Designing and Implementing a High-Level Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington
|
|
- Melvin Ford
- 5 years ago
- Views:
Transcription
1 Designing and Implementing a High-Level Multicolor Flow Cytometry Assay Brent Wood MD PhD Department of Laboratory Medicine University of Washington
2 Define Purpose of Assay Most important question What information is required? What information is most important? Prioritize Compromises are inevitable Simplest assay is best
3 Outline Instrument optimization Compensation Reagent selection and optimization Panel validation Data analysis
4 Detector Optimization
5 Detector Optimization
6 PMT Voltage Green = 400 V, Blue = 450 V, Red = 500 V, Violet = 550 V
7 Detector Voltage 475 volts 575 volts 675 volts 775 volts 875 volts Too low reduces sensitivity Too high pushes bright signals off scale Dilute with unlabeled antibody May need to compromise
8 Optimization Daily optimization not practical Once optimized, do NOT change instrument settings unless change in performance New optical filters New PMT detector Monitor performance daily Use stable bright beads Assess MFI and CV for each detector Levy-Jennings plots Define acceptable ranges of performance
9 Fluorochrome Intensity CD4 FITC = Green PE = Dark Blue PE-TR = Light blue PerCP-Cy5.5 = Magenta PE-Cy7 = Orange Pacific Blue = Red A594 = Yellow green APC = Light green A700 = Blue APC-Cy7 = Purple
10 Tandem Fluorochromes Phycoerythrin (PE) PE-Texas Red
11 Tandem Breakdown Whole blood lysis Cytometry Part A (2009) 75A: NH 4 Cl prelyse and wash 1 hour prior
12 Fluorochrome Options PE-Cy5 vs. PerCP-Cy5.5 vs. PE-Cy5.5 vs. PerCP PerCP dim PerCP-Cy5.5 brighter but dims with increased laser power PE-Cy5 brighter but significant comp with APC PE-Cy5.5 bright, less comp with APC, more comp with PE-Cy7 and APC-A700 A700 vs. APC-A700 APC-A700 significantly brighter but more comp with APC APC-Cy7 vs. APC-H7 vs. APC-A750 APC-Cy7 undergoes photodegradation but is brightest APC-H7 dimmer but more stable, low comp with APC APC-A750 intensity in between but higher comp with APC
13 Compensation Spectral overlap between fluorochromes Critical to success of method For 10 color experiment Need to determine 90 values for Comp Matrix Software compensation required Maximum flexibility Non-destructive
14 FITC = Green PE = Orange Excitation = Dotted Emission = Solid
15 Compensation - Method Single stained controls used One for each individual fluorochrome One for each individual tandem As bright as brightest reagent to be used Samples run without compensation Compensation calculated in software Applied either at acquisition or analysis
16 Compensation Correct Undercompensated Overcompensated
17 Compensation
18 Compensation
19 Compensation Don t worry unduly about PMT voltage 245% 103% 47.5% 23.0% 12.2% 1.9% 4.7% 10.3% 21.0% 41.0% PE = 400 volts PE-TR = 550 volts 450 volts 500 volts 550 volts 600 volts Compensation values should reflect relative spectral overlap, i.e. detector gains should be equal
20 Compensation Validation Fluorescence minus one FMO controls
21 Compensation Validation Fluorescence minus one FMO controls
22 Compensation Quality Control Purist Set at the time of each experiment Important when: Reagents and specimen preparation are inconsistent Accurate MFI to be measured Practical Quality control instrument and reagents using consistent specimen preparation Compensation is dependent on relative fluorescence intensity Slight manual adjustment may be required
23 0.9% 0.9% 1.0% June 2008 Nov 2008 Sept % 1.0% 1.1% May 2008 Dec 2008 May % 4.5% 6.1% May 2008 Dec 2008 May 2009
24 Stem cell evaluation Antigens to be evaluated: Early progenitors - CD34, CD117, CD38 Erythroid - CD71 B cell - CD19 pdcs, basophils - CD123, HLA-DR Myelomonocytic - CD15, HLA-DR Progenitor gating - CD45 Test reagents - CD45RA, CD133, CD7, etc.
25 Surface Antigen Titration 2 ul + 5 ul + 10 ul Titer for signal to noise Saturation desirable Use same total volume as assay Use 5 ul
26 Cytoplasmic Antigen Titration 10 ul, 5 ul, 2 ul, 1ul of neat Titer for signal to noise Not saturation Particularly important for cytoplasmic antigens 10 ul, 5 ul, 2 ul, 1ul of 1:10 Use 5 ul of 1:10 Use 10 ul of 1:10
27 Possibilities PB / V450 DR CD15 CD19 CD123 CD117 CD38 CD34 CD71 CD45 X FITC PE PE-TR P-X PE- Cy7 A594 APC APC- A700 APC- X7
28 Possibilities PB / V450 DR CD15 CD19 CD123 CD117 CD38 CD34 CD71 CD45 X FITC PE PE-TR P-X PE- Cy7 A594 APC APC- A700 APC- X7
29 Fluorochrome Matching Match level of expression with fluorochrome intensity Bright expression = Dim fluorochrome Dim expression = Bright fluorochrome
30 CD34
31 CD123
32 Background Autofluorescence varies between populations Fluorochrome may give non-specific binding Cyanine containing PE tandems Monocytes Fc-receptor mediated Beckman-Coulter reagents block PE-Cy5 PE-Cy5.5 PE-Cy7 Idiosyncratic CD30 APC-A700 Important gating reagents should be clean
33 Possibilities PB / V450 DR CD15 CD19 CD123 CD117 CD38 CD34 CD71 CD45 X FITC PE PE-TR P-X PE- Cy7 A594 APC APC- A700 APC- X7
34 Compensation Background Avoid increased background due to fluorochromes Adjacent with longer wavelength emission PE / PE-TR, PE-TR / PE-Cy5, PE-Cy5.5 or PerCP-Cy5.5/ PE-Cy7 APC / APC-A700, APC-A700 / APC-Cy7 Primary fluorochrome of tandem PE and PE-TR, PE-Cy5, PE-Cy5.5, or PE-Cy7 APC and APC-A700 or APC-Cy7 Interlaser excitation and emission PE-Cy5 and APC PE-Cy5.5 or PerCP-Cy5.5 and APC-A700 PE-Cy7 and APC-Cy7 PE-TR and A594
35 Adjacent fluorochromes
36 Primary of Tandems 10.5% 1.6%
37 Interlaser compensation
38 Strategies to deal with compensation background Avoid bright fluorescence Put fluorochromes on different populations Put fluorochromes brightly on same population Avoid detection of dim expression in presence of high background
39 Avoid bright fluorescence
40 Different populations
41 Bright dual positive
42 Possibilities PB / V450 DR CD15 CD19 CD123 CD117 CD38 CD34 CD71 CD45 X FITC PE PE-TR P-X PE- Cy7 A594 APC APC- A700 APC- X7
43 Possibilities PB / V450 DR CD15 CD19 CD123 CD117 CD38 CD34 CD71 CD45 X FITC PE PE-TR PECy5 PE- Cy7 A594 APC APC- A700 APC- X7
44 Possibilities PB / V450 DR CD15 CD19 CD123 CD117 CD38 CD34 CD71 CD45 X FITC PE PE-TR PECy5 PE- Cy7 A594 APC APC- A700 APC- X7
45 Final PB / V450 DR CD15 CD19 CD123 CD117 CD38 CD34 CD71 CD45 X FITC PE PE-TR PECy5 PE- Cy7 A594 APC APC- A700 APC- X7
46 Compensation Compromises
47 Validation Fluorescence minus one controls Antibody interactions Avoid IgG2 class unless plasma removed Understand background Troubleshooting Run many samples Under same conditions as to be used Positive and negative controls
48 PE background
49 Data analysis Need better software tools Multiparametric analysis is essential Population identification Compensation background Informative use of color Density plots very helpful Use many dot plots Not every possible combination
50 Erythroid gating
51 Information Opportunities
52 Information Opportunities
53 Information Opportunities
54 Conclusion Instrument optimization Reagent selection Fluorochrome matching Fluorochrome options Background evaluation Compensation background Develop compensation strategy Validation Develop data analysis strategy
Designing and Validating a Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington
Designing and Validating a Multicolor Flow Cytometry Assay Brent Wood MD PhD Department of Laboratory Medicine University of Washington Specimen Handling Sample Requirements 5 ml Peripheral blood (EDTA,
More informationUnderstanding Flow Cytometry
Understanding Flow Cytometry The Basic Concepts Maree Bagnara Products Sales Specialist/Account Manager Flow Cytometry PN775136 1 Successful Flow Cytometry data is driven by.. Understanding the Biology
More informationAnatomy of a flow cytometer
Anatomy of a flow cytometer Fluidics Optics Electronics Cells in suspension flow in single-file through an illuminated volume where they scatter light and emit fluorescence that is collected, filtered,
More informationPrinciples of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting. Shoreline Community College BIOL 288
Principles of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting Shoreline Community College BIOL 288 Flow Cytometry What is Flow Cytometry? Measurement of cells or particles
More informationEach question may have MULTIPLE correct answers. Select all that are correct.
Knowledge Assessment Flow Cytometry Workshop, Part 1 April 20, 2015 Each question may have MULTIPLE correct answers. Select all that are correct. 1. Tandem dyes are a. highly stable fluorophores after
More informationPrinciples of Multicolor Panel Design BD. BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company.
1 Principles of Multicolor Panel Design 2 Common Multicolor Applications Intracellular cytokine staining Regulatory T cells (Tregs) Protein phosphorylation (BD Phosflow) Leukemia and lymphoma phenotyping
More informationIncorporating New, Bright Fluorochromes into Multicolor Panel Design
Incorporating New, Bright Fluorochromes into Multicolor Panel Design Maria C. Jaimes, MD Senior Staff Scientist BD Biosciences 23-14684-00 Overview Multicolor flow: successful application prerequisites
More informationAMREP FLOW CYTOMETRY CORE FACILITY. Flow Cytometry Data Analysis Workshop Friday 24 th October
AMREP FLOW CYTOMETRY CORE FACILITY Flow Cytometry Data Analysis Workshop Friday 24 th October Polychromatic Flow Cytometry Data Analysis: Immunophenotyping Lymphocyte sub-populations and targeting their
More informationWhat you need to know before designing a panel
Design What you need to know before designing a panel For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life Technologies Corporation.
More informationSelecting Reagents for Multicolor Flow Cytometry
HotLines Platinum Edition f a l l 0 0 6 Selecting Reagents for Multicolor Flow Cytometry By Holden Maecker and Joe Trotter The availability of flow cytometers capable of detecting 6, 8, and more colors
More informationFlow Cytometry - The Essentials
Flow Cytometry - The Essentials Pocket Guide to Flow Cytometry: 1. Know your Cytometer 2. Understanding Fluorescence and Fluorophores 3. Gating Process 4. Controls 5. Optimization 6. Panel Building 7.
More informationA previous ICCS Module entitled Instrument optimization - Adjusting PMT voltages and compensation 1 should be read as a prerequisite to this module.
Sponsored and reviewed by ICCS Quality and Standards Committee Title: Compensation Tips for Beckman Coulter 10-Color Navios Platform Written by: Salima Janmohamed-Anastasakis Ph.D., Applications Scientist,
More informationCompensation: Fundamental Principles
Flow Cytometry Seminar Series 2017 : Fundamental Principles Spillover correction in multicolor flow cytometry 28.02.2017 http://www.cytometry.uzh.ch Contents Fluorescence and its detection Absorption and
More informationFlow cytometry has become an increasingly important
9-Color and 10-Color Flow Cytometry in the Clinical Laboratory Brent Wood, MD, PhD Context. The development of commercial flow cytometers capable of detecting more than 10 simultaneous fluorescent signals
More informationExpanding Multicolor Options in Flow Cytometry with Novel Brilliant Violet TM Fluorophores. Carsten Wiethe Scientific Application Manager BioLegend
Expanding Multicolor Options in Flow Cytometry with Novel Brilliant Violet TM Fluorophores Carsten Wiethe Scientific Application Manager BioLegend Seminar Outline BrilliantViolet TM (BV)fluorophores Instrument
More informationBeyond Colorful Science. A Practical Approach to Setting Up Multicolor Panels
Beyond Colorful Science A Practical Approach to Setting Up Multicolor Panels ISCT Paris April 2014 Layout: Multicolor Panel Design Identify the pitfalls Have a look at issues with spectral overlap How
More informationSpherotech, Inc. 1. SPHERO TM Technical Note STN-8 Rev C
SPHERO TM Technical Note STN- Rev C. 0070 CALIBRATION AND PERFORMANCE TRACKING OF FLOW CYTOMETERS USING SPHERO TM CALIBRATION PARTICLES Introduction The SPHERO TM Calibration Particles are versatile, stable,
More informationSelected Topics in Electrical Engineering: Flow Cytometry Data Analysis
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Experimental design
More informationDesigning and executing a successful flow cytometry experiment
Designing and executing a successful flow cytometry experiment 1x1 0 5 Orange Fluorescent Protein 10000 1000 100 10 0 1000 2000 3000 4000 Forward Scatter This presentation is based on true events, however,
More informationApplication Note. Assay Portability on the BD FACSVerse System. Summary. Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar
September Assay Portability on the BD FACSVerse System Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar Contents Summary Introduction 3 Objective 4 Methods 6 Results Discussion Conclusions
More informationVisualization of digital data Interpretation of the visual
Technical aspects of 8-color flow cytometry in the diagnosis and classification of hematopoietic malignancies Tomas Kalina!"#$%&'()*+,&$'+-./(0 *1 (2#34%-.(56(7&1+3+*&/( 8$#94&/(!:&3"(;&
More informationBD Multicolor CompBeads
4/2015 23-9955-01 IVD BD Multicolor CompBeads 100 compensation setups Catalog No. 644204 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2015 BD Becton, Dickinson and
More informationFACS Core Facility User Group Meeting. 25. April 2018 Janine Bögli
FACS Core Facility User Group Meeting 25. April 2018 Janine Bögli Overview Basis: Forensic Flow Cytometry by Jennifer Wilshire, PhD, Stemcell Technologies You will suggest the crime in each case 8 Practical
More informationa Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper Long Term Stabilization of Tandem Dyes for Use in High Content, Multi Variant Flow Cytometry Authors: Snehita Sattiraju 1, Tewfik Miloud 2, Neha Girish 1, Murthy
More informationBest practices in panel design to optimize the isolation of cells of interest
Sort Best practices in panel design to optimize the isolation of cells of interest For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life
More informationFLOW CYTOMETRY. CyAn ADP. Analyzer
FLOW CYTOMETRY CyAn ADP Analyzer Experience the Power of the CyAn ADP and its optimal performance The Power of Detection The Power of Speed The Power of Ease The CyAn ADP Analyzer is the next step in Advanced
More informationIdentification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer. Application Note
Identification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer Application Note Sylvie Veriac Valérie Perrone Madeleine Avon Abstract Agilent Equipment: 2100 bioanalyzer
More informationBoundary-breaking acoustic focusing cytometry
Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory
More informationMinimum Information about a Flow Cytometry Experiment (MIFlowCyt) Annotation
Minimum Information about a Flow Cytometry Experiment (MIFlowCyt) Annotation 1. Experiment Overview 1.1 Purpose The purpose of these sets of experiments is to develop a methodology of identifying and quantifying
More informationSelecting Fluorochrome Conjugates for Maximum Sensitivity
2004 Wiley-Liss, Inc. Cytometry Part A 62A:169 173 (2004) Technical Primer Selecting Fluorochrome Conjugates for Maximum Sensitivity Holden T. Maecker,* Tom Frey, Laurel E. Nomura, and Joe Trotter BD Biosciences,
More informationAttune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry
Attune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry Maybelline Giam Field Application Scientist The world leader in serving science Attune NxT Flow Cytometer Attune NxT Acoustic
More informationFlow Cytometry in the Diagnosis of Hematopoietic Neoplasia. Brent Wood MD, PhD Professor, Laboratory Medicine University of Washington, Seattle
Flow Cytometry in the Diagnosis of Hematopoietic Neoplasia Brent Wood MD, PhD Professor, Laboratory Medicine University of Washington, Seattle 1 Flow Cytometer 2 The Power of Flow Cytometry Single cell
More informationQuality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre
Quality Control in Flow Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre Aims Quality Assurance Quality Control Literature In house competencies SHOT DATA 1996-2009 Ref: SHOT Annual
More informationFlow Cytometry SOP: Monocytes from Frozen Cells
Flow Cytometry SOP: Monocytes from Frozen Cells Purpose This SOP standardizes the procedure for measuring immune cells using flow cytometry in ACTG Immunology Laboratories. Materials 1. 12x75mm flow tubes
More informationHigh-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates
APPLICATION NOTE Attune NxT Flow Cytometer with Autosampler High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates Introduction The emerging field
More informationCalibración de equipos de citometría para empleo de paneles EuroFlow
Calibración de equipos de citometría para empleo de paneles EuroFlow Departamento de Medicina, Centro de Investigación del Cáncer y Servicio de Citometría. Universidad de Salamanca, Salamanca, España.
More informationUsing Fluorescence Spillover to Advantage
Using Fluorescence Spillover to Advantage Boston June 9, 2011 Compensation: Why and When is it Necessary? Clare Rogers, Marketing and Applications Accuri Cytometers, Inc. Ann Arbor/St.Ives 1 June 11 Summary
More informationDetecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer
APPLICATION NOTE Attune Acoustic Focusing Cytometer Detecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer Circulating endothelial cells (CECs) are mature cells shed
More informationANAT 3231 Cell Biology Lab12 Stem Cell Analysis
ANAT 3231 Cell Biology Lab12 Stem Cell Analysis 2 June 2010 Dr Antonio Lee Neuromuscular & Regenera9ve Medicine Unit School of Medical Sciences, UNSW Introduction to Flow Cytometry Contributed by Vittoria
More informationFlow Cytometry. Flow Cytometry Basics Guide
Flow Cytometry Flow Cytometry Basics Guide Table of Contents Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Principles of the Flow Cytometer Fluidics System.... 3 Optics and Detection.... 4 Signal and
More informationRuud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth
Sponsored and reviewed by ICCS Quality and Standards Committee Title: Quality of Reagents Monoclonal Antibodies Written by: Date: March 6, 2018 Ruud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth
More informationBASICS OF FLOW CYTOMETRY
BASICS OF FLOW CYTOMETRY AUTHOR: Ana Isabel Vieira APPROVAL: Henrique Veiga Fernandes Ana Sílvia Gonçalves SOP.UCF.002 03-09-2015 Pag. 1/9 Overview Flow: Fluid Cyto: Cell Metry: Measurement Flow cytometry
More informationApplication Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
More informationFlow Cytometry SOP: 14 color flow for immune activation, senescence, and exhaustion
Flow Cytometry SOP: 14 color flow for immune activation, senescence, and exhaustion Purpose This SOP standardizes the procedure for measuring immune cells using flow cytometry in ACTG Immunology Laboratories.
More information11/19/2013. Janine Zankl FACS Core Facility 13. November Cellular Parameters. Cellular Parameters. Monocytes. Granulocytes.
DEPARTEMENT BIOZENTRUM Janine Zankl FACS Core Facility 13. November 2013 Cellular Parameters Granulocytes Monocytes Basophils Neutrophils Lymphocytes Eosinophils Cellular Parameters 1 What Is Flow Cytometry?
More informationa Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper CytoFLEX Instrument Evaluation Using Biological Specimens Authors: James Tung 1, Dan Condello 3, Albert Donnenberg 4, Erika Duggan 3, Jesus Lemus 1, John Nolan
More informationINTRODUCTION TO FLOW CYTOMETRY
DEPARTEMENT BIOZENTRUM INTRODUCTION TO FLOW CYTOMETRY F ACS C ore F acility Janine Zankl FACS Core Facility 3. Dezember 2015, 4pm Cellular Parameters Granulocytes Monocytes Basophils Lymphocytes Neutrophils
More informationNovel Developments in Cytometry. Martin Adelmann Marketing Manager Cellular Analysis & Life Science Beckman Coulter EMEAI
Novel Developments in Cytometry Martin Adelmann Marketing Manager Cellular Analysis & Life Science Beckman Coulter EMEAI 1 Innovation in Flow Cytometry Cellios EPICS 1V 2 lasers, 3 colors EPICS Profile
More informationa Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper DuraClone improves standardization of compensation workflow for high content multicolor applications by flow cytometry Authors: Neha Girish 2, Sudharsan Sathyamurthy
More informationab pdc Subset Phenotyping Kit
ab177772 pdc Subset Phenotyping Kit Instructions for Use A set of antibodies designed to identify pdcs and subsets of pdcs by multi-color flow cytometry. This product is for research use only and is not
More informationAttune TM Acoustic Focusing Cytometer Training. Manik Punj Attune Training
Attune TM Acoustic Focusing Cytometer Training Manik Punj Attune Training Attune Training Agenda Section 1 An Introduction to Flow Cytometry Section 2 An Introduction to Acoustic Focusing Hydrodynamic
More informationIntroduction to. BD FACSAria TM Cell Sorter. Flow = Fluid Cyto = Cell Metry = Measurement
What is Flow Cytometry? Introduction to BD FACSAria TM Cell Sorter Flow = Fluid Cyto = Cell Metry = Measurement BD Biosciences Application Specialist 產品應用專員 Daisy Kuo 郭正佼 A variety of measurements are
More informationFlow Cytometry Immune Activation SOP
Flow Cytometry Immune Activation SOP Purpose This SOP standardizes the procedure for measuring immune activation of T cells using flow cytometry in ACTG Immunology Laboratories. Materials 1. 12x75mm flow
More informationQdot nanocrystal. wide range of biological investigations, Qdot nanocrystals are powerful complements
Feature nanocrystal conjugates for flow cytometry Take the easy route to multicolor flow cytometry. With applications across a wide range of biological investigations, nanocrystals are powerful complements
More informationdetermine optimum instrument settings for their own instruments and establish their own daily values.
PC7 (770/488) SETUP KIT 6607121 PN 4299504-C FLOW CYTOMETER ALIGNMENT VERIFICATION FLUOROSPHERES FLOW CYTOMETER DETECTOR STANDARDIZATION FLUOROSPHERES INTENDED USE For Research Use Only. Not for use in
More informationJ. Philip McCoy, Jr., Ph.D., H.C.L.D. J. Philip McCoy, Jr., PhD
J. Philip McCoy, Jr., Ph.D., H.C.L.D. Email: mccoyjp@mail.nih.gov 1 Introduction 2 How It All Works A very brief description of how a flow cytometer works. Please read it it will help you design your experiments.
More informationMulticolor flow cytometry analysis of human pluripotent stem cell cultures
Multicolor flow cytometry analysis of human pluripotent stem cell cultures ackground Thorough characterization of pluripotency is mandatory during the establishment of newly derived human embryonic or
More informationACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man CD38 Quantitation 07 May 2004
1. Principle Flow Cytometric quantitation of CD38 expression on CD8+ T lymphocytes The purpose of this assay is to quantitate the surface expression of the CD38 molecule on CD8 positive T lymphocytes.
More informationNovoCyte Flow Cytometer
NovoCyte Flow Cytometer The Flow Cytometer for Everyone 2 Experience the NovoCyte Advantage Focus on advancing your research. Let the flow cytometer do the rest. NovoCyte Flow Cytometer High Performance
More informationReview of techniques in flow cytometry. Peter Meeus Onze-Lieve-Vrouwziekenhuis, Aalst 5 may 2011, SCK/CEN Mol
Review of techniques in flow cytometry Peter Meeus Onze-Lieve-Vrouwziekenhuis, Aalst 5 may 2011, SCK/CEN Mol Cytometry definitions The counting and measuring of cells,... Mosby's Medical Dictionary, 8th
More informationYour Research, Revolutionized
Your Research, Revolutionized Drive Your Research Forward Your research needs are evolving and with the CytoFLEX flow cytometer you ll see just how far your data can take you. CytoFLEX has the advanced
More informationFlowcytometry Dirk Pacholsky
Flowcytometry Dirk Pacholsky Flowcytometry: Overview 1 2 Overview Flow Cyto Metry Fluid Cell Measurement measuring cell properties of cells in suspension Most Flow Cytometer have Spatially separated Lasers
More informationLower spectral overlap. Optimized stain index. Innovative antibodies. Superior multicolor flow cytometry. Violet Laser Reagents: The efluor Solution
Lower spectral overlap Optimized stain index Innovative antibodies Superior multicolor flow cytometry Violet Laser Reagents: The efluor Solution www.ebioscience.com ebioscience provides innovative high
More informationThe NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee
January, 1999 From: To: Subject: The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee NIAID DAIDS Flow Cytometry Laboratories Comparison study information for labs wishing to switch
More informationApplication Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents
Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents Compensation setup for high content DuraClone reagents INTRODUCTION High content flow cytometry
More informationefluor Organic Dyes 450/50 BP Fluorescence Intensity Wavelength (nm)
efluor Organic Dyes efluor Organic Dyes Catalog No. Description Clone Application Anti-Mouse efluor 450 Products 48-0042 Anti-mouse CD4 RM4-5 Flow Cytometry 48-0081 Anti-mouse CD8 53-6.7 Flow Cytometry
More informationYour Research, Revolutionized
Your Research, Revolutionized Drive Your Research Forward Your research needs are evolving and with the CytoFLEX flow cytometer you ll see just how far your data can take you. CytoFLEX has the advanced
More informationPowERFUL. INTUITIvE. customizable. ThE NEw STAR of. BENchToP FLow cytometry
PowERFUL. INTUITIvE. customizable. ThE NEw STAR of BENchToP FLow cytometry ABoUT AcEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
More informationarc lamp is substituted. Before
CE update [cytology hematology generalist] The Principles of Flow Cytometry Antony C. Bakke, PhD From the Department of Pathology, Oregon Health Sciences University, Portland, OR On completion of this
More informationFlow Cytometry For New PhDs
Flow Cytometry For New PhDs 2012 Simon Monard SCRM and CIR smonard@staffmail.ed.ac.uk Derek Davies CRUK derek.davies@cancer.org.uk Monday 20th and Tues 21st Feb 2012 Program Day 1: Day 2: Basics of flow
More informationWelcome to our first webinar. A beginner's guide to flow cytometry and all you ever need to know about preparing fluorescent conjugated antibodies.
in association with Welcome to our first webinar A beginner's guide to flow cytometry and all you ever need to know about preparing fluorescent conjugated antibodies. Speakers Prof Graham Pockley What
More informationIntroduction to Flow Cytometry. -- BD FACSCanto II TM. Daisy Kuo Application Specialist BDBiosciences
Introduction to Flow Cytometry -- BD FACSCanto II TM Daisy Kuo Application Specialist E-mail: daisy_kuo@bd.com BDBiosciences Outline Basic Concept of Flow Cytometry FACSCanto II System Introduction Application
More information7-amino actinomycin D (7ADD) was added to all samples 10 minutes prior to analysis on the flow cytometer in order to gate 7AAD viable cells.
Antibody staining for Ho uptake analyses For HSC staining, 10 7 BM cells from Ho perfused mice were stained with biotinylated lineage antibodies (CD3, CD5, B220, CD11b, Gr-1, CD41, Ter119), anti Sca-1-PECY7,
More informationFlow Cytometry Support Reagents
Excite and inspire Flow Cytometry Support Reagents Introduction Miltenyi Biotec is a leading supplier of flow cytometry products, offering one of the broadest ranges of antibodies, kits, assays, and support
More informationTECHNICAL BULLETIN. QUANTUM SIMPLY CELLULAR KIT Product Numbers QSC-20 AND QSC-100 Storage Temperature 2-8 C. Do Not Freeze
QUANTUM SIMPLY CELLULAR KIT Product Numbers QSC-20 AND QSC-100 Storage Temperature 2-8 C. Do Not Freeze TECHNICAL BULLETIN Product Description The Quantum Simply Cellular Kit provides a convenient method
More informationSupplementary Figure 1 (Page1)
Supplementary Figure 1 (Page1) A B FSC-A 45 Fitc-A 133 APC-A 19 APC-Cy7-A 34 PE-A 45 Fitc-A FSC-A 45 Fitcs-A C 38 PE-Cy7-A 45 Fitc-A 45 Fitc-A 7AAD FSC-A Supplementary Figure 1 (Page2) D (1) (6) (2) (3)
More informationBD FACSCanto II. A proven research platform for maximum reliability and the highest quality results
BD FACSCanto II A proven research platform for maximum reliability and the highest quality results A proven platform for maximum reliability and the highest quality results Built on more than 25 years
More informationValidation of Cell-based Fluorescence Assays: Practice Guidelines from the ICSH and ICCS Part III Analytical Issues
Cytometry Part B (Clinical Cytometry) 84B:291 308 (2013) Validation of Cell-based Fluorescence Assays: Practice Guidelines from the ICSH and ICCS Part III Analytical Issues Shabnam Tanqri, 1 * Horacio
More informationDA-Cell Washer and Plate for Flow Cytometry. Accelerating Life Sciences
DA-Cell Washer and Plate for Flow Cytometry Accelerating Life Sciences Conventional Centrifuge-based protocol for staining cells Challenges of staining cells for flow cytometry by centrifuge Losing cells
More informationThe Keys to Successful Polychromatic Flow Cytometry
Dale and Betty Bumpers Vaccine Research Center National Institute of Allergy and Infectious Diseases National Institutes of Health The Keys to Successful Polychromatic Flow Cytometry (Design and Implementation
More informationBD Human Pluripotent Stem Cell Transcription Factor Analysis Kit
BD Human Pluripotent Stem Cell Transcription Factor Analysis Kit Instruction Manual Catalog No. 560589 ii BD Human Pluripotent Stem Cell Transcription Factor Analysis Kit 2009, Becton, Dickinson and Company.
More informationSpherotech, Inc. 1. SPHERO TM Technical Note STN-9 Rev D
SPHERO TM Technical Note STN-9 Rev D. 00 MEASURING MOLECULES OF EQUIVALENT FLUOROCHROME (MEF) USING SPHERO TM RAINBOW AND ULTRA RAINBOW CALIBRATION PARTICLES Introduction The Molecules of Equivalent Fluorochrome
More informationInhibiScreen Kinase Inhibitor Assay
InhibiScreen Kinase Inhibitor Assay Flow Cytometric assessment of efficacy of PI3Kγ, PI3Kδ, BTK and SYK pathway inhibitors using the measurement of basophil activation in human EDTA and Heparin Whole Blood.
More informationBD LSRFortessa. Performance without peer, choice without compromise
BD LSRFortessa Performance without peer, choice without compromise Performance without peer, choice without compromise The BD LSRFortessa cell analyzer offers the ultimate in choice for flow cytometry,
More informationQuality Control in polychromatic Flow Cytometry - EuroFlow interlaboratory standardization
IV. International Symposium Flow Cytometry/Molecular Biology HOSPITAL ISRAELITA ALBERT EINSTEIN São Paulo 2012 Quality Control in polychromatic Flow Cytometry - EuroFlow interlaboratory standardization
More informationSANTA CRUZ BIOTECHNOLOGY, INC.
TECHNICAL SERVICE GUIDE: Western Blotting 2. What size bands were expected and what size bands were detected? 3. Was the blot blank or was a dark background or non-specific bands seen? 4. Did this same
More informationSA3800 Spectral Analyzer
SA38 Spectral Analyzer Sony Biotechnology Inc. SA38 Spectral Analyzer The Sony SA38 spectral cell analyzer incorporates advanced electronics, patented optical technologies and many points of automation
More informationPURPOSE: To delineate the subsets of human lymphocytes based on the expression profiles of different phenotypic markers by FACS analysis
LABORATORY PROCEDURE: IMMUNOPHENOTYPING: Lymphocyte Staining for FACS Analysis Date: April 29 2014 Authors: Jennifer Hossler PURPOSE: To delineate the subsets of human lymphocytes based on the expression
More informationSA3800 Spectral Analyzer
SA38 Spectral Analyzer Sony Biotechnology Inc. SA38 Spectral Analyzer The Sony SA38 spectral cell analyzer incorporates advanced electronics, patented optical technologies and many points of automation
More informationFlow Cytometric Devices Draft Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 36 Flow Cytometric Devices Draft Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE This guidance document
More informationDifferent Potential of Extracellular Vesicles to Support Thrombin Generation: Contributions of Phosphatidylserine, Tissue Factor, and Cellular Origin
Different Potential of Extracellular Vesicles to Support Thrombin Generation: Contributions of Phosphatidylserine, Tissue Factor, and Cellular Origin Carla Tripisciano 1, René Weiss 1, Tanja Eichhorn 1,
More informationLab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations.
Lab 2 Isolation of mononuclear cells from peripheral blood and separation into subpopulations. Supervisors: Ulrika Andreasson ulrika.andreasson@immun.lth.se tel: 222 9264 Niclas Olsson niclas.olsson@immun.lth.se
More informationBD Mouse Pluripotent Stem Cell Transcription Factor Analysis Kit
BD Mouse Pluripotent Stem Cell Transcription Factor Analysis Kit Instruction Manual Catalog No. 560585 ii BD Mouse Pluripotent Stem Cell Transcription Factor Analysis Kit Trademarks Cy is a trademark of
More informationReagent Titration for the ICS Assay
Purpose This standard operating procedure (SOP) describes how to titrate new lots of fluorochromeconjugated antibody reagents and the cell viability marker for use in the intracellular cytokine staining
More informationFlexible, Intuitive and Affordable
Data Sheet guava easycyte Flow Cytometry Systems Flexible, Intuitive and Affordable The guava flow cytometry systems are easy to use and deliver complete and comprehensive cell analysis right on your benchtop.
More informationSpherotech, Inc Irma Lee Circle, Unit 101, Lake Forest, Illinois
78 Irma Lee Circle, Unit, Lake Forest, Illinois 00 SPHERO TM Technical Note STN-9 Rev D. 0 MEASURING MOLECULES OF EQUIVALENT FLUOROCHROME (MEF) USING SPHERO TM RAINBOW AND ULTRA RAINBOW CALIBRATION PARTICLES
More informationReview of techniques in imaging and cytometry. Peter Meeus Onze Lieve Vrouw Ziekenhuis, Aalst 7 may 2009, SCK/CEN Mol
Review of techniques in imaging and cytometry Peter Meeus Onze Lieve Vrouw Ziekenhuis, Aalst 7 may 2009, SCK/CEN Mol Cytometry: The counting and measuring of cells,... Mosby's Medical Dictionary, 8th edition.
More informationMichael Waring Associate Director Ragon Institute Imaging Core Facility Cambridge, Massachusetts, USA
Michael Waring Associate Director Ragon Institute Imaging Core Facility Cambridge, Massachusetts, USA Ragon Institute Imaging Core Maintain and oversee operation of the Institute s flow cytometry equipment
More informationSpherotech, Inc Irma Lee Circle, Lake Forest, IL 60045
Spherotech, Inc. 36 7845 Irma Lee Circle, Lake Forest, IL 60045 SPHERO TM Ultra Rainbow Fluorescent Particles Consists of a single peak for optical alignment of any flow cytometer in all channels from
More informationImmunophenotyping UNIT
Immunophenotyping UNIT 6.2 There are four basic methods for staining cells with antibodies for immunophenotyping by flow cytometry. The first method (see Basic Protocol 1) is an indirect one that employs
More information