Lumen claims of the STERRAD 100NX Sterilizer: Testing performance limits of this technology

Size: px
Start display at page:

Download "Lumen claims of the STERRAD 100NX Sterilizer: Testing performance limits of this technology"

Transcription

1 Lumen claims of the STERRAD 100NX Sterilizer: Testing performance limits of this technology Magda Diab-Elschahawi, Alexander Blacky, Walter Koller Clinical Institute of Hospital Hygiene Medical University of Vienna

2 STERRAD 100NX Sterilizer

3 STERRAD 100NX Full Cycles Standard Cycle 47 minutes Sterilization of most surgical instruments including Inox lumened devices AND Flex Cycle 42 minutes Sterilization of flexible endoscopes only

4 Study Background A main drawback in the past has been the reliable sterilization of devices featuring long narrow lumens. According to the manufacturers information the STERRAD 100NX Sterilizer can adequately process 0.7mm X 500mm Inox lumened devices. We placed inoculated carriers at the midpoint position of lumens, which were then submitted to half-cycle sterilizing conditions in the STERRAD 100NX Sterilizer. In order to simulate insufficient cleaning of instruments or crystalline residues we added organic and inorganic challenges respectively to our inoculated carriers.

5 Aim of the study The aim of this study was to qualify the performance of the STERRAD 100NX Sterilizer under different experimental settings* representing worst case conditions. Protocol: 1. Stainless steel lumens of 0.7mm x 500mm 2. Wrapping (1 X) 3. Flex Scope conditions 4. Organic and inorganic burdens * These conditions do not correspond to the manufacturer s instructions for use

6 Test Conditions Manufacturer s IFU University Worst Case Test Condition Lumen Claim 0.7 mm X 500 mm Inox 0.7 mm X 500 mm Inox* Instrument Condition Must be clean and dry Contaminated with inorganic and organic material Load Arrangement Arranged pouches on edge Sterilization Cycle STANDARD FLEX Pouches stacked on top of each other * 0.7 mm X 500 mm Inox are validated for sterilization only in the STERRAD 100NX STANDARD cycle. Do not process these devices in the FLEX cycle.

7 Rationale for Worst Case Conditions Cleaning process not validated >> sub-standard washing or rinsing Simulated by use of inorganic and organic soils Human Error >>Improper loading Simulated by stacking pouched lumens on top of each other Human Error >> selecting the wrong cycle for the devices to be processed Simulated by use of FLEX cycle for sterilization of a rigid lumen

8 Test load Test load upper shelf Test load lower shelf

9 Manfacturer s Recommendation Place pouches on edge

10 Experimental setting

11 Titer verification BI (condition) Mean spore count/ml SD Final titer (mean) unchallenged 5,08E+00 ± 2,81 5,08E ppm 1,23E+01 ± 4,01 1,23E ppm 1,77E+01 ± 4,63 1,77E ppm 1,37E+01 ± 2,77 1,37E+06 5%FBS 1,88E+01 ± 4,07 1,88E+06

12 Qualitative Results condition position number untreated cycle 2 ng G ng ng G G ng ng ng ng cycle 3 ng G ng ng ng ng ng G ng ng cycle 4 ng G ng ng ng ng ng ng ng ng negative (%) 100% 0% 100% 100% 66% 66% 100% 66% 100% 100% 300ppm cycle 2 G G G G G G G G G G cycle 3 G G G G G G G G G G cycle 4 G G G G G G G G G G negative (%) 0% 0% 0% 0% 0% 0% 0% 0% 0% 0% 600ppm cycle 2 G ng ng G ng ng G G ng ng cycle 3 G G G G G G G G G G cycle 4 G G G G G G G G G G negative (%) 0% 33% 33% 0% 33% 33% 0% 0% 33% 33% 1200ppm cycle 2 G G G G G G ng G G G cycle 3 G G G G G G G G G G cycle 4 G G G G G G G G G G negative (%) 0% 0% 0% 0% 0% 0% 33% 0% 0% 0% 5% FBS cycle 2 G G G G G G G G G G cycle 3 G G ng G ng G G G G G cycle 4 ng G G ng G G G G G G negative (%) 33% 0% 33% 33% 33% 0% 0% 0% 0% 0%

13 log reduction/wire Quanitative results wires 7,00 6,00 5,00 4,00 3,00 2,00 untreated 300ppm 600ppm 1200ppm 5%FBS 1,00 0, position numbers

14 log reduction/bi Besides the protocol: Untreated wires: Heatshock vs. non Heatshock 7,00 6,00 5,00 4,00 untreated 3,00 untreated HS 2,00 1,00 0, position numbers

15 Discussion The titer determination of the inoculated wires in unchallenged condition provided by ASP showed a titer of only 5.08 x As our quantitative results showed that CFU counts of unchallenged wires found after sterilization were below detection limit, the actual log reduction rate can only be stated as Though one can assume, that a SAL of 10-6 could be achieved, a positive verification would have required a higher inoculation of the BI s.

16 Discussion We also cannot rule out that an effect known as inoculum effect would not influence the test results had we had a higher initial test inoculum. The inoculum effect refers to a decrease of the effect of a microbicide associated with a higher inoculum as described for antibiotics and certain disinfectants. The increase in bio-burden going along with an increasing inoculum size would decrease the amount of available sterilant per bacterial cell. Hence in our case H2O2 might fail to kill all test-organisms, thus resulting in inadequate sterilization especially in situations with heavy contamination and insufficient previous cleaning.

17 Discussion Inoculum effects must be considered relevant in all situations where biocides are not abundant in relation to microbial inocula.

18 Recommendations regarding the STERRAD 100NX strictly adhere to the manufacturer s recommendations specified in their user guide correct cycle (Standard or Flex Scope) and permitted dimensions for the processing of medical devices with lumens in the STERRAD 100NX Sterilizer.

19 Nelson DB. Infection control during gastrointestinal endoscopy. J Lab Clin Med 2003 Mar; 141(3): Babb JR, Bradley CR. Endoscope decontamination: where do we go from here? J Hosp Infect 1995 Jun; 30 Suppl: Ishino Y, Ido K, Sugano K. Contamination with hepatitis B virus DNA in gastrointestinal endoscope channels: risk of infection on reuse after on-site cleaning. Endoscopy 2005 Jun; 37(6): Kyi MS, Holton J, Ridgway GL. Assessment of the efficacy of a low temperature hydrogen peroxide gas plasma sterilization system. J Hosp Infect 1995; 31: Rutala WA, Gergen MF, Weber DJ. Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid. Am J Infect Control 1998 Aug; 26(4): Berube R, Oxborrow GS, Gastaud JW. Sterility testing: Validation of sterilization processes and sporicide testing. In: Block SS, Ed. Disinfection, Sterilization and Preservation, 5th edn. Lippincott Williams & Wilkins 2001; Anonymous. EN ISO Sterilization of health care products. General requirements for characterization of sterilizing agent and the development, validation and routine control of a sterilization process. Berlin: Beuth; Koller W and Lessky E. Microbiological test results and observations with an H2O2 plasma sterilizer. Zentr Steril 1996; 4: Pflug IJ. Microbiology and Engineering of Sterilization processes, 10th edn. Minneapolis, MN, USA: University of Minnesota Environmental Sterilization Laboratory Okpara J. Requirements fort the testing of sterilisation processes with Bacillus spores. Zentr Steril 1998; 6(2): Smith DF. Technical information. The Sterrad 100NX Sterilization System. ASP, Johnson and Johnson company. Chan-Myers H, McAlister D, Antonoplos P. Natural bioburden levels detected on rigid lumened medical devices before and after cleaning. Am J Infect Control 1997 Dec; 25(6): Alfa MJ, Olson N, Al-Fadhaly A. Cleaning efficacy of medical device washers in North American healthcare facilities. J Hosp Infect 2010 Feb; 74(2): Epub 2009 Aug 27. Chu NS, Favero M. The microbial flora of the gastrointestinal tract and the cleaning of flexible endoscopes. Gastrointest Endosc Clin N Am 2000 Apr; 10(2): Diab-Elschahawi M, Fürnkranz U, Blacky A, Koller W. Re-evaluation of current A0 value recommendations for thermal disinfection of reusable human waste containers based on new experimental data. J Hosp Infect 2010 May; 75(1):62-5. Deutsche Gesellschaft für Krankenhaushygiene: Mitteilung des Vorstandes der DGKH zum aktuellen Erkenntnisstand der Validierung des Sterrad/E-Plasma-Sterilisationsverfahrens mit den sich ergebenden Schlußfolgerungen für die Praxis. Hyg Med 1995; 20: Li RC and Ma HH. Parametrization of inoculum effect via mathematical modeling: aminoglycosides against Staphylococcus aures and Escherichia coli. J Chemother 1998 Jun; 10(3): Koller W, Wewalka G and Rotter M. Special features in the evaluation of aldehyde-free surface and instrument disinfectants. Hyg Med 1995; 20:8-18. Okpara-Hofmann J, Knoll M, Dürr M, Schmitt B, Borneff-Lipp M. Comparison of low-temperature hydrogen peroxide plasma sterilization for endoscopes using various Sterrad models. J Hosp Infect 2005; 59:

20 Thank you for your attention!

Challenging the STERRAD 100NX Sterilizer under experimental clean and dirty conditions

Challenging the STERRAD 100NX Sterilizer under experimental clean and dirty conditions Challenging the STERRAD 100NX Sterilizer under experimental clean and dirty conditions Magda Diab-Elschahawi, Alexander Blacky, Walter Koller Clinical Institute of Hospital Hygiene Medical University of

More information

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities Sterilization Practices in Healthcare Facilities Virtual Tour of Central Processing William A. Rutala, Ph.D., M.P.H. Director, Hospital Epidemiology, Occupational Health and Safety Program, UNC Health

More information

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope DEPARTMENT: Central Service/ Endoscopy APPROVED BY: EFFECTIVE: REVISED: 6/2016 PURPOSE: To test for detection

More information

TEST CODE DEFINITION STERRAD H202. Test Code Test Name Testing Description. Cycle. Efficacy. SE-200 STERRAD 200 Efficacy. Cycle. STD Cycle Efficacy

TEST CODE DEFINITION STERRAD H202. Test Code Test Name Testing Description. Cycle. Efficacy. SE-200 STERRAD 200 Efficacy. Cycle. STD Cycle Efficacy Code Name ing Description MC-100S SE-100S STERRAD 100S STERRAD 100S TEST CODE DEFINITION STERRAD H202 125 Highpower Road. Rochester, NY 14623 (888) -722-1529 www.highpowervtls.com Materials Compatibility

More information

Low Temperature Processing. An Update on Guidelines, Standards and Requirements

Low Temperature Processing. An Update on Guidelines, Standards and Requirements Low Temperature Processing An Update on Guidelines, Standards and Requirements Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve successful completion

More information

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation STERILIZATION VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial,

More information

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope SUBJECT: Detection of Blood residue inside the biopsy channel of a scope DEPARTMENT: Central Service/ Endoscopy APPROVED BY: EFFECTIVE: REVISED: 2/2012 PURPOSE: To test for detection of blood residue inside

More information

How is Sterilization Validation Performed by the Medical Device Manufacturer?

How is Sterilization Validation Performed by the Medical Device Manufacturer? How is Sterilization Validation Performed by the Medical Device Manufacturer? Rod Parker, Ph.D. Sr. Principal Scientist Stryker Instruments Division May 9, 2016 DISCLOSURE I am an employee of Stryker Corporation.

More information

Microfibre cleaning: is it all its been cracked up to be?

Microfibre cleaning: is it all its been cracked up to be? Microfibre cleaning: is it all its been cracked up to be? Disclaimers: Sponsored to give invited presenta/ons at various Na/onal and Interna/onal conferences by; STERIS, 3M, J&J, Healthmark, APIC, CACMID,

More information

disinfectionandsterilization.org

disinfectionandsterilization.org Best Practices for Cleaning, Disinfection and Sterilization in Healthcare Hosted by Benedetta Allegranzi WHO Technical Lead Clean Care is Safer Care Sponsored by WHO Patient Safety Challenge Clean Care

More information

Biological Monitoring

Biological Monitoring Biological Monitoring Prof. Duygu PERÇİN, MD Department of Clinical Microbiology Erciyes University Faculty of Medicine, Kayseri-TURKEY duygu.percin@hotmail.com Learning Objectives The attandee will be

More information

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities Understanding ANSI/AAMI ST 91 Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve

More information

ENDOSCOPE REPROCESSING AND INFECTION CONTROL FOR ENDOSCOPY

ENDOSCOPE REPROCESSING AND INFECTION CONTROL FOR ENDOSCOPY ENDOSCOPE REPROCESSING AND INFECTION CONTROL FOR ENDOSCOPY Assist. Prof. Pochamana Phisalprapa, M.D., M.Sc. Department of Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University GI ENDOSCOPY

More information

Elevator (Scope Issues) Going Up! Frank Myers, MA, CIC Infection Prevention Clinical Epidemiology UCSD Health

Elevator (Scope Issues) Going Up! Frank Myers, MA, CIC Infection Prevention Clinical Epidemiology UCSD Health Elevator (Scope Issues) Going Up! Frank Myers, MA, CIC Infection Prevention Clinical Epidemiology UCSD Health Objectives Identify two reasons why elevator scopes have a higher risk than other scopes of

More information

Clostridium difficile Update Dr. Michelle Alfa, Winnipeg Sponsored by ARJO

Clostridium difficile Update Dr. Michelle Alfa, Winnipeg Sponsored by ARJO Clostridium difficile: Update Dr. Michelle Alfa Ph.D, FCCM Diagnostic Services of Manitoba St. Boniface General Hospital Site Hosted by Paul Webber paul@webbertraining.com Sponsored by www.arjo.com www.webbertraining.com

More information

Update on Duodenoscope Related CRE Transmission. Disclosures. Outline 9/2/2016. None

Update on Duodenoscope Related CRE Transmission. Disclosures. Outline 9/2/2016. None Update on Duodenoscope Related CRE Transmission Mary Beth Graham, MD, FIDSA,FACP Professor of Medicine Medical College of Wisconsin None Disclosures Outline Overview of multidrug resistant organisms including

More information

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023 AS/NZS 4187 RELEVANT STANDARDS Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023 Relevant Standards AS 2514 AS 2774 AS 2773 AS 2773.1 Drying cabinets for medical equipment Drying cabinets

More information

Environmental disinfection: Bucket and mop to robots and rays.

Environmental disinfection: Bucket and mop to robots and rays. Environmental disinfection: Bucket and mop to robots and rays. A look at environmental decontamination and potential lessons from other sectors. Kevin Griffin Director Of Services Bioquell Asia Pacific

More information

Disinfectant Qualification A Multifaceted Study

Disinfectant Qualification A Multifaceted Study Disinfectant Qualification A Multifaceted Study In this article, Ziva Abraham, Microrite, Inc., gives us a brief overview of the disinfectant qualification study process. An effective cleaning and disinfection

More information

Medline Gemini Sterilization Wrap. Test Studies

Medline Gemini Sterilization Wrap. Test Studies Medline Gemini Sterilization Wrap Test Studies Steam Efficacy : Steam PreVac Steam Efficacy: Steam Gravity Shelf Life: 180 Days Plasma Efficacy: STERRAD 100NX Plasma Efficacy: STERRAD NX Plasma Efficacy:

More information

DEFENDO. Sterile Single-use Valves. Procedure

DEFENDO. Sterile Single-use Valves. Procedure DEFENDO Sterile Single-use Valves Procedure YOUR PARTNER IN THE FIGHT AGAINST HEALTHCARE-ASSOCIATED INFECTIONS (HAIs) Reducing the risk of infections for endoscopy patients is critically important. More

More information

NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish

NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish Evan Goulet, Ph.D., VP of Microbiology Applications Laboratory PDA Midwest: Sterility Assurance and Risk Mitigation October

More information

Endoscopy What s New Up There? Frank Myers, MA, CIC, FAPIC

Endoscopy What s New Up There? Frank Myers, MA, CIC, FAPIC Endoscopy What s New Up There? Frank Myers, MA, CIC, FAPIC Disclosures Frank Myers is a member of ASP J&J scientific advisory committee for the development of new high level disinfection or sterilization

More information

REUSABLE MEDICAL DEVICE VALIDATION

REUSABLE MEDICAL DEVICE VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial, sub-lethal, or repetitive

More information

Disinfection & Sterilization

Disinfection & Sterilization Disinfection & Sterilization Objectives Define important components in cleaning, disinfection, and sterilization Identify methods for cleaning, disinfection, or sterilization List risks of failure to properly

More information

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016 Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016 Ensuring our patients get clean, sterile instruments every time. Objectives Define the perioperative nurse and surgical tech

More information

Residual bioburden in reprocessed side-view endoscopes used for endoscopic retrograde cholangiopancreatography (ERCP)

Residual bioburden in reprocessed side-view endoscopes used for endoscopic retrograde cholangiopancreatography (ERCP) E12 Residual bioburden in reprocessed side-view endoscopes used for endoscopic retrograde cholangiopancreatography (ERCP) Authors D. L. N. L. Ubhayawardana 1, J. Kottahachchi 1, M. M. Weerasekera 1, I.

More information

They can be used on washer-disinfectors, ultrasonic cleaners, surgical instruments and endoscopes. Page 5

They can be used on washer-disinfectors, ultrasonic cleaners, surgical instruments and endoscopes. Page 5 Rev. 0 October 2017 Terragene offers a wide range of simple and effective products to precisely control the washing, hygiene and sterilization processes in veterinary clinics. Available in a variety of

More information

THE BASICS OF STERILIZATION

THE BASICS OF STERILIZATION THE BASICS OF STERILIZATION Objectives State the importance of sterilization to patient care Review three key essentials of the sterilization process Describe sterilization methods used, application, limitations,

More information

NEW Faster, more versatile!

NEW Faster, more versatile! NEW Faster, more versatile! > 35min flexible cycle > Process Single Channel surgical flexible endoscopes > Process Dual Channel surgical flexible endoscopes > Process 10.89 kg of Load including 1 surgical

More information

Disinfection, Sterilization and Antisepsis: An Overview

Disinfection, Sterilization and Antisepsis: An Overview Disinfection, Sterilization and Antisepsis: An Overview William A. Rutala, Ph.D., M.P.H. Director, Statewide Program for Infection Control and Epidemiology, Research Professor of Medicine, University of

More information

The Ethylene Oxide Sterilization Association, Inc. Managed by B&C Consortia Management, L.L.C. POSITION STATEMENT. May 20, 2016

The Ethylene Oxide Sterilization Association, Inc. Managed by B&C Consortia Management, L.L.C. POSITION STATEMENT. May 20, 2016 The Ethylene Oxide Sterilization Association, Inc. Managed by B&C Consortia Management, L.L.C. POSITION STATEMENT Use of Ethylene Oxide The Ethylene Oxide Sterilization Association, Inc. (EOSA) would like

More information

MDRAO ONLINE COURSE FOURTH EDITION 2018

MDRAO ONLINE COURSE FOURTH EDITION 2018 MDRAO ONLINE COURSE FOURTH EDITION 2018 Introduction MDRD Function and Structure Virtual tour of MDRD Standardized workflow Organizational Structure Chapter 1: Quality Systems Creating a Culture of quality

More information

Inactivation of Emerging Pathogens and Continuous Room Decontamination Technologies

Inactivation of Emerging Pathogens and Continuous Room Decontamination Technologies Inactivation of Emerging Pathogens and Continuous Room Decontamination Technologies William A. Rutala, Ph.D., M.P.H. Director, Statewide Program for Infection Control and Epidemiology, Research Professor

More information

Sterilization Policy. Georgia Regents Medical Center Policy Library. Policy Owner: Epidemiology POLICY STATEMENT

Sterilization Policy. Georgia Regents Medical Center Policy Library. Policy Owner: Epidemiology POLICY STATEMENT POLICY STATEMENT The ability to sterilize instruments and equipment for use during operative or other invasive procedures is critical to promoting successful patient outcomes and preventing infections.

More information

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ Disclaimer: The Society of Gastroenterology Nurses and Associates, Inc. assume no responsibility for the

More information

Plasma Science and Technology

Plasma Science and Technology ISSN:1009-0630 Plasma Science and Technology Volume 8 Number 4 July 2006 A Publication of Institute of Plasma Physics, Chinese Academy of Sciences and the Chinese Society of Theoretical and Applied Mechanics

More information

Ortho-phthalaldehyde: a possible alternative to glutaraldehyde for high level disinfection

Ortho-phthalaldehyde: a possible alternative to glutaraldehyde for high level disinfection Journal of Applied Microbiology 1999, 86, 1039 1046 Ortho-phthalaldehyde: a possible alternative to glutaraldehyde for high level disinfection S.E. Walsh, J-Y. Maillard and A.D. Russell Welsh School of

More information

Monitoring Flexible Endoscope Channels to Assure Cleaning

Monitoring Flexible Endoscope Channels to Assure Cleaning Monitoring Flexible Endoscope Channels to Assure Cleaning Dr. Michelle J. Alfa, Ph.D., FCCM Medical Director, Clinical Microbiology, Diagnostic Services of Manitoba, Winnipeg, Canada Disclaimers: Dr. M.

More information

VHP Sterilization and Prion Inactivation

VHP Sterilization and Prion Inactivation WFHSS Workshop Opatija Croatia March 27 th 2009 VHP Sterilization and Prion Inactivation Dr. Georgia Alevizopoulou Clinical Specialist Eastern Europe, Middle East & Africa VHP Sterilization and Prion Inactivation

More information

Monitoring your decontamination and reprocessing cycle

Monitoring your decontamination and reprocessing cycle Monitoring your decontamination and reprocessing cycle 1er Journee National sur la Sterilisation La Societe Marocaine de Sterilisation Maroc 2011 Ahmed Ghatwary/SteriGMS/Pharcomedic Index 1-Cleaning step,

More information

Raven Labs. Sterility Assurance for Industry. Sterility Assurance Products Since A division of Mesa Laboratories, Inc.

Raven Labs. Sterility Assurance for Industry. Sterility Assurance Products Since A division of Mesa Laboratories, Inc. Raven Labs A division of Mesa Laboratories, Inc. Sterility Assurance for Industry Sterility Assurance Products Since 1949 Industrial Use Biological Indicators Most co mmercially available Biologi cal Indicators

More information

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar MONITORING THE SYSTEM 1E PROCESSOR Q: I want to confirm that there is no need for a Biological Indicator (BI) and that use

More information

TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING

TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING A NEED FOR A COMPLETE PROCESS Lionel PINEAU PhD France PATHOGEN TRANSMISSION RELATED TO ENDOSCOPE WHAT IS THE CONTAMINATION RISK? Rates of bacteremia and infectious

More information

NEW Faster, more versatile!

NEW Faster, more versatile! NEW Faster, more versatile! > 35min flexible cycle > Process Single Channel surgical flexible endoscopes > Process Dual Channel surgical flexible endoscopes > Process 10.89 kg of Load Plus 1 surgical flexible

More information

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley ENDOSCOPE DECONTAMINATION Christina Bradley Hospital Infection Research Laboratory City Hospital NHS Trust Dudley Road, Birmingham B18 7QH Hosted by Paul Webber paul@webbertraining.com www.webbertraining.com

More information

Evaluation of Dry Fogging System for Microbial Inactivation

Evaluation of Dry Fogging System for Microbial Inactivation Evaluation of Dry Fogging System for Microbial Inactivation Carol Stansfield Senior Biosafety Officer Safety & Environmental Services Canadian Science Centre for Human and Animal Health, Winnipeg, Manitoba

More information

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy! Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy! Dr. Michelle J. Alfa, Ph.D., FCCM Medical Director, Clinical Microbiology, Diagnostic Services of Manitoba, Winnipeg, Canada Disclaimers:

More information

Disinfection Supply Center, the Affiliated Hospital of Yanan University, Yanan City, Shanxi, PR China. Abstract

Disinfection Supply Center, the Affiliated Hospital of Yanan University, Yanan City, Shanxi, PR China. Abstract Biomedical Research 2017; 28 (10): 4680-4687 ISSN 0970-938X www.biomedres.info Evaluation of the direct relationship between bacterial load on contaminated stainless steel surgical instruments and the

More information

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of Contents Glossary of equivalent standards... ix Committee representation... xi Acknowledgments...xiii Foreword... xiv Introduction: Need for the recommended practice...1 1 Scope...5 1.1 General...5 1.2

More information

QUALITY ASSURANCE IN ENDOSCOPY

QUALITY ASSURANCE IN ENDOSCOPY QUALITY ASSURANCE IN ENDOSCOPY Frank Edward Myers III, MA, CIC Kim Delahanty, BSN, MBA/HCM,CIC WHY DO QC ON SCOPES?? The scopes are clean no problem OBJECTIVES Describe the 2 advantages of a testing based

More information

Disinfection Qualification Testing

Disinfection Qualification Testing Disinfection Qualification Testing Considerations for Aseptic & Cleanroom Manufacturing Environments White Paper Considerations for Aseptic and Cleanroom Manufacturing Environments Obtaining the highest

More information

Sterile Services and Theatres Working in Harmony

Sterile Services and Theatres Working in Harmony Sterile Services and Theatres Working in Harmony BES Decon Reduces Hospital Costs BES Decon Improves SSD & Hospital Efficiencies BES Decon Increases SSD Revenues Sterile Services and Theatres: Two Important

More information

DISINFECTION QUALIFI CATION TESTING CONSIDERATIONS FOR THE ASEPTIC AND CLEANROOM MANUFACTURING ENVIRONMENT

DISINFECTION QUALIFI CATION TESTING CONSIDERATIONS FOR THE ASEPTIC AND CLEANROOM MANUFACTURING ENVIRONMENT DISINFECTION QUALIFI CATION TESTING CONSIDERATIONS FOR THE ASEPTIC AND CLEANROOM MANUFACTURING ENVIRONMENT by Dave Rottjakob, M.T. (ASCP) December, 2013 INTRODUCTION Obtaining the highest confidence that

More information

Infection Control. Low Temperature Sterilization (LTS) Critical Aspects on Selection and Validation of LTS-Processes

Infection Control. Low Temperature Sterilization (LTS) Critical Aspects on Selection and Validation of LTS-Processes Infection Control Low Temperature Sterilization (LTS) Critical Aspects on Selection and Validation of LTS-Processes Dr. Tronje von dem Hagen tronje@vondemhagen.com Lübeck (Germany) T. v. dem Hagen, Lübeck/Germany

More information

3/11/ M Sterile U Network 3M Sterile U Web Meeting March 11, Welcome! Topic: An Overview of ANSI/AAMI ST58: 2013

3/11/ M Sterile U Network 3M Sterile U Web Meeting March 11, Welcome! Topic: An Overview of ANSI/AAMI ST58: 2013 3M Infection Prevention Solutions 3M Sterile U Network 3M Sterile U Web Meeting March 11, 2014 Today s meeting times: 9:00 a.m., 11:00 a.m. and 1:00 p.m. CST To hear audio, call 800 937 0042 and enter

More information

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of Contents Glossary of equivalent standards... ix Committee representation... xi Acknowledgments...xiii Background on Amendment 1...xiii Foreword... xiv Introduction: Need for the recommended practice...1

More information

Controversies in Reprocessing Flexible Endoscopes: High Level Disinfection or Terminal Sterilization????

Controversies in Reprocessing Flexible Endoscopes: High Level Disinfection or Terminal Sterilization???? Controversies in Reprocessing Flexible Endoscopes: High Level Disinfection or Terminal Sterilization???? HONG KONG HOSPITAL AUTHORITY - ADVANCED COURSE FOR INFECTION CONTROL NURSES NOVEMBER 2017 1 Learning

More information

A comparison of the efficacy of lumen-cleaning devices for flexible gastrointestinal endoscopes

A comparison of the efficacy of lumen-cleaning devices for flexible gastrointestinal endoscopes A comparison of the efficacy of lumen-cleaning devices for flexible gastrointestinal endoscopes Australian Infection Control Timothy S Charlton B Sc (Hons), PhD Consultant, University of New South Wales

More information

Administrative Policies and Procedures. Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14 Date: Revised: 12/14 Attachment: None Page of 1 of 6

Administrative Policies and Procedures. Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14 Date: Revised: 12/14 Attachment: None Page of 1 of 6 Administrative Policies and Procedures Originating Venue: Infection Control Policy No.: IC 2306 Title: Cystoscope Reprocessing Policy & Procedure Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14

More information

Appendix 2. Revised 510(k) Summary

Appendix 2. Revised 510(k) Summary Appendix 2 Revised 510(k) Summary April 1, 2010 Appendix 2: Page 1 of 11 510(k) Summary For SYSTEM 1E Liquid Chemical Sterilant Processing System STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone:

More information

The Effect of Air Pockets on the Efficiency of Disinfection of Respiratory Equipment by Pasteurization

The Effect of Air Pockets on the Efficiency of Disinfection of Respiratory Equipment by Pasteurization CENORIN MEDICAL INFORMATION SERIES The Effect of Air Pockets on the Efficiency of Disinfection of Respiratory Equipment by Pasteurization Performed by: Bio Research Laboratories, Inc. Cenorin, LLC 6324

More information

Sterilization Service Revolution Service Enhancement in HA Hospitals

Sterilization Service Revolution Service Enhancement in HA Hospitals Sterilization Service Revolution Service Enhancement in HA Hospitals Hong Kong Hospital Authority Convention 2014 7 May 2014 LAW Tat Hong Samuel Cluster Operations Manager Central Sterile Supplies Department,

More information

Prevention of Transmission of the Superbug Carbapenem-Resistant Enterobacteriaceae (CRE) during Gastrointestinal Endoscopy

Prevention of Transmission of the Superbug Carbapenem-Resistant Enterobacteriaceae (CRE) during Gastrointestinal Endoscopy Prevention of Transmission of the Superbug Carbapenem-Resistant Enterobacteriaceae (CRE) during Gastrointestinal Endoscopy Sponsored by: Presentation by: Lawrence F. Muscarella, Ph.D. President LFM Healthcare

More information

Decontamination Effectiveness of Esco Celsafe CO2 Incubator Sterilization Feature using High Heat Temperature By Bekti Tri Sumaryati

Decontamination Effectiveness of Esco Celsafe CO2 Incubator Sterilization Feature using High Heat Temperature By Bekti Tri Sumaryati Decontamination Effectiveness of Esco Celsafe CO2 Incubator Sterilization Feature using High Heat Temperature By Bekti Tri Sumaryati 1. Summary Esco Celsafe CO 2 Incubator sterilization feature, which

More information

Dissecting Outbreaks of Multi-resistant Bacteria in Gastrointestinal Endoscopy

Dissecting Outbreaks of Multi-resistant Bacteria in Gastrointestinal Endoscopy Dissecting Outbreaks of Multi-resistant Bacteria in Gastrointestinal Endoscopy Dr. Michelle J. Alfa, Ph.D., FCCM Professor, Dept of Medical Microbiology, University of Manitoba, Winnipeg, Canada St Boniface

More information

Endoscopy and OR Solutions SYSTEM 1 TM. Express STERILE PROCESSING SYSTEM SAFETY, ASSURANCE AND EFFICIENCY. One Integrated Approach to Healthcare

Endoscopy and OR Solutions SYSTEM 1 TM. Express STERILE PROCESSING SYSTEM SAFETY, ASSURANCE AND EFFICIENCY. One Integrated Approach to Healthcare Endoscopy and OR Solutions SYSTEM 1 TM Express STERILE PROCESSING SYSTEM SAFETY, ASSURANCE AND EFFICIENCY One Integrated Approach to Healthcare Pioneer in liquid chemical sterilization Filtration System

More information

ADVANTAGE PLUS Pass-Thru

ADVANTAGE PLUS Pass-Thru ADVANTAGE PLUS Pass-Thru AUTOMATED ENDOSCOPE REPROCESSOR REPROCESSING Reprocessing Reimagined Cantel is your partner in the fight against healthcareassociated infections. Healthcare-associated infections

More information

Validation of cleaning and sterilization of reusable medical devices

Validation of cleaning and sterilization of reusable medical devices Synergy makes sence Validation of cleaning and sterilization of reusable medical devices MG-FSI72-119 Last revision: June 2012 Aurélien BIGNON 5, Chemin du Catupolan - 69120 Vaulx en Velin - France - Tel.

More information

Guidelines for Selection and Use of Disinfectants

Guidelines for Selection and Use of Disinfectants Guidelines for Selection and Use of Disinfectants Ref: (a) APIC Guidelines for Infection Control Practice, American Journal of Infection Control; April 1990, Vol 18, 99-113. To assist health care professionals

More information

Bio-Burden Reduction in Biological Laboratories

Bio-Burden Reduction in Biological Laboratories Bio-Burden Reduction in Biological Laboratories In biological laboratories samples worked with can harbor microorganisms that could be pathogens. Work is also done with microorganisms that are known pathogens.

More information

STS Directory Accreditation number: STS 0268

STS Directory Accreditation number: STS 0268 International standard: ISO/IEC 17025:2005 Swiss standard: SN EN ISO/IEC 17025:2005 Bioexam AG Maihofstrasse 95a P.O.B. 6858 CH-6000 Lucerne 6 Head: Candid Bucher Responsible for MS: Nathalie Amstein Telephone:

More information

Cleaning Efficiency of The MicroOne Wipe and Mop Cleaning System

Cleaning Efficiency of The MicroOne Wipe and Mop Cleaning System PDA Laboratories, Inc. A Contract ing Laboratory Visit us at www.pdalabsinc.com 17A Kiel Avenue Kinnelon, NJ 07405 (Tel) 973-492-8000 (Fax) 973-492-8002 pdainfo@pdalabsinc.com Cleaning Efficiency of The

More information

SUBJECT: Detection of residual organic soils inside various channels of lumened items.

SUBJECT: Detection of residual organic soils inside various channels of lumened items. ------------------------------------------------------------------------------------------------------------ NOTE: This document is an example policy that may be instituted in a health-care facility for

More information

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 05/2018 PURPOSE: To test for detection of blood residue on

More information

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue on

More information

5 Updates to Flexible Endoscope Reprocessing

5 Updates to Flexible Endoscope Reprocessing 5 Updates to Flexible Endoscope Reprocessing AORN presents and explains what it means for hospitals. Presented by: Sponsored by: Background Background Real-time video and small cameras led to the development

More information

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY Lesson No. CRCST 162 (Technical Continuing Education - TCE) Sponsored by: by Susan Klacik, BS, CRCST, ACE, CIS, FCS IAHCSMM Clinical Educator Validation and Verification LEARNING OBJECTIVES 1. Describe

More information

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS Lionel PINEAU PhD CONTENT General considerations Endoscope reprocessing, channel brushing, cleaning claims Description of the 2013 study

More information

Overview: High Level Disinfection

Overview: High Level Disinfection High Level Disinfection Who you gonna call?. Bug Busters! Dr. Michelle J. Alfa, FCCM St. Boniface General Hospital, Winnipeg, MB Disclaimers: Dr. M. Alfa - Sponsored to give invited presentations at various

More information

Stericool H 2 O 2. plasma sterilizers Affordable low temperature solution

Stericool H 2 O 2. plasma sterilizers Affordable low temperature solution Stericool H 2 plasma sterilizers Affordable low temperature solution Getinge Stericool H 2 plasma sterilizers Meeting everyday demands of the CSSD An affordable low temperature solution Since many components

More information

Is contamination of bronchoscopes really a problem?

Is contamination of bronchoscopes really a problem? Is contamination of bronchoscopes really a problem? 1 Is contamination of bronchoscopes really a problem? High Level Disinfection does not methodically clean bronchoscopes What we know 210 More than 210

More information

POLICY FOR THE DECONTAMINATION OF FLEXIBLE ENDOSCOPES

POLICY FOR THE DECONTAMINATION OF FLEXIBLE ENDOSCOPES POLICY FOR THE DECONTAMINATION OF FLEXIBLE ENDOSCOPES Version 2.0 August 2012 Name of Policy: Purpose of Policy: Policy for the Decontamination of Flexible Endoscopes To provide guidance on the decontamination

More information

PRODUCT D-VALUE STUDIES: A CRITICAL TOOL WHEN DEVELOPING A STERILIZATION PROCESS

PRODUCT D-VALUE STUDIES: A CRITICAL TOOL WHEN DEVELOPING A STERILIZATION PROCESS Volume 1, Number 3 June 2004 PRODUCT D-VALUE STUDIES: A CRITICAL TOOL WHEN DEVELOPING A STERILIZATION PROCESS Kurt McCauley R & D Lab Manager The objective of a sterilization process is to kill the naturally

More information

Best Practice Monitoring for Sterile Processing Professionals

Best Practice Monitoring for Sterile Processing Professionals Best Practice Monitoring for Sterile Processing Professionals Sterile processing departments are faced with greater challenges than ever. Increasing demands for quick turnarounds. Pressure to streamline

More information

Antimicrobial effectiveness evaluation of Isocide powder coating versus stainless steel plate

Antimicrobial effectiveness evaluation of Isocide powder coating versus stainless steel plate Antimicrobial effectiveness evaluation of Isocide powder coating versus stainless steel plate By: Dian Susanti and Alexander Atmadi Abstract Undesirable bioburdens can cause contamination on equipment

More information

One such problem is blood residual inside the channels of a robotic arm.

One such problem is blood residual inside the channels of a robotic arm. SUBJECT: Robotic Arm Testing for residue blood DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue inside and outside of a robotic arm and

More information

Indigo-Clean White Paper: #1 Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species. Healthcare

Indigo-Clean White Paper: #1 Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species. Healthcare Indigo-Clean White Paper: #1 Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species Healthcare HC Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species Clifford J. Yahnke,

More information

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms Learning Objectives Compare and contrast the available methods of sterilization for the healthcare setting Describe the appropriate use and monitoring for sterilization methods Purpose of Sterilization

More information

ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment

ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment 3M Attest TM TM Products Sterile U Network T U T O R I A L S ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment Background: A second amendment (A2:2009) to the Association for the Advancement of

More information

Endoscope Microbial Surveillance Testing Made Easy

Endoscope Microbial Surveillance Testing Made Easy Endoscope Microbial Surveillance Testing Made Easy Alpa Patel, B.S.RM (NRCM) Senior Scientist Nelson Laboratories Kaumudi Kulkarni, M.S. Manager of Research and Development Healthmark Industries Disclosure

More information

The importance of cleaning for the overall results of processing endoscopes

The importance of cleaning for the overall results of processing endoscopes Journal of Hospital Infection (2004) 56, S16 S22 www.elsevierhealth.com/journals/jhin The importance of cleaning for the overall results of processing endoscopes H. Martiny*, H. Floss, B. Zühlsdorf Charité

More information

applying a unique combination of disciplines consulting, preclinical, toxicology,

applying a unique combination of disciplines consulting, preclinical, toxicology, NAMSA is a Medical Research Organization (MRO), accelerating product development through integrated laboratory, clinical and consulting services. Driven by our regulatory expertise, NAMSA's MRO Approach

More information

Surgical Instruments - Instructions for use

Surgical Instruments - Instructions for use 1. General Notice Before using the instruments for the first time, as well as before each subsequent use and before returning them for repair or servicing, they must be cleaned, disinfected and sterilized

More information

BIOCOTE. BioCote treatment reduces contamination of laboratory equipment. Introduction

BIOCOTE. BioCote treatment reduces contamination of laboratory equipment. Introduction BIOCOTE BioCote treatment reduces contamination of laboratory Introduction Safety in the laboratory is based on the principles of good laboratory practice (GLP) (1). GLP is modified according to the type

More information

Room Decontamination with UV Radiation

Room Decontamination with UV Radiation infection control and hospital epidemiology october 2010, vol. 31, no. 10 original article Room Decontamination with UV Radiation William A. Rutala, PhD, MPH; Maria F. Gergen, MT (ASCP); David J. Weber,

More information

Validating reusable medical and dental devices for over 30 years

Validating reusable medical and dental devices for over 30 years Validating reusable medical and dental devices for over 30 years TABLE OF CONTENTS About Us 2 History Who We Are Today 3 Site Visits/Audits Certifications/Quality GLP/GMP Contact Us Protocols & Procedures

More information

HyperGrip Series Product Sterilization White Paper

HyperGrip Series Product Sterilization White Paper HyperGrip Series Product Sterilization White Paper 1 Contents 3 HyperGrip (HG) Series Product Sterilization 4 Autoclaving 4 EtO Sterilization 4 Sterrad 5 Tests on HyperGrip Connectors 6 Autoclave Sterilization

More information

Surface Compatibility and the Use of the SteraMist BIT Systems:

Surface Compatibility and the Use of the SteraMist BIT Systems: Surface Compatibility and the Use of the SteraMist BIT Systems: Written by: Dr. Helene Paxton, MS, MT(ASCP), PhD, CIC, Infection Preventionist, Bio Guidance, LLC. Surface Compatibility and the Use of the

More information

Surface Compatibility and the Use of the SteraMist BIT Systems:

Surface Compatibility and the Use of the SteraMist BIT Systems: Surface Compatibility and the Use of the SteraMist BIT Systems: Written by: Dr. Helene Paxton, MS, MT(ASCP), PhD, CIC, Infection Preventionist, Bio Guidance, LLC. Surface Compatibility and the Use of the

More information