1 Attune NxT Acoustic Focusing Cytometer The next generation in acoustic cytometry Maybelline Giam Field Application Scientist The world leader in serving science
2 Attune NxT Flow Cytometer Attune NxT Acoustic Focusing Cytometer Acoustic-assisted hydrodynamic focusing to achieve freakishly fast flow rates 2
3 Introduction to Flow Cytometry Topics Module 1: Flow Cytometry Principles 3
4 What is Flow Cytometry? CYTOMETRY is the measurement of physical or chemical characteristics of cells or particles FLOW CYTOMETRY measurements are made as cells or particles in suspension pass individually through a flow cytometer instrument Performed on single cell suspensions Provides discrete measurements from each cell Provides a distribution of the measured characteristics 4
5 Advantages of flow cytometry Single cell data (great for heterogeneous populations) Measurements on large numbers of cells Rich statistical analysis on cell populations Ability to multiplex Thousands of cells can be analyzed rapidly Small sub-populations of cells can be identified In specific instruments, cells can be physically sorted. Sub-populations can be purified for further study 5
6 Some applications of flow cytometry Immunophenotyping (eg. B/T/NK cells in blood, immune organs) Immune assays (eg. T cell killing, phagocytosis) Cytokine production Stem cell characterisation (eg. CD34 HSCs) Analysis of transfected cells (eg. % GFP positivity) Phenotypic assays (eg. proliferation, apoptosis, ROS) Drug screening Protein expression (even phospho-proteins) RNA expression (eg. PrimeFlow) Small particle analysis (eg. bacterial, yeast) 6
7 The Flow Cytometer A Different kind of Microscope Data Analysis Laser Sample Flow cytometry is similar to a microscope. Microscope produces an image of a cell Light A flow cytometer collects and quantifies scattered light and fluorescence. The output produced is numbers. 7
8 8 Comparing flow cytometry and microscopy
9 Flow Workflow Sample prep Blood, tumor, cells Staining Flow Run 9
10 Parameters that Flow cytometry can measure Relative size Cell complexity/granularity Fluorescence Amount of DNA Levels of proteins (on cell surface or intracellular) 10
11 When the light hits the cells, two phenomena occur Before we talk about absorption and fluorescence emission, there is scattering of light! When laser light interacts with a cell, light is scattered in all directions. The light scatter depends on size and internal complexity of the cell. laser 11
12 Light scattering in human whole blood Large and granular Large, less granular Small, not granular 12
13 Light scattering differs when cells have different physiology Aged culture of Jurkat T cells: green are live cells & red are dead cells 13
14 What if 2 cell types are the same size and complexity? Side Scatter (SSC) Neutrophils Lymphocytes Monocytes Forward Scatter (FSC) 14
15 What if 2 cell types are the same size and complexity? CD19 CD3 B T Use cell markers to differentiate between cell types Markers can be: intracellular enzymes, surface receptors, adhesion molecules ANYTHING A FLUORESCENT DYE OR ANTIBODY CAN DETECT 15
16 Separate T and B lymphocytes by their cell surface markers B Cells Side Scatter (SSC) Neutrophils Lymphocytes Monocytes CD19 PE T Cells Forward Scatter (FSC) CD3 Alexa Fluor
17 What is spillover and compensation? BL1 BL2 17
18 What is spillover and compensation? BL1 BL2 What is the real R-PE fluorescence? Need to subtract A488 signal 18
19 What is spillover and compensation? Compensation is the process which corrects the detected "spillover" of the emission of one fluorochrome into the detector designed to collect the emission from another fluorochrome. The primary purpose is to allow the measurement of the true fluorescence in the fluorescence channel contaminated by the spillover. Use single color samples are used to set compensation 19
20 What is spillover and compensation? Un-Compensated Alexa 488 Single Color Control Alexa 488 is being detected by the PE channel as well! 20
21 What is spillover and compensation? 21
22 What is spillover and compensation? Properly Compensated Alexa 488 Single Color Control 22
23 Un-Compensated 1e6 Un-compensated FITC PE APC PerCP-Cy5.5 4_Intensity 1e5 1e Emission Wavelength (nm) Detector (nm) e4 1e5 3_Intensity FITC 1e6 Un-compensated PE 1e5 6_Intensity 1e4 APC e4 1e5 1e6 5_Intensity PerCP-Cy5.5 Data Courtesy Cancer Research UK 23
24 Compensated 1e6 Compensated FITC PE APC PerCP-Cy5.5 1e5 4_Intensity 1e e4 1e5 3_Intensity Emission Wavelength (nm) Detector (nm) FITC 1e6 Compensated PE 1e5 6_Intensity 1e e4 1e5 5_Intensity APC PerCP-Cy5.5 Data Courtesy Cancer Research UK 24
25 Auto-compensation in the Attune NxT 25
26 Auto-compensation in the Attune NxT 26
27 Principles of Panel Design 1. Understand fluorochromes and know your instrument. 2. Check for reagent availability. 3. Match fluorochromes by brightness 4. Minimize spillover 5. Use appropriate controls 27
28 Introduction to Flow Cytometry Topics Module 2: Flow Cytometer Hardware 28
29 What does a flow cytometer consist of? Flow Cytometry is a technology that simultaneously measures and analyses multiple characteristics of single particles as they flow in a fluid stream through a beam of light. 1. Fluidics - To introduce the cells for interrogation with laser 2. Optics - To generate and collect the light signals 3. Electronics - To convert the optical signals to electronic signals 29
30 What does a flow cytometer consist of? 30
31 Traditional flow cytometers use only hydrodynamic focusing Particle positioning in laser is important Low sample flow rate (e.g., 12 µl/min) Hydrodynamic core High sample flow rate (e.g., 200 µl/min) Focused laser Focused laser Narrow particle focus = Narrow distribution Broad particle focus = Broad distribution Intensity sheath sheath sheath sheath Count Count Intensity 31
32 Attune NxT uses Acoustic Assisted hydrodynamic focusing Hydrodynamic Sample Core High sample flow rate > 100 µl/min Acoustic Acoustic Focusing Assisted Hydrodynamic Followed Focusing by Hydrodynamic Focusing Focused laser Narrow Data distribution Sheath Count Intensity 32
33 Acoustic focusing End-on view of capillary 33
34 Comparable Results at All Flow Rates Traditional Cytometers Attune NxT µl/min µl/min 200 Up to10x Faster than Traditional Cytometers 12 µl/min 35 µl/min 60 µl/min 12.5 µl/min 25 µl/min 100 µl/min 200 µl/min 500 µl/min 1,000 µl/min CV= 4.83 CV=6.12 CV=7.76 CV= 2.99 CV= 3.03 CV= 2.76 CV= 2.94 CV= 2.70 CV= 2.96 Hydrodynamic Focusing Only Acoustically Enhanced Hydrodynamic Focusing 34
35 Acoustic Focusing Dilution is the solution 35
36 Rare Event Detection How many events are enough? For an acceptable level of significance, 5% CV, you need 400 positive cells In a sample with 1 x 10 6 cells/ml Frequency Total events Time for speed = 100ul/min Time for speed = 1000ul/min 10% 4, % 40, % 400, % 4,000, % 40,000,
37 Some applications involving Rare Cell Analysis Counting of CD34+ hematopoietic progenitor cells Data courtesy of Prof. David Cousins, University of Leicester Dendritic cell analysis (eg. pdcs) 37
38 Sample preparation workflow for whole blood analysis Collect Sample Generic sample preparation workflow Total sample prep time ~ hrs Process to Single cell Suspension 5-10 min Lyse red blood cells or enrich for white blood cells 30-60min Stain min Wash 5-10 min Wash 5-10 min Wash 5-10 min Add 1-2 hours for intracellular markers Acquire Samples No Lyse/No Wash sample preparation workflow Collect Sample dilute Lyse red X blood cells or enrich for Stain white blood min cells 30-60min Wash X 5-10 min Wash X 5-10 min Acquire Samples 38
39 Intact whole blood detection of WBC Rational: Hemoglobin in RBCs absorbs light at 405nm while WBC and Platelets do not, allowing resolution using light scatter alone (no antibody staining needed) Strategy: Install Attune NxT No-Wash, No-Lyse Filter kit to allow dual-laser light scatter (SSC) detection using Blue and Violet laser 39
40 Results WBC detection in whole blood Normal FSC/SSC 40
41 Results WBC detection in whole blood SSC/VSSC Normal FSC/SSC 41
42 Results WBC detection in whole blood SSC/VSSC Normal FSC/SSC SSC (Violet 488 nm) (x10 6 ) FSC (Blue 488 nm) (x10 6 ) 42
43 Combining NWNL with unparalleled speed of Attune NxT No Lyse/No Wash assessment of phagocyte function in human blood 1. Incubate whole blood with phrodo BioParticles 2. Stain with Vybrant DyeCycle Ruby (threshold on nucleated WBCs) 3. Dilute and analyse on Attune NxT (flow rates 200µl/min) phrodo BioParticle Conjugates: ph-sensitive reagents that fluoresce upon their ingestion into acidic phagosomes 43
44 Combining NWNL with unparalleled speed of Attune NxT No Lyse/No Wash assessment of phagocyte function in human blood 44
45 On Board Fluidics Fluid storage: All fluids stored within instrument with active fluid level sensing Fluid Lines Sensor Connections Focusing Fluid Filters Standard fluidic tanks: 1.8 L focusing fluid tank 1.8 L waste tank 175 ml shutdown solution tank 175 ml wash solution tank Nominal fluid consumption: 1.8 L/day Waste Focusing Fluid Wash Fluid Shutdown Fluid Easy access to tanks for refilling 45
46 Smooth and Accurate Sample Delivery Sample delivery: Sample delivered by positive displacement syringe pump for volumetric analysis Volumetric cell counts: Live/dead analysis Sample analysis volume: 20 µl 4 ml Sample rates: ,000 µl/min Unused sample return option: Sample returned into plate or tubes Syringe Pump Wide dynamic range Long life = 8 million cycles Easy syringe replacement 46
47 Attune NxT Cytometer is Clog Resistant Attune NxT uses a positive displacement syringe pump Samples are pushed through the system with a direct displacement (and volumetric) pump Pump pushes potential clogs through with pressure that builds steadily if there is a restriction Attune NxT flow cell is 200 um in diameter. This is relatively large in the industry Attune NxT flow path is back-flushed after every sample Back-flush is into an orifice that is larger than the sample path Any accumulation in the forward path is cleared back into a larger orifice after every sample. 47
48 What does a flow cytometer consist of? 48
49 Optical Configuration 49
50 Flat Top Lasers Maintain High Resolution When Stream Shifts Flat System top provides is misaligned stable and optical producing alignment large CV s Flow Cell 15 um beam shift 10 um Core Stream Measured CV > 1.7% 1.1% 10% Gaussian Laser Profile Plus: Lasers Only on During Acquisition! 10X Increase in Lifetime 50
51 51 Violet Blue Yellow Red - Standard Configuration 417LP 654LP 577LP Blank Blank Blank 710/50 780/60 780/60 Blank 503LP 695/40 603/48 Blank 695/40 512/25 720/30 620/15 440/50 670/14 585/16 Blank 590/40 530/30 488/10+OD2
52 695/40 574/26 530/30 488/10+OD2 52 Attune NxT Blue Violet Configuration Blue laser (488nm) 780/60 Blank 503LP 780/60 695/40 574/26 530/30 488/10
53 Utilizing four excitation sources limits spillover Having four excitation sources allows the user to select spectrally distinct fluorochromes, thus limiting spillover. CD62L Pacific Blue TM Violet 405 nm Blue 488 nm CD8 FITC CD4 PerCP Cy 5.5 CD56 PE Yellow 561 nm Red 638 nm CD3 APC-Alexa Fluor
54 Utilizing the full spectrum limits the need for compensation CD3 APC-Alexa Fluor 750 (RL-3) αβt cells NKT cells NK cells CD62L Pacific Blue TM (VL-1) CD62L Pacific Blue TM (VL-1) CD56 PE (YL-1) CD8 FITC (BL-1) CD4 PerCP Cy 5.5 (BL-3) Naïve and Central Memory Effector Memory Compensation Matrix Pacific Blue TM (VL-1) FITC (BL-1) PerCP Cy 5.5 (BL-3) PE (YL-1) APC Alexa Fluor 750 (RL-3) Pacific Blue TM (VL-1) FITC (BL-1) PerCP Cy 5.5 (BL-3) PE (YL-1) APC Alexa Fluor 750 (RL-3)
55 What does a flow cytometer consist of? 55
56 Electronics Convert optical signals (photons) to proportional electronic signals (voltage pulses) Electronic signals are processed by the on-board processor Analyze voltage pulse height, area, and width Interface with the computer for data transfer Schematic of a photomultiplier tube. The photon is converted to an electron when it enters the PMT at the cathode. It then travels through the PMT, being amplified at the dynodes (electrodes) throughout and end up at the anode which is a collecting electrode. 56
57 Electronics Sample Flow Voltage Pulse in PMT Laser Voltage Time Laser Voltage Time Laser Voltage Time 57
58 Adjust PMT Voltages Expand the Voltage section Adjust FSC & SSC voltages to position cell population on the scatter plot Adjust Fluorescence Channels voltages to position the auto fluorescence signal (unstained population) to ~10 3 Notes: If compensation controls are recorded, all the fluorescence channels voltages are disabled (i.e. grayed out) 58
59 Optimizing PMT voltage If PMT too low, might not be able to differentiate dim population from bright population If PMT too high, background will be very high and you might have spillover from other dyes PMTs need to be just right! Voltage walking 59
60 Sample Presentation: Voltage Pulse Pulse Area Volts Pulse Height 0 Pulse Width Time (µ Seconds) Attune default: All parameters (A-H-W) on all channels 60
61 Doublet discrimination In most cases, data analysis should include gating on single cells. Three Methods of Determination Pulse width double pulse height Pulse width double pulse area Linear Relationship of Pulse Height to Area 61
62 Thresholds Electronic threshold can be used to limit the number of events that the flow cytometer acquires Only signals with an intensity greater than or equal to the threshold will be processed and analysed X X 62
63 Introduction to Flow Cytometry Topics Module 3: Attune NxT Software 63
64 Main Application Workspace Ribbons Experiment Explorer Tabs Workspace 64
65 Experiment Explorer User name Experiment name Compensation controls Group name Sample name (highlighted in orange = active) 65
66 Tabs Instrument Settings Tab Select and label channels Change PMT voltage 66
67 Two Ribbons of Note Workspace Ribbon Plot Options Gating Tools Statistics box Format Layout Instrument Ribbon Fluidic functions Access PT results 67
68 Histograms A histogram tells us the number of events (or cells) at a particular measurement value 68
69 Dual Parameter Plots Dot Plots Back-gating 69
70 Dual Parameter Plots Density Plots Precedence density plots (Density + Back-gating) 70
71 Sequential gating to identify populations Gate Sample name Y axis X axis Statistics Box 71
72 Easy data comparison using overlays Selected plots within a single experiment are superimposed to allow direct visual comparison of data Single parameter: Histograms Percent of Max Dual parameter: Dot/Density Plot 72
73 Easy data comparison using overlays Selected plots within a single experiment are superimposed to allow direct visual comparison of data Single parameter: Histograms Percent of Max Dual parameter: Dot/Density Plot 73
74 Analysis using heatmaps 74
75 Shutdown with one click 75
76 Attune NxT Acoustic Focusing Cytometer Modular Design in a small footprint 4 lasers with 16 detection channels FAST Unprecedented Acoustic Focusing Technology Highest speed is ~10X faster than max speed of traditional flow cytometers DOES NOT CLOG Robust performance Reduced Clogging and Flat top lasers (less downtime) Simplifies sample prep CELL COUNTING Volumetric, no beads Sample Recovery High-throughput Plate Autosampler Compatible with standard and deep well 96/384-well plates 76
77 77
78 Questions?
Boundary-breaking acoustic focusing cytometry
Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory
More informationHigh-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates
APPLICATION NOTE Attune NxT Flow Cytometer with Autosampler High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates Introduction The emerging field
More informationAttune TM Acoustic Focusing Cytometer Training. Manik Punj Attune Training
Attune TM Acoustic Focusing Cytometer Training Manik Punj Attune Training Attune Training Agenda Section 1 An Introduction to Flow Cytometry Section 2 An Introduction to Acoustic Focusing Hydrodynamic
More informationIntroduction to Flow Cytometry. -- BD FACSCanto II TM. Daisy Kuo Application Specialist BDBiosciences
Introduction to Flow Cytometry -- BD FACSCanto II TM Daisy Kuo Application Specialist E-mail: daisy_kuo@bd.com BDBiosciences Outline Basic Concept of Flow Cytometry FACSCanto II System Introduction Application
More informationPrinciples of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting. Shoreline Community College BIOL 288
Principles of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting Shoreline Community College BIOL 288 Flow Cytometry What is Flow Cytometry? Measurement of cells or particles
More informationCancer inflammation research applications and products
Cancer inflammation research applications and products Flow cytometry Immunoassays Cell imaging Instrumentation Invitrogen Attune NxT Flow Cytometer Antibodies RNA flow Conjugated antibodies for flow cytometry
More informationFlow Cytometry - The Essentials
Flow Cytometry - The Essentials Pocket Guide to Flow Cytometry: 1. Know your Cytometer 2. Understanding Fluorescence and Fluorophores 3. Gating Process 4. Controls 5. Optimization 6. Panel Building 7.
More informationNo-wash, no-lyse detection of phagocytic cells via a phrodo BioParticles functional assay in human whole blood on the
APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of phagocytic cells via a phrodo BioParticles functional assay in human whole blood on the Attune NxT Flow Cytometer Introduction Analysis
More information11/19/2013. Janine Zankl FACS Core Facility 13. November Cellular Parameters. Cellular Parameters. Monocytes. Granulocytes.
DEPARTEMENT BIOZENTRUM Janine Zankl FACS Core Facility 13. November 2013 Cellular Parameters Granulocytes Monocytes Basophils Neutrophils Lymphocytes Eosinophils Cellular Parameters 1 What Is Flow Cytometry?
More informationEach question may have MULTIPLE correct answers. Select all that are correct.
Knowledge Assessment Flow Cytometry Workshop, Part 1 April 20, 2015 Each question may have MULTIPLE correct answers. Select all that are correct. 1. Tandem dyes are a. highly stable fluorophores after
More informationflow cytometry reinvented
Flow cytometry flow cytometry reinvented The Attune Acoustic Focusing Cytometer Precision and sensitivity at all speeds for rare events and precious samples The Attune Acoustic Focusing Cytometer Precise
More informationAnatomy of a flow cytometer
Anatomy of a flow cytometer Fluidics Optics Electronics Cells in suspension flow in single-file through an illuminated volume where they scatter light and emit fluorescence that is collected, filtered,
More informationNovoCyte Flow Cytometer
NovoCyte Flow Cytometer The Flow Cytometer for Everyone 2 Experience the NovoCyte Advantage Focus on advancing your research. Let the flow cytometer do the rest. NovoCyte Flow Cytometer High Performance
More informationPowERFUL. INTUITIvE. customizable. ThE NEw STAR of. BENchToP FLow cytometry
PowERFUL. INTUITIvE. customizable. ThE NEw STAR of BENchToP FLow cytometry ABoUT AcEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
More informationIntroduction to. BD FACSAria TM Cell Sorter. Flow = Fluid Cyto = Cell Metry = Measurement
What is Flow Cytometry? Introduction to BD FACSAria TM Cell Sorter Flow = Fluid Cyto = Cell Metry = Measurement BD Biosciences Application Specialist 產品應用專員 Daisy Kuo 郭正佼 A variety of measurements are
More informationINTRODUCTION TO FLOW CYTOMETRY
DEPARTEMENT BIOZENTRUM INTRODUCTION TO FLOW CYTOMETRY F ACS C ore F acility Janine Zankl FACS Core Facility 3. Dezember 2015, 4pm Cellular Parameters Granulocytes Monocytes Basophils Lymphocytes Neutrophils
More informationIntroduction to Flow Cytometry. -- BD FACSCanto II TM. Daisy Kuo Assistant Product Manager BDBiosciences
Introduction to Flow Cytometry -- BD FACSCanto II TM Daisy Kuo Assistant Product Manager E-mail: daisy_kuo@bd.com BDBiosciences Outline Basic Concept of Flow Cytometry FACSCanto II System Introduction
More informationBASICS OF FLOW CYTOMETRY
BASICS OF FLOW CYTOMETRY AUTHOR: Ana Isabel Vieira APPROVAL: Henrique Veiga Fernandes Ana Sílvia Gonçalves SOP.UCF.002 03-09-2015 Pag. 1/9 Overview Flow: Fluid Cyto: Cell Metry: Measurement Flow cytometry
More informationDetecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer
APPLICATION NOTE Attune Acoustic Focusing Cytometer Detecting human circulating endothelial cells using the Attune Acoustic Focusing Cytometer Circulating endothelial cells (CECs) are mature cells shed
More informationANAT 3231 Cell Biology Lab12 Stem Cell Analysis
ANAT 3231 Cell Biology Lab12 Stem Cell Analysis 2 June 2010 Dr Antonio Lee Neuromuscular & Regenera9ve Medicine Unit School of Medical Sciences, UNSW Introduction to Flow Cytometry Contributed by Vittoria
More informationFlowcytometry Dirk Pacholsky
Flowcytometry Dirk Pacholsky Flowcytometry: Overview 1 2 Overview Flow Cyto Metry Fluid Cell Measurement measuring cell properties of cells in suspension Most Flow Cytometer have Spatially separated Lasers
More informationApplication Note. Assay Portability on the BD FACSVerse System. Summary. Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar
September Assay Portability on the BD FACSVerse System Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar Contents Summary Introduction 3 Objective 4 Methods 6 Results Discussion Conclusions
More informationFlexible, Intuitive and Affordable
Data Sheet guava easycyte Flow Cytometry Systems Flexible, Intuitive and Affordable The guava flow cytometry systems are easy to use and deliver complete and comprehensive cell analysis right on your benchtop.
More informationFLOW CYTOMETRY. CyAn ADP. Analyzer
FLOW CYTOMETRY CyAn ADP Analyzer Experience the Power of the CyAn ADP and its optimal performance The Power of Detection The Power of Speed The Power of Ease The CyAn ADP Analyzer is the next step in Advanced
More informationPrinciples of Multicolor Panel Design BD. BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company.
1 Principles of Multicolor Panel Design 2 Common Multicolor Applications Intracellular cytokine staining Regulatory T cells (Tregs) Protein phosphorylation (BD Phosflow) Leukemia and lymphoma phenotyping
More informationBest practices in panel design to optimize the isolation of cells of interest
Sort Best practices in panel design to optimize the isolation of cells of interest For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life
More informationDesigning and Implementing a High-Level Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington
Designing and Implementing a High-Level Multicolor Flow Cytometry Assay Brent Wood MD PhD Department of Laboratory Medicine University of Washington Define Purpose of Assay Most important question What
More informationWhat you need to know before designing a panel
Design What you need to know before designing a panel For Research Use Only. Not for use in diagnostic or therapeutic procedures. Alexa Fluor is a registered trademark of Life Technologies Corporation.
More informationApplication Note. Yeast Analysis with a life - dead staining kit (Yeast control - viability)
Application Note Yeast Analysis with a life - dead staining kit (Yeast control - viability) (CyFlow Cube 8 and CyFlow Cube 6 with CyView TM version 1.5 CyFlow Space) Introduction The Partec Yeast control
More informationSpherotech, Inc. 1. SPHERO TM Technical Note STN-8 Rev C
SPHERO TM Technical Note STN- Rev C. 0070 CALIBRATION AND PERFORMANCE TRACKING OF FLOW CYTOMETERS USING SPHERO TM CALIBRATION PARTICLES Introduction The SPHERO TM Calibration Particles are versatile, stable,
More informationTheory and Applications for Cell Sorter
Theory and Applications for Cell Sorter Tony Liu 劉聖德 Product Manager BD Biosciences Taiwan Sep. 07, 2016 1 Outline Basic Concept of Flow Cytometry FACSAria II System Introduction Cell Sorting Theory Application
More informationAMREP FLOW CYTOMETRY CORE FACILITY. Flow Cytometry Data Analysis Workshop Friday 24 th October
AMREP FLOW CYTOMETRY CORE FACILITY Flow Cytometry Data Analysis Workshop Friday 24 th October Polychromatic Flow Cytometry Data Analysis: Immunophenotyping Lymphocyte sub-populations and targeting their
More informationThe Basics of Flow Cytometry
The Basics of Flow Cytometry F ACS C ore F acility Janine Bögli, Biozentrum, 29. January 2018 The functions of the FACS Core Facility Centralization of equipment and expertise Train users Sorter operation
More informationBasic Principles in Flow Cytometry. Flow Cytometry
Basic Principles in Flow Cytometry Flow Cytometry» Flow Cytometry is the technological process that allows for the individual measurements of cell fluorescence and light scattering. This process is performed
More informationFlow Cytometry. Flow Cytometry Basics Guide
Flow Cytometry Flow Cytometry Basics Guide Table of Contents Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Principles of the Flow Cytometer Fluidics System.... 3 Optics and Detection.... 4 Signal and
More informationUsing Fluorescence Spillover to Advantage
Using Fluorescence Spillover to Advantage Boston June 9, 2011 Compensation: Why and When is it Necessary? Clare Rogers, Marketing and Applications Accuri Cytometers, Inc. Ann Arbor/St.Ives 1 June 11 Summary
More informationIncorporating New, Bright Fluorochromes into Multicolor Panel Design
Incorporating New, Bright Fluorochromes into Multicolor Panel Design Maria C. Jaimes, MD Senior Staff Scientist BD Biosciences 23-14684-00 Overview Multicolor flow: successful application prerequisites
More informationFlow Cytometry For New PhDs
Flow Cytometry For New PhDs 2012 Simon Monard SCRM and CIR smonard@staffmail.ed.ac.uk Derek Davies CRUK derek.davies@cancer.org.uk Monday 20th and Tues 21st Feb 2012 Program Day 1: Day 2: Basics of flow
More informationCyFlow Cube series Appealing from every angle
CyFlow Cube series Appealing from every angle www.sysmex-partec.com CyFlow Cube 6 and Cube 8: compact, economic flow cytometers with a great performance Panta rhei a flexible solution for demands in flow
More informationALL INFECTOLOGY. and now. Gucker, 1947 Wallace Coulter Coulter Orifice, Kamentsky, 1965 Mack Fulwyler sorter BLOOD:
ALL BLOOD: RBC, PLT, WBC ( /-/ ) (+ LDH, carbamide, ESR, periphery blasts, liver function deviation, renal function dev.) Classification: Morphology + genetics+ immunology! B-cell: T-cell: Dedifferentiated:
More informationALL INFECTOLOGY COMMERCIALLY AVAILABLE DEVICES. and now BLOOD:
ALL BLOOD: RBC, PLT, WBC ( /-/ ) (+ LDH, carbamide, ESR, periphery blasts, liver function deviation, renal function dev.) Classification: Morphology + genetics+ immunology! B-cell: T-cell: Dedifferentiated:
More informationdetermine optimum instrument settings for their own instruments and establish their own daily values.
PC7 (770/488) SETUP KIT 6607121 PN 4299504-C FLOW CYTOMETER ALIGNMENT VERIFICATION FLUOROSPHERES FLOW CYTOMETER DETECTOR STANDARDIZATION FLUOROSPHERES INTENDED USE For Research Use Only. Not for use in
More informationBD FACSCanto II. A proven research platform for maximum reliability and the highest quality results
BD FACSCanto II A proven research platform for maximum reliability and the highest quality results A proven platform for maximum reliability and the highest quality results Built on more than 25 years
More informationReview of techniques in imaging and cytometry. Peter Meeus Onze Lieve Vrouw Ziekenhuis, Aalst 7 may 2009, SCK/CEN Mol
Review of techniques in imaging and cytometry Peter Meeus Onze Lieve Vrouw Ziekenhuis, Aalst 7 may 2009, SCK/CEN Mol Cytometry: The counting and measuring of cells,... Mosby's Medical Dictionary, 8th edition.
More informationYour Research, Revolutionized
Your Research, Revolutionized Drive Your Research Forward Your research needs are evolving and with the CytoFLEX flow cytometer you ll see just how far your data can take you. CytoFLEX has the advanced
More informationBD FACSMelody. Cell Sorter. The simple solution for consistent, quality results
BD FACSMelody Cell Sorter The simple solution for consistent, quality results The right instrument is needed for the best results Simplicity, affordability and quality are key Cell sorting is fast becoming
More informationGUAVA EASYCYTE SYSTEMS
GUAVA EASYCYTE SYSTEMS Expanding the potential of flow cytometry The life science business of Merck KGaA, Darmstadt, Germany operates as MilliporeSigma in the U.S. and Canada. Guava easycyte Systems Expanding
More informationDefinitions. What functions does a flow cytometer be able to do? The build of the Flow cytometry and sorting. Flow cytometry
The build of the Flow cytometry and sorting Flow cytometry Seminar Mónika Tóth 11-12-13.04.2011 Definitions Flow cytometry Process or measurement method can measure discrete properties physical, chemical,
More informationCompensation: Fundamental Principles
Flow Cytometry Seminar Series 2017 : Fundamental Principles Spillover correction in multicolor flow cytometry 28.02.2017 http://www.cytometry.uzh.ch Contents Fluorescence and its detection Absorption and
More informationDesigning and executing a successful flow cytometry experiment
Designing and executing a successful flow cytometry experiment 1x1 0 5 Orange Fluorescent Protein 10000 1000 100 10 0 1000 2000 3000 4000 Forward Scatter This presentation is based on true events, however,
More informationIntroduction to Flow Cytometry:
Introduction to Flow Cytometry: A Learning Guide Manual Part Number: 11-11032-01 April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 1-800-448-2347 Introduction to Flow Cytometry: A Learning
More informationNovoCyte Quanteon F. Flow Cytometer. When Exceptional Performance Meets Simplicity
NovoCyte Quanteon F Flow Cytometer When Exceptional Performance Meets Simplicity A Quantum Leap In Benchtop Flow Cytometry The NovoCyte Quanteon The NovoCyte Quanteon flow cytometer builds on its successful
More informationYour Research, Revolutionized
Your Research, Revolutionized Drive Your Research Forward Your research needs are evolving and with the CytoFLEX flow cytometer you ll see just how far your data can take you. CytoFLEX has the advanced
More informationFlow cytometry within reach.
Affordable, but more than capable. The Accuri C6 Flow Cytometer is flow cytometry at its best. Its affordable price makes it accessible beyond compare. The C6 has a foot print small enough to fit most
More informationJ. Philip McCoy, Jr., Ph.D., H.C.L.D. J. Philip McCoy, Jr., PhD
J. Philip McCoy, Jr., Ph.D., H.C.L.D. Email: mccoyjp@mail.nih.gov 1 Introduction 2 How It All Works A very brief description of how a flow cytometer works. Please read it it will help you design your experiments.
More informationGuava easycyte Systems Expanding the potential of flow cytometry
Guava easycyte Systems Expanding the potential of flow cytometry Guava easycyte Systems Flexible Intuitive Affordable Merck Millipore is a business of Unleash what s possible Fifteen years ago, Guava Technologies
More informationApplication Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents
Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents Compensation setup for high content DuraClone reagents INTRODUCTION High content flow cytometry
More informationAttune NxT Flow Cytometer. Efficient. Flexible. Transformative. Flow cytometry instrument
Attune NxT Flow Cytometer Efficient. Flexible. Transformative. Flow cytometry instrument Building a legacy of advancing flow cytometry technology Never settle for average when striving for significant
More informationAttune NxT Flow Cytometer
SPECIFICATION SHEET Attune NxT Flow Cytometer Attune NxT Flow Cytometer On the quest for significant breakthroughs, never settle for average Run samples faster and achieve greater resolution with little
More informationA previous ICCS Module entitled Instrument optimization - Adjusting PMT voltages and compensation 1 should be read as a prerequisite to this module.
Sponsored and reviewed by ICCS Quality and Standards Committee Title: Compensation Tips for Beckman Coulter 10-Color Navios Platform Written by: Salima Janmohamed-Anastasakis Ph.D., Applications Scientist,
More informationPhagocytosis Assay Kit (IgG PE)
Phagocytosis Assay Kit (IgG PE) Item No. 600540 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION
More informationGuava EasyCyte Systems Expanding the potential of flow cytometry
Guava EasyCyte Systems Expanding the potential of flow cytometry Guava easycyte Systems Flexible Intuitive Affordable Unleash what s possible Fifteen years ago, Guava Technologies introduced the first
More informationUnderstanding Flow Cytometry
Understanding Flow Cytometry The Basic Concepts Maree Bagnara Products Sales Specialist/Account Manager Flow Cytometry PN775136 1 Successful Flow Cytometry data is driven by.. Understanding the Biology
More informationNavios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY
Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications with optimized workflows for
More informationIntroduction to Flow Cytometry: A Learning Guide
11-11032-03 rev. A December 2002 Introduction to Flow Cytometry: A Learning Guide Introduction to Flow Cytometry: A Learning Guide Copyright 2002 Becton, Dickinson and Company. All rights reserved. No
More informationThe Worlds First Smart Flow Cytometer
A Flow Revolution. I N T R O D U C I N G The Worlds First Smart Flow Cytometer & Dual Use Flow Cell Cassette. ORFLO has done it again. Flow cytometry has just been revolutionized. By integrating fluorescence
More informationIdentification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer. Application Note
Identification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer Application Note Sylvie Veriac Valérie Perrone Madeleine Avon Abstract Agilent Equipment: 2100 bioanalyzer
More informationSA3800 Spectral Analyzer
SA38 Spectral Analyzer Sony Biotechnology Inc. SA38 Spectral Analyzer The Sony SA38 spectral cell analyzer incorporates advanced electronics, patented optical technologies and many points of automation
More informationCRC Flow Cytometry Core Facility Knut och Alice Wallenbergs Stiftelse
CRC Flow Cytometry Core Facility CRC Flow Cytometry Core Facility Knut och Alice Wallenbergs Stiftelse CRC Flow Cytometry Core facility Instruments: 1. FACSCalibur: analyser, 6 parameters, 4 colors (BD),
More informationApplication Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
More informationCyFlow Space Your flexible flow cytometer
CyFlow Space Your flexible flow cytometer www.sysmex-partec.com CyFlow Space its flexibility gives you the space you need for your work Analysing cells and particles, be it from blood, plasma, tissue,
More informationFACS Core Facility User Group Meeting. 25. April 2018 Janine Bögli
FACS Core Facility User Group Meeting 25. April 2018 Janine Bögli Overview Basis: Forensic Flow Cytometry by Jennifer Wilshire, PhD, Stemcell Technologies You will suggest the crime in each case 8 Practical
More informationCyFlow Space Your flexible flow cytometer
CyFlow Space Your flexible flow cytometer www.sysmex-partec.com CyFlow Space its flexibility gives you the space you need for your work Analysing cells and particles, be it from blood, plasma, tissue,
More informationNavios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY
Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY BECAUSE EVERY EVENT MATTERS The Navios EX flow cytometer offers a solution for advanced cytometry applications optimized for the clinical
More informationseminar
seminar 02.26.-28.2013. Flow cytometry Process or measurement method can measure discrete properties physical, chemical, biochemical, biological parameters of separate particles, e.g. biological cells
More informationBeyond Colorful Science. A Practical Approach to Setting Up Multicolor Panels
Beyond Colorful Science A Practical Approach to Setting Up Multicolor Panels ISCT Paris April 2014 Layout: Multicolor Panel Design Identify the pitfalls Have a look at issues with spectral overlap How
More informationSee what you ve been missing
See what you ve been missing Seeing is Believing, but ImageStream is Proof If you rely on flow cytometry or microscopy, you need the power of the new ImageStream X A breakthrough intersection of technologies
More informationSelected Topics in Electrical Engineering: Flow Cytometry Data Analysis
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Experimental design
More informationExpanding Multicolor Options in Flow Cytometry with Novel Brilliant Violet TM Fluorophores. Carsten Wiethe Scientific Application Manager BioLegend
Expanding Multicolor Options in Flow Cytometry with Novel Brilliant Violet TM Fluorophores Carsten Wiethe Scientific Application Manager BioLegend Seminar Outline BrilliantViolet TM (BV)fluorophores Instrument
More informationHematopoietic Progenitor Cell Product Characterization
Hematopoietic Progenitor Cell Product Characterization Carolyn A. Taylor, Ph.D. Professor of Medicine Director of BMT Program Cell Processing Laboratory Product Testing and Characterization Goals Required
More informationarc lamp is substituted. Before
CE update [cytology hematology generalist] The Principles of Flow Cytometry Antony C. Bakke, PhD From the Department of Pathology, Oregon Health Sciences University, Portland, OR On completion of this
More informationProceedings of Meetings on Acoustics
Proceedings of Meetings on Acoustics Volume 19, 2013 http://acousticalsociety.org/ ICA 2013 Montreal Montreal, Canada 2-7 June 2013 Physical Acoustics Session 1pPAb: Acoustics in Microfluidics and for
More informationPrinciples of Immunophenotyping
Principles of Immunophenotyping % of Cell types? Immune activation? Changes based on health state? Moving from a heterogeneous population of blood cells to identifying the presence and proportion of different
More informationCELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
More informationab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
More informationab Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
More informationSOPVII-7. Panel X: NK-characterization
Created by judith.eckl Page 1 of 8 09/06/2011 SOPVII-7 Panel X: NK-characterization Date: Author: Petra Prinz, Judith Eckl Experimenter: Date: 08/06/2011 Experiment description: Version: 1.0 Start: End:
More informationBeadless Absolute Counting
WHITE PPER Beadless bsolute Counting pplication of the unique properties of the peristaltic pump fluidic based system for volumetric cell counting Pierre Grenot, C.Cy (ISC), Hervé Luche, Ph.D., Centre
More informationBD Multicolor CompBeads
4/2015 23-9955-01 IVD BD Multicolor CompBeads 100 compensation setups Catalog No. 644204 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2015 BD Becton, Dickinson and
More informationBD LSRFortessa. Performance without peer, choice without compromise
BD LSRFortessa Performance without peer, choice without compromise Performance without peer, choice without compromise The BD LSRFortessa cell analyzer offers the ultimate in choice for flow cytometry,
More informationFlow Cytometry Support Reagents
Excite and inspire Flow Cytometry Support Reagents Introduction Miltenyi Biotec is a leading supplier of flow cytometry products, offering one of the broadest ranges of antibodies, kits, assays, and support
More informationCellometer Vision CBA
Features of the Vision CBA Image Cytometry System All-in-One System Basic cell counting, primary cell viability, and cellbased assays. See for Yourself Why the Top Ten Pharmaceutical Companies Trust Cellometer
More informationa Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper DuraClone improves standardization of compensation workflow for high content multicolor applications by flow cytometry Authors: Neha Girish 2, Sudharsan Sathyamurthy
More informationWVU FLOW CYTOMETRY & SINGLE CELL CORE FACILITY
WVU FLOW CYTOMETRY & SINGLE CELL CORE FACILITY Newsletter Volume 4, issue 2 January 2018 Helpful Hints for Proper Gating of Flow Cytometry Data Inside this Issue 1-2 Helpful Hints for Proper Gating of
More informationFlow cytometry analysis of transcription factor expression during differentiation of hpsc-derived cardiomyocytes
APPLICATION NOTE Attune NxT Flow Cytometer Flow cytometry analysis of transcription factor expression during differentiation of hpsc-derived cardiomyocytes Introduction The ability to direct human pluripotent
More informationSA3800 Spectral Analyzer
SA38 Spectral Analyzer Sony Biotechnology Inc. SA38 Spectral Analyzer The Sony SA38 spectral cell analyzer incorporates advanced electronics, patented optical technologies and many points of automation
More informationEdU Flow Cytometry Kit. User Manual
User Manual Ordering information: (for detailed kit content see Table 2) EdU Flow Cytometry Kits for 50 assays: Product number EdU Used fluorescent dye BCK-FC488-50 10 mg 6-FAM Azide BCK-FC555-50 10 mg
More information