DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion

Size: px
Start display at page:

Download "DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion"

Transcription

1

2

3

4 DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion: Once approved, the new or revised document or the notice of deleted document(s) should be distributed to all persons identified below: ( ) All RP staff members ( X ) All holders of the following manual(s): SLAC Radioanalysis Laboratory Procedures ( ) The following individuals: November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 3

5 TABLE OF REVISIONS REVISION DATE SECTION(S) REASON FOR REVISION # 001 November 27, 2006 All Provide steps for the calibration process November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 4

6 TABLE OF CONTENTS Liquid Scintillation Counter Calibration Procedure 1 PURPOSE REFERENCES RESPONSIBILITIES Rad Lab Operator REP Manager PREREQUISITES PROCEDURES Self Normalization and Calibration Check Accuracy check Quench Curve Verification RECORDS ATTACHMENTS/FORMS Attachment 1: 2005 Annual Calibration Results (LSC)... 9 November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 5

7 1 PURPOSE As part of SLAC Radioanalysis Laboratory s (Rad Lab) Quality Assurance Controls, periodic calibration checks should be performed in order to verify system performance. Calibrations should be performed after system repairs or maintenance services. An annual calibration check should be performed at a minimum. The purpose of this procedure is to provide the necessary steps to perform the annual calibration. 2 REFERENCES 2.1 Packard Tri-Carb Operational Manuals 2.2 SLAC Rad Lab Manual, LSC Counting Procedure RE#009 3 RESPONSIBILITIES 3.1 Rad Lab Operator The Rad Lab Operator assigned to perform activities described in this chapter is responsible for: ensuring that prerequisites are met as required, performing calibrations, reporting any problems or unexpected events to the Radiological Environmental Protection (REP) Manager, generating and providing calibration reports to the REP Manager, and Performing other duties assigned by the REP Manager. 3.2 REP Manager The REP Manager is responsible for: ensuring that the activities in this chapter are conducted in accordance with this written procedure, assigning the tasks in this procedure to a qualified Operator, and November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 6

8 providing final approval or disapproval of all calibration results/reports. 4 PREREQUISITES 4.1 Must be familiar with DOS and Windows software menus. 4.2 Must be familiar with Rad Lab general sample analysis procedures. 4.3 Must be familiar with Liquid Scintillation Counting theory 4.4 Must be familiar with laboratory equipment (pipettes, scales, etc.) 5 PROCEDURES 5.1 Self Normalization and Calibration Check The Packard Tri-Carb Liquid Scintillation Counter (LSC) utilizes a Self- Normalization and Calibration protocol (SNC). The protocol is run daily when the LSC is in use and should be verified during the annual calibration. See Packard Instruments LSC manuals for specific details on the SNC protocol Run the SNC according to the Source Check section of the LSC Sample Counting Procedure (SLAC-I-760-2A39C-008) Efficiencies for the LSC should be compared to the parameters for small vials. If any of the above parameters are not satisfied, notify the Radioactive Environmental Protection (REP) Manager. Otherwise no further action regarding the SNC protocol is required. 5.2 Accuracy check Machine performance can be verified by checking its accuracy. This is accomplished by using a solution with a known specific activity. A stock solution from a NIST traceable standard can be made. Additional replicates should be made in order to reduce statistical error Create stock solution from a NIST traceable standard solution. A prepared solution with known specific activity can also be used Pipette 5 ml of the stock solution into a LSC vial Add 15 ml of Liquid scintillation cocktail November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 7

9 5.2.4 Count the sample(s) using the LSC Counting procedure (SLAC-I-760-2A39C-008). Assure that the samples are counted under LCW and Environmental protocols Determine the accuracy of the machine by comparing the output with the known activity of the sample(s). Perform any error calculation and statistical analysis as needed. Notify the REP Manager if the difference between the known activity and the reported activity is greater than 20%. 5.3 Quench Curve Verification The current quench curves are based on quenched 3 H standards. Rerun a similar set in order to validate the current quench curves. If a quench standard set cannot be located, replicates from a stock solution with known specific activity can be substituted. The replicates will have the same activity, but will contain varying amounts of quenching agent Load a quenched 3 H set into the LSC and count using LCW 30 minute count (Protocol 1) Plot tsie vs. absolute efficiency. For efficiency calculations, be sure to decay correct when using quenched 3 H standards Print the current quench curves (see Tri Carb manual for details). Compare the current quench curves with the quenched 3 H set. If the trends of the curves do not appear to match, notify the REP Manager. Otherwise replace the old quench curves with the new one and complete the calibration report. 6 RECORDS 6.1 Calibration results generated from this procedure must be documented into a report and submitted for approval. A signed copy will be filed in the Rad Lab archives. A sample calibration report has been provided (see attachment #1) for reference. 7 ATTACHMENTS/FORMS 7.1 Attachment 1: 2005 Annual Calibration Results (LSC) November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 8

10 7.1 Attachment 1: 2005 Annual Calibration Results (LSC) Attached are the results of the annual calibration performed on the Packard Tri-Carb Liquid Scintillation Counter (LSC) by Henry Tran and Henry Brogonia. As per the SLAC Rad Lab LSC Calibration Procedures a Self Normalization and Calibration (SNC) check, an accuracy check, and validation of quench curves were performed to assure SLAC Rad Lab Quality Assurance controls. The results of the three checks are summarized below. The Self-Normalization and Calibration Check was performed as per the SLAC Rad Lab LSC Calibration procedure. The efficiencies for 3 H and 14 C are 59.10% and 95.10% respectively, which are within machine specifications (Appendix A). The accuracy check results of the LSC are shown in Table 1 (below). Two 500 ml bottles of stock solution were prepared by H. Tran from a 3 H source with a specific activity of 6.2 E+05 pci/l (Decay corrected, Appendix B). 5 ml of the standard was diluted with 200 ml of water, yielding a specific activity of pci/l per bottle. Samples were prepared and counted using the routine LSC counting procedure. 5 ml of the stock solution was added by pipette into a LSC vial. 15 ml of cocktail fluid was then added. Two replicates from bottle #1 and #2 were counted on protocol 1 (30 minute count) and 13 (150 minute count). Protocol #1 LCW - 30 min Protocol #13 Environmental min Sample DPM Sample DPM 1A 213 1A 210 1B 234 1B 228 2A 203 2A 215 2B 214 2B 218 Average: 216 Average: 218 Bkg: 25 Bkg: 24 Net: 190 Net: 193 Pci/L pci/l: STD: STD: Ratio 1.14 Ratio 1.17 Table 1 The results show that the percent difference between the known activity and the reported activity is 17%, which is within the 20% range for quality assurance controls. The results of the quench curve validation are shown below. Two sets of quenched 3 H standards (Appendix C) were counted on protocol 1 (30 min count). Each quench standard set consists of 10 glass vials with solutions of equal activity. The amount of quenching November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 9

11 agent is the only difference between the vials. The efficiency was calculated based on the reported activity and the known activity of the quenched standards (Appendix D). The tsie vs. efficiency curves were compared against the original stored quench curve. The new quench curves are relatively consistent with the stored quench curve on the LSC (Figure 1, below). The results are also consistent with the 2004 annual calibration. t-sie Quench Curve Efficiency t-sie H H-3 Stored Quench Curve Figure 1. Quench curve validation. Old Quench A represents the quench curve currently stored on the LSC machine. The remaining curves are two separate quench standards, designated by their reference date. The new quench curves are consistent with the stored curve. No adjustment to the LSC is needed at this time. Another test was performed to verify the LSC s response to a quenching agent. Because the quenched standards in the quench curve validation were contained in glass vials, a similar check was needed in order to verify the quench response in polyethylene vials used for Rad Lab LSC samples. Two polyethylene vials, 1B and 2B from the stock solution created during the accuracy check, were used for this quench response check. A quenching agent was added to the vials and subsequently counted to verify a response to quench. The process of adding a quenching agent and counting was repeated twice. The results are shown in the Table (below). As expected the quench indicating parameter (tsie) varied with the concentration of the quenching agent. November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 10

12 Quench ID CPM (in ROI) DPM Efficiency tsie 1B B B B B B B B Table 2 Quench response check for polyethylene vials used in Rad Lab analysis. The Quench ID corresponds to the sample vial and the amount of quench added to each vial (ul of red food coloring). The table shows the increase in quenching agent shows a lowering of tsie, the quench indicating parameter. The result is a lower efficiencies for lower tsie values, which is expected. ( ) It was determined that no adjustment to the LSC was needed. ( ) The following adjustments have been completed and the LSC is operating within satisfactory conditions: ( ) The LSC is not performing within satisfactory conditions. Will require service from vendor. Copy Only Calibration performed by Date Approved by Date Attached: Appendix A: SNC Protocol, SNC Calibration Report Appendix B: Certificate of Standard Appendix C: Certificate of Quenched Standards Appendix D: Table of Quench curve validation results November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 11

13 Appendix A November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 12

14 Appendix B November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 13

15 Appendix C November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 14

16 November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 15

17 Appendix D Quench ID CPM (in ROI) DPM Efficiency SIS tsie UG UG UG UG UG UG UG UG UG UG AX BX CX DX EX FX GX HX IX JX Data results from the quench curve validation. The quench ID corresponds to the reference date on the standard. For each set, vials are placed in order by increasing quench. Increasing the quench would correspond to a decreasing tsie, which is the quench indicating parameter used for the counting process. Efficiency vs. tsie is plotted and compared to original quench curve. As the trends for all curves appear to be similar, no adjustment in the LSC is required at this time. November 27, 2006 SLAC-I-760-2A39C-013-R-001 Page 16

Counting Aqueous Samples with the TopCount Microplate Scintillation Counter

Counting Aqueous Samples with the TopCount Microplate Scintillation Counter TCA-005 Counting Aqueous Samples with the TopCount Microplate Scintillation Counter Abstract The TopCount TM Microplate Scintillation Counter quantitates radioisotopes by liquid scintillation counting

More information

Use and Preparation of Quench Curves in Liquid Scintillation Counting

Use and Preparation of Quench Curves in Liquid Scintillation Counting APPLICATION NOTE Liquid Scintillation Authors: Jock Thomson PerkinElmer, Inc. Waltham, MA Use and Preparation of Quench Curves in Liquid Scintillation Counting Introduction The liquid scintillation process

More information

STUDY OF THE PERFORMANCE OF EFFICIENCY TRACING TECHNIQUE ON A TRICARB 2100TR LIQUID SCINTILLATION ANALYZER

STUDY OF THE PERFORMANCE OF EFFICIENCY TRACING TECHNIQUE ON A TRICARB 2100TR LIQUID SCINTILLATION ANALYZER NUCLEAR PHYSICS RADIATION PROTECTION STUDY OF THE PERFORMANCE OF EFFICIENCY TRACING TECHNIQUE ON A TRICARB 2100TR LIQUID SCINTILLATION ANALYZER R. I. DOBRIN, C. N. DULAMA Institute for Nuclear Research,

More information

Application Note Gross Alpha Counting Of Air Filters Using A Pulse-Shape Discriminating Alpha Liquid Scintillation Counter

Application Note Gross Alpha Counting Of Air Filters Using A Pulse-Shape Discriminating Alpha Liquid Scintillation Counter Application Note Alpha/Beta ABA-001 Gross Alpha Counting Of Air Filters Using A Pulse-Shape Discriminating Alpha Liquid Scintillation Counter by E. R. Hinton Jr., T. T. Adams, and L. A. Burchfield Plant

More information

Quench & Quench Curves. James Thomson Meridian Biotechnologies Ltd.

Quench & Quench Curves. James Thomson Meridian Biotechnologies Ltd. Quench & Quench Curves James Thomson Meridian Biotechnologies Ltd. What is Quench? Basically, the liquid scintillation process is the conversion of the energy of a radioactive decay event into photons

More information

Safe Operating Procedure

Safe Operating Procedure Safe Operating Procedure (Reviewed 8/14) CONTAMINATION SURVEYS FOR RADIOACTIVE MATERIAL LABORATORIES The purpose of this document is to provide guidance for the performance of contamination surveys in

More information

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: CHARCOAL LIQUID SCINTILLATION by Scintillation Counting. Method Number: ELAP method number 7033

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: CHARCOAL LIQUID SCINTILLATION by Scintillation Counting. Method Number: ELAP method number 7033 LAB ID and/or LABORATORY NAME: ASSESSOR NAME: DATE: CHARCOAL LIQUID SCINTILLATION by Scintillation Counting Method Number: ELAP method number 7033 SOP Number: Revision Number: SOP Date: Personnel / Data

More information

RADON MEASUREMENT OA - PRACTICE vs PROTOCOLS

RADON MEASUREMENT OA - PRACTICE vs PROTOCOLS RADON MEASUREMENT OA - PRACTICE vs PROTOCOLS Raymond Johnson, Sandy Colon, and Douglas Heim, Key Technology, Inc. Jonestown, PA ABSTRACT The EPA radon measurement device protocols of July 1992 specify

More information

OREGON HEALTH & SCIENCE UNIVERSITY RADIATION SAFETY OPERATING PROCEDURE 1101 LABORATORY RADIATION SURVEY

OREGON HEALTH & SCIENCE UNIVERSITY RADIATION SAFETY OPERATING PROCEDURE 1101 LABORATORY RADIATION SURVEY OREGON HEALTH & SCIENCE UNIVERSITY RADIATION SAFETY OPERATING PROCEDURE 1101 LABORATORY RADIATION SURVEY I. PURPOSE: This procedure gives the laboratory worker guidance in performing and documenting the

More information

How to overcome the challenges presented by bioassays

How to overcome the challenges presented by bioassays How to overcome the challenges presented by bioassays Tony Mire-Sluis Vice President, North America, Singapore, Abingdon, Contract and Product Quality Amgen Inc Why are Bioassays Considered so Difficult

More information

A comparison of two liquid scintillation instruments for analysis of highly quenched samples

A comparison of two liquid scintillation instruments for analysis of highly quenched samples A comparison of two liquid scintillation instruments for analysis of highly quenched samples Sarah Mullins Radiochemist, Radiation & Nuclear Sciences LSC217, Copenhagen,Denmark. 5 th May 217 Acknowledgements

More information

INSPECTION SYSTEM REQUIREMENTS

INSPECTION SYSTEM REQUIREMENTS MILITARY SPECIFICATIONS MIL-I-45208A INSPECTION SYSTEM REQUIREMENTS This specification has been approved by the Department of Defense and is mandatory for use by all Agencies. 1. SCOPE 1.1 Scope. This

More information

SURVEY GUIDE FOR UNSEALED RADIOACTIVE MATERIAL

SURVEY GUIDE FOR UNSEALED RADIOACTIVE MATERIAL SURVEY GUIDE FOR UNSEALED RADIOACTIVE MATERIAL Purpose The goal of any radiation safety program is to keep dose to individuals and contamination levels As Low As Reasonably Achievable (ALARA). The radiation

More information

SYSTEMATIC STUDY OF THE DIFFERENT QUENCHING PHENOMENA IN ORGANIC SCINTILLATORS

SYSTEMATIC STUDY OF THE DIFFERENT QUENCHING PHENOMENA IN ORGANIC SCINTILLATORS SYSTEMATIC STUDY OF THE DIFFERENT QUENCHING PHENOMENA IN ORGANIC SCINTILLATORS Luz Santiago, Hector Bagán, Alex Tarancón Sanz, Gemma Rauret, Jose Francisco García Departmento de Química Analitica de la

More information

Nuclear Engineering Seibersdorf GmbH

Nuclear Engineering Seibersdorf GmbH Nuclear Engineering Seibersdorf GmbH TDCR-LSC and TDCR-Cerenkov-Counting in a Radwaste Treatment Facility Andreas Vesely Mile Djuricic, Wolfgang Jaiczay 1 NES as Part of AIT (Austrian Institute of Technology)

More information

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups.

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups. Created Revised Reviewed Approved Lupus Insensitive aptt on the ACL TOP HEM 4.15.1 7/13/2012 4/11/2013 4/11/2013 4/11/2013 I. PRINCIPLE The activated partial thromboplastin time is a global screening procedure

More information

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis Certified Reference Material - Certificate of Analysis S(-)-Nicotine, Primary Standard N-008 Page 1 of 7 Catalog Number: N-008 H Lot: Expiration: October 2019 N Description: S(-)-Nicotine in Methanol.

More information

Biological Applications of Microplate Scintillation Counting using the TopCount

Biological Applications of Microplate Scintillation Counting using the TopCount Biological Applications of Microplate Scintillation Counting using the TopCount MICROPLATE SCINTILLATION & LUMINESCENCE COUNTER C A S E S T U D Y Abstract The microplate format has become increasingly

More information

BENTLEY BACTOCOUNT IBCm INDUSTRY OPERATOR (BCMlO) APPROVAL PROCEDURES PROTOCOL

BENTLEY BACTOCOUNT IBCm INDUSTRY OPERATOR (BCMlO) APPROVAL PROCEDURES PROTOCOL A. Training: BENTLEY BACTOCOUNT IBCm INDUSTRY OPERATOR (BCMlO) APPROVAL PROCEDURES PROTOCOL 1. BactoCount IBCm Industry Operators (BCMIO) are to receive one week of training conducted by a certified BactoCount

More information

CSHEMA Innovation Award Application 2017 Resource Enhancement: Reduced Costs and

CSHEMA Innovation Award Application 2017 Resource Enhancement: Reduced Costs and California State University, Fullerton CSHEMA Innovation Award Application 2017 Resource Enhancement: Reduced Costs and Student Success Preparing Students for Environmental Health and Safety Careers While

More information

Application Notes. MicroBeta INTRODUCTION.

Application Notes. MicroBeta INTRODUCTION. Drug Discovery Research Clinical Screening MicroBeta Application Notes Colour quench correction in Scintillation Proximity Assays Using ParaLux Count Mode K.T. Hughes 1, J.C. Ireson 1, N.R.A. Jones 2 and

More information

EASA Part 21 Compliance Matrix

EASA Part 21 Compliance Matrix Weston Aerospace CE8 Issue 2 EASA Part 21 Compliance Matrix Approved By: Peter Rush, Quality Director Date: 31 March 2011 This document is maintained on digital media. The current issue is located on the

More information

Ti-Pure TITANIUM DIOXIDE. CONCENTRATED ACID SOLUBILITY OF TiO 2 METHOD: T WP

Ti-Pure TITANIUM DIOXIDE. CONCENTRATED ACID SOLUBILITY OF TiO 2 METHOD: T WP Ti-Pure TITANIUM DIOXIDE CONCENTRATED ACID SOLUBILITY OF TiO 2 METHOD: T4400.315.07.WP 1 TITANIUM DIOXIDE (Ti-Pure ) Determination of Acid Solubility of TiO 2 by Spectrophotometer (TP-109.2) I. Principle

More information

A2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017

A2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017 Laboratory Page 1 of 17 Laboratory December 6, 2017 2017 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written permission of A2LA.

More information

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water NWTPH-Gx Volatile Petroleum Products Method for Soil and Water Summary The NWTPH-Gx method is intended to replace the Oregon s TPH-G and Washington s WTPH-G methods and to present a more comprehensive

More information

Quality Assurance Procurement Clauses and Flow Down Requirements

Quality Assurance Procurement Clauses and Flow Down Requirements Rev: U Page: 2 of 12 5.3 The Purchasing Department will insure all Purchase Orders include the following: 5.4 5.3.1 Vendor is on the Rogerson Kratos Approved Vendor List (AVL). Located on the server in

More information

2/07. PROCEDURE for DISPOSAL of RADIOACTIVE MATERIALS

2/07. PROCEDURE for DISPOSAL of RADIOACTIVE MATERIALS 2/07 CORNELL UNIVERSITY PROCEDURE for DISPOSAL of RADIOACTIVE MATERIALS This procedure has been developed to ensure the safety of those individuals who handle radioactive waste and to comply with University,

More information

Lead Sampling in School Facilities

Lead Sampling in School Facilities Lead Sampling in School Facilities Laboratory Roundtable October 2016 LEAD SAMPLING IN SCHOOL FACILITIES 1 HOW DOES LEAD GET INTO A SCHOOL S DRINKING WATER? SOURCE WATER RARELY FOUND IN SOURCE WATER IN

More information

Personal Temperature Drift Monitoring System For Thermal Cycling Instruments Product Info

Personal Temperature Drift Monitoring System For Thermal Cycling Instruments Product Info Personal Temperature Drift Monitoring System For Thermal Cycling Instruments Product Info Copyright 2002 GENO-tronics BV. All rights reserved. Page 1 of 8 INTRODUCTION This brochure provides general and

More information

PROCEEDINGS. Introduction. Olga Stawarz, Maria Karpińska, Kalina Mamont-Ciesla

PROCEEDINGS. Introduction. Olga Stawarz, Maria Karpińska, Kalina Mamont-Ciesla NUKLEONIKA 2006;51(Supplement 2):S53 S57 PROCEEDINGS Investigation of the influence of high humidity and exposure duration on the measurement results of radon concentration by means of PicoRad system in

More information

A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018

A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018 A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018 Laboratory Name: Appendix Name: Appendix Number: Assessor: Date: Rev. 4.1 CALA CHECKLIST FOR MICROBIOLOGY Item 01 SAMPLING AND SAMPLE

More information

FEC QUALITY MANUAL. Federal Equipment Company River Rd. Cincinnati, Ohio

FEC QUALITY MANUAL. Federal Equipment Company River Rd. Cincinnati, Ohio QMS-20 2016 FEC QUALITY MANUAL Federal Equipment Company 5298 River Rd Cincinnati, Ohio 45233 www.federalequipment.com www.fecheliports.com www.usdrillhead.com www.orionseals.com www.tkf.com REVISION:

More information

CHAPTER 5 INFORMATION TECHNOLOGY SERVICES CONTROLS

CHAPTER 5 INFORMATION TECHNOLOGY SERVICES CONTROLS 5-1 CHAPTER 5 INFORMATION TECHNOLOGY SERVICES CONTROLS INTRODUCTION In accordance with Statements on Auditing Standards Numbers 78 and 94, issued by the American Institute of Certified Public Accountants

More information

Overview of a sterility assurance program for PET drugs

Overview of a sterility assurance program for PET drugs Coalition for PET Drug Approval Radiopharmaceutical Sciences Council Overview of a sterility assurance program for PET drugs Eric Webster, PETNET Solutions Disclosures Employee of PETNET Solutions, a Siemens

More information

Good Pipetting Practice GPP Improve data consistency Risk based Life cycle oriented Application specific

Good Pipetting Practice GPP Improve data consistency Risk based Life cycle oriented Application specific Good Pipetting Practice GPP Improve data consistency Risk based Life cycle oriented Application specific Minimize Risk, Maximize Repeatability Improve the Quality of Your Data Good Pipetting Practice Minimize

More information

Tracking and Tracing in laboratories

Tracking and Tracing in laboratories Tracking and Tracing in laboratories General information History of KIESTRA Started in 1961 by Philippus Pobe Kiestra Roots: Lab Automation and Robotics (HP-ORCA robot system integrator) Since 1995 full

More information

MODERNIZATION PROJECT. December 2008 LC R0

MODERNIZATION PROJECT. December 2008 LC R0 RESEARCH SUPPORT BUILDING AND INFRASTRUCTURE MODERNIZATION PROJECT LIFE CYCLE COST ANALYSIS DEPARTMENT OF ENERGY SLAC SITE OFFICE December 2008 LC 522 000 01 R0 I ii Research Support Building and Infrastructure

More information

a. Are the foundry s employees trained and familiar with portions of the system applicable to their position?

a. Are the foundry s employees trained and familiar with portions of the system applicable to their position? NAV 29 FOUNDRY OPERATIONS 1. Does the foundry have a documented quality system? a. Are the foundry s employees trained and familiar with portions of the system applicable to their position? b. What recognized

More information

Installation Qualification/Operational Qualification Protocols and Instructions

Installation Qualification/Operational Qualification Protocols and Instructions Installation Qualification/Operational Qualification Protocols and Instructions AquaLab PRE Decagon Devices, Inc July 20, 2016 4:05 PM 1 Table of Contents Section 1 Section 2 Section 2.1 Section 2.2 Section

More information

MASTER PLANNING FOR VALIDATION

MASTER PLANNING FOR VALIDATION MASTER PLANNING FOR VALIDATION By: Gamal Amer, Ph.D. Principal, Premier Compliance Services, Inc. 1 Process Validation: General Principles and Practices Guidance to industry issued by the FDA in January

More information

COLORIMETRIC SANDWICH ELISA KIT INSTRUCTION MANUAL

COLORIMETRIC SANDWICH ELISA KIT INSTRUCTION MANUAL Page 1 of 7 COLORIMETRIC SANDWICH ELISA KIT INSTRUCTION MANUAL This product is for research use ONLY and not for human or animal therapeutic or diagnostic use. Page 2 of 7 Contents Page 3 I. Supplied Materials:

More information

Urban Forest Project Verification Protocol. Version 1.0

Urban Forest Project Verification Protocol. Version 1.0 Urban Forest Project Verification Protocol Version 1.0 August 2008 The TABLE OF CONTENTS I. Introduction...1 II. Standard of Verification...2 III. Core Verification Activities...2 Step 1: Identifying Emission

More information

HI EPA Compliant Turbidity Meter HI EPA Compliant Turbidity and Chlorine Meter

HI EPA Compliant Turbidity Meter HI EPA Compliant Turbidity and Chlorine Meter HI 98703 EPA Compliant Turbidity Meter HI 93414 EPA Compliant Turbidity and Chlorine Meter SPECIFICATIONS Turbidity (HI 93414 & HI 98703) Range Range Selection ACCESSORIES HI 710005 Voltage adapter HI

More information

ELECTRONIC REPAIR STATION MANUAL

ELECTRONIC REPAIR STATION MANUAL UNCONTROLLED WHEN COPIED OR PRINTED CONTROLLED FORMS LOCATED AT: http://www.aero-weld.com/forms.htm AEROWELD, INC 433 Breesport San Antonio, Texas 78216 ELECTRONIC REPAIR STATION MANUAL Including: QUALITY

More information

TECHNICAL INFORMATION

TECHNICAL INFORMATION T-1860A TECHNICAL INFORMATION Capillary Electrophoresis PRECISION IN CAPILLARY ELECTROPHORESIS Harry Whatley and Jeff Chapman Beckman Coulter, Inc. Introduction Instrumentation for routine chemical analysis

More information

QuantiFluor ONE dsdna System

QuantiFluor ONE dsdna System TECHNICAL MANUAL QuantiFluor ONE dsdna System Instructions for Use of Products E4871 and E4870 Revised 6/17 TM405 QuantiFluor ONE dsdna System All technical literature is available at: www.promega.com/protocols/

More information

Regulatory Overview Annex 11 and Part 11. Sion Wyn Conformity +[44] (0)

Regulatory Overview Annex 11 and Part 11. Sion Wyn Conformity +[44] (0) Regulatory Overview Annex 11 and Part 11 Sion Wyn Conformity +[44] (0) 1492 642622 sion.wyn@conform-it.com 1 Two Key Regulations Annex 11 21 CFR Part 11 Apply to the regulated company, but often have a

More information

Tech support: Luciferase Assay System. Protocol. LightSwitch Transfec tion Optimization Kit. Genome In. Function Out.

Tech support: Luciferase Assay System. Protocol. LightSwitch Transfec tion Optimization Kit. Genome In. Function Out. Luciferase Assay System Protocol LightSwitch Transfec tion Optimization Kit LightSwitch Transfection Optimization Kit Instructions for use in 96-well format, 4 day protocol OVERVIEW The LightSwitch Transfection

More information

Quality Supplier Requirements for Beckwith Electric Co., Inc.

Quality Supplier Requirements for Beckwith Electric Co., Inc. Quality Supplier Requirements for Beckwith Electric Co., Inc. This Page Left Intentionally Blank To Our Valued Suppliers: One of the primary goals of Beckwith Electric is to provide the highest Quality

More information

REPAIR STATION MANUAL

REPAIR STATION MANUAL UNCONTROLLED WHEN COPIED OR PRINTED CONTROLLED FORMS LOCATED AT: AEROWELD, INC 433 Breesport San Antonio, Texas 78216 REPAIR STATION MANUAL Including: QUALITY CONTROL SYSTEM FAA Certificated Repair Station

More information

Cyfra 21-1 IRMA. Product information Information about other products is available at: Userś Manual DE52100

Cyfra 21-1 IRMA. Product information Information about other products is available at:  Userś Manual DE52100 Product information Information about other products is available at: www.demeditec.com Userś Manual Cyfra 21-1 IRMA The CYFRA 21.1 IRMA system provides a direct in vitro quantitative determination of

More information

QUALITY MANUAL FOR Department of Materials Testing and Evaluation. Miami-Dade County Public Schools

QUALITY MANUAL FOR Department of Materials Testing and Evaluation. Miami-Dade County Public Schools Quality Manual Title/Signature Page Page 1 of 28 QUALITY MANUAL FOR Original March 3, 2003 Revised August 25, 2006 Revised January 13, 2005 Revised November 30, 2006 Revised June 19, 2006 Revised February

More information

Swette Center for Environmental Biotechnology Radiation Safety Policy

Swette Center for Environmental Biotechnology Radiation Safety Policy Reviewed: BR date RKB date CIT date Swette Center for Environmental Biotechnology Radiation Safety Policy Scope, Application and Purpose: The purpose of the Center for Environmental Biotechnology Radiation

More information

Lumi-Scint Liquid Scintillation Counter

Lumi-Scint Liquid Scintillation Counter DOE/EM-0596 Lumi-Scint Liquid Scintillation Counter Deactivation and Decommissioning Focus Area Prepared for U.S. Department of Energy Office of Environmental Management Office of Science and Technology

More information

Vaccine Storage and Handling INTERIM GUIDANCE

Vaccine Storage and Handling INTERIM GUIDANCE Vaccine Storage and Handling INTERIM GUIDANCE Introduction In response to recent scientific studies i on equipment used for vaccine storage and a better understanding of best practices for vaccine storage

More information

TÜV SÜD BABT Production Quality Certification Scheme

TÜV SÜD BABT Production Quality Certification Scheme TÜV SÜD BABT Production Quality Certification Scheme The Production Quality Certification Scheme for Manufacturers A Certification Body of Copyright TÜV SÜD BABT 2014 Page 1 of 38 CONTENTS Page AMENDMENT

More information

t Environmental Wipe Testing t Surface and Equipment Decontamination Protocols for the use of National Diagnostics Nuc-Wipes and Nuclean

t Environmental Wipe Testing t Surface and Equipment Decontamination Protocols for the use of National Diagnostics Nuc-Wipes and Nuclean t Environmental Wipe Testing t Surface and Equipment Decontamination Protocols for the use of National Diagnostics TM Nuc-Wipes and Nuclean TM USA: 1-800-526-3867 EUROPE: 0115 982 1111 NUC-WIPES (NW-300)

More information

USP Chapter 823 USP 32 (old) vs. USP 35 (new)

USP Chapter 823 USP 32 (old) vs. USP 35 (new) USP Chapter 823 USP 32 (old) vs. USP 35 (new) Sally W. Schwarz, MS, BCNP Research Associate Professor of Radiology Washington University School of Medicine St. Louis, MO Why USP Chapter ? FDA has

More information

Exposing Myths, Legends and Tales About Reference Materials. Presented by: Matt Sica - ANAB Steve Arpie Absolute Joe Konschnik - RESTEK

Exposing Myths, Legends and Tales About Reference Materials. Presented by: Matt Sica - ANAB Steve Arpie Absolute Joe Konschnik - RESTEK Exposing Myths, Legends and Tales About Reference Materials Presented by: Matt Sica - ANAB Steve Arpie Absolute Joe Konschnik - RESTEK Overview Reference Materials Accreditation Reference Materials Production

More information

Standard Operating Procedure Title: Operation and Calibration of Sievers 820 TOC Analyser

Standard Operating Procedure Title: Operation and Calibration of Sievers 820 TOC Analyser Table of Contents 1. 25 ppm IC as Sodium Carbonate - standard preparation procedure 2 2. 25 ppm TOC as KHP standard preparation procedure 3 3. Calibration Procedure 4 4. Performing Calibration - Prepare

More information

Quality Assurance / Quality Control (QA / QC)

Quality Assurance / Quality Control (QA / QC) Quality Assurance / Quality Control (QA / QC) DOCUMENTATION DOCUMENTATION Provides a complete history of any sample from initiation to project completion Includes all associated activities Documentation

More information

By Ginger Bailly 18 Feb For Official Use Only

By Ginger Bailly 18 Feb For Official Use Only By Ginger Bailly 18 Feb 2014 For Official Use Only 1 Discuss: Why Quality Assurance is Important How to Implement a Quality Management System Explore the Following Key Elements: Traceability Document Control

More information

PAB PRINCIPLE. Kit Reorder # ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PAB PRINCIPLE. Kit Reorder # ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON CX System(s) Chemistry Information Sheet Copyright 2008 Beckman Coulter, Inc. Prealbumin Kit Reorder # 475106 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE

More information

Human IgG ELISA Quantitation Set

Human IgG ELISA Quantitation Set Human IgG ELISA Quantitation Set Cat. No. E80-104 Components Supplied Affinity purified Goat anti-human IgG-Fc Coating Antibody A80-104A, 1 ml at 1 mg/ml Human Reference Serum, RS10-110-4, 0.1 ml HRP Conjugated

More information

Human IgG Antigen ELISA Kit

Human IgG Antigen ELISA Kit Human IgG Antigen ELISA Kit Catalog No: IHUIGGKT Lot No: SAMPLE INTENDED USE This human immunoglobulin G antigen assay is intended for the quantitative determination of total human IgG antigen in serum,

More information

Required Documentation for PA State (Chapter 252) Accreditation

Required Documentation for PA State (Chapter 252) Accreditation Required Documentation for PA State (Chapter 252) Accreditation Disclaimer: The information in this guidance document does not supplant the provisions of the Environmental Laboratory Accreditation Regulations,

More information

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856 , TNTplus856, 10220 DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and Application: For Water and Wastewater Test preparation Before starting the test: DR 2800:

More information

NR CHECKLIST Rev. 1. QAM IMP References NBIC Part 3, 1.8 Y N Y N a. Organization. Company Name/Certificate Number: Page 1 of 26

NR CHECKLIST Rev. 1. QAM IMP References NBIC Part 3, 1.8 Y N Y N a. Organization. Company Name/Certificate Number: Page 1 of 26 Company Name/Certificate Number: Page 1 of 26 a. Organization a.1. Has the Organizational Structure of the program identified the levels of management responsible for the Quality System Program, including

More information

Vaccine Storage and Handling INTERIM GUIDANCE

Vaccine Storage and Handling INTERIM GUIDANCE Vaccine Storage and Handling INTERIM GUIDANCE Introduction In response to recent scientific studies i on equipment used for vaccine storage and a better understanding of best practices for vaccine storage

More information

Human Peptide YY EIA

Human Peptide YY EIA KAMIYA BIOMEDICAL COMPANY Human Peptide YY EIA For the quantitative determination of Peptide YY in human serum and plasma. Cat. No. KT-378 For Research Use Only. 1 Rev. 9822378 PRODUCT INFORMATION Human

More information

HACCPEUROPA PUBLICATIONS ISO 22000:2005 FOOD SAFETY QUALITY MANUAL. ISO 22000:2005 Quality Manual

HACCPEUROPA PUBLICATIONS ISO 22000:2005 FOOD SAFETY QUALITY MANUAL. ISO 22000:2005 Quality Manual HACCPEUROPA PUBLICATIONS ISO 22000:2005 FOOD SAFETY QUALITY MANUAL ISO 22000:2005 Quality Manual QUALITY MANUAL ISO 22000:2005 Food Safety Management HACCPEuropa Publications 2012 Table of Contents Introduction...

More information

I.QC (quality control of your qpcr)

I.QC (quality control of your qpcr) qpcr data analysis I.QC (quality control of your qpcr)! 1. Check the melt curve if reactions with the same primer pair give only one peak -> proceed with the analysis if you see more than one peak, consider

More information

Bacterial Counts - Quantitative Analysis of Microbes

Bacterial Counts - Quantitative Analysis of Microbes Bacterial Counts - Quantitative Analysis of Microbes Introduction: It is often important to know not only what types of bacteria are in a sample but also how many of them are present. Food manufacturers

More information

PeliClass human IgG subclass ELISA kit Enzyme-linked immunosorbent assay

PeliClass human IgG subclass ELISA kit Enzyme-linked immunosorbent assay PeliClass human IgG subclass ELISA kit Enzyme-linked immunosorbent assay Catalog No: M1551 Size: six pre-coated 8-well strips for each of the four IgG subclasses Test description The PeliClass human subclass

More information

Growth Promotion Test Guide for Media Used in Sterility Tests

Growth Promotion Test Guide for Media Used in Sterility Tests Growth Promotion Test Guide for Media Used in Sterility Tests The Sterility Test The purpose of the Sterility Test is to determine if a sample of a pharmaceutical product or component is sterile. A satisfactory

More information

University of Maryland Baltimore. Radiation Safety Procedure

University of Maryland Baltimore. Radiation Safety Procedure University of Maryland Baltimore Procedure Number: 2.1 Radiation Safety Procedure Title: Ordering, Receiving, Opening, And Transferring Packages Containing Revision Number: 0 Technical Review and Approval:

More information

IDTox Lactate Dehydrogenase (LDH) Enzyme Cytotoxicity Kit For Cell Culture Supernatant Samples

IDTox Lactate Dehydrogenase (LDH) Enzyme Cytotoxicity Kit For Cell Culture Supernatant Samples Rev Date: Mar 26 2015 Rev: 1 1 of 1 IDTox Lactate Dehydrogenase (LDH) Enzyme Cytotoxicity Kit For Cell Culture Supernatant Samples 192 Wells Enzyme Immunoassay for the determination of the Lactate Dehydrogenase

More information

Validation Strategies for Equipment from Multiple Vendors

Validation Strategies for Equipment from Multiple Vendors Welcome to our E-Seminar: Validation Strategies for Equipment from Multiple Vendors Page 1 Content FDA guidelines and inspectional observations Validation & Qualification Approaches Instrument Qualification

More information

SECTION Coriolis Mass Flow Meters

SECTION Coriolis Mass Flow Meters SECTION 40 71 73 Coriolis Mass Flow Meters PART 1- GENERAL 1.01 SUMMARY A. Coriolis flow meter for permanent installation above ground. The meter shall utilize a measuring principle based on the controlled

More information

Precision Automatic Vacuum Control (PAVC+)

Precision Automatic Vacuum Control (PAVC+) USER GUIDE UGE105-1216 www.conairgroup.com Precision Automatic Vacuum Control (PAVC+) Corporate Office: 724.584.5500 l Instant Access 24/7 (Parts and Service): 800.458.1960 l Parts and Service: 814.437.6861

More information

Form Fit & Function Engineering, LLC 350 Seminole Drive Spartanburg, SC Supplementary Quality Assurance Clauses

Form Fit & Function Engineering, LLC 350 Seminole Drive Spartanburg, SC Supplementary Quality Assurance Clauses Supplementary Quality Assurance Clauses The Supplementary Quality Assurance Clauses specified on the face of the attached Purchase Order are listed below. Your compliance with these requirements is necessary

More information

PROGRAM FOR THE MANAGEMENT AND DISPOSAL OF RADIOACTIVE WASTE

PROGRAM FOR THE MANAGEMENT AND DISPOSAL OF RADIOACTIVE WASTE RADIATION SAFETY DEPARTMENT West Virginia University Health Sciences Center WVU Hospitals Jefferson Memorial Hospital G-139 Health Sciences Center PO Box 9006 Morgantown, WV 26506-9006 Phone: 304-293-3413

More information

Human Plasmin Activity Assay ELISA Kit

Human Plasmin Activity Assay ELISA Kit Human Plasmin Activity Assay ELISA Kit Catalog No. CSI12527A 1 x 96 tests CSI12527B 5 x 96 tests Intended Use: Background: Assay Principle: Reagents Provided: This human plasmin activity assay is for the

More information

CHARACTERIZATION OF RADIOACTIVE WASTES SPENT ION-EXCHANGE RESINS AND CHARCOAL FILTER BEDS

CHARACTERIZATION OF RADIOACTIVE WASTES SPENT ION-EXCHANGE RESINS AND CHARCOAL FILTER BEDS 2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 CHARACTERIZATION

More information

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2017-08-10 Effective

More information

Human IgG ELISA Kit. Strip well format. Reagents for up to 96 tests

Human IgG ELISA Kit. Strip well format. Reagents for up to 96 tests Human IgG ELISA Kit Strip well format. Reagents for up to 96 tests Catalog No. CS222A Quantity: 1 x 96 tests CS222B 5 x 96 tests Intended Use: Background: Assay Principle: This human immunoglobulin G antigen

More information

Competency Assessment: Is Your Program Competent? Judy Sullivan, MS, MT(ASCP)SBB HAABB Meeting April 30, 2013

Competency Assessment: Is Your Program Competent? Judy Sullivan, MS, MT(ASCP)SBB HAABB Meeting April 30, 2013 Competency Assessment: Is Your Program Competent? Judy Sullivan, MS, MT(ASCP)SBB HAABB Meeting April 30, 2013 Objectives Discuss requirements related to competency assessment Describe what assessors, inspectors,

More information

SOP: Human tpa Total Antigen ELISA Kit

SOP: Human tpa Total Antigen ELISA Kit Page 1 of 6 Approvals Preparer: Jason McMillan Reviewer: Dr. Margaret Bryans Date: 09APR14 Date: 10APR14 1. Purpose 1.1. Quantitative determination of total tpa 2. Scope and Applicability 2.1. Human tpa

More information

Validation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System

Validation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System User Bulletin Quantifiler Kits April 2006 SUBJECT: Validation Using SDS Software Version 1.2.3 on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System Note:

More information

OREGON HEALTH & SCIENCE UNIVERSITY RADIATION SAFETY OPERATING PROCEDURE 1801 RADIOACTIVE WASTE HANDLING & STORAGE

OREGON HEALTH & SCIENCE UNIVERSITY RADIATION SAFETY OPERATING PROCEDURE 1801 RADIOACTIVE WASTE HANDLING & STORAGE OREGON HEALTH & SCIENCE UNIVERSITY RADIATION SAFETY OPERATING PROCEDURE 1801 RADIOACTIVE WASTE HANDLING & STORAGE I. PURPOSE: To describe the manner in which radioactive waste shall be stored and/or disposed

More information

Executive Instruments Inc.

Executive Instruments Inc. Self Audit Report Agency Executive Instruments Inc. Phone 972 239 0231 Address 4141 Lindbergh Drive Fax 972 991 3271 Addison State T Zip 75001 Check applicable boxes below Original Equipment Manufacturer

More information

Application Note for Gentian Cystatin C Immunoassay on. ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011

Application Note for Gentian Cystatin C Immunoassay on. ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011 Page 1 of 5 Application Note for Gentian Cystatin C Immunoassay on For in vitro diagnostic use only ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011 Intended Use The Gentian Cystatin C Immunoassay is

More information

INTRODUCTION TO ELECTROCHEMISTRY: CURRENT, VOLTAGE, & BATTERIES. Introduction. Electrochemistry Revised 4/28/14

INTRODUCTION TO ELECTROCHEMISTRY: CURRENT, VOLTAGE, & BATTERIES. Introduction. Electrochemistry Revised 4/28/14 INTRODUCTION TO ELECTROCHEMISTRY: CURRENT, VOLTAGE, & BATTERIES Introduction Electrochemical Cells In this part of the experiment, four half cells are created by immersing metal strips of zinc, copper,

More information

TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009

TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009 SANAS Accr. No/s. TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009 Date/s of evaluation Assessor/s & Observers Facility Area / field of Representative operation This report covers the following:

More information

SAE AS9100/ISO 9001:2008

SAE AS9100/ISO 9001:2008 Table 0: Revision History SAE AS9100/ISO 9001:2008 Rev. Date DCN Description A 10/8/2003 465 Initial Release B 11/18/2003 482 Removed SP and SWI from Section 3.1. Updated terms in Section 3.2. Corrected

More information

RAD H2O With Stainless Steel Aerator Radon in Water Accessory

RAD H2O With Stainless Steel Aerator Radon in Water Accessory RAD H2O With Stainless Steel Aerator Radon in Water Accessory 2016 DURRIDGE Company DURRIDGE Company Inc. 524 Boston Rd Billerica, MA 01821 Tel: (978)-667-9556 Fax: (978)-667-9557 service@durridge.com

More information

QUALITY MANAGEMENT SYSTEM POLICIES AND PROCEDURES

QUALITY MANAGEMENT SYSTEM POLICIES AND PROCEDURES Your Company Name QUALITY MANAGEMENT SYSTEM POLICIES AND PROCEDURES Origination Date: XXXX Document Identifier: Date: Document Revision: QMS-00 QMS Policies and Procedures Latest Revision Date Abstract:

More information

Hach Method Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater

Hach Method Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater Hach Method 10360 Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater LED Photo Diode LED Hach Method 10360 Luminescence Measurement of Dissolved Oxygen in Water and Wastewater

More information

QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY

QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY Richard L. Hodinka, Ph.D. University of South Carolina School of Medicine Greenville Greenville Health System, Greenville, SC

More information

NAG-1 (GDF-15, MIC-1) Prostate Cancer ELISA kit

NAG-1 (GDF-15, MIC-1) Prostate Cancer ELISA kit NAG-1 (GDF-15, MIC-1) Prostate Cancer ELISA kit Catalog Number: NG1 Store at -0 C. FOR RESEARCH USE ONLY v. 1081 Introduction This sandwich ELISA kit is for determination of NAG-1 (GDF-15, MIC-1) levels

More information