A New Tool for Benchmarking Cardiovascular Fluoroscopes a

Size: px
Start display at page:

Download "A New Tool for Benchmarking Cardiovascular Fluoroscopes a"

Transcription

1 Benchmarking Fluroroscopes Balter Sep 2000 page 1 of 6 A New Tool for Benchmarking Cardiovascular Fluoroscopes a Stephen Balter, Ph.D. b, Frederick A. Heupler, M.D. c Pei-Jan Paul Lin, Ph.D. d, Merill Wondrow e ABSTRACT This paper reports the status of a new cardiovascular-fluoroscopy-benchmarking-phantom. A joint working group of the Society for Cardiac Angiography and Interventions and the National Electrical Manufacturers Association (NEMA) developed the phantom. The device was adopted as NEMA standard XR , "Characteristics of and Test Procedures for a Phantom to Benchmark Cardiac Fluoroscopic and Photographic Performance in August The test ensemble includes imaging-field geometry, spatial resolution, low-contrast Iodine detectability, working thickness range, visibility of moving targets, and phantom entrance dose. The phantom tests systems under conditions simulating normal clinical use for fluoroscopically guided invasive and interventional procedures. Test procedures rely on trained human observers. FOREWORD Defining and maintaining the quality of X-ray systems used in the cardiac catheterization laboratory (cath lab) has always been a priority for the Society for Cardiac Angiography and Interventions (SCA&I). The society ran a mail-order testing program until the early 1990s. Some consulting groups continue to perform this test set in individual laboratories. The transition of imaging laboratories from cinefilm to digital acquisition along with the increasing integration of computer controlled imaging systems necessitated the development of a new set of tools and procedures. In 1997, to obtain the broadest possible consensus, SCA&I requested the participation of the X-ray section of the National Electrical Manufacturers Association (NEMA). Virtually all of the manufacturers of cath lab equipment are represented in this section. Thus there was strong technical participation from industry. In addition, several SCA&I members, cath lab service staff members, and medical physicists contributed to this project. The performance of any medical imaging system can be divided into two categories: Suitability of the images for the clinical procedure and the amount of energy administered to the patient while acquiring the images. Measurable differences in test performance may or may not reflect meaningful differences in clinical utility. Uncertainties included the variety of clinical tasks for which the equipment might be used, differences in the skills of operators, and the lack of congruence between the phantom and patient tissue. All tests are performed using the imaging system configured for normal clinical use. The phantom and test procedures described here test systems under conditions simulating a range of fluoroscopically guided invasive and interventional procedures. The phantom loads the system, as does a similarly sized patient. Image quality test targets are placed at the center of the phantom (the clinically relevant region). In principle, all of the tests are performed simultaneously. The major tests provide imaging targets with a range of visibilities. No system is expected a Presented on behalf of the NEMA SCA&I working group b New York, NY c Cleveland OH d Chicago, IL e Rochester, MN

2 Benchmarking Fluroroscopes Balter Sep 2000 page 2 of 6 to image the least visible target in all or most of the tests. Indeed, the test ensemble and the requirement for simultaneous viewing of all tests precludes excessive optimization of any one set of targets. For example, too much image smoothing to improve low contrast detectability will degrade spatial resolution. There is limited availability of optical, electronic, or digital techniques for evaluating imaging performance. Therefore, the test procedures in this standard are based on trained human observers. The United States Food and Drug Administration (FDA) has reached a similar conclusion in conjunction with their Mammography regulations. The system level tools described in this paper are used to screen for inappropriate performance. Such tools are not always helpful in diagnosing the causes of such behavior. Supplementary tools can provide additional information about system or subsystem behavior. Equipment manufacturers have their own tools and procedures relevant to the setup and service of specific imaging systems. The NEMA standard does not assign passing or optimum values for test performance. The equipment s owner should determine an acceptable level of performance. As an alternative, recommended values might be given by experienced consultants or by independent testing organizations. Perhaps, these might eventually be established by entities such as equipment manufacturers, service organizations, professional societies, or regulatory agencies. PHANTOM and TEST DESCRIPTIONS f The basic phantom material is Plexiglas (PMMA). This material has X-ray absorption and scatter properties that are similar to soft tissue. The approximation used in this document is that the PMMA thickness is f The phantom shown in this paper is a prototype. Production pieces will be slightly different in construction and appearance. equivalent to the same path-length in the patient. Configuring the phantom to different thickness simulates the range of patient sizes and imaging projection angles (Figures 1,2). Figure 1: Phantom configurations. The numbers refer to the types of plates described in the NEMA Standard Figure 2: The phantom in its 20 cm configuration. Note space under the phantom for the dosemeter and the 5 cm gap between the phantom and the image intensifier. The phantom s working thickness of 5, 10, 20, and 30 cm correspond to PA projections through an infant, large child or small adult, an average size adult, and a heavy adult. Angled projections have longer radiation path-lengths. Imaging systems that perform very well when tested with 20 cm of PMMA may degrade when tested with 5, 10 or 30 cm of PMMA. Imaging tests are performed with the relevant test-object in the center of the phantom. The test-object (inside the phantom) is placed on the system at a position corresponding to the patients organ of interest. (This is usually at the isocenter of an isocentric system.) These test conditions simulate normal X-ray system loading, attenuation, scatter, and imaging geometry.

3 Benchmarking Fluroroscopes Balter Sep 2000 page 3 of 6 Actual Field of View A plate with suitable scales is placed on the entrance surface of the image intensifier. The plate is fluorographed using each available field of view. This determines the field-of-view of the image receptor The effective field-of-view for any source to image receptor distance (SID) is determined by placing the test plate at isocenter and obtaining fluorographs. Geometric magnification will reduce the field-of-view to a value less than that of the image receptor. Congruence of Irradiated and Visualized Fields A large sheet of film is placed between the field of view plate and the image receptor. The size of the irradiated field is determined by inspecting the processed film. A simultaneous fluorographic image confirms the size of the visualized field-of-view. This may be done with the test plate either at isocenter or at the entrance surface of the image receptor. Figure 3: Photograph of the central test plate A- Iodine concentration check target (1 of 2) B- Iodine contrast detail target (1 of 4) C- Air test pin cylinder (1 of 4) D- Spatial resolution test plate E- Lead test pin (1 of 4) F- Alignment pin (1 of 4) G- Orientation Groove Spatial Resolution and Low Contrast Detectability A standard bar pattern insert is included in the central test plate (Figures 3,4). This pattern is used to test spatial resolution. A target with discreet bars was selected to minimize the influence of ailiasing on the observations. This 0.1 mm thick lead pattern provides sufficient contrast for use in the 30 cm phantom configuration. The lead content of this pattern has a small effect on the X-ray system's operating point. Four sets of holes with diameters (4, 3, 2, 1 mm) are filled with elemental iodine imbedded in epoxy. The areal densities of iodine in the four patterns are 20, 10, 5, 2.5 mg/cm 2. The observer is required to identify the smallest visible pair of targets in each pattern. Figure 4: Central test plate drawing The depth and iodine concentration of contrast targets are given Working Thickness Range Video and digital signals may be either white or black clipped. The working thickness range tool provides a measure of this effect. Aluminum is used as a bone substitute for this test. Air cylinders are used to simulate lung. A cylinder is resolved if the target (lead or air respectively) can be seen within the cylinder (Figures 5,6). The specifications for the cylinders are given in Table 1.

4 Benchmarking Fluroroscopes Balter Sep 2000 page 4 of 6 Figure 5: Photograph of the central sections of the phantom. Air cylinders (H), air test pins (C), aluminum cylinders (J) and lead test pins (E) are shown. Figure 7: Rotating spoke device at isocenter. Note that the drive motor is out of the field of view. Figure 6: Schematic drawing of the Working Thickness Range test. In this drawing there are two instances of black clipping and one instance of white clipping Table 1 Contents of working thickness range cylinders (200 mm total thickness phantom) A1* A2 A3 A4 M1 M2 M3 M4 mm of the specified material Air Al PMMA * A1 A4 and M1 M4 refer to the four air cylinders and four aluminum cylinders found in the phantom. Visibility of Moving Structures A rotating spoke target allows visual evaluation of motion unsharpness and the effects of temporal averaging (Figures 7,8). The device contains five steel wires of different diameters (0.56, 0.41, 0.30, 0.23, 0.13 mm). Two lead dots are used to evaluate lag and recursive filtering. Rotation speed is 30 revolutions per minute. The linear velocity of the outer lead dot is 200 mm/s. The rotating disk replaces the central test plate at isocenter. Figure 8: Fluorograph of rotating spoke wheel Lag is seen as a faint image angularly behind the current image. Motion unsharpness is more evident at larger radial distances Dosimetry Tools Dosimetry adapters allow measurement of the phantom entrance exposure (rate) at a standardized position in front of the entrance surface of the phantom (25 mm). The configuration of the phantom and probe location is shown in Figure 9. The reading is decreased because of loss of scatter from the phantom and increased because the measuring point is closer to the source than the entrance surface of the phantom. This position is considered an acceptable choice for this particular benchmarking phantom. The phantom can also be configured to generate the FDA measuring point (30 cm in front of the image receptor). Additional plastic and lead is added to drive the system

5 Benchmarking Fluroroscopes Balter Sep 2000 page 5 of 6 This paper introduces an example of a new class of phantoms for evaluating fluoroscopic systems. It is based on an International Electrotechnical Commission draft standard on dose and image quality for all X-ray imaging systems. It is noteworthy that the design of the phantom reported in this paper represents a consensus including cardiologists, medical physicists, service engineers, and technical staff from all of the major imaging companies. Figure 9: Phantom configuration for input dose measurements to maximum output. In this case, the dosemeter probe is in exactly the prescribed position. The readings will be increased above 'in-air' measurements due to scatter from the phantom. IMAGE ANALYSIS Ideally, all test images should be computer analyzed. This can be done in principle for those systems with fluoroscopic and fluorographic DICOM outputs. The necessary software tools are under discussion. A digital test solution is impossible given today s mix of analog and digital fluoro systems. The next best thing would be to incorporate tests that minimize operator bias. A forced-choice low-contrast detectability task is an example of such a test. Unfortunately, such a test is far too time consuming to incorporate in a phantom designed for rapid use. The Society for Cardiac Angiography and Interventions (SCA&I) initiated this project because of a need to reestablish a series of benchmarks for imaging system performance. Their intention is to establish a system performance registry based on phantom testing. A similar SCA&I registry, based on component testing, terminated in the early 1990s. Laboratories participating in this registry will receive reports showing their performance relative to the distributions in the entire database. The entire database (with due regard to the confidentiality of the original data) will be periodically published. Over time, this project should provide useful information for reference doses and for reference levels of imaging performance. The test results are dependent on the skill and training of the test operator. The best approach is to enroll testers in a formal training program before allowing them to work independently. This is precisely the route chosen by the FDA in their mammography testing regulations. Societies, such as SCA&I are expected to approve appropriate training programs for use of the NEMA phantom described in this paper. DISCUSSION

6 Benchmarking Fluroroscopes Balter Sep 2000 page 6 of 6 Further Reading Balter, S., and Shope, T.B. (ed) (1995) Syllabus: Physical and Technical Aspects of Angiography and Interventional Radiology, Radiological Society of North America, Oak Brook IL. Chotas, H. G., C. E. Floyd, Jr., et al. (1997). Quality control phantom for digital chest radiography. Radiology 202(1): Gould, R. G., B. Belanger, et al. (1980). Objective performance measurements versus perceived image quality in intensified fluoroscopic or photospot images. Radiology 137(3): Green, S., J. E. Palethorpe, et al. (1999). Performance assessment of patient dosimetry services and X-ray quality assurance instruments used in diagnostic radiology. Appl Radiat Isot 50(1): Holmes, D.R., Wondrow, M.A., et al. (1988) Variability of Radiation Output Dynamic Range in Modern Cardiac Catheterization Imaging Systems, Catherization and Cardiovascular Diagnosis 44: ICRU Report 54 Medical Imaging: The Assessment of Image Quality 1996 Jaffe, C. C., S. C. Orphanoudakis, et al. (1982). The effect of a television digital noise reduction device on fluoroscopic image quality and dose rate. Radiology 144(4): Kamm, K. F. (1997). The future of digital imaging. Br J Radiol 70 Spec No: S Kundel, H. L. (1979). Images, image quality and observer performance: new horizons in radiology lecture. Radiology 132(2): Lin, P. P. (1998). "A Comprehensive Quality Assurance Phantom For Cardiovascular Imaging Systems," in the Proceedings of Medical Imaging 1998 "Physics of Medical Imaging", Published by the Society of Photo-Optical Instrumentation Engineers, Bellingham, WA. Vol. 3336, pp Martin, C. J., D. G. Sutton, et al. (1999). Balancing patient dose and image quality. Appl Radiat Isot 50(1): Martin, C. J., P. F. Sharp, et al. (1999). Measurement of image quality in diagnostic radiology. Appl Radiat Isot 50(1): McParland, B. J. (1998). A study of patient radiation doses in interventional radiological procedures. Br J Radiol 71(842): Moore, R. (1993). Imaging Principals of Cardiac Angiography. Rockville (MD), Aspen. Nickoloff, E.L. and Strauss, K.J., (ed) (1998) Syllabus: Cardiac Catheterization Imaging, Radiological Society of North America, Oak Brook IL. Oresegun, M., J. LeHeron, et al. (1999). Radiation protection and quality assurance in diagnostic radiology--an IAEA coordinated research project in Asia and Eastern Europe. Appl Radiat Isot 50(1): Verdun, F. R., R. Moeckli, et al. (1996). Survey on image quality and dose levels used in Europe for mammography. Br J Radiol 69(824): Wagner, L. K., D. P. Fontenla, et al. (1992). Recommendations on performance characteristics of diagnostic exposure meters: report of AAPM Diagnostic X-ray Imaging Task Group No. 6. Med Phys 19(1): Wondrow, M. A., Laskey, W.L., et al. Cardiac Catheterization Laboratory Imaging Quality Assurance Program, accepted for publication in Catheterization and Cardiovascular Interventions.

Disclaimer and Notices.

Disclaimer and Notices. Fluoroscopy Dose Reduction and Clinical Management Fluoroscopy Trajectories (Curves), Global Dose Monitoring and Tracking of Patient Dose. Pei Jan Paul Lin, Ph.D. Division of Radiation Physics and Biology

More information

Radiography Curriculum Analysis

Radiography Curriculum Analysis Program Number Program Name Date / /20 Radiography Curriculum Analysis DIRECTIONS: Determine the course(s) in which each of the following content area is covered and enter the course number(s) and/or title(s).

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels 2033-7 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dose Reference Levels Peter Homolka EFOMP Training Course on Medical Physics in Diagnostic

More information

PATIENT DOSIMETRY IN INTERVENTIONAL RADIOLOGY

PATIENT DOSIMETRY IN INTERVENTIONAL RADIOLOGY 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 PATIENT DOSIMETRY

More information

New York State Department of Health Bureau of Environmental Radiation Protection

New York State Department of Health Bureau of Environmental Radiation Protection New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Program in Small Facilities Part II Fluoroscopic Equipment Introduction A.

More information

Quality Assurance for particle beam therapy

Quality Assurance for particle beam therapy Quality Assurance for particle beam therapy PTCOG Educational Workshop, Essen 2013 Heidelberg Ion Beam Therapy Center at the University Hospital German Cancer Research Center, Heidelberg, Germany Outline

More information

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy Jean-Pierre Bissonnette, PhD, MCCPM Disclosure Work supported, in part, by Elekta Oncology Systems Commercial Interest in Penta-Guide

More information

TENDER # SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH

TENDER # SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH TENDER # 0571-1707 SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH Amendment tice issued April 13,2017 1. The closing date time is amended to May 5 th,2017 2:00PM 2. The specifications

More information

Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide.

Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide. Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide. Fluoroscopy units used in angiographic and cardiac applications are normally

More information

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring CT QA SOLUTIONS Ensure Accurate Screening, Diagnosis and Monitoring ACCREDITATION ADVANCED CT IMAGE QUALITY QA AUTOMATIC EXPOSURE CONTROL CT CHARACTERIZATION MULTI-ENERGY CT CT PERFUSION QA COMPLETE CT

More information

Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children

Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children Mervyn D. Cohen, MB, ChB, MD Radiologists desire to keep radiation dose as low as possible. Pulsed fluoroscopy provides

More information

Making the difference with Live Image Guidance Dose management in Allura systems

Making the difference with Live Image Guidance Dose management in Allura systems DoseWise Allura Making the difference with Live Image Guidance Dose management in Allura systems The importance of radiation dose management The harmful effects of X-rays on human tissues had already become

More information

Fluoroscopy Testing Workshop

Fluoroscopy Testing Workshop Hands-on Fluoroscopy Testing Workshop 2-day Workshop February 24-25, 2018 in Wichita, KS 4.5 SAMs offered A continuing education division of HERZING UNIVERSITY Workshop Schedule ~~~~ Day One ~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~

More information

Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging

Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging Hanna Piwowarska-Bilska 1, Leszek J. Hahn 2, Bozena Birkenfeld 1, Katarzyna Cichon-Bankowska 1, Maria H. Listewnik 1, and Piotr

More information

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS Rutgers Robert Wood Johnson Medical School, New Brunswick, NJ Abstract

More information

Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services.

Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services. Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services. Action plan A program or method that explains the actions or steps to be taken. Advanced-practice

More information

E2E SBRT Thorax Phantom

E2E SBRT Thorax Phantom E2E SBRT Thorax Phantom Model 036A-CVXX-xx SBRT END-TO-END SBRT TESTING SOLUTION SCAN PLAN LOCALIZE TREAT This product is available through: JRT Associates 800-22-0 2428 Almeda Avenue Suite 36 Norfolk,

More information

Function of this control

Function of this control Page 1 of 7 Introduction The goals of this training are to develop a better appreciation of equipment controls, scatter radiation in the interventional suite, the radiation output capabilities of the angiography

More information

Introductory Presentation: EURAMED s Vision on Medical Radiation Protection (Research)

Introductory Presentation: EURAMED s Vision on Medical Radiation Protection (Research) Introductory Presentation: EURAMED s Vision on Medical Radiation Protection (Research) Christoph Hoeschen Otto-von-Guericke University Magdeburg on behalf of the executive board What is EURAMED?!!!Since

More information

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL 2 nd Edition

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL 2 nd Edition Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL 2 nd Edition New Jersey Department of Environmental Protection Bureau of X-ray Compliance PO Box 420, MC 25-01 Trenton NJ 08625-0420 FAX 609-984-5811

More information

Streamline workflow for higher productivity and patient throughput. DirectView. Classic CR System

Streamline workflow for higher productivity and patient throughput. DirectView. Classic CR System Streamline workflow for higher productivity and patient throughput. DirectView Classic CR System DirectView Classic CR System Improve your workflow and boost productivity with compact, single-cassette

More information

Radiation Physics and Safety Considerations in the EP Lab

Radiation Physics and Safety Considerations in the EP Lab Cardiorhythm 2009 Radiation Physics and Safety Considerations in the EP Lab Joel A. Kirsh, MD, FRCPC, CCDS Cardiology and Critical Care, Hospital for Sick Children Associate Professor of Pediatrics, University

More information

R/F. Experiences Using SONIALVISION safire and the Utility of Tomosynthesis. 1. Introduction. 2. Basics of Tomosynthesis.

R/F. Experiences Using SONIALVISION safire and the Utility of Tomosynthesis. 1. Introduction. 2. Basics of Tomosynthesis. R/F Experiences Using SONIALVISION safire and the Utility of Tomosynthesis Radiology Division, Dokkyo Medical University Koshigaya Hospital Masahiro Nakajima Mr. Masahiro Nakajima 1. Introduction The hospital

More information

Multi-source inspection of fiber-reinforced materials

Multi-source inspection of fiber-reinforced materials 19 th World Conference on Non-Destructive Testing 2016 Multi-source inspection of fiber-reinforced materials Tomasz CHADY 1, Krzysztof GORĄCY 1 1 West Pomeranian University of Technology, Szczecin, Poland

More information

DoseWise Portal. Take control of dose management. across your organization

DoseWise Portal. Take control of dose management. across your organization DoseWise Portal Take control of dose management across your organization Enable dose management across your organization Take control with DoseWise Portal Hospitals and health systems are discovering new

More information

Artis Q and ArtisQ.zen

Artis Q and ArtisQ.zen Artis Q and ArtisQ.zen Visionary GIGALIX GIGALIX Focused power Focused power Flat Emitter Conventional Filament Emitter Optimized focal spots Be certain Imaging excellence even with steep angulations enabled

More information

2017 ACR Computed Tomography Quality Control Manual FAQS

2017 ACR Computed Tomography Quality Control Manual FAQS Updated 11-15-2017 2017 ACR Computed Tomography Quality Control Manual FAQS Q. The updated 2017 ACR Computed Tomography Quality Control Manual has been released. (Visit www.acr.org/education/education-catalog.)

More information

Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp Advance Access publication 25 September 2008

Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp Advance Access publication 25 September 2008 Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp. 18 24 Advance Access publication 25 September 2008 doi:10.1093/rpd/ncn247 CORRELATION OF PATIENT MAXIMUM SKIN DOSES IN CARDIAC PROCEDURES WITH

More information

Take control of dose management

Take control of dose management DoseWise Portal Take control of dose management across your organization Key benefits Make data-informed decisions to improve patient care with support from dose tracking and advanced analytics Enhance

More information

Validation of an Ultrasonic-Phased-Array-Method for Testing of Circumferential Welds at Thin-walled Pipes

Validation of an Ultrasonic-Phased-Array-Method for Testing of Circumferential Welds at Thin-walled Pipes 19 th World Conference on Non-Destructive Testing 2016 Validation of an Ultrasonic-Phased-Array-Method for Testing of Circumferential Welds at Thin-walled Pipes Susanne HILLMANN 1, Frank SCHUBERT 1, David

More information

Progress in X-Ray & MR

Progress in X-Ray & MR Progress in X-Ray & MR Michiel Manuel Analyst Meeting June 15 th, 2005 X-Ray & MR: Agenda Introduction General X-Ray Cardio/Vascular X-Ray Magnetic Resonance China growth opportunity Conclusion 2 X-Ray

More information

Healthcare. Philips DoseWise Portal. Radiation exposure management software

Healthcare. Philips DoseWise Portal. Radiation exposure management software Healthcare Philips DoseWise Portal Radiation exposure management software The first step in building an effective radiation dose management strategy is collecting complete and accurate data so you can

More information

General Guidelines on Drop Size Measurement Techniques and Terminology

General Guidelines on Drop Size Measurement Techniques and Terminology General Guidelines on Drop Size Measurement Techniques As presented at the 47th Chemical Processing Industry Exposition, Javits Convention Center, New York, November 1997 Rudolf J. Schick Spray Analysis

More information

8/12/2011. Calculating Patient-specific Skin Dose. Call to Action! Fluoroscopy time poor measure of PSD Outline

8/12/2011. Calculating Patient-specific Skin Dose. Call to Action! Fluoroscopy time poor measure of PSD Outline Call to Action! Calculating Patient-specific Skin Dose William Pavlicek PhD Dept of Radiology Mayo Clinic Arizona pavlicek.william@mayo.edu 2011 MFMER slide-1 2011 MFMER slide-2 Fluoroscopy time poor measure

More information

GENESIS Edition. Transforming CT

GENESIS Edition. Transforming CT GENESIS Edition Transforming CT Transforming clinical confidence Transforming patient experience Transforming your workspace GENESIS Edition Transforming CT Brought to you by the leaders in area detector

More information

MEDICAL PHYSICS (MED PHYS)

MEDICAL PHYSICS (MED PHYS) Medical Physics (MED PHYS) 1 MEDICAL PHYSICS (MED PHYS) MED PHYS/PHYSICS 265 INTRODUCTION TO MEDICAL PHYSICS Primarily for premeds and other students in the medical and biological sciences. Applications

More information

A New Standard for Radiographic Acceptance Criteria for Steel Castings

A New Standard for Radiographic Acceptance Criteria for Steel Castings Paper 09-118.pdf, Page 1 of 11 A New Standard for Radiographic Acceptance Criteria for Steel Castings Copyright 2009 American Foundry Society M. Blair, R. Monroe Steel Founders Society of America, Crystal

More information

Physicists Quality Control for MR Equipment

Physicists Quality Control for MR Equipment Physicists Quality Control for MR Equipment Geoffrey D. Clarke, Ph.D. University of Texas Health Science Center at San Antonio 1 Overview ABR and the role of the Qualified Medical Physicist/ MR Scientist

More information

Low Dose, Lightning Speed, Latest Applications

Low Dose, Lightning Speed, Latest Applications RXL EDITION Low Dose, Lightning Speed, Latest Applications The New Standard for Radiology Toshiba Medical Systems is committed to the development of new technologies to minimize radiation dose while maintaining

More information

Delivering top performance. Outstanding value. Diagnostic scalability. ACUSON X150 Ultrasound System.

Delivering top performance. Outstanding value. Diagnostic scalability. ACUSON X150 Ultrasound System. Delivering top performance. Outstanding value. Diagnostic scalability. ACUSON X150 Ultrasound System www.siemens.com/medical Visualize more. Perform more effectively. Diagnose more confidently. Designed

More information

Taking the LED Pick and Place Challenge

Taking the LED Pick and Place Challenge Taking the LED Pick and Place Challenge Joshua J. Markle Cree, Inc. Joshua_markle@cree.com Abstract For the past few years there has been a shift in the Lighting Industry that has carried over to the surface

More information

Original article Paediatrics Today 2016;12(1):75-80 DOI /p

Original article Paediatrics Today 2016;12(1):75-80 DOI /p Original article Paediatrics Today 2016;12(1):75-80 DOI 10.5457/p2005-114.138 THE EFFECT OF ANTI-SCATTER GRID ON RADIATION DOSE IN CHEST RADIOGRAPHY IN CHILDREN Igor ŠABIČ 1*, Damjana KLJUČEVŠEK 1, Martin

More information

Vascular. Improvement of Functions Supporting Neuro-IVR. 1. Introduction. 2. Current State of Functions Supporting Neuro-IVR

Vascular. Improvement of Functions Supporting Neuro-IVR. 1. Introduction. 2. Current State of Functions Supporting Neuro-IVR Vascular Improvement of Functions Supporting Neuro-IVR Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction The use of techniques in which intravascular treatment is performed

More information

Types of Radiation Units

Types of Radiation Units RADIATION UNITS Types of Radiation Units Exposure in air - amount of ionization produced by or radiation in air. Unit - (R) or / (c/kg) Absorbed Dose energy transfer of / (unit of energy) of any. Used

More information

GENESIS Edition Transforming CT

GENESIS Edition Transforming CT GENESIS Edition Transforming CT 2 Transforming clinical confidence Transforming patient experience Transforming your workspace GENESIS Edition Transforming CT Brought to you by the leaders in area detector

More information

1 Introduction. 2 Material and methods. 2.1 Available methods. Key words: dose conversion factor, organ dose, effective dose, uncertainty

1 Introduction. 2 Material and methods. 2.1 Available methods. Key words: dose conversion factor, organ dose, effective dose, uncertainty Interpretation of measured dose data in X-ray imaging Toroi P. 1, 2, Kelaranta A. 1, Vock P. 2, Siiskonen T. 1, Tapiovaara M. 1 and Kosunen A. 1 STUK Radiation and Nuclear Safety Authority, Laippatie 4,

More information

Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy: Use of Grid-Controlled Pulsed Fluoroscopy

Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy: Use of Grid-Controlled Pulsed Fluoroscopy Giles W. L. Boland 1 Brian Murphy Ron Arellano Loren Niklason Peter R. Mueller Received January 24, 2000; accepted after revision May 3, 2000. Supported by a grant from Philips Medical Systems. 1 All authors:

More information

Patient doses in lower limb angiography in a recently installed flat detector x-ray system

Patient doses in lower limb angiography in a recently installed flat detector x-ray system Patient doses in lower limb angiography in a recently installed flat detector x-ray system Poster No.: C-0939 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper V. Tsapaki, D. Papavdis,

More information

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite Clinical Applications ImagingRite Neuro Intervention 1 ImagingRite ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in

More information

AIR QUENCHING OF ALUMINUM: THE AFFECT OF QUENCH ORIENTATION AND AIR VELOCITY

AIR QUENCHING OF ALUMINUM: THE AFFECT OF QUENCH ORIENTATION AND AIR VELOCITY AIR QUENCHING OF ALUMINUM: THE AFFECT OF QUENCH ORIENTATION AND AIR VELOCITY A Major Qualifying Project Report Submitted to the Faculty of WORCESTER POLYTECHNIC INSTITUTE In partial fulfillment of the

More information

Creating the future of Healthcare through meaningful innovation. Gene Saragnese, CEO Healthcare Imaging Systems

Creating the future of Healthcare through meaningful innovation. Gene Saragnese, CEO Healthcare Imaging Systems Creating the future of Healthcare through meaningful innovation Gene Saragnese, CEO Healthcare Imaging Systems 1 Key takeaways We are executing on our strategy Delivering a quality experience Demonstrated

More information

Prevention of radiation injury from medical imaging

Prevention of radiation injury from medical imaging Prevention of radiation injury from medical imaging Erica L. Mitchell, MD, a and Patricia Furey, MD, b Portland, Ore; and Manchester, NH Introduction: Concerns over radiation safety are valid. Understanding

More information

Motiva: a TV-based platform for remote patient management

Motiva: a TV-based platform for remote patient management Technology news New products Motiva: a TV-based platform for remote patient management a simplified user interface, a secure broadband connection and a standard set-top box. A nurse care manager, using

More information

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services Medical Physics and Informatics Review Vano et al. Online Quality Control for Diagnostic and Interventional Radiology Medical Physics and Informatics Review FOCUS ON: Eliseo Vano 1,2 Jose I. Ten 2,3 Jose

More information

Optimize efficiency with a universal system.

Optimize efficiency with a universal system. wins www.siemens.com/artis-one Optimize efficiency with a universal system.. Designed around you. Answers for life. Consistently efficient. Improve your competitiveness by mastering the balancing act between

More information

MEDICAL EQUIPMENT (1) TOPIC 1: RECORDING AND PROCESSING OF BIOSIGNALS

MEDICAL EQUIPMENT (1) TOPIC 1: RECORDING AND PROCESSING OF BIOSIGNALS MEDICAL EQUIPMENT (1) TOPIC 1: RECORDING AND PROCESSING OF BIOSIGNALS Term 1 2013/14 Prof. Yasser Mostafa Kadah www.k-space.org Measurement Basics Measuring is the experimental determination of a measured

More information

X-RAY EMISSION IN THE HIGH-CURRENT GLOW DISCHARGE EXPERIMENTS

X-RAY EMISSION IN THE HIGH-CURRENT GLOW DISCHARGE EXPERIMENTS Karabut, A.B. X-ray emission in the high-current glow discharge experiments. in The 9th International Conference on Cold Fusion, Condensed Matter Nuclear Science. 2002. Tsinghua Univ., Beijing, China:

More information

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

A NOVEL ELECTRICAL POTENTIAL DROP METHOD FOR THE DETECTION OF NAPHTHENIC ACID CORROSION IN OIL REFINING PROCESSING

A NOVEL ELECTRICAL POTENTIAL DROP METHOD FOR THE DETECTION OF NAPHTHENIC ACID CORROSION IN OIL REFINING PROCESSING A NOEL ELECTRICAL POTENTIAL DROP METHOD FOR THE DETECTION OF NAPHTHENIC ACID CORROSION IN OIL REFINING PROCESSING T. Batzinger, A May, C. Lester, K. Kutty and P. Allison. General Electric, Niskayuna, New

More information

McKesson Cardiology Bringing together all the cardiovascular information you need into a single platform

McKesson Cardiology Bringing together all the cardiovascular information you need into a single platform Care Providers Hospitals and Health Systems Enterprise Image and Information Management McKesson Cardiology Bringing together all the cardiovascular information you need into a single platform Catheterization

More information

Comprehensive Solutions

Comprehensive Solutions Comprehensive Solutions in Ultrasound-guided Breast Biopsy ATEC Vacuum-assisted Breast Biopsy System Celero Vacuum-assisted, Spring loaded Core Breast Biopsy Device Tru-Core II Spring loaded Core Breast

More information

RADIOGRAPHIC EVALUATION OF CORROSION AND DEPOSITS: IAEA CO-ORDINATED RESEARCH PROJECT ON LARGE DIAMETER STEEL PIPES

RADIOGRAPHIC EVALUATION OF CORROSION AND DEPOSITS: IAEA CO-ORDINATED RESEARCH PROJECT ON LARGE DIAMETER STEEL PIPES RADIOGRAPHIC EVALUATION OF CORROSION AND DEPOSITS: IAEA CO-ORDINATED RESEARCH PROJECT ON LARGE DIAMETER STEEL PIPES U. Zscherpel 1, I. Einav 2, S. Infanzon 3 and J. Zirnhelt 4 1 BAM, Berlin, Germany; 2

More information

Reduction of Backscattered Radiation in Enclosure X-ray Radiography

Reduction of Backscattered Radiation in Enclosure X-ray Radiography Abstract Reduction of Backscattered Radiation in Enclosure X-ray Radiography Samir Abdul-Majid, Abdulrahim Kinsara, Abdullah Almasoumi and Mohamed Kallothody Faculty of Engineering, King Abdulaziz University

More information

Draft Guidance for Industry Development and Use of Risk Minimization Action Plans

Draft Guidance for Industry Development and Use of Risk Minimization Action Plans Draft Guidance for Industry Development and Use of Risk Minimization Action Plans Docket Number [2004D-0188] Submitted to the U.S. Department of Health and Human Services Food and Drug Administration Center

More information

Title Using Neutron Radiography and Liqui. Author(s) Ito, Daisuke; Saito, Yasushi; Kawab. Citation Physics Procedia (2015), 69:

Title Using Neutron Radiography and Liqui. Author(s) Ito, Daisuke; Saito, Yasushi; Kawab. Citation Physics Procedia (2015), 69: Title Hybrid Two-phase Flow Measurements Using Neutron Radiography and Liqui Author(s) Ito, Daisuke; Saito, Yasushi; Kawab Citation Physics Procedia (2015), 69: 570-57 Issue Date 2015 URL http://hdl.handle.net/2433/215123

More information

Lubricating grease shear flow and boundary layers in a concentric. cylinder configuration

Lubricating grease shear flow and boundary layers in a concentric. cylinder configuration 45 3rd International Tribology Symposium of IFToMM, March in 2013, Luleå Lubricating grease shear flow and boundary layers in a concentric cylinder configuration J. X. LI 1, LARS G. WESTERBERG 2, E. HÖGLUND

More information

C-Arm Tables BIODEX. The Clinical Advantage. The table adjusts not you, not the patient. Innovative Contour Design Top

C-Arm Tables BIODEX. The Clinical Advantage. The table adjusts not you, not the patient. Innovative Contour Design Top The Clinical Advantage ISO 9001:2000 ISO 13485:2003 CERTIFIED Innovative Contour Design Top C-Arm Tables Surgical C-Arm Tables Pain Management C-Arm Table Urology C-Arm Table Brachytherapy C-Arm Table

More information

NOVEL NONDESTRUCTIVE EVALUATION TECHNIQUES FOR

NOVEL NONDESTRUCTIVE EVALUATION TECHNIQUES FOR NOVEL NONDESTRUCTIVE EVALUATION TECHNIQUES FOR INERTIA-FRICTION WELDS IN A SiC-REINFORCED HIGH-TEMPERATURE ALUMINUM ALLOY Peter B. Nagy, Laszlo Adler and William A. Baeslack III Department of Industrial,

More information

Contact Resistance Measurement System

Contact Resistance Measurement System Contact Resistance Measurement System Authors: Paul Tortora, Ph.D., Senior Engineer, Data Science Automation Quintin R. Stotts, Senior Measurement & Automation Consultant, Data Science Automation NI Products

More information

Mixed-Mode Crack Growth in Ductile Thin Sheet Materials

Mixed-Mode Crack Growth in Ductile Thin Sheet Materials Proceedings of the SEM Annual Conference June 1-4, 29 Albuquerque New Mexico USA 29 Society for Experimental Mechanics Inc. Mixed-Mode Crack Growth in Ductile Thin Sheet Materials J.H Yan 1, M. A. Sutton,

More information

Hybrid OR concepts Getinge and Toshiba Medical

Hybrid OR concepts Getinge and Toshiba Medical Hybrid OR concepts Getinge and Toshiba Medical Getinge and Toshiba Medical: Integrated solutions for the Hybrid OR A Hybrid OR pioneer: Getinge is one of the world s leading providers of medical products

More information

Introducing a new take on efficient workflow: Deeply integrated clinical applications

Introducing a new take on efficient workflow: Deeply integrated clinical applications Clinical Applications Introducing a new take on efficient workflow: Deeply integrated clinical applications The EXPERIENCE of TRUE CLINICAL APPLICATIONS While developments in image acquisition technology

More information

Benchmark & FAQ for NRG BR001: A Phase 1 Study of SBRT for the Treatment of Multiple Metastases

Benchmark & FAQ for NRG BR001: A Phase 1 Study of SBRT for the Treatment of Multiple Metastases Benchmark & FAQ for NRG BR001: A Phase 1 Study of SBRT for the Treatment of Multiple Metastases Credentialing Requirements (SECTION 5) Facility Questionnaire Benchmark planning 3DCRT credentialing IMRT

More information

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design GE Healthcare Introducing Veo * on Discovery * CT750 HD Great care by design Radiologists and clinicians demand the ultimate levels of clarity and detail in their images. In making an informed and confident

More information

Radiation Shielding Glass RD 30 RD 50

Radiation Shielding Glass RD 30 RD 50 Radiation Shielding Glass RD 30 RD 50 2 SCHOTT is a leading international technology group in the areas of specialty glass and glass-ceramics. With more than 130 years of outstanding development, materials

More information

Authors: (see complete list in Appendix A) History and comments: (see complete details in Appendix B) Preface

Authors: (see complete list in Appendix A) History and comments: (see complete details in Appendix B) Preface Diagnostic Radiology Residents Physics Curriculum AAPM Subcommittee of the Medical Physics Education of Physicians Committee May 2009 Supported by: AAPM Education Council and the Academic Council of the

More information

Joint Commission Medical Imaging Requirements. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting

Joint Commission Medical Imaging Requirements. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting Joint Commission Medical Imaging Requirements Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting My team of professionals 2 Medical Physicist s 3 FTE s (Board

More information

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children At Canada s premier pediatric hospital also recognized as one of the world s foremost pediatric healthcare institutions

More information

Optimization of anode-filter combinations in digital mammography: An experimental study

Optimization of anode-filter combinations in digital mammography: An experimental study Optimization of anode-filter combinations in digital mammography: An experimental study Poster No.: C-2915 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: A. Peruzzo Cornetto

More information

MANUFACTURING PROCESSES

MANUFACTURING PROCESSES 1 MANUFACTURING PROCESSES - AMEM 201 Lecture 8: Forming Processes (Rolling, Extrusion, Forging, Drawing) DR. SOTIRIS L. OMIROU Forming Processes - Definition & Types - Forming processes are those in which

More information

Investigating Copper Metallurgy Effects for Sort Process and Cleaning Performance Metrics

Investigating Copper Metallurgy Effects for Sort Process and Cleaning Performance Metrics Jan Martens NXP Semiconductors Germany Simon Allgaier Feinmetall GmbH Jerry Broz, Ph.D. International Test Solutions Investigating Copper Metallurgy Effects for Sort Process and Cleaning Performance Metrics

More information

Vascular and Interventional Radiology Original Research

Vascular and Interventional Radiology Original Research Vascular and Interventional Radiology Original Research Wiesinger et al. Stent Visibility and DAPs for FPD Versus II Angiography Machines Vascular and Interventional Radiology Original Research Benjamin

More information

Merge Unity HIPAA COMPLIANCE STATEMENT. Merge Healthcare 900 Walnut Ridge Drive Hartland, WI 53029

Merge Unity HIPAA COMPLIANCE STATEMENT. Merge Healthcare 900 Walnut Ridge Drive Hartland, WI 53029 Merge Unity Merge Healthcare 900 Walnut Ridge Drive Hartland, WI 53029 Copyright 20XX-20XX Merge Healthcare Incorporated, an IBM Company. The content of this document is confidential information of Merge

More information

Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations

Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations M. Río, S. Pérez, A. Herreros, A. Coll, A. Ruiz, J.Baró and J. Fernández Asesoría en Control

More information

Understanding Extrusion

Understanding Extrusion Chris Rauwendaal Understanding Extrusion 2nd Edition Sample Chapter 2: Instrumentation and Control ISBNs 978-1-56990-453-4 1-56990-453-7 HANSER Hanser Publishers, Munich Hanser Publications, Cincinnati

More information

2017 Global General Radiography Product Line Strategy Leadership Award

2017 Global General Radiography Product Line Strategy Leadership Award 2017 Global General Radiography Product Line Strategy Leadership Award Contents Background and Company Performance... 3 Industry Challenges... 3 Product Line Strength and Customer Impact... 3 Conclusion...

More information

Progress in Healthcare IT

Progress in Healthcare IT Progress in Healthcare IT Stan Smits Medical IT Philips Agenda The Healthcare IT market IT embedded in PMS products Philips is a leader in PACS Alliance with Epic for Enterprise Software 2 Challenges in

More information

Ultrasonic testing of adhesive bonds of thick composites with applications to wind turbine blades

Ultrasonic testing of adhesive bonds of thick composites with applications to wind turbine blades Center for Nondestructive Evaluation Conference Papers, Posters and Presentations Center for Nondestructive Evaluation 7-2011 Ultrasonic testing of adhesive bonds of thick composites with applications

More information

PRECISION MACHINING IN PLASTICS

PRECISION MACHINING IN PLASTICS FROM CONCEPT T O P R O T O T Y P E TO PRODUCTION Medical Manifolds Fine Writing Instruments Micro Fluidics Quality Assurance Phantoms Lenses & Light Guides PRECISION MACHINING IN PLASTICS Why choose Carville

More information

CARESTREAM DIRECTVIEW Max CR System

CARESTREAM DIRECTVIEW Max CR System CARESTREAM DIRECTVIEW Max CR System Boost CR productivity with a workflow leader. Maximize throughput, productivity, and patient satisfaction with the CARESTREAM DIRECTVIEW Max CR System. A next generation

More information

Finite Element Analysis of the Nerve Cuff to Determine Usability and Stress Analysis During Regular Use

Finite Element Analysis of the Nerve Cuff to Determine Usability and Stress Analysis During Regular Use Finite Element Analysis of the Nerve Cuff to Determine Usability and Stress Analysis During Regular Use Vivek Machhi Senior Project May 29, 2013 Table of Contents Abstact........2 Introduction..........3

More information

Image Analysis of Kevlar 49 Fabric at High Strain Rate

Image Analysis of Kevlar 49 Fabric at High Strain Rate Proceedings of the XIth International Congress and Exposition June -5, 8 Orlando, Florida USA 8 Society for Experimental Mechanics Inc. Image Analysis of Kevlar 49 Fabric at High Strain Rate Deju Zhu (),

More information

The entire document shall be read and understood before proceeding with a test. ISTA 3B Page 1 of 35

The entire document shall be read and understood before proceeding with a test. ISTA 3B Page 1 of 35 Packaged-Products for Less-Than-Truckload (LTL) Shipment ISTA 3 Series General Simulation Performance Test PROCEDURE VERSION DATE Last TECHNICAL Change: MARCH 2017 Last EDITORIAL Change: MARCH 2017 For

More information

Quality Control Concerns for New Generations of CT, MRI, & PET

Quality Control Concerns for New Generations of CT, MRI, & PET Quality Control Concerns for New Generations of CT, MRI, & PET David W. Jordan, Ph.D. Sr. Medical Physicist, UH Case Medical Center Clinical Assistant Professor, CWRU Disclosures No financial disclosures

More information