difference with Live Image Guidance Dose management in Allura systems The importance of radiation dose management The harmful effects of X-rays on human tissues had already become

More information

Power of BRANSIST safire in Neuroendovascular Therapy

Power of BRANSIST safire in Neuroendovascular Therapy Vascular Power of BRANSIST safire in Neuroendovascular Therapy Department of Radiology, Kinki University Hospital Suguru Ueda Mr. Suguru Ueda 1. Introduction Kinki University Hospital is located in the

More information

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety 2018 OPTIONS FOR INDIVIDUAL MEASURES: REGISTRY ONLY MEASURE

More information

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Poster No.: C-0674 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Sofue 1, T. Yoshikawa 1, N. Negi 1, Y. Ohno

More information

Radiation Protection in Fluoroscopy

Radiation Protection in Fluoroscopy Radiation Protection in Fluoroscopy Introduction Physicians performing fluoroscopic guided procedures should be aware of the potential for serious radiation-induced skin injury to the patient. Occasionally

More information

7/28/2017. Michael Speidel. University of Wisconsin - Madison

7/28/2017. Michael Speidel. University of Wisconsin - Madison Scanning-Beam Digital X-ray (SBDX) is a technology for real-time fluoroscopy and angiography in the cardiac cath lab Michael Speidel Characterized by high speed beam scanning with a 2D array of focal spots

More information

The X-ray Tube Cathode Assembly Continued

The X-ray Tube Cathode Assembly Continued The X-ray Tube Cathode Assembly Continued C. Focusing Cup: 1. Blooming 2. Saturation Current or Temperature Limited Region 3. Grid Biased X-ray Tubes The Cathode Assembly 2 Focusing Cup The focusing cup

More information

Radiation Exposure During Endovascular Procedures: Methods to Protect Patients, Physicians and OR Personnel

Radiation Exposure During Endovascular Procedures: Methods to Protect Patients, Physicians and OR Personnel Radiation Exposure During Endovascular Procedures: Methods to Protect Patients, Physicians and OR Personnel Melissa L. Kirkwood MD Associate Professor Division of Vascular and Endovascular Surgery UT Southwestern

More information

The Radiation Dose in Interventional Radiology Study: Knowledge Brings Responsibility

The Radiation Dose in Interventional Radiology Study: Knowledge Brings Responsibility Commentary The Radiation Dose in Interventional Radiology Study: Knowledge Brings Responsibility M. Victoria Marx,MD J Vasc Interv Radiol 2003; 14:947 951 Abbreviations: CD cumulative dose, FDA Food and

More information

FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION

FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION Ray Dielman RSO ST ANTHONY S HOSPITAL James Menge CHP ARSO ST ANTHONY S HOSPITAL BAYCARE HEALTH SYSTEM 1 FLUROSCOPY EDUCATIONAL OBJECTIVES I. Understand the

More information

Methods for measuring fluoroscopic skin dose

Methods for measuring fluoroscopic skin dose Pediatr Radiol (2006) 36 (Suppl 2): 136 140 DOI 10.1007/s00247-006-0193-3 ALARA Stephen Balter Methods for measuring fluoroscopic skin dose Published online: 22 July 2006 # Springer-Verlag 2006 Abstract

More information

Introduction to Medical Imaging. Lecture 6: Radiography ??? hc λ. X-Ray Discovery. Discovered by Wilhelm Röntgen in 1895.

Introduction to Medical Imaging. Lecture 6: Radiography ??? hc λ. X-Ray Discovery. Discovered by Wilhelm Röntgen in 1895. X-Ray Discovery Introduction to Medical Imaging Lecture 6: Radiography Discovered by Wilhelm Röntgen in 1895 accidentally, when performing experiments with cathode tubes and fluorescent screens the light

More information

Comprehensive Dose Management for Improved Patient Care.

Comprehensive Dose Management for Improved Patient Care. Comprehensive Dose Management for Improved Patient Care. Set the standard for dose management Toshiba is committed to continuously advancing dose reduction technologies and incorporating these into its

More information

TENDER # SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH

TENDER # SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH TENDER # 0571-1707 SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH Amendment tice issued April 13,2017 1. The closing date time is amended to May 5 th,2017 2:00PM 2. The specifications

More information

Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil

Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil Canevaro, L.V. (Presenting); Oliveira da Silva, M.W.; Dias Rodrigues, B.B. Serviço de Física Médica, Instituto de Radioproteção

More information

Fluoroscopy Credentialing Training Program for Non-Radiologist Physicians. Department of Radiology

Fluoroscopy Credentialing Training Program for Non-Radiologist Physicians. Department of Radiology Fluoroscopy Credentialing Training Program for Non-Radiologist Physicians Department of Radiology Why Physician Credentialing? FDA. That all facilities assure appropriate credentials and training for physicians

More information

1. Executive Summary

1. Executive Summary 1. Executive Summary 1.1 General The fluoroscope is defined as an instrument used chiefly in industry and in the practice of medicine for observing the internal structure of objects (such as the living

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 17. Subpart 18. Subpart 19. Subpart 20. Subpart 21. Subpart 22. Subpart 23. Subpart 24. Subpart 25. Equipment performance

More information

IMGING GUIDANCE IN PCNL. Mohammad Hossein Soltani MD Assistant professor of urology Fellowship of endourology

IMGING GUIDANCE IN PCNL. Mohammad Hossein Soltani MD Assistant professor of urology Fellowship of endourology IN THE NAME OF GOD IMGING GUIDANCE IN PCNL Mohammad Hossein Soltani MD Assistant professor of urology Fellowship of endourology Ultrasonographic guidance Real time diagnostic ultrasonography (US) has been

More information

Radiological Protection of Patient and Operator in Interventional Radiology

Radiological Protection of Patient and Operator in Interventional Radiology Radiological Protection of Patient and Operator in Interventional Radiology JMAJ 44(11): 484 489, 2001 Hironobu NAKAMURA Professor and Chairman, Department of Radiology, Osaka University Graduate School

More information

White Paper. Dose Tracking System. Patient Skin Dose Estimates in Real Time

White Paper. Dose Tracking System. Patient Skin Dose Estimates in Real Time White Paper Dose Tracking System Patient Skin Dose Estimates in Real Time INTRODUCTION In an era of increased utilization of ionizing radiation for medical diagnostic and interventional procedures and

More information

Progress in X-Ray & MR

Progress in X-Ray & MR Progress in X-Ray & MR Michiel Manuel Analyst Meeting June 15 th, 2005 X-Ray & MR: Agenda Introduction General X-Ray Cardio/Vascular X-Ray Magnetic Resonance China growth opportunity Conclusion 2 X-Ray

More information

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures Vano E (1,2), Aviles P (1,2), Prieto C (2), Fernandez JM (1,2), Guibelalde

More information

On the Importance of Computation in Clinical Radiology

On the Importance of Computation in Clinical Radiology On the Importance of Computation in Clinical Radiology Manuel Arreola, Ph.D., DABR Director of Radiological Physics Radiology Department University of Florida/Shands Healthcare My Disclaimer I ve received

More information

Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values

Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values Dose at an interventional reference point: Comparison between displayed values of X-ray units and actual measured values Poster No.: C-1844 Congress: ECR 2012 Type: Authors: Keywords: DOI: Educational

More information

First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector

First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector 01 White Paper No. 02/2009 First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector Leiden University Medical Center (LUMC) in the Netherlands is the first hospital in the world

More information

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals Koos Geleijns, medical physicist CT delivers excellent 3D image quality CT delivers excellent 3D image quality

More information

Disclaimer and Notices.

Disclaimer and Notices. Fluoroscopy Dose Reduction and Clinical Management Fluoroscopy Trajectories (Curves), Global Dose Monitoring and Tracking of Patient Dose. Pei Jan Paul Lin, Ph.D. Division of Radiation Physics and Biology

More information

Making the difference with Live Image Guidance

Making the difference with Live Image Guidance Surgery BV Pulsera Making the difference with Live Image Guidance Surgical imaging powerhouse Making the difference with Live Image Guidance Together we make the difference in surgical procedures to improve

More information

Purpose. Methods and Materials

Purpose. Methods and Materials MR cholangiopancreatography: Comparison between 3D fast recovery fast spin echo and 2D single shot fast spin echo sequences in the evaluation of choledocholithiasis. Poster No.: C-1193 Congress: ECR 2012

More information

Local dose reference levels for abdominal interventional radiology procedures

Local dose reference levels for abdominal interventional radiology procedures Local dose reference levels for abdominal interventional radiology procedures Poster No.: C-2053 Congress: ECR 2014 Type: Scientific Exhibit Authors: L. Hadid-Beurrier 1, M.-J. Waryn 2, Y. Ajavon 1, O.

More information

Final exam of RAD 435 second semester of the year

Final exam of RAD 435 second semester of the year King Saud University Collage of Applied Medical Science Department of Radiology Science Final Exam 2014/1435 second semester Exam date: Wednesday 29/7/1435-28/5/2014 Time: 8-10 am Course: RAD 435 (Level

More information

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety Educating and Engaging your Staff in Radiation Safety Robert Vincent MD;CM FACC Professor of Pediatrics Emory University School of Medicine Ralf Holzer MD Associate Professor of Pediatrics Ohio State University

More information

Prevention of radiation injury from medical imaging

Prevention of radiation injury from medical imaging Prevention of radiation injury from medical imaging Erica L. Mitchell, MD, a and Patricia Furey, MD, b Portland, Ore; and Manchester, NH Introduction: Concerns over radiation safety are valid. Understanding

More information

Table of Contents. Adaptive Diagnostics...7 Integrated Dose Reduction Streamlined Workflow Clinical Images... 26

Table of Contents. Adaptive Diagnostics...7 Integrated Dose Reduction Streamlined Workflow Clinical Images... 26 2 Table of Contents Adaptive Diagnostics...7 Integrated Dose Reduction... 17 Streamlined Workflow... 21 Clinical Images... 26 3 AQUILION TM PRIME PROVIDES CLINICAL FLEXIBILITY, ENHANCED WORKFLOW FEATURES

More information

New Joint Commission Requirements for Fluoroscopy Services

New Joint Commission Requirements for Fluoroscopy Services New Joint Commission Requirements for Fluoroscopy Services Melissa C. Martin, M.S., FAAPM, FACR Past President AAPM Florida Chapter Fall Meeting Orlando, FL September 14, 2018 New Requirements Effective

More information

Dose management at Ziehm Imaging for 2D imaging

Dose management at Ziehm Imaging for 2D imaging 01 Dose management at Ziehm Imaging for 2D imaging An educational paper on general dose principles and concepts as well as further insights on dose management for 2D imaging at Ziehm Imaging. International

More information

This prospective study was approved by our institutional review board.

This prospective study was approved by our institutional review board. Digital subtraction angiography during TIPS creation or revision: data on radiation exposure and image quality obtained using a standard and a low-dose acquisition protocol in a flat-panel detector-based

More information

Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations

Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations M. Río, S. Pérez, A. Herreros, A. Coll, A. Ruiz, J.Baró and J. Fernández Asesoría en Control

More information

Radiation Physics and Safety Considerations in the EP Lab

Radiation Physics and Safety Considerations in the EP Lab Cardiorhythm 2009 Radiation Physics and Safety Considerations in the EP Lab Joel A. Kirsh, MD, FRCPC, CCDS Cardiology and Critical Care, Hospital for Sick Children Associate Professor of Pediatrics, University

More information

Clarity CT Technology

Clarity CT Technology Clarity CT Technology WHITE PAPER January 2013 Using state of the art algorithms Sapheneia Clarity CT allows physicians to lower radiation dose when acquiring CT data while maintaining image quality. The

More information

Equipment Quality Control for Fluoroscopy February 22, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Fluoroscopy February 22, Imaging Physics CancerCare Manitoba Equipment Quality Control for Fluoroscopy February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks system

More information

Radiography Curriculum Analysis

Radiography Curriculum Analysis Program Number Program Name Date / /20 Radiography Curriculum Analysis DIRECTIONS: Determine the course(s) in which each of the following content area is covered and enter the course number(s) and/or title(s).

More information

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT)

NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) CRCPD Publication E-16-2 NATIONWIDE EVALUATION OF X-RAY TRENDS (NEXT) TABULATION AND GRAPHICAL SUMMARY OF THE 2008-2009 CARDIAC CATHETERIZATION SURVEY June 2016 Published by Conference of Radiation Control

More information

Justification for Training

Justification for Training This educational program is being offered to provide training to medical practitioners who perform or supervise the performance of diagnostic procedures utilizing fluoroscopic x- ray units. There are many

More information

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS Rutgers Robert Wood Johnson Medical School, New Brunswick, NJ Abstract

More information

What s the Issue? Radiation Dose in the OR. Exposure to OR Personnel. Effects of Proper Collimation. Image-guided procedures are increasingly complex

What s the Issue? Radiation Dose in the OR. Exposure to OR Personnel. Effects of Proper Collimation. Image-guided procedures are increasingly complex Radiation Dose in the OR Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco What s the Issue? Image-guided procedures

More information

Radiation Dose Terms Made Simple v1

Radiation Dose Terms Made Simple v1 Terms Made Simple 9670970 v1 In Air Absorbed Table In Air Absorbed Table Goal To provide basic term definitions and exhibit how radiation effects the patient, surgeon and staff. In Air Absorbed Table Why

More information

Patient doses in lower limb angiography in a recently installed flat detector x-ray system

Patient doses in lower limb angiography in a recently installed flat detector x-ray system Patient doses in lower limb angiography in a recently installed flat detector x-ray system Poster No.: C-0939 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper V. Tsapaki, D. Papavdis,

More information

There are three general methods for reducing one's personal radiation exposure. The application of each when working around fluoroscopy units will be

There are three general methods for reducing one's personal radiation exposure. The application of each when working around fluoroscopy units will be There are three general methods for reducing one's personal radiation exposure. The application of each when working around fluoroscopy units will be discussed in more detail. The ALARA concept was introduced

More information

1. Executive Summary

1. Executive Summary 1. Executive Summary Diagnostic reference levels (DRLs), which are a form of investigation levels (ICRP, 1996), represent an important tool to optimize image quality and the radiation dose delivered to

More information

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite Clinical Applications ImagingRite Neuro Intervention 1 ImagingRite ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in

More information

Optimization of pulsed fluoroscopy in pediatric radiology using voiding cystourethrography as an example

Optimization of pulsed fluoroscopy in pediatric radiology using voiding cystourethrography as an example Clinical applications Optimization of pulsed fluoroscopy in pediatric radiology using voiding cystourethrography as an example R. Schumacher H. Allmendinger Department of Pediatric Radiology, University

More information

Clinical Applications. ImagingRite. Interventional Radiology

Clinical Applications. ImagingRite. Interventional Radiology Clinical Applications ImagingRite Interventional Radiology ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in optimizing

More information

New York State Department of Health Bureau of Environmental Radiation Protection

New York State Department of Health Bureau of Environmental Radiation Protection New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Program in Small Facilities Part II Fluoroscopic Equipment Introduction A.

More information

Improving protocols and procedures for strengthened radiation protection in interventional procedures

Improving protocols and procedures for strengthened radiation protection in interventional procedures Improving protocols and procedures for strengthened radiation protection in interventional procedures R. Loose German Commission on Radiological Protection (SSK) German Roentgen Society (DRG) Institute

More information

Clinical Experience Using SONIALVISION G4 Multi-Purpose R/F System for Angiographic Examinations

Clinical Experience Using SONIALVISION G4 Multi-Purpose R/F System for Angiographic Examinations R/F Clinical Experience Using SONIALVISION G4 Multi-Purpose R/F System for Angiographic Examinations Department of Central Radiology, Takasago Municipal Hospital Naoki Shioya Mr. Naoki Shioya 1. Introduction

More information

RECONSTRUCTION OF ORGAN DOSES IN CHILDREN UNDERGOING LIFE-SAVING FLUOROSCOPY-GUIDED CARDIAC SURGERY

RECONSTRUCTION OF ORGAN DOSES IN CHILDREN UNDERGOING LIFE-SAVING FLUOROSCOPY-GUIDED CARDIAC SURGERY RECONSTRUCTION OF ORGAN DOSES IN CHILDREN UNDERGOING LIFE-SAVING FLUOROSCOPY-GUIDED CARDIAC SURGERY November 13 th, 2015 Emily Marshall, B.S. J Crayton Pruitt Family Department of Biomedical Engineering

More information

Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT

Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT Responsibilities Worker (BSS 3.83) Workers shall: follow any applicable rules for protection use properly the monitoring devices and

More information

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland Medical Physics and Informatics Original Research Seuri et al. Tracking Radiologic Procedures and Dose Medical Physics and Informatics Original Research FOCUS ON: Raija Seuri 1 Madan M. Rehani 2 Mika Kortesniemi

More information

Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide.

Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide. Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide. Fluoroscopy units used in angiographic and cardiac applications are normally

More information

Technique for Percutaneous Fluoroscopically Guided G-Tube Placement in a High-BMI Patient

Technique for Percutaneous Fluoroscopically Guided G-Tube Placement in a High-BMI Patient Technique for Percutaneous Fluoroscopically Guided G-Tube Placement in a High-BMI Patient Irwin M Best, Emory University Journal Title: Case Reports in Critical Care Volume: Volume 2012, Number 807161

More information

Title. CitationPhysica Medica, 30(5): Issue Date Doc URL. Type. File Information.

Title. CitationPhysica Medica, 30(5): Issue Date Doc URL. Type. File Information. Title Preliminary analysis for integration of spot-scannin Author(s)Shimizu, S.; Matsuura, T.; Umezawa, M.; Hiramoto, K. CitationPhysica Medica, 30(5): 555-558 Issue Date 2014-07 Doc URL http://hdl.handle.net/2115/56596

More information

A Paradigm Shift in Patient Dose Monitoring

A Paradigm Shift in Patient Dose Monitoring VL White Paper A Paradigm Shift in Patient Dose Monitoring Dose Tracking System Andrew Kuhls-Gilcrist, Ph.D., D.A.B.R. Senior XRVL Clinical Science Manager Canon Medical Systems, Inc. Millions of fluoroscopically

More information

The Consistent Quality of Connected Radiology

The Consistent Quality of Connected Radiology The Consistent Quality of Connected Radiology Radiology is Evolving New compliance requirements, heightened radiation awareness and increasing technical complexity present new challenges to radiology professionals.

More information

Optimizing Radiation Dose for Samsung GC85A

Optimizing Radiation Dose for Samsung GC85A White Paper Article # WP201602-DR-Dose/ Issue Date 23 Feb., 2016 Optimizing Radiation Dose for Samsung GC85A Hong Eo, M.D. / Hyesun Hwang, M.D. / Samsung Medical Center Introduction ALARA (As Low As Reasonably

More information

Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp Advance Access publication 25 September 2008

Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp Advance Access publication 25 September 2008 Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp. 18 24 Advance Access publication 25 September 2008 doi:10.1093/rpd/ncn247 CORRELATION OF PATIENT MAXIMUM SKIN DOSES IN CARDIAC PROCEDURES WITH

More information

Monitoring of radiation dose using dose tracking software in a neurovascular intervention room

Monitoring of radiation dose using dose tracking software in a neurovascular intervention room Monitoring of radiation dose using dose tracking software in a neurovascular intervention room Poster No.: ESI-0017 Congress: EuroSafe Imaging 2017 Type: EuroSafe Imaging Authors: H. Acton 1, K. James

More information

Types of Radiation Units

Types of Radiation Units RADIATION UNITS Types of Radiation Units Exposure in air - amount of ionization produced by or radiation in air. Unit - (R) or / (c/kg) Absorbed Dose energy transfer of / (unit of energy) of any. Used

More information

How to Optimize Radiation Protection During Complex Peripheral

How to Optimize Radiation Protection During Complex Peripheral How to Optimize Radiation Protection During Complex Peripheral G. Bartal MD, FCIRSE, FSIR Dept of Medical Imaging Meir MC, Kfar Saba, Affiliated to Sackler Medical School, Tel Aviv University, Tel Aviv

More information

Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures

Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures Kostas Perisinakis, PhD; John Damilakis, PhD; Nicholas Theocharopoulos,

More information

Radiation protection for theatre staff (refer guidelines) DISTANCE, TIME, SHIELDING principle should be applied

Radiation protection for theatre staff (refer guidelines) DISTANCE, TIME, SHIELDING principle should be applied Radiation protection for theatre staff (refer guidelines) DISTANCE, TIME, SHIELDING principle should be applied Monitoring (Refer guidelines) All full time theatre personnel must be monitored with dosimeters.

More information

Making the difference

Making the difference Intuis Interventional X-ray Making the difference with Live Image Guidance Making the difference with Live Image Guidance Philips Intuis Together we make the difference in minimally invasive treatment

More information

A New Tool for Benchmarking Cardiovascular Fluoroscopes a

A New Tool for Benchmarking Cardiovascular Fluoroscopes a Benchmarking Fluroroscopes Balter Sep 2000 page 1 of 6 A New Tool for Benchmarking Cardiovascular Fluoroscopes a Stephen Balter, Ph.D. b, Frederick A. Heupler, M.D. c Pei-Jan Paul Lin, Ph.D. d, Merill

More information

Making the difference

Making the difference Intuis Interventional X-ray Making the difference with Live Image Guidance Making the difference with Live Image Guidance Philips Intuis Together we make the difference in minimally invasive treatment

More information

The Safe Use of Fluoroscopy. A required inservice for people who work with or near fluoroscopy

The Safe Use of Fluoroscopy. A required inservice for people who work with or near fluoroscopy The Safe Use of Fluoroscopy A required inservice for people who work with or near fluoroscopy The FDA says... The Food and Drug Administration Center for Devices and Radiological health has received reports

More information

Multifunctional remote control X-ray system. Adapting to the way you work

Multifunctional remote control X-ray system. Adapting to the way you work OmniDiagnost Eleva Multifunctional remote control X-ray system Adapting to the way you work OmniDiagnost Eleva: A system for Is it possible for a universal X-ray system to meet the needs of everyone in

More information

IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures

IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures CYRIC Annual Report 2004 IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures Ohuchi H., Satoh T. *, Eguchi Y. *, and Mori K. ** Graduate School of Pharmaceutical

More information

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL SAHLGRENSKA ACADEMY DEPARTMENT OF RADIATION PHYSICS RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL M.Sc. thesis Roham D.

More information

Optimization of Parameters in 16-slice CT-scan Protocols for Reduction of the Absorbed Dose

Optimization of Parameters in 16-slice CT-scan Protocols for Reduction of the Absorbed Dose Iranian Journal of Medical Physics Vol. 11, No. 2 & 3, Spring & Summer 2014, 270-275 Received: March 10, 2014; Accepted: April 27, 2014 Original Article Optimization of Parameters in 16-slice CT-scan Protocols

More information

MANAGING PATIENT DOSE IN FLUOROSCOPIC PROCEDURES

MANAGING PATIENT DOSE IN FLUOROSCOPIC PROCEDURES MANAGING PATIENT DOSE IN FLUOROSCOPIC PROCEDURES Keith J. Strauss, MSc MSc,, FAAPM Director, Radiology Physics & Engineering Children s Hospital Harvard Medical School DILEMAS A. Fluoroscopically Guided

More information

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF FLUOROSCOPIC EQUIPMENT

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF FLUOROSCOPIC EQUIPMENT The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

Making the difference

Making the difference Veradius Unity Surgery Making the difference with Live Image Guidance Enjoy straightforward surgical imaging Key advantages Transform your surgical procedures with a system as easy to use as a tablet.

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels 2033-7 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dose Reference Levels Peter Homolka EFOMP Training Course on Medical Physics in Diagnostic

More information

Radiography. 1. Electromagnetic Spectrum. 2. X-ray History. 3. X-ray Physics. 4. Clinical Application

Radiography. 1. Electromagnetic Spectrum. 2. X-ray History. 3. X-ray Physics. 4. Clinical Application Radiographs X-RAYS Vincent Carrasco, MD, MSIS NLM Postdoctoral Fellow & Doctoral Candidate Carolina Health Informatics Program The University of North Carolina Chapel Hill Radiography 1. Electromagnetic

More information

Vascular. Improvement of Functions Supporting Neuro-IVR. 1. Introduction. 2. Current State of Functions Supporting Neuro-IVR

Vascular. Improvement of Functions Supporting Neuro-IVR. 1. Introduction. 2. Current State of Functions Supporting Neuro-IVR Vascular Improvement of Functions Supporting Neuro-IVR Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction The use of techniques in which intravascular treatment is performed

More information

DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR

DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR DIRECT MEASUREMENT OF SKIN DOSE WITH RADIOCHROMIC FILMS IN FLUOROSCOPY GUIDED VESSEL REPAIR J. KALEF-EZRA, EZRA, * 1 E. KATSAROU, 1 S. KARAVASILIS, 1 M. MATSAGKAS, 2 L.K. MICHALIS 3 1 Medical Physics Department,

More information

2017 ACR Computed Tomography Quality Control Manual FAQS

2017 ACR Computed Tomography Quality Control Manual FAQS Updated 11-15-2017 2017 ACR Computed Tomography Quality Control Manual FAQS Q. The updated 2017 ACR Computed Tomography Quality Control Manual has been released. (Visit www.acr.org/education/education-catalog.)

More information

RADIATION SAFETY Analytical X-Ray Systems Indiana University - Bloomington

RADIATION SAFETY Analytical X-Ray Systems Indiana University - Bloomington RADIATION SAFETY Analytical X-Ray Systems Indiana University - Bloomington USER TRAINING REQUIREMENTS Analytical X-Ray Systems Complete online training and read associated Radiation Safety Guide. Review

More information

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

More flexibility More performance Less dose

More flexibility More performance Less dose 2010 General Electric Company All rights reserved. GE and GE Monogram are trademarks of General Electric Company. GE Healthcare ASiR, Performix, microvoxel, Volara, and SmartView are trademarks of General

More information

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE Ph. R. Trueb #, A. Aroua *, H. Jung #, Th. Theiler #, J.-F. Valley

More information

The Effect of CT Scanning With Reduced Dose on the X-ray Tube Lifetime

The Effect of CT Scanning With Reduced Dose on the X-ray Tube Lifetime The Effect of CT Scanning With Reduced Dose on the X-ray Tube Lifetime October 11, 2010 By Armin Marcovitch Executive summary The X-ray tube is the most expensive "consumable" of the CT scanner. Prolonging

More information