Optimization Approaches in the Routine Analysis of Monoclonal Antibodies by Capillary Electrophoresis

Size: px
Start display at page:

Download "Optimization Approaches in the Routine Analysis of Monoclonal Antibodies by Capillary Electrophoresis"

Transcription

1 Optimization Approaches in the Routine Analysis of Monoclonal Antibodies by Capillary Electrophoresis Oscar Salas-Solano, Lynn Gennaro and Chantal Felten, Protein Analytical Chemistry Department, Process Research and Development Genentech, South San Francisco, California, USA. This month s CE Currents describes three innovative approaches for the analysis of protein-based pharmaceuticals, and their role in the overall QC process at Genentech. Electrophoretic techniques, such as sodium dodecyl sulphate polyacrylamide (SDS-PAGE) and isoelectric focusing (IEF) gels, have traditionally been part of release testing in the biopharmaceutical industry. However, these slab gel methods are inconvenient and irreproducible because of the staining/destaining steps used in analyte detection, the use of toxic reagents and high intra- and inter-gel effective mobility variability. As a result, capillary electrophoresis (CE) methods such as capillary electrophoresis sodium dodecyl sulphate (CE SDS), capillary isoelectric focusing (CIEF) and capillary zone electrophoresis (CZE) have emerged as practical replacements of the slab gel format for many applications in biotechnology In addition, with the advances made in reagents and commercially available instrumentation, CE has expanded its important role in biopharmaceutical analysis by providing complementary information of the physicochemical attributes of therapeutic proteins in terms of charge, hydrophobicity and size. Furthermore, with respect to quality control (QC) laboratories, CE has become an integral part of the specifications of therapeutic proteins. The QC specifications consist of a selected list of methods with relevant acceptance criteria, ensuring the safety, identity, purity and composition of a drug substance and drug product. This CE Currents describes the advances in three CE applications for the analysis of protein-based pharmaceuticals, and their role in the overall QC specifications at Genentech Inc. The first application illustrates the optimization of a generic and quantitative CE SDS assay with LIF detection with improved accuracy and precision to assess the size heterogeneity of recombinant monoclonal antibodies (rmabs). The second section focuses on the use of imaged CIEF to determine charge heterogeneity of rmabs. The third section focuses on the practical applications of interfacing CE to mass spectrometry (MS) for the profiling of rmab glycosylation. Finally, we summarize in the last section additional considerations to be taken with respect to CE training and instrument maintenance to ensure high success rates while applying these practical and complex methods in a routine QC laboratory. Optimization of an rmab Sample Preparation Scheme for Quantitative CE SDS Analysis with LIF Detection CE SDS has been recognized as a robust technique to determine and/or confirm the apparent molecular weight of proteins and to evaluate the size heterogeneity and purity of biologics. 1 The successful performance of a commercial rmab CE SDS based method was recently reported when used by eleven independent biopharmaceutical companies and a regulatory authority. 5 The rmab standard employed contained a mixture of light chain (LC), non-glycosylated heavy chain (NGHC) and heavy chain (HC) components. The high separation power of CE SDS was demonstrated by showing baseline resolution between HC and NGHC species present at a low level (less than 5%). The relative standard deviation (RSD%) for the relative migration time of those components was %. Moreover, the percentage of peak area quantification of the rmab sample components was comparable across all organizations with RSD% values of less than 9%. 5 It is important to indicate that an advantage of CE SDS over other size-based separation methods such as high-performance size-exclusion chromatography (HPSEC) and SDS PAGE is the improved resolution of more closely related size-variants of proteins. Therefore, the CE SDS analysis of reduced rmab samples gives more reliable, accurate and reproducible quantification of the glycolysation site occupancy. Because of these features, CE SDS is also expected to be a routine technique for the quantitative analysis of the size heterogeneity or fragmentation of nonreduced rmab samples. 15

2 However, traditional sample preparation conditions to form SDS protein complexes prior to electrophoretic analysis includes heat treatment at elevated temperatures (e.g., 9 ºC). In the case of non-reduced rmabs, this could lead to sample preparation artifacts in the form of thermally induced fragmentation attributed to disulphide reduction and exchange reactions. 1, It was also reported that high ph conditions during heat treatment also enhanced the fragmentation of SDS-rMAb complexes. 3 These artifacts significantly alter the true representation of the size heterogeneity of a protein and also increase the variability of quantitative CE SDS methodologies of non-reduced samples. As a result, Genentech used non-reduced CE SDS analysis as a qualitative assay to demonstrate that the molecular mass distribution of a bulk manufacture run was comparable to that of a well-characterized reference material, similar to the practice used in SDS PAGE. Moreover, to improve the sensitivity of UV absorbance, a common detector in CE SDS, we demonstrated that precolumn labelling of recombinant protein products with 5-carboxytetramethylrhodamine succinimidyl ester (5-TAMRA.SE) and LIF detection can be used for CE SDS analyses as a replacement for silverstained SDS PAGE to detect low-level impurities and rmab size-variants. 1 The analysis of trace-level rmab variants and process impurities at levels as low as 5 ppm were reported using CE SDS with LIF detection. To greatly decrease any thermally induced fragmentation of non-reduced unlabelled and labelled rmabs, the sample preparation conditions before CE SDS analysis were systematically optimized. The inclusion of an alkylation step with mm iodoacetamide (IAM) during heat treatment of the non-reduced rmab sample and optimization of the incubation temperature and time to 7 ºC for 5 min, in the presence of SDS, significantly suppressed the extent of thermallyinduced fragmentation. The quantitative studies of our work also demonstrated that CE SDS sample buffers at ph levels of.5 further decreased the induced fragmentation of non-reduced rmab samples and gave improved sample stability. As shown in Figure 1 (top trace), a significant decrease of the peak areas corresponding to the rmab fragments was observed for a sample treated with the optimized sample preparation scheme when compared with an rmab control sample. The control sample was prepared using traditional sample preparation conditions that included using 1 mm Tris-HCl buffer ph 9. as CE SDS sample buffer and incubating the sample at 9 ºC for 5 min. The corrected percentage peak area (%CPA) of the intact antibody increased from 9.% in the control Relative Fluorescent Units (RFU) Relative Fluorescent Units (RFU) Intact antibody rmab fragments LC NGHC HC Incomplete reduction sample to 98.5 % in the sample containing mm IAM, reconstituted in 8 mm citrate phophate ph.5 and incubated at 7 ºC for 5 min. The final SDS concentration in all the samples was 1% (w/w). Similar results were obtained when iodoacetic acid (IAA) was used as the alkylating agent. However, IAM was the Figure 1: Expanded view of CE SDS separations of non-reduced (top) and reduced (bottom) labelled-rmab samples in the presence of different alkylating agents. The inset show the fullscale view of the electropherograms. Electrophoretic conditions were as follows: ProteomeLab PA8 instrument with LIF detection, effective length 1. cm, total length 31. cm, 5 µm i.d., 375 µm o.d. uncoated fused-silica capillary; both anode and cathode buffers were Beckman CE SDS polymer solutions. The samples were injected at a constant electric field of 1 V/cm for seconds and electrophoresed at 8 V/cm (3.5 µa) RFU RFU HMW Full scale 8 8 Full scale 8 With IAM With Alkylation LC With IAM Intact antibody rmab fragments NGHC HC No Alkylation HMW With IAM With Alkylation With IAM No Alkylation LC GC Europe December 8

3 reagent of choice because solutions of IAA were less stable as indicated by the appearance of a yellow colouration after a few minutes. The addition of the alkylating agent to reduced labelled-rmab samples was not required. As depicted in Figure 1 (bottom Absorbance at 8 nm (AU) trace), upon the addition of 1M dithiothreitol (DTT) the rmab is reduced to light chain (LC), non-glycosylated heavy chain (NGHC), heavy chain (HC) and other small antibody fragments. The CE SDS profiles of the reduced samples prepared with mm IAM or IAA in the 85 mm Table 1: Precision of CE SDS Method with LIF detection for rmabs. Corrected Percentage Peak Area a Experimental Condition Repeatability b Intermediate Precision c (N = ) (N = 1) Monomer Monomer Non-Reduced Mean RSD (%).1.5 LC HC LC HC Reduced Mean RSD (%) a Corrected percentage peak area (% CPA) = (peak area/migration time) /total corrected peak area x 1. b Six individual preparations of non-reduced and reduced samples were analysed by one analyst on one day. c Two analysts in two different laboratories analysed samples on three separate days using different capillaries, reagents, and instruments. Figure : Imaged cief profile of a rmab. Experimental conditions: imaged cief using an ice8 analyser with a fluorocarbon-coated capillary cartridge (1 µm 5 cm). The ampholyte solution consisted of a mixture of Pharmalytes ph 3 1 and ph 8 1.5, with.%(v/v) of each of the pi markers 7. and The ampholyte solution also contained.35% (v/v) methylcellulose (MC). Samples were diluted to 1.5 mg/ml with purified water and then mixed with the ampholye solution at a 1: (v/v) ratio. The analyte was 8 mm phosphoric acid, and the catholyte was 1 mm sodium hydroxide, both in.1% methyl cellulose. The samples were loaded into the cartridge and focused by introducing a potential of 15 volts for one min, followed by a potential of 3 volts for 1 min pl Acidic Variants -Lys Basis Variants 1-Lys -Lys citrate-phosphate ph.5 were comparable with the control sample prepared as indicated above. The results of the precision studies of the optimized CE SDS method are shown in Table 1. For repeatability, six individual sample preparations were analysed under both non-reducing and reducing conditions on one day with one injection per sample. As shown in Table 1, the RSD value of the corrected percentage peak area for the intact antibody of non-reduced samples was only.1%. The RSD value was.85% for the corrected percent peak area of the LC and HC in the reduced samples. The intermediate precision study was performed by two analysts in different laboratories. They assayed the samples under both non-reducing and reducing conditions on three separate days using separate instruments, reagents and capillaries. The RSD values were.5% for the corrected percent peak area of the intact antibody,.39% and 1.% for the LC and HC, respectively. Taken together, the data demonstrate that CE SDS with LIF detection is a reproducible and robust technique to assess the size heterogeneity of rmabs. This methodology is routinely used at GNE for the characterization and lot release of therapeutic rmabs. Imaged CIEF Regulatory agencies require the assessment of the charge heterogeneity of biotechnology products. Charged variants of recombinant protein products often include post-translational modifications such as sialylation, phosphorylation, deamidation and C-terminal lysine residues. Such variants are often monitored by the widely used ion-exchange chromatography (IEC) techniques. However, the development of an IEC method is typically extensive and product-specific. IEC assays require the optimization of complex separation parameters including column, mobile phase composition, ph, salt, temperature and gradient. CE techniques, such as CZE and CIEF, offer the advantages of developing generic methods for multiple products and faster analysis time, which is desirable in a high-throughput process development laboratory. In CZE, analytes are separated from each other based on the differences in their electrophoretic mobilities; therefore, this technique also introduces a size-based element to the separation. Whereas, the ability of CIEF to separate solely based on charge may have an added advantage in isolating charge variants of rmabs. 18 LC GC Europe December 8

4 CIEF is usually performed on commercial CE instruments, which have a cm long capillary and on-column ultraviolet UV absorbance detector. After focusing, the focused protein zones are moved through the detector point by chemical, electrophoretic or hydrodynamic mobilization. In this two-step CIEF mode, problems associated with the mobilization process may be encountered, including electropherogram shows the various isoforms of the antibody between the isoelectric points (pi) markers 7. and 9.7. The pi values of the rmab peaks, shown in the x-axis, were calculated using two-point calibration with the pl markers 7. and 9.7. The main peak or -Lys form of the rmab appears at pi 9.. The more basic species with one and two C-terminal Lys residues at the heavy CE is commonly avoided in routine analysis because it is reputed to be a troublesome technique with high failure rates. However, this is no longer true because instrument manufacturers have drastically improved instrument design and overall CE knowledge has increased. long analysis time, distortion of ph gradient and uneven resolution because of a nonuniform mobilization speed. The requirement of the mobilization step can be eliminated using whole-column imaging detection via charged coupled device (CCD) camera. This technique is called imaged CIEF, or icief, and has been recently commercialized by Convergent Biosciences. This technique could have the added advantages of robustness because of absence of moving parts typical of conventional CIEF instrumentation, reduced analysis time (less than 15 min per sample), and allowing the user to monitor the focusing process in real-time leading to faster method development. Although icief is still a relatively new technique, the advantages of simplicity and fast analysis time make it more suitable for QC environments and the high-throughput needs in process development. The potential of icief to assess the overall rmab charge distribution is demonstrated in Figure. The chain are baseline-resolved from each other and the main peak. In addition, the acidic variants were also observed before the -Lys peak. Table presents a summary of the reproducibility studies of the icief method to assess the charge variants of the rmab in Figure. The low %RSD values for all the peaks demonstrate that icief can be used in QC as a technique to monitor the charge heterogeneity of protein therapeutics. On-Line CE MS for Direct Characterization of N-Linked Glycans Of the many post-translational modifications present on therapeutic proteins, glycosylation has received considerable attention in the field of proteomics. In the biotechnology industry, it is well known that carbohydrates can play a significant role in the activity and efficacy of a glycoprotein and must be carefully monitored. 7 9 Table : Precision of icief Method for rmabs. Percentage Peak Area a Repeatability b Intermediate Precision c (N = ) (N = 1) Acidic Main Basic Acidic Main Basic Peaks Peak Peaks Peaks Peak Peaks Mean RSD (%) a Percentage peak area (% CPA) = peak area /total peak area x 1. b Six individual preparations were analysed by one analyst on one day. c Two analysts in two different laboratories analysed samples on three separate days using different capillaries, reagents, and instruments. 19

5 Figure 3: Full-scale electropherograms of APTS-labelled rmab glycans using (top) conventional CE LIF in a PVA-coated capillary (5 µm i.d. cm total capillary length) and (bottom) CE MS base peak electropherogram using negative ion MS detection (5 µm i.d. 1 cm total capillary length). The CE LIF experiments were performed using a CE system with LIF excitation at 88 nm and 5 nm emission filter. Capillaries were kept at C. Injection was performed hydrodynamically at.5 psi for 1 s. For MS interfacing, the end of the CE capillary was installed into a CE MS sprayer, allowing for a sheath liquid to be added coaxially. The tip of the CE capillary was aligned flush to the stainless steel capillary tip under magnification. A sheath liquid of 5:5 isopropanol:water.% ammonia was then delivered at µl/min. Mass spectra were acquired from m/z in the NI mode. A running buffer of mm -aminocaproic acid, ph.5 (adjusted with glacial acetic acid with an applied voltage of 3 kv was used for both the CE LIF and CE MS separations. Relative Fluorescent Units (RFU) Electrospray Units (Intens. 1 ) GO F GO 1 Even minor changes in glycan distribution (such as increased afucosylation) can have effects on the activity of a biopharmaceutical product. 1 1 Therefore, advances are continually being made in the biopharmaceutical industry to reduce glycan heterogeneity and improve analytical assays used to identify and quantify minor changes across products and/or batches G Time (min) G1 G1 Time (min) G CE LIF CE MS However, the task of glycoprotein characterization remains an analytical challenge because of the immense heterogeneity of these species. Genentech has been using CE-based methods for profiling glycosylation for several years because of the advantages of automation, short analysis times and the ability to separate and quantify isomeric species. The coupling of CE with on-column laser-induced fluorescence (LIF) detection also has the advantage of highly sensitive detection. The primary assay used for profiling rmab glycosylation involves releasing N-linked carbohydrates with PNGase-F, labelling with the fluorophore 8-aminopyrene-1,3,-trisulphonate (APTS) and analysis by CE-LIF. APTS was chosen because of the low-picomole sensitivity attainable for APTS- derivatized carbohydrates. 1 An earlier version of the assay described by Ma and Nashabeh, 15 in which CE-LIF was used for profiling and quantification of rituximab glycans including major species G, G1, G1 and G. Currently, a new version of this assay has been validated and is used routinely in a high-throughput format for product characterization, clone selection and lot release of recombinant monoclonal antibody (rmab) therapeutics. However, one disadvantage of CEbased methods is the lack of routine detection methods that can provide direct structural information on migrating species. For our glycan assay, characterization was primarily performed using one of two methods: 1. Co-migration with commercially available standards.. Exoglycosidase treatment, followed by co-elution with standards or examination of migration shift. These methods can be particularly labour-intensive and do not provide any direct information on glycan structure. As a result of the more routine use of this assay in all stages of product development, we have been exploring new methods that can be used to characterize new or existing peaks that may be present. Coupling CE to on-line mass spectrometry (MS) presents an attractive option, allowing direct mass information to be obtained on migrating components. 1 Direct CE MS coupling is usually performed via elecrospray ionization (ESI) and requires an interface that can provide high sensitivity without degrading the separation quality. Peak broadening in CE MS have been reported as a result of parabolic flow from pressure differences between the capillary ends, dead volume associated with the CE MS interface and the lack of temperature control at the detector end. 17,18 As a result, routine applications of CE MS are relatively few in the biopharmaceutical industry and most reports have shown little emphasis in evaluating this technique as a robust and reproducible analytical technique to establish the identity of the peaks in the LC GC Europe December 8

6 CE UV or CE LIF electropherograms. Our group performed numerous studies aimed at optimizing this technology towards routine use in process development and presented several relevant applications in the analysis of peptides, glycopeptides and carbohydrates. 19 The current research in our group expands the development of a robust CE MS methodology for the characterization of the glycan peaks observed in the CE LIF assay. For MS interfacing, the end of the CE capillary is installed into a CE MS sprayer (Agilent Technologies, Wilmington, Delaware, USA), allowing for a sheath liquid to be added coaxially. Mass spectrometry detection is achieved using a Bruker MicrOTOF accurate mass MS instrument (Bruker Daltonics, Germany), equipped with an orthogonal, at-ground electrospray ionization (ESI) source. It is important to mention that using an MS instrument with the spray needle grounded allows the use of high CE currents without the risk of instrument damage. As a result, concentrated buffer systems can be used leading to high-efficiency separations. In addition, the reproducibility and robustness of the CE MS system is significantly improved. The ultimate goal is to achieve a similar CE separation profile with mass spectrometric detection as that obtained by the original LIF method to facilitate direct peak identification. To accomplish this, the CE LIF method was modified to allow MS coupling while ensuring that any differences between the CE LIF and CE MS profiles could be attributed only to minor changes in resolution. A typical CE LIF glycan profile for an rmab is shown in Figure 3 (top trace). In this profile, major glycan species G, G1, G1 and G can be easily resolved, as well as several unidentified minor peaks at 9.7 min. Because of the nature of MS detection, there are several assay parameters that must be modified to use this technique. These modifications include (1) removing the excess salts/label from APTS-labelled glycan samples, which interfere with MS detection and () increasing the length of the CE capillary from cm total length (5 cm effective) to 1 cm total length (9 cm effective) so that it can exit the CE instrument and extend to the MS source. Figure 3 (bottom trace) illustrates the CE MS electropherograms of desalted, APTS-labelled rmab glycans using the standard CE LIF buffer and 1 cm capillary (modified conditions). As expected, an increase in migration time was observed using the longer 1 cm capillary. Most importantly, the overall peak profile remained consistent so identification of the peaks in the CE LIF profile can be performed by direct comparison with the electropherogram obtained by CE MS. Mass spectral data obtained for the major peaks confirmed the identity of oligosaccharides with zero, one and two terminal galactose residues (G, G1 positional isomers and G, respectively). In addition, the minor peaks were identified as the afucosylated G (G-F) and G-GlcNAc (G-1) species. CE in Routine Analysis CE is commonly avoided in routine analysis because it is reputed to be a troublesome technique with high failure rates. However, this is no longer true because instrument manufacturers have drastically improved instrument design and overall CE knowledge has increased. There are three key contributors to reducing failure rate and producing accurate, precise and robust CE data: Training the operator, system suitablity and instrument maintenance. Training the operator: General knowledge of the theory of CE, as well as instrument function, is not common amongst analysts. Capillary electrophoresis systems differ significantly in design and operation from HPLC systems and require extensive initial indepth training. As well as explaining the overall sample preparation steps, such as labelling (CE SDS LIF) or sample composition (CIEF), special emphasis should be put into building a cartridge and overall knowledge of capillary coating. HPLC columns tend to be robust and almost indestructible, whereas coated capillaries can be easily damaged through misuse. Simple errors include the burning of a capillary window by heating wire or applying voltage to a dry capillary. Both of these actions tend to damage the coating, which will result in loss of resolution for protein separation. Uncoated capillaries, such as those used in CE SDS, tend to be more robust, but special care should be taken with respect to keeping the ends of the capillary clean cut and free of coating because bits of the polyimide coating may clog the capillary during the pressurized gel filling process. Meaningful system suitability: As well as providing the analyst with the knowledge of science and technique it is also critical to develop a meaningful 1

7 system suitability test. A system suitability test should not only ensure accurate and reproducible results but should also help the analyst discover underlying issues and help when troubleshooting. For example, besides general ranges on %CPA for the main peak in reduced CE SDS LIF, a second criteria can be set with respect to %CPA for incomplete reduced area, which will indicate a poorly performed reduction. To monitor overall separation performance a criteria can be set around migration time precision because migration time is directly related to the reproducibility of the gel buffer/capillary length/applied voltage. Instrument maintenance: Most CE instruments are complex and require routine maintenance to reduce failure rates. During the course of a CE separation both the capillary end and the electrode will push through a septum into a vial (sample, separation, wash solution, etc.) multiple times during operation. This mechanical insertion can often lead to bending and breaking of the capillary and/or electrode if the system is misaligned. As the electrode is generally more sturdy than the capillary it is recommended to cut the capillary about 3 mm shorter than the electrode and to keep the capillary and the electrode tightly together so that the electrode will guide the capillary through the septa (Note: do keep in mind that both ends of the capillary must be at equal height to reduce gravity-induced flow). The need for operational care is magnifiied when viscous gels are involved. The vials containing gels are pressurized during the filling process, which often leads to splashing of the gel against vial septa and instrument interface. If left unattended, the gel will dry out and form a thin, conducting film. The presence of this film can result in current spikes, which will affect resolution and create typically non- Gaussian artifact peaks. In severe cases the gel may close off the pressure connection, or create an uneven surface for septa and instrument interface, which will lead to increased pressure failures. It is, therefore, recommended to load the gel with low pressure and to clean the instrument interface block after each session with a wet cotton tip. Conclusions Capillary electrophoresis (CE) methods for the analysis of therapeutic proteins are routinely used at Genentech in all stages of product development. These include process and formulation development, physicochemical characterization, comparability and QC lot-release testing. This article highlights the work conducted at Genentech to advance CE applications in the biopharmaceutical industry. First, we optimized the sample preparation scheme for quantitative CE SDS analysis of non-reduced rmabs. The inclusion of an alkylation step with mm IAM during heat treatment of non-reduced rmab samples at 7 C for 5 min in the presence of CE SDS sample buffers at ph levels of.5 significantly suppressed the extent of thermally induced fragmentation and led to increased sample stability. Second, we proved that imaged CIEF is a reproducible technique to assess the charge heterogeneity of therapeutic proteins with the advantages of (1) faster method development, () rapid analysis time and (3) the potential of using the same method for a wide variety of rmabs. We then demonstrated the utility of CE MS technology to provide accurate mass information on major and minor N-linked glycans of rmabs. The CE MS method has been developed with an emphasis on maintaining the same peak profile to that obtained in the CE LIF assay for more accurate peak identification. Finally, we provided advice and suggestions with respect to CE training and instrument maintenance to ensure high success rates of CE methods in a routine QC laboratory for therapeutic proteins. Acknowledgements The authors gratefully acknowledge James McElroy, Yong Liu and Monica Parker respectively for their contributions to the icief, CE MS and CE SDS work. In addition, we thank Stacey Ma, Wassim Nashabeh, Dieter Schmalzing and Reed Harris for their support to expand CE technology at Genentech. References 1. G. Hunt and W. Nashabeh, Anal. Chem., 71, (1999).. M. Schenerman and S. Bowen, Chromatographia, 53, S-S7 (1) 3. H. Lee, S. Chang and E.J. Fritsche, J. Chromatogr. A, 97, 13 9 ().. O. Salas-Solano et al., Anal. Chem. 78, (). 5. B. Nunnally, et al., Chromatographia,, (7). J. Wu et al., Biotechnol Lab, 17, (1999). 7. W.S. Hancock, J. Proteome Res., 1, (). 8. R. Jefferis, Biotechnol. Prog., 1, 11-1 (5). (9) G. Walsh and R. Jefferis, Nat. Biotechnol.,, (). 1 R. Shields et al., J. Biol. Chem., 77, () 11. R. Niwa et al., Clin. Cancer Res., 1, 8 55 () 1. C. Ferrara et al., J. Biol. Chem., 81, (). 13. G.Walsh, Nat. Biotechnol.,, (). 1. R.A.Evangelista, M.S. Liu and F.T Chen, Anal. Chem., 7, 39 5 (1995). 15. S. Ma and W. Nashabeh, Anal. Chem., 71, (1999). 1. J. A. Olivares et al., Anal. Chem., 59, (1987). 17. J. Henion, A. Mordehai and J. Cai, Anal. Chem.,, (199). 18. J. Axen et al., Electrophoresis, 8, (7). 19. L.A. Gennaro, O. Salas-Solano and S. Ma, Anal Biochem, 355, 9 58 (). Oscar Salas-Solano is a senior scientist in the Protein Analytical Chemistry at Genentech Inc. His work includes physicochemical characterization of proteins, CE technology development, and development and validation of analytical assays for in-process, lot-release and stability testing of recombinant therapeutic proteins. Lynn Gennaro is a scientist in the Protein Analytical Chemistry at Genentech Inc. Her work includes CE MS technology development and analytical assays development and validation for the control systems of therapeutic proteins. Chantal Felten is a consultant at Cardiome Pharma Corp., Canada. Her work focuses on the implementation of CE and HPLC methods for the analysis therapeutic proteins. LC GC Europe December 8

Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System

Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System Application Note Biologics & Biosimilars, Proteomics & Protein

More information

Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis

Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis Procedures Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis Separation and Online Detection of Intact mab Variants and Impurities using CESI-MS Bryan

More information

Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing

Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing Application Note Biotherapeutics & Biosimilars Author Christian

More information

Capillary Electrophoresis in Quality Control PART II: CE-SDS: Method Development and Robustness

Capillary Electrophoresis in Quality Control PART II: CE-SDS: Method Development and Robustness Capillary Electrophoresis in Quality Control PART II: CE-SDS: Method Development and Robustness Chantal Felten *1 and Oscar Salas Solano 2 1 Alpine Analytical Acadamy, Whistler, British Columbia 2 Seattle

More information

Automated and quantitative analysis of biologics

Automated and quantitative analysis of biologics PA 800 PLUS PHARMACEUTICAL ANALYSIS SYSTEM Automated and quantitative analysis of biologics PA 800 PLUS PHARMACEUTIC AL ANALYSIS SYSTEM Designed for the needs of the biopharmaceutical industry Therapeutic

More information

Application Note. Biopharm. Author. Abstract. Christian Wenz Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Biopharm. Author. Abstract. Christian Wenz Agilent Technologies, Inc. Waldbronn, Germany Performance of commercially available gels for protein characterization by capillary gel electrophoresis with UV detection on the Agilent 7100 CE System Application Note Biopharm Author Christian Wenz

More information

Monoclonal Antibody Charge Variants Identification by Fully Automated Capillary Isoelectric Focusing Mass Spectrometry

Monoclonal Antibody Charge Variants Identification by Fully Automated Capillary Isoelectric Focusing Mass Spectrometry Application Note Pharma & Biopharma Monoclonal Antibody Charge Variants Identification by Fully Automated Capillary Isoelectric Focusing Mass Spectrometry Authors Tony Wang, James Xia, and Rena Wang CMP

More information

Analysis of Monoclonal Antibody Charge Variants by Capillary Zone Electrophoresis

Analysis of Monoclonal Antibody Charge Variants by Capillary Zone Electrophoresis Analysis of Monoclonal Antibody Charge Variants by Capillary Zone Electrophoresis Marcia R Santos Global Tactical Marketing, Beckman Coulter Life Sciences, Brea, CA USA INTRODUCTION Charge heterogeneity

More information

Automated and quantitative

Automated and quantitative Automated and quantitative analysis of biologics. PA 800 plus Pharmaceutical Analysis System Genomics Cell Analysis Particle Characterization Capillary Electrophoresis Lab Automation Centrifugation Bioseparation

More information

Rediscover the Strengths of Your PA 800 Plus - A Multi-Attribute Protein Analyzer

Rediscover the Strengths of Your PA 800 Plus - A Multi-Attribute Protein Analyzer Rediscover the Strengths of Your PA 800 Plus - A Multi-Attribute Protein Analyzer Marcia R Santos SCIEX Separations, Brea, CA Introduction Capillary Electrophoresis is a well-accepted analytical separation

More information

Capillary Electrophoresis of Proteins

Capillary Electrophoresis of Proteins Capillary Electrophoresis of Proteins SDS Capillary Gel Electrophoresis SDS-CGE Outline CE-SDS Gel Analysis Description of Technique Method Development Tips PA800 plus kits SDS-MW IgG Purity & Heterogeneity

More information

LabChip GXII: Antibody Analysis

LabChip GXII: Antibody Analysis WHITE PAPER LabChip GXII: Antibody Analysis Antibody Analysis using microfluidic technology in high throughput Quality by Design Experiments Abstract Current initiatives in Process Analytical Technology

More information

Automated and Quantitative Analysis of Biologics

Automated and Quantitative Analysis of Biologics PA 800 PLUS PHARMACEUTICAL ANALYSIS SYSTEM Automated and Quantitative Analysis of Biologics PA 800 PLUS PHARMACEUTICAL ANALYSIS SYSTEM For Research Use Only. Not for use in diagnostic procedures. Flexibility

More information

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Separation of Native Monoclonal Antibodies and Identification of Charge Variants: Separation of Native Monoclonal Antibodies and Identification of Charge Variants: Teamwork of the Agilent 31 OFFGEL Fractionator, Agilent 21 Bioanalyzer and Agilent LC/MS Systems Application Note Biosimilar

More information

Application Note. Abstract. Author. Biopharmaceutical

Application Note. Abstract. Author. Biopharmaceutical Innovator and Biosimilar Monoclonal Antibody-Peptide Map Comparison Using the Agilent 70 Capillary Electrophoresis System and Agilent MatchCompare Software Application Note Biopharmaceutical Author Arunkumar

More information

Assay of IgG Purity and Heterogeneity using High-Resolution Sodium Dodecyl Sulfate Capillary Gel Electrophoresis

Assay of IgG Purity and Heterogeneity using High-Resolution Sodium Dodecyl Sulfate Capillary Gel Electrophoresis Assay of IgG Purity and Heterogeneity using High-Resolution Sodium Dodecyl Sulfate Capillary Gel Electrophoresis Lucy Y. Liu, Chitra Ratnayake, Jeff Chapman, Narasaiah Dontha, Sae Choo, and M. P. Reddy

More information

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup High Throughput Methods to Maximize the Use of the PA 800 Plus system Jose-Luis Gallegos-Perez

More information

Table 1: CE-MS: Unexploited Opportunities in Process Development?

Table 1: CE-MS: Unexploited Opportunities in Process Development? Table 1: CE-MS: Unexploited Opportunities in Process Development? Session 1 Facilitator: Mei Han, Amgen Inc. Scribe: Yunan Wang, Amgen Inc. Session 2 Facilitator: Göran Hübner, Boehringer Ingelheim Pharma

More information

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography Agilent 1260 Infinity Bio-inert Quaternary LC System with Agilent Bio Columns

More information

This is not unique to CE. It took for instance about 25 years for HPLC to establish itself properly in the pharmaceutical industry.

This is not unique to CE. It took for instance about 25 years for HPLC to establish itself properly in the pharmaceutical industry. CE Pharm 2017 Troubleshooting Report: CZE and CE-SDS the Discussion Continues Workshop facilitators: Tim Blanc 1, Cari Sänger-van de Griend 2 Scribes: Nathan Lacher 3, David Michels 4, Zoran Sosic 5 1

More information

APPLICATION INFORMATION

APPLICATION INFORMATION A-1861A APPLICATION INFORMATION Capillary Electrophoresis USING CIEF TO CHARACTERIZE RECOMBINANT HUMAN MONOCLONAL ANTIBODIES L. C. Santosa, 1 I. S. Krull, 1 and K. L. Grant 2 1 Department of Chemistry,

More information

Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance

Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance Sushma Rampal, Sr. Application Scientist, Beckman Coulter Life Sciences, Brea, CA, USA Marcia Santos, Staff

More information

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram

More information

Analysis of Protein Biopharmaceuticals

Analysis of Protein Biopharmaceuticals Analysis of Protein Biopharmaceuticals Comprehensive cgmp Services at Every Stage of Drug Development Amazing where you can go At Solvias, we work closely with you Solvias provides comprehensive services

More information

Assessment of Active Biopharmaceutical Ingredients Prior To and Following Removal of Interfering Excipients

Assessment of Active Biopharmaceutical Ingredients Prior To and Following Removal of Interfering Excipients Assessment of Active Biopharmaceutical Ingredients Prior To and Following Removal of Interfering Excipients Andrew J Reason and Howard R Morris BioPharmaSpec Ltd, and BioPharmaSpec Inc, The EMA guideline

More information

Repeatability of C100ht Biologics Analyzer, for High Throughput Glycan Screening

Repeatability of C100ht Biologics Analyzer, for High Throughput Glycan Screening Repeatability of C100ht Biologics Analyzer, for High Throughput Glycan Screening Marcia Santos, Tingting Li, Mervin Gutierrez, Anna Lou, and Clarence Lew SCIEX Separations, Brea, CA Introduction The ability

More information

"VUPNBUFE BOE RVBOUJUBUJWF

VUPNBUFE BOE RVBOUJUBUJWF "VUPNBUFEBOERVBOUJUBUJWF BOBMZTJTPGCJPMPHJDT 1" plus1ibsnbdfvujdbm"obmztjt4ztufn Genomics Cell Analysis Particle Characterization Capillary Electrophoresis Lab Automation Centrifugation Bioseparation Lab

More information

PA 800 plus Catalog. Systems/Software

PA 800 plus Catalog. Systems/Software PA 800 plus Catalog Built on 20 years of innovation and leadership in the field of capillary electrophoresis, Beckman Coulter s PA 800 plus provides solutions for protein purity analysis, charge heterogeneity

More information

Knock! Knock! Who s there? Maurice Who? SungAe Suhr Park, Mee Ko, Eleanor Le, Janice Chen Amgen Inc.

Knock! Knock! Who s there? Maurice Who? SungAe Suhr Park, Mee Ko, Eleanor Le, Janice Chen Amgen Inc. Knock! Knock! Who s there? Maurice Who? SungAe Suhr Park, Mee Ko, Eleanor Le, Janice Chen Amgen Inc. Abstract SungAe Suhr Park, Mee Ko, Eleanor Le and Janice Chen Amgen Inc. Applications of Capillary Electrophoresis

More information

Monitoring Charge Heterogeneity of Antibody-Maytansinoid Conjugates (AMC) with icief

Monitoring Charge Heterogeneity of Antibody-Maytansinoid Conjugates (AMC) with icief Monitoring Charge Heterogeneity of Antibody-Maytansinoid Conjugates (AMC) with icief Joyce Lin, Rajesh Krishnamurthy, Alexandru Lazar ImmunoGen, Inc., Waltham, MA CE Pharm 2008 utline I. Introduction of

More information

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Application Note Author David L. Wong Agilent Technologies, Inc. Santa Clara, CA, USA Introduction Monoclonal

More information

CE as an analytical tool in industrial biotechnology. David Goodall, CSO, Paraytec Ltd. 14 November Act P x

CE as an analytical tool in industrial biotechnology. David Goodall, CSO, Paraytec Ltd. 14 November Act P x 1 CE as an analytical tool in industrial biotechnology David Goodall, CSO, Paraytec Ltd 14 November 2013 Coverage 2 Modes of CE Industrial sectors for applications Companies and key markets Selected applications

More information

TECHNICAL INFORMATION

TECHNICAL INFORMATION T-1860A TECHNICAL INFORMATION Capillary Electrophoresis PRECISION IN CAPILLARY ELECTROPHORESIS Harry Whatley and Jeff Chapman Beckman Coulter, Inc. Introduction Instrumentation for routine chemical analysis

More information

Get there, faster with ice3

Get there, faster with ice3 Get there, faster with ice3 Method development in hours, not weeks. The ice3 Tomorrow s challenges are always looming, so why not catapult ahead? ice3 lets you move beyond the limits of traditional protein

More information

Fast IEF Method Development Using Whole-Column Detection cief

Fast IEF Method Development Using Whole-Column Detection cief Fast IEF Method Development Using Whole-Column Detection cief Jiaqi Wu 27 Coronet Road, Toronto, Ontario, CANADA M8Z 2L8 Conventional cief (cief Performed on Commercial CE Instruments) - + ph gradient

More information

SDS-CGE Method Development for Purified Protein Vaccine Antigens

SDS-CGE Method Development for Purified Protein Vaccine Antigens SDS-CGE Method Development for Purified Protein Vaccine Antigens Michael Leach, Elena Newman and Bruce Carpick Sanofi Pasteur Analytical R&D, Biochemistry Overview SDS-CGE method development for purified

More information

Comprehensive characterization of three IgG forms using CESI-MS

Comprehensive characterization of three IgG forms using CESI-MS Comprehensive characterization of three IgG forms using CESI-MS Bryan Fonslow 1, Olga V. Friese 2, and K. Steven Cook 2 1 SCIEX, Brea, CA and 2 Pfizer, Chesterfield, MO mabs, ADCs, biosimilars, and biobetters

More information

The Importance of Temperature Control in Capillary Electrophoresis of Carbohydrates

The Importance of Temperature Control in Capillary Electrophoresis of Carbohydrates The Importance of Temperature Control in Capillary Electrophoresis of Carbohydrates Introducing Temperature Gradient Capillary Electrophoresis Andras Guttman, 1 Marton Szigeti 2 and Jeff Chapman 1 1 SCIEX

More information

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column Shane Bechler 1, Ken Cook 2, and Kelly Flook 1 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, Runcorn,

More information

Characterization of intact monoclonal antibody with microfluidic chip electrophoresis mass spectrometry

Characterization of intact monoclonal antibody with microfluidic chip electrophoresis mass spectrometry POSTER NOTE 65 Characterization of intact monoclonal antibody with microfluidic chip electrophoresis mass spectrometry Authors Chien-Hsun Chen, 1 Andreas Krupke, 2 Stephane Houel, 1 Aaron Bailey, 1 Aran

More information

Simplifying CE-SDS Data Processing

Simplifying CE-SDS Data Processing Simplifying CE-SDS Data Processing Approach for Mitigating Product Peak Migration Time Drift Samuel Shepherd 1, KieranKumar Mistry 1, Whitney Lane Smith 2, Nicholas Bond 1, Vivian Lindo 1, Sam Fox 3, Jim

More information

An Industry Perspective on Established Conditions in the Analytical Control System. Christof Finkler, F. Hoffmann-La Roche

An Industry Perspective on Established Conditions in the Analytical Control System. Christof Finkler, F. Hoffmann-La Roche An Industry Perspective on Established Conditions in the Analytical Control System Christof Finkler, F. Hoffmann-La Roche Control System Development INPUT: Process Parameter OUTPUT: Quality Attribute CPPs

More information

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs Application Note Author Suresh Babu C.V. Agilent Technologies India Pvt. Ltd, Bangalore, India Abstract This Application Note describes

More information

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System Marcia R Santos Global Tactical Marketing, Beckman Coulter

More information

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus Robert Birdsall, Thomas Wheat, and Weibin Chen Waters Corporation,

More information

Capillary Electrophoresis in Biotechnology Industry: a Look at the Last Decade Stacey Ma, Ph. D Director ES Pharmaceutical Development

Capillary Electrophoresis in Biotechnology Industry: a Look at the Last Decade Stacey Ma, Ph. D Director ES Pharmaceutical Development Feb. 5, 2009, Boston, MA Capillary Electrophoresis in Biotechnology Industry: a Look at the Last Decade Stacey Ma, Ph. D Director ES Pharmaceutical Development CE Pharm 1999-2008 1999, 1st Symposium CE

More information

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies Combining Size Exclusion Chromatography with Method Development and Light Scattering Application Note Biotherapeutics and Biosimilars

More information

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System GUIDE TO LC MS - December 21 1 Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System Henrik Wadensten, Inger Salomonsson, Staffan Lindqvist, Staffan Renlund, Amersham Biosciences,

More information

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC The Agilent 126 Infinity Bio-Inert Quaternary LC System and the AdvanceBio SEC 3Å, 2.7 µm Column Application Note Biologics

More information

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column Rapid UHP Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 3SB-C8 Column Application Note BioPharma Authors James Martosella, Phu Duong, Susanne

More information

Tony Mire-Sluis Vice President, Corporate, Product and Device Quality Amgen Inc

Tony Mire-Sluis Vice President, Corporate, Product and Device Quality Amgen Inc The Regulatory Implications of the ever increasing power of Mass Spectrometry and its role in the Analysis of Biotechnology Products Where do we draw the line? Tony Mire-Sluis Vice President, Corporate,

More information

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins BioPharma Compass TM is a fully automated solution for the rapid characterization

More information

High-Throughput Glycan Screening to Remove Biopharmaceutical Development Bottlenecks

High-Throughput Glycan Screening to Remove Biopharmaceutical Development Bottlenecks High-Throughput Glycan Screening to Remove Biopharmaceutical Development Bottlenecks Mark Lies, Marcia Santos, and Tingting Li SCIEX Separations, Brea, CA Introduction Development of biopharmaceuticals

More information

HPCE-5 2. A Revolution in Capillary Electrophoresis. High Performance Capillary Electrophoresis

HPCE-5 2. A Revolution in Capillary Electrophoresis. High Performance Capillary Electrophoresis A Revolution in Capillary Electrophoresis High Performance Capillary Electrophoresis Revolutionary Performance Continuous Monitoring of Capillary and Buffer Conditions Sensitive Label-Free 512 Times Better

More information

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Hongwei Xie, Martin Gilar, and Jeff Mazzeo Waters Corporation, Milford, MA U.S. APPLICATION BENEFITS The ACQUITY UPLC

More information

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns Confidently separate and characterize bio-molecules with Agilent BioHPLC columns Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns "It's a struggle to isolate and identify charge variants of

More information

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF Application Note Authors Xianming Liu, Wei Zhang, Yi Du, Sheng Yin, Hong Que, and Weichang Zhou WuXi AppTec iopharmaceuticals

More information

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3 Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3 Application Note BioPharma Authors Phu T Duong and James Martosella Agilent Technologies, Inc. 285 Centerville Rd, Wilmington,

More information

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column APPLICATION NOTE 21845 ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column Authors Julia Baek, Shane Bechler, Shanhua Lin, Stacy Tremintin Thermo Fisher Scientific, Sunnyvale,

More information

Automating the C100ht Biologics Analyzer Sample Preparation on a Biomek i5 Liquid Handler Platform

Automating the C100ht Biologics Analyzer Sample Preparation on a Biomek i5 Liquid Handler Platform Automating the C100ht Biologics Analyzer Sample Preparation on a Biomek i5 Liquid Handler Platform Mervin Gutierrez, Tingting Li, and Marcia Santos SCIEX Separations, Brea, CA Introduction There is an

More information

deltadot ANTIBODY ANALYSIS Application Note: Antibodies using LABEL FREE INTRINSIC IMAGING LONDON United Kingdom

deltadot ANTIBODY ANALYSIS Application Note: Antibodies using LABEL FREE INTRINSIC IMAGING LONDON United Kingdom ANTIBODY ANALYSIS using LABEL FREE INTRINSIC IMAGING Application Note: Antibodies Ana l y s i s o f Im m u n o g l o b u l i n G u s i n g deltadot s PEREGRINE I 512-p i x e l La b e l Fr e e Ca p i l

More information

Determination of Isoelectric Point (pi) By Whole-Column Detection cief

Determination of Isoelectric Point (pi) By Whole-Column Detection cief Determination of Isoelectric Point (pi) By Whole-Column Detection cief Tiemin Huang and Jiaqi Wu CONVERGENT BIOSCIENCE Determination of Isoelectric Point (pi) by Whole Column Detection cief Definition

More information

Instructions for Capillary Electrophoresis Peptide Analysis Kit

Instructions for Capillary Electrophoresis Peptide Analysis Kit Instructions for Capillary Electrophoresis Peptide Analysis Kit Catalog Number 148-4110 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Table of

More information

Chromatography column for therapeutic protein analysis

Chromatography column for therapeutic protein analysis PRODUCT SPECIFICATIONS ProPac Elite WCX Column Chromatography column for therapeutic protein analysis Benefits Superior resolution power for proteins, monoclonal antibodies, and associated charge variants

More information

HT Pico Protein Express LabChip Kit, Frequently Asked Questions

HT Pico Protein Express LabChip Kit, Frequently Asked Questions Protein Detection 1. How does the GXII determine protein concentration? Protein samples contain a single lower marker. Alignment to a single marker does not provide enough constraints to align large proteins,

More information

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles Application Note Biotherapeutics and Biosimilars Authors James Martosella, Alex

More information

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution Clone Selection Using the Agilent 9 Infinity Online D-LC/MS Solution Application Note Biopharmaceuticals Authors Abstract Ravindra Gudihal and This Application Note describes the Agilent 9 Infinity online

More information

Trastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform

Trastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform Trastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform Ying Qing Yu, Joomi Ahn, and Martin Gilar Waters Corporation, Milford, MA, U.S. APPLICATION BENEFITS Used together,

More information

A Robust cief Method: Intermediate Precision for the ph 5-7 Range

A Robust cief Method: Intermediate Precision for the ph 5-7 Range Robust cief Method: Intermediate Precision for the ph 5-7 Range Ingrid. Cruzado-Park, Scott Mack and Chitra K. Ratnayake SCIEX, Brea, C US Introduction Optimization of the numerous variables associated

More information

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for

More information

Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS

Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS Stephen Lock SCIEX UK Overview Who Should Read This: Senior Scientists and Lab Directors with an Interest in Protein Analysis

More information

Chapter 2 Coated Capillaries for CE-MS of Therapeutic Protein

Chapter 2 Coated Capillaries for CE-MS of Therapeutic Protein Chapter 2 Coated Capillaries for CE-MS of Therapeutic Protein James Q. Xia 2.1 Introduction to Capillary Coating Bare fused-silica capillary is so far the most commonly used type of capillary in CE-UV

More information

Antibody Analysis by ESI-TOF LC/MS

Antibody Analysis by ESI-TOF LC/MS Application Note PROTEOMICS METABOLOMICS GENOMICS INFORMATICS GLYILEVALCYSGLUGLNALASERLEUASPARG CYSVALLYSPROLYSPHETYRTHRLEUHISLYS Antibody Analysis by ESI-TOF LC/MS Authors Lorenzo Chen, Merck & Company,

More information

Application Note. Authors. Abstract. Biopharmaceuticals

Application Note. Authors. Abstract. Biopharmaceuticals Characterization of monoclonal antibodies on the Agilent 126 Infinity Bio-inert Quaternary LC by Size Exclusion Chromatography using the Agilent BioSEC columns Application Note Biopharmaceuticals Authors

More information

Columns for Biomolecules BioLC Column Lines

Columns for Biomolecules BioLC Column Lines Columns for Biomolecules BioLC Column Lines Monoclonal Antibodies Glycans GlycanPac Nucleic Acids DNAPac Protein A Accucore Amide-HILIC DNAPac PA1 SEC-1 GlycanPac AXH-1 DNAPac PA2 SCX-1 GlycanPac AXR-1

More information

One-stop cief and CE-SDS. One-day method dev. Results in a snap. FDA, PDQ.

One-stop cief and CE-SDS. One-day method dev. Results in a snap. FDA, PDQ. One-stop cief and CE-SDS. One-day method dev. Results in a snap. FDA, PDQ. Meet Maurice. He s a real problem solver. Need CE-SDS and cief info on mabs, ADCs and vaccines? Done. He ll even up you one and

More information

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI with the Biopharmaceutical Platform Solution with UNIFI Vera B. Ivleva, Ying Qing Yu, Scott Berger, and Weibin Chen Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S The ability to

More information

Detecting Challenging Post Translational Modifications (PTMs) using CESI-MS

Detecting Challenging Post Translational Modifications (PTMs) using CESI-MS Detecting Challenging Post Translational Modifications (PTMs) using CESI-MS Sensitive workflow for the detection of PTMs in biological samples from small sample volumes Klaus Faserl, 1 Bettina Sarg, 1

More information

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis CHROMATOGRAPHY MAbPac RP Column High-performance reverse phase chromatography column for monoclonal antibody analysis Product Specifications The Thermo Scientific MAbPac RP is a reverse phase (RP) liquid

More information

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis Thermo Scientific MAbPac HIC Columns Novel Hydrophobic Interaction HPLC Columns Designed for Monoclonal Antibody Analysis Introduction The Thermo Scientific MAbPac HIC column family is designed for separations

More information

Characterize mab Charge Variants by Cation-Exchange Chromatography

Characterize mab Charge Variants by Cation-Exchange Chromatography Characterize mab Charge Variants by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for

More information

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC Analysis of Intact and Fragmented mabs and ADC Application Note Biotherapeutics and Biologics Author Suresh Babu C.V. Agilent

More information

Answers, Simple and Streamlined. Solutions for Biotherapeutic Intact Mass Analysis

Answers, Simple and Streamlined. Solutions for Biotherapeutic Intact Mass Analysis Answers, Simple and Streamlined Solutions for Biotherapeutic Intact Mass Analysis Expert Functionality for Mass Spec Non-Experts SCIEX OS on the X500B QTOF system dramatically improves the user experience,

More information

A highly sensitive and robust 150 µm column to enable high-throughput proteomics

A highly sensitive and robust 150 µm column to enable high-throughput proteomics APPLICATION NOTE 21744 Robust LC Separation Optimized MS Acquisition Comprehensive Data Informatics A highly sensitive and robust 15 µm column to enable high-throughput proteomics Authors Xin Zhang, 1

More information

Proteomics. Proteomics is the study of all proteins within organism. Challenges

Proteomics. Proteomics is the study of all proteins within organism. Challenges Proteomics Proteomics is the study of all proteins within organism. Challenges 1. The proteome is larger than the genome due to alternative splicing and protein modification. As we have said before we

More information

Charge Variant Assessment of Nanobodies at the Intact Level by CESI-MS

Charge Variant Assessment of Nanobodies at the Intact Level by CESI-MS Charge Variant Assessment of Nanobodies at the Intact Level by CESI-MS Sensitive workflow for the detection of charge variants of nanobodies from small sample volumes Stephen Lock 1, Rob Haselberg 2, Raimond

More information

Lecture 5: 8/31. CHAPTER 5 Techniques in Protein Biochemistry

Lecture 5: 8/31. CHAPTER 5 Techniques in Protein Biochemistry Lecture 5: 8/31 CHAPTER 5 Techniques in Protein Biochemistry Chapter 5 Outline The proteome is the entire set of proteins expressed and modified by a cell under a particular set of biochemical conditions.

More information

Faster, easier, flexible proteomics solutions

Faster, easier, flexible proteomics solutions Agilent HPLC-Chip LC/MS Faster, easier, flexible proteomics solutions Our measure is your success. products applications soft ware services Phospho- Anaysis Intact Glycan & Glycoprotein Agilent s HPLC-Chip

More information

B05 - PEPTIDE MAPPING(Revision 1) INTRODUCTION

B05 - PEPTIDE MAPPING(Revision 1) INTRODUCTION 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 B05 - PEPTIDE MAPPING(Revision 1) INTRODUCTION Proteins can exist as large complex structures, with some molecules in the population

More information

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer Shanhua Lin, 1 Hongxia Wang, 2 Zhiqi Hao, 2 Patrick Bennett, 2 and Xiaodong

More information

Case Study on Application of Analytical Life Cycle Management and Risk Management

Case Study on Application of Analytical Life Cycle Management and Risk Management Case Study on Application of Analytical Life Cycle Management and Risk Management Jianmei, Ph.D. Genzyme, a Sanofi Company AAPS 2015 Dione Pompe disease Brazil www.genzyme.com Outline Introduction of Concepts

More information

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis Application Note Biologics Development AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis Using the Agilent 16 Infinity II Bio-Inert LC Authors Andrew Coffey and Sandeep

More information

Conjugated protein biotherapeutics

Conjugated protein biotherapeutics B i o P r o c e s s TECHNICAL Mass Spectrometric Conjugate Characterization Process Qualification of Recombinant Protein Hapten Conjugation Miao-Fang Lin, Margo Wilson, Michael V. Murray, and Greg W. Adams

More information

Rapid Identification and Characterization of Biologics using CESI 8000-TripleTOF MS Platform

Rapid Identification and Characterization of Biologics using CESI 8000-TripleTOF MS Platform Rapid Identification and Characterization of Biologics using CESI 8000-TripleTOF 5600 + MS Platform High Throughput Qualitative Analysis of Biopharmaceuticals Using CESI-MS Technology Rajeswari Lakshmanan

More information

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

High-Resolution Separation of Intact Monoclonal Antibody Isoforms High-Resolution Separation of Intact Monoclonal Antibody Isoforms Dai Zhenyu, Xu Qun, Jeffrey Rohrer Thermo Fisher Scientific, Shanghai, People s Republic of China; Thermo Fisher Scientific, Sunnyvale,

More information

autoflex max Practical Power, Maximum Utility Innovation with Integrity MALDI-TOF MS

autoflex max Practical Power, Maximum Utility Innovation with Integrity MALDI-TOF MS autoflex max Practical Power, Maximum Utility Innovation with Integrity MALDI-TOF MS Wide application versatility with enhanced dynamic range The autoflex max series is based on Bruker s robust and efficient

More information

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India Quantitation of Oxidation Sites on a Monoclonal Antibody Using an Agilent 1260 Infinity HPLC-Chip/MS System Coupled to an Accurate-Mass 6520 Q-TOF LC/MS Application Note Authors Ravindra Gudihal Agilent

More information

Peptide Mapping: A Quality by Design (QbD) Approach

Peptide Mapping: A Quality by Design (QbD) Approach Peptide Mapping: A Quality by Design (QbD) Approach Application Note Bio-Pharmaceutical Authors Sreelakshmy Menon and Suresh babu C.V. Agilent Technologies, Inc. Richard Verseput S-Matrix Corporation Abstract

More information

mab Glycopeptide Profiling with V-Tag Adding reliable glycoprofiling to your peptide mapping workflow with ease and simplicity

mab Glycopeptide Profiling with V-Tag Adding reliable glycoprofiling to your peptide mapping workflow with ease and simplicity mab Glycopeptide Profiling with V-Tag Adding reliable glycoprofiling to your peptide mapping workflow with ease and simplicity Who is the V-Tag Glycoprofiling Technology for? V = Velocity The Ludger V-Tag

More information

Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography

Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography 1 Presentation Importance of aggregate analysis in BioPharma manufacturing Analytical techniques

More information