A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein Binding

Size: px
Start display at page:

Download "A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein Binding"

Transcription

1 A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein Binding Keeley Murphy, 1 Patrick Bennett, 1 Francois Espourteille, 2 Maciej Bromirski 3 1 Thermo Fisher Scientific, San Jose,CA; 2 Thermo Fisher Scientific, Franklin, MA; 3 Thermo Fisher Scientific, Bremen, Germany

2 Overview Purpose: To evaluate high throughput liquid chromatography (LC) multiplexing and high-resolution, accurate-mass analysis using a single generic method and rapid sample cycle times for protein plasma binding (PPB) assay measurements. Methods: A panel of commercially available drug compounds was incubated in an in vitro serum protein binding assay and were analyzed using a Thermo Scientific TM Exactive TM Plus mass spectrometer and Thermo Scientific TM Transcend TM LC multiplexing system. Sample results from LC multiplexing analysis were compared to results collected with single channel analysis as well as to results previously collected using LCMS/MS analysis with a triple quadrupole mass spectrometer. Results: Results for the incubated compound sets in the PPB assay meet the LOQ requirement of 5 nm for 85% of the compounds analyzed and a linear dynamic range of 3 orders of magnitude. The coefficient of variance between sample replicates across multiple channels is demonstrated to be less than 15% for all compounds and 5% or less for most. The percentage of unbound compound was calculated for all drugs with results correlating well to results previously collected by LCMS/MS for both single channel and multi-channel analysis. An overall cycle time of 19 seconds per sample was achieved indicating that fast LC gradients and multi-channel analysis did not decrease assay performance. Introduction In-vitro screening assays commonly used in the adsorption, distribution, metabolism and excretion (ADME) development phase of drug discovery rely on analytical methods that are straightforward and robust while maintaining a high level of performance for data quality. Currently the majority of these assays are analyzed using triple quadrupole mass spectrometers with LC gradients that are less than 2 minutes in duration. While effective, this approach requires the development of MS/MS conditions and instrument methods that contain idle mass spectrometry time during the desalting and column re-equilibration phases of the LC. The implementation of high-resolution, full scan mass spectrometry coupled with high throughput LC multiplexing eliminates the need for MS/MS method development while more efficiently utilizing instrument analysis time. Methods Sample Preparation A panel of twenty-two commercially available drug compounds was incubated in an in vitro serum protein binding assay, at an incubation concentration of 10 µm. Samples were incubated for 6.5 hours in a dialysis block followed by protein precipitation. Protein precipitation was performed by first adding 150 ml of acetonitrile containing internal standard compound (Alprenolol) to a 96-well 340-mL V-bottomed storage plate followed by addition of 50 ml of each of the assay samples. Calibration curves were also generated for each compound. A working stock solution of 50 mm in DMSO was first made for each assayed compound. A five-point standard curve at concentrations of 5, 50, 500, 1000 and 2000 nm was prepared for each compound by serial dilution from the working stock solution into blank mixed matrix using an eight channel pipette 1. Three groups of assay samples were prepared for each compound analyzed: Buffer and Serum samples from the dialysis block for % Free calculations as well as a Time Zero sample. Liquid Chromatography Gradient elution was accomplished using water (A) + 0.1% Formic Acid (v/v) and Acetonitrile (B) + 0.1% Formic Acid (v/v). The gradient was held at 98% aqueous for 0.25 minutes, ramped to 98% B over 0.35 minutes, and held at 98% B for 0.2 minutes before returning to the starting conditions at 2% B for a 0.4 minute equilibration time. LC multiplexing parameters for the Transcend system were optimized to achieve a 19 second sample collection cycle time. The optimized LCMS method was used for all sample analysis. Assay samples were analyzed using a four channel LC multiplexing system with each individual sample injected onto each LC channel in replicate injections to measure single channel and inter-channel variability across all analytical columns. Chromatographic separation was performed using a C18, 2.1 x 30 mm, 3µm column with 5 µl injections made for each sample. All sample injections were completed using a Thermo Scientific Transcend LX-4 system with a dual injector arm with DLW (Dynamic Load and Wash) at a flow rate of 900 µl/min 2 A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein

3 N L: 1.15E 8 m /z= M S PPB_HTS_Sin gle_set2_357 N L: 2.16E 7 m /z= M S PPB_HTS_Sin gle_set2_357 Mass Spectrometry Samples were analyzed using an Exactive Plus mass spectrometer in Full Scan mode (m/z ) with a resolution setting of 35,000 (FWHM) at m/z 200 and a spectral speed of 7 Hz. Generic ion source conditions were used for all sample collection including vaporizer temperature 550 C, capillary temperature 350 C, sheath gas of 55 arbitrary units, and an auxiliary gas of 20 arbitrary units. The instrument was calibrated in positive ion mode before sample acquisition using Pierce LTQ Velos ESI Positive Ion Calibration Solution. Data Analysis Data was acquired using Thermo Scientific Xcalibur 2.2 and Exactive Tune 2.1 software. Chromatographic data review and calibration curve generation was performed and reported using QuickCalc TM software (powered by Gubbs Inc., GMSU Gubbs Mass Spec Utilities, Atlanta, GA). Peak area measurements in the buffer chamber of the dialysis plate were compared to the peak area measurement in the serum chamber of the dialysis plate to calculate the percent of unbound compound (% Free) at assay equilibrium 1. The % Free values for each compound replicate were compared across LC channels and also compared to values previously obtained using a triple quadrupole mass spectrometer. The coefficient of variation of the area ratio for the Buffer, Serum, and Time Zero samples we also calculated and compared across LC channels to measure variability across LC columns. Results Data File Review The collection of multiple data points stored in a single data file allows for rapid sample injections and short sample to sample cycle times, but also requires software tools that simplify data review and automatically correlate individual chromatographic peaks to their respective injections (Figure 1). QuickCalc TM software was used to automatically correlate all sample injections contained in each data file with their respective chromatographic peaks areas as well as sample types and texts contained in the original sequence list. Additionally, identification of the LC channel to which each sample was injected was included with the sample information and was used to sort and calculate signal response and variation across each channel (Figure 2). FIGURE 1. Data file review of 64 sample injections in a single data file using Qual Browser. Analyte Desipramine, Internal Standard Alprenolol. Spectra display for RT at min, T 0 replicate number 3. Desipramine _Buffer-1 RT: Time (min) Relative Abundance PPB_HTS_Single_Set2_357 #4970 RT: AV: 1 NL: 1.60E7 T: FTMS + c ESI Full ms [ ] Buffer Rep 1 Rep 2 Rep 3 Acebutolol 5.1% 4.1% 2.2% Buspirone 1.0% 1.8% 3.2% Carbamazepine 4.6% 8.4% Chlorpheniramine 5.0% 2.8% 6.2% Clomipramin 6.9% 3.2% 2.9% Clozapine 3.0% 0.9% 3.4% Desipramine 2.6% 4.2% 2.9% Diltiazem 1.2% 1.9% 2.2% Erythromycin 3.7% 2.6% 3.6% Fluphenazine 4.4% 6.8% Haloperidol 8.3% 5.9% 8.1% Imipramine 4.7% 1.3% 3.2% Metoprolol 3.8% 5.2% 4.5% Nefazadone 3.4% % Phentolamine 12.2% 15.4% 9.4% Propranolol 3.4% 2.9% 1.4% Retonavir 14.2% 13.4% n/a Thioridazine 6.3% 2.2% 5.5% Ticlopidine 4.9% 10.6% 7.1% Timolol 3.5% 8.2% Verapamil 5.1% 4.5% Relative Abundance m/z FIGURE 3. %CV for sample replicat representative compounds. FIGURE 2. Data file review of 64 sample injections in a single data file using QuickCalc. Analyte Desipramine, Internal Standard Alprenolol. Display for RT at min, T 0 replicate number 3 Clomipramine %CV Imipramine %CV Ery B T B Thermo Scientifi c Poster Note PN ASMS13_Th228_KMurphy_E 07/13S 3

4 Inter-Channel Variability The overall sample acquisition rate from injection to injection can be greatly improved when variability between LC columns is minimized and a given analyte can be injected across all four LC channels without a major variation in signal response due to column variability. To demonstrate the overall variability between the four LC channels used in the experiment, each sample was sequentially injected directly onto each of the four LC channels and the coefficient of variation calculated using the resulting area ratio values.(table 1.) The calculated %CV for the area ratio response was less than 15% for all samples, the %CV was no greater than 10% for 92% of the samples analyzed and the %CV was no greater than 5% for 73% of the samples analyzed. The variation in response between LC channels is illustrated for six representative compounds below.(figure 3.) Although twenty-two compounds were incubated in PPB assay, one compound did not provide sufficient signal response to accurately measured and has been excluded from all tables and figures. TABLE 1. %CV for sample replicate across each LC channel. Each value in the table represents the variation in response for a single sample injected to each of the four LC channels. All samples are 15% CV or less, 92% of samples 10% CV or less, and 73% of samples 5% CV or less. Rep 1 Rep 2 Rep 3 Rep 1 Rep 2 Rep 3 Rep 1 Rep 2 Rep 3 Acebutolol 5.1% 4.1% 2.2% 0.7% 3.4% 5.1% 2.5% 5.3% Buspirone 1.0% 1.8% 3.2% 1.0% 1.3% 2.3% 6.5% 3.2% 3.6% Carbamazepine 4.6% 8.4% 10.9% 6.4% 7.8% 5.1% 9.5% 5.1% Chlorpheniramine 5.0% 2.8% 6.2% 3.2% n/a 2.6% 3.1% 9.4% 4.5% Clomipramin 6.9% 3.2% 2.9% 4.6% 3.3% 4.9% 5.4% 4.2% 2.3% Clozapine 3.0% 0.9% 3.4% 2.9% 1.8% 2.3% 2.9% 3.7% 2.5% Desipramine 2.6% 4.2% 2.9% 6.2% 2.4% 2.9% 2.6% 3.3% 1.8% Diltiazem 1.2% 1.9% 2.2% 0.9% 2.7% 3.5% 3.6% 2.8% 3.0% Erythromycin 3.7% 2.6% 3.6% 3.6% 2.7% 1.3% 2.9% 1.7% 1.5% Fluphenazine 4.4% 6.8% 2.9% 6.9% 3.0% 6.4% 2.4% 3.4% Haloperidol 8.3% 5.9% 8.1% 4.3% 3.8% 3.4% 3.8% 4.7% 6.4% Imipramine 4.7% 1.3% 3.2% 3.5% 4.1% 1.7% 3.7% 3.0% 1.5% Metoprolol 3.8% 5.2% 4.5% 8.8% 0.9% 6.1% 2.9% 2.8% 4.7% Nefazadone 3.4% % 6.1% 6.7% 9.1% 4.4% 4.9% 5.1% Phentolamine 12.2% 15.4% 9.4% 1.2% 1.1% 2.7% 1.2% n/a Propranolol 3.4% 2.9% 1.4% 9.3% n/a 1.7% 0.6% 2.3% Retonavir 14.2% 13.4% n/a 15.2% 7.7% 12.8% 1 7.6% 13.0% Thioridazine 6.3% 2.2% 5.5% 4.1% 6.6% 4.9% 7.6% 8.6% Ticlopidine 4.9% 10.6% 7.1% 3.3% 4.8% 3.2% 1.4% 2.1% 2.8% Timolol 3.5% 8.2% 6.6% 5.3% 7.8% 3.7% 3.2% 3.8% Verapamil 5.1% 4.5% 3.0% 1.3% 4.5% 2.6% 2.1% 4.7% % Free Ave A FIGURE 3. %CV for sample replicates injected across each LC channel for six representative compounds. Clomipramine %CV Erythromycin %CV Haloperidol %CV Imipramine %CV Timolol %CV Verapamil %CV Buffer Serum T0 4 A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein

5 % Free Calculation Percent free or the unbound amount of compound in the protein binding assay was calculated by determining the ratio of signal response between the buffer and serum samples 1. The % Free was calculated using both signal response from a single channel as well as signal response from across all channels to determine the viability of the multi-channel analysis approach. The % Free for the single channel analysis was calculated using the buffer/serum signal response from the same LC channel for all injection replicates and the average %Free was reported for each compound. The % Free for the multi-channel analysis was calculated by mixing the signal response for buffer and serum within a sample replicate between each of the four LC channels. For example, buffer from channel 1 was compared to serum from channel 2 etc. The calculated %Free values for replicates were averaged and reported. The average % Free values for each compound were plotted in a bar chart to illustrate differences in the % Free values between single channel analysis, multi-channel analysis and results previously collected by LC-MS/MS (Figure 4). The average %Free values for both the single channel and multi-channel correlate well to each other as well as to the LC- MS/MS. Two compounds in the multi-channel analysis were observed with a difference of more than 25% from the LC-MS/MS analysis, Nefazadone and Clomipramine. FIGURE 5. Calibration curve of concentration point analyzed n channels. Y = ( * X) + ( E- 04) Area Ratio R^2 = RSD, RSD, RSD, FIGURE 4. Average % Free for individual compounds across each analysis type. Nineteen of twenty-one compounds analyzed using high resolution mass spectrometry demonstrate a % difference of less than 25% when compare to results collected using LCMS/MS % Free Average % Free Variation AVG %Free Single Channel AVG %Free Multi Channel AVG %Free MS/MS Each compound analyzed in the PPB assay was evaluated in a concentration curve to evaluate overall sensitivity and linear dynamic range. All compounds were serially diluted using PPB matrix blank solution with concentrations ranging from 5 nm to 2000 nm. Each concentration point was injected across each LC channel to determine inter-channel variability. The calibration curves for all compounds were generated using a linear regression and 1/x 2 weighting. Individual calibration points exceeding a percent difference of more than 20% of the regression line fit were excluded from all calculations. Three compounds did not provide adequate signal response for accurate quantitation of the 5 nm concentration point. Although three compounds could not be measured at the 5 nm concentration point only one compound did not provide enough signal response to be measure in the binding assay. The majority of the compounds provided adequate sensitivity for all concentration points in the calibration curve, with all compounds providing a linear response across the concentration range. Additionally, all compounds and their replicate injections at each concentration point across each LC channel demonstrated RSD values of less than 15%. An example calibration curve for the multi-channel analysis is displayed for Desipramine (Figure 5). Thermo Scientifi c Poster Note PN ASMS13_Th228_KMurphy_E 07/13S 5

6 FIGURE 5. Calibration curve of Desipramine from 5nm to 2000 nm. Each concentration point analyzed n=4 with replicates injected across all four LC channels. Y = ( * X) + ( E- 04) Area Ratio R^2 = Desipramine RSD, RSD, RSD, RSD, RSD, Concentration (ng/ml) Conclusion 85% of the compounds analyzed met the assay calibration curve LOQ of 5 nm and 95% provided adequate signal for assay reporting. 92% of the sample replicates injected across multi-channel system have a %CV of 10% or less. Negligible variation was observed for % Free values in the multi-channel analysis. Sample cycle time of 19 seconds per sample. High resolution full scan mass spectrometry coupled LC multiplexing using a single generic method provides suitable alternative to traditional LCMS/MS analysis References 1. Zhang J, Shou WZ, Vath M, Kieltyka K, Maloney J, Elvebak L, Stewart J, Herbst J, Weller HN., Rapid Commun Mass Spectrom Dec 30;24(24):3593 Acknowledgements We would like to thank Dr. Jun Zhang and Jennifer Maloney of Bristol-Myers Squibb, Wallingford, CT for sample preparation and assistance with data processing. 6 A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein

7 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. GMSU Gubbs Mass Spec Utilities is a trademark of Gubbs Inc. Microsoft and Excel are registered trademarks of Microsoft Corporation. All other trademarks are the property of Thermo Fisher Scientific, Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, San Jose, CA USA is ISO 9001:2008 Certified. Africa-Other Australia Austria Belgium Canada China Denmark Europe-Other Finland/Norway/Sweden France Germany India Italy Japan Latin America Middle East Netherlands New Zealand Russia/CIS South Africa Spain Switzerland UK USA ASMS13_Th228_KMurphy_E 07/13S

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition Using HRAM Survey Analysis Combined with Rapid MS Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition Tim Stratton, 1 Caroline Ding, 1 Christoph Henrich, Hans Grensemann,

More information

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass Application Note: 529 Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass Xiang He, Marta Kozak; Thermo Fisher Scientific, San Jose,

More information

DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS

DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS Darcy Shave and Pete Alden Waters Corporation, Milford, MA, U.S.A. INTRODUCTION Plasma protein binding (PPB) can significantly affect the therapeutic action

More information

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development Thermo Scientific Solutions for Intact-Protein Analysis Better, Faster Decisions for Biotherapeutic Development Quickly and Accurately Assess Product Quality and Safety Therapeutic proteins and monoclonal

More information

Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins

Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins Esther Lewis, 1 Kevin Meyer, 2 Zhiqi Hao, 1 Nick Herold,

More information

Thermo Scientific TSQ Quantiva Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative performance. with unprecedented ease

Thermo Scientific TSQ Quantiva Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative performance. with unprecedented ease Thermo Scientific TSQ Quantiva Triple-Stage Quadrupole Mass Spectrometer Extreme quantitative performance with unprecedented ease Unsurpassed quantitation Continual changes in research and regulatory requirements

More information

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics Metabolomic Profiling in Drug Discovery: Understanding the Factors that Influence a Metabolomics Study and Strategies to Reduce Biochemical and Chemical Noise Mark Sanders 1, Serhiy Hnatyshyn 2, Don Robertson

More information

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Carl Fisher, Monika Verma, and Pranathi Perati; Thermo Fisher Scientific,

More information

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance Thermo Scientific EASY-nLC 1200 System Leading in simplicity and performance Peak performance made EASY Effortless ultra high performance for everybody A straightforward LC-MS solution Optimized and integrated

More information

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation Cees Bruggink, Frans Schoutsen Thermo Fisher Scientific,

More information

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS Zhiqi Hao, 1 Yi Zhang, 1 David Horn, 1 Seema Sharma, 1 Shiaw-Lin

More information

Thermo Scientific TSQ Quantum Access MAX

Thermo Scientific TSQ Quantum Access MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific TSQ Quantum Access MAX UNSURPASSED PRICE TO PERFORMANCE Superior

More information

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma Yun Wang Alelyunas, Mark D. Wrona, Jim Murphy, Angela Doneanu, Gregory

More information

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics, USA Application Note

More information

Analysis of Targeted and Non-Targeted Identified Contaminants in Storm Water Retention Ponds Using LC-HRMS With Online Solid Phase Extraction

Analysis of Targeted and Non-Targeted Identified Contaminants in Storm Water Retention Ponds Using LC-HRMS With Online Solid Phase Extraction Analysis of Targeted and Non-Targeted Identified Contaminants in Storm Water Retention Ponds Using LC-HRMS With Online Solid Phase Extraction Gordon Getzinger, 1 P. Lee Ferguson, 1 Jonathan Beck, 2 Charles

More information

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES Marine Beauvir, Application Specialist, Thermo Fisher Scientific, Cambridge, UK Application Note 43161 Key Words High throughput,

More information

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution Rapidly Identify Top Metabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Shaokun Pang and Ian Moore SCIEX,

More information

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data Thermo Scientific icap 7000 Series ICP-OES Low cost ICP-OES analysis high quality data Thermo Scientific icap 7000 Series ICP-OES The NEW Thermo Scientific icap 7000 Series ICP-OES provides low cost multi-element

More information

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES Jose M. Castro-Perez 1, Jane Kirby 1, Laurent Leclercq 2, Claude Delatour 2, Jean-Michel Mineur 2,

More information

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis Shanhua Lin, Julia Baek, and Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis

Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis Mark Wrona, 1 Paul Rainville, 1 Eric Langlois, 2 Nigel Ewing,

More information

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips Next generation immunoaffinity Robust quantitative platform Immunoaffinity sample preparation Thermo Scientific Mass Spectrometric Immunoassay

More information

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines APPLICATION NOTE 64968 Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines Authors Ali Haghani, 1 Andy Eaton, 1 Neloni Wijeratne, Claudia Martins 1 Eurofins Eaton Analytical,

More information

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo James Hannan, Applications Chemist, Thermo Fisher Scientific, Cambridge, UK Application Note 43171 Key Words Environmental, EU, waste,

More information

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal ntibody Charge Variants Ken Cook, Frank Steiner, Mauro De Pra Thermo Fisher Scientific, Hemel Hempstead,

More information

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions Screening and Confirmation Chemical Analysis Solution Unit November, 2009 Page 1 Agilent s New 6540 Ultra High

More information

Lebedev A, 1 Damoc E, 2 Makarov A, 2 Samguina T 1 1. Moscow State University, Moscow, Russia; 2 ThermoFisher Scientific, Bremen, Germany

Lebedev A, 1 Damoc E, 2 Makarov A, 2 Samguina T 1 1. Moscow State University, Moscow, Russia; 2 ThermoFisher Scientific, Bremen, Germany Winning The Last Battle Against Edman Degradation: Reliable Leucine/Iso-leucine Differentiation In Peptide Sequencing Using an Orbitrap Fusion Mass Spectrometer Lebedev A, 1 Damoc E, 2 Makarov A, 2 Samguina

More information

Liver Mitochondria Proteomics Employing High-Resolution MS Technology

Liver Mitochondria Proteomics Employing High-Resolution MS Technology Liver Mitochondria Proteomics Employing High-Resolution MS Technology Jenny Ho, 1 Loïc Dayon, 2 John Corthésy, 2 Umberto De Marchi, 2 Antonio Núñez, 2 Andreas Wiederkehr, 2 Rosa Viner, 3 Michael Blank,

More information

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube Application Note Environmental Authors Don Noot, Ralph Hindle Vogon Laboratory Services Cochrane, AB

More information

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution [ product solution ] Waters µ Elution PlateS Elution volume as low as 25 μl No evaporation and reconstitution Ideal for small sample volumes Up to a 15x increase in concentration Patented Innovation Now

More information

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC Analysis of Intact and Fragmented mabs and ADC Application Note Biotherapeutics and Biologics Author Suresh Babu C.V. Agilent

More information

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution Rapidly Identify Major Catabolites with the SCIEX X500R QTOF System and MetabolitePilot TM 2.0 Software Ian Moore and Jinal Patel

More information

Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing Christopher Crafts, Bruce Bailey, John Waraska, Ian Acworth Thermo Fisher Scientific, Chelmsford, MA, USA Overview Purpose: The

More information

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge G. Scott*, H. Gaus #, B. Rivera*, and M. McGinley* *Phenomenex,

More information

High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science

High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers The world leader in serving science Goals Explore the capabilities of High Resolution Accurate Mass

More information

LTQ Orbitrap XL Hybrid FT Mass Spectrometer Unrivaled Performance and Flexibility

LTQ Orbitrap XL Hybrid FT Mass Spectrometer Unrivaled Performance and Flexibility m a s s s p e c t r o m e t r y LTQ Orbitrap XL Hybrid FT Mass Spectrometer Unrivaled Performance and Flexibility Part of Thermo Fisher Scientific LTQ Orbitrap XL OFFERING OUTSTANDING MASS ACCURACY, RESOLVING

More information

Application Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona

Application Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona Analysis of an Antibody Drug Conjugate using MSIA D.A.R.T. S. Technology, an Integral Part of Ligand Binding Mass Spectrometric Immunoassay (LB-MSIA) Workflow Application Note Kwasi Antwi, Amanda Ribar,

More information

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance Thermo Scientific icap 7000 Plus Series ICP-OES Gain more power experience more performance Thermo Scientific icap 7000 Plus Series ICP-OES Powerful, easy-to-use, solution for multi-element analysis Maximize

More information

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 15EX LC/MS System compact, rugged, The API 15EX LC/MS System is the most rugged and reliable single quadrupole

More information

Application Note. Authors. Abstract

Application Note. Authors. Abstract Automated, High Precision Tryptic Digestion and SISCAPA-MS Quantification of Human Plasma Proteins Using the Agilent Bravo Automated Liquid Handling Platform Application Note Authors Morteza Razavi, N.

More information

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo Analysis of Trace s in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo Christos Spanos, Applications Chemist, Thermo Fisher Scientific, Cambridge, UK Application Note 43185 Key Words

More information

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION Enhanc em ent o f t h e U P L C Am ino Ac i d Ana lysis So lu t io n w it h F l e x ibl e Detector Options Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION The measurement of amino acids

More information

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles Application Note Biotherapeutics and Biosimilars Authors James Martosella, Alex

More information

Bio-Rad Laboratories BIOPLEX 2200 SYSTEM. BioPlex 2200 ANA Screen with MDSS. The first and only fully-automated, multiplexed ANA Screen

Bio-Rad Laboratories BIOPLEX 2200 SYSTEM. BioPlex 2200 ANA Screen with MDSS. The first and only fully-automated, multiplexed ANA Screen Bio-Rad Laboratories BIOPLEX 2200 SYSTEM BioPlex 2200 ANA Screen with MDSS The first and only fully-automated, multiplexed ANA Screen Bio-Rad Laboratories BIOPLEX 2200 SYSTEM Like no other The BioPlex

More information

ENHANCED MS WITH THE TURN OF A KEY

ENHANCED MS WITH THE TURN OF A KEY [ ionkey/ms ] ENHANCED MS WITH THE TURN OF A KEY EXPERIENCE THE FUTURE OF MS, EXPERIENCE ionkey/ms THE LC-MS YOU KNOW, LIKE YOU VE NEVER SEEN BEFORE ionkey/ms Systems integrate the UPLC separation into

More information

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Samples Michael S.Young, Jeremy C. Shia, Kim vantran, Kevin M. Jenkins and Masayo Yabo Waters Corporation 34 Maple Street,

More information

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use Liquid Chromatograph Liquid Chromatograph Mass Spectrometer CL Series for In Vitro Diagnostic Use Consistency. Performance. Reliability. Shimadzu s CL Series LC, LC-MS, and LC-MS/MS for in vitro diagnostic

More information

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS Puzzled About LCMS? Sensitivity in Mass Spec Analysis Sample Prep Adapting LC-UV Optimizing and Maintaining Your Mass Spec To LCMS Paul Altiero Alex Ucci Applications Phone Support Chemistry and Supplies

More information

EDS Phase Mapping of a Contact Metamorphosed Calc-Silicate Rock

EDS Phase Mapping of a Contact Metamorphosed Calc-Silicate Rock EDS Phase Mapping of a Contact Metamorphosed Calc-Silicate Rock Application Note 52452 Key Words COMPASS, Energy Dispersive Spectroscopy (EDS), FESEM, Mineral Phases, Minerals, Multivariate Statistical

More information

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Application Note Author David L. Wong Agilent Technologies, Inc. Santa Clara, CA, USA Introduction Monoclonal

More information

MassPREP On-Line Desalting Cartridge

MassPREP On-Line Desalting Cartridge CONTENTS I. INTRODUCTION II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION

More information

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI with the Biopharmaceutical Platform Solution with UNIFI Vera B. Ivleva, Ying Qing Yu, Scott Berger, and Weibin Chen Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S The ability to

More information

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual AB SCIEX TripleTOF 5600 SYSTEM High-resolution quant and qual AB SCIEX TripleTOF 5600 SYSTEM AB SCIEX is a global leader in the development of life science analytical technologies that help answer complex

More information

Pharmaceutical LC/MS Solutions from Agilent Technologies

Pharmaceutical LC/MS Solutions from Agilent Technologies Pharmaceutical LC/MS Solutions from Agilent Technologies Application Compendium Overview LC/MS plays a key role in the drug discovery and drug development process. Since the introduction of electrospray

More information

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping. Thermo Scientific Peptide Mapping Workflows Upgrade Your Maps Fast, confident and more reliable peptide mapping. Smarter Navigation... Peptide mapping is a core analytic in biotherapeutic development.

More information

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column mab Titer Analysis with the Agilent Bio-Monolith Protein A Column Application Note Biopharmaceuticals and Biosimilars Authors Emmie Dumont, Isabel Vandenheede, Pat Sandra, and Koen Sandra Research Institute

More information

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System A high-throughput method for detecting ultra-low levels (0.5 pg/ml) of a therapeutic peptide in human

More information

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest and Lei Xiong SCIEX, Redwood City,

More information

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover columns ProSwift Reversed-Phase Monolith Columns for Protein Analysis ProSwift reversed-phase columns use a unique monolith technology for fast, high-resolution HPLC and LC/MS separations of proteins.

More information

Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit

Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit Dr. Jerry Zweigenbaum Senior Applications Scientist Food Safety Page 1 Agilent 1260 LC SampliQ Extraction and Dispersive

More information

Application Note. Abstract. Authors. Introduction. Environmental

Application Note. Abstract. Authors. Introduction. Environmental EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma

More information

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module Application Note Environmental Authors Shane Snyder Department of

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.

More information

NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778)

NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778) NOTICE OF INTENDED DECISION TO ENTER INTO A SINGLE SOURCE CONTRACT (PUR 7778) This Notice of Intended Decision to Enter into a Single Source Contract is posted in accordance with section 287.057(3), Florida

More information

High-Throughput Peptide Mapping with the Vanquish UHPLC System and the Q Exactive HF Mass Spectrometer

High-Throughput Peptide Mapping with the Vanquish UHPLC System and the Q Exactive HF Mass Spectrometer High-Throughput Peptide Mapping with the Vanquish UHPLC System and the Q Exactive HF Mass Spectrometer Martin Samonig 1, Kai Scheffler 2, Remco Swart 1, and Jonathan Josephs 3 1 Thermo Fisher Scientific,

More information

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications Chromatography Column Performance and Data Analysis Success Guide Hints and Tips for Better Purifications This Chromatography Success Guide provides practical advice on preparative chromatography and protein

More information

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo Grégory Lecornet, Applications Specialist, Thermo Fisher Scientific Application Note 43149 Key Words

More information

Thermo Scientific Qtegra Intelligent Scientific Data Solution. Delivering quality. Driving productivity

Thermo Scientific Qtegra Intelligent Scientific Data Solution. Delivering quality. Driving productivity Thermo Scientific Qtegra Intelligent Scientific Data Solution Delivering quality Driving productivity There can be economy only where there is efficiency 2 The Thermo Scientific Qtegra Intelligent Scientific

More information

HepG2 Labeling and Cell Growth Protocol for IROA Phenotypic Metabolic Profiling

HepG2 Labeling and Cell Growth Protocol for IROA Phenotypic Metabolic Profiling HepG2 Labeling and Cell Growth Protocol for IROA Phenotypic Metabolic Profiling Candice Z. Ulmer 1, Felice A. de Jong 2, Chris Beecher 1, Timothy J. Garrett 3, Richard A. Yost 1,3 1 Department of Chemistry,

More information

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run Application Note Author Jessica Westland Agilent Technologies, Inc. Abstract A selected target compound list of 195 various pesticides

More information

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes On-Line SPE CARTRIDGES for Rapid Cleanup and Extraction of Analytes User s Guide What is the strata-x on-line SPE cartridge? The strata-x on-line cartridge combines the revolutionary benefits of the patent

More information

ab Porcine VEGF-A ELISA Kit

ab Porcine VEGF-A ELISA Kit Version 1 Last updated 27 January 2017 ab218298 Porcine VEGF-A ELISA Kit For the quantitative measurement of porcine VEGF-A in serum, plasma and cell culture supernatants. This product is for research

More information

High-Throughput, Simplified Protein Precipitation: A Novel 96-Well Format for Precipitation and Filtration EAS 2003

High-Throughput, Simplified Protein Precipitation: A Novel 96-Well Format for Precipitation and Filtration EAS 2003 Kimvan Tran 1, Diane M. Diehl 1, Tim Schultz 2, Wilf Wixwat 2, Brian P. Murphy 1, Jeff Kane 2, Emily Berlin 2, Kevin Seeley 2, Ziling Lu 1 1 Waters Corporation, Milford, MA 2 Pall Life Sciences, Ann Arbor,

More information

Thermo Scientific icap 6000 Series ICP Emission Spectrometer

Thermo Scientific icap 6000 Series ICP Emission Spectrometer e l e m e n t a l a n a l y s i s Thermo Scientific icap 6000 Series ICP Emission Spectrometer Now everyone can experience dramatically different ICP... Environmental Metallurgical Petrochemical Food Safety

More information

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Disulfide Linkage Analysis of IgG1 using an Agilent 126 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column Application Note Biotherapeutics & Biosimilars Author M. Sundaram

More information

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for

More information

ProteoMiner Protein Enrichment Technology

ProteoMiner Protein Enrichment Technology Sample Preparation ProteoMiner Protein Enrichment Technology Digging Deeper in the Proteome Detect More Proteins With ProteoMiner Technology Than With Immunodepletion ProteoMiner Technology vs. an Agilent

More information

Instructions for Capillary Electrophoresis Peptide Analysis Kit

Instructions for Capillary Electrophoresis Peptide Analysis Kit Instructions for Capillary Electrophoresis Peptide Analysis Kit Catalog Number 148-4110 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Table of

More information

SEC-MS with Volatile Buffers for Characterization of Biopharmaceuticals

SEC-MS with Volatile Buffers for Characterization of Biopharmaceuticals SE-MS with Volatile uffers for haracterization of iopharmaceuticals Martin Samonig, Remco Swart Thermo Fisher Scientific, Germering, Germany pplication Note 1133 Key Words Monoclonal ntibodies, MbPac SE-1,

More information

ANALYTICAL STANDARDS AND REAGENTS

ANALYTICAL STANDARDS AND REAGENTS ANALYTICAL STANDARDS AND REAGENTS productivity future growth globalization compliance Waters, the industry leader in liquid chromatography and mass spectrometry, understands that standards and reagents

More information

ab83355 ATP Assay Kit (Colorimetric/ Fluorometric)

ab83355 ATP Assay Kit (Colorimetric/ Fluorometric) Version 18 Last updated 2 January 2018 ab83355 ATP Assay Kit (Colorimetric/ Fluorometric) For the rapid, sensitive and accurate measurement of ATP in a variety of samples. This product is for research

More information

Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction Format

Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction Format Case Study: PPS367 Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Page 1 Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction

More information

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions An Automated System for At-Line cess Analysis of Biopharmaceutical Fermentation Reactions Frank Steiner, Susanne Fabel, Fraser McLeod, Dionex Corporation, Germering, Germany Abstract Monitoring amino acid

More information

AppNote 5/2006 ABSTRACT

AppNote 5/2006 ABSTRACT AppNote 5/26 Stir Bar Sorptive Extraction Based on Restricted Access Material for the Direct Extraction of Drugs and Metabolites in Biological Fluids Wayne M. Mullett and Elizabeth Kwong Merck Frosst,

More information

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study Application note Biotherapeutics and Biologics Authors Suresh Babu C.V., Anne

More information

Chrom-Card Data System. A simple, chemist-oriented data handling system...productive from the first day!

Chrom-Card Data System. A simple, chemist-oriented data handling system...productive from the first day! c h r o m a t o g r a p h y Chrom-Card Data System A simple, chemist-oriented data handling system......productive from the first day! Analyze Detect Measure Control Chrom-Card Data System A simple, chemist-oriented

More information

Measurement of Hepcidin-25 in Serum Using the Agilent 6490 Triple Quadrupole LC/MS with ifunnel Technology

Measurement of Hepcidin-25 in Serum Using the Agilent 6490 Triple Quadrupole LC/MS with ifunnel Technology Measurement of Hepcidin-25 in Serum Using the Agilent 6490 Triple Quadrupole LC/MS with ifunnel Technology Application Note Clinical Research Authors Christophe Hirtz, Constance Delaby, Pauline Bros, Jérôme

More information

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany New Features of the Agilent Method Scouting Wizard for Automated Method Development of Complex Samples Analysis of Large Data Sets by Method Scouting Reports and Automated Adjustment of Flow Rates and

More information

QUANLYNX APPLICATION MANAGER

QUANLYNX APPLICATION MANAGER QUANLYNX APPLICATION MANAGER OVERVIEW INTRODUCTION QuanLynx, an Application Manager included with Waters MassLynx Software, is designed for quantitative analysis. Simply acquiring mass spectrometry data

More information

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water application note On-line SPE-LC/MS/MS of Pesticides On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water API 3200 LC/MS/MS System Overview The feasibility of

More information

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Columns Guidance for column set up and installation Tips to maximize column lifetime EASY-Spray Column Tips and Tricks This document provides guidance for Thermo Scientific EASY-Spray column

More information

Low pg/ml Detection of rhuepos in Horse Plasma Employing High-Resolution MS

Low pg/ml Detection of rhuepos in Horse Plasma Employing High-Resolution MS Low pg/ml Detection of rhuepos in Horse Plasma Employing High-Resolution MS Claire Dauly 1, Patrice Garcia 2, Joel Louette 3, Florence Cormant 2, Madalina Oppermann 4, Marie-Agnès Popot 2, Yves Bonnaire

More information

Application Note USD Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent

Application Note USD Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent Application Note USD 241 Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent What this Study Demonstrates T h i s s t u d y o n C a

More information

PESTICIDE SCREENING APPLICATION SOLUTION

PESTICIDE SCREENING APPLICATION SOLUTION PESTICIDE SCREENING APPLICATION SOLUTION The most innovative pesticide screening solution ever built for food, beverage, and water testing. As the global food trade continues to expand, so does the challenge

More information

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study Featuring the SCIEX BioBA Solution Xi Qu 1, Shuyu Hou 1, Meghan McCann 1, Caroline Becker 1, Xun Wang 1,

More information

MultiQuant Software Version 3.0

MultiQuant Software Version 3.0 MultiQuant Software Version 3.0 Improving Data Quality and Processing Throughput with Better Peak Integration, Quantitative and Qualitative Compound Review for the Analysis of Food, Drinking Water, and

More information

Application Note 00394

Application Note 00394 Application Note 00394 Multi Component Mycotoxin Analysis using LC/MS/MS Mahalakshmi Rudrabhatla, Ph.D., Varian, Inc. Introduction Mycotoxins are secondary metabolites of molds, contaminating a wide range

More information

Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite. Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct.

Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite. Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct. Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct. 12, 211 Outline Orbitrap Mass Spectrometer in mab Characterization Intact

More information

Structure Characterization and Differentiation of Biosimilar and Reference Products Using Unique Combination of Complementary Fragmentation Mechanisms

Structure Characterization and Differentiation of Biosimilar and Reference Products Using Unique Combination of Complementary Fragmentation Mechanisms Structure Characterization and Differentiation of Biosimilar and Reference Products Using Unique Combination of Complementary Fragmentation Mechanisms Zhiqi Hao, 1 Chen Li, 2 Shiaw-Lin Wu, 2,3 David M.

More information