Implement High Quality, Routine GC-MS/MS Multi-Residue Pesticides Analysis with Ease Paul Silcock & Jason Cole
|
|
- Barbara Skinner
- 6 years ago
- Views:
Transcription
1 Implement High Quality, Routine GC-MS/MS Multi-Residue Pesticides Analysis with Ease Paul Silcock & Jason Cole 1 The world leader in serving science
2 Some discussion points for today.. Trend towards GC Triple Quadrupole MS for amenable pesticides Challenges faced when setting up and working with complex pesticides methods with GC triple quadrupole MS How to easily implement and manage high performance GC- MS/MS pesticide methods 2
3 Increasing popularity in triple quadrupole million books: ~4% of all books ever published 3
4 GC-Triple Quad Adoption 80 s, 90 s
5 GC-MS/MS What s so special? Low detection limits Optimized sample preparation Consolidated analytical methods Faster, automated data processing...it s a high selectivity technique... 5
6 GC-MS/MS What s so special? SIM on m/z 182 SRM on m/z 182 > 153 Acibenzolar-S-Methyl in vegetable matrix 6
7 GC-MS/MS What s so special? Extra dimension of MS to drive selectivity SIM on m/z 182 SRM on m/z 182 > 153 Acibenzolar-S-Methyl in vegetable matrix 7
8 Inside the Thermo Scientific TSQ 8000 GC-MS/MS Ion Source 1 st Quadrupole Q1 3 rd Quadrupole Q3 Ion Guide 2 nd Quadrupole Q2 Collision Cell Dynode and Electron Multiplier 8
9 Step 1: Precursor Ion Selection Selected analyte precursor ion Matrix ion with same mass number 9
10 Step 1: Precursor Ion Selection Acibenzolar-S-Methyl 10 ppb in Lettuce Significant contribution from matrix ion causes peak area to be too high 10
11 Step 2: Fragmentation in Collision Cell Fragmentation of both analyte and matrix ion 11
12 Step 3: Production Ion Selection Selection of unique product ion of analyte 12
13 Step 3: Production Ion Selection m/z 182 m/z 153 SRM removes the shoulder on from matrix interference & reduces background SIM on m/z 182 SRM on m/z 182 >
14 Benefits of Selectivity High selectivity Possibility the reduce selectivity in sample preparation Reduced sample prep steps creates a more generic sample prep method more compounds & matrices Consolidated GC-MS methods due to high performance buffer against requirements Compressed chromatography possible Easy peak evaluation auto-integrators 14
15 Productivity Challenges: New GC-MS/MS Pesticide Methods Method implementation Routine result production Dealing with large compound lists GC Performance Maintenance MS/MS method development MS Performance Maintenance Acquisition method optimization Data processing method production Acquisition Method Maintenance Data analysis method maintenance 15
16 Productivity challenges: new GC-MS/MS pesticide methods.. Downtime Experience Resource Method implementation Routine result production Dealing with large compound lists GC Performance Maintenance MS/MS method development MS Performance Maintenance Acquisition method optimization Data processing method production Acquisition Method Maintenance Data analysis method maintenance 16
17 Barriers to Adoption of GC-MS/MS for Pesticides...or adopting a new brand/model... Learning a new technique Moving across existing methods MS/MS Method development Data processing method development System/software familiarization Time away from real result production 17
18 Where we want to be... To be able to rapidly implement our own customized pesticide analysis methods on the new platform and start to be productive. 18
19 Productivity Challenges: New GC-MS/MS Pesticide Methods Downtime Experience Resource Method implementation Routine result production Dealing with large compound lists GC Performance Maintenance MS/MS method development MS Performance Maintenance Acquisition method optimization Data processing method production Acquisition Method Maintenance Data analysis method maintenance 19
20 Ease the Method Implementation Process Method implementation Routine result production Dealing with large compound lists GC Performance Maintenance MS/MS method development MS Performance Maintenance Acquisition method optimization Data processing method production Acquisition Method Maintenance Data analysis method maintenance 20
21 Thermo Scientific TSQ 8000 Pesticide Analyzer A complete pesticide method implementation, management and maintenance solution to drive unstoppable result productivity 21
22 Powering the TSQ 8000 Pesticide Analyzer Preconfigured performance leading TSQ 8000 GC-MS/MS system featuring the award-winning Thermo Scientific TRACE 1310 GC Pre-loaded acquisition methods Pre-loaded Thermo Scientific TraceFinder 3.0 processing methods Thermo Scientific TraceGOLD GC Column and consumable technology TraceFinder 3.0 EFS Data Processing software 600+ Pesticide compound database (CDB) with SRM transitions AutoSRM & timed SRM (t-srm) Pesticide Analyzer Installation Guide 22
23 What s Different? Truly self-customizable solution Full self sufficiency using the integrated software tools and supporting guides TraceFinder 3.0 Compound Database that is workflow integrated No need for advanced spreadsheet knowledge for complex operations and formulas Automatic optimization Timed SRM means there is no need for extensive lists, for time segment calculations or dwell time optimization. Automated SRM development through AutoSRM for when your method needs to move outside the compound database. No need for manual and advanced level SRM method development 23
24 TSQ 8000 Pesticide Analyzer Workflow My target compound list Develop missing compound/ SRM Pesticide Compound Database Add to database Automatic MRM method creation RT Update Self -customized Auto-optimized My completed pesticide method 24
25 Self Customized: Create Your Target Acquisition List The installation guide describes how selecting your compounds from the CDB populates your acquisition and processing methods 25
26 Auto-Optimized : Adding New Compounds Step 1. Automated SRM development through AutoSRM Step-by-step instruction make adding new compounds to the Compound Database easy. Step 2. Easy import into TraceFinder CDB 26
27 AutoSRM: Fast, Simple Route to Optimized SRM 1) Precursor ion selection AutoSRM automates the development of SRM methodology 2) Product ion selection 3) Collision energy optimization 27
28 Highlights of AutoSRM End result showing optimized transition Automates the following: Creation of full scan, product ion scan and SRM methods Creation of sample sequences Creation of data layouts for analyzing results Selection of precursor, product and collision energies 28
29 Auto-Optimized :Timed-SRM Classical segmented SRM methods become problematic with ~100 or more compounds Difficult to find quiet time in TIC for segment break Too many transitions at once can give unacceptable sensitivity Complex calculations and dwell time decisions 29
30 Auto-Optimized :Timed-SRM Method Overview Acquisition windows centered around retention time 30
31 Timed-SRM Method Overview Acquisition windows allowed to overlap 31
32 Analysis of 300 Pesticides Segmented vs. Timed Segmented SRM Closest compound to segment break: 5 seconds Timed SRM Closest compound to segment break: 15 seconds Average number of simultaneous transitions: 55 Average number of simultaneous transitions: 15 (4X higher dwell times- better sensitivity) 32
33 Bendiocarb Furalaxyl Mefenacet Oxadixyl Triazophos Azinphos-methyl Fenamidone Indoxacarb Zoxamide Benalaxyl Buprofezin Carboxin Diethofencarb Dimethomorph Fenbuconazol Flonicamid Flutolanil Imazalil Propham Propoxur Spiroxamine Methiocarb Thiabendazole Pyridaben Pirimicarb Parts per billion Target a Few Compounds or Screen for Many Pesticides Method LOD 600 Pesticides Method LOD Peak sample frequency flutolanil in 600 method 33
34 The Results: Analytical Performance for Pesticides 34
35 Ease the Method Implementation Process Method implementation Routine result production Dealing with large compound lists GC Performance Maintenance MS/MS method development MS Performance Maintenance Acquisition method optimization Data processing method production Acquisition Method Maintenance Data analysis method maintenance 35
36 Still Need to Maintain Productivity in Result Production Routine result production GC Performance Maintenance MS Performance Maintenance Acquisition Method Maintenance Data analysis method maintenance 36
37 TRACE 1300 Series GC Routine result production GC Performance Maintenance MS Performance Maintenance TRACE 1300 PTV & SSL ic Backflush module to maintain column performance in heavy matrix Instant Connect injector modules can be exchanged faster than a GC column for maintenance offline while your instrument is running Acquisition Method Maintenance Data analysis method maintenance 37
38 Dual MS Heated Zones Dual Independent Heaters Routine result production GC Performance Maintenance MS Performance Maintenance Heat keeps source parts clean TSQ 8000 GC-MS/MS heats all three places where ion burn form: Ion volume Lens stack Ion guide Acquisition Method Maintenance Data analysis method maintenance 38
39 Dual MS Heated Zones Dual Independent Heaters Routine result production GC Performance Maintenance MS Performance Maintenance Heat keeps source parts clean TSQ 8000 heats all three places where ion burn form: Ion volume Lens stack Ion guide Acquisition Method Maintenance Data analysis method maintenance 39
40 Heated RF Lens RF Lens Routine result production GC Performance Maintenance MS Performance Maintenance Never clean or replace quads Heated RF Lens contains ion burn Acquisition Method Maintenance Data analysis method maintenance 40
41 Never Vent to Clean Instrument Routine result production GC Performance Maintenance MS Performance Maintenance Full source removal without venting Spare source allows for non-stop analysis with offline cleaning employed Acquisition Method Maintenance Data analysis method maintenance 41
42 Routine Robustness at the Level of Interest %RSD pesticides (10 ppb) before and after 1,400 vegetable matrix injections Routine result production GC Performance Maintenance MS Performance Maintenance Acquisition Method Maintenance Data analysis method maintenance 42
43 Routine Robustness at the Level of Interest Routine result production GC Performance Maintenance MS Performance Maintenance Acquisition Method Maintenance Data analysis method maintenance 43
44 Method Sync With TraceFinder Software TSQ 8000 GC-MS/MS instrument method Routine result production GC Performance Maintenance MS Performance Maintenance Timed-SRM also allows auto-update of acquisition windows through RT updates in processing method Acquisition Method Maintenance Data analysis method maintenance TraceFinder SW data processing method 44
45 Easy Implementation and Maintenance Method implementation Routine result production Dealing with large compound lists GC Performance Maintenance MS/MS method development MS Performance Maintenance Acquisition method optimization Data processing method production Acquisition Method Maintenance Data analysis method maintenance 45
46 Thank You and Acknowledgement Thanks to Mr. Elmar Häfner, Head of Contaminant Analysis, PhytoLab GmbH & Co. KG, Germany for his support with contributions of pesticide data 46
47 Thank you for your attention Stay connected with us! Chromatography Solutions Blog YouTube Questions? Facebook Solutions Pinterest 47
Simplifying Complex Multi-Residue Pesticide Methodology in GC-MS/MS
Simplifying Complex Multi-Residue Pesticide Methodology in GC-MS/MS David Steiniger, Juan Carmona, Stacy Crain, Paul Silcock, and Jason Cole Thermo Fisher Scientific, Austin, TX, USA Technical Note 10319
More informationConfident Quantitation
Confident Quantitation Robust and reliable performance day after day Thermo Scientific TSQ Quantis triple-stage quadrupole mass spectrometer Experience confident quantitation: TSQ Quantis Triple-Stage
More informationMulti-Residue Pesticide Methodologies. Frans Schoutsen
Multi-Residue Pesticide Methodologies Frans Schoutsen Why Pesticide Analysis? Extensively used in agriculture to control pests & improve yields More than 1000 different active substances used Thoroughly
More informationEvolution Triple Quadrupole GC-MS/MS. Transforming the best GC-MS system into something even better.
Evolution Triple Quadrupole GC-MS/MS Transforming the best GC-MS system into something even better. CHROMTECH EVOLUTION TRIPLE QUADRUPOLE GC-MS/MS Transforming the best GC-MS system into something even
More informationKey Words Pesticides analysis, food safety, TSQ Endura, TraceFinder, MRL, tsrm, residue analysis
Increased Productivity in Pesticide Residue nalysis Quantifying 440 Pesticides Following China GB 2763-2014: The Pesticide Explorer Collection Standard Quantitation Zeming Wu 1, Charles T. Yang 2, Zheng
More informationEvolution3 Triple Quadrupole GC-MS/ Transforming the best GC-MS system into something even better.
Evolution3 Triple Quadrupole GC-MS/ Transforming the best GC-MS system into something even better. CHROMTECH EVOLUTION3 TRIPLE QUADRUPOLE GC-MS/MS Transforming the best GC-MS system into something even
More informationThermo Scientific TSQ Quantiva Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative performance. with unprecedented ease
Thermo Scientific TSQ Quantiva Triple-Stage Quadrupole Mass Spectrometer Extreme quantitative performance with unprecedented ease Unsurpassed quantitation Continual changes in research and regulatory requirements
More informationfor water and beverage analysis
Thermo Scientific EQuan MAX Plus Systems Automated, high-throughput LC-MS solutions for water and beverage analysis Pesticides Pharmaceuticals Personal care products Endocrine disruptors Perfluorinated
More informationThermo Scientific VetDrugs Explorer Collection
Thermo Scientific VetDrugs Explorer Collection Start-to-finish targeted quantitation workflows for today and tomorrow s veterinary drugs analysis Transforming Complex Assays into Simple Workflows The Thermo
More informationPesticide Residues in Produce Analyzed by Targeted MRM HR Full- Scan Acquisition and Processing
Pesticide Residues in Produce Analyzed by Targeted MRM HR Full- Scan Acquisition and Processing MRM HR for Concurrent Quantitation, Library Searching, and High-Confidence ID Confirmation KC Hyland SCIEX,
More informationConfident Quantitation
Confident Quantitation Sensitivity and robustness without compromise Thermo Scientific TSQ Altis triple-stage quadrupole mass spectrometer Experience confident quantitation: TSQ Altis Triple-Stage Quadrupole
More informationSGS Delivering Excellence in Compliant Extractables and Leachables Testing
SGS CASE STUDY SGS Delivering Excellence in Compliant Extractables and Leachables Testing A case study in unknown impurity identification & quantification SGS is a world leading inspection, verification,
More informationLC-MS workflows for emerging contaminants Dwain Cardona Environmental Marketing Manager, Thermo Fisher Scientific
LC-MS workflows for emerging contaminants Dwain Cardona Environmental Marketing Manager, Thermo Fisher Scientific The world leader in serving science Agenda Organic Contaminant Analysis Using LC-MS Workflow
More informationIncreasing Throughput and Efficiency with Exactive LC/MS and Triple Quadrupole LC/MS/MS
Increasing Throughput and Efficiency with Exactive LC/MS and Triple Quadrupole LC/MS/MS Nicholas Duczak Thermo Scientific Annual Mass Spectrometry Users Meeting Somerset, NJ October 12th, 2011 Lead Finding
More informationThermo Scientific TSQ Quantum Access MAX
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific TSQ Quantum Access MAX UNSURPASSED PRICE TO PERFORMANCE Superior
More informationAutomated sample preparation followed by sensitive analysis by GC-MS/MS for environmental contaminants in surface waters
APPLICATION NOTE 10591 Automated sample preparation followed by sensitive analysis by GC-MS/MS for environmental contaminants in surface waters Authors Cees Bijsterbosch, 1 Cedric Wissel, 2 Cristian Cojocariu,
More informationThe Agilent Metabolomics Dynamic MRM Database and Method
The Agilent Metabolomics Dynamic MRM Database and Method White Paper Author Mark Sartain Agilent Technologies, Inc. Santa Clara, California, USA Introduction A major challenge in metabolomics is achieving
More informationPesticide & Environmental Pollutant (P&EP) GC/MS/MS Analyzer 2.0
Pesticide & Environmental Pollutant (P&EP) GC/MS/MS Analyzer 2.0 Get on the Analyzer FAST TRACK Identify Analytes in Complex Matrices Obtain Analytical Results on DAY ONE! -- based on the most comprehensive
More informationMinimize Method Development Time for Pesticide Screening Using LC/MS Application Kit
Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit Dr. Jerry Zweigenbaum Senior Applications Scientist Food Safety Page 1 Agilent 1260 LC SampliQ Extraction and Dispersive
More informationCurrent and future trends in Pesticide Residue analysis. Jim Garvey
Current and future trends in Pesticide Residue analysis Jim Garvey A brief history of Pesticide residue analysis Contents 1. Where we have come from 2. Where we are now 3. Where we are going Where we have
More informationThermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease
Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer Extreme quantitative value with unprecedented ease Unsurpassed value From forensic toxicology to environmental analysis to pharmaceutical
More informationAccurately Identify and Quantify One Hundred Pesticides in a Single GC Run
Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract A selected target compound list of 195 various pesticides
More informationThermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease
Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer Extreme quantitative value with unprecedented ease Unsurpassed value From forensic toxicology to environmental analysis to pharmaceutical
More informationDavid Steiniger, 1 Dwain Cardona, 1 Cristian Cojocariu, 2 Paul Silcock, 2 Alexander Semyonov, 1 Jason Cole 1 1
Using Triple Quadrupole GC-MS in Full Scan, SIM, SRM, and Mixed Scan Modes to Provide the Highest Coverage for Target and Non-target Analysis of Contaminants David Steiniger, 1 Dwain Cardona, 1 Cristian
More informationLC-QTOF-MS/MS method for the simultaneous full scan and MS/MS analysis of pesticides in fruit and vegetables
LC-QTOF-MS/MS method for the simultaneous full scan and MS/MS analysis of pesticides in fruit and vegetables CONTENTS 1. Aim and scope... 1 2. Procedure... 1 2.1. Sample extraction... 1 2.2. Instrumentation
More informationThermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease
Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer Extreme quantitative value with unprecedented ease Unsurpassed value From forensic toxicology to environmental analysis to pharmaceutical
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment API 3000 LC/MS/MS Unprecedented Performance: An advanced LC/MS/MS system
More informationResolve Your Search for More Ions: The 5977B, 7000D, and 7010B GC/MS Instruments. Stephan Baumann ASTS Talk November, 2016
Resolve Your Search for More Ions: The 5977B, 7000D, and 7010B GC/MS Instruments Stephan Baumann ASTS Talk November, 2016 1 The Most Powerful EI GC/MS Source Ever Made. More intense electron beam NEW!
More informationAgilent 6400 Series Triple Quadrupole LC/MS. Breakthrough sensitivity now routine for your lab
Agilent 6400 Series Triple Quadrupole LC/MS Breakthrough sensitivity now routine for your lab Agilent 6400 Series Triple Quadrupole LC/MS Unprecedented sensitivity to improve your results every day Whether
More informationAgilent Automated Card Extraction Dried Blood Spot LC/MS System
Agilent Automated Card Extraction Dried Blood Spot LC/MS System Quick Start Guide What is the AACE DBS LC/MS System? 2 System Overview 3 Operational Overview 4 Method Development 5 Using the AACE DBS LC/MS
More informationIntroducing the new Agilent 7800 Solution-Ready ICP- MS. Agilent Technologies 2015
Introducing the new Agilent 7800 Solution-Ready ICP- MS Agilent Technologies 2015 Agilent s 28 Years of ICP-MS Innovation 8800 ICP-QQQ 7700 World s First (and only) ICP-QQQ 7900 7800 7500 PMS series 4500
More informationDetermination of Pesticide Residue in Vegetables Using the AutoMate-Q40. Application Note. Abstract. Introduction
Application Note Abstract Determination of Pesticide Residue in Vegetables Using the AutoMate-Q40 Page 1 QuEChERS is a Quick-Easy-Cheap-Effective-Rugged-Safe extraction method that has been developed for
More informationAgilent 6000 Series LC/MS Solutions. Confidence in Quantitative and Qualitative Analysis
Agilent 6000 Series LC/MS Solutions Confidence in Quantitative and Qualitative Analysis Agilent 6000 Series LC/MS Solutions SOLUTIONS THAT SET THE STANDARD Your work with Agilent instruments benefits from
More informationAgilent New 6470 Triple Quadrupole LC/MS System
Agilent New 6470 Triple Quadrupole LC/MS System ROCK SOLID Performance For Confident Quantitation Na Pi Parra, Ph.D Product Manager for Triple Quadrupole LC/MS 1 Overview of Topics The New 6470 QQQ LC/MS
More informationNOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778)
NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778) This Notice of Intended Decision to Enter into a Single Source Contract is posted in accordance with section 287.057(3), Florida
More informationAPI 3200 LC/MS/MS system
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment API 3200 LC/MS/MS system Reliability, Reproducibility, and Confidence in Your
More informationQuantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System
Application Note Biotherapeutics Quantification of Host Cell Protein Impurities Using the Agilent 9 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System Authors Linfeng Wu and Yanan Yang
More informationAnalysis of Triazine Herbicides in Drinking Water Using LC-MS/MS and TraceFinder Software
Application Note: 478 Analysis of Triazine Herbicides in Drinking Water Using LC-MS/MS and TraceFinder Software Jonathan R. Beck, Jamie K. Humphries, Louis Maljers, Kristi Akervik, Charles Yang, Dipankar
More informationENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note
ENVIRONMENTAL ANALYSIS A TURN-KEY GUARANTEED ANALYSER FOR THE ROUTINE MEASUREMENT OF GLYPHOSATE AND AMPA IN DRINKING WATER (EUROPEAN DRINKING WATER DIRECTIVE 98/83/EC) AND OTHER ENVIRONMENTAL WATER SAMPLES
More informationSimultaneous Quantitation of a Monoclonal Antibody and Two Proteins in Human Plasma by High Resolution and Accurate Mass Measurements
Simultaneous Quantitation of a Monoclonal Antibody and Two Proteins in Human Plasma by High Resolution and Accurate Mass Measurements Paul-Gerhard Lassahn 1, Kai Scheffler 2, Myriam Demant 3, Nathanael
More informationMultiQuant Software Version 3.0
MultiQuant Software Version 3.0 Improving Data Quality and Processing Throughput with Better Peak Integration, Quantitative and Qualitative Compound Review for the Analysis of Food, Drinking Water, and
More informationTurboMass GC/MS Software
TurboMass GC/MS Software GAS CHROMATOGRAPHY / MASS SPECTROMETRY P R O D U C T N O T E Today s demanding laboratory requires software that is easy to learn yet offers sophisticated instrument control, robust
More informationACCURACY THE FIRST TIME. Reliable testing with performance, productivity and value combined AB SCIEX 3200MD SERIES MASS SPECTROMETERS
ACCURACY THE FIRST TIME Reliable testing with performance, productivity and value combined AB SCIEX 3200MD SERIES MASS SPECTROMETERS Reliability, reproducibility, and confidence in your data Your routine
More informationWaters Pesticide Screening Solutions: Nuove Frontiere nell Analisi di pesticidi Miguel Ángel Cortés, Waters Manuela Cuccurullo, Waters
Waters Pesticide Screening Solutions: Nuove Frontiere nell Analisi di pesticidi Miguel Ángel Cortés, Waters Manuela Cuccurullo, Waters MS Technology day, Milan March 12 2013 Waters Corporation 1 The purpose
More informationSummary results using G7000A GC-QQQ for OCPs, Pesticides and PCBs in Marine Sediment and Biota
Summary results using G7000A GC-QQQ for OCPs, Pesticides and PCBs in Marine Sediment and Biota Paul Zavitsanos Agilent Technologies PITTCON 2009 Chicago G7000A GC-QQQ OCPs and PCBs in Marine Sediment and
More informationWith the globalization of trade, food production
November 2008 Comprehensive Screening, Confirmation, and Quantification of Organic Pesticides in Foods by GC MS and LC MS The global focus on keeping food safe has intensified, resulting in major changes
More informationSCION GC Series. The Gas Chromatographers Choice for Separations
SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Bruker s Scion Instrument s long tradition long of tradition innovation of innovation and product and reliability
More informationANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER
ENVIRONMENTAL ANALYSIS ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER Solutions for Your Analytical Business
More informationAnalysis of Triazine Herbicides in Drinking Water Using LC-MS/MS and TraceFinder Software
Application Note: 478 Analysis of Triazine Herbicides in Drinking Water Using LC-MS/MS and TraceFinder Software Jonathan R. Beck, Jamie K. Humphries, Louis Maljers, Kristi Akervik, Charles Yang, Dipankar
More informationDetermination of Pesticide Residue in Apple Juice Using the AutoMate-Q40
Application Note Abstract QuEChERS is a Quick-Easy-Cheap-Effective-Rugged-Safe extraction method that has been developed for the determination of pesticide residues in agricultural commodities. While the
More informationReal-time point-of-need mass spectrometry detection
Real-time point-of-need mass spectrometry detection Optimise productivity with 4500 MiD 4500 MiD One box solution for MS detection Combining the vacuum system electronics and computer inside one box, our
More informationRobustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety
Technical Overview Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety Authors Mark Sartain, Theresa Sosienski, and Dan Hui Dorothy Yang Agilent Technologies, Inc.
More informationDetermination of Pesticides in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes. Application Note. Author. Abstract.
Determination of Pesticides in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application Note Environmental Author Chin-Kai Meng Agilent Technologies, Inc. 850 Centerville Road Wilmington,
More informationAgilent 6490 Triple Quadrupole LC/MS System REACH ULTIMATE SENSITIVITY WITH BREAKTHROUGH IFUNNEL TECHNOLOGY
Agilent 6490 Triple Quadrupole LC/MS System REACH ULTIMATE SENSITIVITY WITH BREAKTHROUGH IFUNNEL TECHNOLOGY AGILENT 6490 TRIPLE QUADRUPOLE LC/MS SYSTEM WITH IFUNNEL TECHNOLOGY CLEARLY BETTER SENSITIVITY
More informationA Novel Approach to the Reduction of False Positive and Negative Identifications in Screening of Pesticide Residues in Food Analysis
A Novel Approach to the Reduction of False Positive and Negative Identifications in Screening of Pesticide Residues in Food Analysis M McCullagh, Severine Goscinny, Ramesh Rao and Vincent Hanot 2013 Waters
More informationExpand Your Research with Metabolomics and Proteomics. Christine Miller Omics Market Manager ASMS 2017
Expand Your Research with Metabolomics and Proteomics Christine Miller Omics Market Manager ASMS 2017 New Additions to Agilent Omics Workflows Acquisition Ion Mobility Q-TOF IM All Ions MS/MS Find Features
More informationDevelopment of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS
TECHNICAL NOTE Pesticides in Tobacco Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS API 4000 LC/MS/MS System Overview This application note
More informationThe analysis of polar ionic pesticides using ion-exchange chromatography coupled to mass spectrometry: turning negatives into positives
The analysis of polar ionic pesticides using ion-exchange chromatography coupled to mass spectrometry: turning negatives into positives Richard Fussell, Vertical Marketing Manager Food & Beverage 11 th
More informationSCIEX OS Software One Touch Productivity
SCIEX OS Software One Touch Productivity Accuracy, Simplicity and Efficiency What SCIEX OS Software is all About Mass spectrometry and the data challenge Mass spectrometry continues to provide more comprehensive
More informationLeading the way in atomic spectroscopy innovation. AA, MP-AES, ICP-OES, ICP-MS, ICP-QQQ Product Positioning
Leading the way in atomic spectroscopy innovation AA, MP-AES, ICP-OES, ICP-MS, ICP-QQQ Product Positioning Agilent Technology Leadership, the first 55 years 1957 Built components for world's first AA (as
More informationQ40. This automated platform will streamline the two part QuEChERS method from the liquid extraction through the cleanup.
Application Note Abstract With the rise in popularity of QuEChERS, due to its reliable multiresidue methods it has driven the need for automation of the QuEChERS extraction to increase productivity and
More informationIMPORTANCE OF CELL CULTURE MEDIA TO THE BIOPHARMACEUTICAL INDUSTRY
IMPORTANCE OF CELL CULTURE MEDIA TO THE BIOPHARMACEUTICAL INDUSTRY Research, development and production of biopharmaceuticals is growing rapidly, thanks to the increase of novel therapeutics and biosimilar
More informationApplication Note. Abstract. Authors. Environmental
Maintaining Compound Retention Times with the Backflush Enabled Pressure Controlled Tee Configuration for Agilent 7890A GCs with Agilent 5975 Series MSD and Agilent 7000 Series Triple Quadrupole MS Systems
More informationA Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis
A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis Sebastian Westrup, Nick Duczak, and Michal Godula Thermo Fisher
More informationTarget and Non-target Pesticide Residue Analysis Using GC/QQQ and GC/Q-TOF with a Discussion on Ways to Use Sandwich Injections
Target and Non-target Pesticide Residue Analysis Using GC/QQQ and GC/Q-TOF with a Discussion on Ways to Use Sandwich Injections Philip L. Wylie Agilent Technologies Wilmington, DE 1 Topics Target vs. Non-target
More informationSimple Run Table Entry Cuts Sample onto Column 2. Method Developers Calculator to Tools correctly set flows and restrictor size
Introduction 2-D GC/MS and the Deans Switch: Theory and Applications Simplified 2-D GC can be used to accomplish the separation of specific analytes from very complex matrices. In complex sample matrices,
More informationSCION GC Series. The Gas Chromatographers Choice for Separations
SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation
More informationSCION GC Series. The Gas Chromatographers Choice for Separations
SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Scion Instrument s tradition of innovation and product reliability have combined to create the next generation
More informationTyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1
Application Note Method Validation of the AutoMate-Q40 for Pesticide Residues in Grapes Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1 Abstract The QuEChERS (Quick-Easy-Cheap-Effective-Rugged-Safe)
More informationOlefin Analyzer Solutions for Polymer Grade Ethylene and Propylene
Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene Angela Smith Henry Kelly Beard George Gonzalez 1 Impurity Monitoring in Olefin Production Analysis of contaminants in ethylene and propylene
More informationAnalysis of all pesticide residues in cannabis regulated by California and Oregon state using LC/MS/MS with Electrospray and APCI source
Analysis of all pesticide residues in cannabis regulated by California and Oregon state using LC/MS/MS with Electrospray and APCI source Avinash Dalmia July 10 th 2018 NACRW 2018 PerkinElmer Challenges
More informationTargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet
TargetQuan 3 Software Leading the way in regulatory POPs quantification Bullet Bullet Bullet Leading the way in regulatory POPs quantification Analyse Samples Open Sequence Process Sequence Analyse Samples
More informationAgilent solutions for contract research and manufacturing organizations
Agilent solutions for contract research and manufacturing organizations Cooperation between pharmaceutical companies and contract research organizations A successful approach to increasing efficiency along
More informationAgilent 6400 Series. System. Concepts Guide. The Big Picture. Agilent Technologies
Agilent 6400 Series Triple Quad LC/MS System Concepts Guide The Big Picture Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means
More informationPlease enclose a item wise compliance report along with the technical bid.
Amendments to Tender Notification for the procurement of GC and MSD with TDU at the Indian Institute of Science, Bangalore (Last date of submission of tenders: 20-MARCH-2017) REF: CES/RMB/2016-17 Dear
More informationThermo Scientific TSQ Quantum XLS Ultra Triple Quadrupole GC-MS
c h r o m a t o g r a p h y Thermo Scientific TSQ Quantum XLS Ultra Triple Quadrupole GC-MS Introducing Ultra Selective GC-MS/MS Persistent Organic Pollutants Food Safety Drug Screening Forensic Evidence
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment QTRAP 5500
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment QTRAP 5500 Powerful synergy results when the world s most sensitive triple
More informationAB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual
AB SCIEX TripleTOF 5600 SYSTEM High-resolution quant and qual AB SCIEX TripleTOF 5600 SYSTEM AB SCIEX is a global leader in the development of life science analytical technologies that help answer complex
More informationUltivo Triple Quadrupole Mass Spectrometer UNBELIEVABLY POWERFUL REMARKABLY SMALL
Ultivo Triple Quadrupole Mass Spectrometer UNBELIEVABLY POWERFUL REMARKABLY SMALL TRIPLE QUADRUPOLE MASS SPECTROMETER MEET ULTIVO IT S A PINT-SIZED POWERHOUSE WELCOME TO A NEW ERA OF MASS SPECTROMETRY
More informationTransitioning to confident quantitation in search of a better tomorrow
SMART NOTE Transitioning to confident quantitation in search of a better tomorrow What is quantitation and why do we need it? For one to identify and understand any unknown substance, the first step is
More informationthermo scientific Find out more at thermofisher.com/lims TECHNICAL INFORMATION BULLETIN
thermo scientific TECHNICAL INFORMATION BULLETIN Thermo Scientific SampleManager LIMS and Thermo Scientific Chromeleon CDS Link Key Words Integrated LIMS and CDS, Secure & Compliant Interface, Workflows
More informationMRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics
Product Bulletin MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics With Unique QTRAP System Technology Overview Targeted peptide quantitation is a rapidly growing
More informationWorld-leading LC/MS/MS technology for quantitation and identification API LC/MS/MS System
World-leading LC/MS/MS technology for quantitation and identification API 3000 LC/MS/MS System rel Accelerate your drug development cycle The API 3000 LC/MS/MS System leads the field in sensitivity, specificity,
More information[ VION IMS QTOF ] BEYOND RESOLUTION
[ VION IMS QTOF ] BEYOND RESOLUTION THE BENEFITS ARE CLEAR and routinely available, Complex samples give complex data with overlapping spectra and background interferences, making compound identification
More informationMethods for Increased Sample Throughput with Dirty Samples
Methods for Increased Sample Throughput with Dirty Samples Alan D Broske, Limian Zhao, and William H. Wilson Agilent Technologies, Inc. 1 Introduction GC inlet liner impacts sample transfer to the GC column
More informationDetermination of Persistent Organic Pollutants (POPs) in Fish
Faculty of Veterinary Medicine Determination of Persistent Organic Pollutants (POPs) in Fish Dipartimento di Scienze Veterinarie e Sanità Pubblica (DIVET) Prof. L.Chiesa, Dr. S. Panseri -Prague, April
More informationACCURACY THE FIRST TIME. Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS
ACCURACY THE FIRST TIME Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS Your routine diagnostic testing lab requires accurate trace level quantitation of multiple compounds
More informationAgilent Intuvo 9000 GC System DREAM BIGGER
Agilent Intuvo 9000 GC System DREAM BIGGER AGILENT INTUVO 9000 GC SYSTEM INNOVATION WITH PURPOSE Truly transformative innovation is much more than a clever idea made real. It must give users the power
More informationAgilent 500 Ion Trap LC/MS with the Agilent 1200 Infinity Series
Agilent 500 Ion Trap LC/MS with the Agilent 1200 Infinity Series Quick Start Guide Step 1. Set up and learn about the system 3 Step 2. Create method and Sequence Sample List 4 Step 3. Start the Sample
More informationImproving Productivity in Your Environmental and Food Testing Labs with Agilent 7890A/5975C GC/MS System
Improving Productivity in Your Environmental and Food Testing Labs with Agilent 789A/5975C GC/MS System Mike Szelewski Senior Applications Chemist Page 1 Overview Capillary Flow Devices A family of tools
More informationLiquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use
Liquid Chromatograph Liquid Chromatograph Mass Spectrometer CL Series for In Vitro Diagnostic Use Consistency. Performance. Reliability. Shimadzu s CL Series LC, LC-MS, and LC-MS/MS for in vitro diagnostic
More informationA Fast, Robust Approach to Measuring Pesticides and Mycotoxins in Dry Cannabis Flower and Concentrates
A Fast, Robust Approach to Measuring Pesticides and Mycotoxins in Dry Cannabis Flower and Concentrates Combine triple quadrupole LC/MS and GC/MS analysis for up to 97 pesticides and five mycotoxins regulated
More informationJonathan R. Beck and Charles T. Yang; Thermo Fisher Scientific, San Jose, CA
EPA Draft Method 543 Quantitation of Organic Pesticides in Drinking Water Using Online Pre-concentration/Solid Phase Extraction and Tandem Mass Spectrometry Jonathan R. Beck and Charles T. Yang; Thermo
More informationAgilent 6430 Triple Quadrupole LC/MS System
Agilent 6430 Triple Quadrupole LC/MS System Ideal Quantitative LC/MS for UHPLC with Dynamic MRM and Fast Polarity Switching, plus Unsurpassed Sensitivity with Chip LC Summary The Agilent 6430 Triple Quadrupole
More informationWaters QUATTRO PREMIER XE
Www.IetLtd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Waters QUATTRO PREMIER XE Looking for a more intelligent way to overcome these
More informationSample Matrix Influence on GC/MS/MS Multiresidue Pesticide Analysis
Application Note Food Testing & Agriculture Sample Matrix Influence on GC/MS/MS Multiresidue Pesticide Analysis Author Jessica Westland Agilent Technologies, Inc. Abstract The complexity of pesticide analysis
More informationClearly better LC/MS solutions
Agilent 6000 Series LC/MS Clearly better LC/MS solutions Our measure is your success. products applications soft ware services Agilent: the clearly better choice for qualitative and quantitative LC/MS.
More informationChemical Analysis. DHA Analyzer Family. Optimized Solutions for Detailed Hydrocarbon Analysis. Gas Chromatography. think forward
Chemical Analysis DHA Analyzer Family Optimized Solutions for Detailed Hydrocarbon Analysis think forward Gas Chromatography The DHA Analyzer is a complete high resolution gas chromatography solution for
More informationGC-MS (Multi-Residue) Screening for Pesticides; Emerging Organic Pollutants (EOP's) and Persistent Organic Pollutants (POP's).
GC-MS (Multi-Residue) Screening for Pesticides; Emerging Organic Pollutants (EOP's) and Persistent Organic Pollutants (POP's). The GC-MS Target Based Screening methods allow for virtually all GC-amenable
More informationThermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.
Thermo Scientific Peptide Mapping Workflows Upgrade Your Maps Fast, confident and more reliable peptide mapping. Smarter Navigation... Peptide mapping is a core analytic in biotherapeutic development.
More information