PAT940 Smart Titrator

Size: px
Start display at page:

Download "PAT940 Smart Titrator"

Transcription

1 PAT940 Smart Titrator User Manual Edition 1.4 Oct 2015

2

3 Contents Chapter 1 Introduction Instrument Description Instrument Connections Chapter 2 Safety Information General Safety Points Electrical Safety Tubing and Connections Solvent and Chemical Safety Recycling and Disposal Chapter 3 Instrument Overview System Schematic Overview Control POD Front of Control POD Back of Control POD Titrate POD Front of Titrate POD Back of Titrate POD Dispense POD Front of Dispense POD Back of Dispense POD Chapter 4 Instrument Set Up Receiving PAT Packaging Checks Location Connecting the Power Supply Connecting the Dispense POD Connecting a Primary Dispense POD Installation of Syringe Removal of Syringe Connecting the Tubing Connecting the Titrate POD Connecting a Primary Titrate POD Connecting Electrodes and Temperature Probe Connecting an Electrode Connecting an Additional Electrode Connecting a Temperature Probe Connecting Additional Pods Connecting an additional Titrate POD Connecting an additional Dispense POD Connecting External Devices Connecting a USB Device Chapter 5 Basics of Operation Switching ON and OFF

4 5.1.1 Switching PODs ON Switching PODs OFF Capacitive Touch Screen Home Page Navigation Electrode Recognition Chapter 6 System Settings Administrators Configuration Manage Software Update Screenshot Sound Data Manage About Chapter 7 Manual Operations Titrate POD Operations Dispense POD Operations Syringe Initialisation Syringe Flushing Syringe Control Dispense Control Dosing Rate Continuous Dosing Calibration Calibration from Automatic Electrode Recognition Dual Electrode Calibration Calibration Parameters Stirrer Settings Performing a Calibration Calibration Report Calibration History mv/ph Measure Chapter 8 Method Centre The Main Screen Loading a Method Favourites Creating a New Method Saving a Method Editing a Method Deleting a Method Parameters Method Information Pre-Dispense Criteria Stirrer Settings Titrant Addition Rate Control Equivalence Point Evaluation Stop Criteria

5 8.7 Calculations Adding a New Formula Using the Formula Editor Keypad Variable Function Saving a Formula Deleting a Formula Chapter 9 Titrants Selecting a Titrant Adding a New Titrant Deleting a Titrant Chapter 10 Running a Method Performing a Single Determination Performing a Run from Loading a Method Performing a Run from Editing a Method Quick Start from Favourites Entering Sample Information Running a Blank Determination Auto Prime Cancelling a Determination End of a Run Performing a Dual Determination Preparing Determination RUN Indicator icons during screen navigation in a single run Preparing Determination RUN Indicator icons during screen navigation in a dual run Chapter 11 Sample Running Screen Live Display Primary Curve First Derivative Second Derivative Data Table Adjusting Stirrer Aborting a Run Screen Navigation During a Run Real Time Curve Chapter 12 Results Centre Result List Result Information Titration Curves Curve Functions Filter Statistics Selecting a Result Recalculate Deleting Results

6 Chapter 13 Technical Specifications Chapter 14 Parts and Accessories Standalone System Standalone System Accessories and Spare Parts Chapter 15 Conformity and Warranty Declaration of Conformity Quality Control Customer Support and Service Warranty Warranty Exemptions

7 Chapter 1 Introduction 1.1 Instrument Description Standalone PAT940 titration system consists of three PODs; Control POD, Titrate POD and Dispense POD that can be fixed into an organised work tray or arranged by the user in their own work area. The materials of each POD are manufactured from chemically resistant surface treated materials including anodised aluminium side panels to repel aggressive solvents. PAT940 is a modular titration system consisting of PODs for potentiometric titration. PAT940 is designed for ease of use, increased sampling productivity and convenience through its capacitance touch screen colour interface. The following titration modes are available within PAT940: (1) Dynamic Equivalence Point Titration ph (potentiometric ph measurement) mv (potentiometric voltage measurement) (2) Incremental Equivalence Point Titration ph (potentiometric ph measurement) mv (potentiometric voltage measurement) (3) ph Calibration ph (automatic buffer recognition of ph electrodes)

8 1.2 Instrument Connections Various ports are located on each POD, essential for instrument set-up. Below is a list of ports that can be found on each POD. 7 (1) Power Supply Ports The power supply ports are for connecting to a 24V mains power supply between the PODs, using the supplied kycon to kycon cables. (2) Dispense POD Ports These ports are for connecting up to 16 Dispense PODs, in series, using 2-pin LEMO connectors. (3) Titrate POD Ports These ports are for connecting up to two Titrate PODS, in parallel, using 9-pin RS232 Cables. (4) Balance Port This port is for connecting a Balance using a 9-pin RS232 Cable. (5) Electrode Ports These ports are for connecting up to two electrodes, using 2-pin LEMO connectors. (6) Temperature Probe Port This port is for connecting a temperature probe, using a 2-pin LEMO connector. (7) USB Ports These are for connecting a USB flash drive or peripheral devices, up to two of these can be connected at once, unless it is a serial device. (8) Network Port This is for connecting the instrument to a network. The instrument is not provided with a network cable and will need to be sourced elsewhere.

9 Chapter 2 Safety Information General Safety Points GR Scientific products are in conformity with the EU Directives 2004/108/EC and 2006/95/EC. This is based upon the compliance of the products with the harmonised standards: BS EN :2007; BS EN :2010; and BS EN :2013. GR Scientific products are manufactured in accordance to RoHS following the EU Directive 2011/65/EU; it does not contain any of the restricted substances in concentrations and applications banned by this directive. 2.2 Electrical Safety 2.3 Tubing and Connections PAT940 complies with the international standard IEC Only operate this instrument within the mains voltage supply that is specified on the back of the Control POD, an incorrect mains supply can damage the instrument. Never open the housing of the instrument, only GR Scientific personnel who are trained to service the instrument can open the housing. There is risk of electrocution if live components are touched. Always ensure the instrument is disconnected from the mains power supply when connecting or disconnecting cables from the back of the PODs, as electrostatic discharge can build up. Leaks in the tubing are a safety risk if left without attention. Ensure that all tubing is screwed into the Dispense POD tightly by hand, without excessive force. Breaks in the tubing connections or on the tubing can lead to spillages; check the connections regularly to avoid this. 2.4 Solvent and Chemical Safety 2.5 Recycling and Disposal When using flammable or irritant solvents and chemicals, follow suitable safety measures to protect yourself and others around. Always use the instrument in a well ventilated area, clean up spilled fluids immediately, and keep flame sources away from the work area. Always follow the safety instructions of the chemical manufacturer and apply the correct countermeasures that are required. PAT940 is covered by the WEEE (Waste Electronic and Electrical Equipment) Directive 2012/19/EU to prevent the negative and harmful effects on the environment and public health. For more information on how to dispose of the PAT940 safely, please contact your local authority, your local waste disposal

10 company or GR Scientific directly. Chapter 3 Instrument Overview 3.1 System Schematic Overview

11 3.2 Control POD Front of Control POD 1 (1) Display Back of Control POD (1) Titrate POD 1 (9) Electrode 2 (combined ph/ref) (2) Titrate POD 2 (10) Dispense POD (RS485 BUS) (3) Balance (11) Temperature Probe (4) Auto POD (12) Power Output (24V DC 7A Max) (5) USB 1 (13) Power Input (6) USB 2 (14) Fuse 0.2A (7) Network (15) Mains Switch (8) Electrode 1 (combined ph/ref) 11

12 3.3 Titrate POD Front of Titrate POD (1) Titration Vessel Holder (3) Magnetic Stirrer Mechanism (2) Electrode Platform Back of Titrate POD (1) Fuse (4) 24V DC Output (2) I/O Switch (5) RS232 Serial Port (3) 24v DC Input (6) RS485 Serial Bus Port 12

13 3.4 Dispense POD Front of Dispense POD (1) Syringe Cylinder (4) Switching Valve (2) Inlet Hole for Bottle to Syringe Tubing (5) Plunger (3) Outlet Hole for dispensing tube and tip (6) Retaining Screw Back of Dispense POD (1) Fuse 0.2A (5) RS485 Serial Bus Port (2) I/O Switch (6) RS232 Serial Port (3) Configuration Access Plate (7) 24V DC Input (4) RS485 Serial Bus Port (8) 24V DC Output 13

14 Chapter 4 Instrument Set Up 4.1 Receiving PAT Packaging PAT940 is packaged in suitable protective packaging to ensure that no damage occurs during transit, only this packaging is suitable for the safe transportation during transit. All PODs are boxed individually within an overpack. Please keep this packaging as transit of the individual PODs may need to occur in the future as well Checks Once you have received the instrument, immediately check to make sure that there is no damage and no parts are missing. A delivery note will be included within the overpack to use to compare with the shipment. Ensure that inside the shipment is a calibration certificate for each POD, if not please request from the manufacturer Location PAT940 has been developed for indoor use only. Locate the instrument within the work area where it is free from: vibrations direct sunlight contamination from chemicals explosive environments corrosive atmospheres major temperature fluctuations 14

15 4.2 Connecting the Power Supply PAT940 is powered by an external mains source, with a power pack to ensure this connection is the correct voltage of 24V DC for the instrument. Damage may occur beyond repair and the user may be suspect to electric static shocks if the incorrect power is supplied. Note: Refer to Chapter 2 for more safety information. To avoid risk of electrocution, make sure all I/O switches area set to OFF before connecting the power supply (1) Connect the 24V 7A power adaptor from a mains power source to the power input port on the back of the Control POD. (2) Attach one kycon to kycon cable from the power output port on the Control POD to the power input port on the Titrate POD. (3) Attach the other kycon to kycon cable from the power output port on the Titrate POD to the power input port on the Dispense POD. 15

16 4.3 Connecting the Dispense POD Ensure the Dispense POD is sitting the correct way up before connecting any cables into the back of the POD. See image below Connecting a Primary Dispense POD (1) Connect one end of the RS485 to RS485 2-pin 500 mm cable into the RS485 serial bus port on the Control POD. (2) Connect the other end of the cable to one of the RS485 serial bus ports on the Dispense POD. Note: Once connected, this will be automatically detected by the Control POD. Dispense POD 1 is set to default and is ready to use within the software. 16

17 4.3.2 Installation of Syringe Installing a syringe to the Dispense POD is necessary when first receiving PAT940. To install a syringe, follow the procedure below: (1) Remove the syringe from its packaging within the Dispense POD box. (2) Pull the plunger down so that it sits halfway within the syringe cavity. (3) Position the top of the syringe into the electrovalve and screw in firmly in an anticlockwise direction. Note: Keep the syringe straight and do NOT overtighten when installing the syringe. (4) Loosen the retaining screw on the Dispense POD. (5) Pull the plunger down and seat into the syringe plunger holder (6) Tighten the retaining screw. Secure the plunger in place using the retaining screw. 17

18 4.3.3 Removal of Syringe To remove the syringe from the Dispense POD for replacement or cleaning, follow the procedure below: (1) Select [Manual Operation] and [Syringe] (2) Empty the contents of the syringe by using the SYRINGE CONTROL icons; Empty to Vessel or Return to Bottle. (3) Remove the bottle tubing from the titrant bottle and press [FILL] icon to fully extend the plunger. (4) Loosen the retaining screw (5) Push the plunger upwards slightly into the syringe cavity and unscrew the syringe from the electrovalve in a clockwise direction Connecting the Tubing (1) Attach the bottle to syringe tubing (700mm) into the inlet hole securely using the black threaded screw fitting. (2) Feed the other end of the tubing through the DIN45 Bottle Cap into the titrant bottle securely using the black threaded screw fitting. (3) Attach the tubing from syringe to dispensing Tip (600mm) to the outlet hole securely using the black threaded screw fitting. (4) Feed the dispensing tip into the electrode platform attached to the Titrate POD. Note: Refer to Chapter 2.4 for solvent and chemical safety information. Leaks in the tubing are a safety risk. Ensure that all tubing is screwed into the Dispense POD tightly by hand, without excessive force. Breaks in the tubing or connections can lead to spillages. Check the connections and tubing before using to avoid this. 18

19 4.4 Connecting the Titrate POD Ensure the Titrate POD is sitting the correct way up before connecting any cables into the back of the POD. Do not rotate as this may cause unnecessary damage Connecting a Primary Titrate POD (1) Connect one end of the RS232 to RS232 Cable (9 pin d-sub) to Titrate POD 1 serial port on the Control POD. (2) Connect the other end of the RS232 Cable to the RS232 serial port on the Titrate POD. Note: Once connected, this will be automatically detected by the Control POD. 19

20 4.5 Connecting Electrodes and Temperature Probe Connecting an Electrode (1) Connect the electrode LEMO connector into Electrode 1 port on the Control POD. (2) The electrode rests in the electrode platform on the Titrate POD Connecting an Additional Electrode (1) Connect the electrode LEMO connector into Electrode 2 port on the Control POD. (2) The electrode rests in the electrode platform on the Titrate POD. 20

21 4.5.3 Connecting a Temperature Probe (1) Connect the temperature sensor connector of the Pt100 type into the temperature probe port on the back of the Control POD (2) The Pt100 rests in the electrode platform on the Titrate POD. (3) Once connected the temperature will be visible at the top of the screen during a run only. By pressing the thermometer icon you can switch between Celsius and Fahrenheit. If the probe is NOT connected then N.C will be present during a live run. Temperature Probe connected CELCIUS Temperature Probe connected - FARHENHEIT Temperature Probe Not Connected 21

22 4.6 Connecting Additional PODs Connecting an additional Titrate POD Place the Titrate POD 2 next to Dispense POD 1. (1) Connect one end of the RS232 to RS232 Cable (9 pin d-sub) to Titrate POD 2 serial port on the Control POD. (2) Connect the other end of the RS232 Cable to the RS232 serial port on Titrate POD 2 Note: refer to 7.1 Titrate POD Operations for syringe activation. 22

23 4.6.2 Connecting an additional Dispense POD 2 Place the Dispense POD 2 next to Titrate POD 2; refer to picture on page 23. (1) Connect one end of the second RS485 to RS485 2-pin 500 mm cable into Dispense POD 1 and the other end into RS485 serial bus port on Dispense POD 2. Note: the red dot marks the orientation of the pins when connecting to the ports. (2) Attach one kycon to kycon cable into POWER OUT port of Titrate POD 2 and connect the other end into the POWER IN port of Dispense POD 2. Note: refer to 7.2 Dispense POD Operations for syringe initialisation. Very important that Dispense POD 2 is connected as the additional POD and not in Standalone 1 System set up. 23

24 4.7 Connecting External Devices Connecting a USB Device A USB device is any peripheral device that will connect to the instrument using either of these ports; USB1 or USB2. Note: Both of these ports are USB/A type connections. (1) Insert the USB Device into one of the USB Ports. The USB device is automatically recognised. A message will appear when navigating though System Settings; USB Stick Inserted. Note: Refer to Chapter 6 System Settings for instructions on how to use the USB device. 24

25 Chapter 5 Basics of Operation This chapter describes the basic operation of PAT940 which includes: - switching PODs ON - switching PODs OFF - using the capacitive touch screen - homepage navigation - electrode recognition system. 5.1 Powering up PAT940 PODS Switching PODs ON Each individual POD has its own I/O switch and is powered by a serial BUS supply system linking one POD to the next. Note: Refer to Chapter 4.2 for power supply connections. Prior to switching the PODs ON, ensure the I/O switches are switched to OFF and then connect the power supply. Press the I/O switch on the Control POD to the ON position. You will hear a loud beep and the screen will power up with the GR Scientific logo for 40 seconds, then the homepage will load. Press the I/O switch on the Dispense POD to the ON position. You will hear a beep which indicates that the Dispense POD is connected correctly and is ready for initialisation. IMPORTANT: Leave the Titrate POD switched OFF until the Homepage has loaded. Once the home page has loaded (refer to Chapter 5.3 on page 26) then press the I/O switch on the Titrate POD to the ON position. You will hear a quiet beep and the Titrate POD will home itself. Proceed to Chapter 7.2 Dispense POD Operations for syringe initialisation. Note: The syringe must be initialised after powering up the PAT system before any other activity can take place. 25

26 5.1.2 Switching PODs OFF Each POD must be turned OFF within the PAT system when powering down. This can be performed in any order. Simply press the I/O switch to the OFF position on each POD. When powering PAT system back up follow Chapter 5.2 Capacitive Touch Screen The capacitive touch screen is very sensitive, pressing too hard will damage the instrument. Use the touch screen with a fingertip, even while wearing gloves or if you prefer you can use a stylus. Avoid using sharp objects on the screen such as a ballpoint pen even though it is robust, the screen can still be damaged from scratches. 26

27 5.3 HOMEPAGE Navigation The HOMEPAGE provides access to the main operations of the titration system. - Method centre, - System settings, - Manual operations, - Result, - Real-time curve. Favourite methods are displayed in a scroll down table for quick start access. [FLUSH] and [EMPTY] functions for the Dispense POD can be activated on the HOMEPAGE. Flushing cycle should commence prior to any new titration. Note: Refer to Chapter for syringe flushing operation. 5.4 Electrode Recognition When a coloured electrode is plugged in, an electrode shaped icon will appear in top the right corner of the screen to identify which electrode type is connected. The same coloured application icon will enlarge automatically. The methods saved within that coloured icon are now ready to be loaded. If two electrodes are connected both associated coloured icons will enlarge. Therefore methods saved within both of the activated coloured icons can also be loaded. 27

28 Chapter 6 System Settings This chapter describes various configurations and system settings: - Administration - Display - Printer - Configuration - Data Output - Device Manager - Electrodes - Error - About To access system settings, press [SYSTEM SETTINGS] from the Homepage. 6.1 Adminstrators USER SETTINGS is the default page when entering SYSTEM SETTINGS. Adding a User To add a new user, press the NAME field and enter the user name using the automatic pop up keypad. The number of characters in this field is unlimited. Press ENTER [ ] to store the name. Activating a User To activate the user select [ON] from the drop down box. Press [ADD USER] to create the listing in the table. A blue tick will appear in the state column which shows the user is active. Deactivating a User To deactivate a user select one from the table, change the state to OFF and press [SAVE USER]. The blue tick will disappear and the user is therefore deactivated. Editing a User To edit a user select one from the table, amend the NAME as desired using the pop up keypad, press ENTER [ ] and then press [SAVE USER]. Deleting a User To delete a user, select a user from the table, press [DELETE USER] and the name will be removed from the table. Note: The number of users you can add is unlimited but only one user can be deleted at a time. 28

29 6.2 Configuration Follow the procedure below to set the date, time and local time. Once you have amended the configuration settings press the [SAVE] button to save the changes and the current time will then be displayed. 1) Date Format Select the date format from the dropdown box, choose from: DD/MM/YYYY (default setting) MM/DD/YYYY YYYY/MM/DD 2) Date Setting Tap on the date setting box and enter the current date using the pop up calendar. Use the arrow keys to navigate through each month. Select the date by pressing a number. The date selected will appear at the bottom of the calendar. Press [OK] to confirm. 3) Time setting Tap on the time setting field. The automatic keypad will appear. Use the blue cursor to move between the characters. Press key to delete numbers if necessary. Press [ENTER] key, to confirm. 4) Language Press the language field to choose from English, Spanish, Russian or Chinese (simplified). Note: Pressing [SAVE] will save ALL the settings within configuration settings. Make sure you have the correct information in all fields before saving. 29

30 6.3 Manage Press [MANAGE] to perform software updates, activate the screenshot function and control sound Software Update To install the latest software: 1) Upload file onto USB device and insert into USB1 on the back of the Control POD 2) Wait a few seconds until you see a message update available. Install now 3) Press the Update icon 4) Press [OK] and [INSTALL] this may take a few minutes 5) Once complete press [OPEN] and PAT will reload to the homepage Screenshot The screenshot function allows you to save any screen of your choice to a USB device which can be downloaded to a PC. To activate press [OPEN SCREENSHOT] and a circular icon will appear at the top of the screen. To deactivate the screenshot function, press [CLOSE SCREENSHOT] Sound To activate a beep at the start and end of a run determination press the START RUN icon to ON and press the END RUN icon to ON. To switch the sound to OFF, press the ON icon again. 30

31 6.4 Data Manage This function allows you to manage results, methods, favourites, calibration data and calculations within the software. Press a box to activate a file. Choose from importing, exporting or deleting the files. 1) Export Files Choose one or more files from file selection and press [EXPORT] to send to a USB device. All of the following files will be exported when ALL is activated and [Export] is pressed: - Settings - Calibration - Result - Crash - PAT_PDF s - Favourite methods - Methods - Formula Note: Files can only be exported to via a USB device which must be inserted in either USB1 or USB2. 2) Import File Choose one or more files from file selection and press [Import File]. Once successfully imported, a message will appear: File Imported successfully. 3) Export Log This is for manufacturer access only. 4) Delete all files Choose one or more files from file selection and press [DELETE] To clear the Control POD memory of all files listed in export file, activate the box ALL and press [Delete all files] All system settings will return to default. 31

32 6.5 About This section contains contact information about GR Scientific, software version, serial number of the instrument and a QR code which can be scanned via your android/ios phone that links to the GR Scientific website. When a software update is performed through System Settings the latest version will appear in [System Settings] [About]. 32

33 Chapter 7 Manual Operations 7.1 Titrate POD Operations The electrode platform can be activated manually along with stirrer speed adjustment through the Control POD. Initialise the Titrate POD (s) that are connected by pressing the [INITIALISE TitratePOD] button. Activate by pressing the box, a tick will then appear. The electrode platform will home itself at the top of the Titrate POD tower. ELEVATION: Select vessel size by pressing the vessel button choosing either 170ml (standard with all standalone PAT units) or 210ml. Press [ZERO] once and the electrode platform will automatically rise to the home position. Press [DOWN] and the electrode platform will automatically lower to the size of the vessel selected. The position of the electrode platform is indicated in steps in STIRRER CURRENT INFO. STIRRER SPEED: Increase the speed of the stirrer bar clockwise by pressing [>]. Decrease the speed by pressing [<]. Each time the arrow keys are pressed the stirring rate is increased or reduced by one level. The stirrer will automatically start when the stirrer speed is activated by pressing the arrow buttons and the rate is set in steps of -10 to Clockwise Rotation Anti-Clockwise Rotation 33

34 7.2 Dispense POD Operations Syringe Initialisation Before the syringes are operational, they must be initialised. (1) Switch each POD, ON, refer to Chapter 5.1 for Switching ON procedure. (2) A message will appear once the homepage has loaded, see below: Note: Dispense POD 1 is activated as default for PAT Standalone System 1. (3) If a dual PAT system is set up simply select Dispense POD 2 to activate both syringes. (4) Press [OK] and the message below will appear. If [CANCEL] is pressed the Control POD recognises if the syringe has been initialised and will prompt you to do so. 34

35 (5) Insert the platform hole plugs (Part no: ) supplied with each PAT system, into the electrode platform. (6) Insert the dispensing tip into the electrode platform, see image below: (7) Insert the vessel into the Titrate POD (8) Select the correct vessel size. (9) Press [INITIALISE]. The Titrate POD will lower to the vessel and the syringe will empty to vessel and refill for one cycle. The electrode platform will automatically home once the initialisation cycle is over. In a dual PAT System if both Dispense POD s are selected then both syringes will empty to vessel and refill for one cycle. Both electrode platforms will automatically home once the initialisation cycle is over. Note: It is very important to have the plugs fitted into the electrode platform to avoid splashback of any titrant during the initialisation procedure. 35

36 7.2.2 Syringe Flushing The flushing function is used to rinse and fill the syringe and tubings attached to the syringe assembly to remove any air bubbles within the system. There are two areas where Flushing can be performed. From the HOMEPAGE or within [MANUAL OPERATION] [SYRINGE] (1) Press [FLUSH] (2) Insert the platform hole plugs (Part no: ) supplied with each PAT system, into the electrode platform. (6) Insert the dispensing tip into the electrode platform, see image below: (7) Insert the vessel into the Titrate POD (8) Select the correct vessel size. (9) Press [FLUSH]. 36

37 Pressing [CANCEL] will return to homepage. Note: It is recommended that you should FLUSH before the first determination, if necessary more than once. The Titrate POD will lower to the vessel and the syringe will empty to vessel and refill for three cycles. The electrode platform will automatically home once the initialisation cycle is over. In a dual PAT System if both Dispense POD s are selected then both syringes will empty to vessel and refill for three cycles. Both electrode platforms will automatically home once the initialisation cycle is over Syringe Control Activate the syringe or multiple syringes by pressing the number correlating to the number of Dispense POD (s) connected in SYRINGES SELECTION. For PAT940 Standalone System 1, [1] will already be activated as this is a default setting. To empty the syringe press [ ] Empty to Vessel, making sure the dispensing tip is located in a vessel. The Dispense POD will empty the contents of the syringe at maximum rate. Press [>] RETURN TO BOTTLE, the syringe will empty completely back into the titrant bottle. Note: Check the tubing lines have been primed and there is no suspected contamination before returning to bottle. 37

38 Press [ ] FILL to lower the syringe and fill to maximum capacity with titrant. For multiple Dispense PODs connected, select the desired quantity you wish to manually operate by pressing the numbers in [SYRINGES SELECTION] Dispense Control Syringe Volume: To manually dispense titrant select the syringe volume which matches the size of the syringe installed. Use the drop down box to select either 5ml or 12.5ml. Note: Ensure the correct number of syringes has been selected in syringes selection. Dispensed Volume (µl): Press this field to enter the desired volume. Use the automatic keypad to enter the volume and press [DISPENSE] to enter. Range: µl Note: Volume limit is maximum syringe volume, 12.5ml. If the next volume to be dispensed is more than the spare volume remaining, then the syringe will not function. The FILL icon must be pressed and then the next volume can be dispensed. Press [DISPENSE] to activate. Spare volume is automatically calculated during dispensation and is displayed in (µl). Note: For installation and removal of the syringe refer to Chapters and

39 7.2.5 Dosing Rate This function describes the rate at which the syringe doses and fills. Press [DOSING RATE] and the screen shown below will appear: The syringe volume, set in dispense control, dictates the dosing range available. Input Range 5ml syringe 12.5ml syringe ml/min ml/min Enter a value using the automatic pop up keypad within the specified syringe range. Press [OK] to accept or [CANCEL] to return to SYRINGE screen. Once the rate has been set, enter in the desired volume to be dispensed within DISPENSE CONTROL and press [DISPENSE]. Note: the syringe will refill at the same speed as the dosing rate Continuous dosing Press the Continuous dosing box, a blue tick will appear indicating it has been activated. Press again to deactivate. Enter in a maximum dosing volume within the range specified using the automatic pop up keypad. Input Range (ml) Press [OK] or [CANCEL] to return to SYRINGE screen. Enter desired dispensed volume in (µl) within DISPENSE CONTROL if the value has not been entered already. Press [DISPENSE] to perform continuous dosing. Spare volume is automatically calculated during dispensation and is displayed in (µl). 39

40 40 Note: when the volume of one syringe has been dosed, the dosing cylinder will refill automatically.

41 7.3 Calibration Calibration from Automatic Electrode Recognition A calibration is required to match the instrument being used to the current characteristics of the used ph sensor. In practice, the response of a real ph sensor does not exactly follow the Nernst Equation. This difference between the theoretical and actual behaviour of a ph electrode must be therefore compensated for. The Control POD will automatically recognise if there is an electrode connected. When the Control POD is switched ON, if an electrode is connected in port 1, the screen shown below will appear. If an electrode is connected in port 2, the screen shown below will appear: Press [YES] and the calibration screen will load. Press [NO] to return to the homepage. When an electrode is disconnected from either port 1 or port 2 the electrode icon(s) will disappear. 41

42 7.3.2 Dual Electrode Calibration If two electrodes are connected and then the Control POD is powered up, a dual electrode message will appear for calibration of both electrodes. Press [YES] to go to calibration screen or [NO] to return to homepage. In the case of a dual electrode calibration, the calibration of electrode 1 will commence first, a prompt will then appear: INSERT ELECTRODE 1 INTO ph BUFFER and follow calibration procedure as described in Performing a Calibration Once the calibration report is generated and [SAVE] is pressed for Electrode 1, a message will appear to calibrate electrode 2. Press [YES] to proceed with the calibration of electrode 2 and follow procedure as described in Performing a Calibration. The electrode no will automatically change to 2 and INSERT ELECTRODE 2 INTO ph BUFFER will appear on the Calibration screen. Press [NO] to return to calibration screen. 42

43 7.3.3 Calibration Parameters These are the parameters that the user can define which influence the sequence of steps during the calibration. Electrode No: Use the drop down box to select numbers 1 or 2 which is an indication of how many electrodes are required to be calibrated. This number is purely for reporting purposes and does not define which electrode is being calibrated. Note: The number of electrode icons in the top right hand corner of the screen is an indication of how many electrodes are connected. The electrode number will only change automatically during a dual electrode calibration. Glassware Type: Select Vessel size from the dropdown box: 170 or 210ml. Buffers No: Choose up to 5 buffers from the dropdown box that can be used for calibration. If more than two buffers are used for a calibration then they can be used repeatedly. The first two buffers must however always be different from one another. Buffers are automatically detected. mv Range: -29 to +29 Min time (s): The measured value will not be accepted until the minimum waiting time has elapsed. Press this field to enter in a new value using the keypad which automatically pops up. Input Range: s Default Value: 30s Max time (s): The measured value will not be accepted until the maximum waiting time has elapsed. Press this field to enter in a new value using the keypad which automatically pops up. Input Range: s Default Value: 60s Temperature: 43

44 The temperature is being continuously measured when a temperature sensor is connected. Input Range: C Default Value: 25.0 C Drift: Drift is the gradual change in electrode response over a period of time. It is described as the maximum change in mv per minute. This value is automatically calculated after the maximum waiting time has elapsed. The rate and extent of drift can vary depending on which particular electrodes are being used and the age and degree of contamination of the electrodes. Once the extent of drift is known then it should be possible to find the optimum time of use or number of samples analysed before a calibration is necessary. Time (s): This is the min and max waiting times that have already been set and appear once the times have elapsed. A timer also appears once the start button is pressed. Signal (mv): The measured potential value in mv will appear after the min and max waiting times have elapsed Stirrer Settings Refer to Chapter 11.2 Adjust Stirrer for stirrer setting details Performing a Calibration 44

45 To perform a calibration, insert the electrode into an electrode port and follow the procedure below: (1) Set parameters. (2) Rinse electrode with selected buffer and place in electrode platform. (3) Press [START]. (4) A reading will appear in signal row after min time has elapsed. (5) A second reading will appear after max time has elapsed and a drift reading. (6) Repeat the procedure above for the next buffer. (7) Once the calibration is finished a message CALIBRATION OVER will appear. Note: Each time a new electrode is inserted the calibration message will appear on screen as a prompt for you to decide if calibration is necessary. (8) Press [CAL REPORT] to accept and view the results or [CANCEL] to begin the calibration procedure again Calibration Report Detailed information concerning the results from the last calibration performed will appear: 45

46 The Slope of the Graph: Once zero point offset has been determined, the second and further buffer solutions are used to determine the slope of the ph electrode. This slope is expressed as a percentage of the theoretical value (100% = 0.059V per ΔpH = 1 at 25 C). Offset: ph at 0mV This parameter is only visible with ph electrodes. The ph 7 or zero point offset of an electrode is the ph value at which the total output electrode voltage is equal to 0. The zero point of a ph electrode happens at the ph value of 7.00 theoretically. However, in practice, there is almost always a zero point offset. It is very important to run a ph 7 buffer prior to starting a ph measurement to find out the zero point offset. A large zero point offset will introduce a large error in your ph measurement, if you calibrate and measure at different temperatures. Serial Number Entry: Enter the serial number of the electrode in the highlighted fields using the pop up keypad. The serial number is located on the electrode lead. Both fields must be completed otherwise the calibration cannot be saved. Slope Limits: You can define your own limit values based on electrode type. Enter the lower limit by using the pop up keypad and follow the same process for the upper limit. Input Range Lower and Upper Limit: to 999.9% Press [SAVE] to store. 46

47 The default limits set below are based on a non-aqueous electrode type: Default Lower Limit: 88% Upper limit: 120% These values are monitored during the calibration. If these limits are out of range Calibration Fail will appear on the screen. Press the [BACK] button to return to the calibration screen. Saving a Calibration Report: Press [SAVE] once the serial number has been entered. The results are then stored in Calibration History. If you try to exit the screen without pressing the [SAVE] button a message will appear prompting you to either save or discard the result. View mv/ph: Press [mv/ph] button to check the electrode response Calibration History The electrode serial number and colour are listed in order of the most recent one last used, depicted in number order, [ 1 ] being the most recent. Up to 1000 electrodes can be stored in the electrode serial number list, simply scroll up and down to view or select an electrode. Calibration Information: To view the calibration data select an electrode from the list. The slope limits for that calibration are visible. 47

48 The slope curve shows the plot of the date of calibration against the % slope. Therefore the history of the slope for that particular electrode can be monitored over time. Data Table: This table documents the No. of calibrations performed, the date, the slope %, and the OFFSET in mv. 48

49 7.4 mv/ph Measure To perform manual measurements follow the procedure below: (1) Insert solution to be measured in the Titrate POD. (2) If using ELECTRODE 1 make sure the blue tick is visible to activate the Titrate POD and deselect ELECTRODE 2 by pressing the blue tick box. (3) SET ELEVATION of the electrode platform by pressing [LOWER]. (4) Select the vessel size from the dropdown box. (5) Press [DOWN] to lower the electrode platform. (6) Press the [ ] button if necessary to raise the platform in small increments. (7) Press the [ ] button if necessary to lower the platform in small increments. (8) Press [CANCEL] to return to previous screen. The measuring modes displayed show: - ph: potentiometric ph measurement - mv: potentiometric voltage measurement Press [STIRRER ON] to begin stirring. Press [STIRRER OFF] to cease stirring. Note: Stirrer speed cannot be adjusted in VIEW mv/ph mode, a default speed of 3 is pre-set. 49

50 Chapter 8 Method Centre 8.1 The Main Screen The method centre contains 5 coloured icons each with their own application titration type. Red Icon Non-Aqueous Titrations Blue Icon Aqueous Titrations Green Icon Redox Titrations Yellow Icon Precipitation Titrations White Icon Chloride Content Each electrode within the GR Scientific range is also colour coded to match the application area of each method. Method Activation: If you enter the method centre and no electrode is connected then access is denied to operate the main method functions. When an electrode is connected, depending on the coloured cap of that electrode, that matching icon will enlarge and activate Example: A red electrode is connected, the red icon is activated; message below appears: [Press RED icon to LOAD or create a NEW method] 50

51 When two electrodes of different colour are connected both icons will enlarge and become activated: Example: Red and blue electrodes are connected, the red and blue icons are activated, message appears: [Press RED or BLUE icon to LOAD or create a NEW method] Note: Only one icon can be selected at a time to load or create a new method. 51

52 8.2 Loading a Method To load a method, proceed as follows: (1) Press the activated icon you wish to load the method from, the icon will reduce in size, then press [LOAD METHOD] button. The method table with the stored methods will load: (2) Select the desired method. (3) Press [LOAD]. The method is now loaded and the parameters can be viewed once more as a check prior to running. 52

53 8.2.1 Favourites This section describes how you can update a method as a favourite on your homepage to enable quick start function and also edit a favourite. (1) In METHOD CENTRE press [LOAD] and tick the boxes associated with the methods you wish to save to HOMEPAGE. Up to 50 methods can be saved as favourites. (2) Press [Update Favourite] and the methods selected will appear in a scroll down list on the HOMEPAGE. (3) To remove a favourite from the HOMEPAGE untick the box in the methods list in LOAD METHOD. Editing a Favourite: Select [METHOD CENTRE] then [EDIT METHOD] and press the [EDIT FAVOURITE] button. The table shown above lists all of the methods saved as favourites which appear on your homepage. (1) Press the Blue tick box once to deactivate. (2) Press [UPDATE FAVOURITE]. The method will be removed from your favourites list. Press [BACK] button to return to EDIT METHOD screen. 53

54 8.3 Creating a New Method (1) In METHOD CENTRE press the enlarged application icon you wish to create the method in. (2) Press [NEW METHOD]. (3) The method parameters can now be edited individually by pressing each field to activate the pop up keypad or dropdown box. Refer to Chapter 8.6 Parameters (4) In the field ELECTRODE the drop down box only lists the colours based on which electrode is connected. This can be identified by the electrode icons in the top corner of the screen. Note: Only values within the ranges set in white font can be entered. Any values outside that range will automatically revert to the maximum value of that specified range. Adding a Formula: To add a formula, scroll down to the bottom of the parameters list and press [ADD]. A table containing a list of calculations will appear, select one and press [YES] to add. Press [NO] to go back to formula list. 54

55 8.4 Saving a Method To save a method, follow the procedures below: Note: When the [NEW METHOD] button is pressed the next sequential number available is automatically created in method no. field. (1) Press [NEW METHOD] or [EDIT METHOD] depending on whether you are creating or editing a method. (2) Press [SAVE METHOD]; view the parameters as a check. (3) Press [SAVE] to store or press [CANCEL] to return to method parameters. (4) Press [BACK] to return to method centre. In [EDIT METHOD], there is an option to run the method after saving. Press the blue box to activate then press [SAVE] to start the running process. 8.5 Editing a Method (1) In METHOD CENTRE press the enlarged application icon you wish to create the method in. (2) Press [EDIT METHOD] The screen is split up into 3 sections: - Electrodes List Shows the electrode application icons associated with the colour coded electrodes. - Methods List Shows all the methods that have been created within each electrode type, with the most recently created method at the top of the list. - Parameters List Once a method from the methods list has been activated the parameters saved in the method are available to edit. Note: You can edit any method regardless of which coloured electrode is connected. To edit a method follow the procedure below: (1) Press the chosen coloured icon from the electrodes list. (2) Press the method you wish to edit from the methods list. (3) Edit the fields required within the parameters list. 55

56 8.5.1 Deleting a method Deleting a single method: (1) Select a method from the methods list and press [DELETE METHOD]. (2) Press [YES] to delete method and the method will be removed from the list. (3) Press [NO] to return to EDIT METHOD screen. Deleting all methods: (1) Press [DELETE ALL] in EDIT METHOD screen (2) Press [YES] and all methods will be deleted from each coloured icon in the electrodes list. (3) Press [NO] to return to EDIT METHOD screen. 8.6 Parameters The parameters within the instrument give instructions as to how the titrant should be added. Depending on the type of sample and chemical reaction that occurs and the type of electrode used for indication, dynamic or incremental titration are selectable modes Method Information The parameters defined under method information are defined in the following order: Method No: This is an automatically generated number when a new method is created and then populated sequentially. This number cannot be changed. Method Name: Enter up to 42 characters to define your method name using the automatic pop up keypad. Titration Type: Choose one of the following from the drop down box: (1) Non-aqueous associated with the RED icon; Non- aqueous titrations are the titrations in which weakly acidic or basic substances are carried out using non aqueous solvents to achieve a sharp inflection point. (2) Aqueous associated with the BLUE icon; Aqueous titrations are an acid-base neutralization reaction in an aqueous solution to determine the concentration of the acid or base, one of which has a known concentration. (3) Redox associated with the GREEN icon; Redox titrations are based on a reaction between the analyte and titrant. A common example of a redox titration is treating a solution of iodine with a reducing agent and using starch as an indicator. (4) Precipitation associated with the YELLOW icon; Precipitation titrations involve the titrant and analyte reacting to form an insoluble salt. 56

57 (5) Chloride content associated with the WHITE icon; Precipitation titration for determination of Chloride in Crude Oil. Titrant Name: Choose one of the pre-programmed titrants from the drop down box or add a titrant to the list. Note: Refer to Chapter 9 for details on adding a new titrant. Titrant Conc.: Enter the titrant concentration using the numbered automatic pop up keypad. Units are in mol/l. Titer: Enter the titer value manually; up to 13 characters can be entered using the pop up numbered keypad. This value will automatically be used in any calculations involving the titer value. Electrode: Select the electrode colour from the drop down box which only lists the colours based on the type of electrode is connected. This can be identified by the electrode icons in the top corner of the screen. 57

58 Titration Mode: Choose either incremental or dynamic from the drop down box. Syringe Selection: Choose either 5 or 12.5ml from the drop down box, checking which size is installed on the Dispense POD beforehand Pre-Dispense Criteria The parameters below are performed before the start of the titration. Pre-Dispense Criteria: Before starting the titration, titrant addition will occur at maximum rate unless specified until the measured value set in the criteria has been reached. If a volume has been set in pre-dispense volume (µl) and the measured value is reached within this pre-dispense volume, then the titration will automatically begin. Default: OFF Measuring Mode ph: Enter a value using the pop up numbered keypad. Input Range: 0 14 Measuring Mode Potentiometric Voltage, mv: Input Range: to Pre-Dispense Rate: This is the rate at which dosing of the titrant occurs until the pre-dispense criteria is reached. Input Range 5ml syringe: ml/min 12.5ml syringe: ml/min Default: Maximum Maximum dose rate depends on syringe volume: 12.5ml syringe ml/min 5ml syringe 75 ml/min 58

59 8.6.3 Stirrer Settings Set the stirrer parameters for titration as follows: Stirrer Speed: The stirring rate can be set in steps of 10 to Clockwise Rotation - Anti-Clockwise Rotation Initial Stir Time: This is the wait time set, in seconds, before the start of the titration to give the electrode time to settle down. The initial stir time will only begin after the pre-dispensing criteria has been met. Press the blank field using the pop up numbered keypad to enter the desired value. Input Range: (s) Glassware Type: Choose your vessel size either 170ml or 210ml from the drop down box. Note: 170ml is the standard vessel size Titrant Addition The addition of a titrant to a sample solution can be added in two ways. One way is the volume of titrant dispensed is determined by an algorithm built in within the titrator which is free to determine (within certain limits) how much to add dynamic addition. The other way is where fixed volumes of titrant are added for every new measuring point incremental addition. Incremental Equivalence Point Titrations Increment Volume: The titrant is dispensed at fixed volume (ml) increments for each new measuring point. Default: OFF Remove the default by pressing the blue tick box. Press the blank field to enter in a value using the pop up keypad. Input Range: ,000ml 59

60 Wait Between Doses: This is the delay time between doses, in seconds (s) before the next measured value is accepted. Press the blank field to enter a value using the pop up keypad. Input Range: (s) Default: NONE Dose Rate: This is the rate at which the volume increments are dispensed. Enter in a value in ml/min using the pop up numbered keypad once the tick box maximum has been deactivated. Input Range 5ml syringe: 12.5ml syringe: Default: ml/min ml/min Maximum Maximum dose rate depends on syringe volume: 12.5ml syringe ml/min 5ml syringe 75 ml/min Rate Control Dynamic Equivalence Point Titrations Signal Drift ( E): This is a measure of how stable the signal is and the potential difference that should be reached with each increment. The instrument looks for the maximum change of the measured value per minute. A constant measured value can only be reached after a certain amount of time has elapsed. In case the titration does not reach equilibrium, a maximum waiting time is defined by t (max) and in order to avoid too quick addition a minimum waiting time is defined as t (min). Enter a drift value by deactivating the blue tick box and using the pop numbered keypad. Input Range: mv/min 60

61 Drift OFF: Turn the drift function OFF by pressing in the box - OFF, a blue tick will appear in the field. A measured value will only be recorded when the maximum waiting time has elapsed. This is relevant for very slow reacting titrations or less responsive electrodes. Note: A consistent and constant measured value is often only reached after the increment of titrant dispensed has reacted with the sample solution. The electrode may also take time to reach a good response which means reaching an accepted measured value can take longer. It is recommended that the drift function is set in this case. Fixed Time Interval ( t): A fixed time interval tells the titrator at which point the measured value needs to be recorded and the next increment to be dispensed in seconds (s). When a fixed time interval is set, the titrator will wait a set time interval before adding the next increment. This parameter is very common for unstable titrations such as non-aqueous. Enter a fixed time interval by pressing the blank field and entering a number using the numbered keypad. Input Range: 1-60 (s) Default: NONE t (min): An accepted measured value will not be recorded until the minimum waiting time has elapsed regardless if the signal is stable and drift has been reached. The minimum waiting time is important for drift controlled measurements to avoid fluctuations in signal changes when an increment is dispensed. Press the blank field using the pop up numbered keypad to enter the desired value. Input Range: 0-99,999 (s) t (max): An accepted measured value will not be recorded until the maximum waiting time has elapsed. This depends on two factors only: (1) If a stable drift reading has NOT been reached (2) If the signal drift has been set to OFF. Input Range: 0-9,999 (s) 61

62 V (min): This is the smallest volume in (µl) that is dispensed at the beginning of the titration and again at the steepest parts of the titration curve in order to accurately evaluate the exact position of the inflection point. This avoids long titration times but bear in mind the accuracy of the evaluation might not necessarily be increased by using smaller increments. Very small increments can give a measured value in the same region as signal noise. Press the blank field using the pop up keypad to enter the desired value. Input Range: ,500 (µl) V (max): This is the largest volume of titrant in (µl) that is dispensed to prevent the titrator from adding too large increments which could cause the end point to be overshot. Larger increments are dispensed in the flatter parts of the titration curve so fewer measured data points are recorded. A maximum volume should be set if there is an excessive jump in signal or its very abrupt otherwise a large volume will be dispensed in the region of the inflection point. Press the blank field using the pop up numbered keypad to enter the desired value after deactivating blue tick box. Input Range: ,500 (µl) Equivalence Point Evaluation The parameters for identification of the curve are shown below: EQP Selection: This function allows you to filter out which equivalence point you are looking for: 1 st - The first defined equivalence point will be identified Last - The last defined equivalence point will be identified Largest - The equivalence point with the greatest Equivalence Point Recognition criteria and the steepest region of the curve will be identified All - All equivalence points will be identified Press the drop down box to select 1 st, last, largest or all from the list. 62

63 Threshold Value: The threshold value is a minimum number that is set in order to recognise any peak in the 1 st derivative. To set the threshold press the blank field and enter a value using the pop up numbered keypad Input Range: 0-32,000 Note: In order for any peak to be recognised, the threshold set is the minimum size that any peak in the 1 st derivative should be. A good check is to look at the 1 st derivative column in the data table of measured values, then the actual size of the peak can be determined and a suitable threshold can be applied. Bear in mind that the purpose of the threshold is to prevent identification of minor peaks, therefore the level set need only be bigger than the next largest peak. Slope Tendency: This identifies the trend of the curve. If the mv signal rises with every increment dispensed the trend is positive. If the mv signal drops with every increment dispensed ie more negative with dosing or positive to negative, then the trend is negative. Press the drop down box to choose from positive or negative to identify the slope tendency. Set Window: Setting a window allows recognition of the correct equivalence point within a correct range on the curve. This range is given in the unit of measure of the electrode used for the titration. If the equivalence points fall outside of these windows then they will not be identified. Multiple inflection points can be identified within a window. Select the measuring mode in Window 1 Start, press either, mv, ph or OFF. Default: OFF When using set windows the START 1 Value must be less than the END 1 Value this applies to both mv and ph. For example SET ph: Input Range: -20 to +20 START 1 Value: ph 12 END 1 Value: ph 10 For example SET mv: Input Range: to START 1 Value: +600 END 1 Value:

64 8.6.7 Stop Criteria Method Centre New Method / Method Parameters Method Centre Edit Method / Parameters List The conditions defined in this chapter are for ending the titration. Stop V (ml): The titration will end when the STOP volume (ml) specified has been dispensed since the start of the titration. Note: In order to prevent the contents in the vessel from overflowing, the STOP V should be adjusted. Input Range: , ml STOP V will not be activated unless a value is entered. Stop E (mv): The titration will end when the specified measured value has been reached since the beginning of the titration. Input Range: mv Default: OFF To enter a value deactivate the blue tick box and enter a value using the automatic keypad. Stop ph: The titration will end when the specified measured value has been reached since the beginning of the titration. Input Range: 0-14 Default: OFF To enter a value deactivate the blue tick box and enter a value using the automatic numbered keypad. Stop Time: The titration will end when the time specified in seconds (s) has been reached since the beginning of the titration. Input Range: 0-999,999 (s) 64

65 To enter a value deactivate the blue tick box and enter a value using the automatic numbered keypad. Stop EQP: The titration will end when the specified number of equivalence points has been identified. Input Range: 1-9 Default: 1 Select a number between the ranges from the drop down box. Volume after End Point: This is a volume that is dispensed when the number of equivalence points defined under Stop EQP has been found. The shape of the curve after the EQP is also much more visible if this parameter is set. Input Range: , (ml) Default: OFF To enter a value deactivate the blue tick box and enter a value using the automatic keypad. 8.7 Calculations Method Centre / Calculations There are calculations pre-programmed in the table relating to ASTM methods, D664 and D2896. You have the option of using these or creating your own and adding them to a method of your choice Adding a New Formula (1) Choose the result name from the dropdown list. (2) Press the FORMULA field and the formula editor keypad will appear. (3) Delete or amend the existing formula and press [ENTER] to disable the keypad. (4) Select the desired units. (5) Press [ADD FORMULA]. A message will appear asking you if you wish to add this formula. (6) Press [YES] and the formula will be added sequentially. (7) Press [NO] the message is removed and you return to the calculations screen. The formula you originally created will remain in the fields until you amend them. 65

66 There are 3 fields within calculations that are required to be populated to create and save a calculation which are: Result Name: This contains a drop down list of common methods within the petroleum industry. Formulae: The formulae for the calculations are entered using the formula editor keypad which opens when the FORMULA field is pressed. The maximum formula length has unlimited characters. Units: Select the appropriate units from the preprogrammed drop down list. A new formula is added to the calculations table in ascending order Using the Formula Editor Keypad The formula editor keypad contains buttons for entering numbers, mathematical functions and variables. Common variables can be added to the formula by pressing [VFUNC] on the keypad and selected from a list to avoid any spelling mistakes. Note: Refer to Chapter for details on variable function. The functions of the keys in formula editor are shown in the table below: KEY DESCRIPTION / Divide by DEL Delete individual characters CONC Value is automatically added from the parameter: Titrant Conc, within the method the formula is linked to TITER Value is automatically added from the parameter: Titer, within the method the formula is linked to Left scroll though the character Right scroll through the characters W Weight of sample in (g), value automatically added from sample 66

67 run screen VFUNC VEQ Addition of a space between characters Variable functions Volume of titrant required to reach equivalence point Variable Function These are variables that are stored and can be used as part of future calculations. Some common variables include: (1) Determination of a blank value used in the sample calculation. (2) Standardisation of a titrant to give a titer value which is used in the sample calculation. Adding a Variable: (1) Press [VARIABLE FUNCTION]. (2) Press the NAME field and enter a name using the automatic pop up keypad. (3) Enter a value using the automatic pop up keypad. (4) Choose the units from the drop down box. (5) Press [ADD VARIABLE] and a message will appear asking you if you wish to add. (6) Press [YES] to add or [NO] to cancel and return to previous screen. Saving a Variable: Saving a variable overwrites an existing variable that is already in the table. For example, if you have a blank value previously added to the table, you run a new one and wish to update the existing value then follow the procedure below: (1) Press the variable line in the table you wish to overwrite. (2) Amend the existing value in the field using the keypad and/or name if necessary. (3) Press [SAVE VARIABLE] and a message will appear asking if you wish to save it. (4) Press [YES] to accept and the previous variable will be overwritten or press [NO] to cancel. 67

68 Deleting a Variable: (1) Press the white box in the table, a blue tick will appear highlighting which variable you wish to delete. (2) Press more boxes for multiple variables to be deleted. (3) Press [YES] to delete or [NO] to cancel Saving a Formula Saving a formula overwrites an existing formula that has been selected from the formula list. (1) Press a formula line in the table, this will activate. (2) Amend the formula as desired and press [SAVE FORMULA]. (3) A message will appear asking if you are sure you wish to save. (4) Press [YES] to save or [NO] to cancel and return to calculations screen Deleting a Formula (1) Press the formula line in the table you wish to delete. (2) Press [DELETE FORMULA]. A message will appear asking if you wish to delete. (3) Press [YES] and the formula will be permanently removed from the table. Press [NO] to return to the calculations screen. Note: Only one formula can be deleted from the table at a time. 68

69 Chapter 9 Titrants Method Centre / New Method / Titrant Name This chapter explains how you can create a list of titrants used in the system to add to existing titrants already pre-programmed within the software. This list includes: KOH in 2-propanol HCL in 2-propanol HClO 4 in Acetic Acid Sodium Acetate in Acetic Acid Bromide Bromate solution AgNO 3 in 2-propanol Sodium Thiosulphate solution 9.1 Selecting a Titrant In METHOD CENTRE select [NEW METHOD]. Press the field TITRANT NAME and select the titrant from the drop down list by pressing down the name until it becomes highlighted. 9.2 Adding a New Titrant If you wish to use a titrant that is not part of the pre-programmed list you will need to add in a new one. Press the field NEW TITRANT under TITRANT NAME, enter in your titrant by using the automatic pop up keypad and press ENTER [ ]. The number of characters is unlimited. Press [ADD] and the titrant will be added to the titrant name list. Note: New titrants will automatically be listed under titrant name in EDIT METHOD. 69

70 9.3 Deleting a Titrant Only a NEW titrant that has been added to the titrant list can be deleted. You cannot delete any of the pre-programmed titrants from the list. Select a titrant that has been added by pressing the titrant name. Press [DELETE] and the titrant will be removed. 70

71 Chapter 10 Running a Method This chapter explains how to perform a single run determination. Entering sample data is also explained and how it is used in calculations. Prior to any determination both the Titrate POD and Dispense POD must be initialised. Refer to Chapter 7.1 Titrate POD Operations and Chapter 7.2 Dispense POD Operations Performing a Single Determination When performing a determination there are three routes that can be taken to get to the sample run screen and then proceed to START RUN. Performing a run from loading a method Performing a run from editing a method Quick start from favourites Performing a Run from Loading a Method Once you have selected the method you wish to run, view the parameters as a final check. Press [RUN] to enter sample data. Note: Refer to Chapter 10.2 for entering sample information. Press the [RUN] button to start the determination Performing a Run from Editing a Method Note: Refer to Chapter 8.5 for details on editing a method. In edit method choose a method you wish to save by pressing [SAVE] and the method parameters will appear to view. An option to run this method after saving will also appear. Press the box and a tick will appear then press [SAVE]. View the parameters as a final check. Press [RUN] to enter sample data. Note: Refer to Chapter 10.2 for entering sample information. Then press the [RUN] button to start the determination. 71

72 Quick Start from Favourites On HOMEPAGE select the method you wish to RUN from your favourites list by pressing the red arrow icon. This feature takes you straight to Sample Information. Note: Refer to Chapter 10.2 for details on sample information. Press [RUN] to begin the determination Entering Sample Information (1) Begin by preparing the sample for analysis. (2) Weigh the sample in the titration vessel. (3) Add suitable titration solvent as instructed by the method. (4) Place the sample vessel in the Titrate POD. (5) Position the electrode and dispensing tip in the electrode platform. Set Information: Sample ID: Press the blank field and enter in your sample ID using the automatic pop up keypad. User Name: Select an activated user from the dropdown box. Note: Refer to Chapter 6.1 for user settings. Syringe Volume: Check the syringe volume matches the syringe that is installed. If the wrong syringe size is shown in SET INORMATION then this will need to be amended in the parameters in EDIT METHOD. Dispense POD Selection: The choice of which Dispense POD is to be used during the run determination is dependent on which electrode(s) are connected. - Electrode connected in Port 1 Dispense-POD 1 box is automatically activated only and a message [ELECTRODE 1 ACTIVATED] will appear. - Electrode connected in Port 2 Dispense-POD 2 box is automatically activated only and a message [ELECTRODE 2 ACTIVATED] will appear. 72

73 - If both electrode ports are occupied Dispense POD 1 and 2 will be active. Dispense POD 1 is activated as default. Select the correct Dispense POD by pressing the empty box and a green tick will appear once activated Note: when both electrodes are connected Dispense POD 1 can only used with electrode connected in Port 1 and Dispense POD 2 can only be used with an electrode connected in Port 2. For a single determination you cannot activate more than one Dispense POD. Sample Size: Entering a value manually: Press the SAMPLE box, a green tick will appear once activated. Enter in your sample weight by using the automatic pop up numeric keypad. Default Value: Select the appropriate units. Choose from: g, mg, µg, ml, µl Default Unit: grams (g) Press [CANCEL] to return to the previous screen. Adding a blank value to a sample determination: Entering a value manually: Enter a value using the pop up numbered keypad and then select the units from the drop down list. Adding a saved blank value: Press [ADD] and select a blank value from the variable function table. Press [YES] to confirm or [NO] to select another value. Press [CLEAR] to remove the selected blank value. Note: for storing a blank result refer to Running a Blank Determination The blank result will automatically be added to the sample calculation associated with the method used for running the sample Running a Blank Determination In SAMPLE INFORMATON under SAMPLE SIZE, select BLANK by pressing the box. A green tick will appear which activates the blank function. Press [RUN] and the titration will start. Note: The blank result will be saved automatically in the Result Centre after a blank determination. Storing a Blank Result Select [CALCULATIONS] from METHOD CENTRE. Press [VARIABLE FUNCTION]. Enter the name using the pop keypad, then the result the numbered pop up keypad and select the units from the drop down list. The blank result can now be used in sample calculations. 73

74 10.3 Auto Prime This priming function is programmed to flush the titrant tubing and dispensing tip through in full syringe cycles which can be set by the user. Any air bubbles that may be present just before the start of a determination will be removed. Auto-prime can be activated from loading a method or running after saving an edited method. (1) Press [AUTO-PRIME]. (2) Enter your dosing rate, default is set to maximum. (3) Choose the number of flush cycles from the drop down box. Input Range: 0-20 cycles (4) Press [START] to begin or [CANCEL] to return to previous screen Cancelling a Determination A run can be cancelled at any point by pressing [ABORT RUN]. Once this is activated, the run will stop, the electrode platform will rise to the home position and the stirrer will cease. Note: If you are navigating through other screens during a live run press the R1 icon next to the homepage icon in the top right corner of any screen and then press [ABORT RUN] End of a Run Sample information: Date and Time: User Identification: Sample ID: Method No.: Syringe Vol (ml): Vol Dispensed (µl): Starting mv: Run State: Run Time: Exact date and time the run was performed The name of the operator assigned to the determination Batch number or Lot number of the sample being analysed Current method number automatically assigned once saved Volume size of the syringe The total volume of titrant dispensed from the start of the titration Initial reading taken in mv before the first increment is dispensed The status of the run before during and after analysis The time taken in seconds to reach equivalence point Result: EP1 (ml) is displayed as the volume of titrant dispensed to reach equivalence point. 74

75 Result List: This displays all formulae and associated results that have been determined based on the method parameters set. FORMULA 1: Displays the formula used to calculate the result. RESULT 1, 2, etc: Displays the calculated result for the determination. Note: If more than one formula and result has been determined this will be displayed within the result list. Repeat Run: Once the run has finished and the results obtained press [REPEAT RUN]. The repeat run screen will display: Blank Sample If a blank is run is to be repeated the tick box will automatically be activated shown in the screenshot above. The same applies to a sample run. When repeating a blank run press [RUN] to begin the determination. When repeating a sample enter in new sample information including the weight and press [RUN] to begin the determination. Press [CANCEL] to return to RUNNING METHOD SCREEN. Note: The user cannot be changed during a repeat run. Refer to Chapter 6.1 for details on how to change a user. 75

76 10.6 Performing a Dual Determination Prior to any determinations check that the Dispense POD and Titrate PODs have been initialised. Refer to Chapter 7.1 Titrate POD operations and 7.2 Dispense POD Operations. Note: each determination can only be started one at a time. It is important to select which Dispense POD you wish to start your RUN from first when in the sample information screen. Once this determination has begun, sample information for the second Dispense POD can be entered and the determination started Preparing Determination RUN 1 (1) Load the method you wish to run first, following Chapter 10.1 Performing a Single Determination. (2) Enter sample ID and user name as required in sample information screen (3) Activate the desired Dispense POD for determination 1 by pressing the box. A green tick will appear. Note: Dispense POD 1ufrt is activated as default Refer to Chapter 10.2 for full details on Dispense POD selection. (4) When running a sample enter the weight and blank value as required and press [RUN] Indicator icons during screen navigation during a single run A running icon R1 will appear next to the homepage icon whilst navigating through other screens within the software. R1 F1 Run 1 in progress indicator icon Run 1 finished indicator icon When the run has reached completion or has been aborted, a finished icon F1 will appear whilst navigating through other screens within the software. - Press R1 icon to return to live running screen of RUN 1. - Press F1 icon to return to finished RUN 1 screen. Note: during a live a determination run, method parameters cannot be changed for that live run. 76

77 Preparing Determination RUN 2 (1) Whilst determination RUN 1 is in progress, press the Homepage icon and then load the next method following Chapter 10.1 Performing a Single determination. (2) Enter sample ID and User name as required in Sample Information screen. (3) Dispense POD 2 will automatically be selected prior to commencement of the determination. (4) When running a sample, enter the weight and blank value as required and press [RUN] Indicator icons during screen navigation in a dual run When dual determination runs are in progress, screen navigation is still possible. Both running icons R1 and R2 will appear at the top of the screen next to the homepage icon. R1 R2 Homepage icon Run 1 in progress Run 2 in progress Indicator icon indicator icon - Press R1 icon to return to live running screen of RUN 1. - Press R2 icon to return live running screen of RUN 2. As mentioned in Chapter Indicator icons during screen navigation in a single run, when RUN 1 has reached completion or has been aborted an F1 icon will appear during screen navigation. When RUN 2 has reached completion or has been aborted an F2 icon will appear during screen navigation. F2 Run 2 finished indicator icon 77

78 Chapter 11 Live Running Screen 11.1 Live Display The display is split into two sections, the graphical display and sample information. The following functions are available whilst a determination is running: ABORT RUN: Terminate the run at any point during the titration. Note: Refer to Chapter 10.5 for details on cancelling a determination. RESULT CENTRE: View result history and titration curves. Note: Refer to Chapter 13 for details on result history. ADJUST STIRRER: Adjust the stirrer speed at any point during the titration. REPEAT RUN: This function is activated once a run has finished and NOT been aborted. Note: Refer to Chapter 10.5 for details on repeating a run. Sample Information This contains the system status, method that is running and information about the sample being analysed. Once the run has completed, the result will be displayed with EP1 (ml). Note: Refer to Chapter 8.6 for explanation of parameters. If more than one endpoint has been identified then each EP (ml) with the calculated result will be displayed Primary Curve The settings for the primary curve cannot be changed these are set: X Axis Volume (µl) Y Axis Measured Value (mv) When an increment is dispensed, a measuring point is automatically displayed in red, on the curve First Derivative The primary curve is evaluated by plotting the first derivative de/dv versus titrant consumption V. The maximum of the derivative is at the inflection point and indicates the equivalence point. 78

79 Second Derivative The equivalence point is found at the volume corresponding to the x-intercept of the second derivative curve. This point should be coincident with the inflection point (the point at which the curve changes direction) of the original data plot and the peak of the first derivative plot Data Table Press the [Data Table] button and you can view the individual measured values as they are recorded during a live titration and they can also be viewed at the end. These results are also saved in the RESULTS CENTRE Adjusting Stirrer The stirrer speed can be adjusted at any point during a titration. Press [ADJUST STIRRER]. 79

80 Use the arrow keys to increase or decrease the speed clockwise or anticlockwise. The stirring rate can be set in steps of 10 to Clockwise Rotation [>] - Anti-Clockwise Rotation [<] Press the ON icon to begin stirring and press the OFF icon to cease stirring. Press [CANCEL] to return to previous screen Aborting a Run A run can be cancelled at any point by pressing [ABORT RUN]. Once this is activated, the run will stop, the electrode platform will rise to the home position and the stirrer will cease Screen Navigation During a Run During a live RUN, navigation through all sections of the software is possible. You have the capability to create and edit methods, manual operations and calibrate other electrodes when running a titration. A blue R1 icon will appear during a live run when you have exited the running screen. End of a Run When a run has finished a blue F1 icon will appear informing you that the run has finished. Press this icon to return to the running screen. Note: The sample parameters cannot be changed during a live determination, this can only be done when recalculating a result after a run. Refer to Chapter

81 11.5 Real Time Curve The real time curve is a measurement application which records and plots measured values from a run over time and is an indication of the response time of an electrode. X-axis: time (s) Y-axis: potential (mv) The response of the electrode can be monitored and if necessary dosing parameters can be adjusted or the electrode reconditioned in order to improve the titration curve. View the primary curve in full screen mode for dual titrations by pressing [REAL-TIME CURVE] on the homepage. There are three curves available to view when running a sample: Electrode 1 Electrode 2 Temperature Note: If a dual titration is running then both electrode 1 and 2 curves will be active to view in real time Swipe the screens left or right to view. Use the magnifying glass icons to zoom in and out of the curve at any point. Return to the homepage at any point by pressing the HOME icon or press R1 to return to the running screen. 81

82 Chapter 12 Results Centre HOMEPAGE / RESULT The result centre displays a history of all results stored in numerical order with the most recent result at the top in RESULT LIST Result List You can save up to 1000 determinations within the RESULTS LIST located on the left side of the screen. Six results can be displayed on the screen within the results list at a time, use your finger to scroll up and down the results at the speed you desire. Each result within the result list displays the following information: Result number: Method: Date: Result: This is automatically assigned to a result determination that has been generated and proceeds in sequential order. The highest number being the most recent result The method with which the determination was performed The date at which the determination was performed The first calculated result is displayed here. If more results have been calculated these are listed in Sample Information. To view a results press in the result field and once activated the result will highlight orange Result Information Detailed specifications concerning the current determination, including sample information, results, calculations, method parameters and titration curves are displayed within result information. These are displayed in red. 82

83 Sample Information Sample ID: Date: Syringe Vol. (ml): Run time: Method No: User: Time: Starting mv: Run State: Weight (g): The name of the determination this could be the batch or lot no. of the sample. The date at which the determination was performed. The volume size of the syringe used during the determination. The time taken from the start of the analysis to reach the end point determined by STOP conditions. The method number with which the determination was performed. The name of the operator selected from the sample information screen prior to a determination. The time the determination was performed. The initial mv reading taken when the electrode is immersed in the vessel solution. The status of the run situation, either: running, run aborted or run finished. The sample size, entered in (g), (mg), (µg), (ml), (µl) Default is (g) View Method Press the view method button to scroll through the method parameters for that determination. 83

84 Calculate Result Result: The numbering of results, (up to 9) are displayed based on the parameters set within the method and correspond to the order in which they were calculated in the determination run. EP1 (ml): The volume of titrant dispensed to reach reaction completion known as the end point or equivalence point. Formula: The formula added to a method used to calculate the result. Note: Refer to Chapter 8.7 for calculations. 84

85 Titration Curves The titration curves of the current determination are displayed in three forms; the primary curve, first derivative and second derivative. They are made up of individual data points which correspond to each increment of titrant dispensed. Primary Curve This is a symmetric titration curve where the equivalence point is defined as the inflection point of the curve. The equivalence point occurs when the slope of the titration curve is at a maximum. The point of inflection can be found from the primary curve but it is much easier to read the point of maximum slope from a plot of the first derivative of the data. 1 st Derivative Curve The curve is evaluated by plotting 1 st derivative dmv/titrant consumption V (ml). Therefore the maximum of that derivative is the point of inflection and therefore the equivalence point. 85

86 2 nd Derivative Curve The equivalence point is found at the volume corresponding to the x-intercept of the second derivative curve. This point should coincide with the inflection point (the point at which the curve changes direction) of the original data plot and the peak of the first derivative plot Curve Functions Enlarge Press [ENLARGE] to activate. This function allows you to zoom in and out of the graph. To do this simply use two fingers on the screen to enlarge or reduce areas of the curve. Press [RESET] to restore the curve to its original size. Press [CLOSE] to return to RESULT CENTRE screen. 86

87 12.3 Filter The filter function allows you to select the results you want to evaluate within a specific range of parameters which you can set. Press [FILTER] and you can search results through setting the following parameters: COLOUR Choose to filter results within one colour or a choice of colours depending on the electrode and application. Colours: RED BLUE YELLOW GREEN WHITE DATE You may wish to evaluate a set of results within a certain time period. Press from and use the automatic pop up calendar to select the date you wish to filter from. Press to and use the automatic pop up calendar to select the date you wish to filter to. If you select dates that are out of range, and no results are present within that date range then the RESULT LIST will be blank. 87

88 METHOD NAME You may wish to evaluate set of results from a specific method. Select a method from the drop down list to filter. Choose ALL to select all methods. Note: Other methods will automatically update into this list when a determination has finished and the subsequent result is stored in the RESULT CENTRE. Method Name / Starts with Select ALL from the dropdown list and enter in characters that the particular method you wish to filter begins with, using the automatic pop up keypad. Press [OK] to filter or [CANCEL] to return to RESULT CENTRE Statistics RESULT CENTRE / STATISTICS You can statistically evaluate an unlimited number of result determinations from the result list or filter out specific results. The most recent 10 results will be displayed, when scrolling down to the bottom another 10 results will appear and so on. The screen is divided into three sections: (1) RESULT LIST (2) FILTER LIST (3) RESULT STATISTICS Up to 1000 determinations can be stored within the RESULTS LIST. These are located on the left side of the screen. Ten results can be displayed on the screen within the results list at a time. A wait icon will appear for a few seconds then you can scroll through the next set of ten results. 88

89 FILTER The filter function allows you to select a specific range of results. Note: Refer to Chapter 13.3 for more details on Filtering Selecting a Result Activate the results you wish to perform statistics on by pressing the white box within each result in the RESULT LIST. A blue tick will appear once activated. Press [LOAD ITEMS] and the results will appear in the RESULT STATISTICS. The following details are displayed: Result Number The result and units N=x total number of single results from which the mean was calculated Mean value Absolute standard deviation SD Relative standard deviation RSD % To unload result items press the blue tick again. Press [CLEAR ALL] to deactivate all of the selected results. Press [BACK] to return to Result Centre. 89

90 12.5 Recalculate The recalculation function can be used to re-evaluate a particular result for the following reasons: - If the parameters have been altered in a way that changes the recognized equivalence points - If the variable functions have been altered e.g. weight, blank or titer The following parameters can be modified during recalculation of a determination, using the automatic pop up keypad: - Titrant conc. (mol/l) - Threshold Value - Weight of sample (g) - Blank (ml) - Titer To view the changes after modifying a parameter press [APPLY]. Press [SAVE] to store the changes generated in the RESULT CENTRE. Press [CLOSE] to return to RESULT CENTRE. All saved recalculated results will be stored in the result list unless they are deleted. 90

91 12.6 Deleting Results Select the result you wish to delete and press [DELETE RESULT]. Press [YES] to delete or [NO] to return to RESULT CENTRE. To clear the result memory press [DELETE ALL]. Press [YES] to delete all results or [NO] to return to RESULT CENTRE. Note: You cannot select more than one result to delete at a time. 91

92 Chapter 13 Display Resolution: Technical Specifications Capacitance Touch Screen Width 152mm Height 89mm Dispense POD Multiple linking PODs: Up to 16 Dispense PODs can be linked simultaneously Syringe volumes: 12.5 ml and 5 ml Drive Design: Maintenance-free, direct drive stepper motor Stroke Speed: 4.2 secs to 800 mins per stroke (3200 minutes in micro step mode) Step Resolution: 24,000 steps Imprecision (full stroke): 0.1% 50µl to 100 µl 0.05% 250µl to 5ml Inaccuracy (full stroke): 1% syringe volume Measurement Modes: Potentiometry mv Range: Accuracy: Resolution: Amplifier Input Impendence: Equilibrium Equivalence-point Titration Incremental Equivalence-point Titration Manual Titration Calibration with automatic buffer recognition mv 0.1 mv ( ph). 0.1mV > Ω Temperature Measuring Range Pt100: C Resolution Pt100: 0.1 C Measuring accuracy Pt100: ± 0.3 C (with linear calibration) Interfaces USB ports: Titrate POD: Balance: Auto POD: Network Connection: Electrodes: Dispense POD: Ambient Temperature Operating Temp Range: Operating humidity: Storage Temp Range: Storage Humidity: External Power Supply Supply Voltage Frequency Power Consumption Electrode Recognition: Dialog Languages: 2 USB (Type A sockets) for connection of USB devices 2X Titrate POD 9 PIN D-type (female) RS232 1x 9 PIN D-type (female) RS232 1x 9 PIN D-type (female) RS232 LAN RJ-45 connector, Ethernet 2x 6 pin LEMO (male) BUS RS485 Kycon 4 pin for series connections -10 to +45 C 8-90% RH non-condensing -20 to +70 C 5-95% RH no-condensing 110V/220V 50-60Hz 160W Colour coded E-Chem electrodes with auto recognition technology English, Chinese, Russian, Spanish 92

93 Chapter 14 Parts and Accessories 14.1 PAT940 Smart Titrator Standalone System 1 Part no: Control POD with 24VDC 7A PSU Power Pack Comprises of: Control POD VDC 7A PSU Power Pack Titrate POD with RS232 Cable 500mm and Cable Kycon 500mm Comprises of: Titrate POD RS232 to RS232 Cable (9Pin d-sub) 500mm Cable Kycon to Kycon 500mm Dispense POD with one cable Kycon 500mm, RS485 Cable (2pin) 500mm and syringe assembly kit Comprises of: Dispense POD Cable Kycon to Kycon 500mm RS485 to RS485 Cable (2pin) 500mm ml Syringe Assembly Kit ml Syringe Assembly Kit Comprises of: Dispensing Tip (1pk) Bottle/Syringe Tubing (700mm) Tubing from syringe to Dispensing tip (600mm) Syringe 12.5ml Bottle Cap with Drying Tube, desiccant filled Comprises of: Titrant Bottle Cap and Insert Drying Tube with Desiccant ml Standard Titration Vessel PTFE Stirrer 6x20mm (Pack of 5) E-Chem PATrode Red (for non aqueous titrations) 93

94 14.2 PAT940 Smart Titrator Standalone System 2 Part no: Control POD with 24VDC 7A PSU Power Pack Comprises of: Control POD VDC 7A PSU Power Pack x Titrate POD with RS232 Cable 500mm and Cable Kycon 500mm Comprises of: x Titrate POD x RS232 to RS232 Cable (9Pin d-sub) 500mm x Cable Kycon to Kycon 500mm x Dispense POD with one cable Kycon 500mm, RS485 Cable (2pin) 500mm and Syringe Assembly Kit Comprises of: x Dispense POD x Cable Kycon to Kycon 500mm x RS485 to RS485 Cable (2pin) 500mm x 12.5ml Syringe Assembly Kit x 12.5ml Syringe Assembly Kit Comprises of: x Dispensing Tip (1pk) x Bottle/Syringe Tubing (700mm) x Tubing from syringe to Dispensing tip (600mm) x Syringe 12.5ml x Bottle Cap with Drying Tube, desiccant filled Comprises of: x Titrant Bottle Cap and Insert x Drying Tube with Desiccant ml Standard Titration Vessel PTFE Stirrer 6x20mm (Pack of 5) E-Chem PATrode Red (for non aqueous titrations) 94

Water-Water Titration for the MicroCal ITC200

Water-Water Titration for the MicroCal ITC200 Water-Water Titration for the MicroCal ITC200 Goal: Water-Water titrations are a good way to check for possible problems, especially the cleanliness of the ITC and your skills in cell filling. If a water-water

More information

Calcium-EDTA Tutorial for the MicroCal ITC200

Calcium-EDTA Tutorial for the MicroCal ITC200 Calcium-EDTA Tutorial for the MicroCal ITC200 Getting Started 1. The computer and the ITC200 are normally left on. If they have been switched off for any reason makes sure that you boot them in the order

More information

Normalization, Dilution and Transfer Methods for the Maxprep TM Liquid Handler Technical Manual

Normalization, Dilution and Transfer Methods for the Maxprep TM Liquid Handler Technical Manual Normalization, Dilution and Transfer Methods for the Maxprep TM Liquid Handler Technical Manual Instructions for use of Products AS9100, AS9101, AS9200, AS9201 TM530 11/17 Table of Contents 1 Description...2

More information

Instruction Manual HI Portable ph/ec/tds/ C Meters

Instruction Manual HI Portable ph/ec/tds/ C Meters Instruction Manual HI 9811-5 Portable ph/ec/tds/ C Meters www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the meter.

More information

Agilent 1260 Infinity RID User Manual. 9 Maintenance

Agilent 1260 Infinity RID User Manual. 9 Maintenance Agilent 1260 Infinity RID User Manual 9 Maintenance Introduction to Maintenance 130 Warnings and Cautions 131 Detector Maintenance Procedures 133 Cleaning the Module 134 Flow Cell Flushing 135 Correcting

More information

Camera Eyepiece. User s Manual. KS035200G4-Ver1.0

Camera Eyepiece. User s Manual. KS035200G4-Ver1.0 Camera Eyepiece User s Manual Content CHAPTER 1 NOTES AND SAFETY REQUIREMENTS ------------------------- - 1-1.1 Cautions and Notes ------------------------------------------------------------------- -

More information

Industrial flue gas probes. Instruction manual

Industrial flue gas probes. Instruction manual Industrial flue gas probes Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 4 2.3. Protecting the environment...

More information

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+ An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+ LAB ADV COMP 8 From Advanced Chemistry with Vernier, Vernier Software & Technology, 2004 INTRODUCTION A titration, as you recall, is a

More information

HOBO Plug Load Logger (UX ) Manual

HOBO Plug Load Logger (UX ) Manual HOBO Plug Load Logger (UX120-018) Manual The HOBO Plug Load logger is designed to monitor energy consumption of AC-powered plug in loads. This compact device can be used as a power meter with its built-in

More information

TotalTime Plus+ Startup and User Guide. USB Memory Stick Version

TotalTime Plus+ Startup and User Guide. USB Memory Stick Version TotalTime Plus+ Startup and User Guide USB Memory Stick Version 2 Part one Startup Guide Unpack the TotalTimePlus+ terminal, you should find the following: One TotalTimePlus+ terminal, One wall mounting

More information

EGT-AF08 USER GUIDE WHEEL WEIGHING SYSTEMS WITH VARIOUS PLATFORMS

EGT-AF08 USER GUIDE WHEEL WEIGHING SYSTEMS WITH VARIOUS PLATFORMS EGT-AF08 USER GUIDE WHEEL WEIGHING SYSTEMS WITH VARIOUS PLATFORMS Revision 1.00 Last update 01/02/2013 Page intentionally left blank. EGT-AF01 USER GUIDE Table of contents 1 INTRODUCTION... 5 2 TECHNICAL

More information

Instruction Manual Dräger MSI EM200-s

Instruction Manual Dräger MSI EM200-s Dräger MSI GmbH Rohrstraße 32 58093 Hagen Tel.: +49-2331 / 9584-0 Fax: +49-2331 / 9584-29 e-mail: info@draeger-msi.de D 950; Edition 2011-01-01 Content 1. Reference notes Page 3 1.1 Approvals 1.2 Information

More information

Bio-Well GloveManual

Bio-Well GloveManual Bio-Well Glove Manual Ver. 10-2017 Bio-Well GloveManual October 2017 Contents Terms...3 Purpose...4 Important note...4 Technical Information about Bio-Well device...5 Important Remarks and Safety Measures...5

More information

ecosphere water purifier 3in1 USER MANUAL

ecosphere water purifier 3in1 USER MANUAL ecosphere water purifier 3in1 USER MANUAL TABLE OF CONTENTS System Components 3 Function Overview 4 How to Set up Your Purifier 6 3in1 Cartridge Set Up 7 Initial Run 8 Use 8 Purified Water 8 Tap Water

More information

Winzer Corporation 1 Revision: 4.0

Winzer Corporation 1 Revision: 4.0 Table of Content Section 1: Getting Started... 2 1.1 Overview... 2 Section 2: Installation Overview... 3 2.1 Install CD / Start.exe... 3 2.2 Welcome Screen... 4 2.3 Device Selection... 4 2.4 Auto or Industrial...

More information

Version Software User Guide

Version Software User Guide 06 Version 0.0. Software User Guide Long Range LLC. 6 Tannery Street Franklin, NH 05 Table of Contents Introduction Features... Included... Installation 5 Migrating Your Personal Access File... 7 Set Range

More information

Instructions for. Pro AND. Pro. Indoor Air Quality Monitors. Cropped picture here

Instructions for. Pro AND. Pro. Indoor Air Quality Monitors. Cropped picture here Instructions for Pro AND Pro Indoor Air Quality Monitors Cropped picture here 053-244 Rev C July 2002 Quest Service Policy Congratulations! You have purchased one of the finest instruments available,

More information

DATA LOGGER Model DL-SA11

DATA LOGGER Model DL-SA11 DATA LOGGER Model DL-SA11 User s Manual Felix Storch, Inc. An ISO 9001:2008 registered company 770 Garrison Avenue Bronx, New York 10474 www.summitappliance.com R3-081016 1 Product Overview This data logger

More information

Sales Ledger - Version 1.00 SALES LEDGER. Page-1

Sales Ledger - Version 1.00 SALES LEDGER. Page-1 SALES LEDGER Page-1 Document Version History Version 1.00 Description Draft Author Des Cullen Page-2 Table of Contents Introduction... 7 File Menu... 8 New... 8 Invoice... 8 Creating a New Customer...

More information

To whom it may concern, February 21, 2011

To whom it may concern, February 21, 2011 To whom it may concern, February 21, 2011 The following is an evaluation of the GEMS grant that I received to start my own C-MORE lab kits, specifically ocean acidification, plankton, marine debris and

More information

Instruction Manual. HI HI Portable ph/ec/tds/ C Meters

Instruction Manual. HI HI Portable ph/ec/tds/ C Meters Instruction Manual HI 9812-0 HI 9812-5 Portable ph/ec/tds/ C Meters w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully

More information

Experiment 13: Determination of Molecular Weight by Freezing Point Depression

Experiment 13: Determination of Molecular Weight by Freezing Point Depression 1 Experiment 13: Determination of Molecular Weight by Freezing Point Depression Objective: In this experiment, you will determine the molecular weight of a compound by measuring the freezing point of a

More information

Best Selection for Your Business. Electronic Pricing Scale. User s Manual. (Model: TP-31)

Best Selection for Your Business. Electronic Pricing Scale. User s Manual. (Model: TP-31) Best Selection for Your Business Electronic Pricing Scale User s Manual (Model: TP-31) CATALOG 1. Foreword... 2 1.1. Introductions... 2 1.2. Main functions and features... 2 1.3. Specification... 2 1.4.

More information

Manual Rotary Heat Sealers. Type: F108TX. List of content : Introduction 2. General description 3. Application 4. Safety precautions 5

Manual Rotary Heat Sealers. Type: F108TX. List of content : Introduction 2. General description 3. Application 4. Safety precautions 5 Manual Rotary Heat Sealers Type: F108TX List of content : Page: Introduction 2 General description 3 Application 4 Safety precautions 5 Transport and storage 6 Installation 7 First Set-up 8 Connection

More information

8000 / 8100 ph Meters

8000 / 8100 ph Meters 8000 / 8100 ph Meters Operating Instructions GETTING STARTED To begin taking measurements, connect the probe(s) to the socket(s) on the instrument and remove the cap or bottle covering the electrode. Please

More information

NAB EFTPOS User Guide. for Countertop & Mobile Terminals

NAB EFTPOS User Guide. for Countertop & Mobile Terminals NAB EFTPOS User Guide for Countertop & Mobile Terminals YOUR NAB EFTPOS TERMINAL 2 NAB EFTPOS User Guide TABLE OF CONTENTS Getting to know your NAB EFTPOS Ingenico terminal 5 Contactless Tap & Go 8 Sale

More information

nual k Excella E-1 Open Air Shaker N) manual New Brunswick Excella E-1 Open Air Shaker Operating manual

nual k Excella E-1 Open Air Shaker N) manual New Brunswick Excella E-1 Open Air Shaker Operating manual nual k Excella E-1 Open Air Shaker N) manual New Brunswick Excella E-1 Open Air Shaker Operating manual Copyright 2014 Eppendorf AG, Germany. No part of this publication may be reproduced without the prior

More information

Imperial Series. Model IMP-425/525/625/825/1000AP IMD-425/525/625/825/1000AP IMP-1025/1200/1500/2000AP IMD-1025/1200/1500/2000AP

Imperial Series. Model IMP-425/525/625/825/1000AP IMD-425/525/625/825/1000AP IMP-1025/1200/1500/2000AP IMD-1025/1200/1500/2000AP Service Manual Imperial Series Model IMP-425/525/625/825/1000AP IMD-425/525/625/825/1000AP IMP-1025/1200/1500/2000AP IMD-1025/1200/1500/2000AP Contents of Service Manual 1. Safety Precautions ------------------------------------------------------------------

More information

GPS Tour User Manual. Introduction. Getting Started. Step 1) Know your Watch:

GPS Tour User Manual. Introduction. Getting Started. Step 1) Know your Watch: User Manual Introduction Thank you for purchasing the GPS Tour Watch. This GPS Watch is packed with personal Training features like speed, trip time, laps, etc. Watch features include but not limited to

More information

Barkeep Instructions for the Escali SmartConnect Scale

Barkeep Instructions for the Escali SmartConnect Scale Barkeep Instructions for the Escali SmartConnect Scale These instructions are for users of the Escali SmartConnect Kitchen Scale purchased from barkeepapp.com. Users of any other digital scales and/or

More information

Lenntech. sension + Meters & Probes. Electrochemistry. Tel Fax.

Lenntech. sension + Meters & Probes. Electrochemistry. Tel Fax. DATA SHEET sension + Meters & Probes Lenntech info@lenntech.com Tel. +31-152-610-900 www.lenntech.com Fax. +31-152-616-289 Electrochemistry Hach sension+, an all-in-one system with guided menu navigation

More information

442 ph/conductivity Meter

442 ph/conductivity Meter 442 ph/conductivity Meter Instruction Manual This manual contains complete instructions for setting up and using the 442 ph/conductivity meter. Applications information is also available. The information

More information

User Manual QProg Lite

User Manual QProg Lite User Manual QProg Lite Dear valued customer! On behalf of our company we wish to welcome you to the QUORiON family of fine business machines. We sincerely hope you will appreciate the many benefits of

More information

INDEX. 1 Introduction. 2 Software installation. 3 Open the program. 4 General parameters. 5 Tuning

INDEX. 1 Introduction. 2 Software installation. 3 Open the program. 4 General parameters. 5 Tuning SET UP MANUAL INDEX 1 Introduction 2 Software installation 3 Open the program 4 General parameters 5 Tuning 2 1 Introduction Here below the instruction to use the FLYJET software. 1.1 Software features

More information

64C Amplification Block. Instruction Manual

64C Amplification Block. Instruction Manual 64C Amplification Block Instruction Manual Caution: All rights reserved. Quidel Corporation reserves the right to modify this manual at any time without notice. Any part of the manual shall not be duplicated,

More information

User Manual. BOD ANALYST PRO Software for BOD Testing Easy Efficient Accurate. Windows XP Windows Vista Windows 7

User Manual. BOD ANALYST PRO Software for BOD Testing Easy Efficient Accurate. Windows XP Windows Vista Windows 7 User Manual BOD ANALYST PRO Software for BOD Testing Easy Efficient Accurate Windows XP Windows Vista Windows 7 Page Intentionally Left Blank YSI BOD Analyst Pro User Manual ii CONTENTS SECTION 1 INTRODUCTION...

More information

Deltek Touch Time & Expense for Vision. User Guide

Deltek Touch Time & Expense for Vision. User Guide Deltek Touch Time & Expense for Vision User Guide September 2017 While Deltek has attempted to verify that the information in this document is accurate and complete, some typographical or technical errors

More information

Instruction Manual ODY-1765

Instruction Manual ODY-1765 INSPIRING IMAGINATION Instruction Manual ODY-1765 We hope you enjoy your purchase of the Stealth NX-2 Drone and use this Instruction Manual to get your drone to take off! Included Contents 1. Stealth NX-2

More information

ALLDAY TIME SYSTEMS LTD. Allday Time Manager Lite User Guide

ALLDAY TIME SYSTEMS LTD. Allday Time Manager Lite User Guide Allday Time Manager Lite User Guide 1 Table of Contents Table of Contents... 2 Starting Allday Time Manager... 3 Logging In... 3 Adding a New Employee... 4 Viewing / Editing an Employees Record... 5 General

More information

SpeedExtractor E-916 / E-914 Technical data sheet

SpeedExtractor E-916 / E-914 Technical data sheet SpeedExtractor E-916 / E-914 Technical data sheet The SpeedExtractor E-916 / E-914 is an automated instrument for parallel extraction of primarily organic compounds from a variety of solid or semi-solid

More information

INSTRUCTION MANUAL. Voice Command Quadcopter with 2MP Camera

INSTRUCTION MANUAL. Voice Command Quadcopter with 2MP Camera Ver. 4 INSTRUCTI MANUAL Voice Command Quadcopter with 2MP Camera SKY2913 CAUTI Please retain these instructions for future reference. Ensure that the drone is correctly assembled to prevent any injuries

More information

RD01 WHAT S IN THE BOX INSTALLATION MANUAL

RD01 WHAT S IN THE BOX INSTALLATION MANUAL INSTALLATION MANUAL RD0 WHAT S IN THE BOX Rain Director control panel with mode indicators, programming buttons and control valves and 0 V AC to V wall adapter. Control panel measures 80mm W x 70mm H x

More information

Universal Time Clock TM. QuickBooks Plug-in User Guide

Universal Time Clock TM. QuickBooks Plug-in User Guide Universal Time Clock TM QuickBooks Plug-in User Guide Table of Contents Table of Contents...2 Chapter 1: Getting Started...3 Support and Troubleshooting... 3 System Requirements... 3 Installing the QuickBooks

More information

Installation & User Manual

Installation & User Manual Desktop UHF Radio Paging System Installation & User Manual Scope Communications UK Ltd, Quantum House, Steamer Quay, Totnes, Devon TQ9 5AL Tel: +44 1803 860700 Fax: +44 1803 863716 Email: sales@scope-uk.com

More information

Inscale Measurement Technology Ltd Tel: Fax:

Inscale Measurement Technology Ltd Tel: Fax: TCC SERIES Inscale Measurement Technology Ltd 7 Heron Close, St. Leonards-on-Sea. East Sussex TN38 8DX United Kingdom Tel: +44(0)1424 200722 Fax: +44(0)1424 443976 Email: sales@inscale-scales.co.uk Http:www.inscale-scales.co.uk

More information

QUICK Guide. System start-up and shutdown 3. Operating panels. State colors. General process description. Device operating modes

QUICK Guide. System start-up and shutdown 3. Operating panels. State colors. General process description. Device operating modes QUICK Guide System start-up and shutdown 3 Operating panels 4 State colors 5 General process description 6 Device operating modes Recovering from an error situation 15 19 Tips 21 MMS5 rev. 2.0 / FMS Options

More information

BEFORE USING THE COUNTING SCALE

BEFORE USING THE COUNTING SCALE π H-1114, H-1115 H-1116, H-1117 INDUSTRIAL COUNTING SCALES 1-800-295-5510 uline.com WEIGH T UNIT WEIGH T lb/1000 Pc s u 1-800-295-5510 7 8 9 4 5 6 1 2 3 QTY UNIT PST UNIT PST ZERO TOTAL 0. CE TARE TOTAL

More information

PORTABLE ph METER PRICE CODE NUMBER MODEL 100

PORTABLE ph METER PRICE CODE NUMBER MODEL 100 PORTABLE METER MODEL 100 PRICE CODE NUMBER 56-0032 Refer to Bulletin A-301. OPERATION AND SERVICE GUIDE O-670A DEC. 1997 GENERAL INFORMATION Model 100 meter is designed for measuring over the entire 0

More information

WATER QUALITY TESTER. MIC water quality tester standard packing : More MIC water quality combo testers

WATER QUALITY TESTER. MIC water quality tester standard packing : More MIC water quality combo testers WATER QUALITY TESTER 99702 99720 99701 99703 MIC water quality tester standard packing : 99601 ph : ph meter, two calibration solutions, one soaking solution,battery CR2032 and screw driver are installed

More information

4-Beam-Turbidity Meter ITM-4

4-Beam-Turbidity Meter ITM-4 SENSORS FOR FOOD AND BIOPHARMA. Product Information ITM-4 FOOD 4-Beam-Turbidity Meter ITM-4 Application / Specified Usage Turbidity measurement from 0 up to 5000 NTU resp. 0 up to 1250 EBC Filter monitoring

More information

HR Business Partner Guide

HR Business Partner Guide HR Business Partner Guide March 2017 v0.1 Page 1 of 10 Overview This guide is for HR Business Partners. It explains HR functions and common actions HR available to business partners and assumes that the

More information

Scanning Devices Label Counting Table Operations Manual

Scanning Devices Label Counting Table Operations Manual Scanning Devices Label Counting Table Operations Manual This document describes the functions performed by counting table components and suggests procedures for setting up and operating the counting table.

More information

HI EPA Compliant Turbidity Meter HI EPA Compliant Turbidity and Chlorine Meter

HI EPA Compliant Turbidity Meter HI EPA Compliant Turbidity and Chlorine Meter HI 98703 EPA Compliant Turbidity Meter HI 93414 EPA Compliant Turbidity and Chlorine Meter SPECIFICATIONS Turbidity (HI 93414 & HI 98703) Range Range Selection ACCESSORIES HI 710005 Voltage adapter HI

More information

Hazardous Materials Software 2.0

Hazardous Materials Software 2.0 Hazardous Materials Software Hazardous Materials Software 2.0 01970220 Copyright 2004 United Parcel Service of America, Inc. UPS, the UPS brandmark, and the color brown Table of Contents Overview..............................................................

More information

Croner Simplify. Croner Simplify. ~ Manager Training Workbook ~ Version 6.0. P a g e 1

Croner Simplify. Croner Simplify. ~ Manager Training Workbook ~ Version 6.0. P a g e 1 ~ Manager Training Workbook ~ Version 6.0 P a g e 1 Manager View Managers can see information about themselves and their teams. Their role within Croner Simplify is to complete certain tasks for their

More information

HEL Turbidity Probe Manual

HEL Turbidity Probe Manual HEL Turbidity Probe Manual Contents 1. Overview... 3 2. Set up of the reflection probe... 3 3. Calibration... 4 3.1. Calibrating with the analogue electronics... 4 4. Operation... 6 5. Data Analysis...

More information

GIGA Commercial Drone. Owner s Manual. For Owner s Manual updates, warranty information, and support, visit:

GIGA Commercial Drone. Owner s Manual. For Owner s Manual updates, warranty information, and support, visit: GIGA -6000 Commercial Drone Owner s Manual For Owner s Manual updates, warranty information, and support, visit: www.mota.com/giga-6000 Please read the Owner s Manual before your first flight. It has information

More information

USER MANUAL. ExStik CL200A. Waterproof Total Residual Chlorine Tester. Patent Pending. Additional User Manual Translations available at

USER MANUAL. ExStik CL200A. Waterproof Total Residual Chlorine Tester. Patent Pending. Additional User Manual Translations available at USER MANUAL ExStik CL200A Waterproof Total Residual Chlorine Tester Patent Pending Additional User Manual Translations available at www.extech.com Introduction Congratulations on your purchase of the ExStik

More information

SOP: Applikon ez-control Bioreactor Controller Operation

SOP: Applikon ez-control Bioreactor Controller Operation Page 1 of 19 Approvals Preparer: John Buford Reviewer: Jason McMillan Reviewer: Dr. Margaret Bryans Date: 19DEC13 Date: 20DEC13 Date: 21DEC13 1. Purpose 1.1. Cultivate a cell culture using an Applikon

More information

HARBORTOUCH HOSPITALITY USER GUIDE. Harbortouch Technical Support: or

HARBORTOUCH HOSPITALITY USER GUIDE. Harbortouch Technical Support: or HARBORTOUCH HOSPITALITY USER GUIDE INDEX EMPLOYEE FUNCTIONS: Clocking In/Logging In Clocking Out Beginning a Transaction Payment Options Host Mode Customer Database List View Ticket Functions Item Functions

More information

Table of Contents.

Table of Contents. Quick Start Guide Table of Contents Equipment Setup and Break-Down 1-2 Processing a Sale 3 Payment Types 3 Tax Change 3 Scanning Items 4 Price Checks 4 Voids 4 Returns 4 Reprints 4 Gift Certificates 5

More information

Contents OVERVIEW... 3

Contents OVERVIEW... 3 Contents OVERVIEW... 3 Feature Summary... 3 CONFIGURATION... 4 System Requirements... 4 ConnectWise Manage Configuration... 4 Configuration of a ConnectWise Manage Login... 4 Configuration of GL Accounts...

More information

AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd

AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd 1 Content 1. Introduction...3 1.1 Functions...3 1.2 Units of AP5K-C...4 2. The Initial Installation

More information

Manager Dashboard User Manual

Manager Dashboard User Manual Manager Dashboard User Manual Manager User Guide The Manager User Guide is designed to provide a supervisor or a manager with step-by-step instructions for their daily tasks. Although every database will

More information

Instruction Manual Pocket scale

Instruction Manual Pocket scale KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com Tel: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Instruction Manual Pocket scale KERN TAB Version 1.1

More information

Purpose: Enrich ISPs, perform second Qubit assessment, Load Ion Torrent chip.

Purpose: Enrich ISPs, perform second Qubit assessment, Load Ion Torrent chip. Purpose: Enrich ISPs, perform second Qubit assessment, Load Ion Torrent chip. We have 2 fold goals with enrichment: to pull the ISPs that do not have any target attached away from those that do, and then

More information

Lab Exercise #2: Tension Testing (Uniaxial Stress)

Lab Exercise #2: Tension Testing (Uniaxial Stress) Lab Exercise #2: (Uniaxial Stress) Learning Outcomes: 1. Understand the basic concepts of stress and strain 2. Identify the engineering material properties 3. Connect stress and strain through Hooke s

More information

Consignee Guide. Version 1.8

Consignee Guide. Version 1.8 Consignee Guide Version 1.8 TABLE OF CONTENTS 1 Welcome to FlashConsign.com...6 2 Your Consignee Account...7 2.1 Creating a Consignee Account...7 3 Creating a Consignment Sale...10 3.1.1 Consignment Sale

More information

TurboCASH Business Class

TurboCASH Business Class TurboCASH Business Class Guide to Point-of-Sales Philip Copeman The following icons have been placed on specific places throughout the documentation: Keyboard access Press the F1 key when focussed on screens

More information

Quick Reference Guide

Quick Reference Guide Quick Reference Guide Learning and Development Opportunities CWT 2016 Global 20161101 www.cwt-hr-connect.com HR Connect My Learning In this guide Learning at CWT 4 About HR Connect My Learning 6 To Get

More information

3M Ariba Light Account Network Collaboration for Suppliers

3M Ariba Light Account Network Collaboration for Suppliers SAP Training Source To Pay 3M Ariba Light Account Network Collaboration for Suppliers External User Training Version: 1.0 Last Updated: 31-Aug-2017 3M Business Transformation & Information Technology Progress

More information

Completing an Internal Audit User Guide For the Reliance Assessment Database

Completing an Internal Audit User Guide For the Reliance Assessment Database Completing an Internal Audit User Guide For the Reliance Assessment Database Contents Logging into Reliance... 2 Creating an Audit Plan: Planning Your Internal Audit... 4 Approving an Audit Plan... 20

More information

Operator Manual. brite Advanced Retail scale

Operator Manual. brite Advanced Retail scale Operator Manual brite Advanced Retail scale 2 METTLER TOLEDO Operator Manual brite Advanced Order number 30209977 Contents 1 Your new retail scale 5 1.1 Most important functions 5 1.2 Device overview 5

More information

Foreword. Sales Associates Managers

Foreword. Sales Associates Managers Foreword This book is a pocket guide for using the Register module for the InfoTouch Store Manager. It outlines the basic steps and procedures for processing sales, from the beginning to the end of a day.

More information

Tao of Shop Admin User Guide

Tao of Shop Admin User Guide Tao of Shop Admin User Guide Contents Engage... 2 1. Create new post... 2 2. Create new coupon... 6 3. Edit coupon... 9 4. Publishing a Private Post via Customer Import... 11 5. Adding Product Links within

More information

topshelf2.0 QUICK REFERENCE DEVICE MANUAL

topshelf2.0 QUICK REFERENCE DEVICE MANUAL topshelf2.0 QUICK REFERENCE DEVICE MANUAL Scout topshelf2.0 Quick Reference Device Manual January 2007 2007 Scout Inc. Table of Contents 1) Introduction...6 System Requirements Initial Setup Common to

More information

HRO-50 Water Purifier

HRO-50 Water Purifier HRO-50 Water Purifier Product Installation & Operation Manual Honeywell Environmental & Combustion Controls (Tianjin) Co., Ltd. Content Important Notes P2 Product Functions P3-P5 Installation Instructions

More information

Weighloader user manual. Document reference: V2.01

Weighloader user manual. Document reference: V2.01 Weighloader user manual Document reference: 560371-000 V2.01 Note: Prolec Ltd. reserves the right to amend specifications, information and designs without notice. Prolec Ltd. cannot anticipate every possible

More information

U28. Name of Parts 1. Drone. Blade Guard. Battery Cover. Drone Body Blade Motor. Body Light. Camera. Landing Gear. Blade Guard Installation Diagram

U28. Name of Parts 1. Drone. Blade Guard. Battery Cover. Drone Body Blade Motor. Body Light. Camera. Landing Gear. Blade Guard Installation Diagram Name of Parts 1. Drone Blade Guard Drone Body Blade Motor Body Light Camera Landing Gear Blade Guard Installation Diagram U8 TF Card Installation Diagram . Transmitter Power Switch Throttle/ Up and Down

More information

Getting Started With SellerEngine Plus for FBA

Getting Started With SellerEngine Plus for FBA Getting Started With SellerEngine Plus for FBA New SellerEngine Plus users, please visit the Getting Started Guide for a quick start guide to logging in and getting familiar with SellerEngine Plus. If

More information

Operational Manual. Spectrophotometer Model: SP-830 PLUS. Metertech Inc. Version 1.07

Operational Manual. Spectrophotometer Model: SP-830 PLUS. Metertech Inc. Version 1.07 Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc. Version 1.07 Metertech Inc. provides this publication as is without warranty of any kind, either express or implied, including, but

More information

Thermal Value Excellence

Thermal Value Excellence Thermal Value Excellence Excellence Melting Point Systems MP50 MP70 MP90 Automatic Melting Point Determination Simple, Efficient and Video-Recorded Melting Point Determination Melting Point Determination

More information

Instruction Manual. for. gskin U-Value Kit

Instruction Manual. for. gskin U-Value Kit Instruction Manual for gskin U-Value Kit Status: March 1, 2017 gskin U-Value KIT: Instruction Manual 1 / 11 Content 1. The gskin U-Value KIT delivery overview... 1 2. General overview of the gskin U-value

More information

FAST POWER FACTOR REGULATOR. Computer-14df - xx - 144a INSTRUCTION MANUAL ( M / 00A ) (c) CIRCUTOR S.A.

FAST POWER FACTOR REGULATOR. Computer-14df - xx - 144a INSTRUCTION MANUAL ( M / 00A ) (c) CIRCUTOR S.A. FAST POWER FACTOR REGULATOR Computer-14df - xx - 144a INSTRUCTION MANUAL ( M 981 611 / 00A ) (c) CIRCUTOR S.A. -------- POWER FACTOR REGULATOR COMPUTER- 14f --------- Page 2 POWER FACTOR REGULATOR COMPUTER-

More information

Just a few clicks away from a heating system

Just a few clicks away from a heating system Just a few clicks away from a heating system diagnosis. testo 320.The highly efficient flue gas analyzer. testo 320 Efficient measurement with only one instrument. The testo 320 is your reliable partner

More information

Nexsure Training Manual - CRM. Chapter 12

Nexsure Training Manual - CRM. Chapter 12 Nexsure Training Manual - CRM Audits In This Chapter Adding an Audit Updating the form Invoicing the Audit Tracking the Audit Posting the Audit Adding an Audit Audits are processed on policies where an

More information

Instruction Manual. for. gskin U-Value Kit

Instruction Manual. for. gskin U-Value Kit Instruction Manual for gskin U-Value Kit Status: July 12, 2016 gskin U-Value KIT: Instruction Manual 1 / 10 Content 1. The gskin U-Value KIT delivery overview... 1 2. General overview of the gskin U-value

More information

ProODO Optical Dissolved Oxygen Meter for the Lab or Field SPECIFICATIONS

ProODO Optical Dissolved Oxygen Meter for the Lab or Field SPECIFICATIONS ProODO Optical Dissolved Oxygen Meter for the Lab or Field SPECIFICATIONS ProODO Digital Optical Dissolved Oxygen Meter YSI ODO technology reduces the time required to calibrate and maintain your sensors.

More information

ServicePRO + PartsPRO User Guide

ServicePRO + PartsPRO User Guide ServicePRO + PartsPRO User Guide ServicePRO Version 2.0 PartsPRO Version 1.0 Page 1 of 82 1 WHAT IS SERVICEPRO + PARTSPRO?... 4 1.1 What is ServicePRO?... 4 1.2 What are the benefits of using ServicePRO?...

More information

Instruction Manual. gskin U-Value Kit. for. 1 / 7 gskin U-Value KIT: Instruction Manual. Status: 28 th October greenteg AG Technoparkstrasse 1

Instruction Manual. gskin U-Value Kit. for. 1 / 7 gskin U-Value KIT: Instruction Manual. Status: 28 th October greenteg AG Technoparkstrasse 1 1 / 7 gskin U-Value KIT: Instruction Manual Instruction Manual for gskin U-Value Kit Status: 28 th October 2015 2 / 7 gskin U-Value KIT: Instruction Manual Content 1. Overview of gskin U-Value KIT content...

More information

VBS Transport Operators Guide

VBS Transport Operators Guide VBS Transport Operators Guide June 2015 Version 150615 Page 1 OVERVIEW Purpose of the vehicle booking system Napier Port now has an online vehicle booking system (VBS) that transport operators use to book

More information

Operating Instructions. Sniffer Probe

Operating Instructions. Sniffer Probe Operating Instructions Sniffer Probe BG 5268BE/E (0611) LP 503 LP 505 LP 510 Product Identification In all communications with Pfeiffer Vacuum, please specify the information given on the product nameplate.

More information

Opti-Check Handheld Fluorometer

Opti-Check Handheld Fluorometer Opti-Check Handheld Fluorometer November 11, 2011 P/N: 998-0861 Revision A TURNER DESIGNS 1995 N. 1 st Street San Jose, CA 95112 Phone: (408) 749-0994 FAX: (408) 749-0998 Table of Contents 1. Introduction

More information

U846. headless flying quad PRODUCT PARTS. 1. Quadcopter. Clockwise Blade A. Counter-clockwise Blade B. Counter-clockwise Blade B

U846. headless flying quad PRODUCT PARTS. 1. Quadcopter. Clockwise Blade A. Counter-clockwise Blade B. Counter-clockwise Blade B PRODUCT PARTS U846 headless flying quad 1. Quadcopter Clockwise Blade A Counter-clockwise Blade B Counter-clockwise Blade B Clockwise Blade A Clockwise Blade A Counter-clockwise Blade B USB Power 1 Power

More information

Table of Contents.

Table of Contents. Quick Start Guide Table of Contents Equipment Setup and Break-Down 1 2 Processing a Sale 3 Payment Types 3 Tax Change 3 Scanning Items 4 Price Checks 4 Voids/Returns 4 Reprints 4 Gift Certificates 5 All

More information

NAB EFTPOS MOBILE. Terminal Guide

NAB EFTPOS MOBILE. Terminal Guide NAB EFTPOS MOBILE Terminal Guide YOUR NAB EFTPOS MOBILE TERMINAL 2 NAB EFTPOS Mobile Terminal Guide TABLE OF CONTENTS Getting to know your NAB EFTPOS terminal 6 Contactless Tap & Go 8 Understanding your

More information

Quick Guide. CSN950 MultiHead Dimensioner _v2d_MAN_QG_CSN950MH_EN

Quick Guide. CSN950 MultiHead Dimensioner _v2d_MAN_QG_CSN950MH_EN Quick Guide CSN950 MultiHead Dimensioner File: 20160406_v2d_MAN_QG_CSN950MH_EN CSN950 MultiHead Quick Guide Table of Contents 1. Safety Instructions... 2 Safety and General Precautions... 2 2. Specifications

More information

Kronos Workforce Central v8.0 Upgrade Guide for Timecard Approvers

Kronos Workforce Central v8.0 Upgrade Guide for Timecard Approvers Kronos Workforce Central v8.0 Upgrade Guide for Timecard Approvers March 2017 Getting Started... 2 Logging On... 2 Logging Off and Timing Out... 2 Navigating in Workforce Central... 3 Alerts and Notifications...

More information