Significant Dose Reduction for Pediatric Digital Subtraction Angiography Without Impairing Image Quality: Preclinical Study in a Piglet Model

Size: px
Start display at page:

Download "Significant Dose Reduction for Pediatric Digital Subtraction Angiography Without Impairing Image Quality: Preclinical Study in a Piglet Model"

Transcription

1 Pediatric Imaging Original Research Racadio et al. Dose Reduction for Pediatric DSA Pediatric Imaging Original Research John Racadio 1 Keith Strauss 2 Todd Abruzzo 1 Manish Patel 1 Kamlesh Kukreja 1 Neil Johnson 1 Mark den Hartog 3 Bart Hoornaert 3 Rami Nachabe 1 Racadio J, Strauss K, Abruzzo T, et al. Keywords: digital subtraction angiography (DSA), dose reduction, image quality, pediatrics DOI:1.2214/AJR Received October 24, 213; accepted without revision December 15, 213. Cincinnati Children s Hospital Medical Center, Department of Radiology, Division of Interventional Radiology has a master research agreement with Philips Healthcare. J. Racadio has had travel expenses paid for Philips Healthcare sponsored symposiums. He has not received any speaking honorariums, is not paid by Philips Healthcare, and does not have any stock options or other financial incentives in Philips Healthcare or any Philips Healthcare products. R. Nachabe, M. den Hartog, and B. Hoornaert are employees of Philips Healthcare. However, the authors who are not Philips Healthcare employees had full control of inclusion of any data and information that might present a conflict of interest for those authors who are Philips Healthcare employees. 1 Department of Radiology, Division of Pediatric Interventional Radiology, Cincinnati Children s Hospital, 3333 Burnet Ave, Cincinnati, OH Address correspondence to J. M. Racadio (john.racadio@cchmc.org). 2 Department of Radiology, Children s Hospital of Cincinnati Hospital, Cincinnati, OH. 3 Interventional X-Ray Department, Philips Healthcare, Best, The Netherlands. AJR 214; 23: X/14/ American Roentgen Ray Society Significant Dose Reduction for Pediatric Digital Subtraction Angiography Without Impairing Image Quality: Preclinical Study in a Piglet Model OBJECTIVE. The purpose of this study was to validate the hypothesis that image quality of digital subtraction angiography (DSA) in pediatrics is not impaired when using a lowdose acquisition protocol. MATERIALS AND METHODS. Three piglets corresponding to common pediatric population sizes were used. DSA was performed in the aorta and renal, hepatic, and superior mesenteric arteries using both the commonly used reference standard and novel radiographic imaging noise reduction technologies to ensure pairwise radiation dose and image quality comparison. The air kerma per frame at the interventional reference point for each DSA acquisition was collected as a radiation dose measure, and image quality was evaluated by five interventional radiologists in a randomized blinded fashion using a 5-point scale. RESULTS. The mean air kerma (± SD) at the interventional reference point with the novel x-ray imaging noise reduction technology was significantly lower (1.1 ±.8 mgy/frame) than with the reference technology (4.2 ± 3. mgy/frame, p =.5). However, image quality was statistically similar, with average scores of 3.2 ±.4 and 3.1 ±.5 for the novel and reference technologies, respectively (p =.934); interrater absolute agreement was.77. CONCLUSION. The DSA radiation dose for pediatrics can be reduced by a factor of four with a novel x-ray imaging noise reduction technology without deterioration of image quality. D igital subtraction angiography (DSA) is used for a wide range of vascular procedures in pediatric patients; however, the radiation dose from DSA is relatively high. In a study using anthropomorphic phantoms, the effective radiation dose delivered to the head of a 5-year-old child from 1 DSA frames was estimated to be equivalent to approximately 1 minute of fluoroscopy [1]. In that same study, the authors showed that in a typical cerebral angiography procedure, the radiation dose from DSA is nearly three times that from fluoroscopy [1]. Pediatric patients are more sensitive to the effects of radiation than adults and have a greater lifetime risk of radiation-induced cancer [2]. As part of the Image Gently public service campaign, an interventional radiology phase, Image Gently, Step Lightly, was instituted to encourage radiation protection during pediatric interventional radiology procedures [3]. Specific radiation dose reduction measures for DSA that were suggested in the campaign included tight collimation to the anatomic area of interest and appropriate choice of the acqui- sition frame rate (number of exposures) to adequately evaluate the vascular anatomy of interest [3]. Although measures such as these can help to reduce radiation dose as much as possible, further reduction is limited using existing x-ray imaging technology. In a recent study [4], 2 adult patients underwent two consecutive cerebral DSA examinations performed first with a standard reference technology and then with a novel noise reduction technology that was associated with a one quarter radiation dose protocol. Paired comparison of the DSA images revealed no significant difference in image quality, while the radiation dose was reduced by a factor of four. However, these data cannot necessarily be extrapolated to the pediatric population because the smaller size of the patients may make relatively small changes in image quality more significant. The data also cannot necessarily be extrapolated to other anatomic regions. Because it would be ethically unacceptable to perform an additional DSA examination on pediatric patients and it would be difficult to define image quality in a phantom from a clinical perspective, we chose to use an 94 AJR:23, October 214

2 animal model to validate the findings of the previous study in pediatric patients. Materials and Methods Three of the authors are employees of Philips Healthcare. However, the authors who are not Philips Healthcare employees had full control of inclusion of any data and information that might present a conflict of interest for those authors who are Philips Healthcare employees. This was a prospective study comparing radiation dose and image quality during DSA acquisitions performed with a reference technology (, Philips Healthcare) and with a novel noise-reduction technology (, Philips Healthcare). DSA was performed using the reference and then the novel technology sequentially on the aorta, renal artery, hepatic artery, and superior mesenteric artery in each of three different pediatric-sized piglets. Imaging Technology This study was performed using a monoplane flat-panel detector C-arm angiographic system (FD2, version 8.1.2, Philips Healthcare). The system was adapted so that both the reference and novel x-ray technique factors could be used along with the corresponding image processing for DSA imaging. The differences in acquisition factors included increased spectral filtration and decreased energy per pulse with the use of, which yielded lower radiation dose. With respect to image processing, enabled noise reduction by performing automatic pixel shifting for motion correction and also used improved spatial and temporal filtering algorithms for noise reduction and image enhancement [4]. Animal Study The animal experiments were performed on piglets following the guidelines of the local institution animal care and use committee. Three piglets weighing 5, 14, and 19.5 kg were used, corresponding to pediatric patients of approximately.5, 3, and 5 years old, respectively [5]. The animals were sedated with an intramuscular injection of xylazine (2 mg/kg), ketamine (2 mg/kg), and buprenorphine (.2 mg/kg) and placed under general anesthesia with % isoflurane administrated by endotracheal tube. An 8-French Skater biliary drainage catheter (InterV) was inserted in the bladder to drain the excreted contrast material. At the end of the experiments, the piglets were euthanized with an IV injection of pentobarbital sodium (1 ml/1 kg). All interventional DSA examinations were performed by the same interventional radiologist. Femoral arterial access was obtained under ultrasound guidance using the Seldinger technique; selective catheterization of each vessel of interest (aorta, renal artery, hepatic artery, and superior mesenteric Dose Reduction for Pediatric DSA TABLE 1: Protocol According to Arterial Site and Piglet Size Piglet Size (kg) artery) was then performed. Flush aortography was performed with a 4-French pigtail catheter (Soft Vu, Angiodynamics), whereas selective arteriography was performed with a 4-French JB-1 catheter (Soft Vu, Angiodynamics) in the two larger piglets and with a 4-French hook catheter (Soft Vu, Angiodynamics) in the smallest piglet. Iodinated contrast material (Optiray 35, Covidien) was injected with a power injector (Mark V Provis, Medrad) using a predetermined contrast volume and injection rate depending on the piglet size and arterial site (Table 1). All injections were performed during anesthesiainduced breath-hold. Before initiating a subsequent arteriography, the contrast material from the venous phase of the preceding angiography was cleared. DSA was performed using a common clinically used pediatric multiphase acquisition consisting of two frames per second (fps) for 3 seconds, followed by 1 fps for 6 seconds and then.5 fps for the remainder of the acquisition. All DSA acquisitions on all animals were performed at the same table height of 13 cm and Arterial Site Aorta Renal Artery Hepatic Artery Superior Mesenteric Artery TABLE 2: FOV Diagonal Lengths According to Arterial Site and Piglet Size Piglet Size (kg) FOV Diagonal Length (cm) Aorta Renal Artery Hepatic Artery Superior Mesenteric Artery source-to-detector distance of 19 cm, resulting in a source-to-skin distance (SSD) of 77.5 cm. However, the FOV and collimation were adapted to each anatomic site and animal size as would be done clinically to ensure optimal visualization of the vessels (Table 2). Each DSA acquisition was performed first with the reference technology and then repeated with the novel x-ray imaging noise reduction technology with the same conditions and settings to ensure a fair pairwise comparison. Examples of such DSA acquisitions with both technologies are shown in Figure 1. All DSA acquisitions were sent to the local hospital research PACS (Merge Healthcare) along with a Radiation Dose Structured Report in which air kerma was reported for each single DSA acquisition. By using a lock-in technique, the air kerma was identical for every frame of a given DSA run. Image Quality Assessment Each DSA acquisition (eight acquisitions in each of the three piglets) underwent blinded evalu- TABLE 3: Five-Point Scoring Scale Used for Image Quality Assessment of Digital Subtraction Angiography (DSA) Acquisitions Score Interpretation of DSA Acquisition (Nondiagnostic) Failing to depict arterial structures in any meaningful way 1 (Poor) Enabling diagnostic evaluation of first-order branches relative to the injected vessel only 2 (Intermediate) Enabling diagnostic evaluation up to second-order branches relative to injected vessel 3 (Good) Enabling diagnostic evaluation up to third-order branches relative to injected vessel 4 (Excellent) Enabling diagnostic evaluation up to vessels beyond third-order branches relative to injected vessel AJR:23, October

3 Racadio et al. ation twice by each of five interventional radiologists, yielding a total of 24 scores. A scoring system (Table 3) based on diagnostic visualization of vessel branches relative to the injected vessel was used to reduce rater subjectivity. The scoring system was based on a previously published scale that was validated on a rabbit model [6]. A similar scale has also been used for image quality assessment of DSA acquisitions performed on adult patients who underwent transarterial chemoembolization of liver metastases [7]. Each DSA acquisition was presented on the same high-resolution clinical three-megapixel ( ) PACS workstation (Model E3cHB, NDS Surgical Imaging) for scoring. Image evaluations were performed without a time limit in a dimly lit reading room with full capability of image manipulation of the complete DSA acquisition, such as frame-byframe or cine display and window-level manipulation, to simulate typical clinical evaluation. The system settings were deidentified so that the radiologists were blinded to the type of technology that was used to generate the acquisitions. Each DSA acquisition was scored twice by each radiologist, with an interval of at least 1 week to avoid memory bias. Data Collection and Statistical Analysis The air kerma for each DSA acquisition and corresponding image quality scores assigned by each reader were tabulated. The institutional medical physicist calibrated the displayed air kerma value at the interventional reference point; the correction factor was 1%. The reported air kerma at the interventional reference point corresponds to the value displayed by the system after calibration. For each DSA image quality assessment, the mean image quality score was used for statistical comparison. Interrater agreement among the five readers was computed using a two-way intraclass correlation coefficient (ICC) for absolute agreement. An ICC below.4, between.41 and.59, between.6 and.74, and.75 or above corresponded with poor, fair, good, and excellent agreement, respectively [8]. The Wilcoxon signed rank test was applied to identify statistical differences in air kerma and image quality between the reference and the novel x-ray imaging noise reduction technology. In addition, a Levene test was applied to the image quality scores to evaluate variation in scores. A value of p <.5 indicated statistical significance. All statistical calculations were performed with commercially available software (Matlab, version R211b, Mathworks). Results The mean (± SD) air kerma at the interventional reference point was significantly reduced from 4.2 ± 3. mgy/frame with the reference technology to 1.1 ±.8 mgy/frame with the novel x-ray technology, yielding a dose reduction of 72.3% (p =.5) (Fig. 2). Fig. 1 Examples of digital subtraction angiography acquisitions performed using reference (left) and novel ) (right) x-ray imaging noise reduction technology (both systems, Philips Healthcare). From top to bottom, injected vessels were aorta, renal artery, hepatic artery, and superior mesenteric artery, respectively. Note difference in background noise between reference and novel technologies; realtime automatic pixel shift of novel technology reduces noise in parenchyma of kidney and liver. The mean image quality scores and their variation were statistically similar between the two technologies, with an average score of 3.2 ±.4 for the novel technology versus 3.1 ±.5 for the reference technology (p =.934 for comparison of mean and p =.912 for comparison of variance) as depicted in Figure 3. Interrater absolute agreement was excellent at.77. Discussion In this study, we evaluated the possibility of reducing the radiation dose (air kerma) for DSA acquisitions without impairing image 96 AJR:23, October 214

4 Dose Reduction for Pediatric DSA Air Kerma (mgy/frame) A Air Kerma (mgy/frame) Fig. 2 Air kerma rate for each digital subtraction angiography (DSA) acquisition. A and B, Wilcoxon plots of air kerma rate for each DSA acquisition show pairwise comparison between reference and novel x-ray technologies (A) (both systems, Philips Healthcare) and corresponding boxplot (B). B Image Quality Score Fig. 3 Graph shows image quality score mean and SD for reference () and novel () x-ray technologies (both systems, Philips Healthcare). quality in a pediatric animal model. DSA has a wide range of uses in pediatric patients, although depending on local expertise and referral patterns, the types of procedures may differ considerably. Some pediatric indications for DSA include renal arteriography for evaluation and treatment of renovascular hypertension [9, 1], hepatic arteriography to evaluate liver transplant vascular pathology [9], superior mesenteric arteriography for diagnosis and treatment of gastrointestinal bleeding (particularly in immunosuppressed patients) [11], bronchial arteriography to evaluate and treat hemoptysis (particularly in cystic fibrosis patients) [12], evaluation and treatment of arteriovenous dialysis grafts, trauma [13], cerebral arteriography for diagnosis and treatment of arteriovenous malformations and aneurysms [14, 15], and evaluation and treatment of vascular malformations [16]. Use of DSA for such complex interventional procedures can expose pediatric patients to relatively high radiation doses. A review of adult patient doses in interventional neuroradiology showed that approximately 66% of the patient exposure during interventional neuroradiologic procedures came from DSA [17]. In a study of adult patients undergoing DSA of the abdominal aorta and lower limbs, it was noted that although DSA accounted for.1% of the procedure time, it contributed more than 88% of the radiation dose [18]. Pediatric institutions are therefore cautious in using DSA because of the importance of minimizing radiation exposure in children. In the previous study that compared the reference technology and novel noise reduction technology in adults undergoing cerebral DSA [4], two consecutive DSA datasets (one using the reference technology and one using the novel technology) were acquired in each patient and reviewed by three neuroradiologists in a randomized blinded fashion. The authors found that a radiation dose reduction of 75% occurred using the novel technology; there was no decrease in image quality in their adult study population. The current study was designed to perform a pairwise comparison with confirmation that there was no significant image quality degradation associated with the same level of dose reduction using the new technology in pediatric patients with smaller vessels. A similar study could not be performed on children for ethical reasons because of their increased sensitivity to radiation-induced cancer [2]; therefore, three different sized piglets that corresponded to pediatric patients at.5, 3, and 5 years old were used. In this piglet model, the use of the noise-reduction algorithm and one quarter dose protocol did not result in any significant decrease in image quality. Such a significant dose reduction to patients will also benefit the interventional radiology staff by reducing their occupational radiation exposure from x- ray scatter. Because the angiography in this study was performed with a power injector and not hand injected, no bias was introduced to image quality due to differences in injection forces. We chose to perform selective angiography on the aorta, renal artery, hepatic artery, and superior mesenteric artery because they are commonly injected arterial vessels in interventional pediatric radiology. Although pediatric cerebral DSA is also commonly performed, it was not performed in this study because, in our experience, the piglet is a poor cerebrovascular model because the thickness of the skull makes visualization difficult. Conclusion The DSA radiation dose for pediatrics can be reduced by a factor of four with the introduction of the novel x-ray technology without deterioration of image quality. References 1. Miksys N, Gordon CL, Thomas K, Connolly BL. Estimating effective dose to pediatric patients undergoing interventional radiology procedures using anthropomorphic phantoms and MOSFET dosimeters. AJR 21; 194: Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 21; 176: Sidhu M, Goske MJ, Connolly B, et al. Image Gently, Step Lightly: promoting radiation safety in pediatric interventional radiology. AJR 21; 195:[web]W299 W31 4. Soderman M, Holmin S, Andersson T, Palmgren C, Babic D, Hoornaert B. Image noise reduction algorithm for digital subtraction angiography: clinical results. Radiology 213; 269: Heffess CS, Wenig BM, Thompson LD. Metastatic renal cell carcinoma to the thyroid gland: a clinicopathologic study of 36 cases. Cancer 22; 95: Racadio JM, Kashinkunti SR, Nachabe RA, et al. Clinically useful dilution factors for iodine and gadolinium contrast material: an animal model of pediatric digital subtraction angiography using state-of-the-art flat-panel detectors. Pediatr Radiol 213; 43: AJR:23, October

5 Racadio et al. 7. Vogl TJ, Gruber-Rouh T, Eichler K, et al. Repetitive transarterial chemoembolization (TACE) of liver metastases from gastric cancer: local control and survival results. Eur J Radiol 213; 82: Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull 1979; 86: Marshalleck F. Pediatric arterial interventions. Tech Vasc Interv Radiol 21; 13: Tullus K, Brennan E, Hamilton G, et al. Renovascular hypertension in children. Lancet 28; 371: Racadio JM, Agha AK, Johnson ND, Warner BW. Imaging and radiological interventional techniques for gastrointestinal bleeding in children. Semin Pediatr Surg 1999; 8: Roebuck DJ, Barnacle AM. Haemoptysis and bronchial artery embolization in children. Paediatr Respir Rev 28; 9: Gross JL, Woll NL, Hanson CA, et al. Embolization for pediatric blunt splenic injury is an alternative to splenectomy when observation fails. J Trauma Acute Care Surg 213; 75: Wolfe TJ, Hussain SI, Lynch JR, Fitzsimmons BF, Zaidat OO. Pediatric cerebral angiography: analysis of utilization and findings. Pediatr Neurol 29; 4: Lv X, Jiang C, Li Y, Yang X, Wu Z. Endovascular treatment for pediatric intracranial aneurysms. Neuroradiology 29; 51: Cahill AM, Nijs EL. Pediatric vascular malformations: pathophysiology, diagnosis, and the role of interventional radiology. Cardiovasc Intervent Radiol 211; 34: Gkanatsios NA, Huda W, Peters KR. Adult patient doses in interventional neuroradiology. Med Phys 22; 29: Juszkat R, Blaszak MA, Majewska N, Majewski W. Dose-area product of patients undergoing digital subtraction angiography (DSA): abdominal aorta and lower limb examinations. Health Phys 29; 96:13 18 FOR YOUR INFORMATION The AJR has made getting the articles you really want really easy with an online tool, Really Simple Syndication, available at It s simple. Click the RSS button located in the menu on the right side of the page. You ll be on your way to syndicating your AJR content in no time. 98 AJR:23, October 214

This prospective study was approved by our institutional review board.

This prospective study was approved by our institutional review board. Digital subtraction angiography during TIPS creation or revision: data on radiation exposure and image quality obtained using a standard and a low-dose acquisition protocol in a flat-panel detector-based

More information

Radiation Exposure During Endovascular Procedures: Methods to Protect Patients, Physicians and OR Personnel

Radiation Exposure During Endovascular Procedures: Methods to Protect Patients, Physicians and OR Personnel Radiation Exposure During Endovascular Procedures: Methods to Protect Patients, Physicians and OR Personnel Melissa L. Kirkwood MD Associate Professor Division of Vascular and Endovascular Surgery UT Southwestern

More information

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Poster No.: C-0674 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Sofue 1, T. Yoshikawa 1, N. Negi 1, Y. Ohno

More information

Power of BRANSIST safire in Neuroendovascular Therapy

Power of BRANSIST safire in Neuroendovascular Therapy Vascular Power of BRANSIST safire in Neuroendovascular Therapy Department of Radiology, Kinki University Hospital Suguru Ueda Mr. Suguru Ueda 1. Introduction Kinki University Hospital is located in the

More information

Clinical Applications. ImagingRite. Interventional Radiology

Clinical Applications. ImagingRite. Interventional Radiology Clinical Applications ImagingRite Interventional Radiology ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in optimizing

More information

Making the difference with Live Image Guidance

Making the difference with Live Image Guidance OncoSuite Interventional X-ray Making the difference with Live Image Guidance Next generation of Image Guided Therapy in Interventional Oncology Live Image Guidance in Interventional Oncology OncoSuite:

More information

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite Clinical Applications ImagingRite Neuro Intervention 1 ImagingRite ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in

More information

Intracranial stent visualization for image guided interventions and therapy

Intracranial stent visualization for image guided interventions and therapy Intracranial stent visualization for image guided interventions and therapy Daniel Ruijters, Peter van de Haar, Ruben Roijers, Niels J. Noordhoek, Jan Timmer, and Drazenko Babic interventional X-Ray, Philips

More information

Clarity CT Technology

Clarity CT Technology Clarity CT Technology WHITE PAPER January 2013 Using state of the art algorithms Sapheneia Clarity CT allows physicians to lower radiation dose when acquiring CT data while maintaining image quality. The

More information

Radiation Protection in Fluoroscopy

Radiation Protection in Fluoroscopy Radiation Protection in Fluoroscopy Introduction Physicians performing fluoroscopic guided procedures should be aware of the potential for serious radiation-induced skin injury to the patient. Occasionally

More information

Department of Radio Diagnosis

Department of Radio Diagnosis Department of Radio Diagnosis S.No. Name of Equipment Specification Qty required Suggested Manufacturers Name Approx Unit cost (Rs.) Usage 1 Digital substraction angiography (D.S.A.) A single plane, ceiling

More information

First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector

First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector 01 White Paper No. 02/2009 First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector Leiden University Medical Center (LUMC) in the Netherlands is the first hospital in the world

More information

Making the difference

Making the difference Intuis Interventional X-ray Making the difference with Live Image Guidance Making the difference with Live Image Guidance Philips Intuis Together we make the difference in minimally invasive treatment

More information

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children At Canada s premier pediatric hospital also recognized as one of the world s foremost pediatric healthcare institutions

More information

Improving protocols and procedures for strengthened radiation protection in interventional procedures

Improving protocols and procedures for strengthened radiation protection in interventional procedures Improving protocols and procedures for strengthened radiation protection in interventional procedures R. Loose German Commission on Radiological Protection (SSK) German Roentgen Society (DRG) Institute

More information

MANAGING PATIENT DOSE IN FLUOROSCOPIC PROCEDURES

MANAGING PATIENT DOSE IN FLUOROSCOPIC PROCEDURES MANAGING PATIENT DOSE IN FLUOROSCOPIC PROCEDURES Keith J. Strauss, MSc MSc,, FAAPM Director, Radiology Physics & Engineering Children s Hospital Harvard Medical School DILEMAS A. Fluoroscopically Guided

More information

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN

More information

Making the difference

Making the difference Intuis Interventional X-ray Making the difference with Live Image Guidance Making the difference with Live Image Guidance Philips Intuis Together we make the difference in minimally invasive treatment

More information

Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease

Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease Low radiation dose subtraction CT angiography (CTA) for diagnosis of peripheral arterial occlusive disease Poster No.: C-1978 Congress: ECR 2013 Type: Authors: Scientific Exhibit M. Suzuki 1, R. Tanaka

More information

Dose Reduction during EVAR in Hybrid Room. Multicentric Study

Dose Reduction during EVAR in Hybrid Room. Multicentric Study Dose Reduction during EVAR in Hybrid Room Multicentric Study Golden rules to reduce radiation dose in the hybrid room to a minimum Stéphan Haulon, MD, Best Practice from Lille Aortic Center Aotic Centre,

More information

What s the Issue? Radiation Dose in the OR. Exposure to OR Personnel. Effects of Proper Collimation. Image-guided procedures are increasingly complex

What s the Issue? Radiation Dose in the OR. Exposure to OR Personnel. Effects of Proper Collimation. Image-guided procedures are increasingly complex Radiation Dose in the OR Robert G. Gould, Sc.D. Professor and Vice Chair Department of Radiology and Biomedical Imaging University of California San Francisco What s the Issue? Image-guided procedures

More information

Patient doses in lower limb angiography in a recently installed flat detector x-ray system

Patient doses in lower limb angiography in a recently installed flat detector x-ray system Patient doses in lower limb angiography in a recently installed flat detector x-ray system Poster No.: C-0939 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper V. Tsapaki, D. Papavdis,

More information

White Paper. Dose Tracking System. Patient Skin Dose Estimates in Real Time

White Paper. Dose Tracking System. Patient Skin Dose Estimates in Real Time White Paper Dose Tracking System Patient Skin Dose Estimates in Real Time INTRODUCTION In an era of increased utilization of ionizing radiation for medical diagnostic and interventional procedures and

More information

Making the difference with Live Image Guidance

Making the difference with Live Image Guidance Interventional X-ray Making the difference with Live Image Guidance AlluraClarity - Clinically proven Key benefits Dramatically reduces X ray dose, resulting in reduced risk of complications from radiation

More information

2018/7/31. Primary Objective

2018/7/31. Primary Objective Andrew Kuhls-Gilcrist, PhD, DABR July 31, 2018 Primary Objective The purpose of this session is to allow vendors to present their solutions and describe how their product can empower the clinician and

More information

Making the difference with Live Image Guidance Dose management in Allura systems

Making the difference with Live Image Guidance Dose management in Allura systems DoseWise Allura Making the difference with Live Image Guidance Dose management in Allura systems The importance of radiation dose management The harmful effects of X-rays on human tissues had already become

More information

Poster No.: R Page 1 of 12

Poster No.: R Page 1 of 12 Abdominal endovascular aortic repair (EVAR): Comparison of image fusion (IF3D) angiography versus conventional (2D) angiography with respect to intravenous contrast dose, radiation dose and total fluoroscopy

More information

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS Rutgers Robert Wood Johnson Medical School, New Brunswick, NJ Abstract

More information

Setting The setting was secondary care. The economic study was carried out in the Netherlands.

Setting The setting was secondary care. The economic study was carried out in the Netherlands. Contrast-enhanced MR angiography and digital subtraction angiography in living renal donors: diagnostic agreement, impact on decision making, and costs Kock M C, Ijzermans J N, Visser K, Hussain S M, Weimar

More information

Dose management at Ziehm Imaging for 2D imaging

Dose management at Ziehm Imaging for 2D imaging 01 Dose management at Ziehm Imaging for 2D imaging An educational paper on general dose principles and concepts as well as further insights on dose management for 2D imaging at Ziehm Imaging. International

More information

Optimisation of angioplasty workflow for CO2-angiography in patients with kidney impairment

Optimisation of angioplasty workflow for CO2-angiography in patients with kidney impairment Optimisation of angioplasty workflow for CO2-angiography in patients with kidney impairment Disclosure Speaker name: Ulf Teichgräber, MD, MBA Potential conflicts of interest related to the presentation:

More information

Dose comparison between Ziehm Vision RFD 3D, Medtronic O-arm O2 and Samsung NeuroLogica BodyTom

Dose comparison between Ziehm Vision RFD 3D, Medtronic O-arm O2 and Samsung NeuroLogica BodyTom 01 Dose comparison between RFD 3D, Medtronic O-arm O2 and Samsung NeuroLogica BodyTom Discover significantly lower dose levels for the RFD 3D. Today, intraoperative 3D imaging is regarded as the gold standard

More information

Radiation Dose Terms Made Simple v1

Radiation Dose Terms Made Simple v1 Terms Made Simple 9670970 v1 In Air Absorbed Table In Air Absorbed Table Goal To provide basic term definitions and exhibit how radiation effects the patient, surgeon and staff. In Air Absorbed Table Why

More information

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals Koos Geleijns, medical physicist CT delivers excellent 3D image quality CT delivers excellent 3D image quality

More information

Interventional Radiology

Interventional Radiology Angiography Systems Interventional Radiology *Pictured system is INFX-8000C Unprecedented Flexibility and Integration. Today s radiology interventions require speed, flexibility and high performance. Canon

More information

Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT

Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT GE Healthcare Case study Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT Jean-Louis Sablayrolles, M.D. Laurent Macron, M.D. Jacques Feignoux, M.D. Centre Cardiologique du Nord,

More information

Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children

Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children Mervyn D. Cohen, MB, ChB, MD Radiologists desire to keep radiation dose as low as possible. Pulsed fluoroscopy provides

More information

On the Importance of Computation in Clinical Radiology

On the Importance of Computation in Clinical Radiology On the Importance of Computation in Clinical Radiology Manuel Arreola, Ph.D., DABR Director of Radiological Physics Radiology Department University of Florida/Shands Healthcare My Disclaimer I ve received

More information

Radiological Protection of Patient and Operator in Interventional Radiology

Radiological Protection of Patient and Operator in Interventional Radiology Radiological Protection of Patient and Operator in Interventional Radiology JMAJ 44(11): 484 489, 2001 Hironobu NAKAMURA Professor and Chairman, Department of Radiology, Osaka University Graduate School

More information

Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patients

Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patients Neuroradiology (2013) 55:1365 1372 DOI 10.1007/s00234-013-1276-0 INTERVENTIONAL NEURORADIOLOGY Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614

More information

3D Roadmapping in neuroendovascular procedures an evaluation

3D Roadmapping in neuroendovascular procedures an evaluation Clinical applications 3D Roadmapping in neuroendovascular procedures an evaluation H. Okumura T. Terada D. Babic R. Homan T. Katsuma Department of Neurosurgery, Wakayama Rosai Hospital, Wakayama, Japan.

More information

Technical solutions, and clinical utility of CO2 angiography. Ulf Teichgräber, MD, MBA Department of Radiology University Hospital Jena

Technical solutions, and clinical utility of CO2 angiography. Ulf Teichgräber, MD, MBA Department of Radiology University Hospital Jena Technical solutions, and clinical utility of CO2 angiography Ulf Teichgräber, MD, MBA Department of Radiology University Hospital Jena Disclosure Speaker name: Ulf Teichgräber, MD, MBA Potential conflicts

More information

Disclaimer and Notices.

Disclaimer and Notices. Fluoroscopy Dose Reduction and Clinical Management Fluoroscopy Trajectories (Curves), Global Dose Monitoring and Tracking of Patient Dose. Pei Jan Paul Lin, Ph.D. Division of Radiation Physics and Biology

More information

Comprehensive Dose Management for Improved Patient Care.

Comprehensive Dose Management for Improved Patient Care. Comprehensive Dose Management for Improved Patient Care. Set the standard for dose management Toshiba is committed to continuously advancing dose reduction technologies and incorporating these into its

More information

In recent years, the benefits and complexity of interventional

In recent years, the benefits and complexity of interventional PATIENT SAFETY Patient Radiation Dose Management in the Follow-Up of Potential Skin Injuries in Neuroradiology E. Vano, J.M. Fernandez, R.M. Sanchez, D. Martinez, L. Lopez Ibor, A. Gil, and C. Serna-Candel

More information

Function of this control

Function of this control Page 1 of 7 Introduction The goals of this training are to develop a better appreciation of equipment controls, scatter radiation in the interventional suite, the radiation output capabilities of the angiography

More information

8/1/2018. Optimization Strategies for Pediatric CT Imaging. Educational Objectives. Brain CT

8/1/2018. Optimization Strategies for Pediatric CT Imaging. Educational Objectives. Brain CT Image Quality Radiation Dose Contrast Dose Sedation Optimization Strategies for Pediatric CT Imaging Samuel Brady, M.S. Ph.D. DABR samuel.brady@cchmc.org 8-1-218 @CincyKidsRad facebook.com/cincykidsrad

More information

Computed tomography. High. performance. no trade-offs. Philips Ingenuity CT family

Computed tomography. High. performance. no trade-offs. Philips Ingenuity CT family Computed tomography High performance no trade-offs Philips Ingenuity CT family Keeping you ahead Until now, CT scanning has too often been about trade-offs. You ve been forced to choose between high image

More information

Image-Quality Perception as a Function of Dose in Digital Radiography

Image-Quality Perception as a Function of Dose in Digital Radiography Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research Thomas Lehnert 1 Nagy N. N. Naguib Huedayi Korkusuz Ralf W. Bauer J. Matthias Kerl Martin G. Mack Thomas

More information

How to perform multi-modality 3D roadmapping in neurovascular procedures

How to perform multi-modality 3D roadmapping in neurovascular procedures How to perform multi-modality 3D roadmapping in neurovascular procedures Poster No.: C-1183 Congress: ECR 2012 Type: Educational Exhibit Authors: D. Ruijters, R. Homan, P. Mielekamp, D. Babic; Best/NL

More information

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL SAHLGRENSKA ACADEMY DEPARTMENT OF RADIATION PHYSICS RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL M.Sc. thesis Roham D.

More information

Image-Quality Perception as a Function of Dose in Digital Radiography

Image-Quality Perception as a Function of Dose in Digital Radiography Medical Physics and Informatics Original Research Medical Physics and Informatics Original Research PROOF Copyrighted Material For Review Only Not for Distribution Page Copy Image-Quality Perception as

More information

Interventional Tumor Therapy Minimally Invasive, Maximally Effective

Interventional Tumor Therapy Minimally Invasive, Maximally Effective Cover Story Interventional Oncology Interventional Tumor Therapy Minimally Invasive, Maximally Effective There is a continuous expansion of indications for interventional therapies in oncology. Thanks

More information

Making Reliable Low-dose CT a Clinical Reality

Making Reliable Low-dose CT a Clinical Reality Making Reliable Low-dose CT a Clinical Reality By Ramakrishnan RS, MD, DNB, Consultant Radiologist 1 ; George Joseph, MD, DMRD, Consultant Radiologist 2 ; and Sundar RK, BSc, DRT, DAMIT, Clinical Applications

More information

In-vivo Targeting of Liver Lesions with a Navigation System based on Fiducial Needles

In-vivo Targeting of Liver Lesions with a Navigation System based on Fiducial Needles In-vivo Targeting of Liver Lesions with a Navigation System based on Fiducial Needles L. Maier-Hein 1, A. Tekbas 2, A. Seitel 1, F. Pianka 2, S. A. Müller 2, S. Schawo 3, B. Radeleff 3, R. Tetzlaff 1,4,

More information

Minnesota Rules, Chapter 4732 X-ray Revision

Minnesota Rules, Chapter 4732 X-ray Revision Minnesota Rules, Chapter 4732 X-ray Revision FLUOROSCOPIC X-RAY SYSTEMS, 1.0 Subpart 17. Subpart 18. Subpart 19. Subpart 20. Subpart 21. Subpart 22. Subpart 23. Subpart 24. Subpart 25. Equipment performance

More information

Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy: Use of Grid-Controlled Pulsed Fluoroscopy

Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy: Use of Grid-Controlled Pulsed Fluoroscopy Giles W. L. Boland 1 Brian Murphy Ron Arellano Loren Niklason Peter R. Mueller Received January 24, 2000; accepted after revision May 3, 2000. Supported by a grant from Philips Medical Systems. 1 All authors:

More information

Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide.

Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide. Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide. Fluoroscopy units used in angiographic and cardiac applications are normally

More information

Clinical Experience Using SONIALVISION G4 Multi-Purpose R/F System for Angiographic Examinations

Clinical Experience Using SONIALVISION G4 Multi-Purpose R/F System for Angiographic Examinations R/F Clinical Experience Using SONIALVISION G4 Multi-Purpose R/F System for Angiographic Examinations Department of Central Radiology, Takasago Municipal Hospital Naoki Shioya Mr. Naoki Shioya 1. Introduction

More information

New Joint Commission Requirements for Fluoroscopy Services

New Joint Commission Requirements for Fluoroscopy Services New Joint Commission Requirements for Fluoroscopy Services Melissa C. Martin, M.S., FAAPM, FACR Past President AAPM Florida Chapter Fall Meeting Orlando, FL September 14, 2018 New Requirements Effective

More information

Infinix -i 4D CT ONE ViSION

Infinix -i 4D CT ONE ViSION CT Clinical Case Study Infinix -i 4D CT ONE ViSION Seamless Integration of CT and Angiography Follow us: www.medical.toshiba.com @ToshibaMedical +Toshiba Medical Toshiba America Medical Toshiba Medical

More information

The Latest in Radiation Dose Reduction Techniques in CT

The Latest in Radiation Dose Reduction Techniques in CT The Latest in Radiation Dose Reduction Techniques in CT Cynthia H. McCollough, PhD, FAAPM, FACR Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo Clinic,

More information

Future bound. Philips Ingenuity Core

Future bound. Philips Ingenuity Core Future bound Philips Ingenuity Core High reliability Low-dose, high-quality imaging and coverage, and the ability to personalize image quality* patient by patient. Expect excellence in routine imaging,

More information

Optimisation in CT. A case for shared approach. Stephen J. Golding University of Oxford

Optimisation in CT. A case for shared approach. Stephen J. Golding University of Oxford Optimisation in CT A case for shared approach Stephen J. Golding University of Oxford Body CT 1979: 10 mm sections 20 second exposure 60 second reconstruction Body CT 2007 Submillimetre sections instant

More information

Optimize efficiency with a universal system.

Optimize efficiency with a universal system. wins www.siemens.com/artis-one Optimize efficiency with a universal system.. Designed around you. Answers for life. Consistently efficient. Improve your competitiveness by mastering the balancing act between

More information

2017 ACR Computed Tomography Quality Control Manual FAQS

2017 ACR Computed Tomography Quality Control Manual FAQS Updated 11-15-2017 2017 ACR Computed Tomography Quality Control Manual FAQS Q. The updated 2017 ACR Computed Tomography Quality Control Manual has been released. (Visit www.acr.org/education/education-catalog.)

More information

7/28/2017. Michael Speidel. University of Wisconsin - Madison

7/28/2017. Michael Speidel. University of Wisconsin - Madison Scanning-Beam Digital X-ray (SBDX) is a technology for real-time fluoroscopy and angiography in the cardiac cath lab Michael Speidel Characterized by high speed beam scanning with a 2D array of focal spots

More information

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose ipatient is an advanced platform that delivers focused innovations to facilitate patient-centered

More information

A Paradigm Shift in Patient Dose Monitoring

A Paradigm Shift in Patient Dose Monitoring VL White Paper A Paradigm Shift in Patient Dose Monitoring Dose Tracking System Andrew Kuhls-Gilcrist, Ph.D., D.A.B.R. Senior XRVL Clinical Science Manager Canon Medical Systems, Inc. Millions of fluoroscopically

More information

Local dose reference levels for abdominal interventional radiology procedures

Local dose reference levels for abdominal interventional radiology procedures Local dose reference levels for abdominal interventional radiology procedures Poster No.: C-2053 Congress: ECR 2014 Type: Scientific Exhibit Authors: L. Hadid-Beurrier 1, M.-J. Waryn 2, Y. Ajavon 1, O.

More information

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures Vano E (1,2), Aviles P (1,2), Prieto C (2), Fernandez JM (1,2), Guibelalde

More information

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services Medical Physics and Informatics Review Vano et al. Online Quality Control for Diagnostic and Interventional Radiology Medical Physics and Informatics Review FOCUS ON: Eliseo Vano 1,2 Jose I. Ten 2,3 Jose

More information

Enteric Contrast Media for Pediatric Imaging

Enteric Contrast Media for Pediatric Imaging Enteric Contrast Media for Pediatric Imaging Michael J. Callahan, M.D. Department of Radiology Boston Children s Hospital Associate Professor of Radiology Harvard Medical School What is a contrast medium?

More information

Reducing Fluoroscopic Doses in Routine Musculoskeletal Joint Related Procedures

Reducing Fluoroscopic Doses in Routine Musculoskeletal Joint Related Procedures Reducing Fluoroscopic Doses in Routine Musculoskeletal Joint Related Procedures Mital Patel MD, Pranav Chitkara MD, Vikram Kinni MD, Rati Patel MD Naveen Subhas MD, Ryan Fisher PhD Disclosure No relevant

More information

Interventional Fluoroscopy

Interventional Fluoroscopy NATIONAL CANCER INSTITUTE THE SOCIETY OF INTERVENTIONAL RADIOLOGY Interventional Fluoroscopy Reducing Radiation Risks for Patients and Staff CONTENTS Introduction 1 Increasing use and complexity of interventional

More information

The Consistent Quality of Connected Radiology

The Consistent Quality of Connected Radiology The Consistent Quality of Connected Radiology Radiology is Evolving New compliance requirements, heightened radiation awareness and increasing technical complexity present new challenges to radiology professionals.

More information

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland Medical Physics and Informatics Original Research Seuri et al. Tracking Radiologic Procedures and Dose Medical Physics and Informatics Original Research FOCUS ON: Raija Seuri 1 Madan M. Rehani 2 Mika Kortesniemi

More information

SEMINAR 1. Radiological Protection Of Patients. Radiation Protection in Paediatric Cardiology. Dra. Patricia Miranda González

SEMINAR 1. Radiological Protection Of Patients. Radiation Protection in Paediatric Cardiology. Dra. Patricia Miranda González SEMINAR 1 Radiological Protection Of Patients Radiation Protection in Paediatric Cardiology Dra. Patricia Miranda González CARDIOVASCULAR DEPARTMENT, CALVO MACKENNA S HOSPITAL SANTIAGO CHILE Pediatric

More information

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE

DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY: A STRATEGY FOR INTRODUCING DIAGNOSTIC REFERENCE LEVELS TAKING INTO ACCOUNT THE NATIONAL PRACTICE Ph. R. Trueb #, A. Aroua *, H. Jung #, Th. Theiler #, J.-F. Valley

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels 2033-7 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dose Reference Levels Peter Homolka EFOMP Training Course on Medical Physics in Diagnostic

More information

The value of Dyna CT in the hepatocellular carcinoma treated by transcatheter arterial chemoembolisation

The value of Dyna CT in the hepatocellular carcinoma treated by transcatheter arterial chemoembolisation The value of Dyna CT in the hepatocellular carcinoma treated by transcatheter arterial chemoembolisation Poster No.: C-0030 Congress: ECR 2014 Type: Scientific Exhibit Authors: Y. C. wang, Q. Wang, H.

More information

Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT

Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT Responsibilities Worker (BSS 3.83) Workers shall: follow any applicable rules for protection use properly the monitoring devices and

More information

Methods for measuring fluoroscopic skin dose

Methods for measuring fluoroscopic skin dose Pediatr Radiol (2006) 36 (Suppl 2): 136 140 DOI 10.1007/s00247-006-0193-3 ALARA Stephen Balter Methods for measuring fluoroscopic skin dose Published online: 22 July 2006 # Springer-Verlag 2006 Abstract

More information

Siemens Healthineers introduces innovative robot supported Artis pheno angiography system

Siemens Healthineers introduces innovative robot supported Artis pheno angiography system Press Erlangen, October 25, 2016 DKOU 2016 in Berlin: Hall 4.2, Booth 44 introduces innovative robot supported Artis pheno angiography system Enabling personalized, minimally invasive surgery for multimorbid

More information

#Patients develop allergic or nephrotoxic reactions to iodinated media # carbon dioxide (CO2) gas angiography can be used.

#Patients develop allergic or nephrotoxic reactions to iodinated media # carbon dioxide (CO2) gas angiography can be used. 320-row multidetector CT angiography for hepatocellular carcinoma using CO2 gas instead of iodinated contrast agents: Experiment and preliminary clinical study Poster No.: C-1123 Congress: ECR 2011 Type:

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Multi-Energy Computed Tomography - New Opportunities In Imaging the Abdomen Materials can be

More information

Precision in Quantitative Imaging: Trial Development and Quality Assurance

Precision in Quantitative Imaging: Trial Development and Quality Assurance Precision in Quantitative Imaging: Trial Development and Quality Assurance Susanna I Lee MD, PhD Thanks to: Mitchell Schnall, Mark Rosen. Dan Sullivan, Patrick Bossuyt Imaging Chain: Patient Data Raw data

More information

Size-based protocol optimization using automatic tube current modulation and automatic kv selection in computed tomography

Size-based protocol optimization using automatic tube current modulation and automatic kv selection in computed tomography JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Size-based protocol optimization using automatic tube current modulation and automatic kv selection in computed tomography Robert

More information

EndoVascular hybrid Trauma and bleeding Management (EVTM)

EndoVascular hybrid Trauma and bleeding Management (EVTM) EndoVascular hybrid Trauma and bleeding Management (EVTM) Hands-on workshop Örebro, 7-8 Sept 2017 Endovascular and hybrid solutions for the bleeding patient; Aorta balloon occlusion (REBOA) usage, vascular

More information

Optimizing Radiation Dose for Samsung GC85A

Optimizing Radiation Dose for Samsung GC85A White Paper Article # WP201602-DR-Dose/ Issue Date 23 Feb., 2016 Optimizing Radiation Dose for Samsung GC85A Hong Eo, M.D. / Hyesun Hwang, M.D. / Samsung Medical Center Introduction ALARA (As Low As Reasonably

More information

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014 CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014 Answers for life. Answers for life. Is 1 msv the right dose for every patient? Female, 54 Abdominal CT Male,

More information

Trusted Performance. Smart Investment. 80 detector row Ultra Helical CT

Trusted Performance. Smart Investment. 80 detector row Ultra Helical CT TM Trusted Performance. Smart Investment. 80 detector row Ultra Helical CT 2 High performance, highly economical Increased productivity and patient safety Maximum clinical capabilities Are you looking

More information

Not for publication in the USA Erlangen, November 26, 2017

Not for publication in the USA Erlangen, November 26, 2017 Press Not for publication in the USA Erlangen, November 26, 2017 RSNA 2017 in Chicago: South Building, Hall A, Booth 1937 strengthens its CT portfolio by improving patient experience and expanding precision

More information

Poster No.: C-1217 Congress: ECR Scientific Exhibit. Authors:

Poster No.: C-1217 Congress: ECR Scientific Exhibit. Authors: Is rotational cholangiogram a feasible investigation for obstructive jaundice? A comparison study with conventional percutaneous transhepatic cholangiogram. Poster No.: C-1217 Congress: ECR 2013 Type:

More information

COMPREHENSIVE FAMILY OF INTERVENTIONAL SYSTEMS

COMPREHENSIVE FAMILY OF INTERVENTIONAL SYSTEMS COMPREHENSIVE FAMILY OF INTERVENTIONAL SYSTEMS When excellence means saving lives, installing the right lab is absolutely vital. Unlike other imaging systems, the cardiovascular lab is literally an extension

More information

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety Educating and Engaging your Staff in Radiation Safety Robert Vincent MD;CM FACC Professor of Pediatrics Emory University School of Medicine Ralf Holzer MD Associate Professor of Pediatrics Ohio State University

More information

Automatic Bone Removal Technique in Whole-Body Dual-Energy CT Angiography: Performance and Image Quality

Automatic Bone Removal Technique in Whole-Body Dual-Energy CT Angiography: Performance and Image Quality Vascular and Interventional Radiology Original Research Schulz et al. Bone Removal in CT Angiography Vascular and Interventional Radiology Original Research Boris Schulz 1 Katharina Kuehling Wolfgang Kromen

More information

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design GE Healthcare Introducing Veo * on Discovery * CT750 HD Great care by design Radiologists and clinicians demand the ultimate levels of clarity and detail in their images. In making an informed and confident

More information

European Journal of Vascular and Endovascular Surgery

European Journal of Vascular and Endovascular Surgery European Journal of Vascular and Endovascular Surgery 43 (2012) 393e397 Contents lists available at SciVerse ScienceDirect European Journal of Vascular and Endovascular Surgery journal homepage: www.ejves.com

More information

FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION

FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION FLUOROSCOPIC PATIENT & OPERATOR RISK REDUCTION Ray Dielman RSO ST ANTHONY S HOSPITAL James Menge CHP ARSO ST ANTHONY S HOSPITAL BAYCARE HEALTH SYSTEM 1 FLUROSCOPY EDUCATIONAL OBJECTIVES I. Understand the

More information