Evaluation of water equivalency of Plastic Water for high-energy electron beams using IAEA TRS-398 Code of Practice

Size: px
Start display at page:

Download "Evaluation of water equivalency of Plastic Water for high-energy electron beams using IAEA TRS-398 Code of Practice"

Transcription

1 Radiol Oncol 2004; 38(1): Evaluation of water equivalency of Plastic Water for high-energy electron beams using IAEA TRS-398 Code of Practice Božidar Casar 1, Urban Zdešar 2 and Vlado Robar 1 1 Institute of Oncology, Ljubljana, Slovenia 2 Institute of Occupational Safety, Ljubljana, Slovenia Introduction. In the International Code of Practice for dosimetry TRS-398 published by International Atomic Energy Agency (IAEA), water is recommended as the reference medium for the determination of absorbed dose for high-energy electron beams. Plastic phantoms may be used under certain circumstances for electron beam dosimetry for beam qualities R50 < 4g/cm 2 (E 0 below 10 MeV). In our study, water equivalency of Plastic Water was evaluated in order to determine fluence scaling factors h pl for Plastic Water. Extended set of measurements in water and in Plastic Water were performed. Material and methods. The absorbed dose was determined according to IAEA TRS-398 dosimetry protocol following recommendations for all relevant parameters involved. Water equivalency of Plastic Water was evaluated for five electron beams with nominal energies from 6 MeV to 18 MeV generated by linear accelerator Varian Clinac 2100 C/D. Adequate dosimetry equipment was used throughout the measurements and reference conditions, set by IAEA TRS-398, were followed carefully. Results. The results are presented as ratios D pl of absorbed dose in Plastic Water and water. Upon the selection of electron energy, the ratios vary from with combined uncertainties (1SD) of 0.46% %. From the measured data, the fluence scaling factors h pl were determined and found to be in the range from to Measurements were taken over a period of 18 months, within the frame of a Coordinated Research Project of the International Atomic Energy Agency. Conclusions. Our results are compatible with previously published data. Key words: dosimetry; electron bcams; Plastic Water; IAEA TRS-398 Introduction Received 20 January 2004 Accepted 2 February 2004 Correspondence to: Božidar Casar, Head of Radiophysics Department, Institute of Oncology, Zaloška 2, 1000 Ljubljana; Phone ; Fax: ; bcasar@onko-i.si In the International Code of Practice for dosimetry TRS-398 published in the year 2000 by International Atomic Energy Agency (IAEA), 1 water is recommended as the reference medium for the determination of absorbed dose for high energy photon and electron beams. Plastic phantoms should not be

2 56 Casar B et al. / Evaluation of water equivalency of Plastic Water used for reference dosimetry in photon beams, however, they can be used for routine quality assurance measurements (daily or weekly output checks), provided a transfer factor between plastic and water has been established. According to IAEA TRS-398 dosimetry protocol, plastic phantoms in the form of slabs may be used under certain circumstances for electron beam dosimetry for the beam qualities R 50 < 4 g/cm 2 (E 0 below 10 MeV); their use is permitted when no waterproof chamber is available or when accurate positioning in water is not possible. Presently, many different plastic materials are used for dosimetry purposes in radiotherapy and radiophysics departments: white and clear polystyrene, PMMA, Solid water WT1, Solid water RMI-457, Virtual water, Plastic water and possibly a few others. Several articles comparing the equivalency of various plastics as phantom material to water for electron beam dosimetry have been published. 2-8 Ideally, the phantom material should be water equivalent; that is, it should have the same absorption and scatter properties as water for selected range of photon or electron energies used clinically. In our study we evaluated the water equivalency of Plastic Water developed by Computerized Imaging Reference Systems Inc. Norfolk, VA, USA, also marketed by Nuclear Associates, Inc. Carle Place, NY, USA. We limited our evaluation only to five electron beams within a range of energies from 6 MeV to 18 MeV. The aim of the study was to determine the energy fluence scaling factor h pl for Plastic Water at the selected five electron energies and to compare this factor to the recommended one in the IAEA TRS- 398 dosimetry protocol. Material and methods A. Theoretical background According to IAEA TRS-398 Code of Practice, the calculation of absorbed dose by water D w,q (z ref,w ) for high-energy electron beams at the reference depth z ref,w in water, for the reference beam quality Q, and in the absence of the chamber, is given by the equation [1] where M Q is electrometer reading M 1 corrected for temperature and pressure k T,p, as well as for other influencing quantities - polarity k pol and recombination effects k s. N D,W,Qo is the calibration factor of the selected ionisation chamber in terms of absorbed dose by water in 60 Co beam (reference quality Qo), and k Q,Qo is chamber specific factor correcting for the difference between the beam of reference quality Q o and user quality Q. In TRS- 398, electron beam quality is characterized in terms of half-value depth in water R 50, which is a depth in water where the absorbed dose in water is 50 % of the maximum absorbed dose. The reference depth z ref,w is also specified by R 50 and is given by the equation [2] To determine the absorbed dose in water at z ref,w using a plastic phantom, the reference point of the chamber must be positioned at a scaled reference depth z ref,pl in plastic. For particular beam quality, the measurement depth in plastic expressed in g/cm 2 is obtained from the equation [3] where c pl is a depth-scaling factor. The c pl is the ratio of the average depth of electron penetration in water z av,w and in plastic z av,pl, where these depths are expressed in g/cm 2 [4] Additonally to depth scaling, the electrometer reading M Q,pl at the reference depth in Radiol Oncol 2004; 38(1):

3 Casar B et al. / Evaluation of water equivalency of Plastic Water 57 plastic z ref,pl must be converted to the equivalent reading M Q,w at the reference depth in water z ref,w using the relation [5] where h pl is the fluence scaling factor and is generally energy dependant. The uncertainty associated with this scaling factor is the main reason for avoiding the use of plastic phantoms. B. Experimental conditions and setup Experimental equipment In the study, well guarded waterproof plane parallel ionisation chamber PPC 40 was used together with DOSE 1 electrometer (both produced by IBA Scanditronix - Wellhöfer). The ionisation chamber was calibrated at the IAEA Standard Dosimetry Laboratory in Seibersdorf. The comparison was done for five high-energy electron beams with the energies of 6 MeV, 9 MeV, 12 MeV, 15 MeV and 18 MeV generated by linear accelerator Varian Clinac 2100 C/D. The temperature was monitored with a digital thermometer of the resolution of 0.1 C and the pressure with a digital barometer of resolution of 0.1 mbar. 1D water phantom (produced by MED- Table 1. Chemical composition in terms of fractional weight, nominal density ρ[g/cm 3 ], mean atomic number Ζ and depth scaling factor c pl for Plastic Water. Liquid water data are included for comparison Liquid water Plastic Water H C N O Cl Ca Br ρ[g/cm 3 ] Z c pl TEC) with PMMA walls was used for the measurements in water. The phantom was equipped with a fine mechanical depth adjustment mechanism with the resolution of 0.1 mm. Distilled water was used throughout the measurements. For the measurements in plastic phantom, Plastic Water (cream coloured) in the form of slabs of the size of 30 x 30 cm 2 was used. The thickness of the slabs varied from 1 mm up to 60 mm. Chemical composition (fraction by weight), nominal density, mean atomic number and depth scaling factor for Plastic Water are given in Table 1. For comparison, the data for liquid water are included. Reference conditions and set-up Measurements were done in four sessions on four different days. In each session five measurements were done in water as well as in plastic for five high-energy electron beams. Chamber, water and plastic were left in a bunker for several hours before measurements in order to reach as thermal equilibrium. Before we started with measurements in water, the chamber was dipped into water for at least 15 minutes. Reference conditions were always the same: SSD = 100 cm, 10 x 10 cm 2 electron applicator was used and the irradiation time was 200 MU at a constant dose-rate of 300 MU/min. Gantry and collimator were set to 0 and all the measurements were made along the central axis of the beam. Polarising voltage of the chamber was +300 V - the same as during the chamber calibration. The reference depths were set according to the expressions [2] and [3] for water and plastic, respectively. R 50 was determined in separate relative dosimetry measurements using a computer controlled water phantom (Blue Phantom made by IBA Scanditronix - Wellhöfer), where the data were collected in 0.4 mm increments. As the thinnest available slab was 1 mm thick, the actual depth of the Radiol Oncol 2004; 38(1):

4 58 Casar B et al. / Evaluation of water equivalency of Plastic Water chamber reference point in plastic was at the depth that was nearest to the calculated one and not exactly at the calculated one. However, the differences were small. As defined in TRS-398 dosimetry protocol, the reference point of the chamber (effective point of measurement) is at the inner surface of the entrance window. Reference depths and some other chamber parameters for selected set of energies are presented in Table 2. When dipping the chamber into water we were careful not to trap any air bubble at the chamber bottom because it could lower the absorbed dose. For setting up the chamber in plastic, a special cylindrical disc made of white polystyrene was fitted in the chamber hole at its bottom. This was to ensure that no scattered radiation would be lacking due to the absence of scattered material at the chambers bottom. One of the blocks was machined to fit exactly to ionisation chamber PPC40 so that the entrance window of the chamber was at the level of one surface of the block. Under the point of measurement, 6 cm of Plastic Water was always kept to provide an adequate backscatter conditions. Results and discussion The results are presented in Table 3 and Figure 1, as the dose ratios Plastic Water/water as a function of nominal beam energy [6] Depending upon the beam energy, the ratios varied from to From the calculated ratios, the fluence scaling factors h pl can be determined. The measured doses in Plastic Water are within 1% of those measured in water for all electron beam energies, and all the ratios are higher than 1.0 (apart from the ratios for 18 MeV electron beam, where ratios are lower than 1.0). Combined measurement uncertainties (1SD) of type A and type B (detailed explanation about uncertainties is given in reference1), joining the uncertainties from re- Table 3. Ratios D pl of absorbed doses measured in Plastic Water and in water for five high energy electron beams produced by Varian Clinac 2100 C/D linear accelerator. Combined measurement uncertainties (1SD) of type A and type B (detailed explanation about uncertainties is given in reference1), joining the uncertainties from repeated measurements from four sessions and estimated uncertainties due to setup and other influencing quantities (pressure, temperature.) Corresponding fluence scaling factors h pl for each electron energy are presented without standard deviations. D pl h pl 6 MeV ± MeV ± MeV ± MeV ± MeV ± Table 2. Various beam and ionisation chamber (PPC40) parameters used for calculation and measurements of the absorbed dose in water for high energy electron beams generated by linear accelerator Varian Clinac 2100 C/D 6 MeV 9 MeV 12 MeV 15 MeV 18 MeV R50 [cm] Rp [cm] z a ref,w [cm] z b ref,pl [cm] z c ref,pl [cm] a Reference depth in water of the effective measurement point of the chamber according to TRS-398 b Reference depth in Plastic Water of the effective measurement point of the chamber according to TRS-398 obtained from formula expression [3] c Actual depth in PW of the effective measurement point of the chamber due to limitation of minimal slab thickness of 1 mm Radiol Oncol 2004; 38(1):

5 Casar B et al. / Evaluation of water equivalency of Plastic Water 59 Figure 1. Ratios D pl of absorbed doses determined with measurements in Plastic Water and in water for five high-energy electron beams - electron beam qualities. Measurements were performed with plane parallel ionisation chamber PPC40. Beam data are in Table 2. peated measurements from four sessions and the estimated uncertainties due to setup and other influencing quantities (pressure, temperature) are presented in Table 3 and are in the range from 0.46% to 0.68%. In our study, the average value for h pl for energies below 10 MeV is , which is in line with the value published in TRS-398, where h pl is Average value of h pl for all electron energies in our study is A slight disagreement with results obtained by Tello et al. 5 was observed, but well within the reported uncertainties. We can conclude that our results confirm previously published data for h pl for Plastic Water. However, we must emphasize, that the temperature of the air in the chamber cavity was probably not the same as the temperature of water when the measurements in water phantom were performed. Due to specific temperature conditions in the accelerators bunker, we could assume that the temperature of the air in the chamber cavity was always at least a little bit higher than the measured water temperature; this, sometimes large difference (up to 3 C) between the temperature measured in water and the room temperature was due to a slow but permanent rising of room temperature (air condition didn't work optimally). The measured and reported ratios D pl could thus be too high by up to 0.3%, which corresponds to the temperature difference of 1 C. As it was not possible to measure actual temperature of the air in the chamber cavity, we included 0.3% of possible temperature variation in the uncertainties of our measurements, rather that in the systematic errors. We can conclude, that when no water phantom is available in the clinic, or when the use of plastic phantom would be less time consuming or, from any other reason, more appropriate for physicists, the Plastic Water phantom can be used for routine constancy checks of high energy electron beams within the energy range checked in this study and also taking into account the fluence scaling factors as suggested in this study. Even if we take an average fluence scaling factor for all beams in the energy range from 6 MeV - 18 MeV, which is in our case , the estimated difference of the absorbed dose determination in Plastic Water should be below 1% comparing to the absorbed dose in water. However, before Plastic Water is to be used as water substitute for reference dosimetry, a careful comparison with measurement in water should be performed. Acknowledgements Present study was done in the framework of Co-ordinated Research Project (CRP E ) of the International Atomic Energy Agency. The main goal of the extended project was testing of the procedures recommended in TRS-398 dosimetry protocol for different types of radiation beams and ionisation chambers as well as comparison of the results obtained with the existing protocols. Authors would like to thank colleagues in the project group: Mariella Affonseca, Ivaldo Ferreira, Günther Hartmann, Saiful Huq, Govinda Rajan and specially to Scientific Officer of the project Stanislav Vatnitsky. Radiol Oncol 2004; 38(1):

6 60 Casar B et al. / Evaluation of water equivalency of Plastic Water References 1. IAEA International Atomic Energy Agency. Absorbed dose determination in external beam radiotherapy: An international Code of Practice for Dosimetry Based Standards of Absorbed Dose to Water. IAEA Technical Reports Series No Vienna: IAEA; Thwaites DI. Measurements of ionisation in water, polystyrene and 'solid water' phantom material for electron beams. Phys Med Biol 1985; 30: Bruinvis IAD, Heukelom S, Mijnheer BJ. Comparison of ionisation measurements in water and polystyrene for electron beam dosimetry. Phys Med Biol 1985; 30: Thomadsen B, Constantinou C, Ho A. Evaluation of water-equivalent plastics as phantom material for electron-beam dosimetry. Med Phys 1995; 22: Tello VM, Tailor RC, Hanson WF. How water equivalent are water-equivalent solid materials for output calibration of photon and electron beams? Med Phys 1995; 22: Nilsson B, Montelius A, Andreo P, Johansson J. Correction factors for parallel-plate chambers used in plastic phantoms in electron dosimetry. Phys Med Biol 1997; 42: Björeland A, Svensson H. Testing of the Code of Practice at different electron energies and phantom material. IAEA-TECDOC-1173: Review of data and methods recommended in the international code of practice for dosimetry IAEA Technical Reports Series No. 381, The use of plane parallel Ionization chambers in high energy electron and photon beams. Vienna: IAEA; p Saitoh H, Tomaru T, Fujisaki T, Abe S, Myojoyama A, Fukuda K. A study on properties of water substitute solid phantom using EGS code. Proceedings of the Tenth EGS4 Users Meeting in Japan, KEK Proceedings 2002; 18: Radiol Oncol 2004; 38(1):

Kilovoltage X-ray dosimetry in the clinic. Philip Mayles The Clatterbridge Cancer Centre

Kilovoltage X-ray dosimetry in the clinic. Philip Mayles The Clatterbridge Cancer Centre Kilovoltage X-ray dosimetry in the clinic Philip Mayles The Clatterbridge Cancer Centre Kilovoltage Codes of Practice Germany DIN 1988 and 1996 UK IPEMB (IPEM) 1996 and 2005 IAEA TRS 277 2nd edn 1997 Holland

More information

Quality Assurance for particle beam therapy

Quality Assurance for particle beam therapy Quality Assurance for particle beam therapy PTCOG Educational Workshop, Essen 2013 Heidelberg Ion Beam Therapy Center at the University Hospital German Cancer Research Center, Heidelberg, Germany Outline

More information

Ionisation chamber array, for setting up of a linac, and comparison to a 3D Water Tank. Presented by Neil McCann Systems Engineer for Elekta

Ionisation chamber array, for setting up of a linac, and comparison to a 3D Water Tank. Presented by Neil McCann Systems Engineer for Elekta Ionisation chamber array, for setting up of a linac, and comparison to a 3D Water Tank Presented by Neil McCann Systems Engineer for Elekta Why new dosimetry equipment? Replace our ageing IBA Buddelship

More information

Design and dosimetry characteristics of a commercial applicator system for intra-operative electron beam therapy utilizing ELEKTA Precise accelerator

Design and dosimetry characteristics of a commercial applicator system for intra-operative electron beam therapy utilizing ELEKTA Precise accelerator JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 4, fall 2010 Design and dosimetry characteristics of a commercial applicator system for intra-operative electron beam therapy utilizing ELEKTA

More information

Dosimetry of Small IMRT Fields

Dosimetry of Small IMRT Fields Dosimetry of Small IMRT Fields Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 IMRT and Small Fields IMRT beams typically contain small

More information

ArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools

ArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools ArcCHECK & 3DVH The Ultimate 4D Patient QA Solution Your Most Valuable QA and Dosimetry Tools THE 4D QA GOLD STANDARD The ArcCHECK is the world s most selected independent 4D measurement array. Simply

More information

Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment

Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment Anup Kumar Bhardwaj et al Study of 2D ion chamber array for angular response and QA of dynamic MLC... Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment IMRT plans

More information

Clinically relevant investigation of flattening filter-free skin dose

Clinically relevant investigation of flattening filter-free skin dose JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016 Clinically relevant investigation of flattening filter-free skin dose Christopher L. Guy, 1 Kishor Karki, 1 Manju Sharma, 1,2 and

More information

Co γ -ray attenuation coefficient of barite concrete

Co γ -ray attenuation coefficient of barite concrete Iran. J. Radiat. Res., 2006; 4 (2): 71-75 60 Co γ -ray attenuation coefficient of barite concrete F. Bouzarjomehri 1*,T. Bayat 2, M.H. Dashti R 3, J. Ghisari 4, N.Abdoli 5 INTRODUCTION 1 Department of

More information

Shielding evaluation for IMRT implementation in an existing accelerator vault

Shielding evaluation for IMRT implementation in an existing accelerator vault JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 3, SUMMER 2003 Shielding evaluation for IMRT implementation in an existing accelerator vault R. A. Price,* O. Chibani, and C.-M. Ma Department

More information

Implementation of a quality assurance program for computerized treatment planning systems

Implementation of a quality assurance program for computerized treatment planning systems Implementation of a quality assurance program for computerized treatment planning systems Priscilla R. T. L. Camargo and Laura N. Rodrigues Centro de Metrologia das Radiações, Instituto de Pesquisas Energéticas

More information

Patient doses from CT examinations in region of Prishtina, Kosovo

Patient doses from CT examinations in region of Prishtina, Kosovo Patient doses from CT examinations in region of Prishtina, Kosovo Sehad KADIRI, Gëzim HODOLLI, Kostandin DOLLANI Institute of Occupational Medicine, Radiation Protection Service, Obiliq, Kosovo Institute

More information

VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry

VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry Journal of Radiotherapy in Practice Journal of Radiotherapy in Practice (2013) Page 1 of 7 & Cambridge University Press 2013 doi:10.1017/s1460396912000556 Technical Note VMAT monthly QA using two techniques:

More information

Initial Certification

Initial Certification Initial Certification Therapeutic Medical Physics (TMP) Content Guide Part 2 Content Guide and Sample Questions PLEASE NOTE: List of Constants and Physical Values for Use on the Part 2 Physics Exams The

More information

Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system

Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 1, WINTER 2008 Experimental evaluation of the accuracy of skin dose calculation for a commercial treatment planning system Laurence E. Court,

More information

Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover

Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover Karl H Rasmussen, Ph.D.. khrasmussen@uthscsa.edu Nema Bassiri bassirin@livemail.uthscsa.edu Division of Medical Physics Cancer

More information

SPECIFICATION STATEMENT OF WORK. Performance Specification for a radiotherapy linear accelerator

SPECIFICATION STATEMENT OF WORK. Performance Specification for a radiotherapy linear accelerator Specifications / Statement of Work / Terms of Reference SPECIFICATION STATEMENT OF WORK Performance Specification for a radiotherapy linear accelerator 1. Scope These specifications describe the requirements

More information

Uncertainties in treatment planning dose computation. Jeffrey V. Siebers

Uncertainties in treatment planning dose computation. Jeffrey V. Siebers Uncertainties in treatment planning dose computation Jeffrey V. Siebers Acknowledgements Dr David Followill, RPC for access to RPC data Drs. Catharine Clark, Pablo Carrasco, Warren Kilby, Guanghua Yan

More information

Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple

Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple Disclosure UAB has Varian research contracts Honorarium from Varian Not your father s Chevy Not your father s Chevy

More information

Title. CitationPhysica Medica, 30(5): Issue Date Doc URL. Type. File Information.

Title. CitationPhysica Medica, 30(5): Issue Date Doc URL. Type. File Information. Title Preliminary analysis for integration of spot-scannin Author(s)Shimizu, S.; Matsuura, T.; Umezawa, M.; Hiramoto, K. CitationPhysica Medica, 30(5): 555-558 Issue Date 2014-07 Doc URL http://hdl.handle.net/2115/56596

More information

Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems

Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 3, SUMMER 2005 Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems Charles M. Able and Michael D. Thomas

More information

Proposal on Guideline for Quality Assurance of Radiation Treatment Planning System

Proposal on Guideline for Quality Assurance of Radiation Treatment Planning System Original Article PMP Progress in Medical Physics 28(4), December 2017 https://doi.org/10.14316/pmp.2017.28.4.197 pissn 2508-4445, eissn 2508-4453 Proposal on Guideline for Quality Assurance of Radiation

More information

Reduction of Backscattered Radiation in Enclosure X-ray Radiography

Reduction of Backscattered Radiation in Enclosure X-ray Radiography Abstract Reduction of Backscattered Radiation in Enclosure X-ray Radiography Samir Abdul-Majid, Abdulrahim Kinsara, Abdullah Almasoumi and Mohamed Kallothody Faculty of Engineering, King Abdulaziz University

More information

Radiation Protection Dosimetry (2008), Vol. 129, No. 4, pp Advance Access publication 30 October 2007

Radiation Protection Dosimetry (2008), Vol. 129, No. 4, pp Advance Access publication 30 October 2007 Radiation Protection Dosimetry (2008), Vol. 129, No. 4, pp. 419 425 Advance Access publication 30 October 2007 doi:10.1093/rpd/ncm456 IMAGE QUALITY EVALUATION AND PATIENT DOSE ASSESSMENT OF MEDICAL FLUOROSCOPIC

More information

An improvement in IMRT QA results and beam matching in linacs using statistical process control

An improvement in IMRT QA results and beam matching in linacs using statistical process control JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 An improvement in IMRT QA results and beam matching in linacs using statistical process control Justin D. Gagneur a and Gary A. Ezzell

More information

Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy

Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy UCRL-JC-133076 PREPRINT Clinical Implementation of the PEREGRINE Monte Carlo Dose Calculations System for Photon Beam Therapy C.L. Hartmann Siantar, P.M. Bergstrom T.P. Daly, M. Descalle D. Garrett, R.K.

More information

RapidArc: Clinical Implementation

RapidArc: Clinical Implementation RapidArc: Clinical Implementation Fang-Fang Yin, PhD Q. Jackie Wu, PhD Acknowledgements Team efforts from staff at Duke Radiation Oncology, especially to Dr. J Chang, Dr. J O Daniel for providing slide

More information

I N a recent paper Rawlinson and Johns6 showed that a 2.5 MeV linear accelerator

I N a recent paper Rawlinson and Johns6 showed that a 2.5 MeV linear accelerator VOL. II, No. 4 DESIGN OF X-RAY TARGETS FOR HIGH ENERGY LINEAR ACCELERATORS IN RADIOTHERAPY* By E. B. PODGORAK, PH.D., J. A. RAWLINSON, M.Sc., M. I. GLAVINOVI, B.SC., and H. E. JOHNS, PH.D. I N a recent

More information

Imaging/Imagine Needs for Proton Therapy: Treatment Planning. Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA

Imaging/Imagine Needs for Proton Therapy: Treatment Planning. Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA Imaging/Imagine Needs for Proton Therapy: Treatment Planning Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA AAPM Annual Meeting Indianapolis, Aug. 07, 2013 Disclosure Software licensing agreement

More information

Proton Therapy: Application and Quality Assurance

Proton Therapy: Application and Quality Assurance Proton Therapy: Application and Quality Assurance IMRT IMPT Tony Lomax Centre for Proton Radiotherapy, Paul Scherrer Institute, Switzerland Treatment delivery 1. Basics 2. Passive scattering 3. Active

More information

Absorbed Dose and Air Kerma Primary Standards

Absorbed Dose and Air Kerma Primary Standards Absorbed Dose and Air Kerma Primary Standards Paris, 9-11 May, 2007 Procedure for absorbed dose to water determination in high energy photon and electron beams by ferrous sulphate dosimeter at M. Pimpinella,

More information

IAEA-TECDOC-CD Intercomparison of Measurements of Personal Dose Equivalent Hp(10) in Photon Fields in the West Asia Region

IAEA-TECDOC-CD Intercomparison of Measurements of Personal Dose Equivalent Hp(10) in Photon Fields in the West Asia Region IAEA-TECDOC-CD-1567 Intercomparison of Measurements of Personal Dose Equivalent Hp(10) in Photon Fields in the West Asia Region August 2007 IAEA SAFETY RELATED PUBLICATIONS IAEA SAFETY STANDARDS Under

More information

Complex Treatments in Proton Therapy - Selected Topics. Martin Bues, PhD Proton Therapy M.D. Anderson Cancer Center

Complex Treatments in Proton Therapy - Selected Topics. Martin Bues, PhD Proton Therapy M.D. Anderson Cancer Center Complex Treatments in Proton Therapy - Selected Topics Educational Workshop PTCOG 46, May 18-20, Wanjie, China Martin Bues, PhD Proton Therapy M.D. Anderson Cancer Center Outline Topic 1: Topic 2: Topic

More information

Methodology for Quality Control of Treatment Planning Systems for Use in Radiation Teletherapy

Methodology for Quality Control of Treatment Planning Systems for Use in Radiation Teletherapy The Open Medical Devices Journal, 29, 1, 29-34 29 Open Access Methodology for Quality Control of Treatment Planning Systems for Use in Radiation Teletherapy E. M. Simão 1, L. E. A. Macedo 2 and D. N. Souza

More information

Jon B. Hansen, M.S. Under the supervision of Dr. Larry DeWerd, Ph.D. NCCAAPM Spring Meeting 4/7/2017

Jon B. Hansen, M.S. Under the supervision of Dr. Larry DeWerd, Ph.D. NCCAAPM Spring Meeting 4/7/2017 Jon B. Hansen, M.S. Under the supervision of Dr. Larry DeWerd, Ph.D. NCCAAPM Spring Meeting 4/7/2017 Eye cancer considered rare 2,810 new cases for 2016 in U.S. 1 Can metastasize to liver Formerly treated

More information

Monte Carlo Analysis of the Dose Enhancement in the Scalp of Patients due to Titanium Plate Backscatter during Post-Operative Radiotherapy

Monte Carlo Analysis of the Dose Enhancement in the Scalp of Patients due to Titanium Plate Backscatter during Post-Operative Radiotherapy Monte Carlo Analysis of the Dose Enhancement in the Scalp of Patients due to Titanium Plate Backscatter during Post-Operative Radiotherapy Michael T. Hardin Motivation It has been reported that patients

More information

Monte Carlo Study of a New Mobile Electron Accelerator Head for Intra Operative Radiation Therapy (IORT)

Monte Carlo Study of a New Mobile Electron Accelerator Head for Intra Operative Radiation Therapy (IORT) Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.181-186 (2011) ARTICLE Monte Carlo Study of a New Mobile Electron Accelerator Head for Intra Operative Radiation Therapy (IORT) Anna WYSOCKA-RABIN

More information

biij Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy

biij Dosimetric evaluation of radiation dose rate effect in respiratory gated intensity modulated radiation therapy Available online at http://www.biij.org/2012/1/e5 doi: 10.2349/biij.8.1.e5 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Dosimetric evaluation of radiation dose rate effect in respiratory

More information

Shielding evaluation and acceptance testing of a prefabricated, modular, temporary radiation therapy treatment facility

Shielding evaluation and acceptance testing of a prefabricated, modular, temporary radiation therapy treatment facility JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 5, NUMBER 4, FALL 2004 Shielding evaluation and acceptance testing of a prefabricated, modular, temporary radiation therapy treatment facility Gary A.

More information

Effect of Change in Orientation of Enhanced Dynamic Wedges on Radiotherapy Treatment Dose

Effect of Change in Orientation of Enhanced Dynamic Wedges on Radiotherapy Treatment Dose International Journal of Advancements in Research & Technology, Volume 2, Issue 5, May-2013 496 Effect of Change in Orientation of Enhanced Dynamic Wedges on Radiotherapy Treatment Dose Saeed Ahmad Buzdar

More information

Preliminary tests of a commercial extrapolation chamber for utilization as a beta radiation standard system

Preliminary tests of a commercial extrapolation chamber for utilization as a beta radiation standard system Preliminary tests of a commercial extrapolation chamber for utilization as a beta radiation standard system Patrícia L. Antonio, Marcos Xavier and Linda V. E. Caldas Instituto de Pesquisas Energéticas

More information

PATIENT DOSIMETRY IN INTERVENTIONAL RADIOLOGY

PATIENT DOSIMETRY IN INTERVENTIONAL RADIOLOGY 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 PATIENT DOSIMETRY

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 705: Clinical Rotation 3 Inclusive dates of rotation: Aug. 25, 2015 Feb. 25, 2016 Director or Associate

More information

Numerical Simulation of the Medical Linear Accelerator Electron Beams Absorption by ABS-Plastic doped with Metal

Numerical Simulation of the Medical Linear Accelerator Electron Beams Absorption by ABS-Plastic doped with Metal Journal of Physics: Conference Series PAPER OPEN ACCESS Numerical Simulation of the Medical Linear Accelerator Electron Beams Absorption by ABS-Plastic doped with Metal To cite this article: S G Stuchebrov

More information

MONTE CARLO SIMULATION OF THE ELECTRON BEAMS PRODUCED BY A LINEAR ACCELERATOR FOR INTRA- OPERATIVE RADIATION THERAPY *

MONTE CARLO SIMULATION OF THE ELECTRON BEAMS PRODUCED BY A LINEAR ACCELERATOR FOR INTRA- OPERATIVE RADIATION THERAPY * Romanian Reports in Physics, Vol. 66, No. 1, P. 61 74, 2014 MONTE CARLO SIMULATION OF THE ELECTRON BEAMS PRODUCED BY A LINEAR ACCELERATOR FOR INTRA- OPERATIVE RADIATION THERAPY * D. MIHAILESCU, C. BORCIA

More information

NIIEFA Accelerators for Industry and Medicine

NIIEFA Accelerators for Industry and Medicine FSUE D.V. Efremov Scientific Research Institute of Electrophysical Apparatus ( NIIEFA ) Saint-Petersburg,, Russia NIIEFA Accelerators for Industry and Medicine M.F. Vorogushin vorogushin@luts.niiefa.spb.su

More information

AAPM IGRT Recommendations

AAPM IGRT Recommendations AAPM IGRT Recommendations Joint ICTP-IAEA International Workshop on the Implementation of Image Guided Radiotherapy (IGRT) 8-12 May 2017 J. Daniel Bourland, PhD, DABR Professor, Departments of Radiation

More information

Computational benchmarks - proposal - Robert Jeraj

Computational benchmarks - proposal - Robert Jeraj Computational benchmarks - proposal - Robert Jeraj Departments of Medical Physics, Human Oncology and Biomedical Engineering University of Wisconsin Madison, WI Increased importance of computer modeling

More information

RPC Prostate Phantom. Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010

RPC Prostate Phantom. Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010 RPC Prostate Phantom Guidelines for Planning and Treating the RPC IMRT Prostate Phantom. Revised November 2010 Credentialing for this protocol requires three steps: (1) submission of the Facility Questionaire

More information

Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm

Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm Session V: New Therapy Technologies Patient DVH based QA metrics using the Planned Dose Perturbation Algorithm March 31, 2012 Hosang Jin, Ph.D. Assistant Professor University of Oklahoma Conflict of Interest

More information

8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)*

8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)* and Quality Assurance Considerations for Volumetric Modulated Arc Therapy * What is? Report of writing group from TETAWG under Therapy Physics Committee (TPC) of AAPM Writing Group members: James M. Galvin,

More information

ASSESSMENT OF THE EFFECT OF BEAM MODIFIERS ON SKIN DOSE FOR EXTERNAL BEAM RADIOTHERAPY USING GAFCHROMIC EBT2 FILMS

ASSESSMENT OF THE EFFECT OF BEAM MODIFIERS ON SKIN DOSE FOR EXTERNAL BEAM RADIOTHERAPY USING GAFCHROMIC EBT2 FILMS i ASSESSMENT OF THE EFFECT OF BEAM MODIFIERS ON SKIN DOSE FOR EXTERNAL BEAM RADIOTHERAPY USING GAFCHROMIC EBT2 FILMS By Komakech Ignatius (10435670) THIS THESIS IS SUBMITTED TO THE DEPARTMENT OF MEDICAL

More information

Preliminary Study for Dosimetric Characteristics of 3Dprinted. Materials with Megavoltage Photons

Preliminary Study for Dosimetric Characteristics of 3Dprinted. Materials with Megavoltage Photons Preliminary Study for Dosimetric Characteristics of 3Dprinted Materials with Megavoltage Photons Seonghoon Jeong and Myonggeun Yoon Department of Bio-Convergence Engineering, Korea University, Seoul, Korea

More information

Radiation protection CRISTER CEBERG

Radiation protection CRISTER CEBERG Radiation protection CRISTER CEBERG International standards in RP» International Commission on Radiological Protection (ICRP)» International Atomic Energy Agency (IAEA)» The European Union (EU) Science,

More information

Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A

Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A Confidence Safety Accuracy Efficiency The innovative Delta4 Discover transmission detector provides confidence and patient safety based on real

More information

Dose uncertainties associated with a set density override of unknown hip prosthetic composition

Dose uncertainties associated with a set density override of unknown hip prosthetic composition Received: 12 February 2017 Revised: 3 April 2017 Accepted: 17 July 2017 DOI: 10.1002/acm2.12167 RADIATION ONCOLOGY PHYSICS Dose uncertainties associated with a set density override of unknown hip prosthetic

More information

DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM

DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM Dee-Ann Radford, David S. Followill, Peter A. Balter, William F. Hanson Department of Radiation Physics The

More information

RT Workspace from Standard Imaging

RT Workspace from Standard Imaging RT Workspace from Standard Imaging In 1895 Wilhelm Röntgen discovers the X-ray (receives Nobel prize in 1901) From 1900 1950 radium was chiefly used for radiation treatment (based on work of Marie Curie

More information

Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2

Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2 Experience and Suggestions on Commissioning and QA Procedures for Halcyon Version 1.2 Contributors: MD Anderson Cancer Center University of Pennsylvania Laurence Court Dimitris Mihailidis Peter Balter

More information

The quenching effect in PRESAGE dosimetry of proton beams: Is an empirical correction feasible?

The quenching effect in PRESAGE dosimetry of proton beams: Is an empirical correction feasible? The quenching effect in PRESAGE dosimetry of proton beams: Is an empirical correction feasible? S J Doran 1,2, T Gorjiara 3, J Adamovics 4,5, Z Kuncic 3, A Kacperek 6 and C Baldock 3 1 CRUK Cancer Imaging

More information

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods Accounting for kv Imaging Dose Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota 1 2016 AAPM Annual Meeting - SAM Therapy Educational Course, TU-B-201-2 Washington, DC, August

More information

Optimization of anode-filter combinations in digital mammography: An experimental study

Optimization of anode-filter combinations in digital mammography: An experimental study Optimization of anode-filter combinations in digital mammography: An experimental study Poster No.: C-2915 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: A. Peruzzo Cornetto

More information

TECHNICAL ADVANCE. Dosimetry for Quantitative Analysis of the Effects of Low-Dose Ionizing Radiation in Radiation Therapy Patients

TECHNICAL ADVANCE. Dosimetry for Quantitative Analysis of the Effects of Low-Dose Ionizing Radiation in Radiation Therapy Patients RADIATION RESEARCH 165, 240 247 (2006) 0033-7587/06 $15.00 2006 by Radiation Research Society. All rights of reproduction in any form reserved. Dosimetry for Quantitative Analysis of the Effects of Low-Dose

More information

Assessment of Gamma Dose Rate for Hypothetical Radioactive Waste Container

Assessment of Gamma Dose Rate for Hypothetical Radioactive Waste Container Research Article imedpub Journals http://www.imedpub.com International Journal of Applied Science - Research and Review DOI: 10.21767/2394-9988.100055 Assessment of Gamma Dose Rate for Hypothetical Radioactive

More information

Part 3 Oral Exam Content Guide

Part 3 Oral Exam Content Guide Initial Certification in Medical Physics Part 3 Oral Exam Content Guide The oral examination is designed to test your knowledge and fitness to practice applied medical physics in the specified specialty(ies).

More information

Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT

Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Three-dimensional dose prediction based on two-dimensional verification measurements for IMRT Iori Sumida, 1a Hajime Yamaguchi, 2

More information

ECONOMICAL ASPECTS OF RADIATION STERILIZATION WITH ELECTRON BEAM

ECONOMICAL ASPECTS OF RADIATION STERILIZATION WITH ELECTRON BEAM IAEA-SM-350/35 ECONOMICAL ASPECTS OF RADIATION STERILIZATION WITH ELECTRON BEAM Z. ZIMEK, I. KALUSKA XA9847737 Department of Radiation Chemistry and Technology, Institute of Nuclear Chemistry and Technology,

More information

Approved [insert date]*

Approved [insert date]* Page 1 of 33 1 The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education and professional practice of medical

More information

Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology

Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology A. Neocleous *1, E. Yakoumakis 1, G. Gialousis 2, A. Dimitriadis 1, N. Yakoumakis 3 and E. Georgiou 1 1 Medical Physics Department,

More information

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.

More information

Commissioning and initial experience with the first clinical gantry-mounted proton therapy system

Commissioning and initial experience with the first clinical gantry-mounted proton therapy system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Commissioning and initial experience with the first clinical gantry-mounted proton therapy system Tianyu Zhao, a Baozhou Sun, Kevin

More information

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy Jean-Pierre Bissonnette, PhD, MCCPM Disclosure Work supported, in part, by Elekta Oncology Systems Commercial Interest in Penta-Guide

More information

Disclosures. Outline 8/2/2017. Emerging Technologies for IORT: Unidirectional Planar Brachytherapy Sources. Julius V Turian PhD 8/2/2017

Disclosures. Outline 8/2/2017. Emerging Technologies for IORT: Unidirectional Planar Brachytherapy Sources. Julius V Turian PhD 8/2/2017 Emerging Technologies for IORT: Unidirectional Planar Brachytherapy Sources 8/2/2017 59 th AAPM Annual Meeting Denver CO Session Number: WE-B-702-04 Disclosures Rush University is participating in a pancreatic

More information

Transitioning to PBS Proton Therapy From IMRT or Passive Scattered Based Proton Therapy

Transitioning to PBS Proton Therapy From IMRT or Passive Scattered Based Proton Therapy Transitioning to PBS Proton Therapy From IMRT or Passive Scattered Based Proton Therapy Mark Pankuch, PhD Northwestern Medicine Chicago Proton Center July 31, 2017 Progression to better technology 1 Adoption

More information

Going Clinical with the Accuray Radixact System. The Montefiore Experience

Going Clinical with the Accuray Radixact System. The Montefiore Experience Going Clinical with the Accuray Radixact System The Montefiore Experience Disclaimer The views expressed in the presentation are those of the presenters and do not necessarily reflect the views or policies

More information

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures Vano E (1,2), Aviles P (1,2), Prieto C (2), Fernandez JM (1,2), Guibelalde

More information

To the Graduate Council:

To the Graduate Council: To the Graduate Council: I am submitting herewith a thesis written by Michelle Lynn Neeley entitled Patient- Specific Quality Assurance for IMRT. I have examined the final electronic copy of this thesis

More information

2017 ACR Computed Tomography Quality Control Manual FAQS

2017 ACR Computed Tomography Quality Control Manual FAQS Updated 11-15-2017 2017 ACR Computed Tomography Quality Control Manual FAQS Q. The updated 2017 ACR Computed Tomography Quality Control Manual has been released. (Visit www.acr.org/education/education-catalog.)

More information

RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS

RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS Jean-Claude ROSENWALD Institut Curie, Paris Consultant DOSIsoft Earlier use of computers in radiotherapy 1950 1960 : replacement of tedious

More information

Iranian Journal of Medical Physics

Iranian Journal of Medical Physics Iranian Journal of Medical Physics ijmp.mums.ac.ir Absorbed Dose Calculation In Irregular Blocked Radiation Fields: Evaluation of Clarkson s Sector Integration Method for Radiation Fields Commonly Used

More information

Effects of 16 bit CT imaging scanning conditions for metal implants on radiotherapy dose distribution

Effects of 16 bit CT imaging scanning conditions for metal implants on radiotherapy dose distribution ONCOLOGY LETTERS 15: 2373-2379, 2018 Effects of 16 bit CT imaging scanning conditions for metal implants on radiotherapy dose distribution LIUGANG GAO 1, HONGFEI SUN 1, XINYE NI 1, MINGMING FANG 2 and

More information

E2E SBRT Thorax Phantom

E2E SBRT Thorax Phantom E2E SBRT Thorax Phantom Model 036A-CVXX-xx SBRT END-TO-END SBRT TESTING SOLUTION SCAN PLAN LOCALIZE TREAT This product is available through: JRT Associates 800-22-0 2428 Almeda Avenue Suite 36 Norfolk,

More information

Michael Gillin, PhD Professor, Chief of Clinical Services, Dept. of Radiation Physics

Michael Gillin, PhD Professor, Chief of Clinical Services, Dept. of Radiation Physics Michael Gillin, PhD Professor, Chief of Clinical Services, Dept. of Radiation Physics Michael Gillin - Grandfather My twin 12 month old grandsons, Arthur and Charles, live in Minneapolis. UT MDACC > 8,000

More information

Melting properties of radiation-induced Na and Cl 2. precipitates in ultra-heavily irradiated NaCl

Melting properties of radiation-induced Na and Cl 2. precipitates in ultra-heavily irradiated NaCl phys. stat. sol. (c), No. 1, 3 3 (5) / DOI 1.1/pssc.15 Melting properties of radiation-induced Na and Cl precipitates in ultra-heavily irradiated NaCl A. V. Sugonyako 1, D. I. Vainshtein 1, H. P. den Hartog

More information

Uncertainties and correction methods when modeling passive scattering proton therapy treatment heads with Monte Carlo

Uncertainties and correction methods when modeling passive scattering proton therapy treatment heads with Monte Carlo Uncertainties and correction methods when modeling passive scattering proton therapy treatment heads with Monte Carlo The Harvard community has made this article openly available. Please share how this

More information

Reactor Core Internals Replacement of Ikata Units 1 and 2

Reactor Core Internals Replacement of Ikata Units 1 and 2 IAEA-CN-194-1P51 Reactor Core Internals Replacement of Ikata Units 1 and 2 K. Ikeda,T. Ishikawa,T. Miyoshi,T. Takagi Shikoku Electric Power Co., Inc., Takamatsu City, Japan Abstract. This paper presents

More information

Aluminium and silicon determination on two Si±Al sputter targets used for magnetron sputtering

Aluminium and silicon determination on two Si±Al sputter targets used for magnetron sputtering Nuclear Instruments and Methods in Physics Research B 158 (1999) 683±688 www.elsevier.nl/locate/nimb Aluminium and silicon determination on two Si±Al sputter targets used for magnetron sputtering G. Terwagne

More information

Clinical implementation and evaluation of the Acuros dose calculation algorithm

Clinical implementation and evaluation of the Acuros dose calculation algorithm Received: 18 November 2016 Revised: 4 May 2017 Accepted: 12 June 2017 DOI: 10.1002/acm2.12149 RADIATION ONCOLOGY PHYSICS Clinical implementation and evaluation of the Acuros dose calculation algorithm

More information

Impact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy

Impact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy Takahashi and Hui Radiation Oncology 2013, 8:123 RESEARCH Open Access Impact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy Yutaka Takahashi 1*

More information

SPECIFICATION DIGITAL MAMMOGRAPHY UNIT AND ASSOCIATED SERVICES

SPECIFICATION DIGITAL MAMMOGRAPHY UNIT AND ASSOCIATED SERVICES 1 Digital Mammography Unit 1. Scope SPECIFICATION DIGITAL MAMMOGRAPHY UNIT AND ASSOCIATED SERVICES This specification describes the requirements for a digital mammography equipment and associated services

More information

Mevex Equipment for Phytosanitary Irradiation. Chapman University 2016 Tony Brown Mevex Corporation

Mevex Equipment for Phytosanitary Irradiation. Chapman University 2016 Tony Brown Mevex Corporation Mevex Equipment for Phytosanitary Irradiation Chapman University 2016 Tony Brown Mevex Corporation Introduction to Mevex Manufacturers of medium sized electron beam accelerators E-beam systems from 2-40

More information

IMRT PHANTOMS CIRS IMRT HOMOGENEOUS PHANTOM CIRS IMRT PHANTOMS C-1

IMRT PHANTOMS CIRS IMRT HOMOGENEOUS PHANTOM CIRS IMRT PHANTOMS C-1 IRS IMRT PHANTOMS IMRT PHANTOMS The IRS IMRT Phantoms for Film and Ion hamber Dosimetry are designed to address the complex issues surrounding commissioning and comparison of treatment planning systems

More information

TG142 Quality Assurance of Medical Accelerators

TG142 Quality Assurance of Medical Accelerators TG142 Quality Assurance of Medical Accelerators Joe Hanley, Ph.D. Hackensack University Medical Center, Hackensack NJ TG142 MEMBERS Chair: Eric E. Klein, Ph.D., Washington University Joe Hanley, Ph.D.,

More information

SPECIFICATION CT SIMULATOR FOR BLACK LION HOSPITAL, ADDIS ABEBA, ETHIOPIA

SPECIFICATION CT SIMULATOR FOR BLACK LION HOSPITAL, ADDIS ABEBA, ETHIOPIA SPECIFICATION CT SIMULATOR FOR BLACK LION HOSPITAL, ADDIS ABEBA, ETHIOPIA 1. Scope This specification describes the requirements for the supply, delivery, installation, and acceptance testing of a CT Simulator

More information

Phys. Low-Dim. Struct. 3/4, 223 (2001). Surface modifications of sapphire single crystals and PET polymer induced by swift heavy ions.

Phys. Low-Dim. Struct. 3/4, 223 (2001). Surface modifications of sapphire single crystals and PET polymer induced by swift heavy ions. 1 Phys. Low-Dim. Struct. 3/4, 223 (2001). Surface modifications of sapphire single crystals and PET polymer induced by swift heavy ions. D.L.Zagorski a, A.E.Efimov b, A.I.Vilensky a. a Shubnikov Institute

More information

Improvement of construction of recombination chambers for mixed radiation dosimetry at work places

Improvement of construction of recombination chambers for mixed radiation dosimetry at work places Improvement of construction of recombination chambers for mixed radiation dosimetry at work places Dosimetry at work places Q n *(10) 20 15 10 5 0 1meV 10meV 100meV 1eV 10eV 100eV 1keV 10keV 100keV E n

More information

Comparison of measured and computed portal dose for IMRT treatment

Comparison of measured and computed portal dose for IMRT treatment JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 3, SUMMER 2006 Comparison of measured and computed portal dose for IMRT treatment Savino Cilla, 1 Pietro Viola, 1 Luigi Azario, 3 Luca Grimaldi,

More information

Development of a Dose Profiler for range monitoring in particle therapy

Development of a Dose Profiler for range monitoring in particle therapy Università degli studi di Roma La Sapienza Dottorato di ricerca in Fisica, Development of a Dose Profiler for range monitoring in particle therapy Ph.D. candidate: Giacomo Traini Supervisor: Prof. Riccardo

More information

Quotation No: 001 REV B Date: 10 Feb., 2011

Quotation No: 001 REV B Date: 10 Feb., 2011 Sameer IIT campus Powai Mumbai 400092 Attn.: Dr. S.T. Chavan, Scientist Quotation : 00 REV B Date: 0 Feb., 20 Sr. Catalog Description Qty Total in Euros Blue Phantom2 Dosimetry System Relative Dosimetry

More information

Good Practice Guide for nanoindentation of nanoparticles embedded in a layer using an SEM in situ technique

Good Practice Guide for nanoindentation of nanoparticles embedded in a layer using an SEM in situ technique Sample preparation The particles to be tested were two sizes of silica spherical particles, nominally sized 300 nm and 100 nm in diameter, gold spherical particles, nominally sized 60 nm in diameter, and

More information