1 Bacterial adherence to separated modular components in joint prosthesis. A clinical study. Gómez-Barrena E 1, Esteban J 2, Medel F 3, Molina-Manso D 2, Ortiz-Pérez A 2, Cordero- Ampuero J 4, Puértolas JA 3,5. 1 Department of Orthopaedic Surgery, IdiPaz-Hospital La Paz, Autónoma University of Madrid, Madrid (Spain). 2 Department of Clinical Microbiology, IIS-Fundación Jiménez Díaz. Madrid (Spain) 3 Department of Material Science and Technology, CPS-I3A, Zaragoza University, Spain 4 Department of Orthopaedic Surgery, Hospital La Princesa, Autónoma University of Madrid, Madrid (Spain). 5 Instituto Ciencia de Materiales de Aragón, ICMA, Zaragoza University-CSIC, Spain Corresponding author: Prof. Dr. Enrique Gómez Barrena Cirugía Ortopédica y Traumatología A Hospital La Paz, HRT 1ª planta Universidad Autónoma de Madrid Pº Castellana Madrid (Spain) Ph: Fax: enrique.gomezbarrena@uam.es
2 ABSTRACT (250 words) Background: Bacterial adherence on total joint replacement implants may lead to biofilm formation and implant-related osteoarticular infection. It is unclear if different biomaterials in the prosthetic components are more prone to facilitate this bacterial adherence, although polyethylene component exchange in the modular systems has been clinically utilized in the early management of these infections. To clarify if the amount of clinically adhered microorganisms was related to the material or the component, we investigated retrieved implants from infected joint replacements. Material and methods: 32 patients were revised after confirmed implant-related infection through positive cultures. A number of 87 total joint components (hip and knee) were obtained and separately sonicated after surgical retrieval, following a previously published protocol. Cultures were quantified, and detected CFUs were adjusted according to the component surface, and compared based on the component material and location. Results: Variable adherence of bacteria to chrome cobalt alloys, UHMWPE, hydroxyapatite coated components and titanium alloys. The commonest isolated organisms were Staphylococcus epidermidis (23 out of 87 components) and Staphylococcus aureus (10/87). Twelve components did not show any microorganism adhered despite location in an infected joint, with positive cultures in other components. A mixed linear model adjusted for random effects (the random effect being the infected patient) obtained convergence for the CFU/mm 2 variable but could not confirm a significantly higher adherence to a particular component or to a particular biomaterial. Therefore, the bacterial adherence primarily depends on the infective microorganism and the response of each individual patient, rather than materials or components.
3 INTRODUCTION Implant related infection is a dreadful menace to the survivorship and successful outcome of total joint replacement. Although the overall rate of implant infection is under 2-3% in most reports and registries [1], the total number of infected implants is high, and the clinical and economic consequences of significant importance [2]. Increasing interest of the Orthopaedic community is being placed on implant related infection in view of the yearly growing rates of total joint replacement both USA and European countries [3, 4] and future predictions [5, 6]. Bacterial adherence and subsequent biofilm formation on total joint replacement implants are at the origin and maintenance of implant-related osteoarticular infection [7, 8], and are currently considered the key phenomenon in the pathogenesis of these infections, with further implications in the diagnosis and management of the patients [6, 9, 10]. Efforts to diagnose the causative agent on colonized implants have recently been renewed with the introduction of retrieved implant sonication protocols in some laboratories [9, 11, 12]. With this technique, the presence of microorganisms has been diagnosed with higher sensitivity and specificity [12] in certain settings, and even a quantitative approach has been suggested as a new criterion for infection diagnosis [9]. Nevertheless, the component with the predominant bacterial adherence in the prosthetic joint is currently unknown, even if it may be the origin and the cause of initiating and maintaining the joint infection. Different biomaterials in the components are claimed to differently suffer from microorganism adherence and biofilm formation, even justifying clinical decisions such as component exchange in some cases of acute prosthetic joint infection. However, only experimental early studies have considered the preferred adherence of microorganisms to polymethylmetacrilate (PMMA) [13] compared with polyethylene, stainless steel and chromecobalt alloys. Chrome-cobalt alloys also showed experimentally a significantly higher infection susceptibility than titanium alloys [14], and rough titanium alloys were more prone to infection
4 than polished [15]. Yet clinical studies confirming or refuting a preferred bacterial adherence to a certain material or component are lacking. Although significant efforts are being placed by the material scientists to increase the biomaterial resistance to bacterial adherence [16], clinical data about the bacterial colonization of the different biomaterials that constitute the implanted prosthetic components are not available. This knowledge would certainly help in directing biomaterial scientists and implant designers towards the development of infection-resistant orthopaedic implants. In this study, we aimed to isolate and quantify the adherent microorganisms to each individual component retrieved from infected total hip and knee replacements, so as to analyze the different bacterial adherence to each one of the retrieved parts, and, in consequence, the different bacterial adherence to each biomaterial. MATERIAL AND METHODS Patients and samples A total of 87 total joint components (51 hip and 36 knee components) from 32 patients (20 hip and 12 knee arthroplasties) with clinical and microbiological diagnosis of implant-related infection were included in the study. Components under study included 6 femoral heads, 18 femoral stems, 14 metal cup shells, 13 acetabular liners, 9 femoral knee components, 4 allpolyethylene patellas, 11 tibial trays, and 12 tibial polyethylene components. Material in the component surface was chrome-cobalt (CC) alloys in 33, ultra-high molecular weight polyethylene (UHMWPE) in 27, hydroxyapatite (HA) in 17 (5 of which were fully coated), and titanium (Ti) alloys in 10. Sample processing
5 Patients were revised with a diagnosis of infected total joint (hip or knee) replacement at Hospital Fundación Jiménez Díaz or at Hospital La Princesa, both being University Hospitals in Madrid (Spain). All patients gave informed consent to the surgery and to the study of any obtained material from the joint. No antibiotic was used preoperatively, and any antibiotic given before admission were stopped at least 24h before surgery. Implants retrieved at revision surgery were gently irrigated to rinse rests of blood in the surgical table under sterile conditions, separated into parts, and individually placed in sterile bags (the part inside one bag, and this inside a second one that was handled to the circulating nurse). The material was then submitted to the Microbiology laboratory for processing. In case processing was not immediately available, the specimens were kept overnight at 4ºC in a freezer located at the surgical area during less than 24 hours. At the reference laboratory (FJD), samples were aseptically removed from the original bags and located into new sterile plastic bags using sterile material throughout the process. Fifty ml of buffer phosphate were added per component and bag, and bags were then closed. Samples were then sonicated according to a previously described protocol [12]. To avoid contamination problems, water in the sonicator was discharged and replaced after each sonication, and bags were also carefully investigated for any leakage before and after sonication, and the component discarded in case of any damage in the sterile bag. Samples were inoculated quantitatively (10 µl/plate) onto Tryptic-soy 5 % sheep blood agar (TS), Chocolate agar (CH), Schaedler 5 % sheep blood agar (SCH), McConkey agar (MC), Middlebrook 7H10 agar (MIDD) and Sabouraud-Chloramphenicol agar (SC), all from biomérieux (Marcy L Etoile, France). TS, CH, SCH and MC were incubated at 37ºC in 5 % CO2 atmosphere during 7 days, MIDD was incubated under the same conditions during 3 weeks, and SC was incubated at 30ºC in normal atmosphere during 4 weeks, as previously reported [12]. Cultures were quantified by counting the number of colonies that grew in the plate, adjusted to number of Colony Forming Units (CFU)/ml, and then adjusted to
6 CFU/sample. Isolated organisms were identified according to commonly reported criteria using biochemical tests (coagulase, oxidase) and commercial identification galleries (API System galleries, biomérieux, Marcy L Etoile, France). We defined mixed anaerobic microbiota as the presence of more than 2 different species of anaerobic bacteria. Antimicrobial susceptibility testing was performed using a disc-plate assay according to CLSI standards. Surface quantification The individual components of 6 retrieved joint implants (3 hip replacements and 3 knee replacements) were separately scanned using a Picza 3D Laser Scanner LPX-60 (Roland DG Corporation, Japan). All the components were scanned in the plane mode at 0.6 mm pitch, except for femoral heads, which were scanned in the rotary mode. 3D point cloud data were converted into polygon meshes using Dr. PICZA3 software for further file conversion and analysis. Measurements of the scanned surfaces (in mm 2 ) were obtained for each individual component working with PixformTM Pro software (Figures 4 and 5), and the average of measured components included different designs and sizes. Roughness or surface porosity varying among implants could slightly influence the surface in a micrometric scale, not the average measurements used in the study. Statistical analysis Considering the event of bacterial adherence an independent effect, descriptive and comparative (Kruskal-Wallis, Mann-Whitney, and Chi square tests) statistics were used (CFU/mm2 variables did not follow a normal distribution in the Kolmogorov-Smirnov test) in a SPSS 17.0 software (SPSS, Chicago IL, USA). However, in view that the event of bacterial adherence was not independent from the patient with the infection (different components are at risk of adherence in a single patient), mixed
7 linear models with random effects were prepared with SAS software 9.3 (SAS Institute Inc., Cary NC, USA), the infected patient being considered the random effect. For both TKA and THA components and for each of them, the models are adjusted for the number of CFU/mm2 and the component and the material were considered fixed effects. For samples with >100,000 CFU/ml we have used 100,000 CFU/sample for the different calculations. For samples with <500 CFU/sample we used 0 CFU/sample. RESULTS Components with positive cultures Culture was positive in 65 of the 87 components, of which 12 showed more than one microorganism. The average delay before processing was 7.6 ± 7.3 hours. The overall CFU/mm 2 averaged 2.45 with a SD of This occurred in 32 patients with one infected joint (single hip or knee). The etiology of these 32 infections was Gram-positive cocci in 23 cases, aerobic Gram-negative bacilli in 11 cases, anaerobes in 3 cases, Mycobacterium sp. in 2 cases and fungi in one case. Microbiological and component information is displayed in Table 1. S. epidermidis (present on 23 components from 9 patients) and S. aureus (present on 11 components from 5 patients) were the most frequently isolated microorganisms. Polymicrobial isolates were found with S. epidermidis, S. aureus, P. aeruginosa and S. lugdunensis. Components with negative cultures Culture of sonicate was negative in 18 components retrieved from 12 patients (5 infected hips and 7 infected knees). Among these 18 components without isolated bacteria, 9 of them were UHMWPE components (out of 27 UHMWPE components in the whole series), 7 manufactured with chrome cobalt alloys (out of 33 chrome cobalt components), 2 HA coated (from 17 HA coated components), and none of titanium alloys (all 10 titanium alloy components in the series were found with adherent microorganisms). Maximum bacterial counts in the 5 infected hips
8 with at least one component without bacteria were found in stems and cups of chrome-cobalt, and titanium alloys with and without HA coating.the highest bacterial counts in the 7 infected knees with at least one component without bacteria were found in chrome-cobalt tibial trays. Table 2 displays a microbiological descriptive analysis of infected joints that presented one or two components without microorganisms after sonication and culture. Influence of prostheses type, component and biomaterial The distribution of isolated bacteria per surface unit of implant (CFU/mm 2 ) is shown in Figure 1, separating hips and knees in our series. The distribution of individual components is shown in Figure 2; significant differences were found in the adhered CFU/mm 2 among different components (p=0.018, Kruskal-Wallis). The distribution separated for the different biomaterials is shown in Figure 3; the presence of positive cultures was also different among different biomaterials when independently considered (p=0.005, Kruskal-Wallis). Figures 1, 2, and 3 display the great variability found in bacterial adherence. However, taking into account that these were not independent events, but rather related to specific microorganisms infecting individual patients, the association of different adherence per surface (CFU/mm 2 ) to components and materials was studied individually within each infected joint and patient through a mixed linear model, where the random effect was the patient sustaining the infection. Models for all components, for hip components alone and for knee components alone were run, complying with the convergence criterion after 3 iterations. After statistical analysis by means of this linear model, both component type and biomaterial type failed to prove a significant difference in the conjoined and the one-to-one comparison. These final results confirmed that fixed effects such as component type or biomaterial in the component did not differ when a powerful random effect generated by an individual infected patient with a specific microorganism was considered.
9 DISCUSSION The development of implant sonication as a diagnostic tool for prosthetic joint infection may be an important advance in the management of the patients, increasing the sensitivity of conventional techniques. The knowledge of the etiology of the infection is of great importance because it allows selecting the best possible antimicrobial therapy [9, 10]. Sonication even allows a quantitative assay that, theoretically, could lead to establish quantitative criteria for a definitive evaluation of the microbiological results [6, 9, 11]. Nevertheless, some data suggest that clinical infection could be diagnosed with low bacterial counts [12], and even in some cases no organisms can be detected using techniques based on conventional cultures, so other approaches, like detection of organisms using molecular biology techniques, will be necessary [17]. Many factors could influence these results, but delay in processing (if samples were properly refrigerated) had not shown a statistically significant effect on the recovery of organisms [12]. Results of our sonication of disassembled components were statistically analyzed by means of a mixed linear model and highlight the dominant influence of each joint infection in each individual patient when analyzing bacterial counts. In spite of this conclusion about the prevailing effect of the individual patient and microorganism, hip infections included in our series were apparently more severe cases with more isolated microorganisms per surface unit when compared to knees (Figure 1). Moreover, hip infections produced almost no bacterial adherence to femoral head components (Figure 2), while knee infections basically seeded on the tibial trays in our series (Figure 2). Additionally, bacterial adherence on each type of biomaterial (Figure 3) also displayed a trend towards less adherence to UHMWPE in the knee and to HA in the hip retrieved implants, as suggested by independent comparisons. As a study limitation, we observed the high variability of microorganisms and patients that did not permit in our series to conclude any further. Another
10 limitation could be the lack of knowledge about the sonication effect in the different materials found in clinical samples. However, sonication has been largely used with in vitro and in vivo models of bacterial adherence and biofilm [18-21], without proving differences among materials. We performed fluorescent stains after sonication (as a test of sonication efficacy in titanium alloy and polyethylene, data not published) and no remaining bacteria were detected. Our results hold a relevant research consequence, as many new proposals of materials with a supposed resistance to infection based on blocking the adherence of certain microorganisms to a modified material may be unsupported. The variability in the adherence of microorganisms even from the same species is probably too high to conclude on the efficacy of a material modification without testing numerous bacteria strains of the same species, also considering that the patient susceptibility may be influential too. Results of our study also confirm that clinical studies require a large number of infection cases to consider the epidemiology and pathogenesis of different strains from each species under scrutiny. This is the only way to assess the potential risks. Furthermore, the isolated role of biomaterials and of components cannot be assessed without taking into account, at least statistically, the severity of the infection, the patient susceptibility, and the pathogenicity of the infecting microorganisms. Of particular interest is the fact that our clinical series did not confirm basic experimental knowledge [13] about microorganisms being more adherent to polymers, which supposedly could present a higher risk of infection than metals. Furthermore, experimental infection may not mimic clinical variability because each particular bacterial strain used in any experimental study may show a high or a low adherence to a certain material or component while another strain of the same species may show different behaviour. This intraspecies variability is clearly established and is the reason for a recent trend that uses both collection and clinical strains of
11 microorganisms when performing a laboratory test of antibacterial activity of biomaterials [21], so as to achieve more realistic results. This conclusion could have also relevant clinical consequences, as the polyethylene exchange performed in cases of early infection to decrease the microbiological load is unsupported. The substituted polyethylene component may even not hold adhered microorganisms, so unless all the components are revised and analyzed by sonication, it is impossible to know where most of the adhered bacteria are. Therefore, large, multicentric series are required to definitely confirm the role, even secondary, of materials and components in the total joint infection. Despite its limitations, the major determinants in our series were not materials or components, but the patient undergoing a particular infection, and the pathogen microorganism. ACKNOWLEDGMENTS Dr. Rosario Madero and Elia Pérez, statisticians at the Hospital La Paz, directed and performed the statistical analysis on the mixed linear models. DMM was funded by a grant from the Fundacion Conchita Rabago de Jimenez Diaz. This study was partly funded by a grant from the Spanish Ministry of Science (FUNCOAT-CSD ). FIGURE LEGENDS: Figure 1: Distribution of isolated bacteria per surface unit (CFU/mm 2 ) separated for hip and knee implants (mean with error bars showing SD). Figure 2: Distribution of isolated bacteria per surface unit (CFU/mm 2 ) separated for type of component (mean with error bars showing SD). Figure 3: Distribution of isolated bacteria per surface unit (CFU/mm 2 ) separated for type of biomaterial and implant (mean with error bars showing SD). Figure 4: Femoral stem component (Versys, Zimmer, Warsaw IN, USA) result of surface measurement analysis. Figure 5: Femoral component of a knee system (Scorpio, Stryker, Kalamazoo MI, USA) result of surface measurement analysis.
12 REFERENCES 1. Gomez E, Patel R. Laboratory Diagnosis of Prosthetic Joint Infection, Part I. Clinical Microbiology Newsletter. [doi: /j.clinmicnews ] 2011;33(8): Darouiche RO. Treatment of Infections Associated with Surgical Implants. New England Journal of Medicine 2004;350: Kurtz SM, Ong KL, Schmier J, et al. Primary and revision arthroplasty surgery caseloads in the United States from 1990 to J Arthroplasty 2009 Feb;24(2): Kurtz SM, Ong KL, Lau E, et al. International survey of primary and revision total knee replacement. Int Orthop. [ /s [doi]] Mar 15;[Epub ahead of print] PMID: Kurtz SM, Lau E, Ong K, et al. Future young patient demand for primary and revision joint replacement: national projections from 2010 to Clin Orthop Relat Res 2009 Oct;467(10): Gomez E, Patel R. Laboratory Diagnosis of Prosthetic Joint Infection, Part II. Clinical Microbiology Newsletter. [doi: /j.clinmicnews ] 2011;33(9): Costerton JW. Biofilm theory can guide the treatment of device-related orthopaedic infections. Clin Orthop Relat Res 2005 Aug(437): Gristina AG. Biomaterial-centered infection: Microbial adhesion versus tissue integration. Science 1987;237: Del Pozo JL, Patel R. Clinical practice. Infection associated with prosthetic joints. N Engl J Med 2009 Aug 20;361(8): Esteban J, Cordero-Ampuero J. Treatment of prosthetic osteoarticular infections. Expert Opin Pharmacother 2011 Apr;12(6): Trampuz A, Piper KE, Jacobson MJ, et al. Sonication of removed hip and knee prostheses for diagnosis of infection. N Engl J Med 2007 Aug 16;357(7):
13 12. Esteban J, Gomez-Barrena E, Cordero J, et al. Evaluation of quantitative analysis of cultures from sonicated retrieved orthopedic implants in diagnosis of orthopedic infection. J Clin Microbiol 2008 Feb;46(2): Petty W, Spanier S, Shuster JJ, et al. The influence of skeletal implants on incidence of infection. Experiments in a canine model. J Bone Joint Surg Am 1985 Oct;67(8): Cordero J, Munuera L, Folgueira MD. The influence of the chemical composition and surface of the implant on infection. Injury 1996;27 Suppl 3:SC Cordero J, Munuera L, Folgueira MD. Influence of metal implants on infection. An experimental study in rabbits. J Bone Joint Surg Br 1994 Sep;76(5): Katsikogianni M, Missirlis YF. Concise review of mechanisms of bacterial adhesion to biomaterial and of techniques used in estimating bacteria-material interactions. European Cells and Materials 2004;8: Achermann Y, Vogt M, Leunig M, et al. Improved diagnosis of periprosthetic joint infection by multiplex PCR of sonication fluid from removed implants. J Clin Microbiol 2010 Apr;48(4): Kinnari TJ, Esteban J, Gomez-Barrena E, et al. Bacterial adherence to SiO2-based multifunctional bioceramics. J Biomed Mater Res A 2009 Apr;89(1): Kinnari TJ, Esteban J, Zamora N, et al. Effect of surface roughness and sterilization on bacterial adherence to ultra-high molecular weight polyethylene. Clin Microbiol Infect 2010 Jul;16(7): Esteban J, Molina-Manso D, Spiliopoulou I, et al. Biofilm development by clinical isolates of Staphylococcus spp. from retrieved orthopedic prostheses. Acta Orthop 2010 Dec;81(6): Gomez-Barrena E, Esteban J, Molina-Manso D, et al. Bacterial adherence on UHMWPE with vitamin E: an in vitro study. J Mater Sci Mater Med 2011 Jul;22(7):
Use of Quantitative Scanning Electron Microscopy of S. aureus Biofilm Formation in vitro to Identify Strain and Implant Material Specificities
Use of Quantitative Scanning Electron Microscopy of S. aureus Biofilm Formation in vitro to Identify Strain and Implant Material Specificities Werasak Sutipornpalangkul, MD, PhD 1, Kohei Nishitani, MD,
More informationFOCUS ON MRSA/SA SSTI ASSAY FAILURE IN PROSTHETIC JOINT. Institute of Microbiology, Lille University Hospital, F Lille, France
JCM Accepted Manuscript Posted Online 16 November 2016 J. Clin. Microbiol. doi:10.1128/jcm.01658-16 Copyright 2016, American Society for Microbiology. All Rights Reserved. FOCUS ON MRSA/SA SSTI ASSAY FAILURE
More informationInt.J.Curr.Microbiol.App.Sci (2014) 3(10)
ISSN: 2319-7706 Volume 3 Number 10 (2014) pp. 810-815 http://www.ijcmas.com Original Research Article Comparison of Tissue Culture plate method, Tube Method and Congo Red Agar Method for the detection
More informationEvaluation of a Novel Non-Sterile Glove Dispensing System Uyen Nguyen
Evaluation of a Novel Non-Sterile Glove Dispensing System Uyen Nguyen Edited by: Michael Diamond Reviewed by: Andrew Duong Disclaimer: The author of this report and declare no conflict of interest with
More informationComparison of the speed of kill of pathogenic bacteria using ACTICOAT and AQUACEL Ag analysed using Confocal Laser Scanning Microscopy
Smith & Nephew Wound Management Data on file report - 0505004 May 2005 Comparison of the speed of kill of pathogenic bacteria using ACTICOAT and AQUACEL Ag analysed using Confocal Laser Scanning Microscopy
More informationBiofilm Protocol Optimization For Pseudomonas aeruginosa. Introduction. Materials and Methods. Culture Media, Incubation Time, and Biofilm Measurement
Biofilm Protocol Optimization For Pseudomonas aeruginosa Culture Media, Incubation Time, and Biofilm Measurement Introduction In addition to the conventional arsenal of antibiotic resistance mechanisms
More informationGENUS STAPHYLOCOCCUS: Isolation and Identification
GENUS STAPHYLOCOCCUS: Isolation and Identification Staphylococcus is a genus of Gram +, nonspore-forming cocci belonging to the family Micrococcaceae that are often found as normal human microbiota of
More informationBIP Central Venous Catheter
Bactiguard Infection Protection BIP Central Venous Catheter For prevention of healthcare associated infections Bactiguard benefits Reduced healthcare costs Reduced use of antibiotics Save lives Bloodstream
More informationBD GeneOhm MRSA ACP Lysis Kit
REF 441638 100 Tests DOPS09-09-V1E1 USA Date: 2010-01 TABLE OF CONTENTS INTENDED USE...3 SUMMARY AND EXPLANATION OF THE PROCEDURE...3 REAGENTS...3 PRECAUTIONS...3 MATERIALS PROVIDED...4 STORAGE, HANDLING
More informationRapid Bacterial Identification Using a Mass Spectrometry Based Molecular Diagnostics Approach: Evaluation of the Iridica Platform
Rapid Bacterial Identification Using a Mass Spectrometry Based Molecular Diagnostics Approach: Evaluation of the Iridica Platform Alec Saitman, PhD, Jane Y. Yang PhD, Sharon Reed, David Pride, Michele
More informationInoculate: Media. Physical State of Media: Liquid. The Five I s: Basic Techniques to Culture Microbes Tools of the Microbiology Laboratory
The Five I s: Basic Techniques to Culture Microbes Tools of the Microbiology Laboratory 1. Inoculate 2. Incubate 3. Isolate 4. Inspect 5. Identify The Five I s: Inoculate Inoculate: Media Classified according
More informationá61ñ MICROBIOLOGICAL EXAMINATION OF NONSTERILE PRODUCTS: MICROBIAL ENUMERATION TESTS
USP 40 Microbiological Tests / á61ñ Microbiological Examination 1 á61ñ MICROBIOLOGICAL EXAMINATION OF NONSTERILE PRODUCTS: MICROBIAL ENUMERATION TESTS INTRODUCTION The tests described hereafter will allow
More informationIndigo-Clean White Paper: #1 Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species. Healthcare
Indigo-Clean White Paper: #1 Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species Healthcare HC Bactericidal Performance Testing of Indigo-Clean Upon Bacterial Species Clifford J. Yahnke,
More informationFinal text for addition to The International Pharmacopoeia
March 2012 3.3.1 MICROBIOLOGICAL EXAMINATION OF NON-STERILE PRODUCTS: MICROBIAL ENUMERATION TESTS Final text for addition to The International Pharmacopoeia This monograph was adopted at the Forty-sixth
More informationDuraCer. Ceramic Components. The Strong Choice In Ceramics
The Strong Choice In Ceramics is a magnesia-stabilized zirconia (MgPSZ) ceramic material whose combination of physical properties is attractive for use in orthopaedic surgery. Mg-PSZ ceramic often is referred
More informationDetection of Prosthetic Hip Infection at Revision Arthroplasty by Immunofluorescence Microscopy and PCR Amplification of the Bacterial 16S rrna Gene
JOURNAL OF CLINICAL MICROBIOLOGY, Oct. 1999, p. 3281 3290 Vol. 37, No. 10 0095-1137/99/$04.00 0 Copyright 1999, American Society for Microbiology. All Rights Reserved. Detection of Prosthetic Hip Infection
More informationSonication of Removed Hip and Knee Prostheses for Diagnosis of Infection
T h e n e w e ng l a nd j o u r na l o f m e dic i n e original article Sonication of Removed Hip and Knee Prostheses for Diagnosis of Infection Andrej Trampuz, M.D., Kerryl E. Piper, M.S., Melissa J.
More informationABC. Methods for Determining Bactericidal Activity of Antimicrobial Agents; Approved Guideline. Volume 19 Number 18
M26-A ISBN 1-56238-384-1 September 1999 ISSN 0273-3099 Methods for Determining Bactericidal Activity of Antimicrobial Agents; Approved Guideline Volume 19 Number 18 Arthur L. Barry, Ph.D. William A. Craig,
More informationPersonal Check Money Order Credit Card Visa MC Discover Amex LYMPHOCYTE TRANSFORMATION TESTING (LTT) AND/OR METAL ION TESTING SERVICE (ALL FIELDS REQUIRED) Patient Last, First Name, M.I. Mail or e-mail
More informationMicrobicidal Properties of a Silver-Containing Hydrofiber Dressing Against a Variety of Burn Wound Pathogens
Microbicidal Properties of a Silver-Containing Hydrofiber Dressing Against a Variety of Burn Wound Pathogens P. G. Bowler, MPhil, S. A. Jones, BSc, M. Walker, PhD, D. Parsons, PhD Partial-thickness burns
More informationINSTRUCTIONS FOR USE LYFO DISK KWIK-STIK KWIK-STIK PLUS. n LYFO DISK Microorganisms n KWIK-STIK Microorganisms n KWIK-STIK Plus Microorganisms
INSTRUCTIONS FOR USE n LYFO DISK Microorganisms n KWIK-STIK Microorganisms n KWIK-STIK Plus Microorganisms LYFO DISK INTENDED USE LYFO DISK, KWIK-STIK and KWIK-STIK Plus Microorganisms are lyophilized,
More informationInterim Technical Note The Use of Cholera Rapid Diagnostic Tests November 2016
Objective Interim Technical Note The Use of Cholera Rapid Diagnostic Tests November 2016 To provide interim guidance to Ministries of Health and other organizations on the potential use of commercially
More informationNIH Public Access Author Manuscript J Antimicrob Chemother. Author manuscript; available in PMC 2007 November 14.
NIH Public Access Author Manuscript Published in final edited form as: J Antimicrob Chemother. 2005 August ; 56(2): 413 415. Colicins prevent colonization of urinary catheters Barbara W. Trautner 1,3,5,*,
More informationBASE 128 and BASE CONTENTS
BASE 128 and BASE CONTENTS INTRODUCTION RATIONALE FOR PRODUCT DEVELOPMENT & FORMULAS DECONTAMINATION EFFICACY OF BASE 128 RINSING EFFICACY OF BASE TISSUE SAFETY PRODUCT VALIDATION & COLLABORATIONS COMPLIANCE
More informationAntibiofilm agents against MDR bacterial strains: is bioactive glass BAG-S53P4 also effective?
J Antimicrob Chemother 2016; 71: 123 127 doi:10.1093/jac/dkv327 Advance Access publication 12 October 2015 Antibiofilm agents against MDR bacterial strains: is bioactive glass BAG-S53P4 also effective?
More informationPURE CULTURE TECHNIQUES
PURE CULTURE TECHNIQUES Most specimens (from animal tissue, plant tissue, or environmental samples) will be mixed, with a variety of bacteria (or other microorganisms). A single gram of feces, for example,
More informationAir Quality Monitoring of Operation Theaters in Government Medical College and Hospital, Aurangabad, India
International Journal of Current Microbiology and Applied Sciences ISSN: 2319-7706 Volume 5 Number 6 (2016) pp. 42-49 Journal homepage: http://www.ijcmas.com Original Research Article http://dx.doi.org/10.20546/ijcmas.2016.506.005
More informationOrthophenylphenol in healthcare environments: a trial related to a new administration method and a review of the literature*
Turkish Journal of Medical Sciences http://journals.tubitak.gov.tr/medical/ Research Article Turk J Med Sci (2013) 43: 805-809 TÜBİTAK doi:10.3906/sag-1208-4 Orthophenylphenol in healthcare environments:
More information320 MBIO Microbial Diagnosis. Aljawharah F. Alabbad Noorah A. Alkubaisi 2017
320 MBIO Microbial Diagnosis Aljawharah F. Alabbad Noorah A. Alkubaisi 2017 Primary Media for Isolation of Microorganisms As we know, many clinical specimens contain a mixed flora of microorganisms. Thus
More informationQuestions and Answers on allogenic stem cell-based products for veterinary use: specific questions on sterility
15 June 2017 EMA/CVMP/ADVENT/751229/2016 Committee for Medicinal Products for Veterinary Use (CVMP) Questions and Answers on allogenic stem cell-based products for veterinary use: Draft problem statement
More informationIQCP for Commercially Prepared CLSI-Exempt Media
Please note that some references to protocol, publications, performance data etc. are fictitious in this EXAMPLE. Please use your own DATA for your IQCP. The following represents one example of how you
More informationEffect of Storage of Mueller-Hinton Agar Plates on
APPUED MICROBIOLOGY, Sept. 1970, p. 293-297 Copyright 1970 American Society for Microbiology Vol. 20, No. 3 Printed in U.S.A. Effect of Storage of Mueller-Hinton Agar Plates on Zone Sizes for Antimicrobial
More informationStandard Operating Procedure
Subject Blood Culture Collection Index Number Lab-1325 Section Laboratory Subsection Specimen Collection and Processing Category Departmental Contact Regina Collins Last Revised 3/17/2017 References Required
More informationStudy Summary. Results: All tested media-filled vials were negative for growth of any microorganisms.
A 7 Days Microbial Ingress Test of Single Use Vials Utilizing the EQUASHIELD Closed System Drug Transfer Device Study by Nelson Laboratories (Salt Lake City, UT) 2013, updated in 2014* Study Summary Abstract:
More informationCeramics in Orthopaedic and Neurosurgery. B. Sonny Bal, MD MBA JD PhD University of Missouri-Columbia Amedica Corporation
Ceramics in Orthopaedic and Neurosurgery B. Sonny Bal, MD MBA JD PhD University of Missouri-Columbia Amedica Corporation Historic Concern: Wear Bearing Failures of the Past The Rationale for Alternatives
More informationCase 6. Fight Against Nosocomial Infections
Case 6 Fight Against Nosocomial Infections Decontamination and microbiological surveillance of endoscopes Prof. dr. Isabel Leroux-Roels Laboratory of Medical Microbiology & Infection Control Team UZ Gent
More informationCLIA Complexity: MODERATE
CLIA Complexity: MODERATE INTENDED USE The is intended for the rapid detection of Group A Streptococcal antigen directly from throat swabs and beta-hemolytic colonies recovered from culture. This test
More informationEzy MIC Strip FEATURES AND ADVANTAGES
Imipenem with & without EDTA Ezy MIC Strips (IPM+EDTA/IPM) (Imipenem + EDTA: 1-64 mcg/ml) (Imipenem : 4-256 mcg/ml) Antimicrobial Susceptibility Testing For In Vitro Diagnostic use EM078 Not for Medicinal
More informationChapter 02 Tools of the Laboratory: Methods for the Culturing and Microscopic Analysis
Chapter 02 Tools of the Laboratory: Methods for the Culturing and Microscopic Analysis Multiple Choice Questions 1. The Five I's of studying microorganisms include all of the following except A. inoculation.
More informationANAEROBIC DENTAL INFECTIONS AND ADVANCE LABORATORY DIAGNOSIS: A MINI REVIEW
Review Article Biswajit Batabyal,, 2012; Volume 1(5):533-538 ISSN: 2277-8713 ANAEROBIC DENTAL INFECTIONS AND ADVANCE LABORATORY DIAGNOSIS: A MINI REVIEW BISWAJIT BATABYAL 1,SUKANTA CHAKRABORTY 2, SHIBENDU
More informationVerification of Environmental Cleaning & the Impact on Patient Safety
SM 3M Health Care Academy Verification of Environmental Cleaning & the Impact on Patient Safety July 20, 2016 Steve Zeplin, RN, MSN, CSPDT House Keeping Continuing Education Each 1 hour web meeting qualifies
More informationIDEXX Summary. IDEXX Laboratories. Date: April Report Highlights:
IDEXX Summary 17A Topic: Title: Author: Beta Trial Study report comparing Legiolert* versus the standard method in Germany, described in Trinkwasserverordnung 2001 in der Fassung v. 14.12.2012 Anlage 5,
More informationShort Aramid Fibre Reinforced Ultra-High Molecular Weight Polyethylene Composite. A new hip-prosthesis Hofsté, Joanna Maria
University of Groningen Short Aramid Fibre Reinforced Ultra-High Molecular Weight Polyethylene Composite. A new hip-prosthesis Hofsté, Joanna Maria IMPORTANT NOTE: You are advised to consult the publisher's
More informationHardyVal TM CSP RANDOM TEST KIT
HardyVal TM CSP RANDOM TEST KIT Cat. no. HVR1 Random Test Kit 5 tests/kit INTENDED USE Each kit contains: Tryptic Soy Broth, Double Strength (DS), 12ml Syringe, 5.2ml Whirl-Pak Bag Results Log Sheet 5
More informationTransmission Electron Microscopic Study of Antibiotic Action on Klebsiella pneumoniae Biofilm
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, Aug. 2002, p. 2679 2683 Vol. 46, No. 8 0066-4804/02/$04.00 0 DOI: 10.1128/AAC.46.8.2679 2683.2002 Copyright 2002, American Society for Microbiology. All Rights Reserved.
More informationAntibiofilm Efficacy of Advanced Formula Zymox Otic Plus with 1% Hydrocortisone
Antibiofilm Efficacy of Advanced Formula Zymox Otic lus with 1% Hydrocortisone Rajvinder Atwal, hd¹ Anim Omar, h.d., B. harma² Merle Oson, DVM, MSc.² Nicole Butler, B.Sc.² Yanira Caberra, B.Sc.² ¹Laclede
More informationOverview of a sterility assurance program for PET drugs
Coalition for PET Drug Approval Radiopharmaceutical Sciences Council Overview of a sterility assurance program for PET drugs Eric Webster, PETNET Solutions Disclosures Employee of PETNET Solutions, a Siemens
More informationTransfusion-transmitted Virus(TTV) Real Time PCR Kit
Revision No.: ZJ0003 Issue Date: Aug 7 th, 2008 Transfusion-transmitted Virus(TTV) Real Time PCR Kit Cat. No.: HD-0013-02 For use with ABI Prism 7000/7300/7500/7900; Smart CyclerII; icycler iq 4/iQ 5;
More informationMICROBIOLOGICAL ANALYSIS OF FOOD & FOOD PRODUCTS
MICROBIOLOGICAL ANALYSIS OF FOOD & FOOD PRODUCTS By Angshuman Saha Assistant Director Export Inspection Agency-Mumbai, PTH & T Maheswara Rao Technical Officer Export Inspection Agency-Mumbai, PTH 1 Introduction
More informationTRYPTIC SOY AGAR (TSA) WITH LECITHIN AND TWEEN 80
TRYPTIC SOY AGAR (TSA) WITH LECITHIN AND TWEEN 80 Cat. no. P45 TSA with Lecithin and Tween 80, 15x60mm Contact Plate, 15ml Cat. no. Q13 TSA with Lecithin and Tween 80, 20x125mm Tube, 18ml Deep Cat. no.
More informationOne Surgeon. One Patient. FLH /11
FLH 219 03/11 This publication has been issued by: European Central Marketing Waterton Industrial Estate Bridgend South Wales CF31 3XA +44 [0] 1656 655221 +44 [0] 1656 645454 Responsible Manufacturer:
More informationEstablishing a Pseudomonas aeruginosa Biofilm in a Murine Model
University of North Texas Health Science Center UNTHSC Scholarly Repository Theses and Dissertations Spring 5-2010 Establishing a Pseudomonas aeruginosa Biofilm in a Murine Model David Valtierra University
More informationTests to Support Sterility Claim. Imtiaz Ahmed
Tests to Support Sterility Claim Imtiaz Ahmed Sterile Product As per TGO 77, a sterile product must comply with the requirements of the following tests: Sterility Test Bacterial Endotoxins Test Appendix
More informationThe University of Ferrara Study At the St. Anne Hospital
The University of Ferrara Study At the St. Anne Hospital The Research Centre For Pollution Control In High Sterile Rooms Conducted A Year-Long Study (2010 2011) On The TESTING OF BIO-STABILIZATION TECHNIQUES
More informationLaboratory Experiments in Corrosion Engineering II
Lecture - 40 Laboratory Experiments in Corrosion Engineering II Keywords: Polarization Experiments, Pitting Potentials, Microbial Corrosion. A. Electrochemical tests in a given environment Polarization
More informationMethod Suitability Report Membrane Filtration Sterility Test with QTMicro Apparatus
Method Suitability Report Membrane Filtration Sterility Test with QTMicro Apparatus Bevacizumab (Avastin) 1.25 mg/0.05ml This report provides details on the specifics of a Membrane Filtration Sterility
More information6/28/2016. Control of Microbial Growth. Method. Terminology. Disinfectants and Antiseptics
Control of Microbial Growth Disinfectants and Antiseptics 1 Method Three approaches for the control of microbial growth Chemical Disinfectants and antiseptics Physical Heat Ultraviolet Irradiations Mechanical
More informationLab Exercise: Examining Water Quality: Most Probable Number & Colilert Test Kit Lab
Lab Exercise: Examining Water Quality: Most Probable Number & Colilert Test Kit Lab OBJECTIVES 1. Understand the use of MPN to determine likely fecal water contamination. 2. Understand the use of MUG,
More informationLaminar flow in orthopaedic operations - Essential if you can afford it. Peter Hoffman Consultant Clinical Scientist Public Health England
Laminar flow in orthopaedic operations - Essential if you can afford it Peter Hoffman Consultant Clinical Scientist Public Health England Up until the 1950s in the UK (and elsewhere) Surgical instruments
More informationCitation for published version (APA): Free, R. H. (2004). Exogenous factors influencing voice prosthetic biofilm Groningen: s.n.
University of Groningen Exogenous factors influencing voice prosthetic biofilm Free, Rolina Henderika IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to
More informationHot Topics in Drug Product Process Validation: A Reviewer s Perspective
Hot Topics in Drug Product Process Validation: A Reviewer s Perspective Colleen Thomas, Ph.D. Quality Assessment Lead (Acting) FDA/CDER/OPQ/OPF Division of Microbiology Assessment CASSS CMC Strategy Forum
More informationá62ñ MICROBIOLOGICAL EXAMINATION OF NONSTERILE PRODUCTS: TESTS FOR SPECIFIED MICROORGANISMS
USP 40 Microbiological Tests / á62ñ Microbiological Examination 1 á62ñ MICROBIOLOGICAL EXAMINATION OF NONSTERILE PRODUCTS: TESTS FOR SPECIFIED MICROORGANISMS INTRODUCTION The tests described hereafter
More informationLewis White Department of Medical Microbiology and NPHS Cardiff, University Hospital of Wales, Cardiff, UK Friday, February 24, 2006, 10:05-10:25 am
show slides PCR PLATFORMS, STRENGTHS AND WEAKNESSES Lewis White Department of Medical Microbiology and NPHS Cardiff, University Hospital of Wales, Cardiff, UK Friday, February 24, 2006, 10:05-10:25 am
More informationDentistry Test methods for dental unit waterline biofilm treatment
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 16954 First edition 2015-07-01 Dentistry Test methods for dental unit waterline biofilm treatment Médecine bucco-dentaire Méthodes d essais pour
More informationFINAL REPORT Efficacy of a Cold Plasma System
FINAL REPORT Efficacy of a Cold Plasma System ORDER Number PREPARED FOR: 714 Mall Blvd., Suite 3 Savannah, GA 31406 EMSL Analytical, Inc. Web: http://www.emsl.com - 1 - Certificate of Analysis Client:
More informationHuman Pluripotent Stem Cell Growth Medium DXF
Human Pluripotent Stem Cell Growth Medium DXF Instruction Manual Product Size Catalog Number hpsc Growth Medium DXF 500 ml C-28060 Recommended for Human Pluripotent Stem Cells (hpsc) Product Description
More informationDisinfectant Qualification A Multifaceted Study
Disinfectant Qualification A Multifaceted Study In this article, Ziva Abraham, Microrite, Inc., gives us a brief overview of the disinfectant qualification study process. An effective cleaning and disinfection
More informationEM021. Co-Trimoxazole Ezy MIC TM Strip (COT)( mcg/ml) (Trimethoprim/ Sulphamethoxazole) Antimicrobial Susceptibility Testing
Co-Trimoxazole Ezy MIC TM Strip (COT)(0.002-32 mcg/ml) (Trimethoprim/ Sulphamethoxazole) Antimicrobial Susceptibility Testing For In Vitro Diagnostic use EM021 Not for Medicinal Use It is a unique MIC
More informationa solution for reducing the
Outsourcing the precision cleaning, a solution for reducing the environmental impact Trade fair Hanover 2015 Heiko ZSCHIEDRICH Hanover April 15, 2015 Summary 1. Introduction and presentation of ECP 2.
More informationLab Exercise #4 Microbial Control Lab Exercise #4 Control of Microorganisms: Physical, Chemical and Chemotherapeutic
Lab Exercise #4 Control of Microorganisms: Physical, Chemical and Chemotherapeutic I. OBJECTIVES: Investigate the effectiveness various agents of control. Assess the effectiveness of heat in killing vegetative
More informationWHY DO THEY PUT MINT IN TOOTHPASTE? WOULD GARLIC BE BETTER?
Activity 4.22 Student Sheet WHY DO THEY PUT MINT IN TOOTHPASTE? WOULD GARLIC BE BETTER? Purpose To investigate the antibacterial properties of plants. To develop practical skills. YOU NEED Agar plate seeded
More informationDesigning an effective cleaning procedure for medical devices through laboratory studies
Designing an effective cleaning procedure for medical devices through laboratory studies Elizabeth Rivera Technical Service Specialist STERIS Corporation elizabeth_rivera@steris.com October 2012 Agenda
More informationEfficacy Report Summarization for SoClean 2
Efficacy Report Summarization for SoClean 2 October 2017 SoClean Inc 36 Town Forest Road Oxford, Massachusetts 01540 USA Tel. 508-363-0418 Email info@soclean.com SoClean 2 is USA FDA Registered 3009534409
More informationAqua 65JL. Water-Based Hydrophilic Coating. Highly Lubricious, Non-Eluting, surface coating for the reduction of biofilm adhesion.
Biofilm Resistant Water-Base for PU & Silicone Water-Based Hydrophilic Coating Highly Lubricious, Non-Eluting, surface coating for the reduction of biofilm adhesion. US FDA MAF-1785 11/05/2013 CE Approval
More informationVanessa Cris D. Blancaflor 1, Jocelyn T. Zarate 2 and Celia B. Dela Viña 3 1, 3
Morphological, Biochemical and Molecular Characterization Of Gibberellic Acid- Producing Bacteria, Isolated from Saccharum sp. and Eleusine sp. and Their Effect On The Growth Of Corn and Lettuce Vanessa
More informationCreating a Culture of Data Integrity Using an Automated Detection and Enumeration Method
Creating a Culture of Data Integrity Using an Automated Detection and Enumeration Method Novartis, Schaffhauserstrasse, 4332 Stein AGStein, Switzerland Sanofi Pasteur, Marcy l Etoile, France Leo Pharma,
More informationChapter 03 - Tools of the Laboratory: Methods for the Culturing of Microscopic Analysis of microorganisms
Microbiology: A Systems Approach 4th Edition Cowan Test Bank Completed download: https://testbankreal.com/download/microbiology-systems-approach-4thedition-test-bank-cowan/ (Downloadable package TEST BANK
More informationHistoacryl A revolution in mesh fixation
Histoacryl A revolution in mesh fixation Closure Technologies Histoacryl A revolution in mesh fixation A new indication for a classic product Histoacryl has been used for more than 40 years in operating
More informationSection III Research Articles and Short Communications
231 Section III Research Articles and Short Communications 230 233 Screening of Edible Fish Pathogens Deepali Patil Department of Environmental Science B. N. Bandodkar College of Science Abstract : A microbial
More informationAbstract. Mary Jane Ferraro, PhD, MPH Jana M. Swenson, MMSc
January 2009 Vol. 29 No. 2 Replaces M07-A7 Vol. 26 No. 2 Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard Eighth Edition This document addresses
More informationProviding clear solutions to microbiological challenges TM. cgmp/iso CLIA. Polyphasic Microbial Identification & DNA Fingerprinting
Providing clear solutions to microbiological challenges TM Cert. No. 2254.01 Polyphasic Microbial Identification & DNA Fingerprinting Microbial Contamination Tracking & Trending cgmp/iso-17025-2005 CLIA
More informationBenefits and limitations of steam cleaner for the biocleaning of surfaces
Benefits and limitations of cleaner for the biocleaning of surfaces Olivier Meunier, Françoise Salles, Sandrine Burger Hygiene operational team Hospital Centre Hospitalier de Haguenau 64 av. Professor
More informationBacterial Isolation, Identification, and Storage. Lila Ruangpan
Chapter 1. Bacterial isolation, identification and storage Ruangpan, Lila & Tendencia, Eleonor A. Date published: 2004 To cite this document : Ruangpan, L., & Tendencia, E. A. (2004). Chapter 1. Bacterial
More informationSwab type, moistening, and pre-enrichment for Staphylococcus aureus on environmental. Timothy F. Landers* 1. Armando Hoet 2. Thomas E.
JCM Accepts, published online ahead of print on April 00 J. Clin. Microbiol. doi:0./jcm.0-0 Copyright 00, American Society for Microbiology and/or the Listed Authors/Institutions. All Rights Reserved.
More informationebds ebds SAMPLE SET English REF E HAEMONETICS CORPORATION 400 Wood Road, Braintree, Massachusetts 02184, USA
HAEMONETICS CORPORATION 00 Wood Road, Braintree, Massachusetts 0, USA EC REP HAEMONETICS UK LTD Ashley Drive, Bothwell, Scotland, G BS, UK Produced by HAEMONETICS MANUFACTURING INC 0 Industrial Park Street,
More informationCatalog # Product Size PRIME-XV Hematopoietic Cell Basal XSFM 500 ml (liquid) Additional package sizes are available at request
PRIME-XV HEMATOPOIETIC CELL BASAL XSFM PRIME-XV Hematopoietic Cell Basal XSFM is an optimized xeno- and serum free media recommended for use in the expansion of human hematopoietic cells, including hematopoietic
More informationInternal Comparative Evaluation Study of Alternative Enrichment Protocols for Detection of Salmonella in Untreated Spices
Internal Comparative Evaluation Study of Alternative Enrichment Protocols for Detection of Salmonella in Untreated Spices Tech Note Introduction Between 1973 and 2010, there were 14 outbreaks of foodborne
More informationStandard Operating Procedure (SOP) Specimen processing of CSF, lymph nodes and other tissues for Xpert MTB/RIF
Page: 1 of 7 Content 1. Scope 2. Definitions and abbreviations 3. Procedure 3.1 Principle 3.2 General considerations 3.3 Specimen processing 3.3.1 Lymph nodes and other tissues (Xpert MTB/RIF only) 3.3.2
More informationAbout Us. A young organisation, but with deep roots.
About Us Adler Ortho is an Italian Company totally devoted to development, manufacturing and marketing of surgical medical devices for Orthopaedics. Innovation is the foundation of our Company, pursued
More informationBiological Monitoring
Biological Monitoring Prof. Duygu PERÇİN, MD Department of Clinical Microbiology Erciyes University Faculty of Medicine, Kayseri-TURKEY duygu.percin@hotmail.com Learning Objectives The attandee will be
More informationAntimicrobial Surfaces to Prevent Healthcare-Associated Infection
Antimicrobial Surfaces to Prevent Healthcare-Associated Infection Matthew P. Muller, MD, PhD, FRCPC Medical Director, Infection Prevention and Control, St. Michael s Hospital Chair, Provincial Infectious
More informationLab 2-Microbial Enumeration
Lab 2-Microbial Enumeration 2/19/08 CE 573 Introduction There are many different techniques that can be utilized when trying to quantify microorganisms found in a given sample. The purpose of this lab
More information