1 a Beckman Coulter Life Sciences: White Paper DuraClone improves standardization of compensation workflow for high content multicolor applications by flow cytometry Authors: Neha Girish 2, Sudharsan Sathyamurthy 2, Riddhi Bhide 2, Neema Jayadas 2,Valerie Mallet 1, Tewfik Miloud 1, Nathalie Dupas 1, Mike Carlson 3, Michael Kapinsky 1, Julie Vernier 1, Sylvain Monseaux 1, Soraya Kadour 1, Frederic Monsonis 1, James Tung 1, Vishakha Mangale 2, William Godfrey 3, Sridhar Ramanathan 2, Emmanuel Gautherot 1, Affiliation: 1 Beckman Coulter Life Sciences, Marseille, FRANCE., 2 Beckman Coulter India Pvt Limited, Bangalore Development Centre, Bangalore, INDIA 3 Beckman Coulter Life Sciences, Miami, FL, USA. This work was originally presented as a poster at ESCCA Conference, 2014, Lisbon.
2 DuraClone improves standardization of compensation workflow for high content multicolor applications by flow cytometry ABSTRACT Introduction: Multiparametric flow cytometry is a valuable tool for clinical research as it allows the use of multiple dyes, including tandems, to monitor several markers simultaneously. However, degradation of tandem dyes over time as a result of inconsistent reagent handling, light exposure, inappropriate storage conditions and/or manufacturing lot-to-lot variability can severely affect their spectral spillover and result in frequent adjustment in compensation. Beckman Coulter recently introduced DuraClone, a new proprietary format that allows antibody conjugate stabilization in a dry, unitized format enabling long-term storage at room temperature. The DuraClone format accommodates both multicolor panel and single color conjugate formats to enable easy compensation set-up while also improving standardization of compensation workflow in high content multicolor applications by flow cytometry. Study: The instrument was set-up using the Beckman Coulter Navios flow cytometer Autosetup scheduler to generate a robust and testable compensation matrix for routine use in multicolor panels, including those containing PE- and APC- tandems. The compensation matrix was set-up using a whole blood sample stained with DuraClone single color conjugates. The PMT settings are established by setting the X-Median of each single stain at 0.3 ± Positively stained cells are used by the Autosetup scheduler to estimate the spill-over and generate the compensation matrix. When the antigen is rarely or dimly expressed in biological samples, VersaComp positive beads will be spiked in the biological samples to obtain sufficient positive signal. PMT settings can be recorded using Flow-Set Pro Fluorosphere beads, allowing recurrent use of the established settings. Results: Taken together, the use of Beckman Coulter s proprietary DuraClone reagent technology and application of workflow tools such as the AutoSetup Scheduler, using the Navios software platform, enabled automatic compensation matrices to be established as applicable to multicolor panel design. These touch points reduce, or in some cases eliminate altogether, the need for reiterative fine adjustments in compensation outcomes. Conclusions: This study demonstrates the potential of our novel DuraClone reagent technology to deliver stable reproducible results for high content multicolor flow acquisition and analysis. BECKMAN COULTER WHITE PAPER 2
3 MATERIAL AND METHODS To demonstrate the compensation set-up and standardization (Fig 1) of the multicolor panels, two panels were used, DuraClone IM B cell (Table 1a) and DuraClone IM Dendritic cell panels (Table 2a). Generic single colors, with CD4 or CD8 specificities were used to set compensation for Non-Tandem dyes and Lot-matched conjugates from the panels were used for tandem dyes (Table 1b, 2b). The study was conducted using fresh human blood. Samples were processed according to a Stain-Lyse-Wash protocol using 2 ml VersaLyse lysing buffer for singles and DuraClone IM Dendritic Cell panel. One drop of positive VersaComp Beads was added to the CD123-APCA700 single tube with blood for staining. For the DuraClone IM B Cell panel, Wash-Stain-Lyse-Wash protocol was used. Processed samples were analyzed on a Navios flow cytometer equipped with 3 lasers capable of detecting 12 parameters at 20-bit resolution. These data were analyzed using Kaluza Software. Table 1 DuraClone IM B Cell Panel q Compensation kit for B Cell Panel q DuraClone IM B Cell Panel * Dye FITC PE ECD PC7 APC APC A750 (1) Pacific Blue* Marker IgD CD21 CD19 CD27 CD24 CD38 IgM CD45 Compensation Kit - B Cells (single color tubes) * Dye FITC PE ECD PC7 APC APC A750 (1) Pacific Blue* Marker CD4 CD4 CD19 CD27 CD4 CD38 CD4 CD8 Table 2 DuraClone IM Dendritic Cell Panel q DuraClone IM Dendritic Cell Panel * (coming soon) Dye FITC PE PC5.5 PC7 APC Marker CD16 Lineage (CD3/CD19/ CD20/CD14/ CD56) Compensation kit for Dendritic Cell panel q APC Pacific A700 (2) Blue* CD1c CD11c Clec9A CD123 HLA-DR CD45 Compensation Kit - Dendritic Cells (single color tubes) * Dye FITC PE PC5.5 PC7 APC APC Pacific A700 (2) Blue* Marker CD4 CD4 CD1c CD11c CD4 CD123 CD4 CD8 (1) APC-Alexa Fluor* 750 (2) APC-Alexa Fluor* 700 * For Research Use Only. Not for use in diagnostic procedures. BECKMAN COULTER WHITE PAPER 3
4 Fig 1: Instrument setup for PMT (Photomultiplier Tube) and Compensation Standardization on Navios Flow Cytometer (a) Initial setup (b) Setup for routine use (a) INITIAL SETUP (b) ROUTINE USE Set PMT Adjust PMT voltage to place negative populations of the single color stained conjugates in each channel at X-Median=0.3 ± 0.01 Record PMT for Standardized use Standardize PMTs Run Flow-Set PRO beads. Target intensities match requirements? NO Adjust PMT of Flow-Set Pro Protocol Adjust PMT voltage on Flow-Set Pro Standardized protocol to match target value requirements. Record New PMT settings. Flow-Set Pro Fluorospheres are run to acquire target intensity values for the established PMT settings. Settings saved as Flow Set Pro Standardized protocol. YES Adjust PMT settings of DuraClone Panel Protocol to match Flow-Set Pro Standardized protocol. Do not adjust compensation. Generate Compensation Matrix Use compensation tubes to generate matrix on the Auto Setup Scheduler for the Navios Flow Cytometer. Save Protocol as DuraClone Panel Protocol. Acquire Sample on DuraClone Panel Protocol Acquire Sample on DuraClone Panel Protocol BECKMAN COULTER WHITE PAPER 4
5 RESULTS Fig 2: Use of VersaComp Beads for dimly or rarely expressed antigen Example shown for compensation setup of DuraClone IM Dendritic cell panel, CD123-APC Alexa700 is dimly stained in blood. Therefore, VersaComp beads, spiked during staining of sample, were used as positive stained population during compensation setup. For PMT setup, negative of the Lymphocyte population was used to set X-Median at 0.3 Fig 3: DuraClone B Cell panel on Automated Compensation setup Figure 3 depicts typical gating strategy for the DuraClone IM B cell panel. All major B cell subtypes are identified with no adjustment in compensation. BECKMAN COULTER WHITE PAPER 5
6 Fig 4: DuraClone IM Dendritic cell panel on Automated Compensation Setup Figure 4 depicts DuraClone IM Dendritic cell panel when Autosetup is used to generate the compensation matrix. All circulating Dendritic cells are identified with no compensation adjustment required. Fig 5: Overlay of DuraClone IM B cell Panel across two Navios flow Cytometers Figure 5 depicts that the multicolor cocktail shows similar performance in terms of recruitment (i.e. % gated), across both of the flow cytometers, when the flow cytometers are setup using the DuraClone Standardized AutoSetup. BECKMAN COULTER WHITE PAPER 6
7 SUMMARY The use of the DuraClone reagent technology with the Navios AutoSetup Scheduler, or similar work tools, enables easy and accurate generation of a compensation matrix. The use of Flow-Set-Pro Fluorospheres and AutoSetup tools with the dry DuraClone reagent format enables standardization of high content multicolor flow cytometry. ORDERING INFORMATION Premixed, dry reagent cocktails to standardize and streamline flow cytometry workflow in clinical research studies involving the identification of principal populations of the human immune system. PN Panels FITC PE ECD PC5.5 PC7 APC A700 (1) APC- A700 (2) APC- A750 (3) Pacific Blue B53309 DuraClone IM Phenotyping Basic Tube CD16 CD56 CD19 CD14 CD4 CD8 CD3 CD45 B53318 DuraClone IM B cell Tube IgD CD21 CD19 CD27 CD24 CD38 IgM CD45 B53328* DuraClone IM T cell subsets Tube CD45RA CCR7 CD28 PD1 CD27 CD4 CD8 CD3 CD57 CD45 B53351* DuraClone IM Dendritic cell Tube CD16 Lineage (CD3/CD19/ CD20/CD14/CD56) CD1c CD11c Clec9A CD123 HLA-DR CD45 (1) Alexa Fluor*700 (2) APC-Alexa Fluor* 700 (3) APC-Alexa Fluor* 750 * PN B53328 and B53351 are available from December Available in convenient 25 test package. Each package includes three complete compensation kits with single color set of compensation reagents specific for each Panel. For Research Use Only. Not for use in diagnostic procedures. Navios is CE marked for 10-color in-vitro diagnostic use. In the U.S., Navios is intended for use as an in-vitro diagnostic device for immunophenotyping with Navios tetra software and CYTOSTAT tetrachrome reagents. All other uses are for research use only. * Pacific Blue and Alexa Fluor are registered trademarks of Molecular Probes, Inc. DuraClone, Navios, Kaluza, Flow-Set PRO Fluorospheres and are trademarks of Beckman Coulter, Inc. Beckman Coulter; the stylized logo are registered trademarks of Beckman Coulter, Inc and are registered in the USPTO. BECKMAN COULTER WHITE PAPER 7 FLOW-520WP10.14-A 2014 Beckman Coulter, Inc. All rights reserved.
a Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper Long Term Stabilization of Tandem Dyes for Use in High Content, Multi Variant Flow Cytometry Authors: Snehita Sattiraju 1, Tewfik Miloud 2, Neha Girish 1, Murthy
More informationApplication Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents
Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents Compensation setup for high content DuraClone reagents INTRODUCTION High content flow cytometry
More informationDURACLONE IM ACCELERATE YOUR PACE IN IMMUNE SYSTEM RESEARCH. For Reseach Use Only - Not for use in Diagnostic procedures
DURACLONE IM ACCELERATE YOUR PACE IN IMMUNE SYSTEM RESEARCH Your clinical research trial companion For Reseach Use Only - Not for use in Diagnostic procedures ACCELERATE YOUR PACE IN IMMUNE SYSTEM RESEARCH
More informationA previous ICCS Module entitled Instrument optimization - Adjusting PMT voltages and compensation 1 should be read as a prerequisite to this module.
Sponsored and reviewed by ICCS Quality and Standards Committee Title: Compensation Tips for Beckman Coulter 10-Color Navios Platform Written by: Salima Janmohamed-Anastasakis Ph.D., Applications Scientist,
More informationNavios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY
Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications optimized for the clinical
More informationNavios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY
Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications with optimized workflows for
More informationBeyond Colorful Science. A Practical Approach to Setting Up Multicolor Panels
Beyond Colorful Science A Practical Approach to Setting Up Multicolor Panels ISCT Paris April 2014 Layout: Multicolor Panel Design Identify the pitfalls Have a look at issues with spectral overlap How
More informationIncorporating New, Bright Fluorochromes into Multicolor Panel Design
Incorporating New, Bright Fluorochromes into Multicolor Panel Design Maria C. Jaimes, MD Senior Staff Scientist BD Biosciences 23-14684-00 Overview Multicolor flow: successful application prerequisites
More informationCLEARLLAB LS LYMPHOID SCREEN REAGENT
CLEARLLAB LS LYMPHOID SCREEN REAGENT CE MARKED ANTIBODY COMBINATION FOR LEUKEMIA / LYMPHOMA ANALYSIS Because Your Patient is Her Everything BECAUSE YOUR PATIENT IS HER EVERYTHING ClearLLab LS Lymphoid
More informationApplication Note. Assay Portability on the BD FACSVerse System. Summary. Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar
September Assay Portability on the BD FACSVerse System Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar Contents Summary Introduction 3 Objective 4 Methods 6 Results Discussion Conclusions
More informationWhat you need to know before designing a panel
Design What you need to know before designing a panel For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life Technologies Corporation.
More informationa Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper CytoFLEX Instrument Evaluation Using Biological Specimens Authors: James Tung 1, Dan Condello 3, Albert Donnenberg 4, Erika Duggan 3, Jesus Lemus 1, John Nolan
More informationPrinciples of Multicolor Panel Design BD. BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company.
1 Principles of Multicolor Panel Design 2 Common Multicolor Applications Intracellular cytokine staining Regulatory T cells (Tregs) Protein phosphorylation (BD Phosflow) Leukemia and lymphoma phenotyping
More informationYour Research, Revolutionized
Your Research, Revolutionized Drive Your Research Forward Your research needs are evolving and with the CytoFLEX flow cytometer you ll see just how far your data can take you. CytoFLEX has the advanced
More informationNavios Now Because Every Event Matters
Navios Now Because Every Event Matters Powerful, Dependable 10-Color Flow Cytometry Because we understand the power inside every cell, we designed the Navios flow cytometer to capture the information you
More informationYour Research, Revolutionized
Your Research, Revolutionized Drive Your Research Forward Your research needs are evolving and with the CytoFLEX flow cytometer you ll see just how far your data can take you. CytoFLEX has the advanced
More informationApplication Information Bulletin: Lymphocyte Subset Analysis
Application Information Bulletin: Lymphocyte Subset Analysis Use of Precision Profiles to Estimate Precision of Lymphocyte Subset Analysis by Flow Cytometry. Robert Magari 1, Diana, B. Careaga 1, Karen,
More informationAnatomy of a flow cytometer
Anatomy of a flow cytometer Fluidics Optics Electronics Cells in suspension flow in single-file through an illuminated volume where they scatter light and emit fluorescence that is collected, filtered,
More informationdetermine optimum instrument settings for their own instruments and establish their own daily values.
PC7 (770/488) SETUP KIT 6607121 PN 4299504-C FLOW CYTOMETER ALIGNMENT VERIFICATION FLUOROSPHERES FLOW CYTOMETER DETECTOR STANDARDIZATION FLUOROSPHERES INTENDED USE For Research Use Only. Not for use in
More informationNovoCyte Flow Cytometer
NovoCyte Flow Cytometer The Flow Cytometer for Everyone 2 Experience the NovoCyte Advantage Focus on advancing your research. Let the flow cytometer do the rest. NovoCyte Flow Cytometer High Performance
More informationHigh-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates
APPLICATION NOTE Attune NxT Flow Cytometer with Autosampler High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates Introduction The emerging field
More informationWelcome to More Choice. Mouse Panels
Welcome to More Choice Mouse Panels Choose from our extensive portfolio of high-quality fluorescent-conjugated reagents to build your multicolor flow cytometry panels. Welcome to a More Colorful World
More informationAQUIOS CL LOAD & GO AND MAKE FLOW CYTOMETRY SIMPLICITY ITSELF
LOAD & GO AND MAKE FLOW CYTOMETRY SIMPLICITY ITSELF FOCUS ON YOUR PATIENT. NOT ON SAMPLE PREPARATION Sample preparation and data management usually are the major bottlenecks in routine flow cytometry.
More informationNovel Developments in Cytometry. Martin Adelmann Marketing Manager Cellular Analysis & Life Science Beckman Coulter EMEAI
Novel Developments in Cytometry Martin Adelmann Marketing Manager Cellular Analysis & Life Science Beckman Coulter EMEAI 1 Innovation in Flow Cytometry Cellios EPICS 1V 2 lasers, 3 colors EPICS Profile
More informationQuality Control in polychromatic Flow Cytometry - EuroFlow interlaboratory standardization
IV. International Symposium Flow Cytometry/Molecular Biology HOSPITAL ISRAELITA ALBERT EINSTEIN São Paulo 2012 Quality Control in polychromatic Flow Cytometry - EuroFlow interlaboratory standardization
More informationSelecting Reagents for Multicolor Flow Cytometry
HotLines Platinum Edition f a l l 0 0 6 Selecting Reagents for Multicolor Flow Cytometry By Holden Maecker and Joe Trotter The availability of flow cytometers capable of detecting 6, 8, and more colors
More informationFlow Cytometry SOP: 14 color flow for immune activation, senescence, and exhaustion
Flow Cytometry SOP: 14 color flow for immune activation, senescence, and exhaustion Purpose This SOP standardizes the procedure for measuring immune cells using flow cytometry in ACTG Immunology Laboratories.
More informationDesigning and Implementing a High-Level Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington
Designing and Implementing a High-Level Multicolor Flow Cytometry Assay Brent Wood MD PhD Department of Laboratory Medicine University of Washington Define Purpose of Assay Most important question What
More informationUnderstanding Flow Cytometry
Understanding Flow Cytometry The Basic Concepts Maree Bagnara Products Sales Specialist/Account Manager Flow Cytometry PN775136 1 Successful Flow Cytometry data is driven by.. Understanding the Biology
More informationDetecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer
APPLICATION NOTE Attune Acoustic Focusing Cytometer Detecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer Circulating endothelial cells (CECs) are mature cells shed
More informationBD Multicolor CompBeads
4/2015 23-9955-01 IVD BD Multicolor CompBeads 100 compensation setups Catalog No. 644204 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2015 BD Becton, Dickinson and
More informationWelcome to More Choice. Human Panels
Welcome to More Choice Human Panels Choose from our extensive portfolio of high-quality fluorescent-conjugated reagents to build your multicolor flow cytometry panels. Welcome to a More Colorful World
More informationFlow Cytometry Immune Activation SOP
Flow Cytometry Immune Activation SOP Purpose This SOP standardizes the procedure for measuring immune activation of T cells using flow cytometry in ACTG Immunology Laboratories. Materials 1. 12x75mm flow
More informationImmunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A.
Immunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A. * Department of Haematology & Blood Transfusion,College of Health Science, Ladoke Akintola University
More informationFLOW CYTOMETRY. CyAn ADP. Analyzer
FLOW CYTOMETRY CyAn ADP Analyzer Experience the Power of the CyAn ADP and its optimal performance The Power of Detection The Power of Speed The Power of Ease The CyAn ADP Analyzer is the next step in Advanced
More informationBest practices in panel design to optimize the isolation of cells of interest
Sort Best practices in panel design to optimize the isolation of cells of interest For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life
More informationLONDON HEALTH SCIENCES CENTRE (LHSC) WORKFLOW COMPARISON STUDY WITH THE AQUIOS CL FLOW CYTOMETER
CASE WHITE STUDY PAPER LONDON HEALTH SCIENCES CENTRE (LHSC) WORKFLOW COMPARISON STUDY WITH THE AQUIOS CL FLOW CYTOMETER Laboratory Profile London Health Sciences Centre (LHSC), Flow Cytometry Laboratory,
More informationHigh-dimensional flow-cytometric analysis of human B-cell populations
High-dimensional flow-cytometric analysis of human B-cell populations The BD FACSCelesta cell analyzer and FlowJo software together enable deep analysis of B-cell biology Features High-resolution analysis
More informationApplication Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
More informationPrinciples of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting. Shoreline Community College BIOL 288
Principles of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting Shoreline Community College BIOL 288 Flow Cytometry What is Flow Cytometry? Measurement of cells or particles
More informationEach question may have MULTIPLE correct answers. Select all that are correct.
Knowledge Assessment Flow Cytometry Workshop, Part 1 April 20, 2015 Each question may have MULTIPLE correct answers. Select all that are correct. 1. Tandem dyes are a. highly stable fluorophores after
More informationThe NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee
January, 1999 From: To: Subject: The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee NIAID DAIDS Flow Cytometry Laboratories Comparison study information for labs wishing to switch
More informationCancer inflammation research applications and products
Cancer inflammation research applications and products Flow cytometry Immunoassays Cell imaging Instrumentation Invitrogen Attune NxT Flow Cytometer Antibodies RNA flow Conjugated antibodies for flow cytometry
More informationab pdc Subset Phenotyping Kit
ab177772 pdc Subset Phenotyping Kit Instructions for Use A set of antibodies designed to identify pdcs and subsets of pdcs by multi-color flow cytometry. This product is for research use only and is not
More informationFlow Cytometry SOP: Monocytes from Frozen Cells
Flow Cytometry SOP: Monocytes from Frozen Cells Purpose This SOP standardizes the procedure for measuring immune cells using flow cytometry in ACTG Immunology Laboratories. Materials 1. 12x75mm flow tubes
More informationExpanding Multicolor Options in Flow Cytometry with Novel Brilliant Violet TM Fluorophores. Carsten Wiethe Scientific Application Manager BioLegend
Expanding Multicolor Options in Flow Cytometry with Novel Brilliant Violet TM Fluorophores Carsten Wiethe Scientific Application Manager BioLegend Seminar Outline BrilliantViolet TM (BV)fluorophores Instrument
More informationBD FACSCanto II. A proven research platform for maximum reliability and the highest quality results
BD FACSCanto II A proven research platform for maximum reliability and the highest quality results A proven platform for maximum reliability and the highest quality results Built on more than 25 years
More informationUsing Fluorescence Spillover to Advantage
Using Fluorescence Spillover to Advantage Boston June 9, 2011 Compensation: Why and When is it Necessary? Clare Rogers, Marketing and Applications Accuri Cytometers, Inc. Ann Arbor/St.Ives 1 June 11 Summary
More informationIdentification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer. Application Note
Identification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer Application Note Sylvie Veriac Valérie Perrone Madeleine Avon Abstract Agilent Equipment: 2100 bioanalyzer
More informationFlow Cytometry - The Essentials
Flow Cytometry - The Essentials Pocket Guide to Flow Cytometry: 1. Know your Cytometer 2. Understanding Fluorescence and Fluorophores 3. Gating Process 4. Controls 5. Optimization 6. Panel Building 7.
More informationBeadless Absolute Counting
WHITE PPER Beadless bsolute Counting pplication of the unique properties of the peristaltic pump fluidic based system for volumetric cell counting Pierre Grenot, C.Cy (ISC), Hervé Luche, Ph.D., Centre
More informationBD FACSMelody. Cell Sorter. The simple solution for consistent, quality results
BD FACSMelody Cell Sorter The simple solution for consistent, quality results The right instrument is needed for the best results Simplicity, affordability and quality are key Cell sorting is fast becoming
More informationFlexible, Intuitive and Affordable
Data Sheet guava easycyte Flow Cytometry Systems Flexible, Intuitive and Affordable The guava flow cytometry systems are easy to use and deliver complete and comprehensive cell analysis right on your benchtop.
More informationNo-wash, no-lyse detection of phagocytic cells via a phrodo BioParticles functional assay in human whole blood on the
APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of phagocytic cells via a phrodo BioParticles functional assay in human whole blood on the Attune NxT Flow Cytometer Introduction Analysis
More informationCalibración de equipos de citometría para empleo de paneles EuroFlow
Calibración de equipos de citometría para empleo de paneles EuroFlow Departamento de Medicina, Centro de Investigación del Cáncer y Servicio de Citometría. Universidad de Salamanca, Salamanca, España.
More informationBoundary-breaking acoustic focusing cytometry
Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory
More informationKaluza Analysis Software SOPHISTICATED FLOW CYTOMETRY, SIMPLIFIED.
Kaluza Analysis Software SOPHISTICATED FLOW CYTOMETRY, SIMPLIFIED. SOPHISTICATED FLOW CYTOMETRY, SIMPLIFIED KALUZA ANALYSIS SOFTWARE As the complexity of flow cytometry increases, experiments require more
More information2. SUMMARY AND EXPLANATION
English Stem-Trol Control Cells 1-10 1. INTENDED USE 2 2. SUMMARY AND EXPLANATION 2 3. PRINCIPLE OF TEST 2 4. REAGENT CONTENTS 2 5. STATEMENT OF WARNINGS 2 6. STORAGE CONDITIONS AND STABILITY 3 6.1 Evidence
More informationPURPOSE: To delineate the subsets of human lymphocytes based on the expression profiles of different phenotypic markers by FACS analysis
LABORATORY PROCEDURE: IMMUNOPHENOTYPING: Lymphocyte Staining for FACS Analysis Date: April 29 2014 Authors: Jennifer Hossler PURPOSE: To delineate the subsets of human lymphocytes based on the expression
More informationFEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT
FEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY FLOW CYTOMETRY CD34+ STEM
More informationSCIENTIFIC INSTITUTE OF PUBLIC HEALTH QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT
SCIENTIFIC INSTITUTE OF PUBLIC HEALTH QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY FLOW CYTOMETRY CD34+
More informationAMREP FLOW CYTOMETRY CORE FACILITY. Flow Cytometry Data Analysis Workshop Friday 24 th October
AMREP FLOW CYTOMETRY CORE FACILITY Flow Cytometry Data Analysis Workshop Friday 24 th October Polychromatic Flow Cytometry Data Analysis: Immunophenotyping Lymphocyte sub-populations and targeting their
More informationPowERFUL. INTUITIvE. customizable. ThE NEw STAR of. BENchToP FLow cytometry
PowERFUL. INTUITIvE. customizable. ThE NEw STAR of BENchToP FLow cytometry ABoUT AcEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
More informationVisualization of digital data Interpretation of the visual
Technical aspects of 8-color flow cytometry in the diagnosis and classification of hematopoietic malignancies Tomas Kalina!"#$%&'()*+,&$'+-./(0 *1 (2#34%-.(56(7&1+3+*&/( 8$#94&/(!:&3"(;&
More informationCompensation: Fundamental Principles
Flow Cytometry Seminar Series 2017 : Fundamental Principles Spillover correction in multicolor flow cytometry 28.02.2017 http://www.cytometry.uzh.ch Contents Fluorescence and its detection Absorption and
More informationSelecting Fluorochrome Conjugates for Maximum Sensitivity
2004 Wiley-Liss, Inc. Cytometry Part A 62A:169 173 (2004) Technical Primer Selecting Fluorochrome Conjugates for Maximum Sensitivity Holden T. Maecker,* Tom Frey, Laurel E. Nomura, and Joe Trotter BD Biosciences,
More informationInterest in any of the products, request or order them at Bio-Connect Diagnostics.
CE IVD Minicatalog Interest in any of the products, request or order them at Bio-Connect Diagnostics. Bio-Connect Diagnostics B.V. T NL +31 (0)26 326 44 60 T BE +32 (0)2 502 12 53 Begonialaan 3a F NL +31
More informationAttune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry
Attune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry Maybelline Giam Field Application Scientist The world leader in serving science Attune NxT Flow Cytometer Attune NxT Acoustic
More informationACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man CD38 Quantitation 07 May 2004
1. Principle Flow Cytometric quantitation of CD38 expression on CD8+ T lymphocytes The purpose of this assay is to quantitate the surface expression of the CD38 molecule on CD8 positive T lymphocytes.
More informationA high recovery and high purity data analysis strategy for rare abnormal plasma cell events using the DuraClone RE PC Tube
APPLICATION NOTE A high recovery and high purity data analysis strategy for rare abnormal plasma cell events using the DuraClone RE PC Tube Dr Agnieszka Blum Charité Virchow Hospital, Stem Cell Facility,
More informationMinimum Information about a Flow Cytometry Experiment (MIFlowCyt) Annotation
Minimum Information about a Flow Cytometry Experiment (MIFlowCyt) Annotation 1. Experiment Overview 1.1 Purpose The purpose of these sets of experiments is to develop a methodology of identifying and quantifying
More informationResolving Stem Cell Heterogeneity Using Flow Cytometry
Resolving Stem Cell Heterogeneity Using Flow Cytometry Mirko Corselli, PhD Senior Scientist We have not fully utilized the power of flow cytometry to address biological questions in other systems Flow
More informationBD Accuri C6. Flow cytometry within reach.
BD Accuri C6 Flow cytometry within reach. Flow Cytometry Within Reach The BD Accuri C6 is a personal flow cytometer that brings flow cytometry within reach by being easy to use, simple to maintain, and
More informationRuud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth
Sponsored and reviewed by ICCS Quality and Standards Committee Title: Quality of Reagents Monoclonal Antibodies Written by: Date: March 6, 2018 Ruud Hulspas, Mike Keeney, Ben Hedley and Andrea Illingworth
More informationefluor Organic Dyes 450/50 BP Fluorescence Intensity Wavelength (nm)
efluor Organic Dyes efluor Organic Dyes Catalog No. Description Clone Application Anti-Mouse efluor 450 Products 48-0042 Anti-mouse CD4 RM4-5 Flow Cytometry 48-0081 Anti-mouse CD8 53-6.7 Flow Cytometry
More informationTECHNICAL BULLETIN. QUANTUM SIMPLY CELLULAR KIT Product Numbers QSC-20 AND QSC-100 Storage Temperature 2-8 C. Do Not Freeze
QUANTUM SIMPLY CELLULAR KIT Product Numbers QSC-20 AND QSC-100 Storage Temperature 2-8 C. Do Not Freeze TECHNICAL BULLETIN Product Description The Quantum Simply Cellular Kit provides a convenient method
More informationa Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper Flow Cytometric Analysis of Endothelial Progenitor Cells Authors: Affiliation: Dorota Sadowicz, Vasilis Toxavidis, John Tigges Beth Israel Deaconess Medical
More informationMulticolor flow cytometry analysis of human pluripotent stem cell cultures
Multicolor flow cytometry analysis of human pluripotent stem cell cultures ackground Thorough characterization of pluripotency is mandatory during the establishment of newly derived human embryonic or
More informationBD FACSCanto II Flow Cytometry System
Regional Offices Asia Pacific BD Singapore Tel 65.6861.0633 Fax 65.6860.1593 Australia/New Zealand Australia Toll Free: 1800 656 100 Tel 61.2.8875.7000 Fax 61.2.8875.7200 bd_anz@bd.com New Zealand Toll
More informationBD LSRFortessa. Performance without peer, choice without compromise
BD LSRFortessa Performance without peer, choice without compromise Performance without peer, choice without compromise The BD LSRFortessa cell analyzer offers the ultimate in choice for flow cytometry,
More informationSA3800 Spectral Analyzer
SA38 Spectral Analyzer Sony Biotechnology Inc. SA38 Spectral Analyzer The Sony SA38 spectral cell analyzer incorporates advanced electronics, patented optical technologies and many points of automation
More informationSpherotech, Inc. 1. SPHERO TM Technical Note STN-8 Rev C
SPHERO TM Technical Note STN- Rev C. 0070 CALIBRATION AND PERFORMANCE TRACKING OF FLOW CYTOMETERS USING SPHERO TM CALIBRATION PARTICLES Introduction The SPHERO TM Calibration Particles are versatile, stable,
More informationInhibiScreen Kinase Inhibitor Assay
InhibiScreen Kinase Inhibitor Assay Flow Cytometric assessment of efficacy of PI3Kγ, PI3Kδ, BTK and SYK pathway inhibitors using the measurement of basophil activation in human EDTA and Heparin Whole Blood.
More informationGUAVA EASYCYTE SYSTEMS
GUAVA EASYCYTE SYSTEMS Expanding the potential of flow cytometry The life science business of Merck KGaA, Darmstadt, Germany operates as MilliporeSigma in the U.S. and Canada. Guava easycyte Systems Expanding
More informationFor detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe.
ab112133 JC-10 Mitochondrial Membrane Potential Assay Kit Flow Cytometry Instructions for Use For detecting mitochondrial membrane potential changes in cells using our proprietary fluorescence probe. This
More informationChallenges for Flow Cytometry in Regulated Bioanalysis
Challenges for Flow Cytometry in Regulated Bioanalysis Minesh Patel Merck Millipore Discovery & Development Solutions. Oxford, UK. Overview Flow cytometry principles Current uses and regulatory environments
More informationLower spectral overlap. Optimized stain index. Innovative antibodies. Superior multicolor flow cytometry. Violet Laser Reagents: The efluor Solution
Lower spectral overlap Optimized stain index Innovative antibodies Superior multicolor flow cytometry Violet Laser Reagents: The efluor Solution www.ebioscience.com ebioscience provides innovative high
More informationApplication Note. Introduction. Robbie Narang, Zhaoping Liu, Kim Luu IntelliCyt Corporation
pplication Note High Throughput Flow ssays using the MultiCyt Cell Proliferation Reagent Dye Panel: Multiplexing with Membrane Integrity, Cell Cycle, and Immunophenotyping Endpoints Robbie Narang, Zhaoping
More informationBD FACSCalibur. The flow cytometer for your routine cell analysis needs
BD FACSCalibur The flow cytometer for your routine cell analysis needs A system with a rich application basis and a modular approach that continues to meet evolving needs of cell analysis worldwide. The
More informationRef: CYT-MM-MRD. For Research Use Only. Not for use in diagnostic procedures
Ref: CYT-MM-MRD For Research Use Only. Not for use in diagnostic procedures MM MRD VIALS AND COMPENSATION TUBES ARE LYOPHILIZED PRODUCTS. READ CAREFULLY THE FOLLOWING INSTRUCTIONS FOR RECONSTITUTION: The
More informationPERFECT-COUNT MICROSPHERES
PERFECT-COUNT MICROSPHERES Perfect-Count Microspheres-Product code PCB-100 for 100 tests Introduction In recent years, the determination of absolute cell counts has been shown to be relevant in different
More informationFlow Cytometry in the Diagnosis of Hematopoietic Neoplasia. Brent Wood MD, PhD Professor, Laboratory Medicine University of Washington, Seattle
Flow Cytometry in the Diagnosis of Hematopoietic Neoplasia Brent Wood MD, PhD Professor, Laboratory Medicine University of Washington, Seattle 1 Flow Cytometer 2 The Power of Flow Cytometry Single cell
More informationDA-Cell Washer and Plate for Flow Cytometry. Accelerating Life Sciences
DA-Cell Washer and Plate for Flow Cytometry Accelerating Life Sciences Conventional Centrifuge-based protocol for staining cells Challenges of staining cells for flow cytometry by centrifuge Losing cells
More informationFlow Cytometric Devices Draft Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 36 Flow Cytometric Devices Draft Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE This guidance document
More informationPractical Guidelines and Tips for Sensitive and Accurate Identification of PNH Clones
Practical Guidelines and Tips for Sensitive and Accurate Identification of PNH Clones AFCG Meeting Nov 28 Dec 1, 2013 Wellington, New Zealand Andrea Illingworth, MS, H(ASCP), QCYM Dahl-Chase Diagnostic
More informationGuava EasyCyte Systems Expanding the potential of flow cytometry
Guava EasyCyte Systems Expanding the potential of flow cytometry Guava easycyte Systems Flexible Intuitive Affordable Unleash what s possible Fifteen years ago, Guava Technologies introduced the first
More informationGuava easycyte Systems Expanding the potential of flow cytometry
Guava easycyte Systems Expanding the potential of flow cytometry Guava easycyte Systems Flexible Intuitive Affordable Merck Millipore is a business of Unleash what s possible Fifteen years ago, Guava Technologies
More informationSelected Topics in Electrical Engineering: Flow Cytometry Data Analysis
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Experimental design
More informationFEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT
FEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY FLOW CYTOMETRY CD34+ STEM
More informationSA3800 Spectral Analyzer
SA38 Spectral Analyzer Sony Biotechnology Inc. SA38 Spectral Analyzer The Sony SA38 spectral cell analyzer incorporates advanced electronics, patented optical technologies and many points of automation
More information