In Vitro Evaluation of Current Thoracic Aortic Stent- Grafts for Real-time MR-Guided Placement

Size: px
Start display at page:

Download "In Vitro Evaluation of Current Thoracic Aortic Stent- Grafts for Real-time MR-Guided Placement"

Transcription

1 62 J ENDOVASC THER EXPERIMENTAL INVESTIGATION In Vitro Evaluation of Current Thoracic Aortic Stent- Grafts for Real-time MR-Guided Placement Holger Eggebrecht, MD 1,2 ; Michael Zenge 2 ; Mark E. Ladd, PhD 2 ; Raimund Erbel, MD 1 ; and Harald H. Quick, PhD 2 Departments of 1 Cardiology and 2 Diagnostic and Interventional Radiology and Neuroradiology, West German Heart Center Essen, University of Duisburg-Essen, Germany. Purpose: To systematically evaluate the magnetic resonance imaging (MRI) characteristics of current thoracic aortic stent-graft devices before, during, and after in vitro deployment as a step toward real-time MRI-guided stent placement. Methods: Six stent-graft devices used for thoracic aortic repair were examined in a dedicated phantom model using a 1.5-T MRI scanner. First, the delivery systems with the mounted stent-graft were examined using real-time fast imaging with steady-state precession (TrueFISP) with Cartesian and radial k-space filling. TrueFISP imaging was subsequently used for real-time monitoring of stent-graft expansion. The deployed stent-grafts were then examined in a water bath containing gadolinium (1:40) with high-resolution T 1 - weighted 3D fast low-angle shot (FLASH) sequences. The images were analyzed for artifacts, radiofrequency caging effects, and device visualization quality. Results: Three delivery systems with mounted stent-grafts did not contain ferromagnetic elements and were well visualized. Imaging with radial k-space filling showed fewer artifacts than Cartesian imaging. Movement of the delivery system and stent-graft expansion of these devices were successfully demonstrated at a rate of up to 6 frames per second. Evaluation of the expanded stent-grafts revealed only minor susceptibility artifacts without relevant signal attenuation in the stent-graft lumen for 5 nitinol-based stent-grafts. Only a stainless steel based stent-graft was associated with severe artifacts, thwarting visualization of its lumen or surroundings. Conclusion: The present study shows that 3 nitinol-based thoracic stent-graft devices are potentially suited for real-time MRI-guided placement with respect to both the delivery system and the stent-graft itself. These observations provide the basis for the evaluation of MRI-guided stent-graft placement in vivo. J Endovasc Ther Key words: magnetic resonance imaging, real-time MRI, MR guidance, thoracic aorta, stent-graft, susceptibility artifact Endovascular stent-graft placement is emerging as a promising alternative to medical therapy and surgery in the treatment of patients with diseases of the descending thoracic aorta. 1,2 Stent-graft implantation is usually performed under fluoroscopy, which has several shortcomings in this setting beyond the ionizing radiation and nephrotoxic contrast material. For example, fluoroscopy does not allow consistent differentiation of the false from Dr. Eggebrecht is recipient of a research grant from the University Duisburg-Essen (IFORES 102). The authors have no commercial, proprietary, or financial interest in any products or companies described in this article. Address for correspondence and reprints: Dr. Holger Eggebrecht, Department of Cardiology, University of Duisburg- Essen, Hufelandstraße 55, Essen, Germany. Fax: ; holger.eggebrecht@uni-essen.de 2006 by the INTERNATIONAL SOCIETY OF ENDOVASCULAR SPECIALISTS Available at

2 J ENDOVASC THER MR EVALUATION OF THORACIC STENT-GRAFTS 63 Figure 1The stent-grafts examined in this study: (A) Relay, (B) Zenith, (C) TAG, (D) Evita, (E) Talent, and (F) Valiant. the true lumen or adequate visualization of the false lumen. Immediate evaluation of procedural success after stent-graft placement (i.e., thrombosis of the false lumen) is thus not possible. These limitations may be overcome by magnetic resonance imaging (MRI) based procedural guidance. MRI guidance of vascular interventional procedures offers several potential advantages over fluoroscopically-guided techniques, including image acquisition in any desired orientation, superior 3-dimensional (3D) soft tissue contrast with simultaneous visualization of the interventional device, absence of ionizing radiation, and avoidance of nephrotoxic contrast media. 3,4 The feasibility of MRIguided vascular interventions has been demonstrated for a variety of procedures, including peripheral and coronary stent placement, 5,6 selective embolization, 7 and implantation of atrial septal closure devices. 8 MRI guidance may be particularly useful for aortic stent-graft placement because it can provide all relevant information for the preinterventional planning of such procedures and can also be used for postinterventional evaluation of treatment success. 9,10 Several studies have suggested that MRI and MR angiography (MRA) techniques may even be superior to standard techniques (i.e., computed tomography [CT]) for detecting endoleaks during follow-up, as well as for evaluation of aneurysm sac exclusion and graft patency. 11,12 An important prerequisite for performing stent-graft placement under MR guidance is MR compatibility of the delivery systems and endografts with respect to device visibility, image artifacts, and postdeployment visualization of the expanded stent-grafts and their lumens. 3,13 In the present study, we evaluated the MRI characteristics of 6 current thoracic aortic stentgrafts in phantom model experiments. METHODS Stent-Graft Devices Five commercially available thoracic stentgrafts (Relay [Bolton Medical, Sunrise, FL, USA], Zenith [Cook, Bloomington, IN, USA], TAG (W.L. Gore & Associates, Flagstaff, AZ, USA], Evita [Jotec, Hechingen, Germany], and Talent [Medtronic Vascular, Santa Rosa, CA, USA]), as well as a prototype device (Valiant; Medtronic Vascular), were examined (Fig. 1). The individual device specifications are given in Table 1. In Vitro Phantom Setup Scanning was performed on a 1.5-T wholebody MRI scanner (Magnetom Sonata, Siemens Medical Solutions, Erlangen, Germany)

3 64 MR EVALUATION OF THORACIC STENT-GRAFTS J ENDOVASC THER TABLE 1 Characteristics of the Stent-Grafts Assessed Metal Support Markers Graft Fabric Relay Nitinol Platinum Dacron Zenith Stainless steel Gold Dacron TAG Nitinol Gold eptfe Evita Nitinol Nitinol Polyester Talent Nitinol Nitinol Polyester Valiant Nitinol Nitinol Polyester eptfe: expanded polytetrafluoroethylene. equipped with gradients capable of a maximum amplitude of 40 mt/m and a slew rate of 200 T/m s 1. In an initial experiment, each stent-graft was mounted on its delivery system and placed in a water bath phantom consisting of a 1-m-long, 10-cm-wide plastic tube with a central hole for linear catheter manipulation over a fishing cord. The phantom (free from any silicone or glass tubes) was designed to allow assessment of the delivery device and stent-graft artifact properties during manual advancement and subsequent stent-graft deployment. The phantom was placed with its longitudinal axis along the axis of the main magnetic field on top of the spine phased-array radiofrequency (RF) coil with 3 coil elements activated for signal reception (Fig. 2). A 2-element body flex phased-array RF coil was placed on top of the phantom. Real-time Imaging of the Delivery Systems Figure 2Experimental setup. The stent-grafts with their delivery systems were contained in a Plexiglas phantom, which consisted of a 1-m-long, 10-cm-diameter water-filled tube that was placed with its longitudinal axis in parallel to the main magnetic field of the scanner on top of the spine phased-array RF coil. A body phased-array RF coil was placed on top of the phantom. The delivery systems were examined with real-time fast imaging with steady-state precession (TrueFISP) sequences using Cartesian and radial k-space filling (Table 2). For the Cartesian sequence, the frequency encoding direction was chosen parallel to the longitudinal direction of the devices and of the phantom; thus, it was also parallel to the direction of the main magnetic field. First, static images of each delivery system with the mounted stent-graft were obtained. Then, the delivery systems were advanced at 2 cm/s as realtime TrueFISP imaging was performed at 1.5 frames per second (fps) for the Cartesian and 6 fps for the radial TrueFISP sequence. For both sequences, images were displayed without detectable image reconstruction delay. Subsequently, stent-graft expansion and deployment were recorded using radial True- FISP sequences. High-Resolution Imaging of the Expanded Stent-Grafts In a second step, the expanded stent-grafts without their delivery systems were examined hanging freely in a plastic container with water and gadolinium (Gd-DTPA) at a concentration of 1:40. This concentration was chosen based on the results of prior in vitro solution series, which approximated the intra-arterial signal intensity of contrast-enhanced 3D MRA of the aorta in patients. Imaging of the stentgrafts was performed using high-resolution

4 J ENDOVASC THER MR EVALUATION OF THORACIC STENT-GRAFTS 65 TABLE 2 Characteristics of the MRI Sequences Evaluated Cartesian TrueFISP Radial TrueFISP T 1 -weighted 3D FLASH Repetition time, ms Echo time, ms Flip angle, Matrix, pixels Field of view, mm Slice thickness, mm Image acquisition time, s (40 slices) Bandwidth, Hz/pixel TrueFISP: real-time steady-state free precession sequences, FLASH: fast low-angle shot. T 1 -weighted 3D fast low-angle shot (FLASH) imaging (Table 2). Minimum intensity projections (MIP) were derived from the 3D FLASH datasets. Image Data Analysis Image quality was evaluated with respect to (1) static visualization of the delivery system, (2) visualization of delivery systems under motion, (3) monitoring of stent-graft expansion, and (4) visualization of the deployed stent-graft. Image quality assessment was performed by a radiologist and a physicist (H.E. and H.H.Q., respectively) in consensus using a 3-point grading scale: 1 indicated severe artifacts with poor visualization of delivery system/stent-graft, 2 denoted minor artifacts allowing fair visualization of delivery system/stent-graft, and 3 referred to excellent visualization of the delivery system/stent-graft with delineation of device details (e.g., stentgraft markers). Assessment of potential signal shielding from the inside of the stent due to RF caging was expressed as percentage signal loss within the stent-graft lumen on the T 1 -weighted 3D FLASH scan with respect to the undisturbed signal in the water/gd-dtpa solution outside of the stent-graft. For this purpose, signal intensity was measured at regions of interest (ROI) within the stent lumen, as well as outside of the stent-graft in coronal and axial orientations. Figure 3Visualization of the TAG delivery system with the mounted stent-graft in static position with real-time TrueFISP with (A) Cartesian and (B) radial k-space filling. (C E) Visualization of catheter movement (2 cm/s) using real-time TrueFISP (5 fps) with radial k-space filling. RESULTS Real-time Imaging of the Delivery Systems Two delivery systems each with their mounted stent-grafts (TAG and Valiant) did not contain ferromagnetic elements and were well visualized by real-time MRI (Figs. 3 5). The Evita delivery system was suitable for MRI examination only after device modification (exchange of a ferromagnetic screw at the catheter tip), whereas the remaining delivery devices (Relay, Talent, and Zenith) caused large susceptibility artifacts and were thus not suited for MRI examination. Back and forth movement of the 3 suitable delivery systems

5 66 MR EVALUATION OF THORACIC STENT-GRAFTS J ENDOVASC THER Figure 4Visualization of the modified Evita delivery system with the mounted stent-graft in static position with real-time TrueFISP with (A) Cartesian and (B) radial k-space filling. (C-E) Visualization of catheter movement (2 cm/s) using real-time TrueFISP (5 fps) with radial k-space filling. in the phantom model was successfully monitored with real-time TrueFISP imaging (Figs. 3 5). The susceptibility artifacts of the stentgrafts enabled clear determination of the position of the loaded stent-graft in relation to the delivery system. Subsequent expansion of the stent-grafts was also successfully visualized under real-time conditions (radial sequence only; Figs. 6 8). Figure 5Visualization of the Valiant delivery system with the mounted stent-graft in static position with real-time TrueFISP with (A) Cartesian and (B) radial k-space filling. (C-E) Visualization of catheter movement (2 cm/s) using real-time TrueFISP (5 fps) with radial k-space filling. All 5 expanded nitinol-based stent-grafts were well visualized by MRI and did not produce disturbing susceptibility artifacts. Selected slices of the high-resolution T 1 -weighted 3D FLASH datasets demonstrated the individual amount of RF signal shielding among the different stent-grafts (Fig. 9). There was no relevant signal attenuation within the lumen of the 5 nitinol-based stent-grafts (Fig. 9). Three stent-grafts (Relay, TAG, and Talent) showed slight regional signal enhancement due to RF resonance effects at longitudinal stent struts (Fig. 9). Quantitative assessment of RF signal caging showed that the percentage attenuation was around 0%, i.e., basically no signal attenuation was observed for the nitinol-based stent-grafts (Table 3). Small variations in the attenuation score might be due to a different location of the measurement ROIs relative to the position of the surface coils (inherent spatial signal variations). Only High-Resolution Imaging of the Expanded Stent-Grafts Figure 6Visualization of the TAG stent-graft deployment using real-time TrueFISP (5 fps) with radial k-space filling (A). The graft is released (B D) in a fraction of a second by retracting a restraining string that prevents the graft from unfolding.

6 J ENDOVASC THER MR EVALUATION OF THORACIC STENT-GRAFTS 67 Figure 7Visualization of the Evita stent-graft deployment using real-time TrueFISP (5 fps) with radial k-space filling (A). The graft is deployed (B D) by retracting the sheath of the delivery system. the stainless steel based Zenith stent-graft caused severe susceptibility artifacts, which did not allow visualization of the stent-graft lumen or its adjacent area. The MIPs derived from the high-resolution T 1 -weighted 3D FLASH scans demonstrated that artifacts from nitinol-based stent-grafts most likely do not compromise the depiction of the tissue directly surrounding the stent-graft (Fig. 10). Again, the stainless steel based stent-graft produced large artifacts, which hindered visualization of the stent-graft lumen and the surrounding tissue. Image Data Analysis Figure 8Visualization of the Valiant stent-graft deployment using real-time TrueFISP (5 fps) with radial k-space filling (A). The graft is slowly released (B D) by retracting a catheter sheath. The position of the retracting catheter sheath is clearly defined by the partly deployed stent-graft (arrows). Comparison of image quality revealed that static imaging with radial k-space filling generally displayed the delivery systems with less pronounced artifacts than imaging using Cartesian k-space filling (Table 4). With radial k-space filling, the artifacts were constrained to the actual device. Images acquired with the radial sequence were thus rated better than images acquired with the Cartesian acquisition scheme (Table 4). Imaging of the delivery systems under motion was associated with less image quality than static imaging independent of radial or Cartesian k-space filling; however, visualization of the position of the loaded stent-graft in relation to the delivery system was still sufficient for device positioning. Here, the higher temporal resolution of the radial sequence (6 fps) compared to the Cartesian sequence (1.5 fps) provided much better depiction of the moving devices, which again resulted in a higher image quality rating for the radial sequence. Owing to the rapid expansion mechanism of the TAG device, its deployment was less well monitored than for the Relay and Valiant stent-grafts. (Figs. 6 8) DISCUSSION The present study evaluated the applicability of 6 current thoracic aortic stent-graft devices for real-time MRI-guided placement. Three nitinol-based stent-grafts in combination with their delivery systems (TAG, Valiant, and with restrictions Evita) were found to be suitable for MRI-based procedural guidance using passive device tracking. Movement and subsequent stent-graft expansion of these devices were well visualized using real-time TrueFISP im-

7 68 MR EVALUATION OF THORACIC STENT-GRAFTS J ENDOVASC THER Figure 9Single coronal slices and corresponding axial slices from the high-resolution T 1 - weighted 3D FLASH sequences: (A) Relay, (B) Zenith, (C) TAG, (D) Evita, (E) Talent, and (F) Valiant. The stainless steel based Zenith stent-graft shows severe susceptibility artifacts, rendering visualization of the stent lumen and stent vicinity impossible. All the other 5 nitinolbased stent-grafts allow detailed evaluation of the stent lumen. Arrows point to slight regional signal enhancement due to RF resonance effects. TABLE 3 RF Signal Attenuation Caused by the Stent-Grafts Lumen Relay 2%0% Zenith NA TAG 5%0% Evita 3%0% Talent 1%0% Valiant 2%3% NA: RF signal attenuations could not be quantified due to severe artifact. aging with a temporal resolution of 6 fps. Moreover, our study demonstrated that expanded nitinol-based stent-grafts were not associated with relevant susceptibility artifacts or relevant RF signal loss within the stent-graft lumen. The only stainless steel based stentgraft was unsuitable for MRI visualization. Collection of precise 3D information about soft tissue anatomy and reliable visualization of catheter instruments in relation to the surrounding anatomy represent the prerequisite for safe and successful performance of vascular interventions under MRI guidance. 14 However, in contrast to ultrasound, fluoroscopy, or CT, visualization of interventional instruments by MRI has proven to be difficult. 15 The optimal MRI technique to visualize intravascular instruments would be characterized by high spatial and temporal resolution and should also provide a high-contrast instrument signature, making it easy to pick out the instrument in the MR image. 16 A number of approaches have been developed for depicting vascular instruments by MRI. Active device tracking is accomplished by incorporating small receiver coils into the tip of a catheter device. The use of active catheter tracking techniques may, however, be potentially hazardous for the patient as long as no safety-tested and commercially available active devices are available. In contrast, passive device tracking by MRI uses susceptibility artifacts of the interventional instrument

8 J ENDOVASC THER MR EVALUATION OF THORACIC STENT-GRAFTS 69 Figure 10Minimum intensity projections from the high-resolution T 1 -weighted 3D FLASH sequences: (A) Relay, (B) Zenith, (C) TAG, (D) Evita, (E) Talent, and (F) Valiant. Except for (B), all stent-grafts allow detailed evaluation of the lumen and surroundings with the high-resolution 3D FLASH sequence. for visualization. Passive tracking requires no hardware or instrument modifications and thus appears to be particularly promising in terms of potential clinical applications. However, passive visualization requires MR-compatible instruments with adequate susceptibility artifacts. The artifact has to allow clear visualization of the interventional device within the aortic lumen, but should not obscure relevant parts of the field of view. 13 Therefore, the artifact ideally should be as small as possible and confined to the device geometry while providing high image contrast against the background. 17 So far, few studies have addressed the suitability of aortic stent-grafts for MRI applica- TABLE 4 Image Quality Assessment With 3-Point Grading* Visualization of Delivery System Cartesian Radial Monitoring of Delivery System Movement Monitoring of Stent-Graft Expansion Visualization of the Expanded Stent-Graft Relay NA NA NA NA 3 Zenith NA NA NA NA NA TAG Evita Talent NA NA NA NA 3 Valiant NA: examination of the delivery system/stent-graft not possible due to ferromagnetic components. * 1: severe artifacts with poor visualization of delivery system/stent-graft, 2: minor artifacts allowing fair visualization of delivery system/stent-graft, 3: excellent visualization of the delivery system/stent-graft with delineation of device details (e.g., stent-graft markers). Evaluation after device modification compared to the standard device.

9 70 MR EVALUATION OF THORACIC STENT-GRAFTS J ENDOVASC THER tions. 3,13,18 Most studies focused on the evaluation of expanded stent-grafts in terms of MRI-based surveillance. 13,18 Only a single study so far has addressed the MRI suitability of both stent-grafts and delivery systems, which is the prerequisite for MRI-guided stent placement. Using various real-time gradientecho MRI sequences of static devices, Mahnken et al. 3 examined bifurcated stent-graft systems designed for abdominal aortic (AA) repair. In contrast, the present study evaluated the MRI visibility of the stent-graft systems and endograft deployment process with both static and real-time imaging under device motion. Furthermore, the present study examined tubular stent-grafts designed for thoracic aortic repair. These differ significantly from their AA counterparts with respect to both the endoprostheses and delivery systems. Similar to the results of Mahnken et al., 3 both the nitinolbased TAG stent-graft and the Valiant prototype (in combination with their delivery systems) were found to be potentially suitable for MRI guidance using passive catheter visualization. Movement of these delivery systems and subsequent stent-graft expansion was well visualized using real-time TrueFISP imaging. Again, in concordance with Mahnken s results, real-time imaging with radial k-space filling showed better image quality than with Cartesian k-space filling. The Evita stent-graft was of limited suitability only after modification of the delivery system, whereas the remaining 3 devices tested (Relay, Zenith, and Talent) contained ferromagnetic parts and were thus not suitable for MRI visualization. To achieve high instrument to background contrast in the dynamic phantom experiments, a fast steady-state free precession (SSFP or TrueFISP) sequence was employed for data acquisition. 19 Such steady-state sequences, when used with high excitation flip angles, provide high blood signal even without administration of a contrast agent. To ensure high temporal resolution for visualization of the dynamic process of catheter movement and subsequent stent-graft deployment, TrueFISP imaging with radial k- space filling provided 6 images per second, which proved sufficient for direct control of catheter movement and monitoring of stentgraft deployment. With respect to MRI surveillance after stentgraft placement, this study demonstrates that the tested nitinol-based stent-grafts did not produce relevant susceptibility artifacts, which allowed visualization of both the aortic lumen and the tissue surrounding the aorta. Quantitative assessment of RF caging basically showed no signal attenuation for the nitinolbased stent-grafts. Postinterventional MRA assessment and follow-up of the implanted stent-grafts with 3D contrast-enhanced FLASH sequences thus will not be impaired by stentgraft related artifacts. Only the stainless steel based Zenith stent-graft produced severe image artifacts, rendering MRI-based follow-up of this stent-graft useless. 18 It is of even greater concern that these artifacts also significantly limit the diagnostic value of future MRI scans performed for other diagnostic purposes (e.g., MR evaluation of cardiac function). Moreover, this stainless steel based stent-graft exerted significant ferromagnetic attraction force, which potentially may impose a risk of stentgraft migration while inside or in the vicinity of the MR scanner. Another potential safety hazard associated with MRI examinations is the possibility of localized increases in the radiofrequency (RF) Specific Absorption Rate (SAR) near metallic implants. The local electric field can be amplified, especially if the implants are composed of long conducting structures that potentially can couple significantly with the RF energy of the body coil. This might lead to excessive local tissue heating if standing resonating waves are generated These resonance effects, however, occur only when the length of wire-like structures or longitudinal implants exceeds one half of the RF wavelength, i.e., around 26 cm for 1.5 T in human tissue. 20 For implants that are small compared to the RF wavelength (e.g., stents, coils, clips, endografts, etc.), RF resonance is considered unlikely. All endografts used in this study were 20 cm in length; consequently, no significant heating of these devices is expected. Accordingly, we have not carried out any experiments with regard to RF heating of tissue surrounding the endografts. Conclusion The present study shows that 3 commercially available nitinol-based stent-grafts in

10 J ENDOVASC THER MR EVALUATION OF THORACIC STENT-GRAFTS 71 combination with their delivery systems (TAG, Valiant, and with restrictions Evita) are readily suitable for real-time MRI-based procedural guidance using passive device visualization. These observations provide the basis for the evaluation of MRI-guided aortic stent-graft placement in vivo. Acknowledgments: The authors thank Dr. Peter Speier, PhD, of Siemens Medical Solutions, Erlangen, Germany, for providing support with the interactive real-time Works-in-Progress imaging protocols. REFERENCES 1. Fattori R, Napoli G, Lovato L, et al. Descending thoracic aortic diseases: stent-graft repair. Radiology. 2003;229: Eggebrecht H, Herold U, Kuhnt O, et al. Endovascular stent-graft treatment of aortic dissection: determinants of post-interventional outcome. Eur Heart J. 2005;26: Mahnken AH, Chalabi K, Jalali F, et al. Magnetic resonance-guided placement of aortic stents grafts: feasibility with real-time magnetic resonance fluoroscopy. J Vasc Interv Radiol. 2004; 15: Manke C, Nitz WR, Lenhart M, et al. Magnetic resonance monitoring of stent deployment: in vitro evaluation of different stent designs and stent delivery systems. Invest Radiol. 2000;35: Spuentrup E, Ruebben A, Schaeffter T, et al. Magnetic resonance guided coronary artery stent placement in a swine model. Circulation. 2002;105: Buecker A, Adam GB, Neuerburg JM, et al. Simultaneous real-time visualization of the catheter tip and vascular anatomy for MR-guided PTA of iliac arteries in an animal model. J Magn Reson Imaging. 2002;16: Fink C, Bock M, Umathum R, et al. Renal embolization: feasibility of magnetic resonanceguidance using active catheter tracking and intraarterial magnetic resonance angiography. Invest Radiol. 2004;39: Buecker A, Spuentrup E, Grabitz R, et al. Magnetic resonance-guided placement of atrial septal closure device in animal model of patent foramen ovale. Circulation. 2002;106: Engellau L, Albrechtsson U, Dahlstrom N, et al. Measurements before endovascular repair of abdominal aortic aneurysms. MR imaging with MRA vs. angiography and CT. Acta Radiol. 2003;44: Thurnher SA, Dorffner R, Thurnher MM, et al. Evaluation of abdominal aortic aneurysm for stent-graft placement: comparison of gadolinium-enhanced MR angiography versus helical CT angiography and digital subtraction angiography. Radiology. 1997;205: Ayuso JR, de Caralt TM, Pages M, et al. MRA is useful as a follow-up technique after endovascular repair of aortic aneurysms with nitinol endoprostheses. J Magn Reson Imaging. 2004; 20: Ersoy H, Jacobs P, Kent CK, et al. Blood pool MR angiography of aortic stent-graft endoleak. AJR Am J Roentgenol. 2004;182: Hilfiker PR, Quick HH, Debatin JF. Plain and covered stent-grafts: in vitro evaluation of characteristics at three-dimensional MR angiography. Radiology. 1999;211: Ladd ME, Quick HH, Debatin JF. Interventional MRA and intravascular imaging. J Magn Reson Imaging. 2000;12: Quick HH, Kuehl H, Kaiser G, et al. Interventional MRA using actively visualized catheters, TrueFISP, and real-time image fusion. Magn Reson Med. 2003;49: Quick HH, Zenge MO, Kuehl H, et al. Interventional magnetic resonance angiography with no strings attached: wireless active catheter visualization. Magn Reson Med. 2005;53: Hilfiker PR, Quick HH, Schmidt M, et al. In vitro image characteristics of an abdominal aortic stent graft: CTA versus 3D MRA. MAGMA. 1999;8: van der Laan MJ, Bartels LW, Bakker CJ, et al. Suitability of 7 aortic stent-graft models for MRI-based surveillance. J Endovasc Ther. 2004;11: Hunold P, Maderwald S, Eggebrecht H, et al. Steady-state free precession sequences in myocardial first-pass perfusion MR imaging: comparison with TurboFLASH imaging. Eur Radiol. 2004;14: Konings MK, Bartels LW, Smits HF, et al. Heating around intravascular guidewires by resonating RF waves. J Magn Reson Imaging. 2000; 12: Ladd ME, Quick HH. Reduction of resonant RF heating in intravascular catheters using coaxial chokes. Magn Reson Med. 2000;43: Nitz WR, Oppelt A, Renz W, et al. On the heating of linear conductive structures as guide wires and catheters in interventional MRI. J Magn Reson Imaging. 2001;13:

Chapter 3. An evaluation of the suitability of endografts for an MRI-based follow-up after EVAR

Chapter 3. An evaluation of the suitability of endografts for an MRI-based follow-up after EVAR Chapter 3 An evaluation of the suitability of endografts for an MRI-based follow-up after EVAR Maarten J. van der Laan Lambertus W. Bartels Chris J.G. Bakker Max A. Viergever Jan D. Blankensteijn J. Endovasc.

More information

A New Endoprosthesis With Sensing Capability

A New Endoprosthesis With Sensing Capability A New Endoprosthesis With Sensing Capability Isa C.T. Santos, Alexandra Sepúlveda, Luís Rocha & João Manuel R.S. Tavares ICABB 2010 October 14-16 Venice, Italy Outline Aneurysm Treatment options Thoracic

More information

Modular branched endograft system for aortic aneurysm repair: evaluation in a human cadaver circulation model

Modular branched endograft system for aortic aneurysm repair: evaluation in a human cadaver circulation model Modular branched endograft system for aortic aneurysm repair: evaluation in a human cadaver circulation model Vascular and Endovascular Surgery. 2007 41: 126 9 M.A.M. Linsen A.W.F. Vos J. Diks J.A. Rauwerda

More information

Magnetic Resonance Imaging (MRI) Safety for Boston Scientific Products. 1. WALLSTENT Iliac Endoprosthesis with Unistep Plus Delivery System

Magnetic Resonance Imaging (MRI) Safety for Boston Scientific Products. 1. WALLSTENT Iliac Endoprosthesis with Unistep Plus Delivery System Magnetic Resonance Imaging (MRI) Safety for Boston Scientific Products Table of Contents 1. WALLSTENT Iliac Endoprosthesis with Unistep Plus Delivery System.1 2. WALLSTENT RP Endoprosthesis (Transhepatic

More information

CTA Throughout the Ages

CTA Throughout the Ages CTA Throughout the Ages Suhny Abbara, MD Associate Professor, Harvard Medical School Director of Education, Cardiac MRCT Program Director Cardiovascular Imaging Fellowship, Massachusetts General Hospital

More information

PER F ORM ANC E b y d e s i g n. Pioneering TEVAR Therapy, Time and Time Again. CONFORMABLE

PER F ORM ANC E b y d e s i g n. Pioneering TEVAR Therapy, Time and Time Again. CONFORMABLE PER F ORM ANC E b y d e s i g n Pioneering TEVAR Therapy, Time and Time Again. CONFORMABLE T H O R A C I C E N D O P R O S T H E S I S Time-Tested Success For more than 16 years, the GORE TAG Device has

More information

Young-Guk Ko, M.D. Severance Cardiovascular Hospital, Yonsei University Health System,

Young-Guk Ko, M.D. Severance Cardiovascular Hospital, Yonsei University Health System, Young-Guk Ko, M.D., First EVAR Endograft in Human 1991 Annals of Vascular Surgery Dr. Juan Carlos Parodi, Surgeon Dr. Julio Palmaz, Radiologist Parodi-Palmaz endograft: Dacron + modified Palmaz stents

More information

Conformability without Compromise

Conformability without Compromise Conformability without Compromise THE STANDARD IN Conformability AND Designed for flexibility and conformability in tortuous anatomy. Optimized aortic wall apposition in angulated arch anatomy without

More information

Conformability without Compromise

Conformability without Compromise Conformability without Compromise THE STANDARD IN Conformability AND Designed for flexibility and conformability in tortuous anatomy. Optimized aortic wall apposition in angulated arch anatomy without

More information

V. Riambau and G. Mestres Vascular Surgery Division. Cardiovascular Institute Hospital Clinic. University of Barcelona

V. Riambau and G. Mestres Vascular Surgery Division. Cardiovascular Institute Hospital Clinic. University of Barcelona V. Riambau and G. Mestres Vascular Surgery Division. Cardiovascular Institute Hospital Clinic. University of Barcelona Consultant/Advisor Bolton Medical Cordis Bard Medtronic Aptus Lombard Medical ivascular

More information

Conformability without Compromise. Now FDA APPROVED

Conformability without Compromise. Now FDA APPROVED Conformability without Compromise Now FDA APPROVED THE STANDARD IN Conformability AND Designed for flexibility and conformability in tortuous anatomy. Optimized aortic wall apposition in angulated arch

More information

GDC Detachable Coils, GDC 360o Detachable Coils, Matrix2 Detachable Coils, and Target Detachable Coils

GDC Detachable Coils, GDC 360o Detachable Coils, Matrix2 Detachable Coils, and Target Detachable Coils GDC Detachable Coils, GDC 0o Detachable Coils, Matrix Detachable Coils, and Target Detachable Coils Magnetic Resonance Imagining (MRI) Compatibility (Neurovascular Use) Non-clinical testing and analysis

More information

PRINCIPLES OF CT AND MR IMAGING Marc-André d Anjou, DMV, DACVR Faculty of Veterinary Medicine, University of Montreal Saint-Hyacinthe, Quebec, Canada

PRINCIPLES OF CT AND MR IMAGING Marc-André d Anjou, DMV, DACVR Faculty of Veterinary Medicine, University of Montreal Saint-Hyacinthe, Quebec, Canada PRINCIPLES OF CT AND MR IMAGING Marc-André d Anjou, DMV, DACVR Faculty of Veterinary Medicine, University of Montreal Saint-Hyacinthe, Quebec, Canada CT and MR imaging offer superior diagnostic possibilities

More information

Fully MR-Guided Hepatic Artery Catheterization for Selective Drug Delivery: A Feasibility Study in Pigs

Fully MR-Guided Hepatic Artery Catheterization for Selective Drug Delivery: A Feasibility Study in Pigs JOURNAL OF MAGNETIC RESONANCE IMAGING 23:123 129 (2006) Original Research Fully MR-Guided Hepatic Artery Catheterization for Selective Drug Delivery: A Feasibility Study in Pigs Jan-Henry Seppenwoolde,

More information

Poster No.: R Page 1 of 12

Poster No.: R Page 1 of 12 Abdominal endovascular aortic repair (EVAR): Comparison of image fusion (IF3D) angiography versus conventional (2D) angiography with respect to intravenous contrast dose, radiation dose and total fluoroscopy

More information

Zenith Alpha System, Designed for Easy and Durable repairs. Prof. Elixène JEAN-BAPTISTE (Nice)

Zenith Alpha System, Designed for Easy and Durable repairs. Prof. Elixène JEAN-BAPTISTE (Nice) Zenith Alpha System, Designed for Easy and Durable repairs Prof. Elixène JEAN-BAPTISTE (Nice) Aortic Disease Can Progress AFTER Repair Resulting in: Stent migration Endoleaks Sac enlargement Secondary

More information

Branched Grafts for Thoracoabdominal Aneurysms: Off-Label Use of FDA-Approved Devices

Branched Grafts for Thoracoabdominal Aneurysms: Off-Label Use of FDA-Approved Devices 471 TECHNICAL NOTE Branched Grafts for Thoracoabdominal Aneurysms: Off-Label Use of FDA-Approved Devices Karthik, MD Emory University School of Medicine, Atlanta, Georgia, USA. Purpose: To report off-label

More information

First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector

First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector 01 White Paper No. 02/2009 First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector Leiden University Medical Center (LUMC) in the Netherlands is the first hospital in the world

More information

Design and Preliminary Clinical Studies of an MRI-Guided Transrectal Prostate Intervention System

Design and Preliminary Clinical Studies of an MRI-Guided Transrectal Prostate Intervention System Design and Preliminary Clinical Studies of an MRI-Guided Transrectal Prostate Intervention System Axel Krieger 1, Peter Guion 2, Csaba Csoma 1, Iulian Iordachita 1, Anurag K. Singh 2,3, Aradhana Kaushal

More information

Clinical Applications. ImagingRite. Interventional Radiology

Clinical Applications. ImagingRite. Interventional Radiology Clinical Applications ImagingRite Interventional Radiology ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in optimizing

More information

Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA

Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA Time-resolved 4D TWIST MR angiography of congenital heart and vessel anomalies: image quality and added diagnostic value compared to 3D MRA Poster No.: C-1663 Congress: ECR 2012 Type: Scientific Paper

More information

Power of BRANSIST safire in Neuroendovascular Therapy

Power of BRANSIST safire in Neuroendovascular Therapy Vascular Power of BRANSIST safire in Neuroendovascular Therapy Department of Radiology, Kinki University Hospital Suguru Ueda Mr. Suguru Ueda 1. Introduction Kinki University Hospital is located in the

More information

ENDOVASCULAR TREATMENT OF COMPLEX ABDOMINAL AND THORACOABDOMINAL

ENDOVASCULAR TREATMENT OF COMPLEX ABDOMINAL AND THORACOABDOMINAL Editors G. TORsello - K. P. DONAs M. Austermann B. Berekoven T. Bisdas Contributors M. Lachat F. Marques de Azevedo G. Panuccio ENDOVASCULAR TREATMENT OF COMPLEX ABDOMINAL AND THORACOABDOMINAL ANEURYSMS

More information

Development of a patient-specific modular vascular phantom with clinically relevant mechanical properties. Academic Year:

Development of a patient-specific modular vascular phantom with clinically relevant mechanical properties. Academic Year: Development of a patient-specific modular vascular phantom with clinically relevant mechanical properties Student: 852517 Supervisor: Co-Supervisor: Prof. Elena De Momi Dr. Helge Wurdemann Academic Year:

More information

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite Clinical Applications ImagingRite Neuro Intervention 1 ImagingRite ImagingRite, a comprehensive suite of imaging tools offered with Infinix -i angiographic systems, was designed to assist clinicians in

More information

Predictive and patient-specific numerical simulations of endovascular surgery for aortic abdominal aneurysms. Michel Rochette

Predictive and patient-specific numerical simulations of endovascular surgery for aortic abdominal aneurysms. Michel Rochette Predictive and patient-specific numerical simulations of endovascular surgery for aortic abdominal aneurysms Michel Rochette 1 Our Partners This work has been partially supported by French national research

More information

Wireless Implantable Micromachined Pressure Sensors: Approach and Clinical Demonstration

Wireless Implantable Micromachined Pressure Sensors: Approach and Clinical Demonstration The Problem: Abdominal Aortic Aneurysms (AAA) Wireless Implantable Micromachined Pressure Sensors: Approach and Clinical Demonstration Mark G. Allen School of Electrical and Computer Engineering Georgia

More information

1. Until the CE Mark suspension has been lifted, the AFX product should not be implanted in patients.

1. Until the CE Mark suspension has been lifted, the AFX product should not be implanted in patients. 2 Musick Irvine, CA 92618 949 595 7200 endologix.com Field Safety Notice AFX TM Endovascular AAA System: Suspension of CE Mark 29744/29731 and Patient Follow-up Advice Type of Action: Product Codes: Customer

More information

Clarity CT Technology

Clarity CT Technology Clarity CT Technology WHITE PAPER January 2013 Using state of the art algorithms Sapheneia Clarity CT allows physicians to lower radiation dose when acquiring CT data while maintaining image quality. The

More information

Progress in X-Ray & MR

Progress in X-Ray & MR Progress in X-Ray & MR Michiel Manuel Analyst Meeting June 15 th, 2005 X-Ray & MR: Agenda Introduction General X-Ray Cardio/Vascular X-Ray Magnetic Resonance China growth opportunity Conclusion 2 X-Ray

More information

You struggle to get into the fenestration: try a steerable guiding

You struggle to get into the fenestration: try a steerable guiding You struggle to get into the fenestration: try a steerable guiding G Pratesi, MD Vascular Surgery University of Rome Tor Vergata Disclosure Giovanni Pratesi, M.D. I have the following potential conflicts

More information

UGG, CT AND MRI- WHATS ON THE MENU CONSULTANT RADIOLOGIST- SURAKSHA DIAGNOSTIC CENTRE, SAINI ENCLAVE DR MONOJIT MONDAL MBBS,MD,DNB

UGG, CT AND MRI- WHATS ON THE MENU CONSULTANT RADIOLOGIST- SURAKSHA DIAGNOSTIC CENTRE, SAINI ENCLAVE DR MONOJIT MONDAL MBBS,MD,DNB UGG, CT AND MRI- WHATS ON THE MENU CONSULTANT RADIOLOGIST- SURAKSHA DIAGNOSTIC CENTRE, SAINI ENCLAVE DR MONOJIT MONDAL MBBS,MD,DNB THREE COURSE MENU COMPLIMENTARY NOT SUBSTITUTION USG/CT/MRI WHICH IS BETTER???

More information

Endovascular Device Testing A Survey of Requirements for Product Validation

Endovascular Device Testing A Survey of Requirements for Product Validation MET A4 1p Endovascular Device Testing Info Sheet-02.qxp_MET A4 1p Endovascular Device Testing Info Sheet.qxp 05/04/2016 09:58 Page 1 Endovascular Device Testing A Survey of Requirements for Product Validation

More information

Table of Contents. Adaptive Diagnostics...7 Integrated Dose Reduction Streamlined Workflow Clinical Images... 26

Table of Contents. Adaptive Diagnostics...7 Integrated Dose Reduction Streamlined Workflow Clinical Images... 26 2 Table of Contents Adaptive Diagnostics...7 Integrated Dose Reduction... 17 Streamlined Workflow... 21 Clinical Images... 26 3 AQUILION TM PRIME PROVIDES CLINICAL FLEXIBILITY, ENHANCED WORKFLOW FEATURES

More information

Replacement Heart Valve Product Description (Stented Tissue) Models Reference

Replacement Heart Valve Product Description (Stented Tissue) Models Reference Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences LLC

More information

EVAR Guided by 3D Image Fusion and CO 2 DSA: A New Imaging Combination for Patients With Renal Insufficiency

EVAR Guided by 3D Image Fusion and CO 2 DSA: A New Imaging Combination for Patients With Renal Insufficiency 605468JETXXX10.1177/1526602815605468Journal of Endovascular TherapyKoutouzi et al research-article2015 Technical Note EVAR Guided by 3D Image Fusion and CO 2 DSA: A New Imaging Combination for Patients

More information

To Whom It May Concern:

To Whom It May Concern: To Whom It May Concern: DeRoyal Foley Catheters with Temperature Sensors have been tested for safe use in magnetic resonance environments at 1.5 and 3.0 Tesla according to ASTM International F2052, Standard

More information

PEAC. May 13 // May Marie Lannelongue Hospital le Plessis-Robinson, FRANCE. Paris Endovascular Aortic Course

PEAC. May 13 // May Marie Lannelongue Hospital le Plessis-Robinson, FRANCE. Paris Endovascular Aortic Course PEAC Paris Endovascular Aortic Course May 13 // May 18 2019 Marie Lannelongue Hospital le Plessis-Robinson, FRANCE PEAC COURSE DIRECTORS Stéphan Haulon, Marie Lannelongue Hospital, France Tara Mastracci,

More information

TREAT SMARTER. ACHIEVE MORE. When Treating Thrombus

TREAT SMARTER. ACHIEVE MORE. When Treating Thrombus TREAT SMARTER. ACHIEVE MORE. When Treating Thrombus Treat Smarter When you want fast, safe, predictable results EKOS Ultrasound Accelerated Thrombolysis Unlocks The Clot Without Ultrasonic Energy With

More information

Intracranial stent visualization for image guided interventions and therapy

Intracranial stent visualization for image guided interventions and therapy Intracranial stent visualization for image guided interventions and therapy Daniel Ruijters, Peter van de Haar, Ruben Roijers, Niels J. Noordhoek, Jan Timmer, and Drazenko Babic interventional X-Ray, Philips

More information

Protect the Neck. Lieven Maene, MD. Marc Bosiers Koen Deloose Joren Callaert. Patrick Peeters Jürgen Verbist. Lieven Maene Roel Beelen.

Protect the Neck. Lieven Maene, MD. Marc Bosiers Koen Deloose Joren Callaert. Patrick Peeters Jürgen Verbist. Lieven Maene Roel Beelen. A.Z. Sint-Blasius, Dendermonde Marc Bosiers Koen Deloose Joren Callaert Imelda Hospital, Bonheiden Protect the Neck Patrick Peeters Jürgen Verbist OLV Hospital, Aalst Lieven Maene Roel Beelen R.Z. Heilig

More information

Portable and Mobile X-ray Units

Portable and Mobile X-ray Units RADIOLOGY Portable and Mobile X-ray Units Tecmed Africa offers a full range of mobile Radiology systems from a very select group of high quality suppliers. Specialised mobile systems with either manual

More information

Fast enough to stop the Capable of delineating Unprecedented imaging power for the. Virtual endoscopy. The gatewa

Fast enough to stop the Capable of delineating Unprecedented imaging power for the. Virtual endoscopy. The gatewa Fast enough to stop the Capable of delineating Unprecedented imaging power for the M U L T I S L I C E Virtual endoscopy The gatewa motion of a beating heart. anatomic structures as small as 0.25mm. earliest,

More information

Replacement Heart Valve Product Description (Stented Tissue) Models Reference

Replacement Heart Valve Product Description (Stented Tissue) Models Reference Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences LLC

More information

Metallic Neurosurgical Implants: Evaluation of Magnetic Field Interactions, Heating, and Artifacts at 1.5-Tesla

Metallic Neurosurgical Implants: Evaluation of Magnetic Field Interactions, Heating, and Artifacts at 1.5-Tesla JOURNAL OF MAGNETIC RESONANCE IMAGING 14:295 299 (2001) Technical Note Metallic Neurosurgical Implants: Evaluation of Magnetic Field Interactions, Heating, and Artifacts at 1.5-Tesla Frank G. Shellock,

More information

Optima IGS 330 Versatility within your reach. gehealthcare.com

Optima IGS 330 Versatility within your reach. gehealthcare.com Optima IGS 330 Versatility within your reach gehealthcare.com Versatility within your reach Whether you are creating a new diagnostic service to improve access to patient care or building an additional

More information

LINC James F. McKinsey, M.D.

LINC James F. McKinsey, M.D. Two-Year Evaluation of Fenestrated and Parallel Branch Endografts for the Treatment of Juxtrarenal, Suprarenal and Thoracoabdominal Aneurysms at a Single Institution LINC 2018 James F. McKinsey, M.D. The

More information

Artifacts Caused by Eddy Current in Diffusion MRI

Artifacts Caused by Eddy Current in Diffusion MRI Artifacts Caused by Eddy Current in Diffusion MRI Xi Tan ABSTRACT Magnetic resonance diffusion imaging is potentially an important tool for the noninvasive characterization of normal and pathological tissue.

More information

Comprehensive Dose Management for Improved Patient Care.

Comprehensive Dose Management for Improved Patient Care. Comprehensive Dose Management for Improved Patient Care. Set the standard for dose management Toshiba is committed to continuously advancing dose reduction technologies and incorporating these into its

More information

Replacement Heart Valve Product Description (Stented Tissue) Models Reference

Replacement Heart Valve Product Description (Stented Tissue) Models Reference Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences LLC

More information

Balancing versatility and value Introducing Optima MR T

Balancing versatility and value Introducing Optima MR T GE Healthcare Balancing versatility and value Introducing Optima MR360 1.5T Optima MR360 1.5T Providing an innovative balance in MR You know what you need from an MR scanner. The new Optima MR360 is engineered

More information

Valiant Navion device enhancements: a case-based approach to Navion s clinical value

Valiant Navion device enhancements: a case-based approach to Navion s clinical value Valiant Navion device enhancements: a case-based approach to Navion s clinical value Luigi Lovato Cardiovascular Radiology Unit Program In collaboration with Cardiac Surgery Unit S. Orsola University Hospital

More information

Failure modes of thoracic endografts: Prevention and management

Failure modes of thoracic endografts: Prevention and management VASCULAR AND ENDOVASCULAR TECHNIQUES Thomas L. Forbes, MD, Section Editor Failure modes of thoracic endografts: Prevention and management W. Anthony Lee, MD, Gainesville, Fla Failure modes of thoracic

More information

Replacement Heart Valve Product Description (Stented Tissue) Models Reference

Replacement Heart Valve Product Description (Stented Tissue) Models Reference Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences LLC

More information

In-vivo Targeting of Liver Lesions with a Navigation System based on Fiducial Needles

In-vivo Targeting of Liver Lesions with a Navigation System based on Fiducial Needles In-vivo Targeting of Liver Lesions with a Navigation System based on Fiducial Needles L. Maier-Hein 1, A. Tekbas 2, A. Seitel 1, F. Pianka 2, S. A. Müller 2, S. Schawo 3, B. Radeleff 3, R. Tetzlaff 1,4,

More information

Septal repair implants: evaluation of magnetic resonance imaging safety at 3 T

Septal repair implants: evaluation of magnetic resonance imaging safety at 3 T Magnetic Resonance Imaging 23 (2005) 1021 1025 Septal repair implants: evaluation of magnetic resonance imaging safety at 3 T Frank G. Shellock4, Samuel Valencerina Department of Radiology and Medicine,

More information

Our Experience with Trinias Biplane and Hybrid Operating Room Systems

Our Experience with Trinias Biplane and Hybrid Operating Room Systems Vascular Our Experience with Trinias Biplane and Hybrid Operating Room Systems Department of Cardiology, Mitsubishi Kyoto Hospital Masashi Kato Masashi Kato, M.D. 1. Overview of the hospital The Mitsubishi

More information

ClinScan. think forward. Small Animal MRI Solution for Molecular Imaging and Translational Research. Preclinical MRI

ClinScan. think forward. Small Animal MRI Solution for Molecular Imaging and Translational Research. Preclinical MRI ClinScan Small Animal MRI Solution for Molecular Imaging and Translational Research think forward Preclinical MRI Be Clinical from the Very Beginning With ClinScan you enter the field of translational

More information

Charles A Mistretta PhD

Charles A Mistretta PhD Charles A Mistretta PhD Digital Subtraction Angiography State of the Art Neuro-angiography Suite 1972 In 1980 only device capable of accessing the intracranial arteries 1 75 year old woman with Pcom aneurysm

More information

CT post processing and low dose scanning

CT post processing and low dose scanning CT post processing and low dose scanning Gabor Szell GE Healthcare CT Modality Manager EE Annual Scientific and Educational Meeting Innovations in Cardiothoracic Imaging 201 13-14 May 2011, Tokuda Hospital

More information

CQIE MRI PROCEDURES. American College of Radiology Clinical Research Center. Centers for Quantitative Imaging Excellence LEARNING MODULE

CQIE MRI PROCEDURES. American College of Radiology Clinical Research Center. Centers for Quantitative Imaging Excellence LEARNING MODULE Centers for Quantitative Imaging Excellence LEARNING MODULE CQIE MRI PROCEDURES American College of Radiology Clinical Research Center Imaging Core Laboratory v.2 Centers for Quantitative Imaging Excellence

More information

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals Koos Geleijns, medical physicist CT delivers excellent 3D image quality CT delivers excellent 3D image quality

More information

A Novel Segmentation and Navigation Tool for Endovascular Stenting of Aortic Aneurysms

A Novel Segmentation and Navigation Tool for Endovascular Stenting of Aortic Aneurysms A Novel Segmentation and Navigation Tool for Endovascular Stenting of Aortic Aneurysms Marco Feuerstein, Konstantinos Filippatos, Oliver Kutter, Eva U. Schirmbeck, Robert Bauernschmitt, Nassir Navab 01

More information

Computer-Aided Surgical Navigation Coding Guide Neurosurgery. May 1, 2009

Computer-Aided Surgical Navigation Coding Guide Neurosurgery. May 1, 2009 Computer-Aided Surgical Navigation Coding Guide Neurosurgery May 1, 2009 Please direct any questions to: Kim Brew Manager, Reimbursement and Therapy Access Medtronic Surgical Technologies (904) 279-7569

More information

Interventional Fluoroscopy

Interventional Fluoroscopy NATIONAL CANCER INSTITUTE THE SOCIETY OF INTERVENTIONAL RADIOLOGY Interventional Fluoroscopy Reducing Radiation Risks for Patients and Staff CONTENTS Introduction 1 Increasing use and complexity of interventional

More information

The Unique, New MRI Philips Ingenia 3 Tesla is now in Ayios Therissos! The first-ever digital broadband MR system has been installed in Ayios

The Unique, New MRI Philips Ingenia 3 Tesla is now in Ayios Therissos! The first-ever digital broadband MR system has been installed in Ayios The Unique, New MRI Philips Ingenia 3 Tesla is now in Ayios Therissos! The first-ever digital broadband MR system has been installed in Ayios Therissos-Nicosia that delivers crystal clear images, remarkable

More information

Section: Magnetic Resonance Imaging

Section: Magnetic Resonance Imaging Section: Magnetic Resonance Imaging Available masters project: Cardiac shape modeling for automated wall motion analysis Overview One of the most popular segmentation methods in medical imaging is the

More information

An ultrasonic volumetric scanner for image-guided surgery

An ultrasonic volumetric scanner for image-guided surgery International Congress Series 1230 (2001) 190 196 An ultrasonic volumetric scanner for image-guided surgery Jeremy Johnson a,b, *,Ömer Oralkan b, Kambiz Kaviani b, Utkan Demirci b, Mustafa Karaman b, Pierre

More information

Contains Nonbinding Recommendations. Draft Not for Implementation

Contains Nonbinding Recommendations. Draft Not for Implementation 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 Assessment of Radiofrequency- Induced Heating in the Magnetic Resonance (MR) Environment for Multi-Configuration

More information

Quality Control Concerns for New Generations of CT, MRI, & PET

Quality Control Concerns for New Generations of CT, MRI, & PET Quality Control Concerns for New Generations of CT, MRI, & PET David W. Jordan, Ph.D. Sr. Medical Physicist, UH Case Medical Center Clinical Assistant Professor, CWRU Disclosures No financial disclosures

More information

3D Imaging: How to make beautiful images

3D Imaging: How to make beautiful images 3D Imaging: How to make beautiful images Sebastian Schafer, PhD Siemens Medical Solutions USA Inc. Acknowledgments All cases are courtesy of RUSH University Hospital (D. Lopes, M. Chen) unless explicitly

More information

Integrated on-board CBCT-US imaging system for soft tissue IGRT and real-time intra-fraction monitoring

Integrated on-board CBCT-US imaging system for soft tissue IGRT and real-time intra-fraction monitoring Integrated on-board CBCT-US imaging system for soft tissue IGRT and real-time intra-fraction monitoring John Wong Radiation Oncology and Molecular Radiation Sciences Johns Hopkins University School of

More information

Interventional Radiology

Interventional Radiology Angiography Systems Interventional Radiology *Pictured system is INFX-8000C Unprecedented Flexibility and Integration. Today s radiology interventions require speed, flexibility and high performance. Canon

More information

12 pericardial bioprostheses. 7300TFX Carpentier-Edwards bioprosthetic valved conduit

12 pericardial bioprostheses. 7300TFX Carpentier-Edwards bioprosthetic valved conduit Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences (formerly

More information

Evolution of Embolization

Evolution of Embolization Evolution of Embolization WITH JAFAR GOLZARIAN, MD; FRANCOIS CORNELIS, MD, PhD; COREY L. TEIGEN, MD; DMITRI E. SAMOILOV, MD; AND FRANK R. ARKO III, MD Jafar Golzarian, MD Professor of Radiology Department

More information

Hybrid Rooms for Interventional Radiology

Hybrid Rooms for Interventional Radiology Francois Nolte, Hybrid Rooms for Interventional Radiology Answers for life. Hybrid Rooms for Interventional Radiology 1. Trends in Angiography 2. Rise of the hybrid operating room 3. Key Deployment Considerations

More information

3000, 3000TFX 12 bioprostheses Carpentier-Edwards PERIMOUNT pericardial aortic and mitral

3000, 3000TFX 12 bioprostheses Carpentier-Edwards PERIMOUNT pericardial aortic and mitral Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences (formerly

More information

V. Riambau, MD, PhD. Professor and Chief of Vascular Surgery Division CardioVascular Institute, Hospital Clinic, University of Barcelona

V. Riambau, MD, PhD. Professor and Chief of Vascular Surgery Division CardioVascular Institute, Hospital Clinic, University of Barcelona V. Riambau, MD, PhD Professor and Chief of Vascular Surgery Division CardioVascular Institute, Hospital Clinic, University of Barcelona vriambau@clinic.ub.es Consultant/Advisor Bolton Medical Cordis Bard

More information

Emerging Applications and Trends Across Medical Imaging

Emerging Applications and Trends Across Medical Imaging MEDICAL DEVICES PHARMACEUTICALS CHEMICALS FOOD & BEVERAGE ELECTRONICS Emerging Applications and Trends Across Medical Imaging VPG Publications, Consulting, Clients www.vpgcorp.com VPG Market Research Reports

More information

Whitepaper: Precision Micro-Braiding for Implantable Devices

Whitepaper: Precision Micro-Braiding for Implantable Devices Whitepaper: Precision Micro-Braiding for Implantable Devices Keys to Design Success When Contemplating Medical Device Braids By: Robert Kiefer and Keith Smith Braided textile structures are among the most

More information

High quality and fast scanning with Compressed SENSE at KCH

High quality and fast scanning with Compressed SENSE at KCH FieldStrength MRI articles High quality and fast scanning with at KCH FieldStrength MRI magazine User experiences - November 2018 www.philips.com/fieldstrength At Kurashiki Central Hospital, use of resulted

More information

MRI. Magnetic Resonance Imaging

MRI. Magnetic Resonance Imaging MRI Magnetic Resonance Imaging Key Points MRI: The magnet is always on Know the essential components of an MRI system Know general idea of how MRI captures an image; three steps: Magnetic Alignment, Proton

More information

Optimize efficiency with a universal system.

Optimize efficiency with a universal system. wins www.siemens.com/artis-one Optimize efficiency with a universal system.. Designed around you. Answers for life. Consistently efficient. Improve your competitiveness by mastering the balancing act between

More information

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences Poster No.: C-0674 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Sofue 1, T. Yoshikawa 1, N. Negi 1, Y. Ohno

More information

INTRODUCTION. Best Practices in MR Imaging: State-of-the-Art Protocols. 5th Edition. MR Protocols for GE Healthcare Scanners.

INTRODUCTION. Best Practices in MR Imaging: State-of-the-Art Protocols. 5th Edition. MR Protocols for GE Healthcare Scanners. INTRODUCTION These magnetic resonance (MR) protocols were developed by an expert consensus panel for use on General Electric (GE) MR imaging machines, and were developed for high-end platform scanners

More information

Contains Nonbinding Recommendations Draft - Not for Implementation

Contains Nonbinding Recommendations Draft - Not for Implementation Reprinted from FDA s website by EAS Consulting Group, LLC Select Updates for Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems Draft Guidance

More information

MRI Conditional Devices and Patient Safety. Outline. MRI Labeling: Objects 8/3/13 1. David W. Jordan, Ph.D. Senior Medical Physicist

MRI Conditional Devices and Patient Safety. Outline. MRI Labeling: Objects 8/3/13 1. David W. Jordan, Ph.D. Senior Medical Physicist MRI Conditional Devices and Patient Safety David W. Jordan, Ph.D. Senior Medical Physicist 8/3/13 1 Outline MRI Conditional Overview Spatial Gradient Hazards RF SAR Hazards Device Registries MRI Labeling:

More information

Contrast media administration - Alternatives to the cubital vein

Contrast media administration - Alternatives to the cubital vein Contrast media administration - Alternatives to the cubital vein Radiologists must be prepared for administration of contrast media via the intraosseous route & CVC Hampus Eklöf MD, PhD CMO Unilabs radiology

More information

A Coordinated Registry Network Based on the Vascular Quality Initiative: VISION. Vascular Implant Surveillance & Interventional Outcomes Network

A Coordinated Registry Network Based on the Vascular Quality Initiative: VISION. Vascular Implant Surveillance & Interventional Outcomes Network A Coordinated Registry Network Based on the Vascular Quality Initiative: VISION Vascular Implant Surveillance & Interventional Outcomes Network Jack L. Cronenwett, MD Medical Director, Society for Vascular

More information

Testing Methods and Standards for Magnetic Resonance (MR) Safety and Compatibility of Medical Devices - Update

Testing Methods and Standards for Magnetic Resonance (MR) Safety and Compatibility of Medical Devices - Update Testing Services for Magnetic Resonance Imaging Therapy Testing Methods Stards for Magnetic Resonance (MR) Safety Compatibility of Medical Devices - Update G. Schaefers MR:comp GmbH MR Testing Laboratory

More information

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014 CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014 Answers for life. Answers for life. Is 1 msv the right dose for every patient? Female, 54 Abdominal CT Male,

More information

ISO Cardiovascular implants Endovascular devices Part 2: Vascular stents Part 2: Vascular stent

ISO Cardiovascular implants Endovascular devices Part 2: Vascular stents Part 2: Vascular stent INTERNATIONAL STANDARD ISO 25539-2 Second edition 2012-12-01 Cardiovascular implants Endovascular devices Part 2: Vascular stents Part 2: Vascular stent Implants cardiovasculaires Dispositifs endovasculaires

More information

CT to Ultrasound Registration: A Porcine Phantom Study. J. Xiang, S. Gill, C. Nguan, P. Abolmaesumi and R. N. Rohling University of British Columbia

CT to Ultrasound Registration: A Porcine Phantom Study. J. Xiang, S. Gill, C. Nguan, P. Abolmaesumi and R. N. Rohling University of British Columbia CT to Ultrasound Registration: A Porcine Phantom Study J. Xiang, S. Gill, C. Nguan, P. Abolmaesumi and R. N. Rohling University of British Columbia 1 Motivation Ultrasound to CT registration can improve

More information

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology

Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Radiology Rounds A Newsletter for Referring Physicians Massachusetts General Hospital Department of Radiology Multi-Energy Computed Tomography - New Opportunities In Imaging the Abdomen Materials can be

More information

Simulator Features. Imaging. Device Simulation. Patient Management. Educational Features

Simulator Features. Imaging. Device Simulation. Patient Management. Educational Features TM ANGIO Mentor Simulator Features Imaging C-arm and patient table maneuvering Real time fluoroscopy, cineangiography, DSA and roadmapping Real time Echocardiography (ICE) and TEE Device Simulation A wide

More information

COMPREHENSIVE FAMILY OF INTERVENTIONAL SYSTEMS

COMPREHENSIVE FAMILY OF INTERVENTIONAL SYSTEMS COMPREHENSIVE FAMILY OF INTERVENTIONAL SYSTEMS When excellence means saving lives, installing the right lab is absolutely vital. Unlike other imaging systems, the cardiovascular lab is literally an extension

More information

extended sewing ring Carpentier-Edwards bioprosthetic valved conduit , 21, 22

extended sewing ring Carpentier-Edwards bioprosthetic valved conduit , 21, 22 Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences (formerly

More information

The Latest in Radiation Dose Reduction Techniques in CT

The Latest in Radiation Dose Reduction Techniques in CT The Latest in Radiation Dose Reduction Techniques in CT Cynthia H. McCollough, PhD, FAAPM, FACR Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo Clinic,

More information

Detection of endograft fractures with multidetector row computed tomography

Detection of endograft fractures with multidetector row computed tomography Detection of endograft fractures with multidetector row computed tomography Justus E. Roos, MD, Jeffrey C. Hellinger, MD, Richard Hallet, MD, Dominik Fleischmann, MD, Christopher K. Zarins, MD, and Geoffrey

More information

3D morphometry using automated aortic segmentation in native MR angiography: an alternative to contrast enhanced MRA?

3D morphometry using automated aortic segmentation in native MR angiography: an alternative to contrast enhanced MRA? Original Article 3D morphometry using automated aortic segmentation in native MR angiography: an alternative to contrast enhanced MRA? Matthias Müller-Eschner 1,2, Tobias Müller 1, Andreas Biesdorf 3,

More information