Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system

Size: px
Start display at page:

Download "Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system"

Transcription

1 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system Laurence E. Court, a Roy B. Tishler, Aaron M. Allen, Hong Xiang, Mike Makrigiorgos, and Lee Chin Department of Radiation Oncology, Dana Farber / Brigham & Women s Cancer Center, Boston, Massachusetts, U.S.A. lcourt@lcroc.harvard.edu Received 30 August 2007; accepted 16 November 2007 The present work uses the Eclipse treatment planning system (TPS) to investigate the accuracy of skin dose calculations. Micro-MOSFETs (metal oxide semiconductor field effect transistors) were used to measure skin dose for a range of irradiation conditions (open fields, physical wedges, dynamic wedges, various source-to-surface distances) for 6-MV and 10-MV beams, and the results were compared with the calculated mean dose to a skin structure 2 mm thick for semicylindrical phantoms (representative of a neck or breast). Agreement between the calculated and measured skin dose values was better than ±20% for 95% of all measured points (6-MV and 10-MV X-ray spectra alike). For a fixed geometry, the TPS correctly calculated relative changes in dose, showing that minimization of skin dose in intensity-modulated radiation therapy will be effective in Eclipse. PACS numbers: Bn, Dq, Pm, Xa Key words: Skin dose, dose calculations, MOSFET I. INTRODUCTION Skin dose can be the limiting factor in radiation therapy treatments, and it is a fairly common cause of interruptions in radiation therapy treatment. Skin dose is of particular concern when intensity-modulated radiation therapy (IMRT) is used to treat head-and-neck cancer. (1) The use of thermoplastic head-and-neck immobilization devices increases skin dose, (1,2) and the use of multiple tangential beams (common in IMRT treatments) can also increase it. The chosen planning and optimization strategy can also affect skin dose, with expansion of the planning target volume to include the skin increasing skin dose, and defining the skin as a sensitive structure reducing skin dose. (1,3) Despite the clinical importance of skin dose, the literature contains little detail concerning the expected accuracy of skin dose calculations. Fraas et al. (4) (American Association of Physicists in Medicine Task Group 53) reported the collective expectations of the Radiation Therapy Committee Task Group members for pass fail criteria in the buildup region during the commissioning of a treatment planning system (TPS) as 20% of the central ray normalization dose for regular open fields, increasing to 40% for source-to-surface distance (SSD) variations, and 50% for wedged fields. Using radiochromic film, Chung et al. (5) reported that two TPSs [Pinnacle 3 (Philips Medical Systems, Andover, MA) and CORVUS (North American Scientific, Chatsworth, CA)] overestimated surface dose by 7.4% 18.5%. Using thermoluminescent dosimeters and parallel-plate ionization chambers, Mutic and Low (6) found that a tomotherapy TPS underestimated doses to the surface and first few millimeters below the surface by approximately 15%. a Corresponding author: Laurence E. Court, Department of Radiation Oncology, Dana Farber / Brigham & Women s Cancer Center, Boston, MA, U.S.A.; phone: ; fax: ; lcourt@lcroc.harvard.edu 29 29

2 30 Court et al.: Experimental evaluation of the accuracy The objective of the present work was to evaluate the accuracy of skin dose calculations in the Eclipse TPS (Varian Medical Systems., Palo Alto, CA) for a wide range of irradiation conditions. In particular, we looked to answer two questions: First, by comparing experiments and calculations for a range of irradiation conditions typical in radiation therapy, we wanted to be able to quote the accuracy of skin dose calculations. Physicians often ask planners this important clinical question, particularly for head-andneck treatments, and it is one that has not been evaluated in the literature. Second, we wanted to try to predict how the calculation uncertainties would affect the minimization of skin dose during IMRT optimization. For optimization to be reliable, calculated and measured skin dose must be correlated, even if the calculation contains a systematic error. The literature has yet to address these important clinical questions for the Eclipse TPS. That system is very popular, with approximately 6000 stations installed at 2000 sites worldwide (personal communication, Varian Medical Systems, November 2007). To answer our questions, we used MOSFET (metal oxide semiconductor field effect transistor) dosimeters (micro-mosfet, model TN-502RDM: Thomson Nielson Electronics, Nepean, ON) to measure skin dose for 6-MV and 10-MV photon spectra, a range of jaw-defined field sizes, physical and dynamic wedges, IMRT fields, a range of SSDs, and a range of incident angles. We then compared this experimental data with Eclipse calculations. The resulting knowledge will be useful when evaluating and comparing patient plans and also when developing optimization strategies for IMRT (for example, whether the use of a hard constraint for skin dose is sensible). II. MATERIALS AND METHODS A. Choice of radiation detector Skin dose measurements can have significant uncertainties. For example, Quach et al. (7) reported that measurements of surface dose by MOSFET, thermoluminescent dosimeter, and film can disagree by up to 50% 60%. In practice, surface (or skin) dose is often taken as the dose measured by whichever dosimeter is used. (8) We recently addressed the question of the accuracy of skin dose measurements made using MOSFETs. (9) To do so, we used carefully commissioned Monte Carlo techniques to calculate the mean dose to a surface volume 2-mm thick in the region of interest, and we compared those results with experimental dose measurements made by MOSFET. We found that, by embedding the MOSFET in 1 mm of bolus material, the angular dependence effect of the MOSFETs could be reduced, giving readings with a total error of approximately 6%. That error is small as compared with the expectations of Fraas et al. (4) regarding the accuracy of the TPS skin dose calculations (up to 50%). Therefore, MOSFETs are an appropriate choice when measuring skin dose for a range of irradiation conditions. The micro-mosfets in the present study were used in standard bias setting, for a normal sensitivity of approximately 1 mv/cgy. (References provide a full description of MOSFETs.) The overall physical size of the sensors is mm, and the actual sensitive volume is 0.2 mm 0.2 mm 0.5 µm. Signals were read using a wireless mobile MOSFET reader (Model TN-RD-16: Thomson Nielson Electronics), controlled by remote doseverification software running on a personal computer. All dose measurements were carried out with the flat side of the MOSFET placed to face the beam. A full description of MOSFETs can be found in references The MOSFETs were calibrated before each experimental session by first placing the MOSFET at the depth of maximum dose (1.5 cm and 2.5 cm for 6-MV and

3 31 Court et al.: Experimental evaluation of the accuracy MV beams respectively) in a Solid Water (Gammex rmi, Middleton, WI) and bolus phantom (bolus sheets were placed on both sides of the MOSFET to minimize air gaps). The MOSFETs were then irradiated with 100 MU in a cm field, and calibration factors in cgy/mv were obtained. B. Phantom The semi-cylindrical (12 cm diameter) Solid Water phantom used for this study (Fig. 1) is representative of a neck or breast. The first centimeter of the phantom surface was bolus material. Marks and radiopaque fiducials were used to identify measurement positions on the surface of the phantom. The phantom then underwent computed tomography (CT) scanning, yielding CT pixels 1.3 mm 1.3 mm 2.5 mm, which is representative of the imaging parameters used in our clinic. FIG. 1. The curved phantom used for the experiments reported here, showing the irradiation geometry and measurement points. C. Irradiation of the phantom Table 1 shows the various parameters investigated. The investigations were conducted using a Varian 21Ex at the Dana Farber Cancer Institute. Calibrated MOSFETs were placed at the measurement points (Fig. 1), embedded in the first millimeter of the surface bolus material. The phantom was irradiated at a gantry angle of 90 degrees, using combinations of the study parameters. Because a curved phantom was used, the effect of irradiation angle is inherently included. For each parameter, the SSD at the central axis was set to 80, 90, and 100 cm, simulating the variations that can be expected in clinical practice. The exact SSD depended on the measurement point, and it varied from 76 cm to 110 cm. All experiments, with the exception of the IMRT fields, were repeated for 6-MV and 10-MV X-ray spectra. The IMRT fields were 6-MV fields, with the central axis set to 90 cm. The IMRT fields originated in a previous study (3) that investigated planning techniques to minimize skin dose in head-and-neck IMRT. The difference between the various plans was the extent to which calculated dose to a skin structure was minimized. TABLE 1. The various parameters used for dose calculations and measurements Parameter Values Jaw-defined field size (cm) 3 3, 5 5, 10 10, 15 15, 17 15, Physical wedge (degrees) 15, 30, 45, 60 Dynamic wedge (degrees) 15, 30, 45, 60 Dynamic IMRT (6 MV only) Cumulative dose from multiple beams IMRT = intensity-modulated radiation therapy.

4 32 Court et al.: Experimental evaluation of the accuracy D. TPS calculations According to reports from the International Commission on Radiological Protection (14) and the International Commission on Radiation Units and Measurements, (15) the recommended depth for practical dose assessments is 0.07 mm. That depth corresponds approximately to the interface between the epidermis and dermis layers of the skin (8,16) ( mm, depending on the anatomic location). That depth is very difficult to measure or to calculate using most TPSs. For practical reasons in the present work, we therefore chose to define the skin dose as the mean dose to the surface volume (2 mm thickness) in the region of interest. Separate plans were created in Eclipse for each experimental setup. Skin doses were calculated by first creating a surface structure 2 mm 10 mm 10 mm centered on the measurement mark (as seen using fiducials in the CT images). Skin dose was calculated as the mean dose to this defined structure. The dose calculation grid was set to 2.5 mm, because that size is the one used for most clinical cases. All calculations were performed using the Eclipse pencil-beam calculation algorithm. That algorithm first calculates dose at the dose grid points, forcing the dose at points outside the body structure to zero and then interpolating between grid points. The measured and calculated doses were compared, and the differences were expressed as percentages of the measured dose. Agreement between the Eclipse calculations and the MOSFET measurements was then investigated further for dynamic IMRT fields for a single central axis SSD (90 cm), looking at a single point. The purpose was to investigate the uncertainties in calculated skin dose that may be experienced when IMRT plans are optimized for a single patient (that is, fixed geometry). III. RESULTS Table 2 compares the mean, standard deviation, and range of differences for the various irradiation parameters. Overall, agreement was better than 10% for 75% of all measurement points at 6 MV and 63% of points at 10 MV. Agreement was better than 20% for 94% of the measurement points at 6 MV and 96% of the points at 10 MV. TABLE 2. Results of the differences between Eclipse skin dose calculations and MOSFET (metal oxide semiconductor field effect transistor) measurements Difference a (%) Mean / median SD Range 6 MV Open field 4.0 / to +17 Physical wedge 3.3 / to +14 Dynamic wedge 7.2 / to +7 IMRT (all SSDs) 4.0 / to +20 Overall 4.1 / to +20 Single point IMRT 15.2 / to 7 10 MV Open field 2.0 / to +26 Physical wedge 6.2 / to +13 Dynamic wedge 4.2 / to +11 Overall 3.9 / to +26 a Negative numbers indicate that measured values were lower than those calculated by the treatment planning system. Percentages are given relative to the measured dose. SD = standard deviation; IMRT = intensity-modulated radiation therapy; SSD = source to surface distance.

5 33 Court et al.: Experimental evaluation of the accuracy Fig. 2 shows the dose difference (Eclipse vs. MOSFET) for three different setup positions (different SSDs) for various MOSFET positions (shown in Fig. 1) for the 6-MV beam. The measurement and calculation position can be seen apparently to have a larger effect on the agreement than the SSD does. Table 2 also shows the results for a single point, with fixed SSD, in 10 IMRT plans. Fig. 3 shows a graph of these results. It can be seen from the graph that, although a systematic difference of approximately 13% exists between the Eclipse calculations and the experimental points, Eclipse models the changes in skin dose for the various plans fairly well. That is, a real increase in dose (that is, measured dose) produces an increase in calculated dose. The difference between the Eclipse calculations and a second-order polynomial fit to the data was 0.3% ± 1.6% (one standard deviation), with a range of 2% to +3%. FIG. 2. Comparison of the calculated and measured doses for multiple points (open fields) for a 6-MV beam at various isocenter source-to-surface distances (SSDs). The position refers to Fig. 1. FIG. 3. Comparison of the calculated and measured dose for a single point for cumulative intensity-modulated radiation therapy fields. The curve is a second-order polynomial fit to the data. Note that the ranges of the two axes are the same, but that an offset occurs between the calculated and the measured data. MOSFET = metal oxide semiconductor field effect transistor.

6 34 Court et al.: Experimental evaluation of the accuracy IV. DISCUSSION For most clinical cases, the data indicate that the Eclipse skin dose calculation is accurate within ±20% (95% of all measurement points). In situations in which more accuracy in the data is needed, on-patient dose measurements are necessary. Application of these data to other centers will depend on differences in how the dose data for their linear accelerator are obtained and how the TPS is commissioned. In our case, the TPS was commissioned using data supplied by the vendor (Varian s golden beam data). Those data were compared with dose distributions [percentage depth doses (PDDs), profiles, isodose curves] measured using a moving ion chamber in a three-dimensional water phantom (Scanditronix Wellhofer, Nuremburg, Germany). We also used a parallel-plate chamber to measure doses near the surface. Table 3 gives the differences between the PDDs in the TPS and those in the experimental parallel-plate measurements. TABLE 3. Differences between percentage depth dose (PDD) measured using a parallel-plate chamber and PDDs used in the treatment planning system Depth Difference a (%) (mm) 6-MV Beam 10-MV Beam a Relative to the maximum dose. Negative numbers indicate that measured values were lower than those calculated by the treatment planning system. Knowing the accuracy of the skin dose calculation in Eclipse allows for a sensible comparison of skin dose between patients or plans. Importantly, Eclipse seems to accurately model changes in skin dose between IMRT plans, meaning that it can be expected to correctly minimize the skin dose when the skin is defined as a critical structure in those plans. Systematic differences which depend on measurement point, SSD, and so on mean that the use of a hard absolute constraint for skin dose (for example, keeping skin dose strictly below 60% of prescription) should be approached with caution. Instead, it may be more advisable to aim for a dose range (for example, 50% 70% of prescription) or to minimize the skin dose as much as possible (without a specific goal dose), while maintaining specific doses to deeper targets (which are more accurately modeled). The steep dose gradient at the surface can make reliable dose calculations difficult. Depending on the CT and calculation grid parameters, the interplay between pixel size, pixel location, exact phantom (or patient) location, contour grid, and dose calculation grid can be expected to have an important effect on the calculated doses. This interplay is probably one of the main causes for the importance of position rather than SSD in the dose calculation accuracy (Fig. 2). Depending on position, the interplay will be different, resulting in different calculations of skin dose. This finding may be particularly true in the case of flat phantoms, in which no opportunity for averaging of the effects occurs. Nevertheless, for the parameters tested here, we found reasonable agreement between the experimental and calculated skin doses. Calculations for parameters outside of those tested here should still be treated with caution. V. CONCLUSION Agreement between skin doses calculated by Eclipse and those measured by MOSFET in hemispheric phantoms was within ±20% for 95% of all measurement points. Eclipse models relative increases and decreases in skin dose from different IMRT plans reasonably well, and so IMRT

7 35 Court et al.: Experimental evaluation of the accuracy optimization can be expected to be successful in reducing skin dose. However, an absolute goal for skin dose should be considered with caution. REFERENCES 1. Lee N, Chuang C, Quivey JM, et al. Skin toxicity due to intensity-modulated radiotherapy for head-and-neck carcinoma. Int J Radiat Oncol Biol Phys. 2002;53(3): Hadley SW, Kelly, R, Lam K. Effects of immobilization mask material on surface dose. J Appl Clin Med Phys. 2005;6(1): Court LE, Tisher RB. Experimental evaluation of the impact of different head-and-neck intensity-modulated radiation therapy planning techniques on doses to the skin and shallow targets. Int J Radiat Oncol Biol Phys. 2007;69(2): Fraas B, Doppke K, Hunt M, et al. American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning. Med Phys. 1998;25(10): Chung H, Jin H, Dempsey JF, et al. Evaluation of surface and build-up region dose for intensity-modulated radiation therapy in head and neck cancer. Med Phys. 2005;32(8): Mutic S, Low DA. Superficial doses from serial tomotherapy delivery. Med Phys. 2000;27(1): Quach KY, Morales J, Butson MJ, Rosenfeld AB, Metcalfe PE. Measurement of radiotherapy X-ray skin dose on a chest wall phantom. Med Phys. 2000;27(7): Devic S, Seuntjens J, Abdel-Rahman W, et al. Accurate skin dose measurements using radiochromic film in clinical applications. Med Phys. 2006;33(4): Xiang HF, Song JS, Chin DW, et al. Build-up and surface dose measurements on phantoms using micro-mosfet in 6 and 10 MV X-ray beams and comparisons with Monte Carlo calculations. Med Phys. 2007;34(4): Chuang CF, Verhey LJ, Xia P. Investigation of the use of MOSFET for clinical IMRT dosimetric verification. Med Phys. 2002;29(6): Ramaseshan R, Kohli KS, Zhang TJ, et al. Performance characteristics of a micro-mosfet as an in vivo dosimeter in radiation therapy. Phys Med Biol. 2004;49: Ramani R, Russell S, O Brien P. Clinical dosimetry using MOSFETs. Int J Radiat Oncol Biol Phys. 1997;37(4): Soubra M, Cygler J, Mackay G. Evaluation of a dual bias dual metal oxide silicon semiconductor field effect transistor detector as radiation dosimeter. Med Phys. 1994;21(4): International Commission on Radiological Protection Recommendations of the International Commission on Radiological Protection: adopted by the Commission in November Oxford (U.K.): Pergamon Press; p. 15. International Commission on Radiation Units and Measurements (ICRU). Determination of dose equivalents resulting from external radiation sources. ICRU Report 39. Bethesda (MD): ICRU; p. 16. International Commission on Radiological Protection. Reference man: anatomical, physiological and metabolic characteristics. Oxford (U.K.): Pergamon Press; p.

Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems

Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 3, SUMMER 2005 Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems Charles M. Able and Michael D. Thomas

More information

Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment

Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment Anup Kumar Bhardwaj et al Study of 2D ion chamber array for angular response and QA of dynamic MLC... Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment IMRT plans

More information

Imaging/Imagine Needs for Proton Therapy: Treatment Planning. Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA

Imaging/Imagine Needs for Proton Therapy: Treatment Planning. Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA Imaging/Imagine Needs for Proton Therapy: Treatment Planning Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA AAPM Annual Meeting Indianapolis, Aug. 07, 2013 Disclosure Software licensing agreement

More information

Proposal on Guideline for Quality Assurance of Radiation Treatment Planning System

Proposal on Guideline for Quality Assurance of Radiation Treatment Planning System Original Article PMP Progress in Medical Physics 28(4), December 2017 https://doi.org/10.14316/pmp.2017.28.4.197 pissn 2508-4445, eissn 2508-4453 Proposal on Guideline for Quality Assurance of Radiation

More information

Dosimetry of Small IMRT Fields

Dosimetry of Small IMRT Fields Dosimetry of Small IMRT Fields Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 IMRT and Small Fields IMRT beams typically contain small

More information

Shielding evaluation for IMRT implementation in an existing accelerator vault

Shielding evaluation for IMRT implementation in an existing accelerator vault JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Shielding evaluation for IMRT implementation in an existing accelerator vault R. A. Price,* O. Chibani, and C.-M. Ma Department

More information

Practical Workflow and the Cost of Adaptive Therapy. Rojano Kashani, Ph.D., DABR Washington University School of Medicine March 7, 2015

Practical Workflow and the Cost of Adaptive Therapy. Rojano Kashani, Ph.D., DABR Washington University School of Medicine March 7, 2015 Practical Workflow and the Cost of Adaptive Therapy Rojano Kashani, Ph.D., DABR Washington University School of Medicine March 7, 2015 Disclosures Travel expenses paid by ViewRay Inc. for on-site software

More information

ArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools

ArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools ArcCHECK & 3DVH The Ultimate 4D Patient QA Solution Your Most Valuable QA and Dosimetry Tools THE 4D QA GOLD STANDARD The ArcCHECK is the world s most selected independent 4D measurement array. Simply

More information

DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM

DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM Dee-Ann Radford, David S. Followill, Peter A. Balter, William F. Hanson Department of Radiation Physics The

More information

Preliminary Study for Dosimetric Characteristics of 3Dprinted. Materials with Megavoltage Photons

Preliminary Study for Dosimetric Characteristics of 3Dprinted. Materials with Megavoltage Photons Preliminary Study for Dosimetric Characteristics of 3Dprinted Materials with Megavoltage Photons Seonghoon Jeong and Myonggeun Yoon Department of Bio-Convergence Engineering, Korea University, Seoul, Korea

More information

8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)*

8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)* and Quality Assurance Considerations for Volumetric Modulated Arc Therapy * What is? Report of writing group from TETAWG under Therapy Physics Committee (TPC) of AAPM Writing Group members: James M. Galvin,

More information

This is probably the kind of radiotherapy that you are used to delivering in your country.

This is probably the kind of radiotherapy that you are used to delivering in your country. This is probably the kind of radiotherapy that you are used to delivering in your country. This slide shows images two patients, one with a T3 N2c nasopharyngeal CA and the other with a T1/2 N1/2 lung

More information

RapidArc: Clinical Implementation

RapidArc: Clinical Implementation RapidArc: Clinical Implementation Fang-Fang Yin, PhD Q. Jackie Wu, PhD Acknowledgements Team efforts from staff at Duke Radiation Oncology, especially to Dr. J Chang, Dr. J O Daniel for providing slide

More information

Clinical experience of the importance of daily portal imaging for head and neck IMRT treatments

Clinical experience of the importance of daily portal imaging for head and neck IMRT treatments JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 3, SUMMER 2008 Clinical experience of the importance of daily portal imaging for head and neck IMRT treatments Laurence E Court, a Luciant

More information

An improvement in IMRT QA results and beam matching in linacs using statistical process control

An improvement in IMRT QA results and beam matching in linacs using statistical process control JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 An improvement in IMRT QA results and beam matching in linacs using statistical process control Justin D. Gagneur a and Gary A. Ezzell

More information

Uncertainties in treatment planning dose computation. Jeffrey V. Siebers

Uncertainties in treatment planning dose computation. Jeffrey V. Siebers Uncertainties in treatment planning dose computation Jeffrey V. Siebers Acknowledgements Dr David Followill, RPC for access to RPC data Drs. Catharine Clark, Pablo Carrasco, Warren Kilby, Guanghua Yan

More information

Assessing dose variance from immobilization devices in VMAT head and neck treatment planning: A retrospective case study analysis

Assessing dose variance from immobilization devices in VMAT head and neck treatment planning: A retrospective case study analysis Assessing dose variance from immobilization devices in VMAT head and neck treatment planning: A retrospective case study analysis Alyssa Olson, B.S., R.T.(T), Kristine Phillips, B.S., R.T.(T), Tamara Eng

More information

N.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio Bologna, Italy

N.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio Bologna, Italy AN ARTIFICIAL NEURAL NETWORK TO PREDICT TIME OF REPLANNING FOR TOMOTHERAPY TREATMENTS N.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio 40 40138 Bologna, Italy

More information

The step-and-shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage

The step-and-shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 The step-and-shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage Heming Zhen, 1 Luo Ouyang, 2 Qinan

More information

VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry

VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry Journal of Radiotherapy in Practice Journal of Radiotherapy in Practice (2013) Page 1 of 7 & Cambridge University Press 2013 doi:10.1017/s1460396912000556 Technical Note VMAT monthly QA using two techniques:

More information

Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy

Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy UCRL-JC-133076 PREPRINT Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy C.L. Hartmann Siantar, P.M. Bergstrom T.P. Daly, M. Descalle D. Garrett, R.K.

More information

Clinically relevant investigation of flattening filter-free skin dose

Clinically relevant investigation of flattening filter-free skin dose JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 Clinically relevant investigation of flattening filter-free skin dose Christopher L. Guy, 1 Kishor Karki, 1 Manju Sharma, 1,2 and

More information

Design and dosimetry characteristics of a commercial applicator system for intra-operative electron beam therapy utilizing ELEKTA Precise accelerator

Design and dosimetry characteristics of a commercial applicator system for intra-operative electron beam therapy utilizing ELEKTA Precise accelerator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 4, fall 2010 Design and dosimetry characteristics of a commercial applicator system for intra-operative electron beam therapy utilizing ELEKTA

More information

Quality Assurance for particle beam therapy

Quality Assurance for particle beam therapy Quality Assurance for particle beam therapy PTCOG Educational Workshop, Essen 2013 Heidelberg Ion Beam Therapy Center at the University Hospital German Cancer Research Center, Heidelberg, Germany Outline

More information

ViewRay Experience. Sasa Mutic, Ph.D. Washington University School of Medicine

ViewRay Experience. Sasa Mutic, Ph.D. Washington University School of Medicine ViewRay Experience Sasa Mutic, Ph.D. Washington University School of Medicine You can observe a lot, just by watching -Yogi Berra Does organ motion matter? A slide often used to illustrate RT capabilities

More information

7/31/2017. Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a. Disclosures

7/31/2017. Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a. Disclosures Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a Dustin Jacqmin, PhD Assistant Professor, Human Oncology University of Wisconsin Madison None Disclosures 1 Outline

More information

Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm

Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm Session V: New Therapy Technologies Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm March 31, 2012 Hosang Jin, Ph.D. Assistant Professor University of Oklahoma Conflict of Interest

More information

Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover

Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover Karl H Rasmussen, Ph.D.. khrasmussen@uthscsa.edu Nema Bassiri bassirin@livemail.uthscsa.edu Division of Medical Physics Cancer

More information

A comparison of two commercial treatment-planning systems for IMRT

A comparison of two commercial treatment-planning systems for IMRT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 3, SUMMER 2005 A comparison of two commercial treatment-planning systems for IMRT M. Peter Petric, 1 Brenda G. Clark, 1 and James L. Robar

More information

Commissioning of the Heidelberg Ion Beam Therapy Centre

Commissioning of the Heidelberg Ion Beam Therapy Centre Commissioning of the Heidelberg Ion Beam Therapy Centre Jacksonville, May 2008 O. Jäkel 1,2, M. Ellerbrock 1, P. Heeg 1, B. Ackermann 1, M. Winter 1, K. Parodi 1, S. Brons 1, A. Mairani 1,2, T. Haberer

More information

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods Accounting for kv Imaging Dose Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota 1 2016 AAPM Annual Meeting - SAM Therapy Educational Course, TU-B-201-2 Washington, DC, August

More information

Title. CitationPhysica Medica, 30(5): Issue Date Doc URL. Type. File Information.

Title. CitationPhysica Medica, 30(5): Issue Date Doc URL. Type. File Information. Title Preliminary analysis for integration of spot-scannin Author(s)Shimizu, S.; Matsuura, T.; Umezawa, M.; Hiramoto, K. CitationPhysica Medica, 30(5): 555-558 Issue Date 2014-07 Doc URL http://hdl.handle.net/2115/56596

More information

Clinical implementation and evaluation of the Acuros dose calculation algorithm

Clinical implementation and evaluation of the Acuros dose calculation algorithm Received: 18 November 2016 Revised: 4 May 2017 Accepted: 12 June 2017 DOI: 10.1002/acm2.12149 RADIATION ONCOLOGY PHYSICS Clinical implementation and evaluation of the Acuros dose calculation algorithm

More information

Complex Treatments in Proton Therapy - Selected Topics. Martin Bues, PhD Proton Therapy M.D. Anderson Cancer Center

Complex Treatments in Proton Therapy - Selected Topics. Martin Bues, PhD Proton Therapy M.D. Anderson Cancer Center Complex Treatments in Proton Therapy - Selected Topics Educational Workshop PTCOG 46, May 18-20, Wanjie, China Martin Bues, PhD Proton Therapy M.D. Anderson Cancer Center Outline Topic 1: Topic 2: Topic

More information

On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors

On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors Drew McVicker, 1 Fang-Fang Yin, 2 Justus D. Adamson

More information

Clinical feasibility of dual tube floor based kilo voltage X-ray localization system for image guided radiation therapy

Clinical feasibility of dual tube floor based kilo voltage X-ray localization system for image guided radiation therapy 2018; 4(3): 384-391 ISSN Print: 2394-7500 ISSN Online: 2394-5869 Impact Factor: 5.2 IJAR 2018; 4(3): 384-391 www.allresearchjournal.com Received: 26-01-2018 Accepted: 27-02-2018 Suresh Rao Department of

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 705: Clinical Rotation 3 Inclusive dates of rotation: Aug. 25, 2015 Feb. 25, 2016 Director or Associate

More information

Commissioning of the Delta 4 Discover An investigation of its influence on the beam quality and of its error detection capabilities Vikren Sarkar,

Commissioning of the Delta 4 Discover An investigation of its influence on the beam quality and of its error detection capabilities Vikren Sarkar, Commissioning of the Delta 4 Discover An investigation of its influence on the beam quality and of its error detection capabilities Vikren Sarkar, Adam Paxton, Bill Salter Questions we aimed to answer

More information

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy Jean-Pierre Bissonnette, PhD, MCCPM Disclosure Work supported, in part, by Elekta Oncology Systems Commercial Interest in Penta-Guide

More information

At that time, do not reference or utilize the initial plan in any manner or study data may be compromised

At that time, do not reference or utilize the initial plan in any manner or study data may be compromised Thank you for participating in this Lung Target Delineation Project being done in conjunction with the SWOG Radiation Oncology Committee. Without your efforts, this project would not be possible. A few

More information

Initial Certification

Initial Certification Initial Certification Therapeutic Medical Physics (TMP) Content Guide Part 2 Content Guide and Sample Questions PLEASE NOTE: List of Constants and Physical Values for Use on the Part 2 Physics Exams The

More information

Transitioning to PBS Proton Therapy From IMRT or Passive Scattered Based Proton Therapy

Transitioning to PBS Proton Therapy From IMRT or Passive Scattered Based Proton Therapy Transitioning to PBS Proton Therapy From IMRT or Passive Scattered Based Proton Therapy Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center July 31, 2017 Progression to better technology 1 Adoption

More information

Implementation of a quality assurance program for computerized treatment planning systems

Implementation of a quality assurance program for computerized treatment planning systems Implementation of a quality assurance program for computerized treatment planning systems Priscilla R. T. L. Camargo and Laura N. Rodrigues Centro de Metrologia das Radiações, Instituto de Pesquisas Energéticas

More information

Computational benchmarks - proposal - Robert Jeraj

Computational benchmarks - proposal - Robert Jeraj Computational benchmarks - proposal - Robert Jeraj Departments of Medical Physics, Human Oncology and Biomedical Engineering University of Wisconsin Madison, WI Increased importance of computer modeling

More information

Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A

Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A Confidence Safety Accuracy Efficiency The innovative Delta4 Discover transmission detector provides confidence and patient safety based on real

More information

Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver

Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver Sotiri Stathakis, PhD, DABR Disclosure ELEKTA Monaco research support I do not have a financial interest in any company or product

More information

Note RADIATION QUALITY OF A TOMOTHERAPY PHOTON FAN BEAM

Note RADIATION QUALITY OF A TOMOTHERAPY PHOTON FAN BEAM Note RADIATION QUALITY OF A TOMOTHERAPY PHOTON FAN BEAM Abstract Tomotherapy, a novel radiotherapy technique, uses narrow fan beams for cancer patient treatment. Photon energy spectra for a rectangular

More information

Going Clinical with the Accuray Radixact System. The Montefiore Experience

Going Clinical with the Accuray Radixact System. The Montefiore Experience Going Clinical with the Accuray Radixact System The Montefiore Experience Disclaimer The views expressed in the presentation are those of the presenters and do not necessarily reflect the views or policies

More information

Treatment planning for metals using an extended CT number scale

Treatment planning for metals using an extended CT number scale JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 Treatment planning for metals using an extended CT number scale John P. Mullins, 1,2a Michael P. Grams, 1 Michael G. Herman, 1 Debra

More information

Approved [insert date]*

Approved [insert date]* Page 1 of 33 1 The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education and professional practice of medical

More information

Clinical trials with hypo-fractionation for breast conserving therapy (BCT): Shorter overall treatment time Effective radiation treatment

Clinical trials with hypo-fractionation for breast conserving therapy (BCT): Shorter overall treatment time Effective radiation treatment AAMD Regional Meeting, March 21st 2015 Clinical trials with hypo-fractionation for breast conserving therapy (BCT): Shorter overall treatment time Effective radiation treatment Whelan et al 2013 & Yu et

More information

Controlli di qualità in SBRT

Controlli di qualità in SBRT Controlli di qualità in SBRT Carmelo Marino 24/25 OTTOBRE 2014 Università degli studi di Milano Each step in the integrated process of RT needs quality control and quality assurance (QA) to prevent errors

More information

Appreciation for discussions with Daniel Low, Kyle Antes, Stephen Kry, and the faculty of Duke University.

Appreciation for discussions with Daniel Low, Kyle Antes, Stephen Kry, and the faculty of Duke University. Jennifer O Daniel, Ph.D. Duke University Medical Center Spring Clinical AAPM Meeting, 2016 Appreciation for discussions with Daniel Low, Kyle Antes, Stephen Kry, and the faculty of Duke University. The

More information

biij Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy

biij Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy Available online at http://www.biij.org/2012/1/e5 doi: 10.2349/biij.8.1.e5 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Dosimetric evaluation of radiation dose rate effect in respiratory

More information

Jon B. Hansen, M.S. Under the supervision of Dr. Larry DeWerd, Ph.D. NCCAAPM Spring Meeting 4/7/2017

Jon B. Hansen, M.S. Under the supervision of Dr. Larry DeWerd, Ph.D. NCCAAPM Spring Meeting 4/7/2017 Jon B. Hansen, M.S. Under the supervision of Dr. Larry DeWerd, Ph.D. NCCAAPM Spring Meeting 4/7/2017 Eye cancer considered rare 2,810 new cases for 2016 in U.S. 1 Can metastasize to liver Formerly treated

More information

This is an Open Access document downloaded from ORCA, Cardiff University's institutional repository:

This is an Open Access document downloaded from ORCA, Cardiff University's institutional repository: This is an Open Access document downloaded from ORCA, Cardiff University's institutional repository: http://orca.cf.ac.uk/76501/ This is the author s version of a work that was submitted to / accepted

More information

Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT

Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT Iori Sumida, 1a Hajime Yamaguchi, 2

More information

RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS

RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS Jean-Claude ROSENWALD Institut Curie, Paris Consultant DOSIsoft Earlier use of computers in radiotherapy 1950 1960 : replacement of tedious

More information

Imaging for Proton Treatment Planning and Verification

Imaging for Proton Treatment Planning and Verification Imaging for Proton Treatment Planning and Verification Jon J. Kruse Mayo Clinic Dept. of Radiation Oncology Rochester, MN Acknowledgements T.J. Whitaker; Mayo Clinic John Mullins; Mayo Clinic Hiroki Shirato,

More information

Software for plan quality control

Software for plan quality control Software for plan quality control D La Russa *1, R Taylor 1, E Henderson 1, K Lekx 1, D Mason 2, (1) The Ottawa Hospital Cancer Centre, Ottawa, ON, (2) R. S. McLaughlin Durham Regional Cancer Centre, Oshawa,

More information

E2E SBRT Thorax Phantom

E2E SBRT Thorax Phantom E2E SBRT Thorax Phantom Model 036A-CVXX-xx SBRT END-TO-END SBRT TESTING SOLUTION SCAN PLAN LOCALIZE TREAT This product is available through: JRT Associates 800-22-0 2428 Almeda Avenue Suite 36 Norfolk,

More information

RPC Prostate Phantom. Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010

RPC Prostate Phantom. Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010 RPC Prostate Phantom Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010 Credentialing for this protocol requires three steps: (1) submission of the Facility Questionaire

More information

AAPM IGRT Recommendations

AAPM IGRT Recommendations AAPM IGRT Recommendations Joint ICTP-IAEA International Workshop on the Implementation of Image Guided Radiotherapy (IGRT) 8-12 May 2017 J. Daniel Bourland, PhD, DABR Professor, Departments of Radiation

More information

Elekta VMAT. Freedom to Treat. Enhanced dose conformance for patient-specific treatment

Elekta VMAT. Freedom to Treat. Enhanced dose conformance for patient-specific treatment Elekta VMAT Freedom to Treat Enhanced dose conformance for patient-specific treatment Speed or Accuracy Why choose? Clinical data has shown that around 50% of patients would benefit from IMRT treatment

More information

Characterization of interplay errors in step-and-shoot IMRT of the lung

Characterization of interplay errors in step-and-shoot IMRT of the lung Characterization of interplay errors in step-and-shoot IMRT of the lung Travis J. McCaw Under the supervision of Dr. Larry DeWerd Medical Radiation Research Center Department of Medical Physics University

More information

IMRT Planning: Concepts and Recommendations of the ICRU report n. 83

IMRT Planning: Concepts and Recommendations of the ICRU report n. 83 School on Medical Physics for Radiation Therapy: Dosimetry and Treatment Planning for Basic and Advanced Applications Trieste - Italy, 27 March - 7 April 2017 IMRT Planning: Concepts and Recommendations

More information

Precise imaging. Personalized treatment. Varian Exact IGRT couch

Precise imaging. Personalized treatment. Varian Exact IGRT couch Precise imaging. Personalized treatment. Varian Exact IGRT couch Exact optimization for IGRT The Exact IGRT couch from Varian is a vital new tool in patient positioning, designed to optimize safety and

More information

Proton Therapy: Application and Quality Assurance

Proton Therapy: Application and Quality Assurance Proton Therapy: Application and Quality Assurance IMRT IMPT Tony Lomax Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Treatment delivery 1. Basics 2. Passive scattering 3. Active

More information

Evaluation of a Ventricular Assist Device: Stability Under X-Rays and Therapeutic Beam Attenuation

Evaluation of a Ventricular Assist Device: Stability Under X-Rays and Therapeutic Beam Attenuation ASAIO Journal 2012 Evaluation of a Ventricular Assist Device: Stability Under X-Rays and Therapeutic Beam Attenuation Michael S. Gossman,* Joel D. Graham, Dan Tamez, Neil Voskoboynikov, and Jeffrey A.

More information

TG-104: In-room kv Computed Tomography for Image-guidance

TG-104: In-room kv Computed Tomography for Image-guidance TG-104: In-room kv Computed Tomography for Image-guidance D.A. Jaffray, Ph.D. Radiation Medicine Program Princess Margaret Hospital/Ontario Cancer Institute Professor Departments of Radiation Oncology

More information

Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy

Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy Mohammad K. Islam a Department of Radiation Physics, Princess Margaret Hospital and Department of Radiation Oncology,

More information

Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218

Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218 Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218 Moyed Miften a) Department of Radiation Oncology, University of Colorado School

More information

Shielding evaluation and acceptance testing of a prefabricated, modular, temporary radiation therapy treatment facility

Shielding evaluation and acceptance testing of a prefabricated, modular, temporary radiation therapy treatment facility JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 5, NUMBER 4, FALL 2004 Shielding evaluation and acceptance testing of a prefabricated, modular, temporary radiation therapy treatment facility Gary A.

More information

Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2

Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2 Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2 Contributors: MD Anderson Cancer Center University of Pennsylvania Laurence Court Dimitris Mihailidis Peter Balter

More information

Thanks to Jeff Williamson, Mike Steinberg, James Purdy

Thanks to Jeff Williamson, Mike Steinberg, James Purdy Thanks to Jeff Williamson, Mike Steinberg, James Purdy Radiation Therapy 1.6M new cancer cases this year in US Approximately 60% of cancer patients receive radiation therapy during the course of their

More information

MV Cone Beam CT Imaging for daily localiza6on: (part II)

MV Cone Beam CT Imaging for daily localiza6on: (part II) MV Cone Beam CT Imaging for daily localiza6on: (part II) Jean Pouliot, Ph.D. Professor UCSF Helen Diller Family Comprehensive Cancer Center AAPM CE-Therapy Series Panel Session 7h30-9h25 am, July 28th,

More information

Monte Carlo Analysis of the Dose Enhancement in the Scalp of Patients due to Titanium Plate Backscatter during Post-Operative Radiotherapy

Monte Carlo Analysis of the Dose Enhancement in the Scalp of Patients due to Titanium Plate Backscatter during Post-Operative Radiotherapy Monte Carlo Analysis of the Dose Enhancement in the Scalp of Patients due to Titanium Plate Backscatter during Post-Operative Radiotherapy Michael T. Hardin Motivation It has been reported that patients

More information

Technical White Paper: An Educational Report On Helical IMRT EXECUTIVE SUMMARY. CRCPD Publication #E-11-3

Technical White Paper: An Educational Report On Helical IMRT EXECUTIVE SUMMARY. CRCPD Publication #E-11-3 CRCPD Publication #E-11-3 Technical White Paper: An Educational Report On Helical IMRT Developed by CRCPD s H-34 Task Force for Helical IMRT EXECUTIVE SUMMARY Helical IMRT (Intensity Modulated Radiation

More information

TG142 Quality Assurance of Medical Accelerators

TG142 Quality Assurance of Medical Accelerators TG142 Quality Assurance of Medical Accelerators Joe Hanley, Ph.D. Hackensack University Medical Center, Hackensack NJ TG142 MEMBERS Chair: Eric E. Klein, Ph.D., Washington University Joe Hanley, Ph.D.,

More information

VMAT Treatment Planning

VMAT Treatment Planning Swedish Medical Center VMAT Treatment Planning David Shepard Swedish Cancer Institute Seattle, WA Founded in 1910 by Dr. Nils Johanson and a group of Seattle s leading Swedish-born businessmen. We treat

More information

Integrating the MR-Linac into Radiation Therapy Practice

Integrating the MR-Linac into Radiation Therapy Practice Integrating the MR-Linac into Radiation Therapy Practice Mikki Campbell MRT(T), MHE Winnie Li MRT(T), MSc UTDRO Evening Journal Club MR-Linac: From Prototype to Clinical January 24, 2019 Objectives To

More information

Disclosures I have provided Physics Consultation Services to Varian and Vision RT

Disclosures I have provided Physics Consultation Services to Varian and Vision RT Surface Imaging to Inform Statistical Quality Control for SRS Treatments Michael J Tallhamer M.Sc. DABR Chief of Radiation Physics Centura Health MichaelTallhamer@Centura.org Quality Improvement and Safety

More information

Treatment planning for scanned proton beams Francesca Albertini, Center for Proton Therapy, Paul Scherrer Institut Switzerland

Treatment planning for scanned proton beams Francesca Albertini, Center for Proton Therapy, Paul Scherrer Institut Switzerland Treatment planning for scanned proton beams Francesca Albertini, Center for Proton Therapy, Paul Scherrer Institut Switzerland Outlook 1. Planning with active scanning protons in practice 2. Single Field

More information

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland AD Award Number: W81XWH- TITLE: PRINCIPAL INVESTIGATOR: CONTRACTING ORGANIZATION: University REPORT DATE: TYPE OF REPORT: Annual PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick,

More information

ASSESSMENT OF THE EFFECT OF BEAM MODIFIERS ON SKIN DOSE FOR EXTERNAL BEAM RADIOTHERAPY USING GAFCHROMIC EBT2 FILMS

ASSESSMENT OF THE EFFECT OF BEAM MODIFIERS ON SKIN DOSE FOR EXTERNAL BEAM RADIOTHERAPY USING GAFCHROMIC EBT2 FILMS i ASSESSMENT OF THE EFFECT OF BEAM MODIFIERS ON SKIN DOSE FOR EXTERNAL BEAM RADIOTHERAPY USING GAFCHROMIC EBT2 FILMS By Komakech Ignatius (10435670) THIS THESIS IS SUBMITTED TO THE DEPARTMENT OF MEDICAL

More information

Spot-Scanning Proton Therapy Patient- Specific Quality Assurance: Results from 309 Treatment Plans

Spot-Scanning Proton Therapy Patient- Specific Quality Assurance: Results from 309 Treatment Plans Spot-Scanning Proton Therapy Patient- Specific Quality Assurance: Results from 309 Treatment Plans Dennis Mackin, PhD 1 ; X. Ronald Zhu, PhD 1 ; Falk Poenisch, PhD 1 ; Heng Li, PhD 1 ; Narayan Sahoo, PhD

More information

Evaluation of water equivalency of Plastic Water for high-energy electron beams using IAEA TRS-398 Code of Practice

Evaluation of water equivalency of Plastic Water for high-energy electron beams using IAEA TRS-398 Code of Practice Radiol Oncol 2004; 38(1): 55-60. Evaluation of water equivalency of Plastic Water for high-energy electron beams using IAEA TRS-398 Code of Practice Božidar Casar 1, Urban Zdešar 2 and Vlado Robar 1 1

More information

Multiple Metastasis Insert for the Lucy 3D QA Phantom

Multiple Metastasis Insert for the Lucy 3D QA Phantom Multiple Metastasis Insert for the Lucy 3D QA Phantom Tobin Hyman, MS DABR Department of Radiation Oncology, McLeod Health, Florence, SC 3120 Deming Way Middleton WI 53562-1461 USA 800-261-4446. ph 608-831-0025.

More information

Optimizing Adaptive Workflows Using RapidPlan at the Beatson West of Scotland Cancer Centre

Optimizing Adaptive Workflows Using RapidPlan at the Beatson West of Scotland Cancer Centre Case Study RapidPlan Adaptive Workflows Optimizing Adaptive Workflows Using RapidPlan at the Beatson West of Scotland Cancer Centre Introduction Adaptive radiotherapy is the practice of adapting to changes

More information

ELECTRON CONFORMAL RADIOTHERAPY USING BOLUS AND INTENSITY MODULATION

ELECTRON CONFORMAL RADIOTHERAPY USING BOLUS AND INTENSITY MODULATION PII S0360-3016(02)02811-0 Int. J. Radiation Oncology Biol. Phys., Vol. 53, No. 4, pp. 1023 1037, 2002 Copyright 2002 Elsevier Science Inc. Printed in the USA. All rights reserved 0360-3016/02/$ see front

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple

Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple Disclosure UAB has Varian research contracts Honorarium from Varian Not your father s Chevy Not your father s Chevy

More information

Effect of Change in Orientation of Enhanced Dynamic Wedges on Radiotherapy Treatment Dose

Effect of Change in Orientation of Enhanced Dynamic Wedges on Radiotherapy Treatment Dose International Journal of Advancements in Research & Technology, Volume 2, Issue 5, May-2013 496 Effect of Change in Orientation of Enhanced Dynamic Wedges on Radiotherapy Treatment Dose Saeed Ahmad Buzdar

More information

Dose uncertainties associated with a set density override of unknown hip prosthetic composition

Dose uncertainties associated with a set density override of unknown hip prosthetic composition Received: 12 February 2017 Revised: 3 April 2017 Accepted: 17 July 2017 DOI: 10.1002/acm2.12167 RADIATION ONCOLOGY PHYSICS Dose uncertainties associated with a set density override of unknown hip prosthetic

More information

TECHNICAL ADVANCE. Dosimetry for Quantitative Analysis of the Effects of Low-Dose Ionizing Radiation in Radiation Therapy Patients

TECHNICAL ADVANCE. Dosimetry for Quantitative Analysis of the Effects of Low-Dose Ionizing Radiation in Radiation Therapy Patients RADIATION RESEARCH 165, 240 247 (2006) 0033-7587/06 $15.00 2006 by Radiation Research Society. All rights of reproduction in any form reserved. Dosimetry for Quantitative Analysis of the Effects of Low-Dose

More information

Recommendations of MPPG #5 and Practical Implementation Strategies

Recommendations of MPPG #5 and Practical Implementation Strategies Recommendations of MPPG #5 and Practical Implementation Strategies (MPPG #5.a.: Commissioning and QA of Treatment Planning Dose Calculations: Megavoltage Photon and Electron Beams) TG 244 Subcommittee

More information

IMRT PHANTOMS CIRS IMRT HOMOGENEOUS PHANTOM CIRS IMRT PHANTOMS C-1

IMRT PHANTOMS CIRS IMRT HOMOGENEOUS PHANTOM CIRS IMRT PHANTOMS C-1 IRS IMRT PHANTOMS IMRT PHANTOMS The IRS IMRT Phantoms for Film and Ion hamber Dosimetry are designed to address the complex issues surrounding commissioning and comparison of treatment planning systems

More information

Part 3 Oral Exam Content Guide

Part 3 Oral Exam Content Guide Initial Certification in Medical Physics Part 3 Oral Exam Content Guide The oral examination is designed to test your knowledge and fitness to practice applied medical physics in the specified specialty(ies).

More information

On the geud biological optimization objective for organs at risk in Photon Optimizer of Eclipse treatment planning system

On the geud biological optimization objective for organs at risk in Photon Optimizer of Eclipse treatment planning system Received: 12 May 2017 Revised: 13 October 2017 Accepted: 16 October 2017 DOI: 10.1002/acm2.12224 RADIATION ONCOLOGY PHYSICS On the geud biological optimization objective for organs at risk in Photon Optimizer

More information

Proton Treatment Planning Systems Daniel Yeung

Proton Treatment Planning Systems Daniel Yeung New Developments in Proton Treatment Planning Systems Daniel Yeung Statement of Disclosure Funded Research & Development: Philips Medical Systems IBA Proton Planning Systems Commercial Systems Academic

More information