Application of UV LEDs for Turbid Wastewater Disinfection

Size: px
Start display at page:

Download "Application of UV LEDs for Turbid Wastewater Disinfection"

Transcription

1 Application of UV LEDs for Turbid Wastewater Disinfection by Chenghui Zeng A thesis submitted in partial fulfillment of the requirements for the degree of Master of Engineering in Environmental Engineering and Management Examination Committee: Prof. Chettiyappan Visvanathan (Chairperson) Prof. Nguyen Thi Kim Oanh Dr. Oleg Shipin Prof. Kensuke Fukushi (External Expert) Nationality: Previous Degree: Scholarship Donor: Chinese Bachelor of Engineering in Water Supply and Sewerage Engineering Harbin Engineering University P.R. China China Scholarship Council (CSC) Asian Institute of Technology School of Environment, Resources and Development Thailand May 2014

2 Acknowledgements Completing my master thesis is a milestone in my academic career that would not have been possible without the support of many people to whom I am indebted. Prof. C. Visvanathan has been an ideal supervisor in every aspect. I learned so much from him and I am very grateful for his professional guidance and strong support. Without his consistent instructions, this thesis could not have reached its present form. I would also like to extend my sincere thanks to Prof. Nguyen Thi Kim Oanh and Dr. Oleg Shipin, my thesis committee members, for their invaluable comments and inputs all the time. My profound thankfulness goes to Prof. Kensuke Fukushi, my supervisor during the participation of UEHAS program, for giving valuable research recommendations and financial support throughout this research. Likewise, I would also like to thank Dr. Pu Jian, for her kind help during my stay in the University of Tokyo. The research group under the supervision of Prof. C. Visvanathan is a wonderful setting for my graduate experience. Here, I want to say thank you to all the team members-paul, Park, Paru, Pik, Milk, Ter, Mov Chinmeng, Plat, Ben, Ellis, and Lina. Part of the experiment was conducted in the nano-lab. My special thanks go to Ms. Mayuree, the research associate in nano-lab, for her kind guidance on microbial experiment and technical support. I also wish to acknowledge the financial support from China Scholarship Council. With this scholarship, I was able to pursue my master degree at AIT. I wish also to express my love and gratitude to my parents, my older sisters and older brother, whose love and support made it possible for me to pursue my interests and dreams. I love you! ii

3 Abstract UV LEDs was thought to be a good alternative to conventional UV lamp. In this study, the impact of turbidity on the disinfection performance of UV LEDs (282 nm) was examined. Inactivation of Escherichia coli (E. coli) and total coliform was compared based the exposure time. Actinometry methods have been applied to measure the UV fluence in the reactor, including iodide-iodate and ferrioxalate actinometer. Both synthetic and real wastewater have been used to conduct the disinfection test. Turbidity of synthetic wastewater was 27, 70, 113, and 156 NTU and that for real wastewater was 57, 72, 86, and 130 NTU. Irradiance of UV LEDs was found to be mw/cm 2 (iodide-iodate) and mw/cm 2 (ferrioxalate). In synthetic wastewater of 27, 70, and 113 NTU, 5 log-reduction of E. coli could be achieved and the inactivation kinetics was a first order reaction. However, in real wastewater of 72, 86 and 130 NTU, only 3 log-reduction of both total coliform and E. coli could be achieved. The reason is that part of the coliform bacteria or E. coli in real wastewater was attached to the particles. These particles can protect the bacteria from being exposed to UV irradiation, so bacteria can still be alive even though more UV fluence is applied. In conclusion, UV LEDs may not be able to disinfect the wastewater to meet the required guidelines for wastewater reclamation unless a proper pretreatment is applied to reduce the turbidity of wastewater to a certain level. iii

4 Table of Contents Chapter Title Page Title Page Acknowledgements Abstract Table of Contents List of Tables List of Figures List of Abbreviations i ii iii iv vi iv ix 1 Introduction 1.1 Background 1.2 Objectives of Study Scope of Study 2 2 Literature Review Introduction Ultraviolet Light-emitting Diodes (UV LEDs) Fundamentals of UV LEDs Advantages and disadvantages of UV LEDs Current status and applications Turbid Water Disinfection by Ultraviolet Light Application of simplified wastewater treatment process in developing countries Another possible application of UV LEDs for turbid water disinfection UV Disinfection Fundamentals Sources of UV UV disinfection mechanism and the followed photoreactivation Advantages and disadvantages of UV disinfection Factors Affecting the Disinfection Efficiency of UV Light Subordinate factors UV fluence Wavelength Absorbance and scattering by particles in water The types of microorganisms 2.6 Methods for UV Fluence Determination Biodosimetry method Chemical actinometry 2.7 Summary and Research Needs 25 3 Methodology Introduction Experimental Set-up UV Fluence Determination by Actinometry 28 iv

5 3.3.1 Iodide-iodate actinometer Ferrioxalate actinometer 3.4 Disinfection Test with Synthetic Wastewater Preparation of synthetic wastewater Preparation of E. coli for synthetic wastewater Enumeration of E. coli Procedure for disinfection test with synthetic wastewater 3.5 Disinfection Test with Real Wastewater Wastewater sampling Selection of challenge organism and its enumeration Procedure for disinfection test 3.6 Summary Results and Discussions Characterization of UV LEDs UV Fluence Determination by Actinometry Method Disinfection Test with Synthetic Wastewater Disinfection Test with Real Wastewater 51 5 Conclusion and Recommendations Conclusions Recommendations for Further Study 56 References 57 Appendix A Appendix B v

6 List of Tables Table Title Page 2.1 Required Dilution Factors for Both E. coli and COD to Reach a Safe Level Comparison between UV LEDs and Conventional UV Lamp Value for Each Parameter in Equation Value Adopted for Each Parameter in Equation 3.8 and Summary of Methods Absorbance of Idide-iodate Actinometer Solution at Different Exposure Time Absorbance of Ferrioxalate Actinometer Solution at Different Exposure Time Inactivation Kinetics of UV Irradiation in Different Turbid Wastewater Properties of Wastewater Samples 52 vi

7 List of Figures Figure Title Page 2.1 Structure of PN junction UV LEDs (282 nm) with 9 chips Distribution by continent of the 1.5 billion people using sewerage facilities with no treatment in a 2.4b A full wastewater treatment process A simplified wastewater treatment process Application of UV LEDs in SWTP Working conditions of WWTP in post disaster period in Japan UV light in the electromagnetic spectrum The spectra of low-pressure and medium pressure UV lamp Spectra of UV LEDs The disinfection mechanism of chemical disinfectant and UV irradiation Disinfection mechanism of UVB and UVC UV disinfection and photoreactivation The UV fluence (UV dose)-response curve of E. coli Absorbance of DNA to UV light with different wavelengths The absorbance and scattering of UV light The dose-response of different microorganisms Three kinds of inactivation kinetics The quasi-collimated beam apparatus A typical UV inactivation-fluence (dose) response curve for B. subtilis spores The actinometry method based on ferrioxalate actinometer Research framework Experimental set-up Quasi-collimated beam apparatus Comparison between reactor and spectroradiometer Procedures for UV fluence determination by iodide-iodate actinometer The procedure for UV fluence determination by ferrioxalate actinometer Procedure for making synthetic wastewater Determination of montmorillonite added into the wastewater The Process for making glycerol stock from E. coli strain The Chromocult Colifrom agar and the colony Series dilution Plating of dilutions Procedure for disinfection test with synthetic wastewater Research framework for disinfection test with synthetic wastewater AIT wastewater treatment process MPN method 3.17 The procedure for conducting disinfection test with real wastewater Research framework for disinfection test with real wastewater The emission spectrum of UV LEDs 44 vii

8 4.2 The procedure for UV fluence determination UV fluence determined by iodide-iodate vs. exposure time UV fluence determined by ferrioxalate vs. exposure time Turbidity of synthetic wastewater vs. concentration of montmorillonite Time-response curve of E. coli in different turbid wastewater Sensitivity of E. coli to UV irradiation in different turbid synthetic wastewater Time-response curve of E. coli in synthetic wastewater of 27 and 70 NTU Time-response curve of total coliform in different turbid wastewater Time-response curve of total coliform within 100 s UV exposure Time-response curve of E. coli in different turbid wastewater 54 viii

9 List of Abbreviations ABS APHA ASFB BOD B. subtilis c CBD CFU COD CT DBP DC DNA DW E. coli I IEQ J LB agar LED LP mercury lamp MDG MP mercury lamp MS2 NTU PBS PR Q β phage QCB RNA SEQ SWTP THM US EPA USPHS UV UVDGM Absorbance American Public Health Association Aerobic Spore-forming Bacteria Biochemical Oxygen Demand Bacillus Subtilis Speed of Light Collimated Beam Device Colony Forming Unit Chemical Oxygen Demand Collection Tank Disinfection Byproducts Direct Current Deoxyribonucleic Acid Distilled Water Escherichia coli Intensity Inlet of Equalization Tank Joule Luria-Bertani Agar Light-emitting Diode Low Pressure mercury lamp Millenium Development Goal Medium Pressure mercury lamp Male-specific-2 Nephelometric Turbidity Unit Phosphate Buffered Saline Photoreactivation F-specific RNA Bacteriaphage Quasi-collimated Beam Ribonucleic Acid Surface of Equalization Tank Simplified Wastewater Treatment Process Trihalomethane United States Environment Protection Agency United States Public Health Service Ultraviolet Ultraviolet Disinfection Guidance Manual ix

10 Chapter 1 Introduction 1.1 Background Disinfection technology has been applied in water and wastewater treatment for a long time. In the past decades, several kinds of disinfection methods have been developed, such as chlorination, UV irradiation, ozonization, etc. When compared to other methods, UV irradiation has many advantages, such as almost zero formation of disinfection byproducts, more user-friendly and high disinfection efficiency. Furthermore, the UV light can even inactivate some chlorine-resistant microorganisms, such as Giardia and Cryptosporidium (Caron et al., 2007), which is also a big concern in current water supply system. These advantages make UV irradiation become a promising technology for water and wastewater disinfection. Currently, most of the UV lamps are low- or medium-pressure mercury lamps (LP and MP lamps) and they have some sustainability issues because they use toxic mercury to generate the UV light and the corresponding lifetime ranges from 4,000 h to 10,000 h (Autin et al., 2013). So the disposal of this kind of lamp may cause some environmental pollution after using up. Another shortcoming of this kind of lamp is the low energy efficiency as most of the electricity was converted to heat, not the desired UV light, which makes it less costcompetitive than chlorination. UV light-emitting diodes (LEDs) offer a possible solution to this problem. UV LEDs do not contain toxic mercury, have a longer lifetime and their flexibility is also very high due to the compacted structure. Last but not the least, UV LEDs has the potential to offer a high energy efficiency (Bowker et al., 2011; Wurtele et al., 2011). It is essentially a PN junction opto-semiconductor that can emit the light with a defined wavelength when the electricity is applied on the semiconductor in a forward biased direction. The UV LEDs lamp remains cool when it works, so only a small amount of energy is converted to heat, which can improve its efficiency significantly as compared to a low- or medium-pressure mercury lamp. Due to these advantages, UV LEDs has attracted a lot of attentions from both industry and academic institutions and becomes a hot spot in the UV disinfection field. Some investigations have already been done on the application of UV LEDs for disinfection purpose (Chevremont et al., 2012; Oguma et al., 2013; Wurtele et al., 2011). Most of them just focus on the comparison of UV LEDs with different wavelengths or their combinations and disinfection performances on different kinds of microorganisms. The indicating microorganisms include Escherichia coli (E. coli), fecal coliform and Bacillus Subtilis spore (B. subtilis spore), etc. Little information has been paid to the investigation of disinfection performance of UV LED under turbid wastewater conditions. Application of UV irradiation is still limited by its high cost and it is even less costcompetitive than chlorination nowadays. So, it is just used to disinfect the secondary effluent in wastewater treatment, whose turbidity is relative lower. But UV LEDs will make UV irradiation become a cheap technology in the future because it has high energy efficiency. When the wastewater has a high turbidity, the disinfection efficiency of UV 1

11 LEDs might be reduced. But it may be still more cost-competitive than chlorination. So, its application to disinfect the turbid wastewater could be expected. In this research, the disinfection performance of UV LED (280 nm) under turbid wastewater condition has been investigated. Currently, there are several kinds of UV LEDs available, and the wavelengths are 255 nm, 265 nm, 280 nm, and 310 nm, etc. Now UV LED (280 nm) has a relative longer lifetime when compared with UV LEDs in deeper UV zone (less than 280 nm) and it also has a highest disinfection efficiency for time-based inactivation (Oguma et al., 2013). This research consists of three stages and total coliform and E. coli were selected as the indicating organism. At first stage, actinometry methods have been applied to determine the UV fluence. Actinometry methods are based on the photochemical reaction and more details are provided in literature review and methodology part. At the second stage, UV LEDs have been applied to disinfect synthetic wastewater with different turbidity. Montmorillonite has been added into the distilled water to synthetize the turbid wastewater and it was thought to be a good representative of particles in the wastewater due to its tendency to swell and its surface active properties (Passantino et al., 2004). The turbidity of synthetic wastewater was dependent on the amount of montmorillonite added into the distilled water. At third stage, disinfection test has been conducted with the real wastewater from AIT wastewater treatment plant. Wastewater samples were taken from different parts of the wastewater treatment process at different time. The quality of each sample is also different. 1.2 Objectives of Study The main objective of this study was to check the disinfection performance of UV LEDs under high turbid wastewater conditions. The specific objectives of this study are: 1) To develop a reactor for conducting the disinfection test with UV LEDs; 2) To determine the UV fluence in the reactor through two different actinometers and compare the results; 3) To check the disinfection performance of UV LEDs with both synthetic turbid wastewater and real turbid wastewater. 1.3 Scope of Study This research was limited to the experimental scale, rather than full scale, as the UV LEDs has not been implemented in industry yet. The UV fluence has been determined by the actinometry methods, and the secondary data in previous literatures have been applied for calculation. As the turbidity of real wastewater may just have short variation, the disinfection performance of UV LEDs under turbid water conditions may not be well understood. So, synthetic turbid water will be applied to get turbidity with a larger variation, and give an aid to better understand the impact of turbidity on the disinfection performance of UV LEDs. Additionally, the quality of wastewater varies seasonally and geographically. So the results of disinfection test with real wastewater are not reproducible. But disinfection test with synthetic wastewater can overcome this shortcoming. So, conducting disinfection test with both synthetic and real wastewater may be the best way to 2

12 conduct this research - disinfection test with synthetic wastewater can supply the reproducible results while the disinfection test with real wastewater can supply some results as reference in real case. 3

13 Chapter 2 Literature Review 2.1 Introduction This chapter provides a background of relevant information that relates to the research topic. The first section provides the fundamentals of ultraviolet light-emitting diodes (UV LEDs), including fundamentals of UV LEDs, its development, potential applications and the advantages of UV LEDs over the conventional UV lamp. The second section explains why UV LEDs have the potential to be applied for turbid wastewater disinfection. This is followed by the introduction of fundamentals of UV disinfection, including the classification of UV, UV sources, UV disinfections mechanism, and the advantages and disadvantages of UV disinfection. The next section presents some factors that could affect the disinfection efficiency of UV, such as UV fluence, wavelength of UV light, the type of microorganism, scattering and absorbance by particles. This is followed by a section introducing the UV fluence determination. Some methods used to determine UV fluence were introduced, including biodosimetry and actinometry. 2.2 Ultraviolet Light-emitting Diodes (UV LEDs) Fundamentals of UV LEDs Light-emitting diode (LED) is a kind of semiconductor. When the electricity is applied on the semiconductor in a forward biased direction, it will emit light with a specific wavelength. The principle behind this kind of light could be described by Figure 2.1. When the voltage is applied on the pn conjunction, the hole from p-type and the electron from the n-type will combine together and generating a photon. The wavelength of the photon is dependent on energy difference the carriers (hole and electron) overcome in order to combine. Figure 2.1 Structure of PN junction The history of LED could be dated back to H. J. Round of MarconiLabs discovered that some inorganic substances can emit light if an electric voltage is impress on them. Nowadays, LED technology has already been applied in many fields successfully, such as television, trafic light, and optical measurement systems, etc. 4

14 The invention of UV LEDs was just several decades later after the invention of LED. But its commercialization has taken a long time. At the early stage of its development, the energy efficiency is very low and the output power also could not satisfy the needs of applications, which is still a limitation for its implementation in industry nowadays Advantages and disadvantages of UV LEDs UV LEDs has attracted a lot of attentions due to its advantages, including: 1. It is expected to have high energy efficiency. The wall plug efficiency of UV LEDs were supposed to achieve 75% in 2020 (Autin et al., 2013). 2. Long lifetime. Its lifetime could be up to 100,000 hours. 3. Compacted structure. Figure 2.2 shows the picture of UV LEDs with 9 chips and it is quite robust. 4. No warm-up time. This makes it become very suitable for the frequent on-off cycling system. 5. Lower voltage. 6. It could emit light with any wavelength ranges from 200 nm to 400 nm. 7. More environmentally friendly. They are made from aluminum nitride (AlN) or gallium and aluminum nitride (AlGaN) and do not contain mercury, so it won t cause any disposal problems. 8. Less space requirement. Figure 2.2 UV LEDs (282 nm) with 9 chips However, UV LEDs still cannot be implemented in the industry due to its disadvantages, including: 1. Low output optical power, especially in deep UV zone. But the increase of its output power could be clearly expected through its physical improvement in a few years. Nowadays, the output power of one UV LED could be several hundred milliwatt (SeoulOptodevice, 2013). However, in 2011, an output power of 66 mw can be regarded as leading light output (Hayward, 2013). 2. The price of UV LEDs is very high. Because now the technology of making UV LEDs is still not mature, so its price is very high. The UV LEDs shown in Figure 2.2 costs about 1,000 $. 5

15 2.2.3 Current status and applications UV LEDs could be applied in many areas, such as water disinfection, wastewater disinfection, and UV curving, etc. Now the technology of making UV LEDs is still not mature so it has not been applied in industry yet. Almost all the published papers on UV LEDs are about research results at lab-scale, not the full-scale (Autin et al., 2013; Chevremont, Farnet, Coulomb, et al., 2012; Hayward, 2013; Oguma et al., 2013; Wurtele et al., 2011). Now the lifetime of UV LEDs also varies. The lifetime of UV LEDs from different companies are quite different due to different technologies applied to make UV LEDs. Before its implementation in industry, it still has a long way to go. Undoubtedly, some investigations on the application of UV LEDs can promote its implementation in industry. 2.3 Turbid Wastewater Disinfection by Ultraviolet Light Generally, UV disinfection is just applied to disinfect the clean water. When turbidity of water is higher than 2 NTU, a filter should be applied before UV irradiation. But in the future, UV may be applied to disinfect the turbid water when the cost of UV irradiation comes down. One good example of the application change due to development of technology is membrane. In the past, membrane technology was just used to filter the clean water because it is a costly technology. But when the price of membrane comes down, it was widely applied to filter the wastewater directly. Following information explain the reason why UV LEDs has the potential to disinfect the turbid wastewater Application of simplified wastewater treatment process in developing countries Supplying improved sanitation and access to safe drinking water for human being has been a goal for many international organizations and governments for a long time. A lot of work has also been done in order to achieve this goal. The information from the United Nations Millenium Development Goal (MDG) show that the target to reduce the proportion of people without access to improved source of water by half was achieved, five years ahead of the schedule. However, another alarming fact is that 2.5 billion people still have no access to the improved sanitation facilities. On the way to reach the MDG, human being still has a long way to go. Currently, the cost of wastewater treatment is still quite high, and many cities or communities in developing countries still cannot afford wastewater treatment. One research reveals that 1.5 billion people used sewage connections without treatment (Baum et al., 2013). The distribution of this number in each continent is illustrated in Figure 2.3. From this figure, it can be found that Asia accounts for most of them, and the percentage is as high as 68.5%. 6

16 5% % 4% Africa Asia 12% Europe North and Central America 68% South America Oceania Figure 2.3 Distribution by continent of the 1.5 billion people using sewerage facilities with no treatment in 2010 (Baum et al., 2013) Another research result also reveals that for towns in developing countries with a population from 2,000 to 50,000 often falls into what has been termed as the management gap: they are large and compact enough to have the centralized sanitation system, but they are not large enough to have the resources to manage these highly mechanized infrastructures (Pilgrim et al., 2008). In many developing countries, the sanitation facilities often fail prematurely due to lack of maintenance and proper operations. This may be caused several reasons, such as high initial investment, low initial revenues, and the complex operation of WWTP. Under such conditions, the full wastewater treatment process (Figure 2.4 a) may be not the best choice for the developing countries as it is very costly and very complex to operate. Some communities prefer to use a simplified wastewater treatment process (SWTP) to treat the wastewater and then discharge it. This simplified wastewater treatment process just consists of a screen and a tank for settling (Figure 2.4 b). Screen 1 st clarifier Aeration 2 nd clarifier Disinfection Figure 2.4 a A full wastewater treatment process Figure 2.4 b A simplified wastewater treatment process 7

17 The simplified wastewater treatment process can remove most of suspended particles in the wastewater. But the removal of dissolved pollutants may be very poor. So, the concentration of pollutants (such as COD, BOD, N, and P) in the effluent may be still quite high and they may cause pollution to the receiving water body. Rather than these organic and inorganic pollutants, more attentions should be paid to the pathogens remained in the effluent. For example, the COD of sewage ranges from mg/l, and it could meet the standard for discharge after around 20-folds dilution. However, the required dilution factor for fecal coliform in sewage could be as high as 1000 times. Furthermore, most of the pollutants contained in the wastewater are biodegradable, while the pathogens may be quite persistent in the environment. So, the risk would be reduced greatly if this kind of wastewater could be disinfected before discharge. Table 2.1 Required Dilution Factors for Both E. coli and COD to Reach a Safe Level Parameters Concentration Standard for discharge Required dilution factors COD (mg/l) E. coli (CFU/100 ml) UV irradiation may be a better choice than chlorination in this simplified process because the disinfection efficiency of chlorination could be affected by the ph, temperature, and many components in the wastewater. It can also lead to the formation of disinfection byproducts, which is carcinogenic. But UV irradiation is a physical method and it won t cause the formation of disinfection byproduct. Furthermore, it is relative easier to operate when compared with chlorination. But the cost may be a limitation factor for the application of UV irradiation. Now low pressure UV lamp and medium pressure UV lamp are the most common UV lamps applied in wastewater treatment field. They consume a large amount of energy and needs frequent replacement. Now the UV irradiation is even less cost-competitive than chlorination. Table 2.2 Comparison between UV LEDs and Conventional UV Lamp Energy saving Long lifetime No mercury UV LEDs Conventional UV lamp High energy consumption and high cost Frequent replacement Contains mercury But UV LEDs could overcome this shortcoming because it is expected to have a higher energy efficiency and lower cost in the future. The comparison between UV LEDs and conventional UV lamp is shown in Table 2.2. It can be concluded from this table that conventional UV lamp will be hopefully replaced by UV LEDs. So, the application of UV LEDs in SWTP is possible in the future (Figure 2.5). 8

18 Disinfection by UV LEDs Figure 2.5 Application of UV LEDs in SWTP However, one thing must be kept in mind is that the effluent from the sedimentation tank may still have a high turbidity. So, the impact of high turbidity in the primary effluent on the disinfection performation of UV LEDs should be well known Another possible application of UV LEDs for turbid water disinfection This simplified process may not just be used to in these cases. During post-disaster period, it may be adopted as the disinfection method in temporary wastewater treatment. It has been reported that some WWTPs in Japan were stopped by the earthquake and tsunami in 2011 because electricity was shut down after the earthquake and some infrastructures and equipments were also damaged by the tsunami (Masaru et al., 2013). However, the wastewater continues coming to the wastewater treatment plant as the every-day-life was going on. So, the wastewater cannot be treated by the full treatment process. Under such conditions, a large amount of wastewater was discharged to the adjacent water body after the simple sedimentation in the primary sedimentation tank (Figure 2.6). Disinfection before discharge can reduce the concentration of pathogens in the effluent, thus protect public health. Discharge Stop working after the earthquake and tsunami Figure 2.6 Working conditions of WWTP in post disaster period in Japan During post-disaster period, chlorination is not a suitable disinfection method because the road is damaged by the earthquake and chemical reagent (such as Cl 2, NaClO) cannot be transported to the WWTPs. Ozonation is also not practical because its operation is quite complex and consumes a large amount of electricity. UV irradiation may be the best choice because it is relative easier to operate and the electricity can be supplied by temporary electricity generator. Based on above information, it can be expected that UV LEDs will be applied to disinfect the turbid wastewater in the future due to its merits, such as less energy consumption, long lifetime, and environmental friendly. 9

19 2.4 UV Disinfection Fundamentals According to the physical definition, the electromagnetic radiation with a wavelength between 100 nm and 400 nm is called UV light (Figure 2.7). It is a nonvisible light. Due to the different biological effects associated with different wavelengths of the UV light, it is further classfied into 4 types: vacuum UV ( nm), UV-C ( nm), UV-B ( nm) and UV-A ( nm). Figure 2.7 UV light in the electromagnetic spectrum (courtesy of US EPA, 2006) As it is known, the energy associated with a photon is inversely proportional to the wavelegnth of light. This could be described by the following expression: 10 Equation 2.1 Where u is the energy (J) of a photon, h is the Planck constant (the value is J s), c is the speed of light (the value is m/s) in the vacuum, and λ is the wavelength (m) of light. Based on this, it can be easily concluded that the UV-A light is less energetic than the UV-B and UV-C. All kinds of UV lights are harmful to human being s skin. UV-A can cause the tanning of skin while the UV-B could cause the skin to burn and is known to eventually cause the skin cancer. UV-C is more powerful than UV-A and UV-B and it can be absorbed by proteins, ribonucleic acid (RNA) and deoxyribobucleic acid (DNA), then cause the cell mutations and/or cell death. It has the highest germicidal ability. Vacuum UV is even more powerful than UV-C. However, as it can be easily absorbed by the air and water, so it is seldom used for disinfection purpose Sources of UV UV light source can be classified into two sources: artificial and natural. The natural UV is mainly from the sunshine. UV light accounts for 3% of the total sunshine on the surface of the earth. Only a part of the UV light emitted by the sun could reach the surface of the earth and most of them are UV-A and UV-B because the UV-C is completely absorbed by the ozone layer and atmosphere. The artificial UV source includes many kinds of UV lamps, such as low-pressure (LP) mercury vapor lamp, medium-pressure (MP) mercury

20 vapor lamp, UV lasers, and light emitting diodes (LED), etc. They use different materials to emit the UV light, such as Argon, Xenon, and mercury vapor, etc. The mercury type UV lamp is the one which is most commonly used in the water and wastewater treatment. The mechanism of this kind of lamp is that when the electric current passes through the mercury vapor, the mercury atoms are excited by the collisions with the electrons flowing between the electrodes. The excited electrons returns to the particular electronic states in the mercury atom and in doing so the electrons will release the energy they have absorbed in the form of UV light. Figure 2.8 The spectra of low-pressure and medium pressure UV lamp (Emperor Aquatics, 2013) Based on the vapor pressure of mercury, the mercury UV lamp can be further classified into 3 types: low pressure mercury lamp, medium pressure mercury lamp, and high pressure mercury lamp. High pressure mercury lamp is mainly used for emitting the visible light after a modification rather than generating the UV light. The spectras of LP and MP UV lamp were shown in Figure 2.8. For low pressure mercury lamp, the mercury vapor pressure ranges from 0.14 to 14 Pa and the temperature of the mercury is 40 and it just produces the monochromatic UV light with wavelength of 254 nm. For medium pressure mercury lamp, the mercury vapor pressure is much higher (from 14 kpa to 1400 kpa) and the operating temperature is also much higher ( ) (USEPA, 2006). UV LEDs is a new kind of UV source, whose spectra is shown in Figure 2.9. The mechanism behind this kind of UV light is totally differently from the conventional mercury lamp. It is a kind of semiconductor that when electricity is applied on it in a forward direction, it can emit UV lights with a specific wavelength. However, at this stage of development, the output power of UV LEDs is still very low and the cost is also quite high, so it has not been fully implemented in industry yet. 11

21 Figure 2.9 Spectra of UV LEDs (DOWA, 2013) UV disinfection mechanism and the followed photoreactivation UV disinfection mechanism Unlike the chemical disinfectants (such as chlorine and ozone), who kill microorganisms mainly through damaging cell wall and some intracellular moleculesm (Figure 2.10), UV light inactivates microorganism in a totally different way. Figure 2.10 The disinfection mechanism of chemical disinfectant and UV irradiation The disinfection mechanism varies based on the wavelength of UV light. For UV light with shorter wavelength (UVB and UVC), the disinfection mechanism is mainly about the formation of cis-syn cyclobutane pyrimidine dimers in the genome DNA or RNA of organism. This result in the genetic disorder and then the replication, transcription and reproduction process are stopped, and eventually lead to inactivation of microorganisms (Figure 2.11). In this process, the photoproducts are also produced, but it is much less important than the damage caused to nucleic acid (Oguma et al., 2002). Figure 2.11 Disinfection mechanism of UVB and UVC 12

22 However, the principle behind the UVA disinfection is totally different. UVA is known to inactivate the microorganism mainly through exciting the photosensitive molecules, such as, H 2 O 2, and, which can damage the genome and other intracellular modecules and cause lethal or sublethal effects, such as mutation and growth delay. UVA LED has already been tested as water disinfection technology by some researchers (Hamamoto et al., 2007). Photoreactivation The past research results show that some microorganisms can repair the UV-damaged DNA or RNA through many ways (Oguma et al., 2002, 2004, 2005; Zimmer et al., 2002). But only the photoreactivation is discussed here because it can impair the UV disinfection efficiency significantly within several hours after UV disinfection test. Other repair mechanism, which is referred as dark repair, is less important than photoreactivation. But in real wastewater or water treatment, photoreactivation was thought to be not as serious as in lab because the water or wastewater will be discharged to the river or into the pipe, so the visible light is isolated or reduced greatly. Without the energy from the light, photolyase was unable to repair the dimmer in DNA. But attentions should be paid to the photoreactivation if the disinfected water will be exposed to the room light. Photoreactivation process is a totally inverse process of disinfection. The relationship between the disinfection and photoreactivation (PR) could be described by Figure When DNA is exposed to the UV light, it can lead to the formation of dimmer, while under white light condition this damage will be repaired by a kind of enzyme called photolyase. The specific wavelength of light that can cause photoreactivation ranges from nm. This means that UVA can disinfect the microorganism and cause photoreactivation simultaneously. T T DNA photolyase White light PR UV disinfection T T DNA with dimer Figure 2.12 UV disinfection and photoreactivation The UV light with different wavelength can suppress the photoreactivation of microorganism differently. Medium-pressure UV lamp was thought to be able to suppress the photoreactivation of E. coli more than the monochromatic low-pressure UV lamp and the UV light with broad spectra was also thought to be more effective on repressing photoreactivation of E. coli. This suggests that medium-pressure UV lamp can offer an advantage over low-pressure UV lamp in drinking water or wastewater treatment. This 13

23 kind of effect might be attributable to the suppressing effect on photolyase caused by UV light with relative longer wavelength. Someone even thought that the UV light at around 280 nm can repress the photolyase most effectively (Hu et al., 2008). Additionally, high salinity was also found to be able to suppress the photoreactivation of E. coli after UV irradiation. The NaCl solution at 2.4% or above (in weight/volume) can suppress the photoreactivation of E. coli after UV irradiation significantly, but the NaCl solution at 1.9% or lower did not shown such effect (Oguma. et al., 2013). This suggests that the photoreactivation of E. coli potentially may occur in brackish and costal area where the salinity is rather low. For different microorganisms, the photoreactivation effects are also different. The photoreactivation of Legionella pneumophila (L. pneumophila) after both low-pressure and medium-pressure UV lamp are almost the same, which is quite different from that of E. coli. This suggests that the E. coli cannot correctly indicate the fate of L. pneumophila in UV disinfection system (Oguma et al., 2004) Advantages and disadvantages of UV disinfection UV disinfection has many advantages over other chemical disinfections methods. One major advantage is that it can inactivate some chlorine-resistant pathogens, such as Giardia cysts and Cryptosporidium oocysts (Craik et al., 2000; Craik et al., 2001). At the same time, the chemical disinfection methods may change the water quality. For instance, chlorination may lead to the formation of disinfection byproducts, such as trihalomethane (THM) while UV disinfection won t has such a problem. Low-and medium-pressure mercury UV lamps did not have a significant impact on the formation of DBP when the dose is less than 500 mj/cm 2 (Liu et al., 2002). The recommended UV dose for disinfection in drinking water treatment plant is 40 mj/cm 2, which is well below 500 mj/cm 2 (Mosher et al., 2012).So UV disinfection won t cause the formation of DBP. Furthermore, the contact time for UV disinfection is very short (generally a few seconds) so the space requirement of the disinfection device is lower. However, UV disinfection may also have some shortcomings. Unlike the chemical disinfectants, UV won t leave any residuals after disinfection in drinking water treatment, so it cannot prevent the reproduction of microorganisms in the water after disinfection. In practice, the UV disinfection is often combined with the chemical disinfection, such as chlorination and ozonization. Additionally, UV disinfection is very sensitive to the turbidity of water. High turbidity could reduce the disinfection efficiency significantly and cause the scaling problem of the lamp envelope at the same time. The scaling problem could be solved through the regular cleaning of lamp envelope. Last but not the least, some microorganisms could be reactivated after the UV disinfection. The mechanisms include photorepair and darkrepair. The reason why the microorganism could be reactivated is that some enzyme system of the microorganism could repair the DNA or RNA damaged by the UV light (USEPA, 2006). 2.5 Factors Affecting the Disinfection Efficiency of UV Light The disinfection efficiency of UV light could be affected by many factors, such as UV fluence, color, wavelength, particles in water, type of microorganism, fluence rate, 14

24 temperature and ph of the water. These factors can impact the disinfection performance of UV light in different ways Subordinate factors Generally, the impacts of UV intensity, temperature and ph of the water on the disinfection performance of UV were thought to be negligible. When the UV intensity ranges from 1mW/cm 2 to 200mW/cm 2, the UV dose-response of microorganism follows the Law of Reciprocity. The UV light with a low intensity can have the same disinfection performance with a highed intense UV light when the UV dose is the same. The UV dose required for a given log reduction of E. coli, Candida parapsilosis, and f2 bacteriaphage increased slightly as the temperature decreased (Severin et al., 1983). ph was also thought to be a negligible factor for UV disinfection because it can impact neither the transmission of UV light in the water nor the intensity of UV light. Generally the ph of wastewater or drinking water is around 7, and it does not have a significant impact on UV disinfection UV fluence Both UV fluence and UV dose are used in the UV disinfection literature. But UV fluence is the appropriate term for UV disinfection. UV dose means the energy absorbed by the microorganism. In the case of microorganisms, almost all incident UV light passes through the organism with only a few percent being absorbed. So, UV fluence is a more appropriate term (Bolton et al., 2003). It is the main factor that affecting the disinfection efficiency of UV light. UV fluence is defined as the product of fluence rate (mw/cm 2 ) and the exposure time (s). In North America, the unit for UV fluence is mj/cm 2 while in Europe people prefer to use J/m 2. Generally, the higher log-reduction could be achieved with the higher UV fluence. A UV fluence-inactivation response curve of E. coli was shown in Figure It can be found from this figure that the log-reduction of E. coli was different at different UV fluence. Figure 2.13 The UV fluence (UV dose)-response curve of E. coli (USEPA, 2006) Wavelength Wavelength is another important factor that could impact the disinfection efficiency of UV irradiation. The UV light inactivates the microorganisms mainly through destroying the DNA or RNA of the microorganism. The DNA absorbs the UV with a wavelength from 200 nm to 300 nm and tend to have a peak at around 260 nm, which is shown in Figure 2.13 (USEPA, 2006). Microorganisms are most sensitive to the UV light with a 15

25 wavelength of 260 nm. So, it means that the UV light with a wavelength of 260 has the highest germicidal ability. However, some kinds of virus may be more sensitive to the UV light with a wavelength below 230 nm (Linden et al., 2001). In the past, the most common UV lamps applied in water and wastewater engineering are monochromatic low-pressure UV lamp and polychromatic medium pressure UV lamp, so only 254 nm UV light and a polyspectra UV light are available in practical engineering. But now, due to the development of light-emitting diodes, more and more monochromatic UV lamps with different wavelength become available. Some investigations on the disinfection performance of UV LEDs with different wavelength have already been done (Chevremont, Farnet, Sergent, et al., 2012; Hamamoto et al., 2007; Oguma et al., 2013; Wurtele et al., 2011). It can be expected that UV LED with a desired wavelength can be used for the disinfection of a specific kind of wastewater, especially when the microorganism requries the UV light with a specific wavelength. Figure 2.14 Absorbance of DNA to UV light with different wavelengths Absorbance and scattering by particles in water UV disinfection can be impacted by the particles in the wastewater significantly. This is done mainly through two mechanisms: scattering and absorbance by particles. The light scattered by the particles can still inactivate the pathogens in the water, while the light absorbed by particles was not able to. Additionally, the particles can also protect the bacterial cells through shielding effects. When the particles size is big enough, the bacterial cells can harbor inside the particles so that UV cannot inactivate them. Another fact that should also be noticed is that the water can also absorb the UV energy. The distilled water can absorb 8% of the UV light energy at a depth of 3 cm. When some solids dissolved in the water, it may also contribute to this kind of blocking effect. Figure 2.15 The absorbance and scattering of UV light 16

26 Particles in the wastewater can be measured in two ways: total solids (TS) and turbidity. TS could be further classified into two types: total dissolved solids (TDS) and total suspended solids (TSS). Both TS and turbidity can be used to monitor the wastewater quality in wastewater treatment. TS is the measure of weight of particles in the wastewater, so it can reflects the amount of particles in the wastewater exactly, regardless of its properties and size. But it takes a long time to measure as the procedure is more complex than that of turbidity. Turbidity is relative easier to monitor, but it cannot reflect the amount of particles in the wastewater exactly, because it can be greatly impacted by the surface properties of particle, particle size, etc. For example, 50 mg/l kaolin clay give a turbidity reading of about 80 NTU, while 50 mg/l humic acid give a turbidity reading slightly greater than 3 NTU. But turbidity can be monitored automatically, so sometimes the technician prefers to use turbidity to describe the wastewater quality. Particles with different size can impact the disinfection performance of UV light through different mechanisms. Goethite particles (0.2 μm 2 μm) can attach to the surface of E. coli (0.5 μm (1-3) μm), then protect the bacterial cell from UV disinfection even at a low turbidity level (1 5 NTU). Some research results suggest that the coliform shielding effect is mainly attributable to the particles with a diameter of 7-10 μm (Jolis et al., 2001; Qualls et al., 1983). But for virus, such as MS2 coliphage (diameter = 23 nm) and bacteriophage T4 (90 nm (25-200) nm), the particles with a diameter < 2 μm is large enough to provide the shielding effect (Templeton et al., 2005). The properties of particles can also affect the disinfection performance of UV significantly. Particles with different composition can protect the bacteria cell differently. For example, the humic acid can be linked with the E. coli through an affinity effect (Cantwell et al., 2008). The humic acid can attach on the surface of E. coli and then protect them from UV disinfection. The same effect was observed on virus. Humic acid and activated sludge floc particles were found to be very effective on shielding the virus from UV disinfection, while the kaolin clay provide no significant protection (Templeton et al., 2005). This can be explained through the property of humic acid. The organic compounds contained in humic acid can absorb the UV light, so UV light was absorbed before reaching the bacteria. But kaolin clay does not absorb the UV heavily relative to humic substances(bitton et al., 1972). To sum it up, two factors should be taken into consideration when investigating the impact of particles on the disinfection performance: (1) the size of particles and target microorganisms; (2) the interaction between the particles and target microorganisms. In practice, the size and composition of particles varies seasonally and geographically, so it is impossible to synthetize a turbid wastewater that can represent all kinds of wastewater. No matter what kind of particles (such as kaolin clay, montmorillonite, humic acid) have been used to synthetize the turbid wastewater, it still has some limitations to reflect the practical conditions. But the benefit of using synthetic wastewater is that the results are reproducible, which is an important factor to get a convincible research result. But if natural water and wastewater sample are used, it would be impossible to carry out controlled and reproducible experiments (Kollu et al., 2012). So, conducting disinfection test with synthetic wastewater is still a reliable and popular way to investigate the impact of turbidity on UV disinfection performance. 17

27 2.5.5 The types of microorganisms Different type of microorganism could show a different fluence-inactivation response. This means that the inactivation efficiency of UV may be also impacted by the type of microorganism. Figure 2.15 shows the fluence-inactivation response of different microorganisms (USEPA, 2006). This figure is quite old, so the fluence is till called dose in this figure. It can be found that the E. coli is more sensitive than other species of microorganisms and B. subtilis is most UV-resistant. Figure 2.16 The dose-response of different microorganisms The fluence-inactivation response of microorganisms is generally described through the inactivation kinetics, which is a first-order model. It is the same with that of chemical disinfection. This model could be described by the following equation: ( ) Equation 2.2 Where N and N 0 are the concentrations of microorganism after and before disinfection, respectively. Some researcher found that this model cannot be used to describe the inactivation kinetics of microorganisms under all conditions. Later, two other models were also developed, which are called shoulder model and tailing model (Figure 2.16). The causes of tailing are still a matter of debate, but attachment of bacteria to particles was supposed to be one of them. Shouldering was hypothesized to be due to the formation of microorganism aggregates, photoreactivation or dark repair. Figure 2.17 Three kinds of inactivation kinetics 18

28 2.6 Methods for UV Fluence Determination Currently, three methods are available for the determination of UV fluence. They are: 1) biodosimetry; 2) chemical actinometry; 3) mathematical model; 4) new validation method. Among them, biodosimetry method and chemical actinometry method are more suitable for bench scale, while the mathematical model method and the new validation method are more widely used in practical engineering. Here, the chemical actinometry method has been applied to determine the UV fluence in the reactor for exposure experiment Biodosimetry method Biodosimetry method is based on the fluence-inactivation response of some microorganisms. A quasi-collimated beam (QCB) bench scale apparatus must be used in this kind of method to get the standard fluence-inactivation response curve. Then this curve could be used as a reference for the UV fluence in other reactors. This kind of method has three steps: (1) Calculating the UV fluence in a QCB apparatus. The QCB apparatus are schematically described in Figure The UV lamp is put in an enclosure. A long tube is connected with the enclosure, which is used to collimate the UV light. So, on the surface of solution in the petri dish, the intensity of UV light is quite uniform and could be measured by the spectroradiometer. UV lamp Lamp enclosure UV light Collimating tube Magnetic stirrer Petri dish Figure 2.18 The quasi-collimated beam apparatus For low-pressure UV lamp, it has a monochromatic emission and the average germicidal fluence rate E avg could be calculated by following equation (Bolton et al., 2003). Equation

29 Where: 1) E 0 is the UV fluence rate measured by the spectrophotometer at the center of the petri dish. 2) Petri factor is the ratio of the average of the incident irradiane over the area of the petri dish to the irradiance at the center of the dish and is used to correct the irradiance reading at the center of the petri dish to more accurately reflect the average incident fluence rate over the surface area. 3) When UV light pass from one medium to another medium, a small part of the light will be reflected off the interface between the two media. For UV light between nm, the reflection factor could be adopted as ) Water factor is used to correct the errors caused by the adsorption of water to UV light. It is proportional to the depth of water. This factor could be calculated by following equation. Equation 2.4 a is the adsorbance for a 1 cm length path. l is the vertical path length (cm) in the petri dish. For the MP UV lamp, the calculation for water factor is more complex because the water absorbance to the UV light with different wavelength is different. So, the correction must be made over a narrow band (1-5 nm) of wavelength. 5) The divergence factor is used to correct the error caused by the unperfect collimation by the beam. For finite distance between the petri dish and the UV lamp, the irradiance falls off as the inverse square of the distance L from the UV lamp to the surface of the cell suspension. Assume the irradiance of UV light at distance L is I 0. So the irradiance at L+X could be expressed as: 20 Equation 2.5 Divergence factor is the average of this function over the path length l of the cell suspension and it could be expressed by following equation: Equation 2.6 Theoretically, the divergence and the water absorbance should be considered together to correct the error happened during the propagation of UV light. However, for path length less than 5 cm, the errors involved in treating them separately are negligible. The UV fluence on the surface of petri dish could be calculated by the following equation. ( ) Equation 2.7 For the medium pressure UV lamp, the determination of the UV fluence will be much more complex because it has a polychromoatic emission and the germicidal ability of UV light at different wavelength is different. So, only low pressure UV lamp is used in biodosimetry.

30 (2) Plotting standard inactivation-fluence response curve of challenge microorganism After getting the fluence rate above the surface of the solution in the petri dish, some kinds of solution contains challenging microorganism will be used to do the exposure test with the QCB. The challenge microorganism include F-specific RNA bacteriophage Q β (Qβ phage), Bacillus Subtilis spore (BS spore), and MS2 coliphage, etc. Calculate the log-reduction of the challenge microorganism and the corresponding UV fluence. Then plot the standard inactivation-fluence response curve. A typical standard curve is shown in Figure The corresponding equation could also be obtained by doing the linear regression with the standard curve. This equation could be expressed as following: ( ) Equation 2.8 ( ) is the log reduction of microorganism. This parameter could be calculated through the concentration of microorganism before (N 0 ) and after (N) the exposure test, which is relatively easy to obtain. Both k and b are the constant, which is determined by the nature of microorganisms selected. Figure 2.19 A typical UV inactivation-fluence (dose) response curve for B. subtilis spores (Qualls et al., 1983) (3) The last step of this method is conducting the UV exposure test with other reactors. Repeat the same procedure in last step to calculate the log-reduction of challenge microorganism. Then the corresponding UV fluence can be infered from the standard curve or the corresponding equation. The UV fluence obtained here is called reduction equivalent fluence. However, the bioassay method is quite time-consuming and relatively expensive. To some extent, this kind of method is not suitable for on-site measurement of UV. Furthermore, this kind of method only provides the mean fluence in bench scale. In practice, the best way for UV fluence determination is to combine the mathematical model with the actinometry method Chemical actinometry Chemical actinometry is another well-developed method for UV fluence measurement. This method is based on the photochemistry. In this kind of method, some kinds of light- 21

31 sensitive chemicals are used to absorb the energy from UV light, including KI/KIO 3 solution, uridine and potassium ferrioxalate, which are called actinometers. In this study, two kinds of actinometers have been applied to determine the UV fluence. One is KI/KIO 3 actinometer, and another one is potassium ferrioxalate actinometer. (1) KI/KIO 3 actinometer The composition of KI/KIO 3 actinometer is 0.6 M KI, 0.1 M KIO 3 and 0.01 M Na 2 B 4 O 7. The role of Na 2 B 4 O 7 in this actinometer solution is to maintain the solution has a constant ph of 9.2, so the solution won t turn to acid condition, which can lead to the oxidation of I -. The principle behind this method is a photochemical reaction, which is shown as below: 8 I + IO H 2 O + h 3 I OH Equation 2.9 In this chemical reaction, the number of I 3 - formed (N) has a linear relationship with the number of photons absorbed by actinometer solution (P). This relationship could be expressed by following formula: Where Φ is the quantum yield, moles I 3 - /mole photon. N = P Φ Equation The number of I 3 formed in this reaction could be inferred from the increase of its concentration ( C, mole/l). Where V is the volume of actinometer solution, L. N = C V Equation 2.11 The concentration of I 3 can be determined spectrophotometrically, and it is proportional to increase of absorbance of actinometer solution at 352 nm, so: C = ABS/ε Equation 2.12 Where ε is the adsorption coefficient, L mole -1 ; ABS= ABS(before UV exposure) ABS (after UV exposure). Energy contained in UV light (E) is the product of moles of photon and the photon energy (U), so: E = P (moles) U (J/mole) Equation 2.13 Assume that the exposure area is A and the exposure time is t, so the formula for the UV fluence calculation could be expressed as following: And the irradiance (I) could be expressed as: Equation

32 Equaton 2.15 When using this to calculate the irradiance of UV lamp, the reflection of UV light by water should also be taken into consideration. Assume the reflection factor is α, so the irradiance of UV lamp (I L ) is: I L Equaton 2.16 For the reflection between air and water, the correction factor is The procedure for conducting this experiment could be described as following: 1. Measure the absorbance of actinometer solution at 352 nm (ABS (blank) ); 2. Calculate the exposure area (A); 3. Add actinometer solution into the reactor, note the volume of actinometer as V; 4. Turn on the UV lamp and let UV irradiate to the actinometer solution; 5. After a period of time, turn off the UV lamp, note the exposure time as t; 6. Measure the absorbance of actinometer solution again at 352 nm (ABS (sample) ); 7. Calculate the UV fluence by Equation The procedure for preparation of actinometer solution and the selection of parameters for calculation is detailedly introduced in methodology part. This actinometer solution has a numerous advantages. Firstly, it is optically opaque to the light with a wavelength shorter than 290 nm. So, all the UV light within the germicidal range could be absorbed. It can be used as the photon counter to measure the UV fluence. Secondly, KI/KIO 3 solution is optically blind to the light with a wavelength longer than 330 nm, which means that it won t absorb the room light. So, the fluence determination experiment can be conducted in the presence of room light. Thirdly, all the chemicals used in this kind of method are commercially available. However, this kind of method also has some shortcomings. For example, KI is not easy to store in the lab because it can absorb the moisture and then decompose to KOH and I 2. Furthermore, due to the slowly thermal reaction happened in this solution, it can just be stored for 4 hours. So, the fresh solution needs to be made frequently. Last but not the least, the quantum yield of KI/KIO 3 at 282 nm still has not been accurately defined. Different results will be obtained by adopting different quantum yield. (2) Potassium ferrioxalate actinometer Potassium ferrioxalate is another kind of actinometer that has been widely accepted as the standard actinometer for UV fluence determination. Involved photochemical reaction can be expressed as follows: Equation 2.17 After exposure of a ferrioxalate solution to UV light, the will be converted to (Figure 2.18). The generated could be determined through colorimetric method in which complexed with o-phenanthroline. 23

33 Figure 2.20 The actinometry method based on ferrioxalate actinometer The principle behind this kind of method is almost the same with that of KI/KIO 3 actinometer. The UV fluence is proportional to the increase of concentration of Fe 2+ formed during the UV exposure. And the concentration of Fe 2+ could be determined spectrophotochemically at 510 nm. So, the UV fluence could be inferred from the change of absorbance of actinometer solution. Following is an example for this calculation. Assume that the absorbance of actinometer solution before UV exposure is ABS 510 (blank) and the absorbance of actinometer solution after UV exposure is ABS 510nm (Sample). So the absorbance of actinometer solution has increased by: ABS= ABS 510nm (Sample) - ABS 510 (blank) Equation 2.18 So the concentration of Fe 2+ has increased by: Where ε is the adsorption coefficient, mole -1 cm -1 ; C (Fe 2+ ) = ABS ε Equation 2.19 Then the total moles of Fe 2+ formed during this chemical reaction are: N = C (Fe 2+ ) V (sample) Equation 2.20 Where V (sample) is the volume of sample exposed to the UV LEDs. Assume that the quantum yield of this actinometer is Φ (moles I 3 - /mole photon.), so the moles of photon (P) is: 24 Equation 2.21 Assume the photon energy is U (J/mole) and the exposure area is A, so the UV fluence (E) is:

34 Equation 2.22 In conclusion, the formula for UV fluence calculation could be expressed as follows: ( ) Equation 2.23 When using this kind of actinometer to measure the UV fluence, there are several precautions: 1. The phenanthroline solution cannot be exposed to the room light because they may decompose and then cause interference to the final result; 2. The conversion of Fe 3+ should be less than 5%; 3. Every operation should be done in red illumination condition. 2.7 Summary and Research Needs In summary, UV LEDs are promising technologies in UV disinfection field because they have a lot of advantages over conventional UV lamps, such as higher energy efficiency, more environmental friendly, and potential longer lifetime, etc. Due to these advantages, UV irradiation will become a cheap method for wastewater disinfection. It can be expected that its implementation will not just be limited to the disinfection of clean water, but also turbid wastewater. But the impact of high turbidity on the disinfection performance of UV LEDs is still not well known. Past research results show that the particles in the wastewater can impact the UV disinfection performance significantly. The mechanism is dependent on the properties of particles, such as composition, size distribution, and surface charge. But in research, synthetic wastewater is always used in ordered to carry out the controlled experiments and obtain reproducible research results. One of the challenges to conduct the disinfection test with the reactor of UV LEDs is UV fluence determination in the reactor. As the size of reactor is quite small and the UV intensity distribution is very inhomogeneous, so spectroradiometer cannot be used to measure the UV intensity. One possible way to measure the UV fluence is the actinometry method. This method is less time-consuming, cheap and simple relative to biodosimetry. In conclusion, as a promising technology for wastewater disinfection, some investigations on the application of UV LEDs for turbid wastewater disinfection can provide some valuable information on their application in the future. 25

35 Chapter 3 Methodology 3.1 Introduction The objective of this study was to investigate the impact of turbidity on the disinfection performance of UV LEDs (280 nm). The research framework is shown in Figure 3.1. At first, the reactor has been built, based on the characteristics of UV LEDs, such as size, life time and emission power, etc. Figure 3.1 Research framework After building the reactor, the actinometry method has been applied to measure the UV fluence in the reactor. The actinometer used is iodide-iodate actinometer and ferrioxalate actinometer. For iodide-iodate actinometer, the fluence determination experiment was conducted under room light condition, while the ferrioxalate actinometer is quite sensitive to the room light, so it has been conducted under the subdued red light condition. At last, both synthetic wastewater and real wastewater have been used to conduct the disinfection test. Synthetic wastewater was synthetized through adding montmorillonite and E. coli into the distilled water. The wastewater with different turbidity was obtained through controlling the amount of montmorillonite added into the distilled water. Additional E. coli was added as the challenge microorganism. 26

36 In practice, the quality of wastewater may vary seasonally or geographically, so one kind of synthetic wastewater cannot represent all kinds of wastewater. In order to check the disinfection performance of UV LEDs with real wastewater, real wastewater with different turbidities has also been used to conduct the disinfection test. 3.2 Experimental Set-up When designing the reactor with UV LEDs, several factors must be taken into consideration, including the emission power of UV LEDs, and the lifetime of UV LEDs. Generally, the emission power of UV LEDs is still quite low and the lifetime of UV LEDs is quite short at current stage of development. In order to overcome such shortcomings of UV LEDs, the distance between UV LEDs and surface of solution must be short enough, so the intensity of UV light won't attenuate too much. Additionally, the volume of solution should also be small enough. A big volume of solution may lead to a long exposure time, which may cause a big challenge to the lifetime of UV LEDs. In this study, the emission power of UV LEDs is 10.8 mw and lifetime is around 100 hours. Based on such facts, the volume of solution was defined as 5 ml and the distance between the UV LEDs and the surface of solution is around 4-6 cm. The reactor was shown in Figure 3.2. UV LEDs is placed above the vessel used to contain the actinometer solution and water sample. A magnetic stirrer has been used to promote the dispersion of solution in the vessel. The power was supplied by the DC power and the voltage and current are 30 V and 0.06 A, respectively. The position of both vessel and UV LEDs are fixed so the UV intensity in the vessel could keep constant all the time. UV LEDs 2.2 cm Power source Stirring bar Volume =5 ml Figure 3.2 Experimental set-up 3.3 UV Fluence Determination by Actinometry In quasi-collimated beam apparatus that composed of conventional UV lamp (Figure 3.3), the UV intensity on the surface of petri dish is quite uniform and could be measured by the spectroradiometer. The UV fluence could also be calculated through the mathematical model, which was already introduced in biodosimetry method part. 27

37 Figure 3.3 Quasi-collimated beam apparatus But UV intensity distribution in this reactor (Figure 3.2) is very inhomogeneous and the diameter of reactor (φ = 2.2 cm) is even smaller than the diameter of probe of spectroradiometer (φ = 2.5 cm) (Figure 3.4), so the UV intensity cannot be measured by the spectroradiometer. Under such conditions, there are two possible ways to measure the UV fluence inside the reactor. One way is biodosimetry method, and another way is actinometry method. Biodosimetry method is more widely accepted, but it is also more time-consuming and requires the cultivation of virus or other microorganisms. So, the adopted one here is actinometry method. This kind of method is based on some photochemical reaction and it could be adapted to vessel with any kinds of geometrical shape. When compared with the biodosimetry method, it is also cheaper and less timeconsuming. UV LEDs Φ=2.2 cm Power source Stirring bar Volume =5 ml Figure 3.4 Comparison between reactor and spectroradiometer Now many kinds of chemical actinometers are available for UV fluence determination, such as uridine, KI/KIO 3, ferrioxalate. In this research, two kinds of actinometers have been applied to determine the UV fluence in UV LEDs reactor: iodide-iodate and ferrioxalate actinometer. The reason why two kinds of actinometers are used to determine the UV fluence is that actinometry is not a standard method for UV fluence determination, so two-times UV fluence determination can make the result become more acceptable. 28

38 3.3.1 Iodide-iodate actinometer The iodide-iodate actinometer solution consists of 0.6 M KI and 0.1 M KIO 3 in 0.01 M Na 2 B 4 O 7 buffered solution. It has been developed for many years and becomes a popular actinometer (Rahn, 1997, 2003, 2006, 2013). When this kind of actinometer solution is exposed to the UV light, following chemical reaction happens: 8 I + IO H 2 O + h 3 I OH Equation 3.1 The UV fluence is proportional to the amounts of chemical reactions happened. In the products of this reaction, the concentration of I 3 could be determined spectrophotometrically. Its concentration is proportional to the change of absorbance of actinometer solution at 352 nm. (1) Preparation of iodide-iodate actinometer solution The preparation of actinometer solution is the critical procedure in actinometry method. Because I - - could be easily oxidized by IO 3 under low ph condition, then lead to the formation of I 2. This is the reason why the 0.01 M Na 2 B 4 O 7 buffered solution has been used to dissolve the KIO 3. Based on the experience obtained from this experiment, following procedure was used to prepare the actinometer solution. 1. Dissolve g Na 2 B 4 O 7 in 60 ml ultrapure water in a beaker. Use a stirring bar to promote the dissolution until it is totally dissolved. This solution should have a ph of around Dissolve 2.14 g KIO 3 in the solution obtained from the first step. The stirring bar also could be used to promote the dissolution. 3. Dissolve 9.96 g KI in the solution of KIO 3 and Na 2 B 4 O Transfer the actinometer solution into a 100 ml volumetric flask, and add ultrapure water into the solution until the total volume is 100 ml. The order of adding chemicals into the water is very critical in this experiment. Chemicals should not be added into the ultrapure water together as they may not dissolve in the water quickly. If the KIO 3 or Na 2 B 4 O 7 cannot dissolve quickly, the ph of solution will be not 9.2 and the IO 3 - may oxidize I -, eventually lead to the formation of I 2. This phenomenon has been observed for several times in this experiment. I 2 could cause interference to final result as they can absorb the UV light (352 nm) strongly. Another important thing that must be kept in mind is that there is a slow thermal oxidation happen in the actinometer solution, so it cannot be kept for a long time. The recommended storage time for actinometer solution is 4 hours. The actinometer solution should be made up freshly each time before experiment. (2) Procedure for UV fluence determination by iodide-iodate actinometer Actinometer solution (0.6 M KI/ 0.01 M KIO 3 ) is quite sensitive to the UV light under 330 nm. Although there is no well-defined wavelength above which absolutely no absorption takes place, it is assumed that light above 330 nm will not contribute the formation of I 3 -. Hence, this experiment could be conducted with the presence of room light. However, it is also suggested that one must avoid doing this experiment under sunlight or fluorescent 29

39 lamps without a plastic cover over the fixture to avoid exposure to light with wavelength less than 330 nm (Rahn, 2013). In order to reduce the impact of light with wavelength longer than 330 nm, all glassware used in this experiment were covered by the aluminium foil. Moreover, the UV exposure experiment was conducted under the yellow light condition. Yellow light has wavelength ranging from 577 nm to 597 nm. Its possibility to cause photochemical reaction in the actinometer solution is quite small. So, the impacts from the room light have been reduced as much as possible. Figure 3.5 Procedures for UV fluence determination by iodide-iodate actinometer The procedure for UV fluence determination by iodide-iodate actinometer was schematically described in Figure 3.5 and it could be described as follows: 1. Measure the absorbance of fresh actinometer solution in a 1.0 cm pathlength quartz cell at 300 nm and 352 nm. These values should be around 0.58 and 0.02, respectively. Call the later value A 352 (blank). 2. Add 5 ml actinometer solution and a small stirring bar into the tube and turn on the magnetic stirrer. 3. Turn on the UV LEDs and let the actinometer solution be irradiated by the UV light. Note the exposure time as t. 4. Measure the absorbance of irradiated actinometer solution at 352 nm. Note this value as A 352 (sample). 5. Calculate the UV fluence by following formula: Equation 3.2 Where V is the volume of actinometer solution (L), U is the photon energy (J/mole), ε is the adsorption coefficient (M -1 cm -1 ), is the quantum yield of (mole/einstein), S is the area exposed to the UV light (cm -2 ). The value of each parameter in this equation is shown in Table 3.1. The photon energy is energy contained in 1 mole of photon, and it can be calculated by following formula: Where h is the Planck constant ( J s), c is the speed of light ( m/s ), N a is the Avogadro number ( mole -1 ), and λ is the wavelength of UV light ( m). So, the photon energy is: 30

40 = J/einstein The quantum yield of has been updated for many times in last several decades due to the development of technology (Bolton et al., 2011; Goldstein et al., 2008; Rahn et al., 2003). The value of this parameter at 282 nm still has not been reported by any literatures yet. The reason is that the actinometry has just been applied to measure the UV fluence from LP UV lamp or calibrate the radiometer. As it is known, LP UV lamp is a kind of monochromatic lamp, and its wavelength is nm. So, most of the data published in literatures are about the quantum yield at nm. But the quantum yields at 280 nm and 284 nm have already been determined. The value at 280 nm and 284 nm are 0.37±0.01 and 0.30, respectively (Goldstein et al., 2008; Rahn et al., 2003). So, an average value of them can be employed as the quantum yield at 282 nm, which is shown in table 3.1. Table 3.1 Value for Each Parameter in equation 3.2 Parameter Value Unit References Volume of sample (V) L - Area (S) 3.14 cm 2 - Adsorption coefficient (ε) 27,636 M -1 cm -1 Bolton et al., 2011 Photon energy (U) 424, Joul/einstein - quantum yield (Φ) * mole /einstein Goldstein et al., 2008; Rahn et al., 2003 Reflection factor Bolton et al., 2011 * Note: the average value of 0.30 and Using these values to replace the symbols in equation 3.2 and do calculations to simplify the equation 3.2, then it becomes: UV fluence = [A 352 (sample) A 352 (blank) ] mj/cm 2 Equation 3.3 Following is an example of calculation for application of this formula (Bolton et al., 2011): 5.0 ml actinometer solution in a 10 ml beaker (cross-sectional area 3.80 cm 2 ), the absorbance at 352 nm (in 1 cm 1 cm quartz cuvette) before irradiation is found to be call this A 352 (blank). After irradiation for 3.0 min, the absorbance at 352 nm is call this value A 352 (sample). The following calculations illustrate how the photon irradiance and the irradiance are calculated: Concentration of triiodide ion [I 3 - ]=[ A 352 (sample)- A 352 (blank)]/adsorption coefficient =[ ]/27,636 = M Moles of I 3 - =[I 3 - ] V(L) = M L = moles The quantum yield this reaction at 254 nm was 0.6 mole einstein -1. Einsteins (moles of photons) = moles of I 3 - / 31

41 = mole/(0.6 mole einstein -1 ) = einsteins photon irradiance (E p ) = einsteins/(area time) = 1, / (3.80 cm s) = 2, einstein s -1 cm -2 Irradiance (E ) = E p photon energy at nm (U ) The irradiance must be corrected for the 2.5% that is reflected from the water surface, so the incident irradiance on the water surface is: E (corrected) = E (uncorrected)/0.975 = ( ,576)/0.975 W cm -2 = W cm -2 = mw cm Ferrioxalate actinometer Ferrioxalate actinometer is another actinometer that has already been widely accepted by researchers. When compared with iodide-iodate actinometer, this actinometer has advantages and disadvantages. The advantage is that the quantum yield of ferrioxalate actinometer is well defined. Unlike the iodide-iodate actinometer whose quantum yield is different at different wavelength, the quantum yield of ferrioxalate is a constant when the wavelength ranges from nm and nm (Goldstein et al., 2008). So, it is unnecessary to exactly determine the quantum yields at each wavelength, while this is really a must when using the iodide-iodate actinometer. The disadvantage is that this kind of actinometer is very sensitive to the room light, so all operations need to be done under red light conditions. (1) Preparation of actinometer solution Before preparing the actinometer solution, following solution should be prepared (Bolton et al., 2011). a. Ferric sulfate solution (0.2 mole/l) in 1 mole/l H 2 SO 4. The ferric sulfate (Fe 2 (SO 4 ) 3 ) should be added into the H 2 SO 4 solution, not the water, because the Fe 3+ hydrolysis at ph>2.3 and the slat does not dissolve. As most of the ferric salt are a bit impure, so the concentration of Fe 3+ should be determined by phenanthroline again after making this kind of solution. b. Potassium oxalate solution with a concentration of 1.2 mole/l. Use an electronic balance to weigh out g K 2 C 2 O 4 (analytical grade) and dissolve it into a 250 ml volumetric flask. c. Sodium acetate buffer solution at ph g of CH 3 COONa 3H 2 O was weighted out and transferred into a 250 ml volumetric flask. Use around 100 ml distilled water to dissolve it and then add 2.5 ml concentrated sulfuric acid (96-98%) into it. d. 1,10-phenanthroline solution with a concentration of 0.2%. Dissolve 0.5 g 1,10- phenanthroline into 250 ml water. Because the 1,10-phenanthroline is quite hard to dissolve, so a magnetic stirrer was used to promote its dissolution. This kind of solution must be kept in the dark place to prevent the photodecomposition of phenanthroline. e. Hydroxylamine hydrochloride (NH 2 OH) solution with a concentration of 1 mole/l. Dissolve 6.95 g hydroxylamine hydrochloride in 100 ml distilled water. This 32

42 solution cannot be kept for a long time and should be prepared freshly before experiment each time. f. Sulfuric acid solution (1 mole/l). Add 14 ml concentrated sulfuric acid into a 250 ml volumetric flask, in which 125 ml distilled water was added in advance. As mentioned before, the ferric salt is always a bit impure, so the concentration of Fe 3+ in the ferric sulfate solution should be determined again. The phenanthroline method has been employed to determine the concentration of ferric ion. The procedure could be described as follows: 1. Exactly 0.3 ml of Fe 2 (SO 4 ) 3 solution was transferred to a 100 ml volumetric flask. Distilled water was added until the mark. Mix them thoroughly. 2. Take 0.8 ml of this solution into a 10 ml volumetric flask. And 2 ml distilled water and 1 ml NH 2 OH solution was added. 4Fe NH 2 OH 4Fe 2+ + N 2 O + H 2 O + 4 H + Equation After 2 min, 2 ml sodium acetate solution and 2 ml phenanthroline solution were added. Then the solution was kept in the dark for 40 min. The ferrous ion can complex with the phenanthroline. Fe C 12 H 8 N 2 [Fe 3 C 12 H 8 N 2 ] 2+ (red brown) Equation Measure the absorbance of this solution at 510 nm. Note as ABS (sample). 5. Repeat above procedure without adding 1 ml NH 2 OH solution. The absorbance was noted as ABS (blank). 6. Calculate the concentration of ferric ion by following formula: [Fe 3+ ]= = [ ] Equation 3.6 Among which, L/mole cm -1 is the adsorption coefficient of complex of ferricphenanthroline, other parts in this formula means the dilution in this procedure. After the concentration of ferric ion was accurately determined, the ferric sulfate solution was ready for being used to make ferrioxalate actinometer. The actinometer solution is quite sensitive to visible and UV light, so all the operations were conducted under the red light condition. A red lamp for developing the film was applied for illumination during experiment. Following is the procedure for making actinometer solution ml of potassium oxalate solution (1.2 mole/l) and 35 ml of sulfuric acid solution (1 mole/l) were added into a 1 L volumetric flask. 2. 6/[Fe 3+ ] ml ferric sulfate solution (0.2 mole/l) was added into this volumetric flask and mixed them thoroughly. [Fe 3+ ] is the concentration of ferric ion determined before. Then the ferrioxalate actinometer solution was ready for use. The final compositions of the actinometer solution could be described as 6 mm potassium ferrioxalate in 0.1 N sulfuric acid (call this FeOx solution). The sulfuric acid is used here in order to maintain the ph of solution at below 2. So the ferric sulfate would not hydrolyze. 33

43 When this actinometer solution was exposed to the UV light, the Fe 3+ will be reduced to Fe 2+, which could be described as following: Equation 3.7 The number of Fe 2+ formed during this reaction is proportional to the UV fluence exerted to the actinometer solution. At nm, the quantum yield Φ ) is 1.39 ± 0.02 mole einstein -1 (Goldstein et al., 2008). (2) Procedure for UV exposure test The procedure for UV fluence determination by ferrioxalate actinometry is described as below: a. Adding 5 ml FeOx Solution into the reactor and turn on the magnetic stirrer, so the actinometer solution could be mixed thoroughly; b. Turn on the UV LEDs and start the UV irradiation; c. After exposure for a period of time (t), turn off the UV LEDs. d. Take 1 ml sample for measuring the concentration of Fe 2+. e. Use relative formula to calculate the UV fluence This procedure is shown in Figure 3.6. The procedure is almost the same with that of KI/KIO 3 actinometer and just use the ferrioxalate actinometer to instead the KI/KIO 3 solution. 1. Addition of Fe x Solution ( ml) 2. UV exposure. Take sample (1 ml) at time t.. Determine the concentration of Fe 2 through colorimetry. Figure 3.6 The procedure for UV fluence determination by ferrioxalate actinometer The method used to measure the concentration of Fe 2+ method could be described as follows: is phenanthroline assay. This 1. Prepare a set of labeled 10 ml volumetric flask before exposure test because the UV fluence determination generally consists of several runs, and in each volumetric flask, 2 ml sodium acetate buffer solution (ph=4.5) and 2 ml 1,10-phenanthroline solution were added; 2. 1 ml FeOx solution is taken from the reactor after UV exposure test and added into the prepared volumetric flask; 3. Add distilled water to the 10 ml mark and put the flask in a dark place for 40 min to ensure that the Fe 2+ can complex with 1,10-phenanthroline completely. 4. Transfer the solution to a quartz cuvette and measure the absorbance at 510 nm. Note this value as ABS (510) (sample). 5. Take 1 ml FeOx solution without UV exposure, and repeat above procedure. Measure the absorbance at 510 nm and note this value as ABS (510) (blank). 34

44 For one single run of this experiment, the Fe 2+ generated can be calculated from following formula: Equation 3.8 Where V is the total volume of FeOx solution irradiated, ml, V 1 is the volume withdrawn from the irradiated solution, ml, 11,110 M -1 cm -1 is the molar absorption coefficient of the Fe-1,10-phenanthroine complex, M -1 cm -1, 10 is the volume of the volumetric flask, ml, and 1000 is the transformation between liter and milliliter. The UV fluence could is proportional to the numbers of Fe 2+ formed in this reaction. Following formula could be used to calculate the UV fluence: Equation 3.9 Where is the quantum yield, mole einstein -1 ; U is the photon energy, J/einstein; Area is the area of solution surface irradiated by the UV light, cm 2 ; Reflection factor is 2.5%, which is caused by the reflection of water surface. The values of each parameter used in equation 3.8 and 3.9 are shown in table 3.2. Table 3.2 Value Adopted for Each Parameter in Equation 3.8 and 3.9 Parameter value unit References V L - V L - U 424, Joul/einstein - Area 3.14 cm mole/einstein Bolton et al., 2011 Reflection factor Using these values to replace the symbols in equation 3.8 and 3.9 to simplify the calculation, then the formula for UV fluence calculation becomes: UV fluence = [A 510 (sample) A 510 (blank)] mj/cm 2 Equation Disinfection Test with Synthetic Wastewater Preparation of synthetic wastewater The method to prepare the synthetic turbid wastewater is described in Figure 3.7. Synthetic wastewater was prepared through adding montmorillonite and E. coli into the RO water. E. coli was added as the challenge organism. Before UV exposure, synthetic wastewater was mixed thoroughly by vortex mixture. Additionally, all operations for synthetizing wastewater were done aseptically in order to avoid the contamination. 35

45 Figure 3.7 Procedure for making synthetic wastewater Different kinds of turbid wastewater were obtained by controlling the amount of montmorillonite added into the RO water. Figure 3.8 illustrates the determination of amounts of montmorillonite added into the RO water for making synthetic wastewater. Figure 3.8 Determination of montmorillonite added into the wastewater Preparation of E. coli for synthetic wastewater The E. coli (ATCC 29214) was obtained from Thailand Institute of Scientific and Technological Research (TISTR). After getting pure culture of E. coli, it was inoculated in LB broth to regain the activity. After cultivation for 24 hours, 1 ml LB solution of this LB broth solution was inoculated in the new LB broth. After cultivation for 5 hours, it was harvested to make the glycerol stock of E. coli. The E. coli was stored in the glycerol solution ( 0%) so that it won t be killed by low temperature. Here, the glycerol stock of E. coli was obtained through mixing 50% E. coli solution and 50% glycerol solution. A vortex mixer has been applied for mixing step so that the E. coli solution and LB broth can be mixed thoroughly. After that, the glycerol stock was distributed into the 1.5mL cyto tubes, and then stored in the freezer. The temperature of freezer was kept at -20 and the storage time is up to 1 month. This procedure was schematically described in Figure

46 Figure 3.9 The process for making glycerol stock from E. coli strain Before using E. coli for disinfection test, 1 ml E. coli stock solution was inoculated in the LB broth. After cultivation in the new LB broth for a certain period, the E. coli at approximately mid-exponential growth phrase was harvested for the disinfection test. 10 ml of the E. coli suspension was centrifuged at 5000 rpm for 20 minutes and the supernatant was aseptically drawn off. Then, the remaining E. coli was re-suspended with the phosphate buffered saline (PBS, 0.01M) and centrifuged again. As described above, the washing procedure was repeated twice to remove any nutrient medium. After that, the E. coli solution in PBS solution was read for use in the disinfection test Enumeration of E. coli Pour plate technique has been applied to enumerate E. coli in synthetic wastewater before and after UV exposure. One kind of selective medium-chromocult Coliform agar (Merck, Germeny)-has been used to cultivate the E. coli. Chromocult Coliform agar can promote the growth of total coliform and E. coli, while suppress the growth of other microorganisms. On this kind of agar, E. coli appear as dark violet colony (Figure 3.10), and coliform bacteria appear as pink to red colonies. Other bacteria appear as colorless or green colony. The benefits of using this kind of agar to count the density of E. coli is that it can help to avoid the contamination in experiment as most of the gram positive bacteria cannot grow on this kind of agar. When combined with pour plate technique, the density determination of E. coli in synthetic wastewater can be quite accurate. As only E. coli was added into the synthetic wastewater, and Chromocult coliform agar is a selective medium for E. coli, so this kind of modified method is quite good for synthetic wastewater. Figure 3.10 The Chromocult Colifrom agar and the colony 37

47 The pour plate technique consists of two parts: series dilution of sample (Figure 3.11) and the plating of dilutions (Figure 3.12). All the operations should be done aseptically. Serial dilution is the stepwise dilution of E. coli solution and the dilution factor for each step is The procedure could be described as following steps: 1. Sterilizing the glassware and other equipments, and put them into the laminar hood. 2. Clean hands and laminar hood with 70% alcohol and add 9 ml of sterile PBS solution into each tube using 10 ml transfer pipette. 3. Labeling all tubes with 10-1 to 10-5 indicating the dilution factor. 4. Transfer 1 ml from the culture sample to the first tube (10-1 ) by 1 ml automicropipette and mix them gently by vortex. 5. Take 1 ml of the diluted E. coli solution from the first tube and add them to the next tube (10-2 ). Thereafter mix them gently. 6. Repeat the same procedure for the left four tubes (10-3 to 10-6 ). Then the E. coli solutions with dilution factors 10-1, 10-2, 10-3, 10-4, 10-5, 10-6 are formed. Figure 3.11 Series dilution After finishing the serial dilution, each solution was inoculated in the Chromocult Coliform agar. The plating of solution could be described as following steps: 1. Label all necessary information on the plate, such as microbes name, date, user s name and dilution factor. 2. Mix the last solution (10-6 ) by using vortex mixer. Take 1 ml solution from it by using a 1 ml micropipette and drop them slowly into the plate. Take other two samples and inoculate them in other two plates, so each dilution was cultivated in triplicates. 3. Repeat the same procedure for all other dilution samples and inoculate them into the corresponding plates. 4. Pouring Chromocult Coliform agar into the plates and shake them gently to mix the E. coli and agar. 5. Incubate them at 37 for 24 h. Then count the number of colonies in each plate and calculate the numbers of E. coli per 1 ml. For example, the numbers of colonies of one dilution are a 1, a 2, a 3. So the average number of colonies is: Equation

48 If the dilution factor is α, so the concentration (CFU/mL) of E. coli in this dilution is: Equation 3.12 It must be noted that this method is just suitable for colonies per petri dish. If the number of colonies exceeds this range, the solution should be diluted (more than 300 CFU/dish) or concentrated (less than 30 CFU/dish). Figure 3.12 Plating of dilutions Procedure for disinfection test with synthetic wastewater The procedure for conducting disinfection test with synthetic wastewater is illustrated in Figure Firstly, the synthetic wastewater was added into the reactor. Then UV LEDs were turned on. After exposure to the UV LEDs for a period of time, the UV LEDs were turned off and 1 ml water sample will be taken for determining the concentration of E. coli. Figure 3.13 Procedure for disinfection test with synthetic wastewater Usually, the turbidity of secondary effluent is quite low, sometimes it is even could be less than 20 NTU. Turbidity of primary effluent may depend on the quality of raw wastewater and the performance of primary sedimentation tank NTU could be a rational range of primary effluent s turbidity. For raw wastewater, the turbidity may vary significantly. Sometimes, it can be as high as 1000 NTU, while it can be also lower than 100 NTU. Based on such information, the turbidity chose for synthetic wastewater are 27, 70, 113, and 156 NTU, which have also been adjusted based on the disinfection performance of UV LEDs in this study. For each kind of turbid wastewater, the applied UV fluence was different in order to get the UV fluence-inactivation response curve in a suitable range. For synthetic wastewater of 27 NTU, the required UV fluence for getting a UV fluenceresponse curve should be much lower than that of wastewater of 156 NTU. The structure of this experiment was shown as below: 39

49 Figure 3.14 Research framework for disinfection test with synthetic wastewater 3.5 Disinfection Test with Real Wastewater Wastewater sampling As the quality of real wastewater varies seasonally and geographically, one kind of synthetic wastewater would not be able to represent all kinds of wastewater. In real case, the quality of wastewater might be very different from synthetic wastewater used here, so the disinfection performance of UV LEDs may be also quite different. In order to give out a reference for real case, the disinfection test has also been conducted with several kinds of real wastewater samples. The wastewater samples were taken from the AIT wastewater treatment plant, whose treatment process was shown in Figure A sequencing batch reactor (SBR) has been employed to treat the wastewater generated by the residents in AIT campus. Raw wastewater flows into the collection tank first. Then it was lifted to the equalization tank by a pump. The equalization tank can homogenize the quality of wastewater flowing into the aeration tank so that the shocking load won t happen in the aeration tank. The aeration tank was operated in aeration-settling-discharge-filling-aeration mode. During aeration, most of organic matters in the wastewater were removed by the activated sludge. When aeration is stopped, the activated sludge would settle down and the anaerobic environment in the tank is created at the same time, so the denitrification may happen, thus nitrogen in the wastewater can also be removed. At last, the wastewater was discharged. Figure 3.15 AIT wastewater treatment process 40

50 The aeration system in equalization tank and SBR tank works at the same time. So during the settling stage of SBR tank, the aeration in equalization tank also stopped and the particles in the wastewater can settle down. The wastewater sample taken from the surface of equalization tank is much less turbid than the wastewater sample taken from the inlet of equalization tank. In real case, the wastewater disinfected by UV irradiation should be as less turbid as possible. As it is known, sedimentation is a cheap process to remove part of the particles and pathogens from the wastewater. Properly designed and operated sedimentation tank should remove 50-70% of the suspended particles and 25-40% of the BOD (Metcalf & Eddy, 2003). So, it is better to disinfect the turbid wastewater after sedimentation because such pretreatment can improve the disinfection performance of UV LEDs. However, primary sedimentation tank was eliminated in AIT wastewater treatment plant in order to simplify the treatment process. But, as mentioned above, settling process also happened in the equalization tank and the wastewater at the surface level is less turbid than the raw wastewater. To some extent, the wastewater at the surface of equalization tank has the same property of wastewater after settling. So, it is reasonable to use this kind wastewater to instead the wastewater after sedimentation. Granted, the turbidity might be different from the wastewater sample after real sedimentation as the time for settling is different. But it will not matter because the turbidity of primary effluent from the real wastewater treatment plants that using conventional activated sludge process may still vary. In real case, the turbidity of primary effluent is dependent on both detention time and initial concentration of total solids in raw wastewater. Additionally, different turbid wastewater was obtained through mixing raw wastewater and wastewater from the surface of equalization tank in different ratio. And all of them have been used to conduct the disinfection test. The result obtained through disinfection test with different turbid wastewater can shows the impact of turbidity of wastewater on disinfection performance of UV LEDs better. The method of taking water sample from the primary effluent is based on the standard method of APHA. Water sample was stored in ice box and take to the lab within half hour. The disinfection test and cultivation of E. coli should be done as soon as possible because the long duration of storage may change the microbial properties of wastewater Selection of challenge organism and its enumeration Both total coliform and E. coli in the wastewater has been selected as the challenge organism. Both of them have been adopted as the microbial parameter for wastewater discharge standard or reuse standard. Unlike the synthetic wastewater, in which only E. coli exists, so that Chromocult Coliform agar and pour plate technique can be used to count the density of E. coli and the obtained results are also quite reliable. In real wastewater, millions of microorganism exists. Although Chromocult Coliform agar with additional antibiotics (cufsulodin) has been reported to be able to detect the fecal pollution very well (Byamukama et al., 2000), the addition of antibiotics has not been approved by US EPA (Olstadt et al., 2007). Furthermore, no articles related to using Chromocult Coliform agar for wastewater 41

51 examination have been published until now. So, it is better to use the standard method to count the density of total coliform and E. coli in wastewater. In this study, the method for counting density of total coliform is the most probable number method (MPN method), which is adapted from the standard methods (APHA, 2005). In standard MPN method, the number of tubes for each set can be 3, 5, and 10. In this study, 5 tubes were selected in order to obtain accurate results and control the workload. This method is schematically described in Figure 16. Figure 3.16 MPN method In MPN method, 1 ml water sample was inoculated into the tube with 10 ml lactose broth. After incubation at 37 for 24 hours, the tube with both turbidity and gas production was reported as positive result. Otherwise, the result was recorded as negative result. All tubes with negative results have been put into the incubator for another 24 hours to ensure that they are negative results. The detection of E. coli was done after getting the result for total coliform detection. The enriched solution of total coliform was inoculated onto the Eosin Methylene Blue (EMB) agar. In this step, only the solution showing positive result was inoculated onto the EMB agar. EMB agar is a kind of selective medium and the color of E. coli colony appears as metallic green sheen. The most probable number of total coliform and E. coli was calculated by following formula: Equation 3.13 The value of table MPN can be read from the table in standard method. 42

52 3.5.3 Procedure for disinfection test The procedure for conducting the disinfection test with primary effluent is shown in Figure The procedure is almost the same with the disinfection test with synthetic wastewater and the only difference is to use the real wastewater to instead the synthetic wastewater. Figure 3.17 The procedure for conducting disinfection test with real wastewater Different UV fluence has been applied to disinfect each kind of wastewater sample. As illustrated in the last section, UV fluence required to obtain the UV fluence-inactivation response curve is different. So, the time of UV exposure for each wastewater sample is also very different. At the same time, the turbidity of each kind of wastewater was noted. The research framework was shown as below: Figure 3.18 Research framework for disinfection test with real wastewater 3.6 Summary All the methods for conducting this research were summarized in Table 3.3. Table 3.3 Summary of Methods Parameters Units Methods Reference E. coli CFU/mL Pour plate technique (Merck, Germeny) MPN/100 ml MPN method (APHA, 2005) Total Coliform MPN/100 ml MPN method (APHA, 2005) Turbidity NTU Nephelometric (APHA, 2005) Absorbance - Spectrophotometer (APHA, 2005) 43

53 Relative emission power Chapter 4 Results and Discussions This chapter presents all the results for this study and the corresponding discussions to the results. Firstly, the characterization of UV LEDs was introduced. This was followed by a section to illustrate the results from two kinds of actinometers, which were used to determine the UV fluence in the reactor. A comparison has been done between the results of two kinds of actinometers and the discrepancy between them was also discussed. At last, the results for UV disinfection test were shown, including disinfection test results for both real wastewater and synthetic wastewater. 4.1 Characterization of UV LEDs Before the fluence determination experiment and disinfection test, spectroradiometer (Ocean Optics USB2000) has been used to measure the wavelength of UV LEDs. The result is shown in Figure 4.1. From this figure, it can be found that the peak emission was not at 280 nm exactly, the measured wavelength is more close to 282 nm. As the properties of 280 nm and 282 nm UV light are almost the same, so it won t cause a big impact to the final results. 1.2 Figure 4.1 The emission spectrum of UV LEDs 4.2 UV Fluence Determination by Actinometry Method 1. Iodide-iodate actinometer wavelength (nm) The procedure for UV fluence determination was shown in Figure 4.2. Generally, a full UV fluence determination process consists of several runs of this procedure. In each run, 5 ml iodide-iodate actinometer solution was added into the reactor. The absorbance of actinometer solution before and after exposure was measured by spectrophotometer at 352 nm. As illustrated in last chapter, the equation for calculating UV fluence can be expressed as follows: UV fluence = [A 352 (sample) A 352 (blank) ] mj/cm 2 Equation

54 Figure 4.2 The procedure for UV fluence determination In this study, the UV fluence determination has been run for 18 times. The results were shown in table 4.1. The absorbance of actinometer solution before UV exposure was Table 4.1 Absorbance of Iodide-iodate Actinometer Solution at Different Exposure Time Exposure time (s) ABS UV fluence (mj/cm 2 ) Exposure time (s) ABS UV fluence (mj/cm 2 ) After doing linear regression between the UV fluence and exposure time with the data obtained after s, following result was obtained (Figure 4.3). The UV fluence was proportional to the exposure time, and the R 2 was The relationship between UV fluence and exposure time was: UV fluence = exposure time Equation 4.2 The reason why the first two data were not used here is that they are abnormal result because the ABS value did not change after UV exposure. This may be caused by unexpected chemical reactions. The triiodide formed during UV exposure reacts with other chemicals in the actinometer solution and then no absorbance change was observed after UV exposure. Figure 4.3 UV fluence determined by iodide-iodate vs. exposure time 45

55 In the reactor, the emission power of UV LEDs is a constant and the position of UV LEDs in the reactor is fixed, so UV fluence in the reactor should be proportional to exposure time. When the exposure time is 0 s, the UV fluence is 0 mj/cm 2. So the line in Figure 4.3 should be set to have an intercept of 0. Then equation 4.2 becomes: UV fluence (mj/cm 2 ) = mw/cm 2 exposure time (s) Equation 4.3 The equation shows that the irradiance the actinometer received from UV LEDs is mw/cm 2. The area exposed to the UV LEDs is 3.14 cm 2, so the total irradiance is 1.34 mw. When compared with 10.8 mw, which is the emission power shown in the instruction of UV LEDs, this is much lower. This attenuation is attributable to several reasons. Firstly, there is attenuation during the transportation of UV light in the air. Secondly, not all UV lights from the UV LEDs irradiated into the actinometer solution, only a part of them were absorbed by the actinometer solution. Thirdly, a small part of the UV lights were reflected by the actinometer solution. 2. Ferrioxalate actinometer Just as illustrated in the methodology part, the procedure for this actinometry is the same with that of iodide-iodate actinometer. The equation for the calculation of UV fluence was shown as below: UV fluence = [A 510 (sample) A 510 (blank)] mj/cm 2 Equation 4.4 In this experiment, ferrioxalate actinometer solution has been used to replace the iodideiodate actinometer solution, and it has been repeated for 12 times. In each run, the ferrous ion concentration was analyzed by phenanthroline assay before and after UV exposure, which was done through the spectrophotometric method. The absorbance measured in this experiment was shown in table 4.2. Table 4.2 Absorbance of Ferrioxalate Actinometer Solution at Different Exposure Time Exposure Time (s) Absorbance UV fluence (mj/cm 2 ) Exposure Time (s) Absorbance UV fluence (mj/cm 2 ) The results shown in above table were used to do linear regression. The obtained result was shown in Figure 4.4. From this figure, it can be found that the relationship between UV fluence and exposure time could be expressed as following: UV fluence (mj/cm 2 ) = exposure time Equation

56 Before using this equation, the intercept of the trendline should be set to zero. So the equation turns to: UV fluence (mj/cm 2 ) = exposure time Equation 4.6 The irradiance of UV LEDs was found to be mw/cm 2. When it is multiplied by the exposure area (3.14 cm 2 ), the irradiance becomes 1.2 mw. This value is even lower than the value measured by iodide-iodate actinometry. The reasons that should be responsible for such kind of attenuation were already discussed in the iodide-iodate actinometry part. Figure 4.4 UV fluence determined by ferrioxalate vs. exposure time 3. Comparison between the results from two actinometers and discussion The irradiance of UV LEDs measured by iodide-iodate actinometer was mw/cm 2, while this value was mw/cm 2 in ferrioxalate actinometry. Based on the experience from this study, two factors should be accountable for this discrepancy. Firstly, the quantum yield used in iodide-iodate was not very accurate. Currently, quantum yield of iodide-iodate is not available at 282 nm. The value applied here is the average value of quantum yields at 280 nm and 284 nm, which is already illustrated in the chapter 3. So, this may be also a resource of the discrepancy between them. Secondly, iodide-iodate actinometry and ferrioxalate actinometry are two different actinometry system. Actinometry methods are quite sensitive to procedural variation, so it is normal to get two results that are slightly different. Based on above discussion, following conclusion can be made: 1. The irradiance of the UV LEDs might be mw/cm 2 or mw/cm UV fluence measured by actinometry method was not constant. The result is very sensitive to the procedural variation and the results obtained through different actinometry are also different. 47

57 Turbidity (NTU) 4.3 Disinfection Test with Synthetic Wastewater Synthetic wastewater was synthetized through adding montmorillonite into the water that purified through reverse osmosis. Different concentration of montmorillonite can result in different kinds of turbid wastewater. The relationship between the turbidity of synthetic wastewater and the concentration of montmorillonite was shown in Figure y = x R² = concentration of montmorillonite (mg/l) Figure 4.5 Turbidity of synthetic wastewater vs. concentration of montmorillonite From this figure, it can be concluded that the turbidity of synthetic wastewater was proportional to the concentration of montmorillonite (C m ) in the water, and the relationship between them can be expressed as following: Turbidity (NTU) = C m Equation 4.7 In this study, the turbidity chose for synthetic wastewater was 0, 27, 70, 113, and 156 NTU and the concentration of montmorillonite was 0, 237, 475.5, 713.9, and 952 mg/l, respectively. Disinfection test has been conducted with each kind of synthetic wastewater and the result for disinfection test was shown in Figure 4.6. As explained in chapter 2, the UV inactivation kinetics can be expressed by following equation: log ( ) = k 1 F t Equation 4.8 Where log ( ) is the log-reduction of microorganism, C 0 is the concentration of challenge organism before UV exposure (CFU/mL), C t is the concentration of challenge organism after UV exposure (CFU/mL), F t is the UV fluence at time t, and k is a constant (cm 2 /mj). In this reactor, UV irradiance from UV LEDs was a constant, and the UV fluence is the product of irradiance and exposure time. So, equation 4.8 can be converted to: log ( ) = k 2 t Equation

58 Log-reduction Where k 2 is a constant (s -1 ), and t is the exposure time. Through doing linear regression with result of each kind of synthetic wastewater, the equation of inactivation kinetics under different turbid water condition was obtained and they are shown in table 4.3. In order to make it simple, log ( ) was expressed as L Exposure time (s) NTU 27 NTU 70 NTU 113 NTU 156 NTU Figure 4.6 Time-response curve of E. coli in different turbid wastewater Table 4.3 Inactivation Kinetics of UV Irradiation in Different Turbid Wastewater Turbidity (NTU) Equation k 2 (s -1 ) R 2 n 0 L = t L = t L = t L = t L = t * Note: R 2 is the coefficient in each linear regression; n is the number of data used for linear regression. A UV fluence of 40 mj/cm 2 is typically the minimum requirement for drinking water disinfection plants, whereas recommended design UV fluence can go up to 100 mj/cm 2 for reclaimed water systems, depending on the upstream treatment application (Kollu et al., 2012). In this reactor, the maximum UV irradiance measured by actinometry was mw/cm 2. The exposure time would be 233 s if a UV fluence of 100 mj/cm 2 was exerted to the synthetic wastewater solution. When E. coli in wastewater is exposed to the UV LEDs for such a long time, the expected log-reduction would be at least 4 in all kinds of turbid wastewater. The absolute value of k 2 in the equation shown in table 4.3 represent the sensitivity of E. coli to UV irradiation. When increased the turbidity, the sensitivity of E. coli to UV irradiation decreased significantly, which was shown in Figure 4.7. The absolute value of 49

59 Log-reduction Sensitivity of E. coli to UV (s -1 ) k 2 was 0.13 when the turbidity was 0 NTU, while it was just 0.02 when the turbidity was increased to 156 NTU, with a decrease of 85%. This means that E. coli became much less sensitive to UV irradiation when the turbidity of wastewater was increased. This can be due to several reasons: absorbance of particles to UV light, scattering effect of particles on UV light and habitation of E. coli inside the particles Turbidity (NTU) Figure 4.7 Sensitivity of E. coli to UV irradiation in different turbid synthetic wastewater In Figure 4.8, it can be found that, when the turbidity of wastewater was 70 NTU, the disinfection performance of UV LED was better that of 27 NTU. This is caused by error in this experiment. The R 2 was just 0.55 when the turbidity is 70 NTU, which can be found in table 4.3. It means that the log-reduction did not show a linear relationship with UV exposure time very well. However, it clearly shows that the disinfection performances of UV LED on E. coli in both two kinds of turbid wastewater are very close to each other Exposure time (s) NTU 27 NTU 70 NTU Figure 4.8 Time-response curve of E. coli in synthetic wastewater of 27 and 70 NTU Based on the information above, it can be concluded that E. coli in turbid wastewater can still be disinfected by UV LED effectively. But the sensitivity of E. coli to UV irradiation 50

60 may decrease greatly when increasing the turbidity of wastewater. However, a slight increase of turbidity of wastewater may not reduce the disinfection efficiency too much. Wastewater reclamation became more and more popular around the world due to the water scarcity. In developing countries, domestic wastewater is even used for agriculture irrigation directly, without any pretreatment, which increases the risk to human being significantly. Pretreatment of wastewater before reuse in strongly recommended. The typical abundance of E. coli in domestic wastewater ranges from 10 6 to 10 8 CFU/100 ml (Kadam et al., 2008; Molleda et al., 2008; Vrhovšek et al., 1996; Zhang et al., 2007). Based on the treatment performance, treated wastewater can be reused for many purposes, such as agricultural irrigation, recreation and even discharged to the environment directly. The requirements for wastewater that used in different purpose are quite different. In Canada, the concentration of fecal coliform of treated wastewater for unrestricted use must be less than 2.2 CFU/100 ml(zhang et al., 2007). To meet such strict standard, disinfecting wastewater directly may be not a good option. However, the standard for reusing wastewater for agriculture irrigation is not so strict. In Spanish, the microbial standard for reusing wastewater in agriculture irrigation is that E. coli concentration should not exceed 100 CFU/100 ml, which requires around 5 logreduction of E. coli in wastewater. In Figure 4.6, it can be found that, in all turbid wastewater samples (27, 70, 113, 156 NTU) except the turbid wastewater of 156 NTU, 5 log-reduction of E. coli is possible when UV exposure time is as long as 233 s. So, wastewater disinfected by UV irradiation might be used for agriculture irrigation. It should also be noted that composition of real wastewater is much more complex than the synthetic wastewater. This may lead to a reduction of disinfection efficiency of UV LEDs. Meanwhile, pathogens in real wastewater may be also more resistant to UV irradiation than E. coli used in synthetic wastewater. Thus, whether it is possible to use UV LEDs to disinfect the wastewater for reuse may depend on two aspects: one aspect is the quality of wastewater and another one is the standard requirement. To some extent, the result shown above proved that UV irradiation might be a possible way to disinfect the turbid wastewater. However, as other pathogens may be more resistant to UV irradiation than E. coli, so the disinfection efficiency of UV LEDs on UV-resistant microorganism should also be checked, especially the virus and protozoa. 4.4 Disinfection Test with Real Wastewater As mentioned in last section, the composition of real wastewater is much more complex than synthetic wastewater. The disinfection efficiency of UV LEDs on real wastewater should also be different from that of synthetic wastewater. So it is necessary to conduct the disinfection test with real wastewater. In this study, the wastewater sample from AIT wastewater treatment plant was used to conduct the disinfection test. Wastewater was sampled at different parts of treatment process at different time. The sampling site includes: surface of equalization tank (SEQ), inlet of equalization tank (IEQ), and collection tank (CT). More detailed information could be found from chapter 3. Properties of wastewater samples were described in table 4.4, including the turbidity, TS, TSS. 51

61 Table 4.4 Properties of Wastewater Samples No. of Sampling Turbidity TS TSS TDS Date sample site (NTU) (mg/l) (mg/l) (mg/l) 1 April 9 IEQ April 11 SEQ April 14 CT April 16 CT *Note: SEQ: surface of equalization tank CT: collection tank From this table, it can be found that turbidity of wastewater did not show a proportional relationship with total solids or total suspended solids in wastewater. This might be caused by different size distribution or composition of particles in the wastewater. Turbidity is a measure of the scatter of visible light ( nm). This means the particles with size ranges from 400 to 700 nm can contribute to the turbidity most, while particles with other size distribution may contribute less. So, different size distribution might be a reason for this phenomenon. Another reason might be the different composition of particles in wastewater. Generally, inorganic particles can contribute more to turbidity than organic particles. For example, 50 mg/l kaolinite gave a turbidity reading of about 80 NTU, while 50mg/L humic acid gave a turbidity reading only slightly greater than 3 NTU (Edzwald, 1987). The results of disinfection test with each kind of wastewater were shown in Figure 4.9 and Unlike the synthetic wastewater, the disinfection test can be repeated for many times and enough amount of data can be obtained to do very accurate analysis. Disinfection test with real wastewater needs to be done within 6 hours after sampling. Here, the disinfection test for each wastewater sample has just been conducted for 4 to 5 times in order to finish the experiments within 6 hours and control the workload. So, only 4 to 5 data for each trial were used in Figure 4.9 and 4.11 to show the trend of result. Figure 4.9 Time-response curve of total coliform in different turbid wastewater 52

62 Log-reduction From Figure 4.9, it can be found that the disinfection performances of UV LEDs on total coliform in wastewater were quite similar to each other even though the turbidity of each wastewater sample is different. When the exposure time was less than 100 s, the log-reduction of total coliform shows a good linear relationship with exposure time, regardless of the turbidity of wastewater. This could be further confirmed by Figure In Figure 4.10, all log-reduction of total coliform and the corresponding exposure time were used to do linear regression. The obtained equation was: Log-reduction = exposure time Equation 4.10 And the trendline also has a R 2 of It means that the log-reduction of E. coli shows a good linear relationship with exposure time. Actually, the size distribution of particles in the four kinds of wastewater should be different, because one wastewater sample was taken from the surface of EQ tank after sedimentation, while another sample was taken from the inlet of EQ tank, which was without sedimentation. But coliform bacteria in all samples showed the similar UV fluence-inactivation response curve. This result suggests that slightly change of particle size distribution may not impact the disinfection performance of UV LEDs significantly. Exposure time (s) y = -0.03x R² = 0.95 Figure 4.10 Time-response curve of total coliform within 100 s UV exposure The maximum log-reduction of total coliform is nearly 3 and it could not be increased further by prolonging the UV exposure time. This is caused by the attachment of coliform on particles, which is also called tailing in UV inactivation kinetics. The coliform bacteria in wastewater can be free or attached to particles. Free coliform bacteria could be inactivated by UV easily. The inactivation kinetics is also a first order reaction. However, coliform bacteria attached to the particles can be very resistant to UV irradiation as they are sheltered inside the particles. Hence, even the UV exposure time has been increased, the coliform bacteria can still be alive in the wastewater. This is the reason why logreduction did not increase further after reaching its maximum value at around 110 s. Special attentions should be paid to the total coliform in the wastewater of 57 NTU as it has reached the maximum log-reduction at s, with a value of This indicates that the tailing effect may not appear or will be retarded if the turbidity of wastewater is low enough. However, this may be also caused by the error of results because only one 53

63 Log-reduction data was obtained here. More investigations should be done to confirm whether it is caused by error or the tailing effect would disappear or be retarded in UV disinfection of less turbid wastewater Exposure time (s) NTU 130 NTU Linear (clean water) Figure 4.11 Time-response curve of E. coli in different turbid wastewater The time-response curve of E. coli in wastewater was shown in Figure Due to the statistically unreliable data, only two sets of data were presented. The result was a bit the same as what has been observed in the disinfection test with total coliform, E. coli in wastewater can be inactivated effectively at first, and then it became very resistant to UV light due to the tailing effect. The maximum log-reduction of E. coli was nearly 3 and it could not be increased further by prolonging the UV exposure time. As mentioned in last section, a 5 log-reduction of E. coli is generally required for wastewater reclamation. But the maximum log-reduction of both total coliform and E. coli achieved in this experiment was about 3. Hence, after disinfection with UV directly, the wastewater may still be unable to meet the required guidelines. However, if the bacteria attached to the particulates could be removed, the maximum log-reduction could be increased further and it is possible to meet the required guidelines. In practice, the particles in wastewater can be removed through sedimentation or other ways. Properly designed primary sedimentation tank can remove 50-70% suspended particles (Metcalf & Eddy, 2003). The turbidity of wastewater after primary sedimentation tank (primary effluent) can be as low as NTU (Ravazzini et al., 2005). Sometimes, even the turbidity of raw sewage can be as low as 51 NTU (Bukhari, 2008). Figure 4.9 shows that the log-reduction of total coliform can be as high as 4.12 when the turbidity of wastewater is 57 NTU. Thus, if a proper pretreatment is applied (e.g. sedimentation), wastewater may reduce the turbidity significantly and UV irradiation might be a possible way to disinfect the wastewater. 54

64 Chapter 5 Conclusions and Recommendations 5.1 Conclusions This study focused on the disinfection performance of UV LEDs on turbid wastewater. Both synthetic wastewater and real wastewater were used to conduct the disinfection test. In synthetic wastewater, only E. coli was selected as the target organism, while both total coliform and E. coli in real wastewater have been selected as the challenge organisms. The conclusions for this study are as follows: 1. The reactor built for this experiment has met all requirements for conducting disinfection test with both synthetic and real wastewater. 5 ml was a suitable wastewater sample size for conducting disinfection test with UV LEDs. 2. The irradiance of UV LEDs measured by iodide-iodate actinometer was mw/cm 2, while that value of ferrioxalate actinometer was mw/cm 2. It suggests that different actinometry system may lead to different results. 3. UV inactivation of E. coli in synthetic wastewater was a first order reaction and there is no tailing effect. The log-reduction of E. coli could be increased by prolonging the UV exposure time, regardless of the turbidity of synthetic wastewater. 4. With the increase of turbidity, E. coli in synthetic wastewater became less sensitive to UV irradiation. 5. Slightly changes for a given turbid wastewater did not change the UV disinfection performance in real wastewater. 6. Measurement of turbidity and suspended solids of real wastewater did not show any direct correlation. This could be mainly due to the variation of particle size distribution and its composition. As disinfection efficiency is more related to particle size distribution than turbidity, there is need to measure this parameter. 7. The UV inactivation kinetics of total coliform in real wastewater was not completely first order reaction. The tailing effect was observed due to the interaction of coliform bacteria with particles. A maximum of ~ 3 log-reductions for total coliform in real wastewater was observed, which could not be further increased by prolonging the UV exposure time. 8. The inactivation kinetics of E. coli in real wastewater was similar with that of total coliform. A maximum ~ 3 log-reduction was achieved and could not be increased further with an increase in the exposure time. 9. The amount of coliform bacteria attached to the particles is the limiting factor for further improvement of disinfection efficiency of UV LEDs. Higher disinfection efficiency could be achieved if the coliform bacteria attached to the particles can be removed by a proper way. 55

65 10. UV LEDs were not able to disinfect the real wastewater effectively at higher turbidities. However, if coupled with proper treatment technologies to reduce wastewater turbidity, UV LEDs should be able to disinfect real wastewater effectively and meet the required WHO guidelines. 11. In some emergent cases (e.g. after earthquake), UV LEDs can be still be used for disinfecting the wastewater before discharge so that the risks exposed to the environment by wastewater can be reduced significantly. 5.2 Recommendations for Further Study Based on the experience from this study, following recommendations are for the further study: 1. UV light have a range of nm wavelength. Thus assessing the effective UV wavelength is vital as their disinfection performance on turbid wastewater might vary greatly. Thus, disinfection performances of various UV lights should be checked, especially the UV LED with wavelength of 265 nm as its wavelength is closer to the germicidal peak (264 nm). 2. Scaling up the reactor from lab scale to bench scale and later full scale is necessary to establish both technological and economical feasibility of the this treatment system. 3. In practice, UV irradiation was applied to disinfect the water or wastewater in flowthrough condition, which is different from the batch reactor used in this study. Previous study has already proved that the disinfection performance of UV LEDs in flow-through conditions is quite different from that of batch mode. Moreover, the behavior of particles in flow condition is also different from the behavior in batch mode. Hence, a study on the disinfection performance of UV LEDs in flow-through reactor is deemed necessary for real world application. 4. Previous studies on conventional UV lamps shown that their disinfection efficiencies on different microorganisms were different. UV LEDs may have the same properties. Hence, other kinds of microorganisms (e.g. protozoa and virus) should also be used to conduct the disinfection test with UV LEDs. 5. Coupling low cost technologies like woven fiber membranes to remove turbidity and UV LEDs to reduce pathogen inactivation should be conducted. As UV LEDs cannot be used as a standalone technology for wastewater treatment. To increase the overall treatment efficiency coupling technologies becomes predominant for economical use. 56

66 References Autin, O., Romelot, C., Rust, L., Hart, J., Jarvis, P., MacAdam, J., Parsons, S. A., and Jefferson, B. (2013). Evaluation of a UV-light emitting diodes unit for the removal of micropollutants in water for low energy advanced oxidation processes. Chemosphere, 92(6), Baum, R., Luh, J., and Bartram, J. (2013). Sanitation: A Global Estimate of Sewerage Connections without Treatment and the Resulting Impact on MDG Progress. Environmental Science & Technology, 47(4), Bitton, G., Henis, Y., and Lahav, N. (1972). Effect of several clay minerals and humic acid on the survival of Klebsiella aerogenes exposed to ultraviolet irradiation. Applied microbiology, 23(5), Bolton, J. R., and Linden, K. G. (2003). Standardization of methods for fluence (UV dose) determination in bench-scale UV experiments. Journal of Environmental Engineering, 129(3), Bolton, J. R., Stefan, M. I., Shaw, P.-S., and Lykke, K. R. (2011). Determination of the quantum yields of the potassium ferrioxalate and potassium iodide iodate actinometers and a method for the calibration of radiometer detectors. Journal of Photochemistry and Photobiology A: Chemistry, 222(1), Bowker, C., Sain, A., Shatalov, M., and Ducoste, J. (2011). Microbial UV fluenceresponse assessment using a novel UV-LED collimated beam system. Water Research, 45(5), Bukhari, A. A. (2008). Investigation of the electro-coagulation treatment process for the removal of total suspended solids and turbidity from municipal wastewater. Bioresource technology, 99(5), Byamukama, D., Kansiime, F., Mach, R. L., and Farnleitner, A. H. (2000). Determination of Escherichia coli contamination with chromocult coliform agar showed a high level of discrimination efficiency for differing fecal pollution levels in tropical waters of Kampala, Uganda. Applied and Environmental Microbiology, 66(2), Cantwell, R. E., Hofmann, R., and Templeton, M. R. (2008). Interactions between humic matter and bacteria when disinfecting water with UV light. Journal of applied microbiology, 105(1), Caron, E., Cheurefils, G., Barbeau, B., Payment, P., and Prevost, M. (2007). Impact of microparticles on UV disinfection of indigenous aerobic spores. Water Research, 41(19), Chevremont, A. C., Farnet, A. M., Coulomb, B., and Boudenne, J. L. (2012). Effect of coupled UV-A and UV-C LEDs on both microbiological and chemical pollution of urban wastewaters. Science of the Total Environment, 426,

67 Chevremont, A. C., Farnet, A. M., Sergent, M., Coulomb, B., and Boudenne, J. L. (2012). Multivariate optimization of fecal bioindicator inactivation by coupling UV-A and UV-C LEDs. Desalination, 285(0), Craik, S. A., Finch, G. R., Bolton, J. R., and Belosevic, M. (2000). Inactivation of Giardia muris cysts using medium-pressure ultraviolet radiation in filtered drinking water. Water Research, 34(18), Craik, S. A., Weldon, D., Finch, G. R., Bolton, J. R., and Belosevic, M. (2001). Inactivation of cryptosporidium parvum oocysts using medium- and low-pressure ultraviolet radiation. Water Research, 35(6), DOWA (2013). Products instruction. Retrieved May 1, 2014, from Eddy, M., Tchobanoglous, G., Burton, F. L., and Stensel, H. D. (2003). Wastewater engineering: Treatment and reuse (4th ed.). Boston: McGraw-Hill. ISBN: X. Edzwald, J. K. (1987). Coagulation-sedimentation-filtration processes for removing organic substances from drinking water. Control of Organic Substances in Water and Wastewater,( pp 26-64). Noyes Data Corporation, Park Ridge New Jersey. Emperor Aquatics, I. (2013). Effective Disinfection and Chlramines removal. Retrieved January 18, 2014, from Goldstein, S., and Rabani, J. (2008). The ferrioxalate and iodide iodate actinometers in the UV region. Journal of Photochemistry and Photobiology A: Chemistry, 193(1), Hamamoto, A., Mori, M., Takahashi, A., Nakano, M., Wakikawa, N., Akutagawa, M., Ikehara, T., Nakaya, Y., and Kinouchi, Y. (2007). New water disinfection system using UVA light emitting diodes. Journal of applied microbiology, 103(6), Hayward, K. (2013). UV LEDs light the way for disruptive technologies. Water 21, Hijnen, W. A. M., Beerendonk, E. F., and Medema, G. J. (2006). Inactivation credit of UV radiation for viruses, bacteria and protozoan (oo)cysts in water: A review. Water Research, 40(1), Hu, J. Y., and Quek, P. H. (2008). Effects of UV radiation on photolyase and implications with regards to photoreactivation following low-and medium-pressure UV disinfection. Applied and Environmental Microbiology, 74(1), Kadam, A. M., Oza, G. H., Nemade, P. D., and Shankar, H. S. (2008). Pathogen removal from municipal wastewater in constructed soil filter. Ecological engineering, 33(1),

68 Kollu, K., and Örmeci, B. (2012). Effect of particles and bioflocculation on ultraviolet disinfection of Escherichia coli. Water Research, 46(3), Linden, K. G., Shin, G., and Sobsey, M. D. (2001). Comparative effectiveness of UV wavelengths for the inactivation of Cryptosporidium parvum oocysts in water. Water Science and Technology, 43(12), Liu, W., Andrews, S. A., Bolton, J. R., Linden, K. G., Sharpless, C., and Stefan, M. (2002). Comparison of disinfection byproduct (DBP) formation from different UV technologies at bench scale. Water Supply, 2(5-6), Mamane, H., Ducoste, J. J., and Linden, K. G. (2006). Effect of particles on ultraviolet light penetration in natural and engineered systems. Applied Optics, 45(8), Masaru, N., and Miho, T. (2013). It will take a few years for wastewater infrastructure rehabilitation, serious damages especially in coastal areas; Japanese: 下水道復旧は数年かかる見通し沿岸部に集中 被害深刻. Retrieved December 10, 2013, from Molleda, P., Blanco, I., Ansola, G., and de Luis, E. (2008). Removal of wastewater pathogen indicators in a constructed wetland in Leon, Spain. Ecological engineering, 33(3), Mosher, J. J., and Gina, M. V. (2012). Ultraviolet disinfection guidelines for drinking water. Fountain Valley, California, USA: National Water Research Institute. ISBN:NWRI Oguma, K., Izaki, K., and Katayama, H. (2013). Effects of salinity on photoreactivation of Escherichia coli after UV disinfection. Journal of Water & Health, 11(3). Oguma, K., Katayama, H., and Ohgaki, S. (2002). Photoreactivation of Escherichia coli after low-or medium-pressure UV disinfection determined by an endonuclease sensitive site assay. Applied and Environmental Microbiology, 68(12), Oguma, K., Katayama, H., and Ohgaki, S. (2004). Photoreactivation of Legionella pneumophila after inactivation by low- or medium-pressure ultraviolet lamp. Water Research, 38(11), Oguma, K., Katayama, H., and Ohgaki, S. (2005). Spectral impact of inactivating light on photoreactivation of Escherichia coli. Journal of Environmental Engineering and Science, 4(S1), S1-S6. Oguma, K., Kita, R., Sakai, H., Murakami, M., and Takizawa, S. (2013). Application of UV light emitting diodes to batch and flow-through water disinfection systems. Desalination, 328, Olstadt, J., Schauer, J., Standridge, J., and Kluender, S. (2007). A comparison of ten USEPA approved total coliform/e. coli tests. Journal of water and health, 5(2),

69 Passantino, L., Malley, J. R., Knudson, M., Ward, R., and Kim, J. (2004). Effect of low turbidity and algae on UV disinfection performance. Journal of American Water Works Association, 96(6), Pilgrim, N., Roche, B., Kalbermatteni, J., Revels, C., and Kariuki, M. (2008). Town water supply and sanitation: challenges, solutions, and guidelines. Retrieved December 3, 2013, from /townwater-supply-sanitation-challenges-solutions-guidelines Qualls, R. G., Flynn, M. P., and Johnson, J. D. (1983). The role of suspended particles in ultraviolet disinfection. Journal (Water Pollution Control Federation), Rahn, R. O. (1997). Potassium Iodide as a Chemical Actinometer for 254 nm Radiation: Use of lodate as an Electron Scavenger. Photochemistry and Photobiology, 66(4), Rahn, R. O. (2013). Fluence Measurements Employing Iodide/Iodate Chemical Actinometry as Applied to Upper Room Germicidal Radiation. Photochemistry and Photobiology, 89(4), Rahn, R. O., Bolton, J. R., and Stefan, M. I. (2006). The lodide/lodate actinometer in UV disinfection: Determination of the fluence rate distribution in UV reactors. Photochemistry and Photobiology, 82(2), Rahn, R. O., Stefan, M. I., Bolton, J. R., Goren, E., Shaw, P. S., and Lykke, K. R. (2003). Quantum Yield of the Iodide Iodate Chemical Actinometer: Dependence on Wavelength and Concentration. Photochemistry and Photobiology, 78(2), Ravazzini, A., Van Nieuwenhuijzen, A., and Van Der Graaf, J. (2005). Direct ultrafiltration of municipal wastewater: comparison between filtration of raw sewage and primary clarifier effluent. Desalination, 178(1), Seoul Optodevice, I. (2013). Product instruction. Retrieved from product/categorys.asp?catecode= Severin, B. F., Suidan, M. T., and Engelbrecht, R. S. (1983). Effect of temperature on ultraviolet light disinfection. Environmental Science & Technology, 17(12), Templeton, M. R., Andrews, R. C., and Hofmann, R. (2005). Inactivation of particleassociated viral surrogates by ultraviolet light. Water Research, 39(15), US EPA. (2006). Ultraviolet disinfection guidance manual for the final long term 2 enhanced surface water treatment rule. Washington D.C., USA: US EPA. Vrhovšek, D., Kukanja, V., and Bulc, T. (1996). Constructed wetland (CW) for industrial waste water treatment. Water Research, 30(10),

70 Wu, Y., Clevenger, T., and Deng, B. (2005). Impacts of goethite particles on UV disinfection of drinking water. Applied and Environmental Microbiology, 71(7), Wurtele, M. A., Kolbe, T., Lipsz, M., Kulberg, A., Weyers, M., Kneissl, M., and Jekel, M. (2011). Application of GaN-based ultraviolet-c light emitting diodes - UV LEDs - for water disinfection. Water Research, 45(3), Zhang, K., and Farahbakhsh, K. (2007). Removal of native coliphages and coliform bacteria from municipal wastewater by various wastewater treatment processes: implications to water reuse. Water Research, 41(12), Zimmer, J. L., and Slawson, R. M. (2002). Potential repair of Escherichia coli DNA following exposure to UV radiation from both medium-and low-pressure UV sources used in drinking water treatment. Applied and Environmental Microbiology, 68(7),

71 Appendix A Experimental data 62

72 The log-reduction of E. coli in clean water at different exposure time was presented in this part. Pour plate technique has been applied to count the density of E. coli, and the agar for pour plate was Chromocult Coliform agar. Table B.1 Log-reduction of E. coli in Clean Water at & s Time for experiment Mar. 4 th, 2014 UV exposure time (s) Density of E. coli ( 10 7 CFU/ ml) log-reduction Table B.2 Log-reduction of E. coli in Clean Water at & s Time for experiment Mar. 6 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction Table B.3 Log-reduction of E. coli in Clean Water at & s Time for experiment Mar. 8 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) error log-reduction error Table B.4 Log-reduction of E. coli in Clean Water at & s Time for experiment Mar. 9 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction Table B.5 Log-reduction of E. coli in Clean Water at 24.31, & s Time for experiment Mar. 11 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction Table B.6 Log-reduction of E. coli in Clean Water at 24.28, & s Time for experiment Mar. 12 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction

73 The log-reduction of E. coli in synthetic wastewater of 70 NTU were presented here and montmorillonite was used to synthetize the turbid wastewater. Table B.7 Log-reduction of E. coli in Synthetic Wastewater of 70 NTU (1 st trial) Time for experiment Mar. 14 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction >7 Table B.8 Log-reduction of E. coli in Synthetic Wastewater of 70 NTU (2 nd trial) Time for experiment Mar. 15 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction Table B.9 Log-reduction of E. coli in Synthetic Wastewater of 70 NTU (3 rd trial) Time for experiment Mar. 17 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) 4.6 > log-reduction 0 < In this section, the data of disinfection test on synthetic wastewater of 156 NTU were presented. Table B.10 Log-reduction of E. coli in Synthetic Wastewater of 156 NTU (1 st trial) Time for experiment Mar. 18 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction Table B.11 Log-reduction of E. coli in Synthetic Wastewater of 156 NTU (2 nd trial) Time for experiment Mar. 20 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction

74 The result of disinfection test with synthetic wastewater of 113 NTU were presented here. Table B.12 Log-reduction of E. coli in Synthetic Wastewater of 113 NTU (1 st trial) Time for experiment Mar. 23 rd, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction Table B.13 Log-reduction of E. coli in Synthetic Wastewater of 113 NTU (2 nd trial) Time for experiment Mar. 25 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction The result of disinfection test with synthetic wastewater of 113 NTU were presented here. Table B.14 Log-reduction of E. coli in Synthetic Wastewater of 27 NTU (1 st trial) Time for experiment Mar. 27 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction Table B.15 Log-reduction of E. coli in Synthetic Wastewater of 27 NTU (2 nd trial) Time for experiment Mar. 29 th, 2014 UV exposure time (s) Density of E. coli ( 107 CFU/ ml) log-reduction

75 In this part, the result of disinfection test with real wastewater was presented, including the result for total coliform and E. coli. The densities of both total coliform and E. coli were counted by MPN method. Table B.16 Log-reduction of total coliform in Real Wastewater of 86 NTU Time for experiment April 10 th, 2014 Turbidity (NTU) 86 exposure time (s) total coliform ( 10 6 MPN/100 ml) log-reduction E. coli detection failed. Table B.17 Log-reduction of total coliform in Real Wastewater of 57 NTU Time for experiment April, 11 th, 2014 Turbidity (NTU) 57 exposure time (s) total coliform ( 10 6 MPN/100 ml) log-reduction E. coli detection failed. Table B.18 Log-reduction of total coliform and E. coli in Real Wastewater of 130 NTU Time for experiment April 14 th, 2014 Turbidity (NTU) 130 NTU exposure time (s) total coliform ( 10 6 MPN/100 ml) log-reduction E. coli ( 10 6 MPN/100 ml) log-reduction Table B.19 Log-reduction of total coliform and E. coli in Real Wastewater of 72 NTU Time for experiment April 17 th, 2014 Turbidity (NTU) 72 exposure time (s) total coliform ( 10 6 MPN/100 ml) log-reduction E. coli ( 10 6 MPN/100 ml) log-reduction

76 Appendix B Experimental set-up 67

77 Figure A.1 Details of experimental set-up 68

UV VALIDATION: HOW TO INTERPRET A VALIDATION REPORT AND CURRENT VALIDATION PRACTICE TRENDS

UV VALIDATION: HOW TO INTERPRET A VALIDATION REPORT AND CURRENT VALIDATION PRACTICE TRENDS UV VALIDATION: HOW TO INTERPRET A VALIDATION REPORT AND CURRENT VALIDATION PRACTICE TRENDS ABSTRACT Peter D Adamo, Ph.D., P.E. and Chengyue Shen, Ph.D., P.E. HDR The USEPA Long Term 2 Enhanced Surface

More information

UV DISINFECTION & VALIDATION

UV DISINFECTION & VALIDATION UV DISINFECTION & VALIDATION Graham Smith Ultraviolet (UV) radiation is being used increasingly to provide an additional barrier to the passage of pathogens in both water and wastewater treatment plants.

More information

Mechanisms and Sustainability of Wavelength-Tailored Ultraviolet Drinking Water Disinfection for Small Systems

Mechanisms and Sustainability of Wavelength-Tailored Ultraviolet Drinking Water Disinfection for Small Systems NWRI GRADUATE FELLOW FINAL PROGRESS REPORT DATE: May 2018 Project Title: Graduate Fellow: Advisor: Institution: Mechanisms and Sustainability of Wavelength-Tailored Ultraviolet Drinking Water Disinfection

More information

Ultraviolet Light for Disinfection of Wastewater. Tom O Connor, PE H 2 O C Engineering

Ultraviolet Light for Disinfection of Wastewater. Tom O Connor, PE H 2 O C Engineering UV Ultraviolet Light for Disinfection of Wastewater Tom O Connor, PE H 2 O C Engineering Solar corona as seen in deep UV light Extreme UV imaging telescope aboard Solar and Heliospheric Observatory spacecraft

More information

Wedeco s iduron INTELLIGENT SOLUTION FOR WASTEWATER DISINFECTION NEEDS

Wedeco s iduron INTELLIGENT SOLUTION FOR WASTEWATER DISINFECTION NEEDS Wedeco s iduron INTELLIGENT SOLUTION FOR WASTEWATER DISINFECTION NEEDS First, let s have a closer Look: Wedeco Production and Services in Herford 1976 Revenues 2017: 127 Mio.$ Employees: 250 Leading global

More information

Drinking Water Treatment Overview Filtration and Disinfection

Drinking Water Treatment Overview Filtration and Disinfection Drinking Water Treatment Overview Filtration and Disinfection April 16 th, 2009 Yousry Hamdy, M.Sc., P.Eng Nicolás s Peleato, EIT 1 Table of Contents Introduction Filtration Cartridge filters Chemical

More information

The small size and ability for UV-LEDs to emit a targeted monochromatic wavelength adds substantial design freedom when configuring a UV-LED reactor.

The small size and ability for UV-LEDs to emit a targeted monochromatic wavelength adds substantial design freedom when configuring a UV-LED reactor. ABSTRACT JENNY, RICHARD MATTHEW. Numerical Optimization and Experimental Validation of a Continuous Flow Point-of-Use UV-LED Disinfection Reactor using Computational Fluid Dynamics. (Under the direction

More information

Scavenger2000 s Technology

Scavenger2000 s Technology Scavenger2000 s Technology THE SYSTEM: When ozone and UV light is added to the aeration/oxygenation process of the Scavenger2000 water decontamination and regeneration vessel, several unique abilities

More information

Ultraviolet Sterilization Technology

Ultraviolet Sterilization Technology Ultraviolet Sterilization Technology Ultraviolet disinfection systems are mysterious to many people how can light kill bacteria? But the truth is it can. Ultraviolet (UV) technology has been around for

More information

Ultraviolet Systems for Water Applications. Water Technologies

Ultraviolet Systems for Water Applications. Water Technologies Ultraviolet Systems for Water Applications Water Technologies With the largest portfolio of innovative products and services, and the most knowledgeable people in the industry, Siemens Water Technologies

More information

Validation Studies and Beyond: Key Implementation Issues

Validation Studies and Beyond: Key Implementation Issues Review of UV Disinfection Validation Studies and Beyond: Key Implementation Issues Paul D. Swaim, P.E., CH2M HILL Denver March 8, 2011 The Burden on State Regulators Has Increased Technology-Specific EPA

More information

Let there be light! - Introduction to USEPA's UV Design Guidance Manual

Let there be light! - Introduction to USEPA's UV Design Guidance Manual Let there be light! - Introduction to USEPA's UV Design Guidance Manual Matthias Boeker Disinfection Market Manager Water & Wastewater UVDGM 2006 Published in November 2006 http://www.epa.gov/safewater/disinfection/lt2/compliance.html

More information

BEET TECHNOLOGY. PART UVC and BEET. - The ultimate disinfection method. BEET AS Tel: (+47)

BEET TECHNOLOGY. PART UVC and BEET. - The ultimate disinfection method. BEET AS Tel: (+47) # 2003 BEET TECHNOLOGY PART 0081 - and BEET - The ultimate disinfection method DNA Hygienic challenges for the food industry and the hospitals Unwanted spread of pathogen microorganisms is perhaps the

More information

ULTRAVIOLET DISINFECTION 2013 Municipal Markets.

ULTRAVIOLET DISINFECTION 2013 Municipal Markets. ULTRAVIOLET DISINFECTION 2013 Municipal Markets www.glascouv.com Glasco Timeline 1940 s NYC based industrial manufacturer - pharmaceutical and food grade stainless steel fabricator 1960 s - Added Industrial

More information

Applied Environmental Microbiology. Copyright McGraw-Hill Global Education Holdings, LLC. Permission required for reproduction or display.

Applied Environmental Microbiology. Copyright McGraw-Hill Global Education Holdings, LLC. Permission required for reproduction or display. 43 Applied Environmental Microbiology Copyright McGraw-Hill Global Education Holdings, LLC. Permission required for reproduction or display. Water Purification and Sanitary Analysis Microbial containment

More information

Linden Laboratory Civil, Environmental and Architectural Engineering University of Colorado, Boulder. November 9,

Linden Laboratory Civil, Environmental and Architectural Engineering University of Colorado, Boulder. November 9, Laboratory MS- 2 Coliphage Testing of the SteriPEN Adventurer Opti against the US EPA Guide Standard and Protocol for Testing Microbiological Water Purifiers Linden Laboratory Civil, Environmental and

More information

Evaluation of Conventional Activated Sludge Compared to Membrane Bioreactors

Evaluation of Conventional Activated Sludge Compared to Membrane Bioreactors Evaluation of Conventional Activated Sludge Compared to Membrane Bioreactors Short Course on Membrane Bioreactors 3/22/06 R. Shane Trussell, Ph.D., P.E. shane@trusselltech.com Outline Introduction Process

More information

Goulburn Mulwaree Council Effluent Irrigation Scheme Medium Pressure UV Disinfection

Goulburn Mulwaree Council Effluent Irrigation Scheme Medium Pressure UV Disinfection Goulburn Mulwaree Council Effluent Irrigation Scheme Medium Pressure UV Disinfection Pieter Groenewegen, Business Development Manager, Contra Shear Technology Abstract Ultraviolet (UV) light has become

More information

UV DISINFECTION. Introduction to TrojanUV

UV DISINFECTION. Introduction to TrojanUV UV DISINFECTION Introduction to TrojanUV Applications for Ultraviolet (UV) Disinfection Disinfection of primary, secondary or tertiary wastewater effluent Disinfection of high quality wastewater for reuse

More information

MAXIMIZING PERFORMANCE OF UV SYSTEMS FOR SMALL COMMUNITY DRINKING WATER DISINFECTION. Scott Bindner NC AWWA-NEA SPRING CONFERENCE April 13, 2015

MAXIMIZING PERFORMANCE OF UV SYSTEMS FOR SMALL COMMUNITY DRINKING WATER DISINFECTION. Scott Bindner NC AWWA-NEA SPRING CONFERENCE April 13, 2015 MAXIMIZING PERFORMANCE OF UV SYSTEMS FOR SMALL COMMUNITY DRINKING WATER DISINFECTION Scott Bindner NC AWWA-NEA SPRING CONFERENCE April 13, 2015 TREATING SMALL COMMUNITY DRINKING WATER Concerns with obtaining

More information

Water Pollution. Chapter 20

Water Pollution. Chapter 20 Water Pollution Chapter 20 Water Pollution Comes from Point and Nonpoint Sources Water pollution Any chemical, biological, or physical change in water quality that harms living organisms or makes water

More information

Advanced disinfection of wastewater ponds effluent by UV irradiation. T. Fuhrmann and K.-U. Rudolph

Advanced disinfection of wastewater ponds effluent by UV irradiation. T. Fuhrmann and K.-U. Rudolph Advanced disinfection of wastewater ponds effluent by UV irradiation T. Fuhrmann and K.-U. Rudolph www.uni-wh-utm.de 1 Pathogen reduction in wastewater ponds Pond systems offer excellent treatment features

More information

RESP ND GROW QUALITY DELIVER. Engineered UV Treatment Systems INNOVATE. Shimon Peretz March 2017 GLOBAL

RESP ND GROW QUALITY DELIVER. Engineered UV Treatment Systems INNOVATE. Shimon Peretz March 2017 GLOBAL GROW Engineered UV Treatment Systems Shimon Peretz March 2017 QUALITY DELIVER RESP ND INNOVATE GLOBAL Presentation Outline What is Disinfection? UV vs. Chlorination What is UV disinfection? Design Requirements

More information

Ultraviolet disinfection. Guidance document

Ultraviolet disinfection. Guidance document Ultraviolet disinfection Guidance document About Australian WaterSecure Innovations Ltd Australian WaterSecure Innovations Ltd (trading as WaterSecure) was established in 2016 to oversee the implementation

More information

Water Pollution. Objective: Name, describe, and cite examples of the eight major types of water pollution.

Water Pollution. Objective: Name, describe, and cite examples of the eight major types of water pollution. Water Pollution Objective: Name, describe, and cite examples of the eight major types of water pollution. Types of Water Pollution Water pollutants are divided into eight categories: 1. Sediment pollution

More information

3 8 COLIFORM BACTERIA AS INDICATOR ORGANISMS Laboratory tests for disease-producing bacteria, viruses, and protozoa are difficult to perform

3 8 COLIFORM BACTERIA AS INDICATOR ORGANISMS Laboratory tests for disease-producing bacteria, viruses, and protozoa are difficult to perform 3 8 COLIFORM BACTERIA AS INDICATOR ORGANISMS Laboratory tests for disease-producing bacteria, viruses, and protozoa are difficult to perform Most utilities have neither qualified personnel nor laboratories

More information

MEMBRANE BIO-REACTOR. Prashanth N 1 1. INTRODUCION

MEMBRANE BIO-REACTOR. Prashanth N 1 1. INTRODUCION International Journal of Latest Trends in Engineering and Technology Vol.(7)Issue(3), pp. 296 301 DOI: http://dx.doi.org/10.21172/1.73.540 e ISSN:2278 621X MEMBRANE BIO-REACTOR Prashanth N 1 ABSTRACT:

More information

UV Sterilizers by Chiyoda Kohan Chiyoda Kohan Co., Ltd. UV Systems Division.

UV Sterilizers by Chiyoda Kohan Chiyoda Kohan Co., Ltd. UV Systems Division. UV Sterilizers by 2017.9.20 Co., Ltd. UV Systems Division. 1 Table of Contents 1. What is ultraviolet? 2. Sterilization (inactivation) by ultraviolet 3. Oxidative decomposition of small amounts of organic

More information

Mutation. ! Mutation occurs when a DNA gene is damaged or changed in such a way as to alter the genetic message carried by that gene

Mutation. ! Mutation occurs when a DNA gene is damaged or changed in such a way as to alter the genetic message carried by that gene Mutations Mutation The term mutation is derived from Latin word meaning to change.! Mutation occurs when a DNA gene is damaged or changed in such a way as to alter the genetic message carried by that gene!

More information

Lecture 7 Water Quality Monitoring: Estimation of fecal indicator bacteria

Lecture 7 Water Quality Monitoring: Estimation of fecal indicator bacteria Lecture 7 Water Quality Monitoring: Estimation of fecal indicator bacteria FECAL INDICATOR BACTERIA Fecal indicator bacteria, which are directly associated with fecal contamination, are used to detect

More information

Wastewater Treatment Works... The Basics

Wastewater Treatment Works... The Basics United States EPA 833-F-98-002 Environmental Protection May 1998 Agency Office of Water (4204) Wastewater Treatment Works... The Basics O ne of the most common forms of pollution control in the United

More information

NO IT WON T MAKE YOU STERILE

NO IT WON T MAKE YOU STERILE NO IT WON T MAKE YOU STERILE ULTRAVIOLET DISINFECTION FROM AN OPERATIONAL PRESPECTIVE SAFE DRINKING WATER WORKSHOP MARCH 22, 2006 GANDER NEWFOUNDLAND BRIAN PETT 5 SQUARE Consulting Inc. CALEDONIA, ON.

More information

Pathogens in Water. Monitoring, Control and Technologies Available for Treatment. Prepared by; Paul O Callaghan M.Sc. Dr.

Pathogens in Water. Monitoring, Control and Technologies Available for Treatment. Prepared by; Paul O Callaghan M.Sc. Dr. Pathogens in Water National Water Summit, Tuesday, 2nd October 2007 Monitoring, Control and Technologies Available for Treatment Prepared by; Paul O Callaghan M.Sc. Dr. Hans Gethke Filtration of drinking

More information

Disinfection of Total Coli-forms in the Effluent from Municipal Wastewater Plant with e- Beam. June 07, EB TECH Co.

Disinfection of Total Coli-forms in the Effluent from Municipal Wastewater Plant with e- Beam. June 07, EB TECH Co. 1 Disinfection of Total Coli-forms in the Effluent from Municipal Wastewater Plant with e- Beam June 07, 2005 Yuri Kim, Bumsoo HAN, Jinkyu Kim EB TECH Co. International Symposium on the Utilization of

More information

Selection of UV Disinfection Systems for Achieving Class A Reuse Water at Small MBR plants

Selection of UV Disinfection Systems for Achieving Class A Reuse Water at Small MBR plants Selection of UV Disinfection Systems for Achieving Class A Reuse Water at Small MBR plants Li Lei, Douglas Berschauer, Elizabeth Hill, and Tom Moore 1 Outline Introduction (UV, WWTPs) Regulations on UV

More information

Calculated Dose Approach for MP systems with low and high wavelength sensors

Calculated Dose Approach for MP systems with low and high wavelength sensors February, 2018 Calculated Dose Approach for MP systems with low and high wavelength sensors logi = 10 A UVA B UVA254 254 + 10 D F SH S0H Q ASCF UVA G UVA220 220 H D 2 C+ D UVA254+ E UVA254 S S0 Q ASCF

More information

UV TREATMENT EFFICIENCY FOR E. COLI IN STORMWATER CONTAINING DIFFERENT SIZE FRACTIONS OF SUSPENDED SOLIDS

UV TREATMENT EFFICIENCY FOR E. COLI IN STORMWATER CONTAINING DIFFERENT SIZE FRACTIONS OF SUSPENDED SOLIDS UV TREATMENT EFFICIENCY FOR E. COLI IN STORMWATER CONTAINING DIFFERENT SIZE FRACTIONS OF SUSPENDED SOLIDS Shawn P. McElmurry 1, Shawn M. Ingram 2, Nafa Khalaf 3, and Gopu Pillai 4 1 Department of Civil

More information

Solar-based UV: Fundamentals and applications. Tamar Kohn Associate Professor Environmental Engineering Insititute

Solar-based UV: Fundamentals and applications. Tamar Kohn Associate Professor Environmental Engineering Insititute Solar-based UV: Fundamentals and applications Tamar Kohn Associate Professor Environmental Engineering Insititute Solar radiation spectrum Wikipedia Solar vs lamp UV spectrum UVC UVB UVA Relative Output

More information

Why Choose BacTerminator?

Why Choose BacTerminator? Technical Note Why Choose BacTerminator? Dear customer, we proudly introduce Why choose BacTerminator? because your care about your water and want the best technology for your water The green alternative

More information

Inactivation of Giardia lamblia cysts by UV irradiation in real field waters. INTRODUCTION

Inactivation of Giardia lamblia cysts by UV irradiation in real field waters. INTRODUCTION Inactivation of Giardia lamblia cysts by UV irradiation in real field waters. Gwy-Am Shin 1 *, Karl G. Linden 2, Gaetan Faubert 3, and Mark D. Sobsey 4 1: Department of Environmental and Occupational Health

More information

Aeration, filtration and disinfection in aquaculture

Aeration, filtration and disinfection in aquaculture Aeration, filtration and disinfection in aquaculture Aeration, filtration and disinfection in aquaculture Boby Ignatius Central Marine Fisheries Research Institute Post Box No. 1603, Ernakulam North P.O,

More information

NC-PC Industry Day Pretreatment 101. Industrial Waste Impacts on POTW Treatment Processes. Dawn Padgett Operations Manager Charlotte Water

NC-PC Industry Day Pretreatment 101. Industrial Waste Impacts on POTW Treatment Processes. Dawn Padgett Operations Manager Charlotte Water NC-PC Industry Day Pretreatment 101 Industrial Waste Impacts on POTW Treatment Processes Dawn Padgett Operations Manager Charlotte Water NC-PC Industry Day Definitions BOD Amount of oxygen consumed by

More information

Microbial Removal Efficiency of UV in Tehran Shahid Mahallati Wastewater Treatment Plant

Microbial Removal Efficiency of UV in Tehran Shahid Mahallati Wastewater Treatment Plant (// // )... /.. / NTU. / /... MPN/100mL MPN/100mL. : Microbial Removal Efficiency of UV in Tehran Shahid Mahallati Wastewater Treatment Plant Reza Dabbagh 1 (Received Feb. 22, 2008 Accepted Feb. 17, 2009)

More information

Inhibiting Microbial Growth in vitro. CLS 212: Medical Microbiology Zeina Alkudmani

Inhibiting Microbial Growth in vitro. CLS 212: Medical Microbiology Zeina Alkudmani Inhibiting Microbial Growth in vitro CLS 212: Medical Microbiology Zeina Alkudmani Microbicidal or Microbistatic? Microbicidal Microbicidal is the process or an agent that kills the microorganism. The

More information

UV Wastewater. Vertical Open Channel Ultraviolet Disinfection

UV Wastewater. Vertical Open Channel Ultraviolet Disinfection UV Wastewater Vertical Open Channel Ultraviolet Disinfection GLOW VC Series dour Company Glasco Ultraviolet has been manufacturing UV water and wastewater disinfection systems for over 50 years. We manufacture

More information

Inhibiting of Microbial Growth in vitro CLS 212

Inhibiting of Microbial Growth in vitro CLS 212 Inhibiting of Microbial Growth in vitro CLS 212 Microbicidal Microbicidal is the process or an agent that kills the microorganism. The suffix -cidal or cide means??( See chapter 8 page 131) Microbistatic

More information

ULTRAVIOLET DISINFECTION FOR THE HETCH HETCHY AQUEDUCT

ULTRAVIOLET DISINFECTION FOR THE HETCH HETCHY AQUEDUCT Operated by the San Francisco Public Utilities Commission ULTRAVIOLET DISINFECTION FOR THE HETCH HETCHY AQUEDUCT Enio Sebastiani, SFPUC Water Quality Division Wholesale Customer Workshop 11-9-11 1 Overview

More information

2009 Summary Report for Municipalities

2009 Summary Report for Municipalities 2009 Summary Report for Municipalities (As per Schedule 22 of O. Reg. 170/03) The Onaping Potable Water System - #220003519 Certificate of Approval 1414-0696HD3 0014-6MQQFQ Prepared by: Jean-Marc Joliat

More information

Comparison of Plant Processes and Operations for Producing Potable Water from the Available Source Water

Comparison of Plant Processes and Operations for Producing Potable Water from the Available Source Water Comparison of Plant Processes and Operations for Producing Potable Water from the Available Source Water Reveal types of source waters and the ones that are common in Canada Look at many processes that

More information

8. E and 4. Water Quality, Indicators, and Usage

8. E and 4. Water Quality, Indicators, and Usage 8. E. 1. 3 and 4 Water Quality, Indicators, and Usage How is the health of a water system determined? Balance between physical, chemical, and biological factors Physical variables include: temperature,

More information

UV-Licht. ... und... und seine Wirkung. HYDRO PUR UV Disinfection system B Series. Application

UV-Licht. ... und... und seine Wirkung. HYDRO PUR UV Disinfection system B Series. Application relative units relative Einheiten relative Einheiten UV Disinfection system B Series Application Our UV disinfection systems of type B are used for disinfection of iron and manganese free drinking water

More information

UV-LED Curing for Industrial Printing

UV-LED Curing for Industrial Printing UV-LED Curing for Industrial Printing (March 2012) posted on Tue Apr 24, 2012 UV-LED curing units have become more efficient in delivering higher energy to the By Richa Anand UV-LED curing refers to a

More information

Treatment Processes for Potable Water

Treatment Processes for Potable Water Treatment Processes for Potable Water Drinking Water Potable water is water that is free of harmful chemicals, microorganisms, and other contaminants and is safe to drink Drinking water also should be

More information

White Paper Adhesives Sealants Tapes

White Paper Adhesives Sealants Tapes Is LED the UV technology of the future? An objective comparison between LED and UV technology For many years UV technology has been a reliable method for the curing of photo-reactive chemicals. In response

More information

Water Quality - Condensed Version 1999

Water Quality - Condensed Version 1999 11.0 DISINFECTION refers to operations in water treatment that kills or renders harmless pathogenic microorganisms but does not refer to sterilization. sterilization; the complete destruction of all living

More information

Inactivation of various bacteriophages by different UV technologies: Development of a reliable virus indicator system for water reuse

Inactivation of various bacteriophages by different UV technologies: Development of a reliable virus indicator system for water reuse Environ. Eng. Res. 2016 Research Article http://dx.doi.org/10.4491/eer.2016.032 pissn 1226-1025 eissn 2005-968X In Press, Uncorrected Proof Inactivation of various bacteriophages by different UV technologies:

More information

Module 1 WATER SUPPLY SYSTEM BASIC OPERATIONS

Module 1 WATER SUPPLY SYSTEM BASIC OPERATIONS Module 1 WATER SUPPLY SYSTEM BASIC OPERATIONS Workbook Financial/Managerial Series This course includes content developed by the Pennsylvania Department of Environmental Protection in cooperation with

More information

Drinking Water Supply and

Drinking Water Supply and Drinking Water Supply and Health Engineered Water Systems Water and Health 80% of sickness in the world is caused by inadequate water supply or sanitation 40% of the world population does not have access

More information

UV Disinfection for a Municipal Wastewater Lagoon Effluent No Prior Filtration Needed? Wayne Wong, M.A.Sc., P.Eng., PMP

UV Disinfection for a Municipal Wastewater Lagoon Effluent No Prior Filtration Needed? Wayne Wong, M.A.Sc., P.Eng., PMP UV Disinfection for a Municipal Wastewater Lagoon Effluent No Prior Filtration Needed? Wayne Wong, M.A.Sc., P.Eng., PMP Acknowledgements 2 Northern Rockies Regional Municipality Northern Rockies Engineering

More information

Wastewater Treatment Processes

Wastewater Treatment Processes Wastewater Treatment Processes (Sep 27 th and 28 th, 2016) by Dr. Arun Kumar (arunku@civil.iitd.ac.in) Objective: To learn about processes used in tertiary treatment Courtesy: Dr. Irene Xagoraraki, MSU,

More information

Microbiology sheet (6)

Microbiology sheet (6) Microbiology sheet (6) Made by marah marahleh corrected by : abd. Salman DATE :9/10/2016 Microbial growth / control of microbial growth 1 The method of counting bacteria is divided into: 1) direct 2) indirect

More information

Multiple Disinfectant Barriers Within A Multiple Barrier System: A New Drinking Water Paradigm for Public Health Protection

Multiple Disinfectant Barriers Within A Multiple Barrier System: A New Drinking Water Paradigm for Public Health Protection Multiple Disinfectant Barriers Within A Multiple Barrier System: A New Drinking Water Paradigm for Public Health Protection James P. Malley, Jr.; Steven E. Farabaugh; David R. Gaithuma and Nicola A. Ballester

More information

Overcoming UV Interference in Leachate Discharges

Overcoming UV Interference in Leachate Discharges Overcoming UV Interference in Leachate Discharges NEW YORK STATE ASSOCIATION FOR SOLID WASTE MANAGEMENT BOLTON LANDING, NY Ivan A. Cooper, PE, BCEE Civil & Environmental Consultants, Inc. Olya S. Keen,

More information

POLY ALUMINIUM CHLORIDE AS AN ALTERNATIVE TO ALUM AS A COAGULANT IN WATER TREATMENT

POLY ALUMINIUM CHLORIDE AS AN ALTERNATIVE TO ALUM AS A COAGULANT IN WATER TREATMENT POLY ALUMINIUM CHLORIDE AS AN ALTERNATIVE TO ALUM AS A COAGULANT IN WATER TREATMENT Shyama Dharmasinghe (Adm. No: 139206L) Degree of Master of Engineering in Environment Engineering and Management Department

More information

Coagulation and Flocculation: Color Removal

Coagulation and Flocculation: Color Removal Coagulation and Flocculation: Color Removal Submitted to: Dr. Hashsham Research Complex Engineering Department of Civil and Environmental Engineering Michigan State University East Lansing, MI 48824 Authors

More information

Can MBR Eliminate Additional Disinfection? A Case Study. Ufuk Erdal, PhD, PE, CH2M Jonathan Vorheis, PE, CH2M July 17, 2015

Can MBR Eliminate Additional Disinfection? A Case Study. Ufuk Erdal, PhD, PE, CH2M Jonathan Vorheis, PE, CH2M July 17, 2015 Can MBR Eliminate Additional Disinfection? A Case Study Ufuk Erdal, PhD, PE, CH2M Jonathan Vorheis, PE, CH2M July 17, 2015 2 Acknowledgement We sincerely thank to co-authors for their valuable contribution

More information

Electrocoagulation. Achieving clean, clear, treated and reusable water: The process, technology and benefits. CALL US (631)

Electrocoagulation. Achieving clean, clear, treated and reusable water: The process, technology and benefits.   CALL US (631) Electrocoagulation Achieving clean, clear, treated and reusable water: The process, technology and benefits. WWW.AWWTCORP.COM CALL US (631) 213-1324 SEWAGE WASTE Clean water is vital to virtually all living

More information

EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method 10258

EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method 10258 EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method 10258 TURBIDITY MEASUREMENT HISTORY 3 TURBIDITY MEASUREMENT OVER THE LAST 50 YEARS TURBIDITY MEASUREMENT Turbidity is the interaction

More information

Waste water treatment Biological treatment

Waste water treatment Biological treatment Leonardo da Vinci Project Sustainability in commercial laundering processes Module 1 Usage of Water Chapter 5 Waste water treatment Biological treatment Module 1 Usage of water Chapter 5 Waste Water Treatment

More information

PROCESS OPTIMIZATION USING SPECTROPHOTOMETERS

PROCESS OPTIMIZATION USING SPECTROPHOTOMETERS www.zapstechnologies.com PROCESS OPTIMIZATION USING SPECTROPHOTOMETERS By Bud Dunbar ZAPS Technologies 617-448-0440 robert.dunbar@zapstechnologies.com 1 2 WHAT S CHANGED? INEXPENSIVE & POWERFUL PHOTON

More information

University of South Florida

University of South Florida ENV 4417: WATER QUALITY & TREATMENT Fall 2015 Problem set #4 Due Tuesday, Oct. 13 University of South Florida Civil & Environmental Eng. Prof. J. A. Cunningham For 2015, answer problem 1, then problem

More information

Chapter 9 Controlling Microbial Growth in the Environment. 10/1/ MDufilho

Chapter 9 Controlling Microbial Growth in the Environment. 10/1/ MDufilho Chapter 9 Controlling Microbial Growth in the Environment 10/1/2017 1 MDufilho Table 91 Terminology of Microbial Control 10/1/2017 MDufilho 2 Number of living microbes Figure 91 A plot of microbial death

More information

VC Series Wastewater. Vertical Open Channel Ultraviolet Disinfection

VC Series Wastewater. Vertical Open Channel Ultraviolet Disinfection VC Series Wastewater Vertical Open Channel Ultraviolet Disinfection VC Series dour Company Glasco has been manufacturing UV disinfection systems for over 50 years. Equipment is manufactured for a variety

More information

Water Chemistry Monitoring

Water Chemistry Monitoring Chicago River Fieldtrip Activity Summary Students use basic to advanced water quality monitoring kits and probes to assess the water quality of the Chicago River. Water chemistry kits available for loan

More information

The Control of Microorganisms LC D R B R I A N B E A R D E N, M S, P E

The Control of Microorganisms LC D R B R I A N B E A R D E N, M S, P E The Control of Microorganisms LC D R B R I A N B E A R D E N, M S, P E U S P U B L I C H E A LT H S E R V I C E / U S E PA R 9 M A R I A N A I S L A N D S WAT E R O P E R ATO R S A S S O C I AT I O N F

More information

Water Pollution & Quality. Dr. Deniz AKGÜL Marmara University Department of Environmental Engineering

Water Pollution & Quality. Dr. Deniz AKGÜL Marmara University Department of Environmental Engineering Water Pollution & Quality Dr. Deniz AKGÜL Marmara University Department of Environmental Engineering IMPORTANCE OF WATER Life on planet Earth would be impossible without water. All life forms, from simple

More information

Disinfection and Disinfection Byproducts (DBP)

Disinfection and Disinfection Byproducts (DBP) Disinfection and Disinfection (DBP) 1 Regulatory Requirements for Disinfection By-Product Management & Disinfection By-Product Monitoring for Water Treatment Plants 2 DPB Rule Requirements Rule/Requirement

More information

Biofouling Control Using UVC LEDs

Biofouling Control Using UVC LEDs APPLICATION NOTE: Biofouling Control Using UVC LEDs July 10, 2015 THIS APPLICATION NOTE DESCRIBES THE USE OF UVC LEDS AS A SOURCE OF GERMICIDAL RADIATION TO PREVENT BIOFILM FORMATION AND BIOFOULING IN

More information

Ultraviolet Disinfection BIO-SECURITY Specification, Operation and Maintenance in a New Era

Ultraviolet Disinfection BIO-SECURITY Specification, Operation and Maintenance in a New Era 2010 AALSO Symposium Moody Gardens Presented By: Dan Dycha Ultraviolet Disinfection BIO-SECURITY Specification, Operation and Maintenance in a New Era Agenda Why UV? UV Basics Types of Hg Lamps UV Design

More information

http://www.photonics.com/article.aspx?aid=60709 HARI VENUGOPALAN, CRYSTAL IS INC. UVC LEDs are fast becoming the technology of choice for designers developing a new breed of spectroscopy devices that are

More information

Overview of Chapter 21

Overview of Chapter 21 21 Water Pollution Overview of Chapter 21 Types of Water Pollution Water Quality Today Municipal Agricultural Industrial Improving Water Quality Laws Controlling Water Pollution Types of Water Pollution

More information

Cincinnati, Ohio 45268

Cincinnati, Ohio 45268 INTRODUCTION A MICROBIOLOGAL SURROGATE FOR EVALUATING TREATMENT EFFICIENCY Eugene W. Rice, Kim W. FOX, Richard J. Miltner, Darren A. Lytle and Clifford H. Johnson Drinking Water Research Division Risk

More information

UV DISINFECTION OF LOW TRANSMITTANCE PHARMACEUTICAL WASTEWATER

UV DISINFECTION OF LOW TRANSMITTANCE PHARMACEUTICAL WASTEWATER UV DISINFECTION OF LOW TRANSMITTANCE PHARMACEUTICAL WASTEWATER Jurek Patoczka, PhD, PE Hatch Mott MacDonald, 27 Bleeker Str., Millburn, NJ 74 jurek.patoczka@hatchmott.com ABSTRACT An UV disinfection system

More information

TITLE 252. DEPARTMENT OF ENVIRONMENTAL QUALITY CHAPTER 626. PUBLIC WATER SUPPLY CONSTRUCTION STANDARDS SUBCHAPTER 1. INTRODUCTION

TITLE 252. DEPARTMENT OF ENVIRONMENTAL QUALITY CHAPTER 626. PUBLIC WATER SUPPLY CONSTRUCTION STANDARDS SUBCHAPTER 1. INTRODUCTION TITLE 252. DEPARTMENT OF ENVIRONMENTAL QUALITY CHAPTER 626. PUBLIC WATER SUPPLY CONSTRUCTION STANDARDS SUBCHAPTER 1. INTRODUCTION 252:626-1-2. Definitions Terms have the meaning assigned in the Environmental

More information

Lowering The Total Cost Of Operation

Lowering The Total Cost Of Operation Lowering The Total Cost Of Operation The system removes more solids than conventional clarification, so filters can run longer between backwash cycles. Fewer backwash cycles means less backwash water,

More information

ADVANCED UV SOURCE FOR BIOLOGICAL AGENT DESTRUCTION

ADVANCED UV SOURCE FOR BIOLOGICAL AGENT DESTRUCTION ADVANCED UV SOURCE FOR BIOLOGICAL AGENT DESTRUCTION Wayne Clark, Joseph Stumpf, Bernard Eastlund, and Michael Ingram Novatron, Inc. 5955 Mira Mesa Blvd., Suite A San Diego, CA 92121 1.0 EXECUTIVE SUMMARY

More information

Importance. Prokaryotes vs. Eukaryotes. Viruses: a form of life or not?

Importance. Prokaryotes vs. Eukaryotes. Viruses: a form of life or not? 1 Importance Microorganisms (esp. bacteria) plays a key role in the decomposition and stabilization of organic matter Control of diseases caused by pathogenic organisms of human origin Prokaryotes vs.

More information

DISCUSSION PAPER. 1 Introduction. 2 Existing Legislation and Guidelines

DISCUSSION PAPER. 1 Introduction. 2 Existing Legislation and Guidelines DISCUSSION PAPER Arrowsmith Water Service Englishman River Water Intake Study Phase 1 Conceptual Planning Discussion Paper 4-2 Drinking Water Quality Criteria Prepared by: Keith Kohut and Matt Henney Issued:

More information

INTRODUCTION TO UV-OXIDATION FOR ENVIRONMENTAL CONTAMINANT TREATMENT

INTRODUCTION TO UV-OXIDATION FOR ENVIRONMENTAL CONTAMINANT TREATMENT INTRODUCTION TO UV-OXIDATION FOR ENVIRONMENTAL CONTAMINANT TREATMENT Trojan: Providing Water Confidence Trojan has a solution to treat water at virtually every point in its cycle of use to give you confidence

More information

As part of its program to help protect the quality of urban waterways and Lake Erie, the

As part of its program to help protect the quality of urban waterways and Lake Erie, the Enhanced Pathogen Reduction of Urban Wet Weather Flows BACKGROUND As part of its program to help protect the quality of urban waterways and Lake Erie, the City of Toledo operates up to 776-ML/d (205-mgd)

More information

TITLE : : SABS/TC 147/SC

TITLE : : SABS/TC 147/SC : 2018-03-22 List of Standards and Abstracts Page No : 1 Of 7 Corr SANS 203 SABS SM 1158 2007 2.00 ST Bactericidal and algaecidal efficacy of swimming pool chemicals Specifies a presumptive test for the

More information

FOOD PRODUCTION WATER SUPPLIES

FOOD PRODUCTION WATER SUPPLIES FOOD PRODUCTION WATER SUPPLIES SAFOODNET (SSA) 3 rd Open seminar Dr. Mehlika Borcakl 5 May 2009 Tallinn Water Sources v Most food processors use water from city and municipal suppliers or from wells. v

More information

Wastewater Treatment clarifier

Wastewater Treatment clarifier Wastewater Treatment Pretreatment During pretreatment, items would normally be removed that would hinder the further processes of treatment. Items commonly removed include roots, rags, cans, or other large

More information

Platinum will ensure the safety of your water...

Platinum will ensure the safety of your water... Sterilight Platinum will ensure the safety of your water... Sterilight s PLATINUM UV disinfection systems offer simple, safe and effective disinfection. If water-borne pathogens, including bacteria, virus,

More information

RELATIONSHIP BETWEEN ULTRAVIOLET SENSITIVITY AND

RELATIONSHIP BETWEEN ULTRAVIOLET SENSITIVITY AND JOURNAL OF BACTERIOLOGY Vol. 88, No. 5, p. 1283-1287 November, 1964 Copyright X 1964 American Society for Microbiology Printed in U.S.A. RELATIONSHIP BETWEEN ULTRAVIOLET SENSITIVITY AND ABILITY TO PROPAGATE

More information

Issues. Issues (cont.) A multiple barriers approach is required to provide potable water

Issues. Issues (cont.) A multiple barriers approach is required to provide potable water EFFECTIVE TECHNOLOGIES TO PROVIDE POTABLE WATER SUPPLY FOR SANITIZING DAIRY FARMS MILKING EQUIPMENTS D.I. Massé, G. Séguin, L. Masse, E. Topp, F. Malouin, E. Pariseau, L. Ortega PRESENTATION OUTLINE Issues

More information

Science Exploration. DHRITI BHATTACHARJEE Class : VII/C Roll No : 31

Science Exploration. DHRITI BHATTACHARJEE Class : VII/C Roll No : 31 Science Exploration DHRITI BHATTACHARJEE Class : VII/C Roll No : 31 INTRODUCTION TO WATER TREATMENT PLANTS Visit to Nimeta Water Treatment Plant, Vadodara INDEX 1. Introduction to Water Treatment Plants

More information

CGN 6933: Drinking Water Treatment Processes Department of Civil & Environmental Engineering University of South Florida Cunningham Spring 2013

CGN 6933: Drinking Water Treatment Processes Department of Civil & Environmental Engineering University of South Florida Cunningham Spring 2013 Homework #6 Due Thurs., April 11 Topic: Disinfection Assignment for 2013: Problems 4, 5, 8, 12 1. Answer question 13-5 in the Howe text book (same as 13-4 in the Crittenden text, 3rd edition). Don t worry

More information

Unit 3c. Microbial Gene0cs

Unit 3c. Microbial Gene0cs Unit 3c Microbial Gene0cs Microbial Genetics! Gene0cs: the science of heredity Genome: the gene0c informa0on in the cell Genomics: the sequencing and molecular characteriza0on of genomes Gregor Mendel

More information

We used these results to calculate the Water Quality Index (WQI) using this formula:

We used these results to calculate the Water Quality Index (WQI) using this formula: Water Quality Study The water quality study of Sungai Way was undertaken by a certified laboratory (EraLab) on Friday, March 7, 2008 from 9.30am to 12pm. The official Certificate of Analysis of the water

More information