VIBRATION AND MECHANICAL SHOCK TEST REPORT
|
|
- Sophia Cox
- 5 years ago
- Views:
Transcription
1 Report No. :DH A-1 Version:A VIBRATION AND MECHANICAL SHOCK TEST REPORT Company :OTO Photonics Inc. Address :9F-5,No. 27, Guanxin Rd., East Dist., Hsinchu City 30072, Taiwan(R.O.C.) Sample Name :Spectrometer Date Received :JUN 07, 2017 Date Tested :JUN 07, 2017 ~ JUN 13, 2017 TESTING LABORATORY IS ACCREDITED BY: IEC/IECQ certificate of independent test laboratory approval Certificate No.: ISO accredited in respect of laboratory is approved by TAF Certificate No.:L0835 ISO 9001 certificate is approved by TUV CERT certification body of TUV NORD Cert GmbH WE HEREBY CERTIFY THAT: The test(s) shown in the attachment were conducted according to the indicating procedures. We assume full responsibility for the accuracy and completeness of these tests and vouch for the qualifications of all personnel performing them. Name Signature Date Test Engineer Marco Yu Jun 19, 2017 Manager Allan Tseng Jun 19, 2017 NOTE: 1. This report will be invalid if reproduced in part or altered in any way. 2. This report refers only to the specimen(s) submitted to test, and is invalid if used otherwise. 3. This report is ONLY valid with the examination seal and signature of this institute. 4. The tested specimen(s) will only be preserved for thirty days from the date issued, if not collected by the applicant.
2 Reliability Testing Laboratory Page 1 of 26 TABLE OF CONTENTS 1. GENERAL INFORMATION DESCRIPTION OF UNIT UNIT OPERATING CONDITION VIBRATION TEST TEST EQUIPMENT LABORATORY AMBIENCE CONDITION REFERENCE DOCUMENT TEST CONDITION SUMMARY OF TEST PHOTO OF TEST GRAPH OF TEST MECHANICAL SHOCK TEST TEST EQUIPMENT LABORATORY AMBIENCE CONDITION REFERENCE DOCUMENT TEST CONDITION SUMMARY OF TEST PHOTO OF TEST GRAPH OF TEST... 24
3 Reliability Testing Laboratory Page 2 of GENERAL INFORMATION 1.1 DESCRIPTION OF UNIT Manufacturer:OTO Photonics Inc. Sample name:spectrometer Sample quantity:6 units 1.2 UNIT OPERATING CONDITION None
4 Reliability Testing Laboratory Page 3 of VIBRATION TEST 2.1 TEST EQUIPMENT Model Serial Number Calibration Date King Design EM-600F2K-40N120 UW MAY 12, LABORATORY AMBIENCE CONDITION Temperature:25 C ± 10 C Relative humidity:50 % ± 25 % (RH) 2.3 REFERENCE DOCUMENT The test is based on customer specification 2.4 TEST CONDITION Units are non-operating. Vibration waveform:sine waveform Vibration frequency:10 to 200 Hz Vibration acceleration:1.5 g Cycle time:30 minutes Number of cycle:1 cycle for each axis Vibration axes:x, Y and Z 2.5 SUMMARY OF TEST 2.5.1, visual inspection showed no physical defect on units , inform OTO Photonics Inc. to take back the units; and the function inspection of all samples was performed by OTO Photonics Inc. (Without 3 rd party lab Dekra-iST person participated) To follow OTO Photonics Inc. requirement, attached the result as below table, this result was provided from OTO Photonics Inc. Sample name Spectrometer Test result, units function was normal.
5 Reliability Testing Laboratory Page 4 of PHOTO OF TEST X axis Y axis Z axis
6 Reliability Testing Laboratory Page 5 of 26 Appearance was normal
7 Reliability Testing Laboratory Page 6 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
8 Reliability Testing Laboratory Page 7 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
9 Reliability Testing Laboratory Page 8 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
10 Reliability Testing Laboratory Page 9 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
11 Reliability Testing Laboratory Page 10 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
12 Reliability Testing Laboratory Page 11 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
13 Reliability Testing Laboratory Page 12 of GRAPH OF TEST X axis
14 Reliability Testing Laboratory Page 13 of 26 Y axis
15 Reliability Testing Laboratory Page 14 of 26 Z axis
16 Reliability Testing Laboratory Page 15 of MECHANICAL SHOCK TEST 3.1 TEST EQUIPMENT Model Serial Number Calibration Date LAB ASII MAY 19, LABORATORY AMBIENCE CONDITION Temperature:25 C ± 10 C Relative humidity:50 % ± 25 % (RH) 3.3 REFERENCE DOCUMENT The test is based on customer specification 3.4 TEST CONDITION Units are non-operating. Pulse shape:half-sine waveform Impact acceleration:50 g Pulse duration:12 ms Number of shocks:6 shocks (1 shock for each face) Orientation:Bottom, top, left, right, front and rear faces 3.5 SUMMARY OF TEST 3.5.1, visual inspection showed no physical defect on units , inform OTO Photonics Inc. to take back the units; and the function inspection of all samples was performed by OTO Photonics Inc. (Without 3 rd party lab Dekra-iST person participated) To follow OTO Photonics Inc. requirement, attached the result as below table, this result was provided from OTO Photonics Inc. Sample name Spectrometer Test result, units function was normal.
17 Reliability Testing Laboratory Page 16 of PHOTO OF TEST Bottom Top Left Right Front Rear
18 Reliability Testing Laboratory Page 17 of 26 Appearance was normal
19 Reliability Testing Laboratory Page 18 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
20 Reliability Testing Laboratory Page 19 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
21 Reliability Testing Laboratory Page 20 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
22 Reliability Testing Laboratory Page 21 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
23 Reliability Testing Laboratory Page 22 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
24 Reliability Testing Laboratory Page 23 of 26 Function was normal Note:This photo was provided by OTO Photonics Inc.
25 Reliability Testing Laboratory Page 24 of GRAPH OF TEST Bottom face Top face
26 Reliability Testing Laboratory Page 25 of 26 Left face Right face
27 Reliability Testing Laboratory Page 26 of 26 Front face Rear face << The Following Blank >>
TESTING / INSPECTION
Reliability & Communication Testing Instruments KING DESIGN INDUSTRIAL CO., LTD. VIBRATION TEST LABORATORY 4F,NO. 3, Lane 270, Pei Shen Road Sec. 3, http: //www.kdi.tw Shen Keng Dist., New Taipei City,
More informationLABORATORY TEST REPORT
Report : 12144 BAE Systems (Operations) Limited The Faraday Test Centre Marconi Way Rochester Kent ME1 2XX Telephone: 01634 203541 Facsimile: 01634 203647 TOTAL NUMBER OF PAGES:- 6 LABORATORY TEST REPORT
More informationVibration & Shock Test Report
Nebraska Center for Excellence in Electronics 4740 Discovery Drive Lincoln, NE 68521-5376 Phone: 402.472.5880 Fax: 402.472.5881 Vibration & Shock Test Report Company: Contact: Crystal Group 850 Kacena
More informationQA Test Report VITA-350. (Product Reliability Test) Charles Chang Manager of QA Department. Knight Hu Test Engineer. Jeff Yang Approval
ISO-9001 Certified ISO-14000 Certified ISO-17025 Certified ADVANTECH CO., LTD. QA Test Report VITA-350 (Product Reliability Test) Report No : 07S092A1 Report Date : January 10, 2008 Issue Stamp Charles
More informationEnvironmental Testing
Nebraska Center for Excellence in Electronics 4740 Discovery Drive Lincoln, NE 68521-5376 Phone: 402-323-NCEE (6233) Fax: 402-323-6238 Environmental Testing Company: Crystal Group 850 Kacena Road Hiawatha,
More informationEnvironmental Testing
Nebraska Center for Excellence in Electronics 4740 Discovery Drive Lincoln, NE 68521-5376 Phone: 402-323-NCEE (6233) Fax: 402-323-6238 Environmental Testing Company: Crystal Group 850 Kacena Road Hiawatha,
More informationEnvironmental Testing
Nebraska Center for Excellence in Electronics 4740 Discovery Drive Lincoln, NE 68521-5376 Phone: 402-323-NCEE (6233) Fax: 402-323-6238 Environmental Testing Company: Crystal Group 850 Kacena Road Hiawatha,
More informationISTA 3A PACKAGE PERFORMANCE TEST
Note: This report is issued subject to the Testing and Certification Regulations of the TÜV SÜD Group and the General Terms and Conditions of Business of TÜV SÜD PSB Pte Ltd. In addition, this report is
More informationAUGUST 14, 2006 TEST REPORT # RANDOM VIBRATION TEST SERIES PROBE SERIES PROBE TEST SOCKET INTERCONNECT DEVICES, INC.
AUGUST 14, 2006 TEST REPORT #206271 RANDOM VIBRATION TEST 101377-000 SERIES PROBE 204130-036 SERIES PROBE TEST SOCKET INTERCONNECT DEVICES, INC. APPROVED BY: MAX PEEL SENIOR FELLOW CONTECH RESEARCH, INC.
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Wichita, KS Sound Transmission Loss CONDUCTED: 2017-03-20 (Report Revised 2017-05-11) Page 1 of 10 ON: Composite Wall Section with PCS-1B Silicone Compression Seal System - Double Installation in
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Wichita, KS Sound Transmission Loss CONDUCTED: 2017-03-14 (Report Revised 2017-05-11) Page 1 of 10 ON: Composite Wall Section with PCS-1C Silicone Compression Seal System - Double Installation in
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Liberty, MO. Sound Transmission Loss CONDUCTED: 2016-02-09 Page 1 of 8 ON: 3/4" Insulated Glazed Unit TEST METHOD Riverbank Acoustical Laboratories is accredited by the U.S. Department of Commerce,
More informationA L I O N Technical Center. Test Report
FOR: Spartanburg, SC Sound Absorption CONDUCTED: 2018-09-14 Page 1 of 8 ON: R_1 76778 D01-0445B EFSL61W Fabric over 2 in. fiberglass insulation TEST METHOD Riverbank Acoustical Laboratories is accredited
More informationA L I O N Technical Center. Test Report
FOR: Spartanburg, SC Sound Absorption CONDUCTED: 2018-09-14 Page 1 of 9 ON: R_1 76437 11634-12A EFFL61W Fabric over 2 in. fiberglass insulation TEST METHOD Riverbank Acoustical Laboratories is accredited
More informationJune 14, TEST REPORT # Rev ºC MECHANICAL SHOCK THERMAL SHOCK. CGAT (7 FIN 2.0 Dia/.892 HT)
June 14, 2004 TEST REPORT # 204158 Rev 1.1 VIBRATION @ 125ºC MECHANICAL SHOCK THERMAL SHOCK CGAT29-0008 (7 FIN 2.0 Dia/.892 HT) CGAT29-0019 (5 Fin 2.0 Dia/.678 HT) CALGREG APPROVED BY: George G. Olear
More informationA L I O N Technical Center. Test Report
FOR: Dublin, OH Sound Transmission Loss CONDUCTED: 2018-03-30 Page 1 of 8 ON: KNM-100B Limp Mass Vinyl TEST METHOD Riverbank Acoustical Laboratories is accredited by the U.S. Department of Commerce, National
More informationAmphenol Amphenol Taiwan Corporation Sheet 1 of 7
Amphenol Amphenol Taiwan Corporation Sheet 1 of 7 Title: Part Number: Description: Micro SD Card Connectors Product Specification GTFP08 SERIES Micro SD Card Connectors push-push type Revisions Control
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Liberty, MO. Sound Transmission Loss CONDUCTED: 2016-02-09 Page 1 of 8 ON: 3/4" Insulated Glazed Unit with Absorbar Impact Absorption Muntins + Absorbud Impact Absorption Intersects TEST METHOD Riverbank
More informationA L I O N Technical Center. Test Report
FOR: Dublin, OH Sound Transmission Loss CONDUCTED: 2018-03-30 Page 1 of 9 ON: KNM-100AL Limp Mass Vinyl with Reinforced Aluminum Foil Face TEST METHOD Riverbank Acoustical Laboratories is accredited by
More informationTEST REPORT. B&W International GmbH Junkendiek Ibbenbühren
1 Cover Sheet TEST REPORT Project-No.: 17-8981 Test Specimen: Shipping Unit - - Client: B&W International GmbH Junkendiek 5 49479 Ibbenbühren Responsible Persons: Waldemar Masur (B&W International GmbH)
More informationPCIEC/PCIE PCIEC EC-EC-AP/PCIE F-D-RA
Project Number: Design Qualification Test Report Tracking Code: 346583 _Report_Rev_2 Requested by: Liam Parkes Date: 1/15/2015 Tech: Qty to test: 14 Test Start: 10/31/2014 Test Completed: 12/9/2014 (Actual
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Cambridge, Ontario. Sound Transmission Loss CONDUCTED: 2015-04-15 Page 1 of 8 ON: dbnr Nois-eNvelope Panel A4 (Perforations Face Source) TEST METHOD Riverbank Acoustical Laboratories is accredited
More informationDESIGN QUALIFICATION TEST REPORT ECUE/UEC5 ECUE C1-FF-01-1/UEC H-D-RA-1-A
Project Number: Design Qualification Test Report Requested by: Liam Parkes Date: 8/25/2016 Tech: Aaron McKim Test Start: 12/10/2015 Test Completed: 12/25/2015 DESIGN QUALIFICATION TEST REPORT ECUE/UEC5
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Elgin, IL Sound Absorption CONDUCTED: 2016-05-31 Page 1 of 7 ON: Rigid PET felt board TEST METHOD Riverbank Acoustical Laboratories is accredited by the U.S. Department of Commerce, National Institute
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: St. Louis, MO. Sound Absorption CONDUCTED: 2015-04-08 Page 1 of 7 ON: Acoustical Wrapped Panel(AWP) - 2 Inch Truetone with Polyester Svelte Face TEST METHOD Riverbank Acoustical Laboratories is accredited
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Beachwood, OH Sound Absorption CONDUCTED: 2016-06-29 Page 1 of 7 ON: eco-c-tex (1" nominal thickness) TEST METHOD Riverbank Acoustical Laboratories is accredited by the U.S. Department of Commerce,
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Margate, FL. Sound Absorption CONDUCTED: 2015-09-22 Page 1 of 8 ON: Micro-Perforated Acoustical Aluminum Ceiling Tile with SoundTex - #MPA-1 Over 3.5" R-13 Fiberglass TEST METHOD Riverbank Acoustical
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: St. Louis, MO. Sound Absorption CONDUCTED: 2015-02-19 Page 1 of 7 ON: 1" Truetone Acoustical Wrapped Panels (On 1/8" Spacers) TEST METHOD Riverbank Acoustical Laboratories is accredited by the U.S.
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Margate, FL. Sound Absorption CONDUCTED: 2015-09-15 Page 1 of 7 ON: Micro-Perforated Acoustical Aluminum Ceiling Tile with SoundTex - #MPA-1 TEST METHOD Riverbank Acoustical Laboratories is accredited
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: AcoustiMac, LLC Tampa, FL. Sound Transmission Loss CONDUCTED: 2015-09-14 Page 1 of 8 ON: Sound Lock (Window+Door Sound Proofing Panel) Fabric Wrapped Panel - Wood Backing - Wood Frame TEST METHOD
More informationA L I O N Technical Center. Test Report
FOR: Lebanon, PA Sound Transmission Loss CONDUCTED: 2017-07-12 Page 1 of 9 ON: Quarry Floor Tiles and Regupol SonusPlus (underlayment) over 8 inch Concrete Slab (no Ceiling) TEST METHOD Riverbank Acoustical
More informationA2LA Promotion of Accreditation Package
A2LA Promotion of Accreditation Package Congratulations on achieving A2LA accreditation! Your organization joins thousands of like-minded, A2LA-accredited companies that understand the value of providing
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that B83 Testing & Engineering, Inc. 3040 West
More informationPurchasing Department Finance Group INVITATION TO BID
Purchasing Department Finance Group March 6, 2014 INVITATION TO BID The City of Norwalk, Norwalk Public Schools, is soliciting bids for the supply of twenty-three (23) Medtronic Lifepak CR Plus devices.
More informationQP Supplier Quality Requirements Manual. Purchasing. Dearborn Inc. Revision: E Page 1 of PURPOSE
Page 1 of 8 1.0 PURPOSE 1.1 The purpose of this manual is to have a single document which conveys the minimum quality requirements and manufacturing practices expected of suppliers for value added products
More informationA L I O N Technical Center. Test Report. RAL -A CONDUCTED: Page 1 of 9 ON: BioMontage Panels, Sustainable Poplar Bark
CLIENT: Sound Absorption Waterbury, CT CONDUCTED: 2019-02-27 Page 1 of 9 ON: BioMontage Panels, Sustainable Poplar Bark TEST METHODOLOGY Riverbank Acoustical Laboratories is accredited by the U.S. Department
More informationEquipment In-house Calibration Requirements and use of Non-Accredited Calibration Service Providers
Issue 5.0: Provided additional clarification on record retention and non-conforming work with In- House calibration. Issue 6.0: Updated UL internal approvers only. No other changes. For Client Labs Purpose
More informationISO General Requirements for Competence of Testing Laboratories Procedure
Competence of Testing Laboratories Page: 1 of 24 ISO 17025 General Requirements for Competence of Testing Laboratories Procedure Competence of Testing Laboratories Page: 2 of 24 Table of Contents Para.
More informationProcedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
More informationA L I O N Technical Center. Test Report
FOR: Beachwood, OH Impact Sound Transmission CONDUCTED: 2017-11-15 Page 1 of 8 ON: LVT flooring over Peacemaker Soundproofing Underlayment over 8" Concrete Slab No Ceiling TEST METHOD Riverbank Acoustical
More informationGMP, GLP OR ISO 17025: HOW DO THESE APPLY TO OUTSOURCED ANALYTICAL TESTING?
GMP, GLP OR ISO 17025: HOW DO THESE APPLY TO OUTSOURCED ANALYTICAL TESTING? Joe Grappin, MS, EAG Laboratories 2018 MD&M West Tech Theater 1 OVERVIEW Background: What Led to the Presentation History of
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 ELEMENT MATERIALS TECHNOLOGY DENVER 1530 Vista View Drive Longmont, CO 80504 Sandra Frank Phone 561 529 1488 MECHANICAL Valid To: July 31, 2018 Certificate
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 NATIONAL TECHNICAL SYSTEMS Massachusetts Division 1146 Massachusetts Avenue Boxborough, MA 01719 David Potpinka Phone: 732 936 0800 MECHANICAL Valid to: September
More informationMaster Air Parts, Inc Skyway Drive, Suite 303 Santa Maria, CA (805) Phone (805) Fax
(Please type or print) Supplier: Address: City: State: Zip: Telephone: Fax: Email: Website: Brief list of Parts, material, or services to be provided to Master Air Parts, Inc. Are you an FAA Certified
More information1 General Information Supplier:
1 General Information Supplier: Company Address Street Postal code and City Country Phone Fax Contact Person Function Department E-Mail Quality Management Quality Manager Phone Fax E-Mail The information
More informationAmerican Association for Laboratory Accreditation
Page 1 of 18 The following pages present the requirements from R104 General s: Accreditation of Field Testing and Field Calibration Laboratories in a checklist format. The policies, procedures and activities
More informationMAY 22, 2006 TEST REPORT # RANDOM VIBRATION TEST SERIES PROBE TEST SOCKET INTERCONNECT DEVICES, INC.
MAY 22, 2006 TEST REPORT #206320 RANDOM VIBRATION TEST 100045-042 SERIES PROBE TEST SOKET INTERONNET DEVIES, IN. APPROVED BY: MAX PEEL SENIOR FELLOW ONTEH RESEARH, IN. ontech Research REVISION HISTORY
More informationCONFIRMATION. In accordance with TOV AUSTRIA AUTOMOTIVE procedures, is hereby certified that
CONFIRMATION Management System with approval relevant requirements In accordance with TOV AUTOMOTIVE procedures, is hereby certified that Aluminum Precision Products, Inc. 3333 W. Warner Avenue Santa Ana,
More informationIPD Quality Management System Supplier Questionnaire Instructions
IPD Quality Management System Supplier Questionnaire Instructions 1. The following Supplier Questionnaire was developed by Industrial Parts Depot (IPD) to assess and document the capability of its supplier
More informationCertificate of Calibration
Page 1 of 9 Accurate Calibration & Repair 1924 Pinnacle Drive Aurora, IL 60504 Instrument Id: IT1609000001 Noun: Handheld Calibrator Mfr: ERO Electronic Accuracy: See manufacturer's specifications Reason
More informationA L I O N Technical Center. Test Report
FOR: Lebanon, PA Impact Sound Transmission CONDUCTED: 2017-07-12 Page 1 of 8 ON: Quarry Floor Tiles and Regupol SonusPlus (underlayment) over 8 inch Concrete Slab (no Ceiling) TEST METHOD Riverbank Acoustical
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Denver, Colorado Impact Sound Transmission RAL-IN16-035 CONDUCTED: 2016-11-01 Page 1 of 7 ON: Porcelain Tile flooring - Thinset Mortar - Whisper Mat underlayment - 8 inch Concrete Slabs - No Ceiling
More informationVendor Qualification Survey
1200 West 96 th St Minneapolis, MN 55431 Ph: 952-888-7900 Fax: 952-888-2719 Vendor Qualification Survey Vendor Information Company Name: Date: Address: City: Phone Number: email address: Product or Service
More informationDEKRA Rail Company brochure Committed to Safety & improving Performance
DEKRA Rail Company brochure Committed to Safety & improving Performance DEKRA Rail in brief Questions we like to answer Safe assets with the required performance Do you want to use and maintain your rail
More informationOCT 16 Rev A
Product Specification 108-151041 10 OCT 16 Rev A Jiffy Splice System 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the TE Connectivity (TE) Jiffy Splice
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: AcoustiMac, LLC Tampa, FL. Sound Absorption CONDUCTED: 2015-09-14 Page 1 of 9 ON: Sound Lock/Space Divider/Partition - Wood Framed Sound Proofing Panel - Fabric Both Sides TEST METHOD Riverbank Acoustical
More informationA L I O N Technical Center. Test Report. RAL -A CONDUCTED: Page 1 of 9 ON: BioMontage Panels, Pole Moss
CLIENT: Sound Absorption Waterbury, CT CONDUCTED: 2019-02-28 Page 1 of 9 ON: BioMontage Panels, Pole Moss TEST METHODOLOGY Riverbank Acoustical Laboratories is accredited by the U.S. Department of Commerce,
More informationLaboratory Airborne Sound Insulation of Stud Walls with Warmcel 500 Insulation. Prepared for:
Laboratory Airborne Sound Insulation of Stud Walls with Warmcel 500 Insulation Prepared for: Excel Industries Limited Maerdy Industrial Estate (South) Rhymney Gwent NP22 5PY 18 th June 2010 0578 Tested
More informationEnvironmental Testing
Nebraska Center for Excellence in Electronics 4740 Discovery Drive Lincoln, NE 68521-5376 Phone: 402-323-NCEE (6233) Fax: 402-323-6238 Environmental Testing Company: Crystal Group 850 Kacena Road Hiawatha,
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Atlanta, GA Sound Absorption CONDUCTED: 2016-09-09 Page 1 of 7 ON: D-GIK (24"x48" panels) TEST METHOD Riverbank Acoustical Laboratories is accredited by the U.S. Department of Commerce, National Institute
More informationA L I O N Technical Center. Test Report. RAL -A CONDUCTED: Page 1 of 9 ON: BioMontage Panels - Reindeer Moss
CLIENT: Sound Absorption Waterbury, CT CONDUCTED: 2019-02-25 Page 1 of 9 ON: BioMontage Panels - Reindeer Moss TEST METHODOLOGY Riverbank Acoustical Laboratories is accredited by the U.S. Department of
More informationAnnex to the Accreditation Certificate D-ZE according to DIN EN ISO/IEC 17065:2013
Deutsche Akkreditierungsstelle GmbH Annex to the Accreditation Certificate D-ZE-12007-01-11 according to DIN EN ISO/IEC 17065:2013 Holder of certificate: TÜV NORD CERT GmbH Langemarckstraße 20, 45141 Essen
More informationUGANDA NATIONAL BUREAU OF STANDARDS CERTIFICATION SCHEME
1. NAME AND ADDRESS OF APPLICANT Company Name Company Tax Identification Number (TIN) A. APPLICANT S INFORMATION Business Registration Please attach certificate of registration of business COMPANY ADDRESS
More informationASM AEROSPACE SPECIFICATION METALS, INC NW 34 th Place #B28 Terrace Pompano Beach, Florida Tel: Fax:
SUBCONTRACTOR QUALITY SYSTEM QUESTIONNAIRE General: Company Name: Address: City: State: Zip: Country: Telephone: Fax: Business Information: Primary Products/Services: Type of Business: Manufacturing Distributor
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. CONTECH RESEARCH INC. 67 Mechanic Street Attleboro, MA Mark Gates MECHANICAL
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 CONTECH RESEARCH INC. 67 Mechanic Street Attleboro, MA 02703 Mark Gates 508 226 4800 MECHANICAL Valid to: February 29, 2016 Certificate Number: 1478.01 In recognition
More informationMIL-STD 461F Test Report
MIL-STD 461F Prepared for: Address: EUT: Crystal Group 850 Kacena Road Hiawatha, IA 52233 Model RCS6450-24 Rugged Switch No.: R20160707-20 Approved By: Nic S. Johnson, NCE Technical Manager inarte Certified
More informationUNITED NATIONS / DOT PERFORMANCE CERTIFICATION 4G DESIGN QUALIFICATION
UNITED NATIONS / DOT PERFORMANCE CERTIFICATION 4G DESIGN QUALIFICATION 6 x 2.5 Liter Plastic Bottle Packaging (PT 04003) with Two Case Sealing Mechanisms: #1) Taped Top and Bottom Flaps #2) Taped Top and
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. CONTECH RESEARCH INC. 750 Narragansett Park Drive Rumford, RI Thomas Lamoureux
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 CONTECH RESEARCH INC. 750 Narragansett Park Drive Rumford, RI 02916-1035 Thomas Lamoureux 401-865-6440 MECHANICAL Valid to: February 28, 2018 Certificate Number:
More informationAmphenol Amphenol Taiwan Corporation Sheet 1 of 6
Amphenol Amphenol Taiwan Corporation Sheet 1 of 6 Title: Part Number: Description: Mini SAS HD Integrated Connector Product Specification G40H series Mini SAS HD Integrated Connector, 0.75 Pitch, R/A,
More informationRIVERBANK ACOUSTICAL LABORATORIES
FOR: Indianapolis, IN. Sound Absorption CONDUCTED: 2015-11-05 Page 1 of 7 ON: Studio 6 Foam Panel TEST METHOD Riverbank Acoustical Laboratories is accredited by the U.S. Department of Commerce, National
More informationPhysikalisch-Technische Bundesanstalt REPORT. No on the. Type Examination of a. Non-Automatic Weighing Instrument
Physikalisch-Technische Bundesanstalt Braunschweig und Berlin REPORT No 1.12-4057916 on the Type Examination of a Non-Automatic Weighing Instrument Type: CHS01 E I; M Manufacturer: seca gmbh & co. kg.
More informationMar11 Rev C
Product Specification 108-1163-4 11Mar11 Rev C Eight Position Right Angle Inverted Multiple Port Modular Jack With & Without LED s 1. SCOPE 1.1. Content This specification covers the performance, tests
More informationSeismic Qualification Test Protocols, Standards and Methodology
Task NA 3.6 Preparatory Course on Seismic Qualification Bristol, 11-12 January 2011 Seismic Qualification Test Protocols, Standards and Methodology Chris Stone, TRaC Global Limited TRaC Global Limited
More informationYOUNG & FRANKLIN / TACTAIR FLUID CONTROLS QUALITY ASSURANCE SURVEY. Company Name: Address: City: State: Zip: Phone: Fax: Response Prepared By:
YOUNG & FRANKLIN / TACTAIR FLUID CONTROLS QUALITY ASSURANCE SURVEY Company Name: Date: Address: City: State: Zip: Phone: Fax: e-mail: Response Prepared By: Title: 1.0 Is your company ISO, QS, AS, or NADCAP
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 JSC RECEIVING INSPECTION AND TEST FACILITY Building 15, M/S NT315 Houston, TX 77058 Deborah Applegate Phone: 281 483 0288 ELECTRICAL Valid Until: December 31,
More informationTECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009
SANAS Accr. No/s. TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009 Date/s of evaluation Assessor/s & Observers Facility Area / field of Representative operation This report covers the following:
More informationSEAC DISCRETE TWINAX CABLE TYPE QUALIFICATION TEST REPORT SEAC/SEAF SEAC TU-TU-2/SEAF L-06-2-RA-LP-TR
Project Number: Design Qualification Test Report Tracking Code: 346584_Report_Rev_1 Requested by: Liam Parkes Date: 1/29/2015 Tech: Aaron Mckim Test Start: 8/15/2014 Test Completed: 1/19/2015 (Actual part
More informationCO-ORDINATION OF NOTIFIED BODIES PPE Regulation 2016/425 RECOMMENDATION FOR USE
CO-ORDINATION OF NOTIFIED BODIES PPE Regulation 2016/425 PPE-R/00.018 Version 1 RECOMMENDATION FOR USE Number of pages: 5 Approval stage : Approved on : Origin : Horizontal Committee, C2D Ad hoc group
More informationTransition Strategy for VDA 6.1, VDA 6.2 and VDA 6.4. Revision January Contents. 1. Foreword
Transition Strategy for.1,.2 and.4 Revision 1.0 --- 06. January 2017 Contents 1. Foreword 2. Transition timing requirements to the relevant.x standards 3. Transition audit requirements to the relevant.x
More informationValidation and Verification: VCS
Validation and Verification: VCS Mr N.VASASMITH GHG Assessor Scope 13.1 Meeting room 3, Zone B, Queen Sirikit National Convention Center 2012 January 24 Prepare by Ms S.THONGSONG 1 Introduction VCS is
More informationBecause People Rely on Your Products, Everyday, You can Rely on Us
Because People Rely on Your Products, Everyday, You can Rely on Us testing Services Environmental Testing Shock & Vibration Testing ACCREDITED 503.1 & 503.2 Ohio and Michigan Laboratories csz testing services
More informationSUPPLY AND INSTALLATION OF TURBINES AND GENERATORS CONTRACT SCHEDULE 8 QUALITY MANAGEMENT TABLE OF CONTENTS
SUPPLY AND INSTALLATION OF TURBINES AND GENERATORS CONTRACT SCHEDULE 8 QUALITY MANAGEMENT TABLE OF CONTENTS 1 INTERPRETATION... 1 1.1 Definitions... 1 2 QUALITY MANAGEMENT SYSTEM... 2 2.1 Quality Management
More informationSupplier Quality Survey. 1. Type of Business: g) Commodities supplied? Supplier Changes/comments: 2. Headcount breakdown by group: Purchasing
Supplier: Phone: Prime Contact/Title: Sales Contact/Title: Address: Fax: e-mail address e-mail address Quality Contact/Title: e-mail address 1. Type of Business: a) Number of years in business? b) Company
More informationETCI announces the following Amendment No. 2.3 to the National Rules for Electrical Installations ET 101:2008 incorporating Amendment No.
25 th February 2016 ETCI announces the following Amendment No. 2.3 to the National Rules for Electrical Installations ET 101:2008 incorporating Amendment No.1:2011 ETCI announces the following Amendment
More informationTECHNICAL REQUIREMENTS (CLAUSE 5)
LECTURE 4 THE ELEMENTS OF ISO/IEC 17025 - TECHNICAL REQUIREMENTS (CLAUSE 5) Slide 1 ISO 17025 - RELATIONSHIP OF EACH REQUIREMENTS Management System Organization Quality system Continual Improvement Management
More informationThis document is a preview generated by EVS
INTERNATIONAL STANDARD ISO 16831 First edition 2012-04-01 Non-destructive testing Ultrasonic testing Characterization and verification of ultrasonic thickness measuring equipment Essais non destructifs
More informationISO INTERNATIONAL STANDARD
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 16831 First edition 2012-04-01 Non-destructive testing Ultrasonic testing Characterization and verification of ultrasonic thickness measuring equipment
More informationMaharashtra State Electricity Distribution Co. Ltd.
Maharashtra State Electricity Distribution Co. Ltd. SPECIFICATION NO. MSEDCL/ DIST/MM-III/02/2008/R1(010111) SPECIFICATION FOR L.T. PIN/SHACKLE INSULATORS Page 1 of 12 MAHARASHTRA STATE ELECTRICITY DISTRIBUTION
More informationENERGY EFFICIENCY ENERGY STAR
ENERGY EFFICIENCY ENERGY STAR UL s Performance Test Capabilities Luminaire Testing Laboratory (Allentown, PA) 2 UL s Performance Test Capabilities Nansha Technology Testing Center - Nansha, China 3 UL
More informationMar11 Rev B
Product Specification 108-1163-7 11Mar11 Rev B Inverted Six and Eight Position Surface Mount Shielded Modular Jack Assembly 1. SCOPE 1.1. Content This specification covers performance, tests and quality
More informationEuropean Packaging Regulations What to expect from a notified body audit of your packaging
What to expect from a notified body audit of your packaging Dr. Ralf Walther TÜV Süd Product Service GmbH March 3-5, 2009 www.healthpack.net Content 1. Short introduction of TÜV Süd Product Service 2.
More informationL.L_.,.till' President and CEO For the Accreditation Council Certificate Number Valid to May 31,2020
Accredited Laboratory A2LA has accredited KEYSTONE COMPLIANCE New Castle, PA for technical competence in the field of Mechanical Testing This laboratory is accredited in accordance with the recognized
More informationAUGUST 18, 2008 TEST REPORT #208315A QUALIFICAITON TESTING PART NUMBERS ERF S-D-RA ERM S-DV-L SAMTEC, INC.
AUGUST 18, 2008 TEST REPORT #208315A QUALIFICAITON TESTING PART NUMBERS ERF8-050-01-S-D-RA ERM8-050-05.0-S-DV-L SAMTEC, INC. APPROVED BY: DOMINIC ARPINO PROGRAM MANAGER CONTECH RESEARCH, INC. Contech Research
More informationMar11 Rev C
Product Specification 108-1949 11Mar11 Rev C Small Form Factor Pluggable, 0.8 mm Centerline Connectors 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the
More informationTHE KENYA POWER AND LIGHTING CO. LTD
THE KENYA POWER AND LIGHTING CO. LTD SPECIFICATION for SINGLE PHASE STATIC METERS FOR ACTIVE ENERGY Compiled by NAME DESIGNATION SIGNATURE DATE Ephantus Nyagah Ag. Senior Engineer Checked by Approved by
More informationIt s Simple. Protect the integrity of your products and your brand during cold chain transportation.
99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com It s Simple. Protect the integrity of your products and your brand during cold
More informationMar11 Rev C
Product Specification 108-1579 11Mar11 Rev C Sealed Circular Plastic Connector With Removable Crimp Contacts 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements
More informationOAİB PRODUCT CONFORMANCE SPECIFICATIONS -PUMPS- e - m a i l : i n f t u r q u m. c o m
PRODUCT CONFORMANCE SPECIFICATIONS -PUMPS- M a h a t m a G a n d h i C a d. N o : 1 0 3 G. O. P. 0 6 7 0 0 A N K A R A P h o n e : 9 0-3 1 2-4 4 7 2 7 4 0 F a x : 9 0-3 1 2-4 4 7 0 1 8 0-4 4 6 9 6 0 5
More informationProcedure for Biomedical Equipment Calibration and Maintenance
Procedure for Biomedical Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of Biomedical
More informationSEMI Draft Document 5204 NEW STANDARD: Test Method for Mechanical Vibration of c-si PV Cells in Shipping Environment
SEMI Draft Document 5204 NEW STANDARD: Test Method for Mechanical Vibration of c-si PV Cells in Shipping Environment 1 Purpose 1.1 For c-si technology, one of the problems is to identify and eliminate
More information