1 November 2008 Comprehensive Screening, Confirmation, and Quantification of Organic Pesticides in Foods by GC MS and LC MS The global focus on keeping food safe has intensified, resulting in major changes in the number of pesticides that are being regulated and monitored, as well as the allowable levels of those pesticides in food. These new regulations drive a need for a wide range of comprehensive and complementary analytical methods and instrumentation that possess the throughput, sensitivity, and breadth of pesticide compound coverage to meet the demand of testing laboratories all over the world. This article provides an overview of the instrument platforms, tools, and workflow for analyzing pesticides. Philip Wylie, Jerry Zweigenbaum, Melissa Churley, Chin-Kai Meng, and Cao Zhe With the globalization of trade, food production and distribution have become truly international businesses. When we dine out, the fish might come from Japan, the rice from Australia, the spices from China, and the strawberries from Mexico. We take it for granted that the food we eat is safe and free from contamination that could make us seriously ill. One of the 20th-century advancements that made this cornucopia possible was the development of pesticides. Crops that were previously devastated by pest damage can now be produced at very high yields to meet worldwide demand. However, all pesticide compounds are potentially harmful to humans, and safeguards must be in place to assure that consumers are not exposed to dangerous levels of these pesticides. Many countries have had regulations regarding the use of pesticides, and their allowable levels in foods, for many years. With the increasing globalization of the food industry, there is greater scrutiny on food safety, resulting in major changes in the number of pesticides that are being regulated and monitored, as well as the allowable levels of those pesticides in food. A Global Effort to Keep Food Safe The Food Quality Protection Act (FQPA) of 1996 updated federal regulation of pesticides in the United States to implement a single health-based standard for all pesticides in foods. It provides special protection for infants and children, expedites approval of safe pesticides, and requires periodic evaluation of pesticide registrations and tolerances to ensure that the scientific data supporting registrations will remain up to date in the future. Under this mandate, the US Food and Drug Administration (FDA) performs regulatory monitoring by sampling individual lots of domestically produced and imported foods

2 their products to assure compliance with FQPA, the EU regulation, and the Japan Positive List, if they hope to sell their products in the U.S., Europe, and Japan, which generate much of the world s demand for imported food. These new regulations drive a global need for analytical methods that possess the throughput, sensitivity, and breadth of pesticide compound coverage to meet the demand of testing laboratories all over the world that must ensure compliance. Figure 1: GC MS analysis of p,p -DDE in spinach. Data were generated using a J&W DB-5ms column, a model 7890A GC system, and a model 5975C MSD mass spectrometer (Agilent Technologies). and analyzes them for pesticide residues to enforce established tolerances ( Domestic samples are collected at the point of production, and import samples are collected at the point of entry into U.S. commerce. Products typically are analyzed as the unwashed, whole, raw commodity. The FDA also performs a total diet study (TDS) on four regional market baskets per year, each basket comprising about 300 foods that represent the average U.S. consumer s diet. The foods are prepared as they would be consumed (table-ready) and tested for regulated pesticides. The FDA also performs focused sampling that is generally used to follow up on suspected problem areas or acquire residue data on food commodities that are not usually covered during regulatory monitoring. There are more than 1000 registered pesticides in the U.S., and approximately 400 with tolerances established by the Environmental Protection Agency and enforced by FDA. The European Union (EU) has also recently updated a harmonized regulation of pesticide residues in food, with the new standard taking effect on September 8, 2008 ( This regulation sets maximum residue levels (MRL) for food and animal feed. The MRL is the highest level of a pesticide residue that is legally tolerated. The regulation covers around 1100 pesticides currently or formerly used in agriculture in or outside the EU. For pesticides that are not specifically mentioned, a general default MRL of 0.01 mg/kg (10 ppb) applies. The EU designates the minimum number of samples to be taken by each member state, and EU reference laboratories coordinate, train staff, develop methods of analysis, and organize tests to evaluate the skills of the various national control laboratories. Japan also adopted a stricter pesticide regulation in 2006, referred to as the Positive List system, which establishes MRLs for about 500 pesticides ( Produce with pesticide residues exceeding these MRLs or the default tolerance of the uniform limit of 0.01 ppm (10 ppb) for those pesticides for which there is no Japanese MRL cannot be marketed in Japan. The regulation also lists 16 pesticides that must not be detected in food. Japan tested more than 200,000 food samples in The increased international focus on pesticide contamination in food impacts food producers in virtually any country, since the food market is a global one. Producers in Thailand, for example, must test A Broad Range of Complementary Analytical Methods Is Needed Food testing laboratories need the ability to detect and quantify hundreds of pesticides, in myriad foodstuffs, at very low levels of contamination. They might be screening for a set of hundreds of known target compounds, using a defined test method for each target in the set. Alternatively, a testing laboratory might wish to look for the presence of any pesticide ( unknowns ), without the need for defining a method for each possible compound. They are often called on to do both types of testing, and no single analytical approach can provide the flexibility required to meet this need. Wide variations in the chemical properties of pesticide contaminants and the necessity to detect a very large number of compounds require a range of chromatography and mass spectrometry (MS) systems. For pesticides that can be vaporized easily without degradation, gas chromatography (GC) MS (single quadrupole) is an ideal analytical tool, due to the availability of large libraries of pesticide spectra and deconvolution software. This approach can be used to screen, confirm, and quantify both target and unknown compounds. Many foodstuffs, such as garlic and ginger, are very complex, or dirty, due to the presence of a very large number of background compounds. GC triple quadrupole MS (QQQ) is very useful for screening, confirming, and quantifying trace-level target compounds in these complex matrices. Pesticides that are quite polar, not easily vaporized, thermally labile, or not easily derivatized are best analyzed using liquid chromatography (LC) methods. LC QQQ is particularly useful for analyzing sets of known target compounds.

3 Abundance Abundance Abundance m/z m/z m/z Figure 2: Confirmation of p,p -DDE in spinach using DRS and the Pesticide and Endocrine Disruptor spectral library (Agilent Technologies). A unique fragment ion for each pesticide is routinely used for screening of each target compound a second precursor to product ion transition can then be used to confirm the identity of the compounds. The transition with the greatest response is typically used to quantify the pesticides. Laboratories may be faced with the challenge of determining all pesticides, if any, present in a food sample. LC time of flight MS (TOF) screening methods provide extremely accurate molecular mass information (typically < 2 ppm) that can be searched against exact mass databases of thousands of compounds. Compounds found in this type of screen can be confirmed and quantified using LC MS-MS analysis on either QQQ or quadrupole time-of-flight (Q-TOF) instruments. Analyzing for Target Pesticides One of the most common activities in a food-testing laboratory is analysis for a set of target pesticides of specific interest. A finite set of pesticides might be used for production of a particular food crop, or a specific set might be especially toxic to children, necessitating the routine analysis for those compounds. GC MS (single quadrupole) utilizing deconvolution software has broad applicability for this type of analysis. The Automated Mass spectral Deconvolution and Identification System (AMDIS) was developed by the National Institute of Standards and Technology, and is incorporated into deconvolution reporting software (DRS) available from Agilent Technologies. Deconvolution separates unresolved peaks at any given retention time into sets of cleaned component spectra that can be matched to library spectra. The analyst runs the method that was used to generate the Pesticide and Endocrine Disruptor Database Library and locks the method, so that retention times will match those stored in the database. DRS then uses deconvolution and locked retention times to identify pesticides from the database. This is particularly useful in the analysis of extracts from foods such as spinach (Figure 1). In this case, one of the pesticides of interest is p,p - dichlorodiphenyldichloroethylene (p,p - DDE), a breakdown product of DDT. The total ion chromatogram (TIC) does not show a discernible peak at the p,p -DDE retention time (24.07 min). Deconvolution analysis found 768 different components over the whole TIC. After a pesticide library search, 12 of these had a match, including p,p -DDE. The black ions in the bottom window of Figure 1 represent the raw scan, before deconvolution. The white ions are a deconvoluted component that matches the known spectrum of p,p -DDE. Plotting the extracted ion chromatograms for this component (246, 248, 316, and 318) in the top window of Figure 1 also helped to confirm the presence of the p,p -DDE peak in the TIC. Figure 2 illustrates the extraction of the p,p - DDE spectrum from the raw spectrum, and its match to the library spectrum, resulting in a positive confirmation of the presence of this pesticide in the spinach sample. The use of powerful integrated deconvolution software such as DRS, in combination with specialized database libraries such as the Japanese Positive List Pesticide Library (1) or the more comprehensive Pesticide and Endocrine Disruptor Library (2) (Agilent Technologies, Palo Alto, California), can increase the efficiency of a food testing laboratory greatly. These GC MS methods can screen for hundreds of compounds and quantify a subset of them in as little as 2 min after the run using DRS. With traditional GC MS data processing, an analyst would need as much as 15 min per sample to review the data and confirm pesticide hits. LC MS methods are also very useful for analysis of target pesticides that are too polar or thermally labile for GC MS. LC QQQ is especially suited to very complex food matrices that can interfere with the analysis. Dealing with Dirty Sample Matrices Sample preparation is a significant challenge with food analysis and makes up a considerable portion of the total pesticide analysis time. Robust analysis methods that can tolerate very complex sample matrices without loss of sensitivity or accuracy are a must to reach throughput goals.

4 y= x R 2 = Figure 3: Calibration curve for diazinon in pepper using a six-point curve from 0.1 to 100 ng/ml with a linear fit and no origin treatment. Data were generated using a ZORBAX SB-C m LC column with a model G6410A QQQ system in the positive ESI mode (Agilent Technologies). Area 1 Area 2 Area 3 Area 4 Area 5 Mean RSD % Figure 4: Precision of GC-QQQ quantification of 500 fg of cyanophos spiked in garlic. Data were generated using an HP 5ms UI (Ultra Inert) 30 m 0.25 mm, m film thickness GC column and a model 7000A QQQ MS system (Agilent Technologies). The pesticide cyanophos was spiked into a garlic matrix at 0.5 ppb and replicate analyses performed at 500 fg on-column. The peaks are superimposed for the five runs, with the replicate peak areas, mean peak area, and RSD listed in the table to the left. Multiple reaction monitoring (MRM) with a QQQ system provides detection of trace amounts of target compounds in complex matrices, providing screening, confirmation, and quantification in one analysis (although a replicate analysis is usually required in most laboratories for confirmation). The precursor ion of interest generated by the source is selected and isolated in the first (Q1) quadrupole. A hexapole collision cell (Q2) fragments the molecule, and the product ions are selected in Q3. By choosing product ions from this transition that are characteristic for the pesticides of interest, chemical noise is separated from signal, providing very high sensitivity and selectivity, even in very dirty matrices. With careful selection of highly specific product ions for each pesticide, it is possible to create an MRM assay that can be used to identify and quantify hundreds of pesticides in a sample simultaneously. Efficient use of LC QQQ MRM methods requires optimization of a few parameters, including voltages on the fragmentor and the collision cell. Parameters must be optimized to generate the most intense product ions representative for each target compound, and automated method optimization software simplifies this task enormously. LC QQQ is an excellent platform for screening, confirmation, and quantification of pesticides. A screening method uses one transition per compound to look for a large number of pesticides on the target list. Confirmation of pesticides detected in the screening is done using two transitions per compound. For example, 301 pesticides have been rapidly identified in one analysis with most having a limit of detection (LOD) of 0.01 mg/kg (10 ppb), which is the MRL for baby food in the U.S. and the general default MRL for foods in

5 the European Union. These levels were reached in a single analysis using positive ion electrospray with 99 transitions per time segment, and a quantifying and qualifying ion for each compound (3). Two types of vegetables (pepper and tomato) were evaluated, and a linearity of response over three orders of magnitude was shown for quantification (r 2 > 0.99; Figure 3). GC QQQ also provides sensitive and Time (min) reproducible screening, confirmation, and quantification of pesticides that are not highly polar, particularly in very complex matrices such as garlic and ginger. Despite the high background of these matrices, pesticide compounds can be detected at levels as low as 0.5 ppb, with very high signal to noise and peak area precision as good as 6% RSD (Figure 4). For certain active pesticides such as acephate, the Using oven backflush it took an additional 33 min at 320 C to remove those boilers Run stopped at 42 min, backflushed at 290 C for 7 min Blank run after backflushed shows the column was clean Figure 5: Backflushing the GC column for higher signal to noise and shorter cycle times. Comparison of backflushing using capillary flow (red chromatogram) versus column bake-out (blue chromatogram) utilizing a model 7890A GC system and an HP-5ms GC column (Agilent Technologies). use of chemical protectants can reduce adsorption of analytes during the analysis, increasing sensitivity and improving peak shape. The result can be a marked improvement in accuracy of quantification in complex matrices (4). Backflushing the GC column, rather than using oven bakeout, also removes many high-boiling interfering substances in the sample (Figure 5) without altering retention times in subsequent runs or shortening the life of the column (5). Backflushing improves data quality while shortening cycle time and increasing column life. It also eliminates ghost peaks and reduces ion source contamination by keeping excess column bleed and heavy residues from being introduced into the mass spectrometer. One key to the successful rapid analysis of large numbers of pesticides is the dwell time of the QQQ mass spectrometer the time spent to measure the signal of any one precursor to product ion transition. The shorter the dwell time, the more transitions that can be analyzed per unit time, and the more pesticides that can be quantified per unit time. Figure 6: Pesticide screening for unknowns using LC QTOF. The top panel shows the full spectrum total ion chromatogram (TIC) using a model 1200 LC system, a ZORBAX Eclipse XDB, 5- m column, and a model 6510 LC MS QTOF system (Agilent Technologies). The Molecular Feature Extractor found 510 compounds in the TIC, 15 of which had hits from the EXACT MASS database search (bottom panel). The three highlighted compounds were further confirmed by MS MS analysis.

6 Analyzing for Unknown Pesticides Often, food-testing laboratories want to determine the entire spectrum of pesticides that might be present in a sample, rather than just a limited set of target compounds. While GC MS using deconvolution is an effective tool to screen for over 900 pesticides (2), LC TOF or LC QTOF can be used to screen for any number of LC MS-amenable compounds. TOF and QTOF are extremely accurate MS techniques, routinely achieving <2 ppm mass accuracy. Analysis software such as the Molecular Feature Extractor (MFE) from Agilent Technologies can find all ions representing real compounds in the sample. Noise and other extraneous ions are excluded. The resulting list of masses is then searched against a database of theoretical exact masses of compounds based on their molecular formulas and selected adducts, to identify pesticide compounds (Figure 6). This method has been used to screen food extracts for 600 pesticides and their degradates. The limit of detection (LOD) was determined for 100 of the 600 pesticides, and varied from <0.01 mg/kg (10 ppb) for 34% of the compounds, to <0.5 mg/kg (500 ppb) for 95% of the compounds (6). Using the MFE algorithm streamlines the data analysis for screening of hundreds of compounds at sensitive levels to minutes, compared with hours or even days using a manual approach. TOF can be used to identify an unlimited number of ionizable compounds, and the sensitivity is not impacted by the number of compounds screened. While LC TOF can provide screening and quantification of both unknown and targeted pesticides, LC QTOF or another MS-MS approach is required to confirm identification. Conclusion One need only look to recent events to appreciate the importance of pesticide testing in food.a series of poisonings in Japan in December of 2007 and January of 2008 that sickened 10 people was linked to methamidophos pesticide contamination of gyoza dumplings imported from China. The impact on Japanese food buying habits was immediate, as importation of Chinese vegetables fell 40% shortly after the source of the poisoning was determined. In another example, pet food vegetable proteins contaminated with melamine were imported into the United States from China in early 2007, resulting in the deaths of several cats and dogs. More recently, the same contaminant has appeared in baby formula in China. Food-testing laboratories require a broad range of complementary MS-based analytical technologies and methods to screen, confirm, and quantify the presence of as many as 1000 pesticide compounds. Screening methods can be used to detect as many contaminants as possible in a single run, and these methods don t necessarily have to quantify all the compounds. The screening method can be followed by target compound analysis that is very selective, sensitive, and quantitative. For LCamenable pesticides, this means running LC TOF or LC QTOF followed by LC QQQ. For GC-amenable pesticides, this means running GC MS with DRS for screening followed by GC QQQ for target compound analysis at low levels. The method used by any given laboratory is determined by its specific pesticide analysis goals, existing instrumentation, the budget for acquisition of new instrumentation, and required LODs. When choosing new instrumentation, laboratories doing pesticide testing also should consider whether a particular manufacturer has a complete solution that includes all of the above platforms, and whether those platforms share common, interchangeable ion sources, a unified software platform, and transferable collision voltages, thus providing a uniform workflow. The robustness of a manufacturer s systems is a key factor, including the tolerance of the ion source to contamination and its required frequency of cleaning and maintenance. Utility of the software and its integration with comprehensive pesticide spectra libraries and exact mass databases are also keys to success. Finally, the manufacturer s reputation for excellence in GC and LC MS instrumentation should be considered. References (1) P.L. Wylie, Screening for pesticides in food using the Japanese Positive List Pesticide Method, Agilent Application Note (2007). (2) P.L. Wylie, Screening for 926 Pesticides and Endocrine Disruptors by GC/MS with Deconvolution Reporting Software and a New Pesticide Library, Agilent Application Note (2006). (3) J.A. Zweigenbaum, Multiresidue analysis of 301 pesticides in food samples by LC/Triple Quaduprole mass spectrometry, Agilent Application Note (2008). (4) M. Anastassiades et al., J. Chromatogr., A 1015, (2003). (5) C.-K. Meng, Improving Productivity and Extending Column Life with Backflush, Agilent Application Note (2006). (6) J.A. Zweigenbaum, Automated screening of 600 pesticides in Food by LC/TOF MS using a molecular feature database search, Agilent Application Note (2006). Philip Wylie, Jerry Zweigenbaum, Melissa Churley, Chin- Kai Meng, and Cao Zhe are Senior Applications Chemists, Agilent Technologies, Santa Clara, California. Reprinted from Current Trends In Mass Spectrometry, a Spectroscopy Supplement, November 2008 Printed in U.S.A.

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract A selected target compound list of 195 various pesticides

More information

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High

More information

TurboMass GC/MS Software

TurboMass GC/MS Software TurboMass GC/MS Software GAS CHROMATOGRAPHY / MASS SPECTROMETRY P R O D U C T N O T E Today s demanding laboratory requires software that is easy to learn yet offers sophisticated instrument control, robust

More information

QUANLYNX APPLICATION MANAGER

QUANLYNX APPLICATION MANAGER QUANLYNX APPLICATION MANAGER OVERVIEW INTRODUCTION QuanLynx, an Application Manager included with Waters MassLynx Software, is designed for quantitative analysis. Simply acquiring mass spectrometry data

More information

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual AB SCIEX TripleTOF 5600 SYSTEM High-resolution quant and qual AB SCIEX TripleTOF 5600 SYSTEM AB SCIEX is a global leader in the development of life science analytical technologies that help answer complex

More information

Implement High Quality, Routine GC-MS/MS Multi-Residue Pesticides Analysis with Ease Paul Silcock & Jason Cole

Implement High Quality, Routine GC-MS/MS Multi-Residue Pesticides Analysis with Ease Paul Silcock & Jason Cole Implement High Quality, Routine GC-MS/MS Multi-Residue Pesticides Analysis with Ease Paul Silcock & Jason Cole 1 The world leader in serving science Some discussion points for today.. Trend towards GC

More information

Thermo Scientific TSQ Quantum Access MAX

Thermo Scientific TSQ Quantum Access MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific TSQ Quantum Access MAX UNSURPASSED PRICE TO PERFORMANCE Superior

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment QTRAP 5500

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment QTRAP 5500 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment QTRAP 5500 Powerful synergy results when the world s most sensitive triple

More information

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Application Note Author David L. Wong Agilent Technologies, Inc. Santa Clara, CA, USA Introduction Monoclonal

More information

Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1

Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1 Application Note Method Validation of the AutoMate-Q40 for Pesticide Residues in Grapes Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1 Abstract The QuEChERS (Quick-Easy-Cheap-Effective-Rugged-Safe)

More information

HIGHER IMPACT LC/MS DIAGNOSTICS AGILENT IN VITRO DIAGNOSTIC MEDICAL DEVICES FOR CLINICAL LABORATORIES

HIGHER IMPACT LC/MS DIAGNOSTICS AGILENT IN VITRO DIAGNOSTIC MEDICAL DEVICES FOR CLINICAL LABORATORIES HIGHER IMPACT LC/MS DIAGNOSTICS AGILENT IN VITRO DIAGNOSTIC MEDICAL DEVICES FOR CLINICAL LABORATORIES 2 AGILENT IN THE CLINIC THE PARTNER YOU TRUST Agilent s commitment to quality When a diagnosis is on

More information

Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs

Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs Technical Overview Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs Authors Stephen Madden, Crystal Cody, and Jungkap Park Agilent Technologies, Inc. Santa Clara, California,

More information

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC Analysis of Intact and Fragmented mabs and ADC Application Note Biotherapeutics and Biologics Author Suresh Babu C.V. Agilent

More information

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB

More information

Xevo G2-S QTof and TransOmics: A Multi-Omics System for the Differential LC/MS Analysis of Proteins, Metabolites, and Lipids

Xevo G2-S QTof and TransOmics: A Multi-Omics System for the Differential LC/MS Analysis of Proteins, Metabolites, and Lipids Xevo G2-S QTof and TransOmics: A Multi-Omics System for the Differential LC/MS Analysis of Proteins, Metabolites, and Lipids Ian Edwards, Jayne Kirk, and Joanne Williams Waters Corporation, Manchester,

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.

More information

The LC-MS/MS Workhorse. SCIEX 4500 Series Mass Spectrometers

The LC-MS/MS Workhorse. SCIEX 4500 Series Mass Spectrometers The LC-MS/MS Workhorse SCIEX 4500 Series Mass Spectrometers The SCIEX 4500 Series takes the legendary API 4000 platform and intelligently re-engineers it to set a new benchmark for reliable quantitation.

More information

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Yun Wang Alelyunas, Mark D. Wrona, Jim Murphy, Angela Doneanu, Gregory

More information

THE SOURCE OF NEW POSSIBILITIES

THE SOURCE OF NEW POSSIBILITIES The Agilent 5977B GC/MSD THE SOURCE OF NEW POSSIBILITIES The Measure of Confidence The Agilent 5977B GC/MSD INSTRUMENT RELIABILITY AND PERFORMANCE: THE KEY TO A GOOD DAY IN THE LAB Most Popular Instrument

More information

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI with the Biopharmaceutical Platform Solution with UNIFI Vera B. Ivleva, Ying Qing Yu, Scott Berger, and Weibin Chen Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S The ability to

More information

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides Application Note 25390206 Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides Keywords Model 5380 Organophosphorus Pesticide PFPD PFPDView Phosphorus

More information

Agilent 6400 Series. System. Concepts Guide. The Big Picture. Agilent Technologies

Agilent 6400 Series. System. Concepts Guide. The Big Picture. Agilent Technologies Agilent 6400 Series Triple Quad LC/MS System Concepts Guide The Big Picture Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means

More information

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Application Note 25370206 Low-level Detection of,, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Keywords Eclipse Fuel Oxygenates Methanol Model 4552 Purge and Trap Sample Concentrator

More information

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools Application Note Abstract The Commonwealth of Massachusetts Department of Environmental Protection

More information

Changing the Game in LC-MS/MS. New Tools for Unprecedented Performance in Residue Analysis. Dr. Thomas Glauner

Changing the Game in LC-MS/MS. New Tools for Unprecedented Performance in Residue Analysis. Dr. Thomas Glauner Changing the Game in LC-MS/MS New Tools for Unprecedented Performance in Residue Analysis Dr. Thomas Glauner EMEA LC/MS Food Segment Scientist Agilent Technologies Waldbronn The New 6490 Triple Quadrupole

More information

Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis

Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis Mark Wrona, 1 Paul Rainville, 1 Eric Langlois, 2 Nigel Ewing,

More information

The clearly better choice

The clearly better choice Agilent Accurate-Mass 6500 Series Q-TOF and 600 Series TOF LC/MS Systems The clearly better choice in TOF and Q-TOF LC/MS Our measure is your success. products applications soft ware services The unmatched

More information

Validation Report 19

Validation Report 19 EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark Validation Report 19 Determination of pesticide residues in oat, rye and wheat by GC-MS/MS and LC-MS/MS (QuEChERS

More information

Tony Mire-Sluis Vice President, Corporate, Product and Device Quality Amgen Inc

Tony Mire-Sluis Vice President, Corporate, Product and Device Quality Amgen Inc The Regulatory Implications of the ever increasing power of Mass Spectrometry and its role in the Analysis of Biotechnology Products Where do we draw the line? Tony Mire-Sluis Vice President, Corporate,

More information

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping. Thermo Scientific Peptide Mapping Workflows Upgrade Your Maps Fast, confident and more reliable peptide mapping. Smarter Navigation... Peptide mapping is a core analytic in biotherapeutic development.

More information

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany Quantification of trace-level herbicides in drinking water by online enrichment with the Agilent Infinity Series Online-SPE Solution and Triple Quadrupole MS Detection Application Note Environmental Author

More information

A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics

A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics Catalin Doneanu,, Ph.D. Biopharmaceutical Sciences, Waters September 16, 2009 Mass Spec 2009 2009 Waters Corporation Host Cell Proteins

More information

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS Zhiqi Hao, 1 Yi Zhang, 1 David Horn, 1 Seema Sharma, 1 Shiaw-Lin

More information

Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis

Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis SWATH Acquisition on the TripleTOF 5600+ System Samuel L. Bader, Robert L. Moritz Institute of Systems Biology,

More information

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications TECHNICAL BRIEF Gas Chromatography/ Mass Spectrometry The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications C 14 scans Introduction As gas chromatography

More information

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution Rapidly Identify Top Metabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Shaokun Pang and Ian Moore SCIEX,

More information

LC/MS/MS Solutions for Biomarker Discovery QSTAR. Elite Hybrid LC/MS/MS System. More performance, more reliability, more answers

LC/MS/MS Solutions for Biomarker Discovery QSTAR. Elite Hybrid LC/MS/MS System. More performance, more reliability, more answers LC/MS/MS Solutions for Biomarker Discovery QSTAR Elite Hybrid LC/MS/MS System More performance, more reliability, more answers More is better and the QSTAR Elite LC/MS/MS system has more to offer. More

More information

Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS

Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS TECHNICAL NOTE Pesticides in Tobacco Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS API 4000 LC/MS/MS System Overview This application note

More information

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use Liquid Chromatograph Liquid Chromatograph Mass Spectrometer CL Series for In Vitro Diagnostic Use Consistency. Performance. Reliability. Shimadzu s CL Series LC, LC-MS, and LC-MS/MS for in vitro diagnostic

More information

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module Application Note Environmental Authors Shane Snyder Department of

More information

The Five Key Elements of a Successful Metabolomics Study

The Five Key Elements of a Successful Metabolomics Study The Five Key Elements of a Successful Metabolomics Study Metabolomics: Completing the Biological Picture Metabolomics is offering new insights into systems biology, empowering biomarker discovery, and

More information

AppNote 5/2006 ABSTRACT

AppNote 5/2006 ABSTRACT AppNote 5/26 Stir Bar Sorptive Extraction Based on Restricted Access Material for the Direct Extraction of Drugs and Metabolites in Biological Fluids Wayne M. Mullett and Elizabeth Kwong Merck Frosst,

More information

High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science

High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers The world leader in serving science Goals Explore the capabilities of High Resolution Accurate Mass

More information

Fusion Analytical Method Validation

Fusion Analytical Method Validation Fusion QbD Software System Fusion Analytical Method Validation The Only Software That Has It All! 100% aligned with FDA/ICH Quality by Design (QbD) guidances! Can be used for LC and Non-LC methods (e.g.

More information

The Agilent 5975C Series GC/MSD Performance, productivity and confidence. Our measure is your success.

The Agilent 5975C Series GC/MSD Performance, productivity and confidence. Our measure is your success. The Agilent 5975C Series GC/MSD Performance, productivity and confidence. Our measure is your success. The Agilent 5975C Series GC/MSD Proven performance, superior productivity and maximum confidence in

More information

Proteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1

Proteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1 Routine Characterization of mabs and Other Proteins Patrick Boyce Biopharmaceutical Marketing Manager Europe and India 2011 Waters Corporation 1 Agenda Why? What scientific challenges? Technology Example

More information

Validation Report 16

Validation Report 16 EURL for Cereals and Feeding stuff National Food Institute Technical University of Denmark Validation Report 16 Determination of pesticide residues in wheat, barley and rice by GC-MS/MS and LC-MS/MS (QuEChERS

More information

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method 321.8 Application ICP-MS Author Jason A. Day, Anne Vonderheide, Joseph A. Caruso University of Cincinnati Cincinnati,

More information

ICP Expert software. Technical Overview. Introduction

ICP Expert software. Technical Overview. Introduction ICP Expert software Technical Overview Introduction The Agilent 5110 ICP-OES provides fast sample analysis, using less gas, without compromising on performance for tough samples. It has been designed for

More information

Application Note: Multi-Residue Pesticides in Fruit and Vegetables (9/16/08)

Application Note: Multi-Residue Pesticides in Fruit and Vegetables (9/16/08) Application Note: Multi-Residue Pesticides in Fruit and Vegetables (9/16/08) Product: DPX-Q (5 ml) INTRODUCTION The analysis of pesticides in fruit and vegetables is very important in the field of food

More information

The Use of LC-MS in Systematic Toxicology Analysis

The Use of LC-MS in Systematic Toxicology Analysis CAT Meeting Phoenix, AZ June 8, 7 Presentation outline Methods used in Systematic Toxicology Analysis The Use of LC-MS in Systematic Toxicology Analysis LC/MS approach ChromaLynx software Chromatogram

More information

Using accurate mass LC/MS/MS in forensic toxicology screening

Using accurate mass LC/MS/MS in forensic toxicology screening Using accurate mass LC/MS/MS in forensic toxicology screening Dr Simon Elliott Consultant Forensic Toxicologist (ROAR) Forensics Ltd Malvern Hills Science Park, Geraldine Rd, Malvern, Worcestershire, WR14

More information

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips Next generation immunoaffinity Robust quantitative platform Immunoaffinity sample preparation Thermo Scientific Mass Spectrometric Immunoassay

More information

EURL FOR PESTICIDE RESIDUES IN FRUITS AND VEGETABLES (EURL-FV) Activity Programme 2015

EURL FOR PESTICIDE RESIDUES IN FRUITS AND VEGETABLES (EURL-FV) Activity Programme 2015 EURL FOR PESTICIDE RESIDUES IN FRUITS AND VEGETABLES (EURL-FV) Activity Programme 2015 FUNCTIONS AND DUTIES The functions and duties of the European Commission Reference Laboratory are described in Article

More information

Electron Capture Detector for GC-2010 Plus

Electron Capture Detector for GC-2010 Plus Electron Capture Detector for GC-2010 Plus C180-E080 The World's Highest Level of Durability and Performance for a Capillary ECD Electron Capture Detector Novel Contact Free Technology was incorporated

More information

Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite. Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct.

Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite. Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct. Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct. 12, 211 Outline Orbitrap Mass Spectrometer in mab Characterization Intact

More information

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS Application Note BioPharma Authors Alex Zhu, James Martosella, and Phu T Duong Agilent Technologies, Inc Introduction Peptide

More information

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution Rapidly Identify Major Catabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Ian Moore and Jinal Patel

More information

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS Authors: Conor Smith, Doug Fryer United Science Corp. 15911 Furuby Rd. Center City, MN 55012 USA Abstract

More information

Metabolic Changes in Lung Tissue of Tuberculosis-Infected Mice Using GC/Q-TOF with Low Energy EI

Metabolic Changes in Lung Tissue of Tuberculosis-Infected Mice Using GC/Q-TOF with Low Energy EI Metabolic Changes in Lung Tissue of Tuberculosis-Infected Mice Using GC/Q-TOF with Low Energy EI Application Brief Authors Mª Fernanda Rey-Stolle 1, Vineel P. Reddy 2, Santiago Angulo 1, Adrie J.C. Steyn

More information

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A high-throughput method for detecting ultra-low levels (0.5 pg/ml) of a therapeutic peptide in human

More information

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation Cees Bruggink, Frans Schoutsen Thermo Fisher Scientific,

More information

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 15EX LC/MS System compact, rugged, The API 15EX LC/MS System is the most rugged and reliable single quadrupole

More information

ENHANCED MS WITH THE TURN OF A KEY

ENHANCED MS WITH THE TURN OF A KEY [ ionkey/ms ] ENHANCED MS WITH THE TURN OF A KEY EXPERIENCE THE FUTURE OF MS, EXPERIENCE ionkey/ms THE LC-MS YOU KNOW, LIKE YOU VE NEVER SEEN BEFORE ionkey/ms Systems integrate the UPLC separation into

More information

NIR Checkmaster Near-infrared spectroscopy On-line analysis of active ingredients during tablet production

NIR Checkmaster Near-infrared spectroscopy On-line analysis of active ingredients during tablet production NIR Checkmaster Near-infrared spectroscopy On-line analysis of active ingredients during tablet production Slash release times with NIR Innovative features Fully automatic assay of tablet weight, hardness,

More information

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry Pre-cleaned and certified vials for mass spectrometry The FIRST and ONLY pre-cleaned, low particle, low background chromatography vial When your instrumentation, sample handling and methodology is pushing

More information

rapiflex Innovation with Integrity Designed for Molecules that Matter. MALDI TOF/TOF

rapiflex Innovation with Integrity Designed for Molecules that Matter. MALDI TOF/TOF rapiflex Designed for Molecules that Matter. Innovation with Integrity MALDI TOF/TOF rapiflex TM The first MALDI-TOF/TOF that adapts to your needs. The rapiflex is the most advanced MALDI TOF/TOF system

More information

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES Jose M. Castro-Perez 1, Jane Kirby 1, Laurent Leclercq 2, Claude Delatour 2, Jean-Michel Mineur 2,

More information

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY AGILENT 7693A SERIES AUTOMATIC LIQUID SAMPLER A powerful combination of productivity and flexibility It

More information

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC UOP 63-13: Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC Fast Analysis in

More information

HYDROGEN AS A CARRIER GAS. Gas Chromatography new frontiers: can Hydrogen substitute Helium?

HYDROGEN AS A CARRIER GAS. Gas Chromatography new frontiers: can Hydrogen substitute Helium? HYDROGEN AS A CARRIER GAS Gas Chromatography new frontiers: can Hydrogen substitute Helium? Manufacturing UHP gas generators for lab analysis since 1979, CLAIND is seating in Lenno, on Como Lake, where

More information

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography Agilent 7693A Series Automatic Liquid Sampler Inject new performance into your gas chromotography Agilent 7693A Series Automatic Liquid Sampler A powerful combination of productivity and flexibility How

More information

Application Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona

Application Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona Analysis of an Antibody Drug Conjugate using MSIA D.A.R.T. S. Technology, an Integral Part of Ligand Binding Mass Spectrometric Immunoassay (LB-MSIA) Workflow Application Note Kwasi Antwi, Amanda Ribar,

More information

PHARMACEUTICAL TESTING

PHARMACEUTICAL TESTING WHITEHOUSE, NJ PHARMACEUTICAL TESTING Pharmaceutical Expertise for GMP & CMC Testing Our Pharmaceutical Expertise With more than 20 years of experience in a variety of industries, our Whitehouse, New Jersey

More information

Conjugated protein biotherapeutics

Conjugated protein biotherapeutics B i o P r o c e s s TECHNICAL Mass Spectrometric Conjugate Characterization Process Qualification of Recombinant Protein Hapten Conjugation Miao-Fang Lin, Margo Wilson, Michael V. Murray, and Greg W. Adams

More information

A Single Analyzer Solution for Refinery Flare Emissions Monitoring (40 CFR 60 Subpart Ja) Using Mass Spectrometry

A Single Analyzer Solution for Refinery Flare Emissions Monitoring (40 CFR 60 Subpart Ja) Using Mass Spectrometry A Single Analyzer Solution for Refinery Flare Emissions Monitoring (40 CFR 60 Subpart Ja) Using Mass Spectrometry Introduction New environmental regulations have recently prompted many oil refineries and

More information

EU Document on Analytical Quality Control and Method Validation Procedures for Pesticide Residues Analysis in Food and Feed An update

EU Document on Analytical Quality Control and Method Validation Procedures for Pesticide Residues Analysis in Food and Feed An update EU Document on Analytical Quality Control and Method Validation Procedures for Pesticide Residues Analysis in Food and Feed An update Antonio Valverde Pesticide Residues Research Group University of Almería,

More information

Clarus 600. Gas Chromatograph/Mass Spectrometers. your GC/MS results. better. faster.

Clarus 600. Gas Chromatograph/Mass Spectrometers. your GC/MS results. better. faster. Clarus 600 Gas Chromatograph/Mass Spectrometers your GC/MS results. better. faster. whatever you need it for the Clarus 600 GC/MS family delivers Clarus 600 Gas Chromatograph/Mass Spectrometer (GC/MS)

More information

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics Metabolomic Profiling in Drug Discovery: Understanding the Factors that Influence a Metabolomics Study and Strategies to Reduce Biochemical and Chemical Noise Mark Sanders 1, Serhiy Hnatyshyn 2, Don Robertson

More information

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY AGILENT 7693A SERIES AUTOMATIC LIQUID SAMPLER A powerful combination of productivity and flexibility Autosampler

More information

Genotoxicity is the property of a compound

Genotoxicity is the property of a compound Impurities Analysis in Pharmaceuticals: Genotoxicity is the property of a compound known to have irreversible effects on the structure and functionality of the DNA in cells and cause DNA loss, DNA replication

More information

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development Thermo Scientific Solutions for Intact-Protein Analysis Better, Faster Decisions for Biotherapeutic Development Quickly and Accurately Assess Product Quality and Safety Therapeutic proteins and monoclonal

More information

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany D-LC/MS Characterization of Charge Variants Using Ion Exchange and Reversed-Phase Chromatography Multiple Heart-Cutting D-LC Analysis of Innovator versus Biosimilar Monoclonal Antibodies Application Note

More information

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column Application Note Bio-Pharmaceutical Authors M. Sundaram Palaniswamy and Andrew Coffey Agilent Technologies,

More information

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF Application Note Authors Xianming Liu, Wei Zhang, Yi Du, Sheng Yin, Hong Que, and Weichang Zhou WuXi AppTec iopharmaceuticals

More information

International Journal of Pharma Research & Review, Feb 2014; 3(2):11-16 ISSN:

International Journal of Pharma Research & Review, Feb 2014; 3(2):11-16 ISSN: Research Article Determination of Methyl Methanesulfonate, Ethyl Methanesulfonate and Isopropyl Methanesulfonate Impurities in Lopinavir API by GC/MS/MS using Electron Ionization Veenaeesh P, *Manikumar

More information

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES THE AGILENT 7820A GC SYSTEM FORWARD-THINKING INNOVATION, EVERYDAY RELIABILITY AND VALUE Is your lab looking for better, more reproducible results

More information

MAX300-RTG PRODUCT NOTE. Real-Time Gas Analysis Process Automation and Control Product Quality Analysis. Industrial Process Mass Spectrometer

MAX300-RTG PRODUCT NOTE. Real-Time Gas Analysis Process Automation and Control Product Quality Analysis. Industrial Process Mass Spectrometer MAX300-RTG Industrial Process Mass Spectrometer PRODUCT NOTE Real-Time Gas Analysis Process Automation and Control Product Quality Analysis www.extrel.com Introducing the MAX300-RTG Optimize Your Manufacturing

More information

Capillary GC Troubleshooting: a Practical Approach

Capillary GC Troubleshooting: a Practical Approach Capillary GC Troubleshooting: a Practical Approach Katherine Stenerson Gas Separations R&D Supelco, Supelco Park, Bellefonte, PA 16823 1999 Sigma-Aldrich Co. Outline Basic Troubleshooting Strategy Preventing

More information

Využití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS

Využití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS Využití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS Michal Godula Ph.D. Thermo Fisher Scientific The world leader in serving science

More information

N- The rank of the specified protein relative to all other proteins in the list of detected proteins.

N- The rank of the specified protein relative to all other proteins in the list of detected proteins. PROTEIN SUMMARY file N- The rank of the specified protein relative to all other proteins in the list of detected proteins. Unused (ProtScore) - A measure of the protein confidence for a detected protein,

More information

Determination of Iodinated Contrast Media in Aqueous Samples by Direct-Injection LC-MS/MS

Determination of Iodinated Contrast Media in Aqueous Samples by Direct-Injection LC-MS/MS Determination of odinated Contrast Media in Aqueous Samples by Direct-njection LC-MS/MS Application ote Environmental Authors Gerd Vanhoenacker, Mieke Steenbeke, Frank David, and Pat Sandra Research nstitute

More information

CHROMATOGRAPHY/MASS SPECTROMETRY

CHROMATOGRAPHY/MASS SPECTROMETRY DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) WITH SELECTED ION MONITORING (SIM)* EPA Method 522 Part Number: EU52112M6

More information

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram

More information

Application Note. Authors. Abstract

Application Note. Authors. Abstract Automated, High Precision Tryptic Digestion and SISCAPA-MS Quantification of Human Plasma Proteins Using the Agilent Bravo Automated Liquid Handling Platform Application Note Authors Morteza Razavi, N.

More information

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations SCION GC Series The Gas Chromatographers Choice for Separations Innovation in Gas Chromatography Bruker s Scion Instrument s long tradition long of tradition innovation of innovation and product and reliability

More information

MasterView Software. Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY

MasterView Software. Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY MasterView Software Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY Mass spectrometers especially high-resolution, accurate mass instruments produce a significant amount

More information

Proteomics: A Challenge for Technology and Information Science. What is proteomics?

Proteomics: A Challenge for Technology and Information Science. What is proteomics? Proteomics: A Challenge for Technology and Information Science CBCB Seminar, November 21, 2005 Tim Griffin Dept. Biochemistry, Molecular Biology and Biophysics tgriffin@umn.edu What is proteomics? Proteomics

More information

Roche Molecular Biochemicals Technical Note No. LC 10/2000

Roche Molecular Biochemicals Technical Note No. LC 10/2000 Roche Molecular Biochemicals Technical Note No. LC 10/2000 LightCycler Overview of LightCycler Quantification Methods 1. General Introduction Introduction Content Definitions This Technical Note will introduce

More information

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy Advances in Bioscience and Biotechnology, 2010, 1, 367-371 doi:10.4236/abb.2010.15049 Published Online December 2010 (http://www.scirp.org/journal/abb/). Quantitative determination of residual 2-(2-chloroethoxy)

More information