Measurement of Hepcidin-25 in Serum Using the Agilent 6490 Triple Quadrupole LC/MS with ifunnel Technology
|
|
- Dorcas Logan
- 6 years ago
- Views:
Transcription
1 Measurement of Hepcidin-25 in Serum Using the Agilent 6490 Triple Quadrupole LC/MS with ifunnel Technology Application Note Clinical Research Authors Christophe Hirtz, Constance Delaby, Pauline Bros, Jérôme Vialaret, and Sylvain Lehmann CHU Montpellier, Institut de Recherche en Biothérapie, Hôpital St Eloi, Laboratoire de Biochimie Protéomique Clinique and CCBHM, Montpellier, France Abstract A novel nano HPLC/MS method has been developed for the rapid, robust, reliable, and accurate determination of hepcidin-25 levels in human serum. It relies on straightforward pre-analytical steps that are cost effective, avoiding the need for solid phase extraction. This Application Note demonstrates how nano HPLC/MS can be used to deliver better reproducibility and a limit of detection (LOD) for hepcidin less than half that of current C-ELISA assays.
2 Introduction Both intestinal iron absorption and macrophage iron recycling are regulated by hepcidin, a 25 amino-acid (aa) cysteine-rich peptide first characterized in Isoforms of the peptide (20, 22, and 24-amino acids) also exist 4, but their clinical relevance is unknown. A number of methods have been developed for this purpose using multiple technologies, including radioimmunoassay (RIA), enzyme-linked immunosorbent assay (ELISA), and more recently, mass spectrometry analysis (SELDI-TOF and LC/MS/MS) 7 9. The immunoassays employ antihepcidin antibody, with the potential to recognize the different isoforms. The methods based on mass spectrometry can specifically detect all the hepcidin isoforms, and they can provide absolute quantitation through the use of a spiked internal standard. All of these methods have demonstrated limited specificity and reproducibility as well as lack of concurrence in quantified levels of hepcidin, across the assay technologies. In an effort to resolve these issues, the different methods have been recently compared through round robin studies 7,8. The results demonstrated that development of a reliable assay remains quite challenging. The use of an internal standard was recommended for mass spectrometry based methods. These studies also called for development of a calibrator mimicking human serum and development of a consensus on calibrator levels. This Application Note describes a recently published liquid chromatography tandem-mass spectrometry (LC/MS/MS) analytical method using a nano HPLC-chip on the Agilent 6490 Triple Quadrupole LC/MS for the reliable quantification of hepcidin-25 in human serum 10. The method was evaluated for specificity, reproducibility, repeatability, and determination of LODs, limits of quantitation (LOQs), and linearity. These results were compared to the well-characterized C-ELISA method used to quantify serum hepcidin-25 in human samples 11. Although discrepancies were observed in the absolute values of hepcidin measured, there was good correlation between the two approaches (R 2 = 0.96). 2
3 Experimental Reagents and standards Reagents were obtained as described 10. The human endogenous hepcidin standard (DTHFPICIFCCGCCHRSKCGMCCKT) and the internal hepcidin standard (DTHFPICIFCCGCCHRSKCGMCCKT) [ 13 C 6, 15 N 4 ] Arg 16 were purchased from Eurogentec (Seraing, Belgium) with a purity > 97 % assessed by reverse phase HPLC and mass spectrometry. Samples The serum samples included in this study were part of a biobank (officially registered # DC ) and all participants gave their informed consent to use their sample for research conducted in accordance with the local ethics committee. Instruments This method was developed using an Agilent 1290 Infinity HPLC-Chip/MS System coupled to an Agilent 6490 Triple Quadrupole LC/MS equipped with a nano ESI source. The instrument conditions are listed in Table 1. Sample preparation Serum samples (50 µl) were extracted using trichloroacetic acid as described 10. Data analysis Agilent MassHunter Quantitative Analysis Software was used for data treatment. For the calibration curve, all replicates results were loaded into the software database, and the automatic quantification method was then used to obtain the appropriate calibration curve type with the best fit to the experimental data. Statistical analyses were performed using MedCalc (7.3) software. Bland and Altman 12 and Deming adjusted regression curves 13 were used to test the commutability of the methods. Table 1. HPLC-Chip and MS instrument conditions for the hepcidin quantification method. LC conditions HPLC-Chip ProtID-Chip-43 (II) 43 mm 300 Å C18 chip with 40 nl trap column (p/n G ) Injection volume 4 µl in 15 % B, followed by 7 µl of flush volume Mobile phase A = 97:3:1 v/v/v (water/acetonitrile/formic acid) B = 10:90:1 v/v/v (water/acetonitrile/formic acid) Run time 9 minutes Flow rate 2.5 µl/min for sample loading, 0.6 µl/min for separation Gradient program 7 minutes linear gradient from 3 % to 100 % B, then 2 minutes wash (100 % B) and 1 minute equilibration (97 % A) MS conditions Mode ESI mode, positive ionization; MRM Capillary tension 1,700 2,100 V Nebulizer gas flow 11 L/min Nebulizer gas temperature 150 C Cell accelerator voltage 4 V Delta EMV 200 V Fragmentor voltage 380 V Dwell time 40 msec MS1 Wide MS2 Wide Ion funnel voltage pressure Low: 80 V; High: 180 V Acquisition parameters The acquisition parameters used in the multiple reaction monitoring (MRM) method for both endogenous hepcidin and the isotopically-labeled hepcidin internal standard are shown in Table 2. Table 2. Acquisition parameters for the hepcidin MRM method. Precursor ion (m/z) Product ion (m/z) Collision energy (ev) Endogenous hepcidin Isotopically-iabeled hepcidin internal standard 1, ,2 11 Quantifier ion; all others are qualifier ions. 3
4 Results and Discussion Analytical validation A typical total ion chromatogram and extracted ion chromatograms using MRM are shown in Figure 1 for serum spiked with endogenous hepcidin standard and isotopically-labeled hepcidin standard. Calibration curves were constructed in MassHunter software over the range ng/ml, based on hepcidin quantification values determined using orthogonal analysis methods. The calibration curve equation was linear (y = x ) when the origin was ignored and the equation was weighted 1/y. The calibration coefficient (R 2 ) was (Figure 2). The specificity of the method was confirmed by the absence of a hepcidin-25 peak at the appropriate retention time in blank samples (normal goat serum containing 0 ng/ml of hepcidin-25). Analysis of four replicates of one calibration curve point on the same day was used to evaluate intra-assay precision and accuracy. The intra-assay precision was 7 % at 50 ng/ml and 28 % at 5 ng/ml. The intra-assay accuracy was approximately 100 % at 20, 50, 100, and 200 ng/ml and almost 80 % at 5 and 10 ng/ml (Table 3). Inter-assay precision and accuracy were evaluated by quantifying the 50 and 200 ng/ml calibration standards on four separate days. Inter-assay precision was around 9 10 %, and the accuracy was close to 80% at both concentrations. The LOD and LOQ were 2 ng/ml and 6 ng/ml, respectively. Total ion chromatogram (TIC) and extracted ion chromatogram (XIC) of hepcidin-25 in serum samples using nano HPLC-MS 10 4 Hepcidin peak Relative response TIC on serum sample MRM endogenous hepcidin (559 & 694) MRM labeled hepcidin (560 & 696) y = x R 2 = Figure 1. Typical TIC and extracted ion chromatograms (EICs) for endogenous hepcidin standard and isotopically-labeled hepcidin spiked into human serum Relative concentration Figure 2. Calibration curves of hepcidin-25 using model matrix (normal goat serum) in the concentration range of 0 to 500 ng/ml (0, 5, 10, 20, 50, 100, 200, ng/ml). The calibration equation was linear when ignoring origin and weighted 1/y. R 2 = Table 3. Intra-assay coefficient of variation and accuracy (%). Hepcidin-25 concentration (ng/ml) Intra-assay coeffcient of variation (% CV) Intra-assay accuracy (%) 4
5 Comparison to competitive ELISA Serum samples previously quantified using the C-ELISA method 11 were analyzed using this nano HPLC/MS MRM method. A positive significant correlation with R 2 = 0.89 and Pearson coefficient P < (Figure 3) However, the slope of the linear correlation (5,0255x) indicates that the MRM assay provided values five times lower than those obtained with the C-ELISA. This result is likely related to the fact that the MS methods quantify only the 25 aa isoform and, therefore, are fully specific. The difference may also be due in part to the differences in the origin and purity of the material used for the standard curves in each of the two methods. Hepcidin concentration by ELISA (ng/ml) y = x R 2 = Hepcidin concentration by MRM (ng/ml) Figure 3. Comparison of hepcidin-25 concentrations measured in serum samples using nano HPLC/MS (MRM) and ELISA methods. A positive significant correlation has been obtained with an R 2 value of 0.89 and a Pearson coefficient P < This nano HPLC/MS method generated a coefficient of variation (CV) of only 9 % for 30 participants, while the CV for the C-ELISA study described by Ganz et al. ranged from 5 19 % 11. The LOD for the nano HPLC/MS method is also less than half of the LOD determined for the C-ELISA method, at 2 ng/ml, versus 5.5 ng/ml. We have, therefore, developed a robust and reliable analytical method for hepcidin-25 suitable for quantification of hepcidin-25 in human serum. Conclusions This nano HPLC/MS method developed on the Agilent 6490 Triple Quadrupole LC/MS is suitable for the accurate, robust, and reproducible measurement of hepcidin in human serum samples. It relies on straightforward pre-analytical steps that are cost effective and avoid the need for solid phase extraction. While the obtained hepcidin levels differed from those determined using the reference C-ELISA assay, this nano HPLC/MS method has the advantage of being able to detect and quantify the truncated 20, 22, and 24-amino acid isoforms of hepcidin, whose clinical relevance is not yet known. 5
6 References 1. C.H. Park, et al. Hepcidin, a urinary antimicrobial peptide synthesized in the liver. J. Biol. Chem. 276, (2001). 2. C. Pigeon, et al. A new mouse liver-specific gene, encoding a protein homologous to human antimicrobial peptide hepcidin, is overexpressed during iron overload J. Biol. Chem. 276, (2001). 3. G. Nicolas, et al. Lack of hepcidin gene expression and severe tissue iron overload in upstream stimulatory factor 2 (USF2) knockout mice Proc. Natl. Acad. Sci. USA 98, (2001). 4. B. Rochat, et al. Validation of hepcidin quantification in plasma using LC-HRMS and discovery of a new hepcidin isoform Bioanalysis 5, (2013). 5. J.J. Kroot, et al. Immunochemical and mass-spectrometry-based serum hepcidin assays for iron metabolism disorders Clin. Chem. 56, (2010). 6. J.J. Kroot, et al. Hepcidin in human iron disorders: diagnostic implications Clin. Chem. 57, (2011). 7. J.J. Kroot, et al. Results of the first international round robin for the quantification of urinary and plasma hepcidin assays: need for standardization Haematologica 94, (2009). 8. J.J. Kroot, et al. Second round robin for plasma hepcidin methods: first steps toward harmonization Am. J. Hematol. 87, (2012). 9. C. Delaby, C. Beaumont, Serum hepcidin assay in 2011: where do we stand? Ann. Biol. Clin. (Paris) 70, (2012). 10. C. Delaby, et al. EuPA Open Proteomics, Article in revision. 11. T. Ganz, et al. Immunoassay for human serum hepcidin Blood 112, (2008). 12. J.M. Bland, D.G. Altman Statistical methods for assessing agreement between two methods of clinical measurement Lancet 1 (8476), (1986). 13. P.J. Cornbleet, N. Gochman Incorrect least-squares regression coefficients in method-comparison analysis Clin. Chem. 25, (1979). 14. B.J. Sasu, et al. Serum hepcidin but not prohepcidin may be an effective marker for anemia of inflammation (AI) Blood Cells Mol. Dis. 45, (2010). For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 2014 Published in the USA, August 5, EN
Application Note. Authors. Abstract
Automated, High Precision Tryptic Digestion and SISCAPA-MS Quantification of Human Plasma Proteins Using the Agilent Bravo Automated Liquid Handling Platform Application Note Authors Morteza Razavi, N.
More informationEPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube
EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB
More informationApplication Note. Abstract. Authors. Introduction. Environmental
EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma
More informationAutomated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module
Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module Application Note Environmental Authors Shane Snyder Department of
More informationApplication Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany
Quantification of trace-level herbicides in drinking water by online enrichment with the Agilent Infinity Series Online-SPE Solution and Triple Quadrupole MS Detection Application Note Environmental Author
More informationAgilent 6430 Triple Quadrupole LC/MS System
Agilent 6430 Triple Quadrupole LC/MS System Ideal Quantitative LC/MS for UHPLC with Dynamic MRM and Fast Polarity Switching, plus Unsurpassed Sensitivity with Chip LC Summary The Agilent 6430 Triple Quadrupole
More informationMonitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS
Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 641 Triple Quadrupole LC/MS Application Note Environmental Author Toni Hall Wessex Water Bath, UK Abstract A method for the
More informationOn-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water
application note On-line SPE-LC/MS/MS of Pesticides On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water API 3200 LC/MS/MS System Overview The feasibility of
More informationPLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC
PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC Analysis of Intact and Fragmented mabs and ADC Application Note Biotherapeutics and Biologics Author Suresh Babu C.V. Agilent
More informationPrecise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF
Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Application Note Author David L. Wong Agilent Technologies, Inc. Santa Clara, CA, USA Introduction Monoclonal
More informationConfident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions
Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High
More informationN-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF
N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF Application Note Authors Xianming Liu, Wei Zhang, Yi Du, Sheng Yin, Hong Que, and Weichang Zhou WuXi AppTec iopharmaceuticals
More informationExploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma
Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Yun Wang Alelyunas, Mark D. Wrona, Jim Murphy, Angela Doneanu, Gregory
More informationPeptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS
Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS Application Note BioPharma Authors Alex Zhu, James Martosella, and Phu T Duong Agilent Technologies, Inc Introduction Peptide
More informationMinimize Method Development Time for Pesticide Screening Using LC/MS Application Kit
Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit Dr. Jerry Zweigenbaum Senior Applications Scientist Food Safety Page 1 Agilent 1260 LC SampliQ Extraction and Dispersive
More informationA Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics
A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics Catalin Doneanu,, Ph.D. Biopharmaceutical Sciences, Waters September 16, 2009 Mass Spec 2009 2009 Waters Corporation Host Cell Proteins
More informationFast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides
Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides High-pH-stable, superficially porous particle columns for LC/UV and LC/MS Application Note Biologics and Biosimilars Authors
More informationSize Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column
Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Application Note Bio-Pharmaceutical Authors M. Sundaram Palaniswamy and Andrew Coffey Agilent Technologies,
More informationAB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual
AB SCIEX TripleTOF 5600 SYSTEM High-resolution quant and qual AB SCIEX TripleTOF 5600 SYSTEM AB SCIEX is a global leader in the development of life science analytical technologies that help answer complex
More informationThermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform
Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips Next generation immunoaffinity Robust quantitative platform Immunoaffinity sample preparation Thermo Scientific Mass Spectrometric Immunoassay
More informationQuantitation of cyanotoxins in drinking water according to EPA 544 guidelines
APPLICATION NOTE 64968 Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines Authors Ali Haghani, 1 Andy Eaton, 1 Neloni Wijeratne, Claudia Martins 1 Eurofins Eaton Analytical,
More informationA Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System
A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A high-throughput method for detecting ultra-low levels (0.5 pg/ml) of a therapeutic peptide in human
More informationInvestigating Biological Variation of Liver Enzymes in Human Hepatocytes
Investigating Biological Variation of Liver Enzymes in Human Hepatocytes SWATH Acquisition on the TripleTOF Systems Xu Wang 1, Hui Zhang 2, Christie Hunter 1 1 SCIEX, USA, 2 Pfizer, USA MS/MS ALL with
More informationMultiQuant Software Version 3.0
MultiQuant Software Version 3.0 Improving Data Quality and Processing Throughput with Better Peak Integration, Quantitative and Qualitative Compound Review for the Analysis of Food, Drinking Water, and
More informationDisulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column
Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram
More informationVyužití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS
Využití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS Michal Godula Ph.D. Thermo Fisher Scientific The world leader in serving science
More informationHeart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System
Heart-cut 2D-LC/M approach for pharmaceutical impurity identification using an Agilent 640 Q-TF LC/M ystem Application ote Pharmaceuticals Author iji Joseph Agilent Technologies, Inc. Bangalore, India
More informationHighly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs
Technical Overview Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs Authors Stephen Madden, Crystal Cody, and Jungkap Park Agilent Technologies, Inc. Santa Clara, California,
More informationCell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS
Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS Application Note Biopharmaceuticals and Biosimilars Authors Emmie Dumont, Isabel Vandenheede, Pat Sandra, Koen Sandra Research
More informationAgilent 6400 Series. System. Concepts Guide. The Big Picture. Agilent Technologies
Agilent 6400 Series Triple Quad LC/MS System Concepts Guide The Big Picture Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means
More informationThe clearly better choice
Agilent Accurate-Mass 6500 Series Q-TOF and 600 Series TOF LC/MS Systems The clearly better choice in TOF and Q-TOF LC/MS Our measure is your success. products applications soft ware services The unmatched
More informationAdvantages of Ion Mobility QTOF for Characterization of BioPharma Molecules
Advantages of Ion Mobility QTOF for Characterization of BioPharma Molecules Add a New Dimension to your Research Capability with Agilent s New Drift Ion Mobility QTOF System David Wong, Ph.D. Sr. Application
More informationAccurately Identify and Quantify One Hundred Pesticides in a Single GC Run
Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract A selected target compound list of 195 various pesticides
More informationApplication Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany
D-LC/MS Characterization of Charge Variants Using Ion Exchange and Reversed-Phase Chromatography Multiple Heart-Cutting D-LC Analysis of Innovator versus Biosimilar Monoclonal Antibodies Application Note
More informationProteomic analysis of biological fluids
Proteomic analysis of biological fluids Anne Gonzalez, CNRS-IPBS, Toulouse Équipe «Analyse Protéomique et Spectrométrie de Masse des Biomolécules» Odile Schiltz-Bernard Monsarrat Workshop «Protéomique
More informationAntibody Analysis by ESI-TOF LC/MS
Application Note PROTEOMICS METABOLOMICS GENOMICS INFORMATICS GLYILEVALCYSGLUGLNALASERLEUASPARG CYSVALLYSPROLYSPHETYRTHRLEUHISLYS Antibody Analysis by ESI-TOF LC/MS Authors Lorenzo Chen, Merck & Company,
More informationQuantitation of Microcystins and Nodularins in Water Samples Using LC-MS/MS
Quantitation of Microcystins and Nodularins in Water Samples Using LC-MS/MS Craig Butt1, April Quinn-Paquet1, and Christopher Borton2 SCIEX (Framingham, MA) 2 SCIEX (Redwood City, CA) 1 Overview This application
More informationSimultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis
Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis Mark Wrona, 1 Paul Rainville, 1 Eric Langlois, 2 Nigel Ewing,
More informationRapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution
Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution Rapidly Identify Top Metabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Shaokun Pang and Ian Moore SCIEX,
More informationRapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution
Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution Rapidly Identify Major Catabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Ian Moore and Jinal Patel
More informationApplication Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona
Analysis of an Antibody Drug Conjugate using MSIA D.A.R.T. S. Technology, an Integral Part of Ligand Binding Mass Spectrometric Immunoassay (LB-MSIA) Workflow Application Note Kwasi Antwi, Amanda Ribar,
More informationCharacterize Fab and Fc Fragments by Cation-Exchange Chromatography
Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for
More informationChanging the Game in LC-MS/MS. New Tools for Unprecedented Performance in Residue Analysis. Dr. Thomas Glauner
Changing the Game in LC-MS/MS New Tools for Unprecedented Performance in Residue Analysis Dr. Thomas Glauner EMEA LC/MS Food Segment Scientist Agilent Technologies Waldbronn The New 6490 Triple Quadrupole
More informationmab Titer Analysis with the Agilent Bio-Monolith Protein A Column
mab Titer Analysis with the Agilent Bio-Monolith Protein A Column Application Note Biopharmaceuticals and Biosimilars Authors Emmie Dumont, Isabel Vandenheede, Pat Sandra, and Koen Sandra Research Institute
More informationEnabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis
Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis SWATH Acquisition on the TripleTOF 5600+ System Samuel L. Bader, Robert L. Moritz Institute of Systems Biology,
More informationIncreasing MS Sensitivity for Proteomics
Increasing MS Sensitivity for Proteomics Discovery to Pathways Analysis: Using Agilent 6550 ifunnel Q-TOF Vadi Bhat, Ph.D. Senior Applications Scientist Agilent Technologies Wilmington, DE, USA 1 April
More informationMetabolic Changes in Lung Tissue of Tuberculosis-Infected Mice Using GC/Q-TOF with Low Energy EI
Metabolic Changes in Lung Tissue of Tuberculosis-Infected Mice Using GC/Q-TOF with Low Energy EI Application Brief Authors Mª Fernanda Rey-Stolle 1, Vineel P. Reddy 2, Santiago Angulo 1, Adrie J.C. Steyn
More informationCharacterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody
Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody Application Note BioPharma Authors James Martosella, Phu Duong, and Alex Zhu Agilent Technologies, Inc.
More informationProteomics And Cancer Biomarker Discovery. Dr. Zahid Khan Institute of chemical Sciences (ICS) University of Peshawar. Overview. Cancer.
Proteomics And Cancer Biomarker Discovery Dr. Zahid Khan Institute of chemical Sciences (ICS) University of Peshawar Overview Proteomics Cancer Aims Tools Data Base search Challenges Summary 1 Overview
More informationImproved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental
Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Samples Michael S.Young, Jeremy C. Shia, Kim vantran, Kevin M. Jenkins and Masayo Yabo Waters Corporation 34 Maple Street,
More informationValidation of a concentration assay using Biacore C
GE Healthcare Application Note 48 Biacore systems Validation of a concentration assay using Biacore C Guideline for development of a GxP - compliant concentration assay Support for informed decision-making
More informationProteomics. Proteomics is the study of all proteins within organism. Challenges
Proteomics Proteomics is the study of all proteins within organism. Challenges 1. The proteome is larger than the genome due to alternative splicing and protein modification. As we have said before we
More informationHigh Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science
High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers The world leader in serving science Goals Explore the capabilities of High Resolution Accurate Mass
More informationRapidly Characterize Antibody- Drug Conjugates and Derive Drug-to-Antibody Ratios Using LC/MS
Rapidly Characterize Antibody- Drug Conjugates and Derive Drug-to-Antibody Ratios Using LC/MS Ning Tang, Ph.D. Agilent Technologies Santa Clara, California, USA Heterogeneity in ADCs Complexity T-DM1 (Genentech)
More informationValidation Report 19
EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark Validation Report 19 Determination of pesticide residues in oat, rye and wheat by GC-MS/MS and LC-MS/MS (QuEChERS
More informationA Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples
A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples Mary Lame, Hua Yang, Sherri Naughton, and Erin Chambers Waters Corporation, Milford,
More informationKey Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics
Metabolomic Profiling in Drug Discovery: Understanding the Factors that Influence a Metabolomics Study and Strategies to Reduce Biochemical and Chemical Noise Mark Sanders 1, Serhiy Hnatyshyn 2, Don Robertson
More informationTechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION
METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES Jose M. Castro-Perez 1, Jane Kirby 1, Laurent Leclercq 2, Claude Delatour 2, Jean-Michel Mineur 2,
More informationThe LC-MS/MS Workhorse. SCIEX 4500 Series Mass Spectrometers
The LC-MS/MS Workhorse SCIEX 4500 Series Mass Spectrometers The SCIEX 4500 Series takes the legendary API 4000 platform and intelligently re-engineers it to set a new benchmark for reliable quantitation.
More informationFeaturing Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System
Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 15EX LC/MS System compact, rugged, The API 15EX LC/MS System is the most rugged and reliable single quadrupole
More informationLaboratory Reagents. for life science research.
Laboratory Reagents Research, quality control, or routine analysis whatever the field of activity, where there is a need for laboratory reagents, Thermo Fisher Scientific has a suitable product. We offer
More informationAutomated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application
Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method 321.8 Application ICP-MS Author Jason A. Day, Anne Vonderheide, Joseph A. Caruso University of Cincinnati Cincinnati,
More informationXevo G2-S QTof and TransOmics: A Multi-Omics System for the Differential LC/MS Analysis of Proteins, Metabolites, and Lipids
Xevo G2-S QTof and TransOmics: A Multi-Omics System for the Differential LC/MS Analysis of Proteins, Metabolites, and Lipids Ian Edwards, Jayne Kirk, and Joanne Williams Waters Corporation, Manchester,
More informationMaximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns
Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns Hongwei Xie, Martin Gilar, and Jeff Mazzeo Waters Corporation, Milford, MA U.S. APPLICATION BENEFITS The ACQUITY UPLC
More informationMassPREP On-Line Desalting Cartridge
CONTENTS I. INTRODUCTION II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION
More informationThermo Scientific TSQ Quantum Access MAX
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific TSQ Quantum Access MAX UNSURPASSED PRICE TO PERFORMANCE Superior
More informationProudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment QTRAP 5500
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment QTRAP 5500 Powerful synergy results when the world s most sensitive triple
More informationFor the quantitative detection of human IL6 in serum, plasma, cell culture supernatants and urine.
m andw da a For the quantitative detection of human IL6 in serum, plasma, cell culture supernatants and urine. general information Catalogue Number Product Name Species cross-reactivity Range (calibration
More informationDevelopment of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS
TECHNICAL NOTE Pesticides in Tobacco Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS API 4000 LC/MS/MS System Overview This application note
More informationPeptide Mapping: A Quality by Design (QbD) Approach
Peptide Mapping: A Quality by Design (QbD) Approach Application Note Bio-Pharmaceutical Authors Sreelakshmy Menon and Suresh babu C.V. Agilent Technologies, Inc. Richard Verseput S-Matrix Corporation Abstract
More informationA Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS
A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS Zhiqi Hao, 1 Yi Zhang, 1 David Horn, 1 Seema Sharma, 1 Shiaw-Lin
More informationPharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study
Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study Featuring the SCIEX BioBA Solution Xi Qu 1, Shuyu Hou 1, Meghan McCann 1, Caroline Becker 1, Xun Wang 1,
More informationDetermination of Iodinated Contrast Media in Aqueous Samples by Direct-Injection LC-MS/MS
Determination of odinated Contrast Media in Aqueous Samples by Direct-njection LC-MS/MS Application ote Environmental Authors Gerd Vanhoenacker, Mieke Steenbeke, Frank David, and Pat Sandra Research nstitute
More informationAnalysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column
Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb µm Column Application Note BioPharma Authors Xiaomi Xu and Phu T Duong Agilent Technologies, Inc. Abstract Nearly all proteins undergo
More informationImmunoassay Kit Catalog # KCA0021. Canine. C-Reactive Protein
Immunoassay Kit Catalog # KCA0021 Canine C-Reactive Protein BioSource International, Inc. 542 Flynn Road Camarillo, California 93012 USA Tel: 805-987-0086 800-242-0607 FAX: 805-987-3385 email: tech.support@biosource.com
More informationCertified Reference Material - Certificate of Analysis
Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.
More informationDETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS
DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS Darcy Shave and Pete Alden Waters Corporation, Milford, MA, U.S.A. INTRODUCTION Plasma protein binding (PPB) can significantly affect the therapeutic action
More informationQUANLYNX APPLICATION MANAGER
QUANLYNX APPLICATION MANAGER OVERVIEW INTRODUCTION QuanLynx, an Application Manager included with Waters MassLynx Software, is designed for quantitative analysis. Simply acquiring mass spectrometry data
More informationThermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance
Thermo Scientific EASY-nLC 1200 System Leading in simplicity and performance Peak performance made EASY Effortless ultra high performance for everybody A straightforward LC-MS solution Optimized and integrated
More informationSupplementary Figure 1. Utilization of publicly available antibodies in different applications.
Supplementary Figure 1 Utilization of publicly available antibodies in different applications. The fraction of publicly available antibodies toward human protein targets is shown with data for the following
More informationA Comparison of AlphaLISA and TR-FRET Homogeneous Immunoassays in Serum-Containing Samples
application Note A Comparison of and Homogeneous Immunoassays in Serum-Containing Samples Authors Anuradha Prasad, PhD, Catherine Lautenschlager, PhD, Stephen Hurt, PhD, David Titus, PhD and Stéphane Parent,
More informationVICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY
The European Agency for the Evaluation of Medicinal Products Veterinary Medicines Evaluation Unit CVMP/VICH/591/98-FINAL London, 10 December 1998 VICH Topic GL2 (Validation: Methodology) Step 7 Consensus
More informationWith the globalization of trade, food production
November 2008 Comprehensive Screening, Confirmation, and Quantification of Organic Pesticides in Foods by GC MS and LC MS The global focus on keeping food safe has intensified, resulting in major changes
More informationBioanalytical method validation: An updated review
Review Article www.phmethods.org Bioanalytical method validation: An updated review Abstract The development of sound bioanalytical method(s) is of paramount importance during the process of drug discovery
More informationAdvanced QA/QC characterization MS in QC : Multi Attribute Method
Advanced QA/QC characterization MS in QC : Multi Attribute Method Global BioPharma Summit The world leader in serving science A Complex Problem: Drug Safety and Quality Safety Is the product safe to use?
More informationAgilent 6500 Series Q-TOF LC/MS System
Agilent 6500 Series Q-TOF LC/MS System Quick Start Guide What s New in Data Acquisition 3 Where to find information 7 Getting Started 9 Step 1. Start the Data Acquisition software 10 Step 2. Prepare the
More informationCurrent practice and future vision on metabolite profiling and quantification in Drug development
Current practice and future vision on metabolite profiling and quantification in Drug development Geert Mannens EBF Focus workshop Metabolite Profiling and Quantification Strategies in Drug R&D Sept 25,
More informationAnalysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation
Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation Cees Bruggink, Frans Schoutsen Thermo Fisher Scientific,
More informationIntegrating an exploratory BM in an early clinical stage in Pharma R&D
Integrating an exploratory BM in an early clinical stage in Pharma R&D Ulrich Kunz, Translational Medicine and Clinical Pharmacology, Boehringer Ingelheim Pharma GmbH & Co. KG, Biberach, Germany Content
More informationRapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins
Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins Esther Lewis, 1 Kevin Meyer, 2 Zhiqi Hao, 1 Nick Herold,
More informationTyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1
Application Note Method Validation of the AutoMate-Q40 for Pesticide Residues in Grapes Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1 Abstract The QuEChERS (Quick-Easy-Cheap-Effective-Rugged-Safe)
More informationAbsolute Quantification of Yeast Kinases by LC/MS/MS using QconCAT and MRM Technologies
Absolute Quantification of Yeast Kinases by LC/MS/MS using QconCAT and MRM Technologies Philip J Brownridge, 1 Victoria Harman, 1 Simon Cubbon, 2 Johannes PC Vissers, 2 Craig Lawless, 3 Simon J Hubbard,
More informationThe Waters strategy for the quantification of proteins
Protein quantification workshop Barcelona, 13 th November 2012 The Waters strategy for the quantification of proteins Robert Tonge Ph.D. Principal Scientist, European Omics Waters Corporation MS Technology
More informationTurboMass GC/MS Software
TurboMass GC/MS Software GAS CHROMATOGRAPHY / MASS SPECTROMETRY P R O D U C T N O T E Today s demanding laboratory requires software that is easy to learn yet offers sophisticated instrument control, robust
More informationChem 321 Lecture 23 - Liquid Chromatography 11/19/13
Chem 321 Lecture 23 - Liquid Chromatography 11/19/13 Student Learning Objectives High Performance Liquid Chromatography With the advent of relatively inexpensive and reliable pumps, the development of
More informationMoving steroid analyses from immunoassay to LC-MS/MS. Arieh Cohen
Moving steroid analyses from immunoassay to LC-MS/MS challenges and pitfalls Arieh Cohen Neonatal screening and ormone analysis Dept. Of Clinical Biochemistry Statens Serum Institut aco@ssi.dk Background
More informationTony Mire-Sluis Vice President, Corporate, Product and Device Quality Amgen Inc
The Regulatory Implications of the ever increasing power of Mass Spectrometry and its role in the Analysis of Biotechnology Products Where do we draw the line? Tony Mire-Sluis Vice President, Corporate,
More informationBioConfirm B Presentation
BioConfirm B.08.00 Presentation 1 Major Enhancements to Increase Your Productivity The Latest Version of our BioConfirm Product, B.08.00 BioConfirm is protein and peptide characterization software using
More informationApplication Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider
Combining small-scale purification and analysis of monoclonal antibodies on one instrument Protein purification with high-volume injection using the Agilent 126 Infinity Bio-inert Quaternary LC System
More information