New Requirements of ISO 17664/AAMI ST 81

Size: px
Start display at page:

Download "New Requirements of ISO 17664/AAMI ST 81"

Transcription

1 New Requirements of ISO 17664/AAMI ST 81 How the instrument manufacturer validate the cleaning process Klaus Roth SMP GmbH Tübingen Germany SMP GmbH

2 Operation, which have been cancelled due to not sufficient reprocessed instruments Year Data of 57 hospitals** Estimated for England and Wales ** 57/340 Datensätze * mit freundl. Genehmigung G.Shapp MP SMP GmbH

3 SMP GmbH

4 ISO (2004) Preparation at the point of use Preparation, cleaning, disinfection Drying Inspection, maintenance and testing Packaging Sterilisation Storage SMP GmbH

5 ISO (2004) SMP GmbH

6 Instrument circulation in a hospital Use Available for use Storage/transport Storage Cleaning/disinfection Sterilisation Service and maintenance Functional testing SMP GmbH

7 ISO (2004) 3.5 Cleaning A validated method of cleaning shall be specified. At least one validated automated method using a washer-disinfector shall also be specified unless the medical device cannot withstand any such process, in which case a warning should be issued. Where appropriate, at least the following information shall be included:.accessories required for cleaning process;.identification and concentration of chemicals required for cleaning;.identification of water quality,.limits and monitoring of chemical residues.limits on temperature, concentration of solution(s), exposure time,.process temperature(s);.techniques to be used including rinsing; SMP GmbH

8 Standard EN/ISO 17664: Chapter 5 allows the building of groups of instruments: Where the manufacturer supplies a number of different medical devices which share common features and attributes, the validation specified may be performed with respect to these medical devices as a group of family, provided that the manufacturer can demonstrate the commonality of the medical devices and that the tests and assessments address the worst case feature or attribute of the group of family SMP GmbH

9 ISO (new) Scope 1.1 Inclusions This International Standard specifies requirements for the information to be provided by the medical device manufacturer for the processing of a medical device that requires cleaning followed by disinfection and/or sterilization to ensure that the device is safe and effective for its intended use. This includes information for processing prior to use or reuse of the medical device. The provisions of this standard are applicable for medical devices that are intended for invasive or other direct patient contact. Processing instructions are not defined in this standard. Rather, this International Standard specifies requirements to assist manufacturers of medical devices in providing detailed instructions for processing that consists of the following activities where applicable: SMP GmbH

10 ISO (new) 1.1 Inclusions a) Pre-treatment at the point of use before processing; b) Preparation before cleaning, c) Cleaning, d) Disinfection; e) Drying; f) Inspection, maintenance and testing; g) Packaging; h) Sterilization; i) Storage j) Transportation. 1.2 This standard excludes: noncritical medical devices not intended for direct patient contact; Textile devices used in patient draping systems or surgical clothing; processing of medical devices specified by the manufacturer for single-use only and supplied sterile SMP GmbH

11 ISO (new) SMP GmbH

12 ISO (new) SMP GmbH

13 ISO (new) If the medical device is intended to be disinfected a validated method of manual disinfection Shall be specified if automated disinfection is not possible SMP GmbH

14 ISO (new) SMP GmbH

15 SMP GmbH

16 Categories of instruments and specific requirements for use: Instruments, to be used outside the body. Instruments, to be used inside the body without penetration of the mucosa or skin. Instruments, penetrating the skin or mucosa. Clean / Disinfected High Level Disinfected Sterile Spaulding Classification by the CDC SMP GmbH

17 SMP GmbH

18 Classification of the instruments groups Group 1: Group 2: Group 3: Group 4: Group 5: Group 6: Group 7: Critical Instruments, like hooks; no hidden surfaces and lumens Critical Instruments Scissors, Clamps Shift shaft instruments Rongeur etc. Shaft instruments for MIS need validation, as the result of the cleaning can not be inspected Micro surgical Instruments need validation, as the result of the cleaning can not be inspected Complex Devices has to be tested, as no analogical conclusions can be made Flexible Instruments need validation, as the result of the cleaning can not be inspected SMP GmbH

19 Comparison of Group 2 Instruments Crile Clamp, 1. Messung nach Kontaminati on 2. Messung nach Vorreinigung + 10 min 40 C Einweichen 4. Messung nach Reinigung 1. Messung nach Kontaminati on 2. Messung nach Vorreinigung + Ultraschall 10 min 40 C 3. Messung nach Reinigung Nr.87/26 Zr Zr Zr Zr Zr Zr Crile clamp Crile clamp Crile clamp SMP GmbH

20 Group 2 (Instruments with hidden surfaces): Side cutter 1. Messung nach Kontaminati on 2. Messung nach Vorreinigung + 10 min 40 C Einweichen 3. Messung nach Reinigung 1. Messung nach Kontaminati on 2. Messung nach Vorreinigung + Ultraschall 10 min 40 C 3. Messung nach Reinigung Nr.87/26 Zr Zr Zr Zr Zr Zr Wire cutter Wire cutter Wire cutter SMP GmbH

21 Group 3: Shift shaft instruments Yes 1 Yes 2 Yes 3 Tischler-Morgan (Not dismountable) Step 1 Yes 1 Yes 2 Yes 3 No Step 2 Yes 1 Yes 2 Yes 3 No Step 3 No Validation not successfull SMP GmbH

22 Group 3 (Shift shaft instruments): Rongeur,Arthroskopiezangen etc. 1. Messung nach Kontaminati on 2. Messung nach Vorreinigung + 10 min 40 C Einweichen 4. Messung nach Reinigung 1. Messung nach Kontaminati on 2. Messung nach Vorreinigung + Ultraschall 10 min 40 C 3. Messung nach Reinigung Nr.87/26 Zr Zr Zr Zr Zr Zr Rongeur Rongeur Rongeur SMP GmbH

23 SMP GmbH

24 Group 3: Shift shaft instruments (dismantable) Yes 1 Yes 2 Yes 3 Bronchoscopy Forceps (Dismountable) Step 1 Yes 1 Yes 2 Yes 3 No Step 2 Yes 1 Yes 2 Yes 3 No Step 3 No Validation not successfull SMP GmbH

25 Group 3: Shift shaft instruments (improved design) SMP GmbH

26 Instruments for Minimally invasive surgery Dismountable Not dismountable with Luer adapter SMP GmbH

27 Parameters of the Cleaning Cycle Interacting Time Mechanics: Amount of Water Pressure Ultrasound Temperature Chemical Agent Water Quality SMP GmbH

28 MIS Instruments dismountable Results: Remaining contamination in the shaft Manual pre-cleaning required Anzahl der Instrumrnte unter 5 Counts/sec Anzahl der Instrumrnte über 5 Counts/sec 100% 100% Hunderte 73% 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% Miele 7836 Oxivario Gesamt Anteill der Ins. 11 0% Miele 7836 Tw /Tb Gesamt Anzahl der Ins. 11 0% 73% Miele 7735 Alkalisch Gesamt Anteil der Ins.11 27% 27% Miele 7735 Enzymatisch Gesamt Anteil der Ins SMP GmbH

29 ISO (new) Chemistry SMP GmbH

30 MIS Instruments, not dismountable with Luer adapter Results: Remaining contamination in the shaft Anzahl der Instrumrnte unter 5 Counts/sec Anzahl der Instrumrnte über 5 Counts/sec Hunderte 100% 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 100% Niagara Enzymatisch Gesammt Anteil der Ins.19 0% 95% Miele 7836 Oxivario Gesamt Anteill der Ins. 19 Manual pre-cleaning required 5% 84% Miele 7836 Tw /Tb Gesamt Anzahl der Ins % 48% 52% Miele 7735 Alkalisch Gesamt Anteil der Ins.19 42% Miele 7735 Enzymatisch Gesamt Anteil der Ins.19 58% SMP GmbH

31 Effort for cleaning Validation not successfull Validation not successfull Effort for cleaning Influence of the detergent Group 2: Alkaline Step 3 Step 2 Step 1 2 A Titan 2 A Ceramic 2 B 2 C 2 D 2 E 2 F Instruments sorted by category Tab. 58: Zusammengefasste Ergebnisse Gruppe 2 Alkalisch maschinell und ggf. manuellen Vorreinigung Group 2: Enzymatic automated and manual if necessary Step 3 Step 2 Step 1 2 A Titan 2 A Ceramic 2 B 2 C 2 D 2 E 2 F Instruments sorted by category Tab. 59: Zusammengefasste Ergebnisse Gruppe 2 Enzymatisch maschinell und ggf. manuelle Vorreinigung SMP GmbH

32 Power of cleaning processes Multicomponent Process Power Flush Multicomponent Enzymatic process Alcaline process Enzymatic process SMP GmbH

33 Parameters of the Cleaning Cycle Mechanics: Amount of Water Pressure Ultrasound Interacting Time Temperature Chemical Agent Water Quality SMP GmbH

34 Validation of the cleaning step Monitoring of water pressure SMP GmbH

35 Foaming SMP GmbH

36 SMP GmbH

37 Vario SMP GmbH

38 Influence of Velocity of Flow Laminar Flow SMP GmbH

39 Influence of Velocity of Flow Turbulent Flow R e r v l m u Reynolds number Density Velocity Length Viscosity kin. Viscosity R e = l r v m SMP GmbH

40 Test procedure: OPA-Testing according to ISO Improved Cycle of the Medisafe PCF with High Pressure Ultrasonic Irrigation Pre-Wash internal & 3 bar Detergent Dispense De-gas 15 minute Ultrasonic Main Wash internal/external 3 bar Pre-Rinse internal & 3 bar Final Rinse Empty each cycle SMP GmbH

41 Guidelines and legal Requirements SMP GmbH

42 Cleaning: Contamination of the artery clamps with 100µl coagulable sheep blood Investigation of Cleaning Performance Following the Standard pren/iso Zentr Steril 2005; 13: SMP GmbH

43 SMP GmbH Validation of Washer/Disinfectors (Cleaning tests) PCD: Vacuum package Real contamination

44 SDS-Elution for OPA-testing SMP GmbH

45 Results of the round robin test 2004 Charge I 1 x und Charge II 2 x = 100 Charge III 1 x = 100 Charge I 1 x = 100 Charge II 1 x = SMP GmbH

46 First results of the year 2005 of automated reprocessing In order to implement the up and coming EN ISO in practice, we worked out the Guidelines of the DGKH, DGSV and AKI for validating and routinely monitoring automatic washing and disinfecting processes for thermo-stable medical devices, and principles for selecting a washer-disinfector (WD) Part 1**. In an inter-hospital trial conducted with 18 hospitals, the practicability and meaningfulness of the test method was investigated. 5 of 18 WD involved in the trial were over threshold. Although the sensitivity of the semi-quantitative Biuret / BCA method used in the tests is fairly low, it can be presumed that the evaluation framework is sufficient for the moment to identify WDs with poor cleaning quality, and to monitor the quality of the measures carried out here. Inter-hospital trials to determine minimal cleaning performance according to the Guidelines from DGKH, DGSV and AKI; Zentr Steril 2005; SMP GmbH

47 Summary of 10 years validation Remaining protein on test instruments (4122) in 2012 Restprotein an Prüfinstrumenten, Anzahl: % 2% 20% unwg 50µg µg µg 73% SMP GmbH

48 Summary of 10 years validation Restprotein an Realinstrumenten, Anzahl: 3780 Remaining protein on real used instruments (3780) in % 6% 18% 64% unwg 50µg µg µg SMP GmbH

49 Summary of 10 years validation Remaining protein on real used MIC instruments (786) in SMP GmbH

50 Protein detection methods Bradford-Method Modified OPA-Method Biuret-Reaction BCA-Protein Assay Kit Ninhydrin-Reaction TOC (Total organic Carbon) ATP Combur 3Test HemoCheck-S SMP GmbH

51 Protein detection methods 0, ,1 Protein: 16 µg/ml 160 µg/ml 1,6 mg/ml 160 mg/ml SMP GmbH

52 SMP GmbH HemoCheck-S Test Kit for a qualitative determination blood on surgical instruments 100µg 1µg 0µg

53 SMP GmbH

54 Dilution of Blood Full blood Visual detection of blood is possible up to 1: SMP GmbH

55 Combur 3 Test SMP GmbH

56 SMP GmbH Combur 3 Test Hemoglobin katalysed the Oxidation of the indicator by organic H 2 O 2 Very sensitive Detection limit at 0,3 µg/ml Extraction of the protein from the instrument can be performed with SDS

57 Criteria for selection of the test method Method Sensitivit y Recovery Training Costs Results Bradford OPA Biuret BCA Ninhydri n TOC ATP Hemo check Combur test SMP GmbH

58 SMP GmbH

59 SMP GmbH

60 Recovery of protein: Group 1 instruments SMP GmbH

61 Recovery of protein Group 2 instruments SMP GmbH

62 Elution of protein: Group 3 instruments SMP GmbH

63 Elution of protein Group 4 instruments SMP GmbH

64 Qualification of the recovery SMP GmbH

65 Products: ADVICE: Reprocessing Instructions Endoscopic Take-Apart Instrument / Company:... Reprocessing procedures have only limited implications to a surgical instrument. The limitation of the numbers of reprocessing procedures is therefore determined by the function / wear of the device. In case of damage the device should be reprocessed before sending back to the manufacturer for repair. Preparation at the Point of Use: Transportation: Remove gross soiling by submerge the instrument into cold water (<40 C) immediately after use. Don t use a fixating detergent or hot water (>40 C) as this can cause the fixation of residua which may influence the result of the reprocessing process. Safe storage and transportation to the reprocessing area to avoid any damage and contamination to the environment. Preparation for The devices must be reprocessed in a disassembled state. Decontamination: Pre-Cleaning: Warning: Do not allow the instruments to rest on the bottom of an ultrasonic cleaner unit during cleaning, as damage or incomplete cleaning could result. 10 minutes at 40 C in an ultrasonic bath with 0,5% detergent. Brushing the instrument under running tap water until all visible residues are removed Flushing the inner lumens of all parts with a water jet pistol (pressure min. 3 bar) with cold tap water for at least 10 seconds. Cleaning: Manual Cleaning Process: Automated Cleaning: 1. Rinsing under running tap water (<40 C) Connect the instrument to a rack for MIS-instruments until all visible soil has been removed. If and start the program needed a soft bristle brush should be used to 4 min pre-washing with cold water (<40 C); remove visible soil; 6 min washing with 0,5% detergent at 55 C; 2. Submerge instruments in an detergent (if 3 min neutralising with warm water (>40 C); ultrasonic bath is used, ultrasonic process of 2 min intermediate rinsing with warm water 3 minutes and ultrasonic frequency of 35 khz (>40 C). have been shown to be effective). Follow the Special instructions of the manufacturer of the instructions of the manufacturer of the automated washing machine have to be followed. detergent; 3. Rinse the instrument under running tap water to remove the detergent SMP GmbH

66 Disinfection: Drying: Manual Disinfection: 1. Submerge instruments in an disinfection detergent according to the instructions of the manufacturer of the detergent; 2. Rinse the instrument with sterile water to remove the detergent. Manual Drying: Dry the instrument with a lint free towel. The instrument may never be heated up >140 C. To avoid water residues we recommend using sterile compressed air to insufflate cavities. Automated Disinfection: Automated Thermal Disinfection in washer/disinfector under consideration of national requirements in regards to A0-Value (see EN 15883) Automated Drying: Drying of outside of instrument through drying cycle of washer/disinfector. If needed, additional manual drying can be performed through lint free towel. Insufflate cavities of instruments by using sterile compressed air. Functional Testing, Maintenance: Packaging: Functional testing, if available according to instructions of use and visual inspection for cleanliness. If necessary perform reprocessing process again until instrument is visibly clean. Appropriate packaging for sterilization. Sterilization: Sterilization of instruments by applying a fractionated pre-vacuum process (according DIN EN 554 / ISO 11134) under consideration of the respective country requirements. Storage: Reprocessing validation study information Parameters for the pre-vacuum cycle: 3 prevacuum phases with at least 60 milli bar Heat up to a minimum sterilization temperature of C Minimum Holding time: 3,5 min Drying time: minimum 10 min Flash sterilization is not allowed on lumen instruments! Storage of sterilized instruments in a dry, clean and dust free environment at modest temperatures of 5 C to 40 C. The following testing test devices, materials & machines have been used in this valdition study; Detergent: deconnex 28 Alka One, (Borer, Zuchwil, Switzerland) deconnex 23 Neutrazym, (Borer, Zuchwil, Switzerland) Washer / Disinfector: Miele 7735 CD Instrument Rack: Miele E450-1 Details: See report SMP SMP GmbH

67 Conclusion ISO is demanding more information from the manufacturer for the user ISO is requiring qualified equipment according to the standard and validated processes Methods for the evaluation and quality control of cleaning processes has already been developed and establish in some countries. They have led to an improvement of the outcome of the cleaning process Sensitive test methods have to be validated when applied in the daily routine SMP GmbH

68 Thank you SMP GmbH Hechingerstrasse Tübingen SMP GmbH

Best Practices for Reprocessing Robotic Instruments

Best Practices for Reprocessing Robotic Instruments Best Practices for Reprocessing Robotic Instruments Nupur Jain Manager, Validation Engineering Services Intuitive Surgical, US Agenda 1. Background on robotic instruments 2. Reprocessing procedure for

More information

Multi-Dimensional Study of Cleaning Efficacy in a Hospital CSSD

Multi-Dimensional Study of Cleaning Efficacy in a Hospital CSSD Multi-Dimensional Study of Cleaning Efficacy in a Hospital CSSD Dr. Jonathan A. Wilder 1*, David M. Jagrosse 2, Donna Swenson 1, Brian J. Fortier 1, Klaus Roth 1, Prof.-Dr. Ludger Schnieder 1 1, Richmond,

More information

Instructions for the Use Surgical Instruments

Instructions for the Use Surgical Instruments Instructions for the Use Surgical Instruments Geister Medizintechnik GmbH Föhrenstraße 2 78532 Tuttlingen / Germany Tel.: +49-7461-96624-0 FAX: +49-7461-96624-22 E-Mail: info@geister.com 0123 First use

More information

What Does Validation Involve?

What Does Validation Involve? Practical instructions for the Validation of cleaning and disinfection methods according to the recommendation of the German Society for Hospital Hygiene (DGKH) and according to the pren ISO 15883. Laws,

More information

Instruction Guide Re-Usable Surgical Instruments. Instruction Guide Re-usable surgical instruments (GA 30.1)

Instruction Guide Re-Usable Surgical Instruments. Instruction Guide Re-usable surgical instruments (GA 30.1) Instruction Guide Re-usable surgical instruments (GA 30.1) Part A: General Information 1. Field of application 2. Basics 3. General 4. Construction and Material 5. Compatibility 6. Labelling label symbols

More information

Entellus Medical Sinuscope INSTRUCTIONS FOR USE

Entellus Medical Sinuscope INSTRUCTIONS FOR USE Entellus Medical Sinuscope INSTRUCTIONS FOR USE Table of contents Entellus Medical Sinuscope Instructions for Use 1. About this document... 1 1.1 Purpose... 1 1.2 Symbols used... 1 2. Intended use... 2

More information

How is Sterilization Validation Performed by the Medical Device Manufacturer?

How is Sterilization Validation Performed by the Medical Device Manufacturer? How is Sterilization Validation Performed by the Medical Device Manufacturer? Rod Parker, Ph.D. Sr. Principal Scientist Stryker Instruments Division May 9, 2016 DISCLOSURE I am an employee of Stryker Corporation.

More information

Endoscope Washer Disinfector Cleaning Process Challenge Devices Research & Validation

Endoscope Washer Disinfector Cleaning Process Challenge Devices Research & Validation Endoscope Washer Disinfector Cleaning Process Challenge Devices Research & Validation Richard Bancroft, B.Sc (Hons), FRSB Science & Technical Director STERIS Corporation Registered Authorising Engineer

More information

Lotus. Reprocessing instructions for reusable instrumentation

Lotus. Reprocessing instructions for reusable instrumentation Lotus Reprocessing instructions for reusable instrumentation Reprocessing instructions These instructions relate to the Lotus Transducers & Waveguides only. Following use remove the Single Use Handpiece

More information

Lotus. Reprocessing instructions for reusable instrumentation. French/Swiss protocol

Lotus. Reprocessing instructions for reusable instrumentation. French/Swiss protocol Lotus Reprocessing instructions for reusable instrumentation French/Swiss protocol Reprocessing instructions These instructions relate to the Lotus Transducers & Waveguides only. Following use remove the

More information

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 05/2018 PURPOSE: To test for detection of blood residue on

More information

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue on

More information

The Significance of Validation to US CSSD Managers

The Significance of Validation to US CSSD Managers The Significance of Validation to US CSSD Managers Jonathan A. Wilder, Ph.D. President, H & W Technology, LLC/Stericert Co. Managing Director, Quality Processing Resource Group, LLC 1 Introduction This

More information

Loaner Program Guidance

Loaner Program Guidance Loaner Program Guidance This document is valid only on the date printed. If unsure of the print date, please re-print to ensure use of the latest revision of this document (available at www.depuysynthes.com).

More information

Guide for Cleaning, Sterilization and Storage of Reusable Introducers

Guide for Cleaning, Sterilization and Storage of Reusable Introducers Guide for Cleaning, Sterilization and Storage of Reusable Introducers M Manufactured by: Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 U.S. Toll Free: 866-4-CALDERA Telephone: 818-879-6555

More information

Competency Guide: Care and Handling of Rigid Endoscopes

Competency Guide: Care and Handling of Rigid Endoscopes Competency Guide: Care and Handling of Rigid Endoscopes This document details the basic steps to correctly process and maintain a rigid endoscope. This document can be used to perform a competency review

More information

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization)

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization) Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization) Date: Office Name: Office Number: Responsible Physician: Assessor: M.1. General Yes No N/A If No Please Comment

More information

Loaner Programme Guidance

Loaner Programme Guidance Loaner Programme Guidance This document is valid only on the date of printing. In case of doubt about the date of printing, please re-print to ensure that the latest revised version of this document is

More information

Loaner Programme Guidance

Loaner Programme Guidance Loaner Programme Guidance This document is valid only on the date of printing. In case of doubt about the date of printing, please re-print to ensure that the latest revised version of this document is

More information

pump pressure conveying quantity chamber geometry flow conditions Washing arms trays water quantity dosing temperature foam performance

pump pressure conveying quantity chamber geometry flow conditions Washing arms trays water quantity dosing temperature foam performance Validation of cleaning and disinfecting processes in WDs: Implementation of the ÖGSV- Guideline M. Gehrer, T. Miorini, V. Buchrieser, A. Gruber, G. Maierl Processing of instruments preferably automated

More information

Getinge da Vinci Solution

Getinge da Vinci Solution Getinge da Vinci Solution A comprehensive cleaning solution for da Vinci EndoWrist instruments This document is intended to provide information to an international audience outside of the US. Validated,

More information

Surgical Instruments - Instructions for use

Surgical Instruments - Instructions for use 1. General Notice Before using the instruments for the first time, as well as before each subsequent use and before returning them for repair or servicing, they must be cleaned, disinfected and sterilized

More information

The processing and cleaning of instruments provided by the company Weber Instrumente GmbH & Co. KG

The processing and cleaning of instruments provided by the company Weber Instrumente GmbH & Co. KG The processing and cleaning of instruments provided by the company Weber Instrumente GmbH & Co. KG SOP Contents 1 Preface... 1 2 Further applicable norms, directives and additional information... 1 3 General

More information

Technical Data Monograph. The Instrument Protection Properties of Prolystica 2X Concentrate Neutral Detergent Versus Other Neutral Detergent Products

Technical Data Monograph. The Instrument Protection Properties of Prolystica 2X Concentrate Neutral Detergent Versus Other Neutral Detergent Products Technical Data Monograph The Instrument Protection Properties of Prolystica 2X Concentrate Neutral Detergent Versus Other Neutral Detergent Products Table of Contents 1. Background.....................................................1

More information

Manner of Assessment Code (check all that apply) & Surveyor Notes. Yes N/A. Yes. Yes N/A N/A. Yes N/A. N/A Yes N/A

Manner of Assessment Code (check all that apply) & Surveyor Notes. Yes N/A. Yes. Yes N/A N/A. Yes N/A. N/A Yes N/A Section 4. L Reprocessing of Semi-Critical Equipment Elements to be assessed Manner of Assessment Code (check all that apply) & Surveyor Notes Manner of Assessment Code (check all that apply) & Surveyor

More information

Monitoring Flexible Endoscope Channels to Assure Cleaning

Monitoring Flexible Endoscope Channels to Assure Cleaning Monitoring Flexible Endoscope Channels to Assure Cleaning Dr. Michelle J. Alfa, Ph.D., FCCM Medical Director, Clinical Microbiology, Diagnostic Services of Manitoba, Winnipeg, Canada Disclaimers: Dr. M.

More information

In use test of a Washer-disinfector reprocessing of flexible endoscopes

In use test of a Washer-disinfector reprocessing of flexible endoscopes of a Washer-disinfector reprocessing of flexible endoscopes BHT INNOVA E4 Clean Endoscope Copenhagen Hospital Corporation Denmark of INNOVA E4 Washer Disinfector from BHT, Germany. An aldehyde free thermo-chemical

More information

The efficacy of cleaning surgical instruments

The efficacy of cleaning surgical instruments 26 ORIGINAL ARTICLE Central Service 1/2011 Investigations into reproducible cleaning of instruments based on a worst-case model Project Group Cleaning (PGR) Contact: G. Kirmse The efficacy of cleaning

More information

Instruction of use. Annastr.25/1 D Fridingen. Reusable surgical Instruments (Steel, Titanium, TITANIT and antimagnetic Titanium Instruments)

Instruction of use. Annastr.25/1 D Fridingen. Reusable surgical Instruments (Steel, Titanium, TITANIT and antimagnetic Titanium Instruments) You get with the purchase this instrument and it is a hi quality product, its proper handling and use are described below. To keep hazards for patients and users as possible, we ask you to follow the instructions

More information

Best Practices for Reprocessing Endoscopes

Best Practices for Reprocessing Endoscopes Best Practices for Reprocessing Endoscopes Reprocessing of Endoscopes Reprocessing Room Standards The resources referenced include: Canadian Standards Association Provincial Infectious Disease Advisory

More information

Testing the Endoscope Journey vs the Decontamination Process. Wayne Spencer

Testing the Endoscope Journey vs the Decontamination Process. Wayne Spencer Testing the Endoscope Journey vs the Decontamination Process Wayne Spencer Historically Validation has focused on the process with only sporadic testing of the device used on the patient Probably with

More information

Professional instrument reprocessing in CSSD

Professional instrument reprocessing in CSSD Professional instrument reprocessing in CSSD The reprocessing cycle in CSSD Fig. 1 The objective of the surgical instrument reprocessing cycle Only proper, professional manual as well as automatic in central

More information

One such problem is blood residual inside the channels of a robotic arm.

One such problem is blood residual inside the channels of a robotic arm. SUBJECT: Robotic Arm Testing for residue blood DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue inside and outside of a robotic arm and

More information

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/ AS Medizintechnik GmbH Sattlerstrasse 15 78532 Tuttlingen, Germany Tel. (+49) 7461-966326 Fax: (+49) 7461-9663288 All rights reserved, any reproduction, adaption or translation, even in part, without the

More information

Test soils, cleaning indicators and protein residual tests in comparison

Test soils, cleaning indicators and protein residual tests in comparison Test soils, cleaning indicators and protein residual tests in comparison V. Buchrieser, T. Freundlinger, M. Gehrer,, H. Getreuer,, A. Gruber, M. Hell, W. Koller,, P. Lachner,, T. Miorini, H. Mittermayer,,

More information

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities Understanding ANSI/AAMI ST 91 Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve

More information

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants M Manufactured by: Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 U.S. Toll Free: 866-4-CALDERA

More information

Experience of soil tests used in two French hospitals : detergent validation

Experience of soil tests used in two French hospitals : detergent validation European Sterilization Congress 18 th 20 th may 2006, N f S and E f H SS Lillehammer Experience of soil tests used in two French hospitals : detergent validation Damien TALON, Axèle REBERGA PharmacySterilization

More information

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of Contents Glossary of equivalent standards... ix Committee representation... xi Acknowledgments...xiii Foreword... xiv Introduction: Need for the recommended practice...1 1 Scope...5 1.1 General...5 1.2

More information

Validation of washer-disinfector and steam sterilizer processes. Iven Kruse. July 2016

Validation of washer-disinfector and steam sterilizer processes. Iven Kruse. July 2016 Validation of washer-disinfector and steam sterilizer processes Iven Kruse July 2016 ebro in Ingolstadt, Germany The company was founded in Freiburg in 1968; It has been located in Ingolstadt since 1976

More information

TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING

TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING A NEED FOR A COMPLETE PROCESS Lionel PINEAU PhD France PATHOGEN TRANSMISSION RELATED TO ENDOSCOPE WHAT IS THE CONTAMINATION RISK? Rates of bacteremia and infectious

More information

CAUTION: FEDERAL LAW (USA) RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER OF A PHYSICIAN.

CAUTION: FEDERAL LAW (USA) RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER OF A PHYSICIAN. Manufacturer Silony Medical GmbH Leinfelder Strasse 60 70771 Leinfelden-Echterdingen Germany US Representative: Silony Medical Corp. 8200 NW 27TH STR, STE#104, DORAL, FL 33122 USA Phone: +1 305 916 0016

More information

Overview: High Level Disinfection

Overview: High Level Disinfection High Level Disinfection Who you gonna call?. Bug Busters! Dr. Michelle J. Alfa, FCCM St. Boniface General Hospital, Winnipeg, MB Disclaimers: Dr. M. Alfa - Sponsored to give invited presentations at various

More information

MDRAO ONLINE COURSE FOURTH EDITION 2018

MDRAO ONLINE COURSE FOURTH EDITION 2018 MDRAO ONLINE COURSE FOURTH EDITION 2018 Introduction MDRD Function and Structure Virtual tour of MDRD Standardized workflow Organizational Structure Chapter 1: Quality Systems Creating a Culture of quality

More information

CLEANING AND HANDLING OF MICROPORT INSTRUMENTS The following languages are included in this packet:

CLEANING AND HANDLING OF MICROPORT INSTRUMENTS The following languages are included in this packet: EN CLEANING AND HANDLING OF MICROPORT INSTRUMENTS 150802-1 The following languages are included in this packet: English (en) Deutsch (de) Nederlands (nl) Français (fr) Español (es) Italiano (it) Português

More information

ATTENTION: READ THESE INSTRUCTIONS VERY CAREFULLY BEFORE YOU PREPARE AND USE THE PRODUCT FOR THE FIRST TIME.

ATTENTION: READ THESE INSTRUCTIONS VERY CAREFULLY BEFORE YOU PREPARE AND USE THE PRODUCT FOR THE FIRST TIME. INSTRUCTION MANUAL Field of application All reusable surgical instruments, which consist of only one piece may contain simple joints or simple movable parts may be composed of several exchangeable parts

More information

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of Contents Glossary of equivalent standards... ix Committee representation... xi Acknowledgments...xiii Background on Amendment 1...xiii Foreword... xiv Introduction: Need for the recommended practice...1

More information

Facility: Sterile Processing Assessment Date: CRITERIA ANSI/AAMI COMPLIANT Number Y N N/A Point of Use Gross contaminant is removed with

Facility: Sterile Processing Assessment Date: CRITERIA ANSI/AAMI COMPLIANT Number Y N N/A Point of Use Gross contaminant is removed with Point of Use 6.2-6.3 Gross contaminant is removed with water Instruments are sorted Sharps removed Instruments placed back in original container Un-used instruments placed in bottom of basket Towel placed

More information

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS Lionel PINEAU PhD CONTENT General considerations Endoscope reprocessing, channel brushing, cleaning claims Description of the 2013 study

More information

Facility: Audit Date Auditor Name/s

Facility: Audit Date Auditor Name/s 1. 2. 3. 4. Facility: Audit Date Auditor Name/s Abbreviations and Definitions See page 10 (abbreviations); and pages 11-20 (definitions) of AS/NZS 4187. Policy or procedure : in this tool means policy

More information

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities Sterilization Practices in Healthcare Facilities Virtual Tour of Central Processing William A. Rutala, Ph.D., M.P.H. Director, Hospital Epidemiology, Occupational Health and Safety Program, UNC Health

More information

ISO/TC 198. Secretariat: ANSI

ISO/TC 198. Secretariat: ANSI DRAFT INTERNATIONAL STANDARD ISO/DIS 17664 ISO/TC 198 Secretariat: ANSI Voting begins on: Voting terminates on: 2016-04-28 2016-07-27 Processing of health care products Information to be provided by the

More information

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope SUBJECT: Detection of Blood residue inside the biopsy channel of a scope DEPARTMENT: Central Service/ Endoscopy APPROVED BY: EFFECTIVE: REVISED: 2/2012 PURPOSE: To test for detection of blood residue inside

More information

Validation of cleaning and sterilization of reusable medical devices

Validation of cleaning and sterilization of reusable medical devices Synergy makes sence Validation of cleaning and sterilization of reusable medical devices MG-FSI72-119 Last revision: June 2012 Aurélien BIGNON 5, Chemin du Catupolan - 69120 Vaulx en Velin - France - Tel.

More information

Not to be reproduced - AGPAL & QIP

Not to be reproduced - AGPAL & QIP Keep it Clean! Practical methods for implementing correct sterilisation processes advanced level Presented by: Lin Lochead, B. App. Sc Validation A Documented Procedure for obtaining, recording and interpreting

More information

Monitoring your decontamination and reprocessing cycle

Monitoring your decontamination and reprocessing cycle Monitoring your decontamination and reprocessing cycle 1er Journee National sur la Sterilisation La Societe Marocaine de Sterilisation Maroc 2011 Ahmed Ghatwary/SteriGMS/Pharcomedic Index 1-Cleaning step,

More information

Sterilization of Schweickhardt Instruments

Sterilization of Schweickhardt Instruments Sterilization of Schweickhardt Instruments INTRODUCTION l Most dental instruments are made of corrosion-resistant highgrade steels. The requirements for grades of steel are defined in national and international

More information

Products made in Switzerland

Products made in Switzerland Products made in Switzerland HYGIENIC AND MEDICALLY CLEAR DIFFERENT PRODUCTS BUT THE SAME QUALITY LABEL HYGIENIC AND MEDICALLY CLEAR 2 MDD TECHNOLOGIES FOR: 1. Cleaning 2. Disinfection 3. High Level Disinfection

More information

Endoscopy Reprocessing Trends. Wayne Spencer United Kingdom

Endoscopy Reprocessing Trends. Wayne Spencer United Kingdom Endoscopy Reprocessing Trends Wayne Spencer United Kingdom The Poor Relation? The Concept Of A level Playing Field? Irrespective of where the decontamination is performed or who it is performed by the

More information

Instructions for use. Reusable surgical instruments. Hersteller:

Instructions for use. Reusable surgical instruments. Hersteller: Hersteller: ROLAN Instruments GmbH Ernst-Thälmann-Str. 50 15859 Storkow / Germany Tel.: 0049 33678 449433 E-Mail: info@rolan-instruments.de Item codes 01-02- 03-04- 05-06- 07-08- 09-10- 11-12- 13-14- 15-16-

More information

CDWA3 / CDWA4 CDWAH. Page 4. BASKET ZONE: water pressure, spray system performance, detergent chemistry and concentration, temperature.

CDWA3 / CDWA4 CDWAH. Page 4. BASKET ZONE: water pressure, spray system performance, detergent chemistry and concentration, temperature. Rev. 2 March 2017 Control of cleaning and washing processes of medical instruments and materials is highly important since its outcome affects the success of subsequent disinfection and/or sterilization

More information

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy! Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy! Dr. Michelle J. Alfa, Ph.D., FCCM Medical Director, Clinical Microbiology, Diagnostic Services of Manitoba, Winnipeg, Canada Disclaimers:

More information

New National Standard of Canada CAN/CSA-Z Canadian medical device reprocessing

New National Standard of Canada CAN/CSA-Z Canadian medical device reprocessing New National Standard of Canada CAN/CSA-Z314.-18 Canadian medical device reprocessing Gale Schultz BN, RN 2018 Canadian Standards Association. All Rights Reserved. Nine CSA MDR standards now in one book!

More information

Low Temperature Processing. An Update on Guidelines, Standards and Requirements

Low Temperature Processing. An Update on Guidelines, Standards and Requirements Low Temperature Processing An Update on Guidelines, Standards and Requirements Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve successful completion

More information

Steam Sterilization Validation for Healthcare Institutions

Steam Sterilization Validation for Healthcare Institutions Steam Sterilization Validation for Healthcare Institutions Jonathan A. Wilder, Ph.D. President, H & W Technology, LLC/Stericert Co. Managing Director, Quality Processing Resource Group, LLC 1 Introduction

More information

Guidelines for processing and sterilising instruments

Guidelines for processing and sterilising instruments Guidelines for processing and sterilising instruments Table of contents Underlying concept of the document 1 5 Purpose, scope, glossary, symbols General information concerning reprocessing 6 13 Preliminary

More information

The Reusable Surgical Instrument Cycle

The Reusable Surgical Instrument Cycle Daily and Periodic Tests for Washer-Disinfectors and Sterilizers: Regulations, Technical and Operational Characteristics, and Updates on Electronic Bowie Dick Tests Richard Bancroft, B.Sc (Hons), FRSB

More information

The cytotoxicity and haemolysis potential

The cytotoxicity and haemolysis potential 334 ORIGINAL ARTICLE Central Service 5/211 Biocompatibility of medical devices after automated reprocessing in washer-disinfectors H. Biering*, R. Glasmacher 1, M. Hermann 2, E. Schrader 1 The cytotoxicity

More information

RIGID AUTOCLAVABLE ENDOSCOPES

RIGID AUTOCLAVABLE ENDOSCOPES IMPORTANT ADVICES PLEASE READ CAREFULLY AND KEEP THE INSTRUCTIONS FOR FUTURE REFERENCE RIGID AUTOCLAVABLE ENDOSCOPES Applications RUDOLF endoscopes serve the purpose of illumination and visualisation during

More information

WASHING / DISINFECTION. Using the AMSCO WASHER / DISINFECTOR. Easy to use. Reliable. Compact LEAN STERISHEALTHCARE

WASHING / DISINFECTION. Using the AMSCO WASHER / DISINFECTOR. Easy to use. Reliable. Compact LEAN STERISHEALTHCARE WASHING / DISINFECTION Using the AMSCO 5053 WASHER / DISINFECTOR Compact Reliable Easy to use LEAN STERISHEALTHCARE 5053 EASY TO USE - RELIABLE - COMPACT - LEAN ONE TOUCH Screen Interface 1EASY TO USE

More information

ANNEX 2. to Guideline. for Testing, Validation and Monitoring of Automated Cleaning and Disinfection Processes for Medical Devices

ANNEX 2. to Guideline. for Testing, Validation and Monitoring of Automated Cleaning and Disinfection Processes for Medical Devices ÖGSV Austrian Society for Sterile Supplies A-8045 Graz, Ursprungweg 160 Tel: +43 (0) 316/69 47 11, Fax: Ext. 4 Email: office@oegsv.com Internet: www.oegsv.com ANNEX 2 to Guideline for Testing, Validation

More information

MDRAO Importance of Bedside & Manual Precleaning CONMED.CA

MDRAO Importance of Bedside & Manual Precleaning CONMED.CA MDRAO Importance of Bedside & Manual Precleaning CONMED.CA AGENDA Role of MDRD Hazards related to reprocessing Importance of Adequate Reprocessing Standards Bedside PreClean Manual PreClean Solutions

More information

Ethicon, Inc All rights reserved. EVOTECH and the EVOTECH logo are registered trademarks owned by Advanced Sterilization Products.

Ethicon, Inc All rights reserved. EVOTECH and the EVOTECH logo are registered trademarks owned by Advanced Sterilization Products. Ethicon, Inc. 2009. All rights reserved. EVOTECH and the EVOTECH logo are registered trademarks owned by Advanced Sterilization Products. Consistent Cleaning with No Brushing. Now, the labor-intensive

More information

Guideline will serve as an orientational guide to those centres entrusted with validation, certification

Guideline will serve as an orientational guide to those centres entrusted with validation, certification DGKH Guidelines (draft) for Validation and Routine Monitoring of Process Steps in Automated Cleaning Processes with Thermal Disinfection for Heat-Resistant Medical Devices as per pren ISO 15883-1 and pren

More information

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope DEPARTMENT: Central Service/ Endoscopy APPROVED BY: EFFECTIVE: REVISED: 6/2016 PURPOSE: To test for detection

More information

DISCOVER INNOVATION DRIVING INCREASED EFFICIENCY

DISCOVER INNOVATION DRIVING INCREASED EFFICIENCY ULTRASONIC CLEANING SYSTEMS AND SONIC IRRIGATORS Central Sterile Services DISCOVER INNOVATION DRIVING INCREASED EFFICIENCY One Integrated Approach to Healthcare. Advancements in surgical instruments have

More information

VALIDATION OF WASHER-DISINFECTOR

VALIDATION OF WASHER-DISINFECTOR VALIDATION OF WASHER-DISINFECTOR MACCHINARIO DI SANIFICAZIONE AUSILI MOD. SANIFIKO S30 REPORT N. 5779-17 Customer: GIALDI S.R.L. Via Cantone, 99-42046 Reggiolo RE TIME SCHEDULE Acceptance N.: 17-5344 Samples

More information

8/23/2012. Continuing Education Contact Hours. How much of your formal education focused on sterilization or sterile processing?

8/23/2012. Continuing Education Contact Hours. How much of your formal education focused on sterilization or sterile processing? Continuing Education Contact Hours The presenter is an employee of STERIS Corporation and receives no direct compensation other than normal salary for participation in this activity. STERIS Corporation

More information

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023 AS/NZS 4187 RELEVANT STANDARDS Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023 Relevant Standards AS 2514 AS 2774 AS 2773 AS 2773.1 Drying cabinets for medical equipment Drying cabinets

More information

Endoscope Decontamination. A Tube in a Hole??

Endoscope Decontamination. A Tube in a Hole?? Endoscope Decontamination A Tube in a Hole?? John Prendergast AE(D) MIHEEM Senior Decontamination Engineer (Wales) May 2016 Note This presentation is my personal interpretation and not NHS Wales policy.

More information

National Endoscopy Programme. Decontamination Standards for Flexible Endoscopes

National Endoscopy Programme. Decontamination Standards for Flexible Endoscopes National Endoscopy Programme Decontamination Standards for Flexible Endoscopes Updated March 2008 Introduction Providing an effective endoscope decontamination service within a safe environment is an essential

More information

Workshop on Medical Device Cleanliness: How Clean is Clean Enough? Sponsored by ASTM Committee F04 on Medical and Surgical Materials and Devices

Workshop on Medical Device Cleanliness: How Clean is Clean Enough? Sponsored by ASTM Committee F04 on Medical and Surgical Materials and Devices Breaking the Myth Case Medical Page 1 Our Position Caustic cleaners are unnecessary for instrument processing and present significant hazards to health and safety Properly formulated ph neutral instrument

More information

Flexible endoscopy in a robotic world Challenges? HONG KONG HOSPITAL AUTHORITY - ADVANCED COURSE FOR INFECTION CONTROL NOVEMBER 2017

Flexible endoscopy in a robotic world Challenges? HONG KONG HOSPITAL AUTHORITY - ADVANCED COURSE FOR INFECTION CONTROL NOVEMBER 2017 Flexible endoscopy in a robotic world Challenges? HONG KONG HOSPITAL AUTHORITY - ADVANCED COURSE FOR INFECTION CONTROL NOVEMBER 2017 Objectives 1. Provide definitions and applications of robotic surgery

More information

ENDOSCOPE DECONTAMINATION- EDUCATION AND TRAINING COMPETENCY FRAMEWORK

ENDOSCOPE DECONTAMINATION- EDUCATION AND TRAINING COMPETENCY FRAMEWORK ENDOSCOPE DECONTAMINATION- EDUCATION AND TRAINING COMPETENCY FRAMEWORK Standardising the manual cleaning and decontamination process of flexible endoscopes throughout Lothian. C:\Users\Scott\Documents\Endoscopy\website

More information

Important information

Important information Version SE_023827 AI Date February 2014 Important information (with Cleaning and Sterilization Instructions) Important information 2 Basic Instructions on the Use of Synthes Implants and Instruments for

More information

The stainless steels used in the manufacture of surgical instruments can be divided into two main classes.

The stainless steels used in the manufacture of surgical instruments can be divided into two main classes. Introduction Instruments represent a significant material asset within the overall investment of a hospital. The care and maintenance these instruments receive is critical to their performance during surgery

More information

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ Disclaimer: The Society of Gastroenterology Nurses and Associates, Inc. assume no responsibility for the

More information

ROP SELF ASSESSMENT AND ACTION PLAN ACCREDITATION CANADA

ROP SELF ASSESSMENT AND ACTION PLAN ACCREDITATION CANADA CLIENT SAFETY AREA: Infection Control ROP: Reprocessing (ROP: pg.63) WRH monitors its processes for reprocessing equipment, & makes improvements as appropriate. VP Champion: Rosemary Petrakos Leader: Barb

More information

Volume 13, Issue 5/ Medtech

Volume 13, Issue 5/ Medtech Volume 13, Issue 5/2011 - Medtech The New Central Sterile Supply Department of University Hospital Authority St. Orsola- Malpighi Polyclinic The new Central Sterile Supply Department (CSSD) of the University

More information

WG 3 - ESR WORKING GROUP ON ULTRASOUND

WG 3 - ESR WORKING GROUP ON ULTRASOUND WG 3 - ESR WORKING GROUP ON ULTRASOUND Minimizing the risk of cross infection: How to keep patients safe in ultrasound ECR 2017 Thursday 2 March 2017 Vienna @ 8:30 INDUSTRY PERSPECTIVE Nicole Denjoy COCIR

More information

2016. All rights reserved.

2016. All rights reserved. DOCUMENT: Future AAMI/ISO 17664, 12 May 2017 Final Draft Standard Sterilization of health care products Information to be provided by the device manufacturer for the processing of medical devices Public

More information

EBI 12 The new data logger generation

EBI 12 The new data logger generation EBI 12 The new data logger generation 3 EBI 12 The new data logger generation High quality stainless steel housing Application range from -90 C to 150 C High temperature accuracy up to 0.05 C Extended

More information

Xerafy Autoclavable XS and XXS User Guide

Xerafy Autoclavable XS and XXS User Guide Xerafy Autoclavable XS and XXS User Guide General Information This user guide provides instructions for attaching Xerafy autoclavable XS and XXS series tags onto surgical instruments. This document is

More information

Data Loggers for Routine Control. CSSD Medical Hygiene Control

Data Loggers for Routine Control. CSSD Medical Hygiene Control Data Loggers for Routine Control Washer-Disinfectors / Washer-Disinfectors for Endoscopes / Steam Sterilizers / Bedpan Washers CSSD Medical Hygiene Control Routine control made easy with ebro data loggers

More information

Hong Kong Sterile Services Management Association Limited. Seminar. 25 th June 2014

Hong Kong Sterile Services Management Association Limited. Seminar. 25 th June 2014 Hong Kong Sterile Services Management Association Limited Seminar 25 th June 2014 1. Bowie-Dick-Test: Use of Compact-PCD vs. Bowie-Dick-Test Pack 2. How to Monitor and Evaluate Steam Sterilization Processes

More information

HTM COMPLIANCE PROGRAMME

HTM COMPLIANCE PROGRAMME HTM COMPLIANCE PROGRAMME Compliance Timeline HTM 01-01 HTM 01-06 1 July 2017 HIGH RISK SURGICAL PROCEDURES 1 July 2018 ALL ACUTE SURGICAL PROCEDURES SOP Guide to Compliance 1. Introduction HTM 01-01 was

More information

Reusable 5x Holmium Laser Fibers Instructions for Use Model: LF0201X200, LF0201X272, LF0201X365, LF0201X550, LF Holmium and Nd:YAG Fiber

Reusable 5x Holmium Laser Fibers Instructions for Use Model: LF0201X200, LF0201X272, LF0201X365, LF0201X550, LF Holmium and Nd:YAG Fiber Instructions for Use (IFU) Intended Use Reusable 5x Holmium Laser Fibers Instructions for Use Model: LF0201X200, LF0201X272, LF0201X365, LF0201X550, LF0201940 Holmium and Nd:YAG Fiber The laser probes

More information

Health Service Executive Code of Practice for Decontamination of Reusable Invasive Medical Devices

Health Service Executive Code of Practice for Decontamination of Reusable Invasive Medical Devices PART 3: RECOMMENDED PRACTICES FOR CENTRAL DECONTAMINATION UNITS Health Service Executive Code of Practice for Decontamination of Reusable Invasive Medical Devices Reader Information Directorate: Title:

More information

Water for Instrument Processing

Water for Instrument Processing Water for Instrument Processing by Marcia Frieze, Case Medical Water, the universal solvent Water can dissolve more substances than any other liquid. It is essentially nonionic or neutral. While alkaline

More information

GEISTER. I.F.U-Re/Processing Reusable Medical Devices. Medical articulated arm with quick fixation + Components

GEISTER. I.F.U-Re/Processing Reusable Medical Devices. Medical articulated arm with quick fixation + Components GEISTER Page 1 of 12 Table of Contents 1 Introduction... 3 1.1 General Information... 3 1.2 Symbols used... 3 1.3 Definitions (according to the norm EN ISO 17664)... 4 2 Intended uses... 4 3 Product description...

More information